K Number
K180318
Date Cleared
2018-04-11

(65 days)

Product Code
Regulation Number
892.1200
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PET Digital Gating generates a respiratory signal which can be used to automate motion correction of PET images with respiratory motion. Images that are motion-corrected using PET Digital Gating are intended to aid physicians in: detection; localization; evaluation; diagnosis; staging; monitoring; and/or follow up of disease, abnormality, and or function; therapy planning, monitoring, and guidance; radiotherapy treatment planning; and for Nuclear Medicine interventional procedures. PET Digital Gating maybe used with PET radiopharmaceuticals approved by the regulatory authority in the country of use, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease.

Areas of the body most impacted by respiratory motion are the chest, abdomen, and pelvis. Disease types in which respiratory motion may have a significant impact, if uncorrected, include:

-Lung Cancer (e.g. Small Cell and Non-Small Cell);

-Liver Cancer;

  • -Colorectal Cancer;
  • -Lymphoma (e.g. Hodgkin's and Non-Hodgkin's);
  • -Cancers that have metastasized to the liver;
  • -Cancers in the Thorax; and
  • -Heart Disease
Device Description

The new PET Digital Gating option (aka Data Driven Gating or DDG) is a software-only device created within an update to the existing PET/CT acquisition and processing software. DDG provides the capability to derive a respiratory signal from the acquired PET data as an alternative to existing device-based respiratory gations that are available on the predicate device, GE's Discovery MI (K161574). The respiratory triggers generated by PET Digital Gating are used in a manner identical to those generated by device-based respiratory gating systems such as the reference device, Varian's "RPM Respiratory Gating System" (K102024).

PET Digital Gating provides the analogous respiratory triggers as the device-based systems, without the use of an external respiratory gating device and is an alternative to any existing external device. PET Digital Gating may be used both retrospectively on previously acquired exams, or prospectively where it operates during the acquisition. As is the case for existing device-based respiratory gating systems, PET Digital Gating's triggers may be used with GE's existing motion compensation techniques (gated, Q.Freeze, Q.Static).

Device-based systems rely on external body motion to determine the respiratory waveform. With PET Digital Gating the respiratory motion determined from internal patient anatomical movement.

PET Digital Gating uses an algorithm that incorporates a principal components analysis (PCA) to compute the spatial-temporal variation of PET list data. Principal components analysis is a data processing technique to find a mathematical basis for a dataset where the basis vectors are ordered to explain the maximum variation within the data. The PCA computes the basis vectors (eigenvectors) of the input data variation. The largest principal components are used along with the input data to generate 1-dimensional eigenvectors. These eigenvectors are subsequently used along with the list data to generate respiratory waveforms.

A fast Fourier transforms of the waveforms are used to determine an "R-value" that is used as the characterization metric for the respiratory motion identified in the dataset.

The R value threshold is user configurable. When DDG is used prospectively in conjunction with Q.Static motion correction, the R value threshold can be set for each bed position along with a "base" (no motion) acquisition time and a "Q.Static" (motion detected) acquisition time. Then, for each bed position, PET Digital Gating will evaluate the R value near the end of the base acquisition time, and if it is greater than the preset R value threshold, extend the acquisition time to the Q.Static acquisition time and generate the respiratory triggers to be used. If the evaluated R value is below the preset value, the acquisition completes and the next bed position is acquired.

AI/ML Overview

The GE Medical Systems PET Digital Gating device aims to provide automated motion correction for PET images affected by respiratory motion. The primary study presented to demonstrate substantial equivalence and meet acceptance criteria involves engineering bench testing using a respiratory motion phantom and analysis of representative clinical examples.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in a table format. Instead, it describes the testing performed and the conclusion drawn from it: that the device is "as safe and effective the legally marketed predicate device." The performance reported is that the device successfully generates respiratory triggers and allows for motion correction, analogous to device-based systems, but using internal patient anatomical movement.

Acceptance Criteria (Implied)Reported Device Performance
Generate analogous respiratory triggers to device-based systems.PET Digital Gating provides analogous respiratory triggers as device-based systems, without the use of an external respiratory gating device.
Determine respiratory motion from internal patient anatomical movement.Respiratory motion is determined from internal patient anatomical movement, utilizing a software algorithm (PCA) for evaluating motion of internal anatomy visualized in the PET image data.
Compatibility with existing GE motion compensation techniques (gated, Q.Freeze, Q.Static).PET Digital Gating's triggers may be used with GE's existing motion compensation techniques (gated, Q.Freeze, Q.Static).
Ability to operate retrospectively or prospectively.PET Digital Gating may be used both retrospectively on previously acquired exams or prospectively during acquisition.
No new risks/hazards, warnings, or limitations compared to predicate.PET Digital Gating does not introduce any new risks/hazards, warnings, or limitations.
Successful completion of design control testing and software verification/validation without unexpected results.PET Digital Gating has successfully completed the required design control testing, and software verification and validation showed no unexpected results.
Performance demonstrated through engineering bench testing and corroborated by clinical examples.Engineering bench testing using a respiratory motion phantom demonstrated performance, which was corroborated by quantitative analysis of representative clinical examples.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document mentions "a representative sample of each of the current GE PET/CT scanner platforms" for the engineering bench testing. For the clinical examples, it states "representative clinical examples." Specific numbers for either are not provided.
  • Data Provenance: The engineering bench testing used a "commercially available respiratory motion phantom." The "representative clinical examples" are not specified for their country of origin or whether they were retrospective or prospective, though it does state that "retrospective application of PET Digital Gating was used" for evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications used to establish ground truth for the test set.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set. The validation relies on engineering bench testing with a phantom and comparison to clinical examples, rather than a multi-reader assessment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or reported in this document. The study described focuses on the technical capability of the device to generate motion correction signals, not on direct human reader performance with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance assessment was conducted through the engineering bench testing. This testing evaluated the algorithm's ability to "determine the respiratory waveform" and "generate respiratory triggers" using a phantom, indicating an algorithm-only evaluation of its core functionality. The analysis of "representative clinical examples" also implies a standalone evaluation of the algorithm's output (motion-corrected images) using retrospective data.

7. The Type of Ground Truth Used

  • For Engineering Bench Testing: The ground truth was established by the precisely controlled and known motion of the "commercially available respiratory motion phantom," which is designed to move inserts with varying speed and amplitude linked to "clinically determined, patient-specific respiratory motion waveforms." This is an engineered/simulated ground truth.
  • For Clinical Examples: The type of ground truth used for "representative clinical examples" is not explicitly defined, but the quantitative analysis would likely compare the motion-corrected images generated by the device against a clinical expectation or standard for motion reduction, potentially based on qualitative assessment or SUV/volume measurements from previous methods, though this is not detailed.

8. The Sample Size for the Training Set

The document does not provide a sample size for the training set. It describes the algorithm (PCA) and its function, but not the data used to train it.

9. How the Ground Truth for the Training Set was Established

The document does not describe how the ground truth for the training set was established, as details about a training set are not provided. The algorithm (Principal Component Analysis) is described as a "data processing technique to find a mathematical basis for a dataset," which is typically unsupervised or uses specific characteristics of the input data rather than explicitly labeled ground truth in the traditional sense of classification.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 11, 2018

GE Medical Systems, LLC % Lee Bush Regulatory Affairs Manager 3000 N. Grandview Blvd. WAUKESHA WI 53188

Re: K180318

Trade/Device Name: PET Digital Gating Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: February 2, 2018 Received: February 5, 2018

Dear Lee Bush:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (it known)

K180318

Device Name PET Digital Gating

Indications for Use (Describe)

PET Digital Gating generates a respiratory signal which can be used to automate motion correction of PET images with respiratory motion. Images that are motion-corrected using PET Digital Gating are intended to aid physicians in: detection: localization: evaluation; diagnosis; staging; monitoring; and/or follow up of disease, abnormality, and or function; therapy planning, monitoring, and guidance; radiotherapy treatment planning; and for Nuclear Medicine interventional procedures. PET Digital Gating maybe used with PET radiopharmaceuticals approved by the regulatory authority in the country of use, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease.

Areas of the body most impacted by respiratory motion are the chest, abdomen, and pelvis. Disease types in which respiratory motion may have a significant impact, if uncorrected, include:

-Lung Cancer (e.g. Small Cell and Non-Small Cell);

-Liver Cancer;

  • -Colorectal Cancer;
  • -Lymphoma (e.g. Hodgkin's and Non-Hodgkin's);
  • -Cancers that have metastasized to the liver;
  • -Cancers in the Thorax; and
  • -Heart Disease
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white.

510(k) SUMMARY OF SAFETY AND EFFECTIVNESS

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):

Date:February 2, 2018
Submitter:GE Medical Systems, LLC3000 North Grandview BlvdWaukesha, WI 53188
Primary Contact:Lee BushRegulatory Affairs ManagerPhone 262-309-9429Email: Lee.Bush@ge.com
Secondary Contacts:John JaeckleChief Regulatory Affairs EngineerGE HealthcareTel: 262-424-9547Email: john.jaeckle@ge.com
Device Trade Name:PET Digital Gating
Common/Usual NamePET/CT Imaging System
Classification NamesEmission Computed Tomography System per 21CFR 892.1200
Device ClassificationClass II
Product Code:KPS
Predicate Device(s):GE Discovery MI (K161574)
Reference Devices:Varian RPM Respiratory Gating System (K102024)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are in a light blue color, and the background is white.

Introduction

The new PET Digital Gating option (aka Data Driven Gating or DDG) is a software-only device created within an update to the existing PET/CT acquisition and processing software. DDG provides the capability to derive a respiratory signal from the acquired PET data as an alternative to existing device-based respiratory gations that are available on the predicate device, GE's Discovery MI (K161574). The respiratory triggers generated by PET Digital Gating are used in a manner identical to those generated by device-based respiratory gating systems such as the reference device, Varian's "RPM Respiratory Gating System" (K102024).

Intended Use

PET Digital Gating is intended to be used to detect and characterize respiratory motion using acquired PET coincidence data without the use of an external gating device.

Indications for Use

PET Digital Gating generates a respiratory signal which can be used to automate motion correction of PET images with respiratory motion. Images that are motion-corrected using PET Digital Gating are intended to aid physicians in: detection; evaluation; diagnosis; staging; restaging; monitoring; and/or follow up of disease, abnormality, and/or function; therapy planning, monitoring, and guidance; radiotherapy treatment planning; and for Nuclear Medicine interventional procedures. PET Digital Gating maybe used with PET radiopharmaceuticals approved by the regulatory authority in the country of use, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease.

Areas of the body most impacted by respiratory motion are the chest, abdomen, and pelvis. Disease types in which respiratory motion may have a significant impact, if uncorrected, include:

  • Lung Cancer (e.g. Small Cell and Non-Small Cell);
  • Liver Cancer;
  • Colorectal Cancer;
  • Lymphoma (e.q. Hodqkin's and Non-Hodqkin's);
  • Cancers that have metastasized to the liver;
  • Cancers in the Thorax; and
  • Heart Disease

Background

Due to the long duration of data acquisition required to produce diagnostic quality PET images, respiratory motion has the potential to negatively impact the image quality and quantitation capability. Respiratory motion during PET imaging can degrade lesion localization and quantification (SUV and Volume) as well as cause significant artifacts, leading to reduced image quality. These effects are particularly significant during imaging of the thorax and abdomen where respiratory motion typically has the largest magnitude. The effect of respiratory motion in PET images is a loss of sensitivity in the detection of disease because of the loss in feature conspicuity and reduced quantitation accuracy associated with motion induced blurring.

A moving lesion can be expected to have a larger apparent size (volume), and lower intensity and lower SUVs.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters 'GE' in a stylized, cursive font inside. The circle is surrounded by decorative flourishes, giving it a classic and recognizable appearance. The blue color is consistent throughout the logo.

Device Description and Technological Characteristic

PET Digital Gating provides the analogous respiratory triggers as the device-based systems, without the use of an external respiratory gating device and is an alternative to any existing external device. PET Digital Gating may be used both retrospectively on previously acquired exams, or prospectively where it operates during the acquisition. As is the case for existing device-based respiratory gating systems, PET Digital Gating's triggers may be used with GE's existing motion compensation techniques (gated, Q.Freeze, Q.Static).

Device-based systems rely on external body motion to determine the respiratory waveform. With PET Digital Gating the respiratory motion determined from internal patient anatomical movement.

PET Digital Gating uses an algorithm that incorporates a principal components analysis (PCA) to compute the spatial-temporal variation of PET list data. Principal components analysis is a data processing technique to find a mathematical basis for a dataset where the basis vectors are ordered to explain the maximum variation within the data. The PCA computes the basis vectors (eigenvectors) of the input data variation. The largest principal components are used along with the input data to generate 1-dimensional eigenvectors. These eigenvectors are subsequently used along with the list data to generate respiratory waveforms.

A fast Fourier transforms of the waveforms are used to determine an "R-value" that is used as the characterization metric for the respiratory motion identified in the dataset.

The R value threshold is user configurable. When DDG is used prospectively in conjunction with Q.Static motion correction, the R value threshold can be set for each bed position along with a "base" (no motion) acquisition time and a "Q.Static" (motion detected) acquisition time. Then, for each bed position, PET Digital Gating will evaluate the R value near the end of the base acquisition time, and if it is greater than the preset R value threshold, extend the acquisition time to the Q.Static acquisition time and generate the respiratory triggers to be used. If the evaluated R value is below the preset value, the acquisition completes and the next bed position is acquired.

In addition to not having to set up an external gating system, the benefit of PET Digital Gating to the clinician and the patient is that the clinician is no longer required to "guess" which bed positions would benefit from motion correction during exam setup and then incur the associated longer acquisition time. PET Digital Gating will assess each bed position and only extended scan time for those bed positions that could benefit from motion correction. It will also identify and apply motion correction for bed positions that that were not thought in need of motion correction during exam set up.

The PET Digital Gating software is portable and could also be made available on additional GE PET based systems (e.g. PET/CT, PET/MR, and standalone) or as a remote application on an AW Workstation/Server, cloud application, or other software platform that hosts advanced image processing applications. The primary consideration for the deployment of this feature is that the host platform can receive and process PET coincidence data (List Mode).

{6}------------------------------------------------

Image /page/6/Picture/6 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The color of the logo is a light blue, and the background is white.

Comparisons

The main change in the proposed device for this submission is the addition of the PET Digital Gating option which serves as an alternative to existing device-based respiratory gating options that are available on the predicate device. The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:

Specification/AttributeDiscovery MI (Predicate Device,K161574) including a third-partyrespiratory gating systemPET Digital Gating (ProposedDevice)
Gating hardwareDedicated gating hardwarerequiredNo additional hardware required
Breathing ratesAble to be used with a widerange respiratory ratesencountered clinically.Same
Compatible motioncompensationtechniques able toutilize respiratorytriggersGated, Q.Static, and Q.FreezeSame
Reference protocolsfor respiratorygatingReference protocols provided onthe systemSame
RespiratoryamplitudeDevice-based systems rely onexternal body motion todetermine the respiratorywaveformRespiratory motion determinedfrom internal patient anatomicalmovement
Setup of gatingdeviceUsers are required to interactwith the external gating devicefor setupNo direct user interaction
Triggering operatingprincipleTriggers generated based on anexternal patient movement as"observed" and determined byexternal device hardware andsoftwareTriggers generated utilizingsoftware algorithm (PCA) forevaluating motion of internalanatomy visualized in the PETimage data
PET acquisitiondurationFixed prior to PET examacquisition at time ofprescriptionVariable during PET acquisitionas determined by level ofrespiratory motion detected

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative flourishes or swirls around its perimeter, giving it a classic or vintage appearance. The overall design is simple, recognizable, and associated with the multinational conglomerate.

PET Digital Gating does not introduce any new risks/hazards, warnings, or limitations. The PET Digital Gating software is a "Moderate" level of concern.

Determination of Substantial Equivalence

Engineering bench testing was performed to support substantial equivalence and the product performance claims. The testing uses the commercially available respiratory motion phantom which is designed to move cylindrical inserts (activity filled spheres) in the superior/inferior direction within a body shaped oval both varying speed and amplitude. The motion of the insert is linked to a moving chest wall platform designed to carry a respiratory tracking device. Clinically determined, patient-specific respiratory motion waveforms are used to drive the phantom controller.

The testing was performed using a representative sample of each of the current GE PET/CT scanner platforms.

Representative clinical examples, where retrospective application of PET Digital Gating was used, were quantitatively analyzed to corroborate the engineering testing in support of substantial equivalence.

PET Digital Gating has successfully completed the required design control testing per GE's quality system. No new hazards were identified and no unexpected test results were obtained. PET Digital Gating was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures have been applied to the development of the system:

  • Risk Analysis
  • Required Reviews
  • Design Reviews
  • Software Development Lifecycle
  • Testing on unit level (Module verification)
  • Integration testing (System verification)
  • Performance testing (Verification)
  • Safety testing (Verification)
  • . Simulated use testing (Validation)

The testing and results did not raise different questions of safety and effectiveness than associated with predicate device using device-based respiratory gating.

Conclusion:

The Substantial Equivalence of PET Digital Gating has been demonstrated by:

  • ♦ Review of the proposed labeling included in this 510k and comparison and review of the proposed Indications for Use demonstrate that PET Digital Gating's Indications for Use fall

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are four white flourishes around the circle. The logo is simple and recognizable.

within the intended use of the predicate device and, therefore PET Digital Gating has the same intended use as the predicate device.

  • ♦ The device description and implementation details and the comparison of device characteristics show that PET Digital Gating has some technical characteristics that remain the same, such as the outputted gating signals, but different technological characteristics for how the final signals are determined.
    • o That the different technological characteristics do not raise different question of safety and effectiveness than the predicate and reference devices, and that the device is as safe and effective the legally marketed predicate device as demonstrated by the:
      • submitted scientific information, including references to peer reviewed published information regarding the use of principle components analysis;
      • software verification and validation without unexpected results;
      • development under GE's quality management system, design control activities including risk management;
      • engineering bench testing using established methods and without unexpected results; and
      • supporting clinical examples that corroborate the results of the engineering bench testing.

GE Healthcare believes that the PET Digital Gating is substantially equivalent to the predicate device, Discovery MI, with supporting evidence and comparisons with its reference device, and hence is safe and effective for its intended use.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.