(193 days)
The physlQ Heart Rhythm Module is intended for use by a physician or other qualified medical professionals for the calculation of heart rate and heart rate variability and the detection of atrial fibrillation using ambulatory ECG data. The physlQ Heart Rhythm Module supports receiving and analyzing single-lead ECG signals recorded in a compatible format from FDA-cleared ECG biosensor devices using "wet" electrode technology when assessment of rhythm is desired. The phys\Q Heart Rhythm Module is for use in subacute clinical settings for remote patient monitoring. The physlQ Heart Rhythm Module is not for use in patients requiring or life-sustaining systems or ECG Alam devices.
The physIQ Heart Rhythm Module (Version 1.0) is a computerized all-software callable function library in the Python programming language that is designed for calculating heart rate and heart rate variability and for detecting atrial fibrillation determined by automated analysis of any single electrocardiogram (ECG) channel collected by commercially-available ECG biosensor devices. This Heart Rhythm Module will be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display) that makes calls into the product, most typically via a Python middleware script. The "middleware" accesses the source ECG data from a customer's data collection system, most likely via its own application programming interface (API), and makes calls to the physIQ Heart Rhythm Module to input ECG for processing into the vital sign outputs of the product. These outputs are returned to the middleware, which may insert these results into a downstream monitoring system for clinical use.
Here's an analysis of the acceptance criteria and supporting studies for the physIQ Heart Rhythm Module (Version 1.0), based on the provided FDA 510(k) document:
phyIQ Heart Rhythm Module (Version 1.0) Acceptance Criteria and Performance
1. Table of Acceptance Criteria and Reported Device Performance
The provided document states that "Performance testing following guidelines of ANS/AAMI EC572012: Testing and Reporting Performance Results of Cardiac Rhythm and ST segment Measurement Algorithms has been applied to each of the algorithms. The performance testing results for all algorithms were compared to physIQ's defined acceptance criteria for performance testing. All algorithms met their corresponding acceptance criteria."
However, the specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) for each algorithm (Heartbeat Detector, Heart Rate, Heart Rate Variability, and Atrial Fibrillation Detector) are not explicitly detailed in the provided text. Similarly, the exact reported performance metrics (e.g., the achieved sensitivity/specificity values) are also not provided in a summarized table within this document. The document only confirms that "All algorithms met acceptance criteria."
Therefore, an exact table with numerical acceptance criteria and reported performance cannot be generated from the given text.
2. Sample Size Used for the Test Set and Data Provenance
The document states:
"further supportive clinical validation testing of the physIQ Heart Rhythm Module was performed using electrocardiography (ECG) signals captured from ambulatory patients using a wearable single-lead biosensor device which were annotated by medical experts in cardiology."
- Sample Size for Test Set: Not explicitly stated. The document only refers to "ambulatory patients" without specifying the number of patients or the duration/amount of ECG data.
- Data Provenance: The ECG signals were "captured from ambulatory patients" using two commercially available FDA-cleared patches: HealthPatch (K152139) manufactured by VitalConnect Inc. and BodyGuardian (K121197; K151188) manufactured by Preventice Inc. The country of origin of the data is not specified, but the use of FDA-cleared devices typically implies data collected in regions where such devices are marketed, often the US. The data appears to be retrospective, as it was "captured from ambulatory patients" and then annotated.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not explicitly stated. The document mentions "medical experts in cardiology."
- Qualifications of Experts: They were described as "medical experts in cardiology." Specific experience level (e.g., "10 years of experience") is not provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. The document only says the data was "annotated by medical experts in cardiology." It does not specify if multiple experts independently annotated and then reached consensus, or if a single expert provided the ground truth, or if a specific adjudication process (like 2+1 or 3+1) was used.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not mentioned or described in the provided text. The testing focused on the standalone performance of the algorithm against expert annotations.
- Effect Size of Human Improvement with AI Assistance: Not applicable, as no MRMC study was described.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
- Standalone Performance: Yes, a standalone performance evaluation was conducted. The document states: "this testing did not use any patch-generated vitals, but instead compared physIQ Heart Rhythm Module outputs to annotations by cardiology experts using ECG captured from two commercially-available patches..." This indicates the algorithm's output was directly compared to the expert-derived ground truth without human intervention in the device's output.
7. Type of Ground Truth Used
- Type of Ground Truth: Expert Consensus (or Expert Annotation). The document explicitly states the ECG signals were "annotated by medical experts in cardiology."
8. Sample Size for the Training Set
- Sample Size for Training Set: Not mentioned in the provided text. The document focuses on the performance testing and clinical validation rather than the development and training details of the algorithms.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not mentioned in the provided text. As with the training set size, the document does not delve into the methodology for establishing ground truth for any training data used for the algorithms.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
August 10, 2018
physIQ Inc. George Hides Vice President, Regulatory and Clinical Affairs 300 E. 5th Avenue, Suite 105 Naperville, Illinois 60563
Re: K180234
Trade/Device Name: physIO Heart Rhythm Module (Version 1.0) Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: July 13, 2018 Received: July 17, 2018
Dear George Hides:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Wilhelm
for Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K180234
Device Name
physIO Heart Rhythm Module (Version 1.0)
Indications for Use (Describe)
The physlQ Heart Rhythm Module is intended for use by a physician or other qualified medical professionals for the calculation of heart rate and heart rate variability and the detection of atrial fibrillation using ambulatory ECG data. The physlQ Heart Rhythm Module supports receiving and analyzing single-lead ECG signals recorded in a compatible format from FDA-cleared ECG biosensor devices using "wet" electrode technology when assessment of rhythm is desired. The phys\Q Heart Rhythm Module is for use in subacute clinical settings for remote patient monitoring. The physlQ Heart Rhythm Module is not for use in patients requiring or life-sustaining systems or ECG Alam devices.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
510(k) Notification K180234
GENERAL INFORMATION
Applicant:
physIQlnc. 300 E. 5th Avenue Suite 105 Naperville IL 60563 USA Phone: (800) 561-7902
Contact Person:
George Allen Hides Vice President, Regulatory and Clinical Affairs physIQ Inc. 300 E. 5th Avenue Suite 105 Naperville IL 60563 USA Email: george.hides@physiq.com Phone: (312) 654-1010
Trade/Proprietary Name:
physIQ Heart Rhythm Module (Version 1.0)
Generic/Common Name:
Electrocardiograph
Classification:
Class II, 21 CFR§870.2340 (Electrocardiograph)
Product Code:
DbS
Predicate Device"
Monebo Automated ECG Analysis and Interpretation Software Library, Version 3.0 (K062282) Monebo Technologies Inc.
Indications for Use:
The physIQ Heart Rhythm Module (Version 1.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate variability and the detection of atrial fibrillation using ambulatory ECG data. The physIQ Heart Rhythm Module supports receiving and analyzing single-lead ECG signals recorded in a compatible format from
{4}------------------------------------------------
FDA-cleared ECG biosensor devices using "wet" electrode technology when assessment of rhythm is desired. The physIQ Heart Rhythm Module is for use in subacute clinical and non-clinical settings for remote patient monitoring. The physIQ Heart Rhythm Module is not for use in patients requiring life-supporting or life-sustaining systems or ECG Alarm devices.
Product Description:
The physIQ Heart Rhythm Module (Version 1.0) is a computerized all-software callable function library in the Python programming language that is designed for calculating heart rate and heart rate variability and for detecting atrial fibrillation determined by automated analysis of any single electrocardiogram (ECG) channel collected by commercially-available ECG biosensor devices. This Heart Rhythm Module will be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display) that makes calls into the product, most typically via a Python middleware script. The "middleware" accesses the source ECG data from a customer's data collection system, most likely via its own application programming interface (API), and makes calls to the physIQ Heart Rhythm Module to input ECG for processing into the vital sign outputs of the product. These outputs are returned to the middleware, which may insert these results into a downstream monitoring system for clinical use.
Performance Testing:
physIQ's Heart Rhythm Module (Version 1.0) contains a collection of algorithms intended to be applied to ECG data collected by commercially-available ECG biosensor devices in an ambulatory setting. The collection consists of Heartbeat Detector, and Heart Rate, Heart Rate Variability and Atrial Fibrillation Detector algorithms. Performance testing following guidelines of ANS/AAMI EC572012: Testing and Reporting Performance Results of Cardiac Rhythm and ST segment Measurement Algorithms has been applied to each of the algorithms. The performance testing results for all algorithms were compared to physIQ's defined acceptance criteria for performance testing. All algorithms met their corresponding acceptance criteria.
In addition, further supportive clinical validation testing of the physIQ Heart Rhythm Module was performed using electrocardiography (ECG) signals captured from ambulatory patients using a wearable single-lead biosensor device which were annotated by medical experts in cardiology. This testing did not use any patch-generated vitals, but instead compared physIQ Heart Rhythm Module outputs to annotations by cardiology experts using ECG captured from two commerciallyavailable patches: HealthPatch (K152139) manufactured by VitalConnect Inc. and BodyGuardian (K121197; K151188) manufactured by Preventice Inc. All algorithms met acceptance criteria.
Substantial Equivalence:
The intended use and indications for use for the physIQ Heart Rhythm Module (Version 1.0) are the same as the predicate device, the Monebo Automated ECG Analysis and Interpretation Software Library Version 3.0 (K062282). Patient populations and monitoring environments for the physlQ Heart Rhythm Module are similar to the predicate device. Performance testing on the physIQ Heart Rhythm Module demonstrates comparable performance to the predicate device using international standard ANSI/AAMI EC57:2012. Any differences in the technological characteristics between the physIQ Heart Rhythm Module and the predicate device do not raise any new issues of safety or effectiveness. Therefore, the physlQ Heart Rhythm Module is substantially equivalent to the predicate device.
{5}------------------------------------------------
| DeviceFunctionality | Monebo Automated ECG Analysisand Interpretation SoftwareLibraryVersion 3.0 | physIQ Heart Rhythm Module(Version 1.0) |
|---|---|---|
| Manufacturer | Monebo Technologies Inc. | physIQ Inc. |
| 510(k) Number | K062282 | K180234 |
| Classification | Class II, 21 CFR §870.2340 | Class II, 21 CFR §870.2340 |
| Product Code | DPS | DPS |
| Indications forUse | The Automatic Analysis andInterpretation Software Library isintended for use by qualified medicalprofessionals for the assessment ofarrhythmias using historicambulatory ECG data. The productsupports downloading and analyzingdata recorded incompatible formatsfrom any device used for thearrhythmia diagnostics such asHolter, Event Monitor, 12 leadambulatory or resting ECG devices,or other similar devices whenassessment of the rhythm isnecessary. The Automatic Analysisand Interpretation Software Librarycan also be electronically interfaced,and perform analysis with datatransferred from other computerbased ECG systems, such as an ECGmanagement system. The AutomaticAnalysis and Interpretation SoftwareLibrary provides ECG signalprocessing and analysis on a beat bybeat basis, QRS and VentricularEctopic Beat detection, QRS featureextraction, interval measurement,heart rate measurement, and rhythmanalysis for up to sixteen (16) leadsof captured data. The library is notfor use in life supporting orsustaining systems or ECGmonitoring and Alarm devices. Theproduct can be integrated intocomputerized ECG monitoringdevices. In this case the medicaldevice manufacturer will identify theindication for use depending on theapplication of their device. | The physIQ Heart Rhythm Moduleis intended for use by a physician orother qualified medicalprofessionals for the calculation ofheart rate and heart rate variabilityand the detection of atrialfibrillation using ambulatory ECGdata. The physIQ Heart RhythmModule supports receiving andanalyzing single-lead ECG signalsrecorded in a compatible formatfrom FDA-cleared ECG biosensordevices using "wet" electrodetechnology when assessment ofrhythm is desired. The physIQHeart Rhythm Module is for use insubacute clinical and non-clinicalsettings for remote patientmonitoring. The physIQ HeartRhythm Module is not for use inpatients requiring life-supporting orlife-sustaining systems or ECGAlarm devices. |
| Level ofConcern | Moderate | Moderate |
| Components | Software only | Software only |
| DeviceFunctionality | Monebo Automated ECG Analysisand Interpretation SoftwareLibraryVersion 3.0 | physIQ Heart Rhythm Module(Version 1.0) |
| Interface | Callable application programminginterface (API) | Callable application programminginterface (API) |
| Display | No primary display | No primary display |
| QRS detection | YES | YES |
| Heart rate non-paced adult | YES | YES |
| Heart ratevariability | YES (deterministic based on R-to-Rinterval derived from QRS detectionwhich may be analyzed usingspectral and time-domainapproaches) | YES (deterministic based on R-to-Rinterval derived from QRSdetection which is analyzed using atime-domain approach) |
| Atrialfibrillationdetection | YES (classification) | YES (classification) |
| ECGmorphologicalanalysis | YES (including PR and QT intervals,and QRS duration | NO (other than QRS location andbeat-to-beat analyses, no ECGmorphological analyses areperformed) |
| Arrhythmiaclassifications(other thanatrialfibrillation) | YES (normal sinus rhythm,ventricular tachycardia, bradycardia,tachycardia, AV block, BBB,ventricular bigeminy, ventriculartrigeminy, AIVR, pause, ventricularflutter) | NO |
| Patientpopulations | Adult | Adult |
| Clinical setting | Subacute (non-life-supporting or life-threatening systems required) | Subacute (non-life-supporting orlife-threatening systems required) |
| Alarm / Trigger | NO | NO |
{6}------------------------------------------------
Conclusion:
The physIQ Heart Rhythm Module (Version 1.0) has the same intended use and patient population and similar technological characteristics as those of the predicate device, the Monebo Automated ECG Analysis and Interpretation Software Library. Any differences in technological characteristics have been analyzed and addressed through performance testing which demonstrated that the physIQ Heart Rhythm Module meets it intended use and that any differences between the physIQ Heart Rhythm Module and the predicate device do not raise any new issues of safety or effectiveness. As such, the physIQ Heart Rhythm Module is substantially equivalent to the predicate device.
Summary:
Based on the information provided and the testing conducted, the physIQ Heart Rhythm Module (Version 1.0) is substantially equivalent to the predicate device.
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).