K Number
K180065
Manufacturer
Date Cleared
2019-04-19

(465 days)

Product Code
Regulation Number
872.6070
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Source of illumination for curing photo-activated dental restorative materials and adhesives.

Device Description

Fusion Plus Curing Light is a portable dental LED curing light system. It provides a source of illumination for polymerization of photo-activated dental materials. The principle of operation is the same as the predicate devices: illumination for polymerization or curing of photo-activated dental materials. It uses our patented Fusion optical technology in combination with high-power LED to deliver high efficiency and high-power curing light.

The system consists of a curing handpiece, a charging stand, 100 barrier sleeves, light shields, a power adapter, and instructions for use. The handpiece is constructed of aluminum and consists of a main control, a light head, and a rechargeable battery assembly. The main control houses the control circuitry with microprocessor and user interface consisting of LCD display and two control buttons. A mode button selects operation timer and two curing modes: plasma or pulse mode. A ON/OFF button activates and deactivates the light. The handpiece is non-sterile utilizing the barrier sleeves for infection control. Similar to the predicate devices, the light shield and curing cap absorb emitted light and perform the function of eye protection. The charging stand allows for the placement of the handpiece in a holder with electrical connection to the power adapter as the source for charging the rechargeable battery. FUSION Plus has an interchangeable LED light head. Depending on the light head connected with the main control or added later, the kit can be configured in three options:

Trade Name: FUSION Plus, LED Peak Wavelength (nm): 395-480
Trade Name: FUSION Plus Standard, LED Peak Wavelength (nm): 440-480
The instructions for use detail the use of each head in curing of dental materials.

AI/ML Overview

The DentLight FUSION Plus Curing Light is a dental LED curing light system designed for the polymerization of photo-activated dental restorative materials and adhesives. The manufacturer asserts its substantial equivalence to the predicate device Valo (K083647) by Ultradent Products Inc. and the reference device FUSION (K052593) by DentLight Inc., based on similarities in indications for use, technology, and performance test results.

Below is a detailed breakdown of the acceptance criteria and the study proving the device meets these criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

ParameterAcceptance Criteria (Predicate/Reference Device Performance)FUSION Plus Performance (Reported Device Performance)
Indications for UseSource of illumination for curing photo-activated dental restorative materials and adhesivesSource of illumination for curing photo-activated dental restorative materials and adhesives
Light SourceHigh-power LEDHigh-power LED
Power SupplyInput: 100-240V - 50/60Hz; Output: 5V 1-2A (FUSION) Input: 100-240VAC@50-60Hz; Output: 12V 5A (Valo)Input: 100-240V - 50/60Hz; Output: 5V 1-2A
Peak Wavelength Range (nm)395-480 nm (Valo) 440-480 nm (FUSION)395-480 nm (Plus head) 440-480 nm (Plus Standard)
Curing Optical Intensity (mW/cm²): Plasma Mode4500 (Valo) 2700 (FUSION)4000
Curing Optical Intensity (mW/cm²): Pulse/High Mode1400 (Valo) 1350 (FUSION)2000
User Control Interface2 Buttons2 Buttons
Weight (grams)75 grams (Valo) 111 (FUSION)112
BatteryLithium ion (FUSION)Lithium ion
StructureErgonomic wandErgonomic wand
MaterialsAluminum body, plastic barrier control and light shieldAluminum body, plastic barrier control and light shield
Compliance StandardsIEC 60601-1; IEC 60601-1-2 (Valo and FUSION)IEC 60601-1; IEC 60601-1-2
Optical Irradiance(Performance comparable to predicate devices)Tested to evaluate functionality and performance
Spectral Irradiance(Performance comparable to predicate devices)Tested to evaluate functionality and performance
Curing Time(Performance comparable to predicate devices)Tested to evaluate functionality and performance
Curing Modes and Safety Controls(Performance comparable to predicate devices)Tested to evaluate functionality and performance
Heat Management(Performance comparable to predicate devices)Tested to evaluate functionality and performance
Software and Hardware V&V(No specific criteria mentioned, implies functional and safe)Verified and validated
Risks to Health and Traceability Analysis(No specific criteria mentioned, implies acceptable risk)Analyzed
Electrical Safety and EMC(Reference to standards)Complies with IEC 60601-1, IEC 60601-1-2, and ANSI/ADA 48-2

2. Sample Size Used for the Test Set and Data Provenance:

The provided document does not specify a separate "test set" in the context of human data or image data for the FUSION Plus Curing Light. The non-clinical performance testing involved evaluating the device's functionality and performance directly, rather than using a dataset of clinical cases. The data provenance is essentially from the internal testing conducted by the manufacturer, DentLight Inc.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

Not applicable. This device is a dental curing light, a physical device, not an AI or diagnostic tool that requires expert-established ground truth from clinical cases. The testing focused on the physical and electrical properties of the device.

4. Adjudication Method for the Test Set:

Not applicable, as no ground truth requiring adjudication was established from clinical or imaging data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI assistance:

Not applicable. This device is a dental curing light and does not involve AI assistance for human readers or interpretation of medical images.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

Not applicable. This device is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

Not applicable in the conventional sense. The "ground truth" for this device's performance is established by direct measurement of its physical and electrical characteristics against established industry standards and the performance of predicate devices. For instance, optical irradiance and spectral irradiance are directly measured and compared.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical medical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set was Established:

Not applicable. As there is no training set for an AI model, there is no ground truth to establish for it.

Study Proving Device Meets Acceptance Criteria:

The study that proves the FUSION Plus Curing Light meets its acceptance criteria is a series of non-clinical performance tests conducted by DentLight Inc. These tests were performed to evaluate the device's functionality, performance, safety, and substantial equivalence to the predicate devices. The study was guided by the FDA's "Dental Curing Lights – Premarket Notification [510(k)] Submissions" guidance document.

The non-clinical tests included:

  • Device performance: This category covered specific technical parameters such as optical irradiance, spectral irradiance, curing time, curing modes and safety controls, and heat management.
  • Software and hardware verification and validation: This ensured that the device's internal systems operate correctly and reliably.
  • Risks to health and traceability analysis: This involved assessing potential hazards and ensuring the device's components and manufacturing processes are traceable.
  • Electrical safety and electromagnetic compatibility (EMC): The device was tested for compliance with relevant electrical and safety standards, specifically IEC 60601-1, IEC 60601-1-2, and ANSI/ADA 48-2.

The conclusion drawn from these tests was that, based on similarities in indications for use and technology, coupled with the results from the performance testing, the FUSION Plus Curing Light is substantially equivalent to the predicate devices. The listed performance values in the table above demonstrate that the FUSION Plus Curing Light performs within acceptable ranges established by the predicate devices, and in some cases, exceeds them (e.g., Pulse/High curing optical intensity). Compliance with relevant IEC and ANSI/ADA standards confirms its safety and electromagnetic compatibility.

{0}------------------------------------------------

510(k) Summary for FUSION Plus Curing Lights

1. APPLICANT

K180065

Dentli

DentLight Inc. 1825 Summit Ave., Suite 210 Plano, TX 75074

Phone: 972-889-8857 Fax: 972-346-6550 Contact Person: Richard Liu Date Prepared: April 19, 2019

2. DEVICE NAME

Proprietary Name: FUSION Plus Curing Light Common/Usual Name: Dental Curing Light Classification Name: Ultraviolet activator for polymerization (872.6070) Product Code: EBZ

3. PRIMARY PREDICATE DEVICE

Valo (K083647), manufactured by Ultradent Products, Inc

REFERENCE DEVICE

FUSION (K052593), manufactured by DentLight Inc.

4. INTENDED USE / INDICATIONS FOR USE

Source of illumination for curing photo-activated dental restorative materials and adhesives.

5. DEVICE DESCRIPTION

Fusion Plus Curing Light is a portable dental LED curing light system. It provides a source of illumination for polymerization of photo-activated dental materials. The principle of operation is the same as the predicate devices: illumination for polymerization or curing of photo-activated dental materials. It uses our patented Fusion optical technology in combination with high-power LED to deliver high efficiency and high-power curing light.

The system consists of a curing handpiece, a charging stand, 100 barrier sleeves, light shields, a power adapter, and instructions for use. The handpiece is constructed of aluminum and consists of a main control, a light head, and a rechargeable battery assembly. The main control houses the control circuitry with microprocessor and user interface consisting of LCD display and two control buttons. A mode button selects operation timer and two curing modes: plasma or pulse mode. A ON/OFF button activates and deactivates the light. The handpiece is non-sterile utilizing the barrier sleeves for infection control. Similar to the predicate devices, the light shield and curing cap absorb emitted light

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for "DentLight". The word "Dent" is in gray, while "Light" is in blue. There is a light blue glow above the "i" in "Light".

and perform the function of eye protection. The charging stand allows for the placement of the handpiece in a holder with electrical connection to the power adapter as the source for charging the rechargeable battery. FUSION Plus has an interchangeable LED light head. Depending on the light head connected with the main control or added later, the kit can be configured in three options:

Trade NameLED Peak Wavelength (nm)
FUSION Plus395-480
FUSION Plus Standard440-480
The instructions for use detail the use of each head in curing of dental materials.

6. DEVICE CHARACTERISTICS AND SUBSTANDTIAL EQUIVALENCE

FUSION Plus is substantially equivalent to the predicate device Valo (K083647) manufactured by Ultradent Products Inc. and FUSION (K052593) designed and manufactured by DentLight Inc. A comparison table with the predicate and reference devices lists key features and characteristics of the subject and predicate devices. It shows the product to be substantially equivalent to the predicate devices.

ParameterFUSION PlusValo (PrimaryPredicate Device)FUSION(Reference Device)
Indications for UseSource of illuminationfor curing photo-activated dentalrestorative materialsand adhesivesSource of illuminationfor curing photo-activated dentalrestorative materialsand adhesivesPhotopolymerization of dental resins,restorative composite materials, andorthodontic brackets, bonding, andsealing that are photo-polymerized inthe 440-480nm wavelength range
510(k) NumberK180065K083647K052593
Product CodeEBZEBZEBZ
Regulation Number21 CFR 872.607021 CFR 872.607021 CFR 872.6070
Device ClassIIIIII
Common Name ofClassificationUltraviolet activatorfor polymerizationUltraviolet activatorfor polymerizationUltraviolet activatorfor polymerization
Light SourceHigh-power LEDHigh-power LEDHigh-power LED
Power SupplyInput: 100-240V -50/60Hz; Output: 5V1-2AInput: 100-240VAC@50-60Hz;Output: 12V 5AInput: 100-240V - 50/60Hz; Output: 5V1-2A
Peak Wavelength Range395-480(Plus head)440-480 nm(Plus Standard)395-480 nm440-480 nm
Curing OpticalPlasma400045002700
IntensityPulse/high200014001350
mW/cm²StandardNA1000NA
User Control Interface2 Buttons2 Buttons2 Buttons
Weight (grams)11275 grams111
BatteryLithium ionNALithium ion
StructureErgonomic wandErgonomic wandErgonomic wand
MaterialsAluminum body,plastic barrier controland light shieldAluminum body,plastic barrier controland light shieldAluminum body, plastic barrier controland light shield
Compliance StandardsIEC 60601-1; IEC60601-1-2IEC 60601-1; IEC60601-1-2IEC 60601-1; IEC 60601-1-2

{2}------------------------------------------------

Similarities between the FUSION Plus Curing Light and Valo:

  • · The FUSION Plus has the same indications for use as the primary predicate device Valo.
  • · The FUSION Plus has substantially equivalent curing modes as the primary predicate device Valo.
  • · The FUSION Plus has substantially equivalent performance characteristics as the primary predicate device Valo.
  • · The FUSION Plus is substantially equivalent to the primary predicate device Valo in standards compliance.

Differences between the FUSION Plus Curing Light and Valo:

The primary predicate Valo has a uni-body construction with a fixed aluminum body. The FUSION Plus has a modular aluminum body construction with a swivel interchangeable light head. Plus Head and Plus Standard Head. The Plus Head has a wavelength range of 395- 480nm that is the same as the primary predicate Valo and wider than the reference device FUSION. These wavelengths cover VLC materials with photoinitiator sensitive to all photoactivated dental materials (395- 480mm). The difference in construction improves the operator's ease of use and does not raise any substantial equivalence concerns as it has the same wavelengths of the primary predicate Valo.

Similarities between the FUSION Plus Curing Light and the reference device FUSION:

  • · The FUSION Plus and the reference device FUSION has the same construction. FUSION Plus Standard Head has the same center wavelengths of 440 nm - 480 nm as the reference device FUSION. The wavelengths cover VLC materials with photoinitiators sensitive to photoactivated dental materials in the 440 nm - 480 nm (CQ and PPD).
  • · The FUSION Plus and the reference device FUSION use the same optoelctronic design, FUSION optical beam technology, and construction materials except the Plus Head LED.

Both FUSION Plus head and FUSION Plus Standard head are powered and controlled by the same main control under the same condition with the only difference being the choice of LED and its placement. These differences do not raise concerns about safety and effectiveness as the performances are similar to the primary predicate device and the reference device.

7. NON-CLINICAL PERFORMANCE TESTING

The following non-clinical tests were conducted to evaluate the functionality, performance, safety, and substantial equivalence of the FUSION Plus Curing Light to the predicate device as suggested by the FDA's guidance for industry and FDA staff titled "Dental Curing Lights – Premarket Notification [510(k)] Submissions" issued on March 27, 2006:

  • · device performance
    • o optical irradiance
    • o spectral irradiance
    • o curing time
    • o curing modes and safety controls
    • o heat management
  • · software and hardware verification and validation,
  • · risks to health and traceability analysis
  • · electrical safety and electromagnetic compatibility

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DentLight. The word "Dent" is in gray, and the word "Light" is in blue. There is a small, light blue circle above the "i" in "Light."

FUSION Plus complies with the electrical and optical safety requirements as the predicate devices: IEC 60601-1, IEC 60601-1-2, and ANSI/ADA 48-2.

8. CONCLUSION

Based on similarities in indications for use and technology together with results from performance testing, we believe that FUSION Plus Curing Light is substantially equivalent to the predicate devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K180062

Device Name FUSION Plus Curing Light

Indications for Use (Describe) Source of illumination for curing photo-activated dental restorative materials and adhesives

Trade Name FUSION Plus FUSION Plus Standard

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 19, 2019

DentLight Inc. Richard Liu President 1825 Summit Ave. Ste 210 Plano, Texas 75074

Re: K180065

Trade/Device Name: FUSION Plus Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: March 19, 2019 Received: March 20, 2019

Dear Richard Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 : medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

§ 872.6070 Ultraviolet activator for polymerization.

(a)
Identification. An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.(b)
Classification. Class II.