K Number
K180041
Date Cleared
2018-09-28

(266 days)

Product Code
Regulation Number
866.3990
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative, gastrointestinal microorganism multiplexed nucleic acid-based assay capable of detecting of nucleic acids from the following organisms in unpreserved stool and Cary-Blair media:

  • Adenovirus 40/41 ●
  • Campylobacter (C. jejuni, C. coli) ●
  • Clostridium difficile (C. difficile) toxin A/B (from fresh specimens only) ●
  • Cryptosporidium (C. parvum, C. hominis) 0
  • Entamoeba histolytica ●
  • Escherichia coli (E. coli) 0157 ●
  • Enterotoxigenic E. coli (ETEC) LT/ST ●
  • Enteroaggregative E. coli (EAEC) 0
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis) ●
  • Norovirus GI/GII ●
  • Rotavirus A ●
  • Salmonella spp. .
  • Shiga-like Toxin producing E. coli (STEC) stx1/stx2 ●
  • Shigella (S. boydii, S. sonnei, S. flexneri, S. dysenteriae)/EIEC ●
  • Vibrio spp. (V. cholerae, V. parahaemolyticus, V. vulnificus), specific identification of V. ● parahaemolyticus
  • Yersinia enterocolitica

The BioCode Gastrointestinal Pathogen Panel (GPP) is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BioCode Gastrointestinal Pathogen Panel (GPP). The agent detected may not be the cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Campylobacter spp., E. coli 0157, Shigella/EIEC, Yersinia enterocolitica, and Adenovirus 40/41 were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio spp. (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Device Description

The BioCode Gastrointestinal Pathogen Panel (GPP) is a multiplex nucleic acid test designed to be used with the BioCode MDx 300 system. The BioCode MDx 3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for nultiple gastrointestinal pathogens from a single stool speciment or in Cary Blair. Stool specimens are processed and nucleic acids extracted with easyMAG, an automated system. Once the PCR plate is set up and sealed, all other operations are automated on MDx 3000. The BioCode Gastrointestinal Pathogen Panel simultaneously tests for 17 pathogens (see table below) from unpreserved stool specimens or stool collected in Cary-Blair transport medium. Results from the BioCode Gastrointestinal Pathogen Panel test are available within less than 5 hours.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the BioCode Gastrointestinal Pathogen Panel (GPP) based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria for the clinical performance metrics (PPA and NPA). Instead, it presents the calculated agreement rates and their 95% confidence intervals from the clinical study. The reproducibility study, however, does have clear acceptance criteria defined (>= 95% detection for positive targets, and >= 95% non-detection for negative targets).

Here's a table summarizing the reported clinical performance. Note that "Acceptance Criteria" for clinical performance are inferred based on general expectations for diagnostic assays submitted to the FDA; explicit numerical targets were not provided in the text for these. The Reproducibility section does state explicit criteria.

CategoryTarget (General)Acceptance Criteria (Explicit if available, otherwise inferred)Reported Device Performance
Clinical PerformancePositive Percent Agreement (PPA)Generally high agreement (e.g., >80-90% for most targets)See tables on pages 14-17 for specific organism PPAs and 95% CIs.
Negative Percent Agreement (NPA)Generally high agreement (e.g., >95% for most targets)See tables on pages 14-17 for specific organism NPAs and 95% CIs.
Reproducibility StudyDetection of positive targets (1.5x LoD and 3x LoD)>= 95% detectionAll specified targets at both concentrations achieved 100% detection, except STEC (98.89% at Low). All >95%. (pg 32)
Non-detection of negative targets (for negative control)>= 95% non-detectionGiardia lamblia: 99.78% non-detection. All others 100%. All >95%. (pg 32)
Limit of DetectionLowest concentration with >=95% detection for 20 replicates>= 95% detection (19/20) at specified LoD (for 20 replicates)All targets met 20/20 or 19/20 detection at their determined LoD for 20 replicates across both unpreserved and Cary-Blair stool. (pg 35-36)
Fresh vs. Frozen Stability95% replicates for positive targets should be valid and detected (>=60/63)63/63 replicates for positive targets were valid and detected. (pg 50)
95% of negative targets should be valid and not detected (>=60/63)63/63 replicates for negative targets were valid and not detected (one was detected but invalid for RNA IC). (pg 50)
Interfering Substances/MicrobesNo inhibition or unexpected resultsNo inhibition or unexpected results observed, with some exceptions/details noted for specific cross-reactivity findings (Vibrio spp., Yersinia enterocolitica, Shigella dysenteriae, Rotavirus vaccines, C. meleagridis). (pg 37-38)
Analytical Reactivity/InclusivityDetection of diverse strains at 3X LoDDetection at 3X LoD for most strains; higher if not.All organisms were detected at the indicated concentrations. (pg 38-45)
Analytical Specificity/Cross-ReactivityNo cross-reactivity with common commensals/pathogensNo cross-reactivity observed/predicted with list of organisms, with some exceptions/details noted. (pg 46-48)
Competitive InhibitionNo competitive inhibition observedNo competitive inhibition observed. (pg 48)
Cross Contamination/Sample CarryoverAbsence of carryover or cross-contaminationNo cross-contamination was observed. (pg 50)

2. Sample Sizes and Data Provenance

  • Clinical Study (Prospective):

    • Sample Size: 1558 leftover, de-identified samples.
    • Data Provenance: Prospectively collected from patients in the United States (Baltimore, MD; Tampa, FL; Memphis, TN; Los Angeles, CA). Samples collected between January 2015 and August 2017.
    • Specimen Breakdown:
      • Unpreserved (Fresh): 237 samples
      • Unpreserved (Frozen): 960 samples
      • Cary-Blair (Fresh): 361 samples
    • Inoculated Cary-Blair Samples: 400 unpreserved stool samples from Sites 1 and 2 were thawed and inoculated into Cary-Blair to supplement numbers for specific analytes (pg 21).
  • Pre-selected Archived Specimens (Category III):

    • Sample Size: 260 preselected archived specimens.
    • Data Provenance: Archived clinical specimens that were previously tested positive by different methods (retrospective). These were spiked with known positives and randomized with negative specimens. (pg 23)
  • Contrived Specimens (Category IV):

    • Sample Size: 612 samples total, with 485 positive samples.
    • Data Provenance: Prepared using specimens previously tested negative for all BioCode GPP analytes, then spiked at levels of up to 3X LOD or greater. (pg 25)
  • Asymptomatic Volunteers (Clinical Specificity):

    • Sample Size: 125 clinical stool samples.
    • Data Provenance: Collected from healthy asymptomatic donors from Tampa General Hospital and University of Maryland. (pg 30-31)
  • Reproducibility Study:

    • Sample Size: 7 contrived samples (6 positive, 1 negative control) extracted in triplicate and each assayed in singlet over 10 runs per site. Total 90 replicates per target. (pg 32)
    • Data Provenance: Performed at 3 sites. (pg 32)
  • Limit of Detection (LoD):

    • Sample Size: 20 replicates for each tested organism and sample type (unpreserved stool and Cary-Blair). (pg 35)
  • Interfering Substances:

    • Each sample extracted in triplicate and tested in singlet. (pg 37)
  • Cross Contamination/Sample Carryover:

    • Five complete runs in "checkerboard" pattern. (pg 50)

3. Number of Experts and Qualifications for Ground Truth

The document does not specify the number or qualifications of experts used to establish ground truth for the clinical test set. Instead, it relies on reference methods for clinical performance evaluation (pg 13).

4. Adjudication Method for the Test Set

The document does not describe an adjudication method involving human experts. For the clinical study, the reference method for each target pathogen was used as the ground truth. In some cases, discrepancies between the BioCode GPP and the initial reference method were subjected to bidirectional sequencing or additional rounds of sequencing for "confirmation" (e.g., footnotes on pages 14, 15, 17, 22), which can be considered a form of analytical adjudication to determine the true state of the specimen.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was mentioned. The device is a "multiplex nucleic acid-based assay," indicating an automated laboratory test, not an imaging AI diagnostic that assists human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

Yes, the studies presented are for the standalone performance of the BioCode Gastrointestinal Pathogen Panel (GPP) system, without human interpretation or assistance beyond typical laboratory operational procedures. The results, as quantitative fluorescence intensity (MFI) values, are processed by the BioCode MDx 3000 system to determine positive or negative results.

7. Type of Ground Truth Used

  • Clinical Study (Prospective):

    • Reference standard methods: Culture, FDA-cleared NAAT (Nucleic Acid Amplification Test), or Composite results of two PCR/sequencing assays (pg 13).
    • For discrepancies, bidirectional sequencing or additional rounds of sequencing were used as further confirmation/adjudication (e.g., footnotes on pages 14, 15, 17, 22).
  • Archived Specimens: Previously tested positive by different methods, and verified using analyte-specific PCR followed by bi-directional sequencing performed by Applied BioCode, Inc. (pg 23).

  • Contrived Specimens: Known concentrations of quantified bacteria, viruses, or parasites spiked into negative clinical matrix. (pg 25)

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning. This device is a molecular diagnostic assay, not a machine learning algorithm that is trained on a dataset. The studies described are performance validation studies for an established assay.

9. How the Ground Truth for the Training Set Was Established

Since this is not a machine learning device trained on a "training set," this question is not applicable. The device's probes and primers for detecting targets are designed based on known genetic sequences, not through a data-driven training process like that used for AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, composed of three overlapping shapes. The figure is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2018

Applied BioCode Inc.

Robert Di Tullio Regulatory Consultant 10020 Pioneer Blvd. Suite 102 Santa Fe Springs. CA 90670

Re: K180041

Trade/Device Name: BioCode Gastrointestinal Pathogen Panel (GPP) Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal Microorganism Multiplex Nucleic Acid-based Assay Regulatory Class: Class II Product Code: PCH, OOI Dated: January 08, 2018 Received: January 09, 2018

Dear Mr. Di Tullio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Steven R. Gitterman -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180041

Device Name

BioCode Gastrointestinal Pathogen Panel (GPP)

Indications for Use (Describe)

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative, gastrointestinal microorganism multiplexed nucleic acid-based assay capable of detecting of nucleic acids from the following organisms in unpreserved stool and Cary-Blair media:

  • Adenovirus 40/41 ●
  • Campylobacter (C. jejuni, C. coli) ●
  • Clostridium difficile (C. difficile) toxin A/B (from fresh specimens only) ●
  • Cryptosporidium (C. parvum, C. hominis) 0
  • Entamoeba histolytica ●
  • Escherichia coli (E. coli) 0157 ●
  • Enterotoxigenic E. coli (ETEC) LT/ST ●
  • Enteroaggregative E. coli (EAEC) 0
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis) ●
  • Norovirus GI/GII ●
  • Rotavirus A ●
  • Salmonella spp. .
  • Shiga-like Toxin producing E. coli (STEC) stx1/stx2 ●
  • Shigella (S. boydii, S. sonnei, S. flexneri, S. dysenteriae)/EIEC ●
  • Vibrio spp. (V. cholerae, V. parahaemolyticus, V. vulnificus), specific identification of V. ● parahaemolyticus
  • Yersinia enterocolitica

The BioCode Gastrointestinal Pathogen Panel (GPP) is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BioCode Gastrointestinal Pathogen Panel (GPP). The agent detected may not be the cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

{3}------------------------------------------------

Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Campylobacter spp., E. coli 0157, Shigella/EIEC, Yersinia enterocolitica, and Adenovirus 40/41 were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio spp. (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

APPLIED BIOCODE, INC. 10020 PIONEER BLVD. SUITE 102 SANTA FE SPRINGS, CA 90670, USA

{5}------------------------------------------------

510(k) SUMMARY

Introduction: According to the requirements of 21 CFR 807.92, the following provides sufficient information to understand the basis for a determination of substantial equivalence.

Submitted by:

Applied BioCode, Inc. 10020 Pioneer Blvd. Suite 102 Santa Fe Springs, CA 90670

Contact:

Robert Di Tullio Regulatory Consultant rditullio@apbiocode.com Telephone: 310 801 1235 Fax: 323 372 3816

Date Submitted:

August 22, 2018

Trade Name:

BioCode Gastrointestinal Pathogen Panel

Classification Name and Regulation Number:

Gastrointestinal microorganism multiplex nucleic acid-based assay (21 CFR 866.3990)

Predicate Device:

K121454 – Luminex xTAG® Gastrointestinal Pathogen Panel (GPP)

Intended Use:

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative, gastrointestinal microorganism multiplexed nucleic acid-based assy capable of detecting of nucleic acids from the following organisms in unpreserved stool and Cary-Blair media:

□ Adenovirus 40/41

{6}------------------------------------------------

Campylobacter (C. jejuni, C. coli)

Clostridium difficile (C. difficile) toxin A/B (from fresh specimens only)

Cryptosporidium (C. parvum, C. hominis)

  • Entamoeba histolytica
  • Escherichia coli (E. coli) 0157
  • Enterotoxigenic E. coli (ETEC) LT/ST
  • Enteroaggregative E. coli (EAEC)
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis)

Norovirus GI/GII

  • Rotavirus A
  • Salmonella spp.
  • Shiga-like Toxin producing E. coli (STEC) stx1/stx2
  • Shigella (S. boydii, S. sonnei, S. flexneri, S. dysenteriae)/EIEC
  • Vibrio spp. (V. cholerae, V. parahaemolyticus, V. vulnificus), specific identification of V. parahaemolyticus
  • Yersinia enterocolitica

The BioCode Gastrointestinal Pathogen Panel in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BioCode Gastrointestinal Pathogen Panel (GPP). The agent detected may not be the cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection.

{7}------------------------------------------------

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Compylobacter spp., E. coll O157, Shigella/Elec, Yersinio enterocoltico, and Adenovirus 40/41 were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoebo histolytica, and Vibrio spp. (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

{8}------------------------------------------------

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Adenovirus 40/41, Campylobacter, E. coli 0157, Shigello/ElEC, Yersinia enterocoltica, and Giardied additionally with retrospective clinical specimens. Performance characteristics for Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. valificus, and V. cholerae) were established primarily using contrived clinical specimens.

A gastrointestinal microorganism multiplex nucleic acids in the detection and identification of acute gastroenteritis in the context of outbreaks.

Device Description:

The BioCode Gastrointestinal Pathogen Panel (GPP) is a multiplex nucleic acid test designed to be used with the BioCode MDx 300 system. The BioCode MDx 3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for nultiple gastrointestinal pathogens from a single stool speciment or in Cary Blair. Stool specimens are processed and nucleic acids extracted with easyMAG, an automated system. Once the PCR plate is set up and sealed, all other operations are automated on MDx 3000. The BioCode Gastrointestinal Pathogen Panel simultaneously tests for 17 pathogens (see table below) from unpreserved stool specimens or stool collected in Cary-Blair transport medium. Results from the BioCode Gastrointestinal Pathogen Panel test are available within less than 5 hours.

BacteriaParasites
▪ Campylobacter spp. (C. jejuni, C. coli)▪ Cryptosporidium spp.(C. parvum/C. hominis)
▪ Clostridium difficile toxin A/B▪ Entamoeba histolytica
▪ Enteroaggregative E. coli (EAEC)▪ Giardia lamblia/intestinalis
▪ Enterotoxigenic E. coli (ETEC): LT/STViruses
▪ Shiga-toxin producing E. coli (STEC): stx1/stx2▪ Adenovirus 40/41
▪ E.coli O157▪ Norovirus GI/GII
▪ Shigella spp. /Enteroinvasive E.coli (EIEC)▪ Rotavirus A
▪ Salmonella spp.RNA Internal Control
▪ Vibrio parahaemolyticus
▪ Vibrio spp (not parahaemolyticus)
▪ Yersinia enterocolitica

Bacteria, Viruses, and Parasites Detected by the BioCode Gastrointestinal Pathogen Panel

{9}------------------------------------------------

Device Comparison:

Comparison of the BioCode Gastrointestinal Pathogen Panel with the Predicate Device

CharacteristicProposed DevicePredicate
NameApplied BioCode, Inc. BioCode Gastrointestinal Pathogen PanelxTAG®GastrointestinalPathogen Panel (GPP)
Common NameGastrointestinal Microorganism Multiplex Nucleic acid-based assayGastrointestinalMicroorganismMultiplex Nucleicacid-based assay
510(k) No.N/AK121454
Regulation21CFR 866.399021CFR 866.3990
Product CodePCH, OOIPCH, OOI
Device ClassIIII
Similarities
CharacteristicProposed DevicePredicate
Intended UseThe BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative multiplexedEquivalent
nucleic acid-based in vitro diagnostic test intended for use with the BioCode MDx
3000 Instrument. The BioCode GPP is capable of the simultaneous detection and
identification of nucleic acids from multiple bacteria, viruses, and parasites
extracted directly from unpreserved stool samples or stool preserved in Cary-Blair
transport medium obtained from individuals with signs and/or symptoms of
gastrointestinal infection. The following bacteria (including several diarrheagenic
E. coli/Shigella pathotypes), parasites, and viruses are identified using the BioCode
GPP: Campylobacter (C. jejuni/C. coli), Clostridium difficile (C. difficile) toxin A/B,
Salmonella spp, Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae) , including
specific identification of Vibrio parahaemolyticus, Yersinia enterocolitica ,
Enteroaggregative Escherichia coli (EAEC), Enterotoxigenic Escherichia coli (ETEC)
It/st, E. coli O157 serogroup, Shiga-like toxin-producing Escherichia coli (STEC)
stx1/stx2, Shigella / Enteroinvasive Escherichia coli (EIEC), Cryptosporidium spp,
Entamoeba histolytica, Giardia lamblia (also known as G. intestinalis and G.
duodenalis ), Adenovirus F 40/41, Norovirus GI/GII, Rotavirus A. The BioCode GPP is
indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and
results are meant to be used in conjunction with other clinical, laboratory, and
epidemiological data. For In Vitro Diagnostic Use Only. For Prescription Use Only.
Positive results do not rule out co-infection with organisms not included in the
BioCode GPP. The agent detected may not be the definite cause of the disease.
Negative results in the setting of clinical illness compatible with gastroenteritis may
be due to infection by pathogens that are not detected by this test or non-infectious
causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
Concomitant culture is necessary for organism recovery and further typing of
bacterial agents. This device is not intended to monitor or guide treatment for C.
difficile infection. Due to the small number of positive specimens collected for
certain organisms during the prospective clinical study, performance characteristics
for Adenovirus 40/41, Campylobacter, E. coli O157, Shigella/EIEC, Yersinia
enterocolitica , and Giardia lamblia were established additionally with retrospective
clinical specimens. Performance characteristics for Entamoeba histolytica, Giardia
lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V.
cholerae) were established primarily using contrived clinical specimens. A
gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the
detection and identification of acute gastroenteritis in the context of outbreaks.
CharacteristicProposed DevicePredicate
InstrumentNucleic Acid Purification SystemBioCode MDx 3000Nucleic AcidPurification SystemPCR ThermocyclerLuminex® 100/200TMor MAGPIXinstruments
Sample TypeUnpreserved stool and stool in Cary- Blair MediaSame
ControlsExternally Sourced -Same
Differences
MethodologyMultiplex RT-PCR and probe hybridization followed by fluorescence detection anddecoding of barcoded magnetic beads (BMB) that are coupled to biotinylatedproducts with streptavidin conjugateMultiplex RT-PCR andmultiplex TSPEfollowed byfluorescence-activated sorting oflabeled beads coupledto streptavidin-conjugatedbiotinylated products.
CalibratorsInternal CalibrationExternal CalibrationKit

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

Summary of Performance Characteristics of the

BioCode GPP Clinical Performance

Testing of Prospectively collected Specimens (Categories I and II)

A clinical investigational study was performed in which a total of 1558 leftover, de-identified samples were prospectively collected from patients who underwent stool sample collection for clinical indications at four (4) investigational sites, with each collecting approximately 400 samples. The sites were located in the United States, with a broad geographic representation. The sites were located in the Northeast (Baltimore, MD), Southeast (Tampa, FL), Midwest (Memphis, TN), and West (Los Angeles, CA). In addition testing was performed on archived known positives and contrived specimens to further augment the sample numbers. Each raw stool specimen was de-identified at the site and the de-identified leftover sample divided into aliquots and either tested freshly on the ABC IUO assay at the site, or frozen and tested at a later date at the site. Each stool specimen in Cary-Blair was similarly de-identified at the site and the de-identified leftover sample aliquoted into 4 aliquots and tested freshly on the ABC IUO assay at the site. All prospectively collected samples had age, sex, and therapy status collected by the site. Enrollment of samples covered multiple calendar seasons beginning January 2015 and ending in August 2017, within which to enroll the specimens displaying due diligence in attempting to cover all panel targets with native specimens. The results of the clinical study follow.

Prospective Study Specimens
Total Specimens1558
Gender (n/N(%))
Female778/1558 (49.9)
Male780/1558 (51.1)
Age Category (n/N(%))
< 5 year140/1558 (9.0)
6-21 yrs237/1558 (15.2)
22-59 yrs718/1558 (46.1)
60+ yrs463/1558 (29.7)
Status (n/N(%))
Inpatient1212/1558 (77.8)
Outpatient346/1558 (22.2)

Demographic data for prospective specimens (fresh and frozen).

{13}------------------------------------------------

Clinical Sites collected both Category I and II see the following table for the breakdown.

Breakdown of prospective specimen collection by site.

Unpreserved(Fresh)Unpreserved(Frozen)Cary-Blair(Fresh)
Site 001503500
Site 002503470
Site 0031372630
Site 00600361
Total237960361

Table. Comparator (reference) methods for prospective clinical study.

Target Pathogen/ToxinReference Method
Adenovirus 40/41Composite result of two PCR/sequencing assays
Campylobacter (C. jejuni, C. coli)Culture
Clostridium difficile (C. difficile) toxin A/BFDA cleared NAAT
Cryptosporidium (C. parvum, C. hominis)Composite result of two PCR/sequencing assays
Entamoeba histolyticaComposite result of two PCR/sequencing assays
Escherichia coli (E. coli) 0157Enrichment culture
Enterotoxigenic E. coli (ETEC) LT/STComposite result of two PCR/sequencing assays
Enteroaggregative E. coli (EAEC)Composite result of two PCR/sequencing assays
Giardia lamblia /intestinalisComposite result of two PCR/sequencing assays
Norovirus GI/GIIComposite result of two PCR/sequencing assays
Rotavirus AComposite result of two PCR/sequencing assays
Salmonella spp.Enrichment culture
Shiga-like Toxin producing E. coli (STEC) stx1/stx2Enrichment culture/FDA cleared antigen test
Shigella (S. boydii, S. sonnei, S. flexneri, S. dysenteriae)/EIECEnrichment culture
Vibrio spp. (V. cholerae, V. parahaemolyticus, V. vulnificus)Culture
Yersinia enterocoliticaCulture

Clinical sensitivity/positive agreement was calculated as TP/(TP + FN). TP = true positive by both the reference and BioCode GPP; FN= false negative or negative by BioCode GPP only. Clinical specificity/negative agreement was calculated as TN/(TN + FP). TN = true negative by both the reference and BioCode GPP; FP = false positive or positive by BioCode GPP only. The exact binomial twosided 95% confidence interval was calculated. The results stratified by sample type and storage method are presented in the tables below.

{14}------------------------------------------------

PPANPA
TargetAgreement Raten/N (%)95% CIAgreement Raten/N (%)95% CI
Campylobacter spp.a1/1 (100.0)(2.5, 100.0)234/236 (99.2)(96.97, 99.9)
Clostridium difficileb26/27 (96.3)(81.0, 99.9)208/210 (99.1)(96.6, 99.9)
E.coli 0157N/AN/A237/237 (100.0)(98.46, 100.0)
EAEC1/1 (100.0)(2.5, 100.0)234/234 (100.0)(98.44, 100.0)
ETECc3/3 (100.0)(29.2, 100.0)229/232 (98.7)(96.27, 99.7)
STECdN/AN/A235/237 (99.2)(96.99, 99.9)
Salmonella sppe3/3 (100.0)(29.2, 100.0)232/234 (99.2)(96.95, 99.9)
Shigella/EIECf1/1 (100.0)(2.5, 100.0)233/236 (98.7)(96.33, 99.7)
Vibrio parahaemolyticusgN/AN/A236/237 (99.6)(97.67, 100.0)
Vibrio sppN/AN/A237/237 (100.0)(98.46, 100.0)
Yersinia enterocoliticahN/AN/A236/237 (99.6)(97.67, 100.0)
Cryptosporidium spp1/1 (100.0)(2.5, 100.0)234/234 (100.0)(98.44, 100.0)
Entamoeba histolyticaN/AN/A235/235 (100.0)(98.44, 100.0)
Giardia lambliaiN/AN/A234/235 (99.6)(97.65, 100.0)
Adenovirus 40/41jN/AN/A233/235 (99.2)(96.96, 100.0)
Norovirus GI/GII1/1 (100.0)(2.50, 100.0)235/235 (100.0)(98.44, 100.0)
Rotavirus A1/1 (100.0)(2.50, 100.0)234/235 (99.6)(97.65, 100.0)

Table. Summary of Clinical Study Results (Prospective specimens) for Unpreserved Stool (Fresh).

a - Campylobacter spp: The 2 false positives compared to the culture reference method were tested by bidirectional sequencing, and 1 of 2 confirmed as positive.

b - Clostridium difficile: The 1 false negative compared to the FDA Cleared NAAT reference test produced high Ct (Ct 35.0).

c - ETEC: The 3 false positives compared to bidirectional sequencing were not confirmed as positives by an additional round of sequencing.

d - STEC: The 2 false positives compared to the culture reference method were tested by bidirectional sequencing, and both confirmed as positive.

e– Salmonella spp: The 2 false positives compared to the culture reference method were tested by bidirectional sequencing, and both confirmed as positives.

f - Shigella/EIEC: The 3 false positives compared to the culture reference method were tested by bidirectional sequencing, and all 3 confirmed as positives.

g - Vibrio parahaemolyticus: The 1 false positive sample compared to the culture reference method was tested by bidirectional sequencing and confirmed as positive.

h - Yersinia enterocolitica: The 1 false positive sample compared to the culture reference method was tested by bidirectional sequencing and could not be confirmed as positive.

ij – Giardia lamblia: The 1 false positive to bidirectional sequencing was not confirmed as positive by 2 additional rounds of sequencing.

j - Adenovirus 40/41: The 2 false positives to bidirectional sequencing were not confirmed as positives by an additional round of sequencing.

{15}------------------------------------------------

PPANPA
TargetAgreement Raten/N (%)95% CIAgreement Raten/N (%)95% CI
Campylobacter spp.a3/3 (100.0)(29.2, 100.0)936/952 (98.3)(97.3, 99.0)
Clostridium difficilebN/AN/AN/AN/A
E.coli 0157c1/2 (50.0)(1.3, 98.7)950/954 (99.6)(98.9, 99.9)
EAECd25/29 (86.2)(68.3, 96.1)916/919 (99.7)(99.1, 99.9)
ETECe7/10 (70.0)(34.8, 93.3)934/939 (99.5)(98.8, 99.8)
STECf3/3 (100.0)(29.2, 100.0)918/919 (99.9)(99.4, 100.0)
Salmonella sppg18/22 (81.8)(59.7, 94.8)926/934 (99.1)(98.3, 99.6)
Shigella/EIECh4/5 (80.00)(28.4, 99.5)940/951 (98.8)(97.9, 99.4)
Vibrio parahaemolyticusiN/AN/A955/957 (99.8)(99.3, 100.0)
Vibrio sppN/AN/A956/956 (100.0)(99.6, 100.0)
Yersinia enterocoliticajN/AN/A951/956 (99.5)(98.8, 99.8)
Cryptosporidium spp7/7 (100.0)(59.0, 100.0)941/941 (100.0)(99.6, 100.0)
Entamoeba histolyticaN/AN/A948/948 (100.0)(99.6, 100.0)
Giardia lambliak2/2 (100.0)(15.8, 100.0)940/946 (99.4)(98.6, 99.8)
Adenovirus 40/41l7/10 (70.0)(34.8, 93.3)935/938 (99.7)(99.1, 99.9)
Norovirus GI/GII39/39 (100.0)(91.0, 100.0)913/917 (99.6)(98.9, 99.9)
Rotavirus A19/20 (95.0)(75.1, 99.9)928/936 (99.2)(98.3, 99.6)

Table. Summary of Clinical Study Results (Prospective specimens) for Unpreserved Stool (Frozen).

a - Campylobacter spp: The 16 false positives compared to the culture reference method were tested by bidirectional sequencing, and 8 of 16 confirmed as positives.

b - Clostridium difficile: C. difficile testing must be performed with fresh specimens only, not with previously frozen specimens.

c - E. coli 0157: The one false negative compared to the culture reference method was tested by bidirectional sequencing and could not be confirmed as positive. The 4 false positive samples compared to the culture reference method were tested by bidirectional sequencing, and 3 of 4 confirmed as positives.

d - EAEC: The 4 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; 3 of the 4 confirmed as positives. 2 of the 3 false positives could not be repeated due to low sample volume. The remaining 1 was not detected by addition rounds of sequencing.

e - ETEC: The 3 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; none were confirmed as positives. 1 of the 5 false positives could not be repeated due to low sample volume. The remaining 4 false positives were not confirmed as positives by an additional round of sequencing.

f - STEC: The 1 false positive compared to the culture reference method was tested by bidirectional sequencing and confirmed as positive.

g – Salmonella spp: The 4 false negatives compared to the culture reference method were tested by bidirectional sequencing, and 1 of 4 could not be confirmed as positive samples compared to the culture reference method were tested by bidirectional sequencing and 6 of 8 confirmed as positives.

h - Shigello/ElEC: The 1 false negative compared to the culture reference method was tested by bidirectional sequencing and could not be confirmed as positive. The 11 false positive samples compared to the culture reference method were tested by bidirectional sequencing, and 10 of 11 confirmed as positives.

i - Vibrio parahaemolyticus: The 2 false positives compared to the culture reference method were tested by bidirectional sequencing

{16}------------------------------------------------

and, 1 of 2 confirmed as positive.

j - Yersinia enterocolitica: The 5 false positives compared to the culture reference method were tested by bidirectional sequencing and, 3 of 5 confirmed as positive.

k – Giardia lamblia: The 4 false positives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing: none were confirmed as positives.

l - Adenovirus 40/41: The 3 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; none were confirmed as positives. The 3 false positives were not confirmed as positives by an additional round of sequencing.

Positive AgreementNegative Agreement
BacteriaAgreement Raten/N (%)95% CIAgreement Raten/N (%)95% CI
Campylobacter spp.a2/3 (66.7)(9.4, 99.2)347/358 (96.9)(94.6, 98.5)
Clostridium difficileb37/38 (97.4)(86.2, 99.9)318/322 (98.8)(96.9, 99.7)
E.coli 0157cN/AN/A359/361 (99.5)(98.0, 99.9)
EAECd17/18 (94.4)(72.71, 99.9)336/341 (98.5)(96.6, 99.5)
ETECe13/14 (92.9)(66.13, 99.8)343/345 (99.4)(97.9, 99.9)
STECfN/AN/A359/361 (99.5)(98.0, 99.9)
Salmonella sppg4/5 (80.0)(28.36, 99.5)354/356 (99.4)(98.0, 99.9)
Shigella/EIECh1/2 (50.0)(1.26, 98.7)356/359 (99.2)(97.6, 99.8)
Vibrio parahaemolyticusN/AN/A361/361 (100.0)(99.0, 100.0)
Vibrio sppN/AN/A361/361 (100.0)(99.0, 100.0)
Yersinia enterocoliticaiN/AN/A357/361 (98.9)(97.2, 99.7)
Cryptosporidium sppj3/3 (100.0)(29.24, 100.0)354/356 (99.4)(98.0, 99.9)
Entamoeba histolyticaN/AN/A359/359 (100.0)(99.0, 100.0)
Giardia lambliak1/1 (100.0)(2.50, 100.0)357/358 (99.7)(98.5, 100.0)
Adenovirus 40/41N/AN/A359/359 (100.0)(99.0, 100.0)
Norovirus GI/GIIl6/7 (85.7)(42.13, 99.6)354/354 (100.0)(99.0, 100.0)
Rotavirus A1/1 (100.0)(2.50, 100.0)360/360 (100.0)(98.98, 100.0)
Table. Summary of Clinical Study Results (Prospective specimens) for Native Cary-Blair Samples.
-----------------------------------------------------------------------------------------------------------

a – Campylobacter spp. The 1 false negative compared to reference culture method was tested by bidirectional sequencing and confirmed positive. The 11 false positives compared to reference culture method were tested by bidirectional sequencing, and 11 of 11 confirmed as positives.

b – Clostridium difficile. The 1 false negative compared to the FDA cleared NAAT reference method produced high Ct (35).

c - E. coli 0157. The 2 false positives compared to reference culture method were tested by bidirectional sequencing, and 2 of 2 confirmed as positives.

d – EAEC. The 1 false negative compared to bidirectional sequencing was tested by 2 additional rounds of sequencing and confirmed as positive. The 4 of 5 false positives were not detected by an addition round of sequencing.

e – ETEC. The 1 false negative compared to bidirectional sequencing was tested by 2 additional rounds of sequencing, and was not confirmed as positive. 1 of 2 false positives was confirmed as positive by 2 additional rounds of sequencing.

f – STEC. The 2 false positives compared to reference culture method were tested by bidirectional sequencing, and 2 of 2 confirmed as positives.

{17}------------------------------------------------

g – Salmonella spp. The 1 false negative compared to the reference culture method was tested by bidirectional sequencing and confirmed as positive. The 2 false positives compared to reference culture method were tested by bidirectional sequencing and 1 of 2 confirmed as positive.

h – Shigello/EIEC. The 1 false negative compared to the reference culture method was tested by bidirectional sequencing and could not be confirmed as positive. The 3 false positives compared to reference culture method were tested by bidirectional sequencing, and all 3 confirmed as positives.

i - Yersinio enterocolitica. The 4 false positives compared to the reference culture method were tested by bidirectional sequencing, and none were confirmed as positive.

j – Cryptosporidium spp. The 2 false positives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing and both confirmed as positives.

k – Giardia lamblia. The 1 false positive compared to bidirectional sequencing was not confirmed as positive by 2 additional rounds of sequencing.

l - Norovirus Gl/Gll. The 1 false negative compared to bidirectional sequencing produced a high Ct (37) which indicates that this sample is low positive.

During the prospective clinical study 2.6% (41/1558) of samples were invalid for lack of RNA-IC signal on initial testing. The invalid rate after reflex testing for the prospective study was approximately 0.2% (3/1558).

{18}------------------------------------------------

Analysis of mixed infections in the prospective study

The BioCode GPP detected a total of 49 samples with mixed infections in the prospective clinical study. This represents 3.1% of the total number of specimens (49/1558). 40 were double infections, and 1 was quadruple infection. The most common pathogens in co-infections were with EAEC (22/49, 44.9%) and ETEC (18/49, 36.7%). The most common co-infected by the BioCode GPP in the prospective clinical study are summarized in the table below.

11 11 11 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Multiple Detection CombinationNumber of Specimens
EAEC + ETEC

Most prevalent multiple detection combinations (5 or more instances) from clinical evaluation.

Clostridium difficile + Salmonella spp5
Clinical co-infection combinations detected by BioCode GPP (unpreserved stool).
Distinct Co-Infection Combinations Detected by BioCode GPP
Analyte_1Analyte_2Analyte_3Analyte_4Total Co-infectionsNumber of Discrepant Co-infectionsDiscrepant Analyte(s)
Adenovirus 40/41Rotavirus AN/AN/A11Adenovirus 40/41 (x1)
Campylobacter sppShigella/EIECN/AN/A11All
Campylobacter sppGiardia lambliaN/AN/A11All
Campylobacter sppNorovirus GI/GIIN/AN/A11Campylobacter spp (x1)
Campylobacter sppSTECN/AN/A11All
Campylobacter sppY. enterocoliticaN/AN/A11All
C. difficileETECN/AN/A11ETEC (x1)
Cryptosporidium sppCampylobacter sppN/AN/A11Campylobacter spp (x1)
Cryptosporidium sppGiardia lambliaN/AN/A11Giardia lamblia (x1)
E.coli 0157Norovirus GI/GIIN/AN/A11E.coli 0157 (x1)
E.coli 0157Shigella/EIECN/AN/A11E.coli 0157 (x1)
EAECShigella/EIECN/AN/A11Shigella/EIEC (x1)

{19}------------------------------------------------

Distinct Co-Infection Combinations Detected by BioCode GPPTotalCo-infectionsNumber ofDiscrepantCo-infectionsDiscrepant Analyte(s)
Analyte_1Analyte_2Analyte_3Analyte_4
EAECShigella/EIECNorovirus GI/GIIN/A11Shigella/EIEC (x1)
EAECETECN/AN/A41ETEC (x1)
EAECETECNorovirus GI/GIIN/A11EAEC (x1);ETEC (x1)
EAECGiardia lambliaN/AN/A11All
EAECRotavirus AN/AN/A11Rotavirus A (x1)
Shigella/EIECETECN/AN/A11Shigella/EIEC (x1)
Norovirus GI/GIIRotavirus AN/AN/A11Norovirus GI/GII (x1)
Norovirus GI/GIIRotavirus ASTECN/A11Rotavirus A (x1);STEC (x1);
Norovirus GI/GIIV. parahaemolyticusY. enterocoliticaN/A11All
Total Co-infections2421
Double Infections2017
Triple Infections44

Clinical co-infection combinations detected by BioCode GPP (Cary-Blair).

Distinct Co-Infection Combinations Detected by BioCode GPPTotal Co-infectionsNumber of Discrepant Co-infectionsDiscrepant Analyte(s)
Analyte_1Analyte_2Analyte_3Analyte_4
C. difficileSalmonella sppN/AN/A31All
E.coli O157EAECETECSTEC11E.coli O157 (x1);STEC (x1)
EAECETECN/AN/A41EAEC (x1)
EAECETECY. enterocoliticaN/A22ETEC (x1);Y. enterocolitica (x2)
EAECNorovirus GI/GIIN/AN/A31EAEC (x1)
EAECSTECN/AN/A11STEC (x1)
Total Co-infections147
Double Infections114
Triple Infections22

{20}------------------------------------------------

Distinct Co-Infection Combinations Detected by BioCode GPPTotal Co-infectionsNumber of Discrepant Co-infectionsDiscrepant Analyte(s)
Analyte_1Analyte_2Analyte_3Analyte_4
Quadruple Infections11

Clinical co-infection combinations detected by reference methods (unpreserved stool).

Distinct Co-Infection CombinationsDetected by Reference Methods
Analyte_1Analyte_2Analyte_3TotalCo-infectionsNumber ofDiscrepantCo-infectionsDiscrepant Analyte(s)
Adenovirus40/41EAECN/A11Adenovirus 40/41 (x1)
EAECETECN/A41EAEC (x1)
EAECETECNorovirusGI/GII22EAEC (x1);ETEC (x2)
Total Co-infections74
Double Infections52
Triple Infections22

Clinical co-infection combinations detected by reference methods (Cary-Blair).

Distinct Co-Infection CombinationsDetected by Reference Methods
Analyte_1Analyte_2Analyte_3TotalCo-infectionsNumber ofDiscrepantCo-infectionsDiscrepant Analyte(s)
EAECShigella/EIECETEC11Shigella/EIEC (x1)
Total Co-infections11
Double Infections00
Triple Infections11

{21}------------------------------------------------

Testing of inoculated Cary-Blair specimens from previously frozen prospective specimens

To supplement the number of prospective Cary-Blair specimens. 400 unpreserved stool samples from sites 1 and 2 were thawed and inoculated into Cary-Blair. 3 were removed from the study for improper storage prior to testing. 2 were invalid for RNA IC failure in the unpreserved stool. The following table summarizes the comparison of the results from the initial unpreserved testing and subsequent inoculated results.

Summary of Inoculated Cary-Blair or unpreserved samples Vs. reference method. Agreements were calculated compared to reference method testing of unpreserved stool. Reference testing was not repeated after samples were inoculated to Cary-Blair.

(n)Positive AgreementNegative Agreement
TargetSpecimen TypePPA (%)95% CINPA (%)95% CI
Campylobacter spp.aUnpreserved3942/2 (100.0)(15.81, 100.0)385/392 (98.21)(96.36, 99.28)
Campylobacter spp.aCary-Blair(Inoculated)3962/2 (100.0)(15.81, 100.0)388/394 (98.48)(96.72, 99.44)
E. coli O157bUnpreserved3951/2 (50.00)(1.26, 98.74)389/393 (98.98)(97.41, 99.72)
E. coli O157bCary-Blair(Inoculated)3971/2 (50.00)(1.26, 98.74)391/395 (98.99)(97.43, 99.72)
Enteroaggregative E. coli(EAEC)cUnpreserved39412/14 (85.71)(57.19, 98.22)378/380 (99.47)(98.11, 99.94)
Enteroaggregative E. coli(EAEC)cCary-Blair(Inoculated)39612/14 (85.71)(57.19, 98.22)382/382 (100.0)(99.04, 100.0)
EnterotoxigenicE. coli (ETEC)dUnpreserved3944/6 (66.67)(22.28, 95.67)387/388 (99.74)(98.57, 99.99)
EnterotoxigenicE. coli (ETEC)dCary-Blair(Inoculated)3964/6 (66.67)(22.28, 95.67)386/390 (98.97)(97.39, 99.72)
Shiga toxin-producing E.coli (STEC)Unpreserved3612/2 (100.0)(15.81, 100.0)359/359 (100.0)(98.98, 100.0)
Shiga toxin-producing E.coli (STEC)Cary-Blair(Inoculated)3632/2 (100.0)(15.81, 100.0)361/361 (100.0)(98.98, 100.0)
Salmonella spp.eUnpreserved3955/6 (83.33)(35.88, 99.58)385/389 (98.97)(97.39, 99.72)
Salmonella spp.eCary-Blair(Inoculated)3976/6 (100.0)(54.07, 100.0)389/391 (99.49)(98.16, 99.94)
Shigella/ EIECfUnpreserved3951/1 (100.0)(2.50, 100.0)389/394 (98.73)(97.06, 99.59)
Shigella/ EIECfCary-Blair(Inoculated)3971/1 (100.0)(2.50, 100.0)391/396 (98.74)(97.08, 99.59)
Vibrio parahaemolyticusUnpreserved395N/AN/A395/395 (100.0)(99.07, 100.0)
Vibrio parahaemolyticusCary-Blair(Inoculated)397N/AN/A397/397 (100.0)(99.08, 100.0)
Vibrio spp. (notparahaemolyticus)Unpreserved395N/AN/A395/395 (100.0)(99.07, 100.0)
Cary-Blair(Inoculated)397N/AN/A397/397 (100.0)(99.08, 100.0)
Yersinia enterocoliticagUnpreserved395N/AN/A394/395 (99.75)(98.60, 99.99)
Yersinia enterocoliticagCary-Blair(Inoculated)397N/AN/A396/397 (99.75)(98.60, 99.99)
Cryptosporidium sppUnpreserved3942/2 (100.0)(15.81, 100.0)392/392 (100.0)(99.06, 100.0)
Cryptosporidium sppCary-Blair(Inoculated)3962/2 (100.0)(15.81, 100.0)394/394 (100.0)(99.07, 100.0)
Entamoeba histolyticaUnpreserved394N/AN/A394/394 (100.0)(99.07, 100.0)
Entamoeba histolyticaCary-Blair(Inoculated)396N/AN/A396/396 (100.0)(99.07, 100.0)
TargetSpecimen Type(n)Positive AgreementNegative Agreement
PPA (%)95% CINPA (%)95% CI
Giardia lambliahUnpreserved394N/AN/A391/394 (99.24)(97.79, 99.84)
Giardia lambliahCary-Blair(Inoculated)396N/AN/A394/396 (99.49)(98.19, 99.94)
Adenovirus 40/41iUnpreserved3943/6 (50.00)(11.81, 88.19)388/388 (100.0)(99.05, 100.0)
Adenovirus 40/41iCary-Blair(Inoculated)3962/6 (33.33)(4.33, 77.72)385/390 (98.72)(97.03, 99.58)
Norovirus(GI/GII)Unpreserved39528/28 (100.0)(87.66, 100.0)364/367 (99.18)(97.63, 99.83)
Norovirus(GI/GII)Cary-Blair(Inoculated)39728/28 (100.0)(87.66, 100.0)364/369 (98.64)(96.87, 99.56)
Rotavirus AUnpreserved39511/12 (91.67)(61.52, 99.79)380/383 (99.22)(97.73, 99.84)
Rotavirus ACary-Blair(Inoculated)39711/12 (91.67)(61.52, 99.79)380/385 (98.70)(97.00, 99.58)

{22}------------------------------------------------

a – Campylobacter spp. Unpreserved: The 6 false positives compared to reference culture method were tested by bidirectional sequencing and 4 of 6 confirmed as positives. Cary-Blair: The 7 false positives compared to the reference culture method were tested by bidirectional sequencing and 4 of 7 confirmed as positives.

b - E. coli 0157. Unpreserved: The one false negative compared to the reference culture method was tested by bidirectional sequencing and could not be confirmed as positives compared to the reference culture method were tested by bidirectional sequencing and 3 of 4 confirmed as positives. Cary-Blair: The one false negative compared to the reference culture method was tested by bidirectional sequencing and could not be confirmed as positives compared to reference culture method were tested by bidirectional sequencing and 3 of 4 confirmed as positives.

c - EAEC. Unpreserved: The 2 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; 1 of the 2 confirmed as positive. Cary-Blair: The 2 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; 1 of the 2 confirmed as positives compared to bidirectional sequencing were not confirmed as positive by 2 additional rounds of sequencing.

d - ETEC. Unpreserved: The 2 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; none were confirmed as positives were confirmed as positive by 2 additional rounds of sequencing. Cary-Blair: The 2 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; none were confirmed as positive compared to bidirectional sequencing was not available for confirmation testing.

e – Salmonella spp. Unpreserved: The 2 false positives compared to the reference culture method were tested by bidirectional sequencing and 1 of 2 confirmed as positives. Cary-Blair: The one false negative compared to the reference culture method was tested by bidirectional sequencing and confirmed as positives compared to the reference culture method were tested by bidirectional sequencing and 2 of 4 confirmed as positives.

f - Shigella/EIEC. Unpreserved: The 5 false positives compared to the reference culture method were tested by bidirectional sequencing and 4 of 5 confirmed as positives. Cary-Blair: The 5 false positives compared to the reference culture method were tested by bidirectional sequencing and 4 of 5 confirmed as positives.

g - Yersinia enterocolitica. Unpreserved: The 1 false positive compared to the reference culture method was tested by bidirectional sequencing and confirmed as positive. Cary-Blair: The 1 false positive compared to the reference culture method were tested by bidirectional sequencing and confirmed as positive.

h - Giardia lamblia. Unpreserved: The 2 false positives compared to bidirectional sequencing were not confirmed as positive by 2 additional rounds of sequencing. Cary-Blair: The 3 false positives compared to bidirectional sequencing were not confirmed as positive by 2 additional rounds of sequencing.

i – Adenovirus 40/41. Unpreserved: The 4 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; 1 of 4 was confirmed as positives were not confirmed as positives by an additional round of sequencing. Cary-Blair: The 3 false negatives compared to bidirectional sequencing were tested by 2 additional rounds of sequencing; none confirmed as positive.

{23}------------------------------------------------

Testing of Pre-selected Archived Specimens (Category III)

Several analytes were not encountered or had low prevalence in the clinical study. To supplement the results of the prospective clinical study, 260 preselected archived specimens were assayed. These were archived clinical specimens that were previously tested positive by different methods. Prior to testing with the Applied BioCode Gastrointestinal Pathogen Panel, the presence of the expected analyte was verified in each specimen using analyte-specific PCR followed by bi-directional sequencing performed at Applied BioCode, Inc. The specimens were randomized with negative specimens, such that the users performing the BioCode GPP were blinded to the expected test result. A summary of the demographic information of the tested samples and the results of the BioCode GPP testing are presented in Tables below.

Prospective Study Specimens
Total Specimens260
Gender (n/N(%))
Female123/260 (47.3)
Male137/260 (52.7)
Age Category (n/N(%))
< 5 year54/260 (20.8)
6-21 yrs46/260 (17.7)
22-59 yrs123/260 (47.3)
60+ yrs37/260 (14.2)

Demographic summary for archived specimens

Summary of Clinical specimen Results (Archived specimens)

Positive AgreementNegative Agreement
TargetAgreementn/N (%)95% CIAgreementn/N (%)95% CI
Campylobacter sppa38/40 (95.0)(83.1, 99.4)152/152 (100.0)(97.6, 100.0)
E.coli 015719/19 (100.0)(82.4, 100.0)152/152 (100.0)(97.55, 100.0)
ETEC20/20 (100.0)(83.2, 100.0)152/152 (100.0)(97.6, 100.0)
STECb30/33 (90.9)(75.7, 98.1)152/152 (100.0)(97.6, 100.0)
Salmonella spp.c29/30 (96.7)(82.8, 99.9)152/152 (100.0)(97.6, 100.0)
Shigella/ EIECd43/45 (95.6)(84.9, 99.5)151/152 (99.3)(96.4, 100.0)
Yersinia enterocolitica3/3 (100.0)(29.24, 100.0)152/152 (100.0)(97.6, 100.0)
Cryptosporidium spp.e16/19 (84.2)(60.4, 96.6)152/152 (100.0)(97.6, 100.0)
Giardia lambliaf25/26 (96.2)(83.2, 99.9)152/152 (100.0)(97.6, 100.0)
Adenovirus 40/4126/26 (100.0)(86.8, 100.0)151/152 (99.3)(96.4, 100.0)

{24}------------------------------------------------

Analytes DetectedSimultaneouslyn/N (%)
Total Mixed Infections84
271/84 (84.5)
311/84 (13.1)
42/84 (2.4)

Distribution of Mixed Infections in Archived positives.

Prevalence of Analytes in Mixed Infections in Archived positives.

Prevalence in Mixed InfectionN = 84
Analyten/N%
Campylobacter spp7/848.3
E.coli 015728/8433.3
EAEC30/8435.7
Shigella/EIEC18/8421.4
ETEC17/8420.2
STEC33/8439.3
Salmonella spp12/8414.3
Yersinia enterocolitica4/844.8
Cryptosporidium spp6/847.1
Giardia lamblia6/847.1
Adenovirus 40/4111/8413.1
Norovirus GI/GII4/844.8
Rotavirus A7/848.3

Most Prevalence Multiple Detection Combinations (5 or more instances) in Archived positives.

Multiple Detection CombinationNumber of Specimens
E.coli 0157+STEC21
EAEC+ETEC10

{25}------------------------------------------------

Testing of Contrived Specimens (Category IV)

For some analytes both prospective and archived testing were insufficient to demonstrate system performance. To supplement the prospective and archived specimens were assayed. The contrived specimens were positive for Giardia, E. histolytica, Yersinia enterocolitica, Vibrio parahaemolyticus, and Vibrio spp. These contrived clinical specimens were prepared using specimens that had previously tested negative for all BioCode GPP analytes. Specimens were spiked at levels of up to 3X LOD (~50% of contrived specimens) or greater using multiple strains for each organism. Positive samples of each were prepared, and randomized by mixing with negative samples before testing. A total of 612 samples, 485 positives, were tested. The results of the BioCode GPP testing are presented in the Table below.

PPANP
TargetAgreementRaten/N (%)95% CIAgreementRaten/N (%)95% CI
Vibrio parahaemolyticus88/96 (91.7)(84.2, 96.3)516/516 (100.0)(99.3, 100.0)
Vibrio spp. (not parahaemolyticus)82/94 (87.2)(78.8, 93.2)518/518 (100.0)(99.3, 100.0)
Vibrio cholerae40/47 (85.1)(72.3, 92.6)518/518 (100.0)(99.3, 100.0)
Vibrio vulnificus42/47 (89.4)(77.4, 95.4)518/518 (100.0)(99.3, 100.0)
Yersinia enterocolitica95/98 (96.9)(91.3, 99.4)514/514 (100.0)(99.3, 100.0)
Entamoeba histolytica96/99 (97.1)(91.4, 99.4)507/513 (98.8)(97.5, 99.6)
Giardia lamblia94/98 (95.9)(89.9, 98.9)513/514 (99.8)(98.9, 100.0)

Summary of contrived specimen results

a - All false negative specimens were tested by PCR/bidirectional sequencing and none could be confirmed as positives (not positive via sequencing). It is likely that these were either prepared during shipping and handling.

Detailed breakdown of contrived testing samples

OrganismStrainIDStraincharacterizationLoD(CFU/mL oroocysts/mL)ConcentrationPrepared(CFU/mL oroocysts/mL)Multipleof LODReplicatestestedFalsenegativeFalsenegativedistribution
VibrioparahaemolyticusATCC17802strain EB1015.00E+011.50E+023x722
1.00E+0320x10
1.50E+0330x400
2.50E+0350x30
VibrioparahaemolyticusBEI NR-2199101:K56 strain102957.50E+021.50E+020.2x103
1.50E+032x330
2.37E+033x50
2.50E+033.15200
5.00E+036.3x300
OrganismStrainIDStraincharacterizationLoD(CFU/mL oroocysts/mL)ConcentrationPrepared(CFU/mL oroocysts/mL)Multipleof LODReplicatestestedFalsenegativeFalsenegativedistribution
VibrioparahaemolyticusBEI NR-21990O4:K12 strain480575.00E+017.90E+0310x10
1.50E+023x800
1.00E+0320x30
2.30E+04460x311
VibrioparahaemolyticusBEI NR-22002O3:K6 strainTX21031.71E+021.50E+020.88x91
5.13E+023x410
2.50E+0314.6x40
5.00E+0329x400
1.71E+051000x40
VibrioparahaemolyticusZepto0801903strain Z1345.00E+011.50E+023x611
5.00E+03100x40
1.50E+04300x400
1.92E+053840x40
Vibrio choleraeATCC25870O:1 Vibrio, vib+4.90E+021.47E+033x900
2.45E+0450x41
4.90E+04100x31
9.80E+04200x131
9.80E+052000x10
Vibrio choleraeBEI NR-146O:1 strain 2125same as ATTC39050 EL Tor1.47E+031.47E+031x611
4.44E+033x310
1.48E+0410x10
4.90E+0433x300
1.47E+05100x40
Vibrio choleraeBEI NR-149O:2 (non-O:1,non- O:139strain) Nanking32/1239.20E+031.47E+030.16x32
9.80E+031.1x31
1.47E+041.6x130
2.45E+042.7x10
2.76E+043x30
9.20E+049x100
Vibrio vulnificusATCC27562(vib+) strain324 CDC B96294.90E+021.47E+033x1322
1.96E+034x1400
Vibrio vulnificusATCC29306strain CDCA14022.45E+037.35E+033x1033
7.35E+0430x30
1.23E+0550x300
2.45E+0510040
YersiniaATCCBilups-1803-68,1.50E+034.50E+033x900
OrganismStrainIDStraincharacterizationLoD(CFU/mL oroocysts/mL)ConcentrationPrepared(CFU/mL oroocysts/mL)Multipleof LODReplicatestestedFalsenegativeFalsenegativedistribution
enterocolitica23715biotype2,serotype 0:84.50E+0430x40
1.50E+05100x400
1.48E+079900x40
YersiniaenterocoliticaATCC29913serotype 0:8CDC 497-70, BEINR-2071.50E+034.50E+033x800
6.00E+034x40
7.50E+0450x300
YersiniaenterocoliticaATCC9610O:8 biovar 1strain NCTC129821.50E+034.50E+033x811
7.50E+0450x70
1.50E+05100x400
YersiniaenterocoliticaBEI NR-206O:8 ATCC 27729strain WA1.50E+034.50E+033x800
3.00E+0420x411
4.50E+0430x40
YersiniaenterocoliticaBEI NR-212O:3 strain NCTC111751.50E+034.50E+033x700
1.50E+05100x40
4.50E+05300x400
YersiniaenterocoliticaBEI NR-213O:9 Strain NCTC111741.50E+034.50E+033x911
1.50E+05100x200
EntamoebahistolyticaBEI NR-176HB-301:NIH3.10E+019.30E-013x1311
1.24E+004x20
9.30E+0030x40
3.10E+01100x410
9.30E+01300x41
EntamoebahistolyticaBEI NR-177200:NIH3.10E+019.30E-013x1200
1.55E+0150x81
3.10E+01100x410
EntamoebahistolyticaBEI NR-178HM-1:IMSS3.10E+019.30E-013x1100
6.20E+0020x40
3.10E+01100x411
9.30E+01300x40
EntamoebahistolyticaBEI NR-179Rahman3.10E+019.30E-013x1311
9.30E+0030x40
1.55E+0150x300
3.00E+01100x40
Giardia lambliaBEI NR-9232Mario strain1.81E+033.62E+032x200
5.42E+033x1200
OrganismStrainIDStraincharacterizationLoD(CFU/mL oroocysts/mL)ConcentrationPrepared(CFU/mL oroocysts/mL)Multipleof LODReplicatestestedFalsenegativeFalsenegativedistribution
9.04E+035x30
1.28E+047x30
1.81E+0410x20
6.92E+0438x20
Giardia lambliaBEI NR-9234D.Hall strain1.81E+035.42E+033x1400
9.94E+035x22
1.63E+049x20
1.81E+0410x530
3.62E+0420x21
4.52E+0425x20
Giardia lambliaBEI NR-9235DAN strain1.81E+035.42E+033x1311
9.04E+035x20
3.62E+0420x20
5.42E+0430x200
6.36E+0435x20
Giardia lambliaWaterborneP101Lot# 1604281.81E+035.42E+033x1200
2.71E+0415x60
3.62E+0420x600
3.75E+0421x20

{26}------------------------------------------------

Premarket Notification 510(k)

{27}------------------------------------------------

{28}------------------------------------------------

Premarket Notification 510(k)

{29}------------------------------------------------

Performance of Controls During Clinical Trials

During clinical evaluation of the BioCode GPP, at least one negative control was included on each run. Negative controls were S.T.A.R. Buffer, or well characterized negative specimens. The negative controls completed all processing steps (pretreatment/extractions/amplification). It was recommended that external controls consisting of 4 pools of inactivated organisms (see table) be assayed on a rotating basis with the exception of Giardia (Waterborne Inc.), all controls were prepared by ZeptoMetrix (cat no. NATGIP-ABC; Giardia is now also available with that catalog number).

PoolOrganismStrainSourceDilution Factor
Pool 1Clostridium difficileNAP1Natrol (ZeptoMetrix)1/4
Rotavirus AWANatrol (ZeptoMetrix)1/4
Shigella sonneiZ004Natrol (ZeptoMetrix)1/4
Pool 2Escherichia coli (EAEC)92.0147; EAECNatrol (ZeptoMetrix)1/5
Entamoeba histolyticaDS4-868Natrol (ZeptoMetrix)1/5
Yersinia enterocoliticaClinical isolateNatrol (ZeptoMetrix)1/5
Norovirus GIIrecombinantNatrol (ZeptoMetrix)1/5
VibrioparahaemolyticusClinical isolateNatrol (ZeptoMetrix)1/5
Pool 3Adenovirus Type 41TAKNatrol (ZeptoMetrix)1/4
Escherichia coliO157/STECEDL933Natrol (ZeptoMetrix)1/4
Giardia lambliaH3Waterborne Inc.1/4
Pool 4SalmonellatyphimuriumZ005Natrol (ZeptoMetrix)1/4
Campylobacter jejuniClinical isolateNatrol (ZeptoMetrix)1/4
CryptosporidiumparvumIowaNatrol (ZeptoMetrix)1/4
Escherichia coli (ETEC)ETEC; ST+, LT+Natrol (ZeptoMetrix)1/4
Norovirus GIrecombinantNatrol (ZeptoMetrix)1/4

Performance of Controls during Clinical Trials (including Reproducibility testing)

Site 001(valid/total)Site 002(valid/total)Site 003(valid/total)Site 006(valid/total)All sites(valid/total)
Pool 15/56/611/14a6/628/31
Pool 25/58/85/55/523/23
Pool 38/85/53/34/420/20
Pool 43/35/53/34/420/20
NC24/27b34/35c38/39d15/15111/116

a – The same pool 1 extract was being repeated freeze-thawed. It failed 3 times in a row for low signals, while the RNA IC signals were normal in the other samples and NC for the plate. When the group performed fresh extractions the PC signals returned to normal.

b – 2 failed for False positive signals and 1 for RNA IC not being detected

c – 1 RNA IC not detected

d — 1 RNA IC not detected

{30}------------------------------------------------

General Performance of Assay During Clinical Trials

Run DescriptionNumber% of Total
Valid runs with complete results5349.5
Valid runs with RNA-IC failures for oneor more samplesª3633.6
Partially or completely invalid runs1816.8
Total107100

Accounting of valid, partially invalid, and invalid runs.

a – All invalid results were reflex tested according to the IFU, and all but 3 were resolved as valid results.

Summary of issues causing partially or completely invalid runs.

Reason for failureNumber% of Total
User error43.7
Instrument/Alignment a43.7
Negative Control contamination21.9
Software installation error b32.8
Reagent storage/handling c21.9
Unknown reason32.8
Total Invalid runs1816.8

a MDx 3000 Alignment error at one clinical site accounted for 3 consecutive failed runs before it was corrected.

b Unapproved Software for remote access was installed that resulted in a software error for 2 runs software was removed and issue did not repeat.

c Reagent storage/handling error at one site accounted for 2 consecutive failed runs.

Clinical Specificity - Microbial Detection in Asymptomatic Volunteers

In order to determine baseline levels for each analyte included in the BioCode Gastrointestinal Pathogen Panel in individuals who are not exhibiting signs and symptoms of infectious gastroenteritis, 125 clinical stool samples were collected from healthy asymptomatic donors. These are defined as donors not exhibiting signs and symptoms, or on antibiotics (for symptoms) during the previous 30 days. Asymptomatic donors from two sites, Tampa General Hospital (clinical site 2) and University of Maryland (clinical site 3) and various age groups were included in this study and the demographic information for the donors is shown in the table below. PCR inhibition, as determined by results of the assay internal control (bacteriophage MS2), was observed for two samples (1.6%). After re-running this sample in accordance with the package insert instructions for use, inhibition was still observed in, so no result was reported. A total of 26 samples were positive. The results are summarized in the Table below.

{31}------------------------------------------------

GenderNumber of Subjects
Male67
Female58
Total125
Age
<1-51
6-213
22-5961
>6060

Table. Demographic information for Asymptomatic Volunteers.

Table. Detections in Asymptomatic Volunteers-Stratified by Age

Analyte< 5 yrs6-21 yrs22-59 yrs60+ yrs
All Negative1 (100.0%)3 (100.0%)49 (80.33%)46 (76.67%)
Clostridium difficile0 (0.00%)0 (0.00%)9 (14.75%)11 (18.33%)
EAEC0 (0.00%)0 (0.00%)0 (0.00%)1 (1.67%)
ETEC0 (0.00%)0 (0.00%)2 (3.28%)0 (0.00%)
Salmonella spp0 (0.00%)0 (0.00%)0 (0.00%)1 (1.67%)
Giardia lamblia0 (0.00%)0 (0.00%)1 (1.64%)0 (0.00%)
Norovirus GI/GII0 (0.00%)0 (0.00%)0 (0.00%)1 (1.67%)

{32}------------------------------------------------

ANALYTICAL PERFORMANCE

The results of the analytical studies summarized in the following paragraphs met the acceptance criteria and successfully demonstrated the analytical performance characteristics of the proposed BioCode Gastrointestinal Pathogen Panel assay.

REPRODUCIBILITY STUDY

A study was performed to assess the Reproducibility of the BioCode Gastrointestinal Pathogen Panel on the BioCode MDx 3000. This study was designed to assess intra-assay (within run), Inter-assay (run-torun), day-to-day and site-to-site reproducibility. One lot of reagents was assayed at 3 sites by 2 operators on 1 instrument per site for 5 days (total of 10 runs per site; 90 replicates total). The reproducibility panel consisted of 7 contrived sample 7 a negative control) extracted in triplicate and each assayed in singlet. The samples consisted of combinations of 12 representative targets at 1.5x LoD (Low) and 3x LoD (Medium). Reproducibility was > 99%.

Concentration Level
Medium PositiveLow PositiveNegative
TargetDetection95% CIDetection95% CIAgreement95% CI
n/N (%)n/N (%)n/N (%)
Salmonella bongori90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Clostridium difficile90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Giardia lambia90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)449/450(99.78)(98.77,99.99)
Adenovirus 40/4190/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Shigella sonnei90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Vibrioparahaemolyticus90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
ETEC90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Yersiniaenterocolitica90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
STEC90/90(100.0)(95.98,100.0)89/90(98.89)(93.96,99.97)450/450(100.0)(99.18,100.0)
Campylobacterjejuni90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Rotavirus A90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)
Cryptosporidiumparvum90/90(100.0)(95.98,100.0)90/90(100.0)(95.98,100.0)450/450(100.0)(99.18,100.0)

Qualitative results from Reproducibility study. All results are as expected with the exception of one false positive for Giardia lamblia and one false negative for STEC.

{33}------------------------------------------------

Quantitative results from Reproducibility study. Fluorescent signals from BMBs with the same barcode and calculated to generate median fluorescence index (MFI) for each analyte (shown below). The presence or absence of a pathogen is determined relative to the validated assay utoff by MFI.

TargetAnalyte/ ProbeAnalyteConcentrationAnalyteMeanRepeatabilitySDRepeatability%CVBetween RunsSDBetween Runs%CVBetween DaysSDBetween Days%CVBetween SitesSDBetween Sites%CVTotalSDTotal%CV
SalmonellaLow9174185320.26226.78146415.9500244226.62
SalmonellaMed11572120410.414694.05198017.11119010.28264722.87
tcdA (C. difficile)Low5144140727.35130625.468913.3988217.15222243.2
tcdA (C. difficile)Med8802238127.05144416.4164518.69133715.19349939.76
tcdB (C. difficile)Low11595200417.298477.31229319.7800316127.27
tcdB (C. difficile)Med16441187811.42221913.49280417.056824.15409624.91
Giardia lambliaLow3020624368.0621647.1625168.33334211.07530217.55
Giardia lambliaMed2048510245382418.6700283913.86487123.78
Adenovirus 40Low17155204311.91285016.62198711.5810306.01416024.25
Adenovirus 40Med2085016998.15266612.7919499.35244711.74444821.33
Shigella sonneiLow6115165427.05148724.31133721.87104317.06279745.74
Shigella sonneiMed9707249225.68196820.2800166917.19358836.96
Vibrio parahaemolyticusLow9927269027.17927.98295629.789669.73418842.18
Vibrio parahaemolyticusMed12421154012.39201816.25312625.1600402632.42
ST-a (ETEC)Low14272346924.31356224.96391027.400632644.32
ST-a (ETEC)Med21453415119.35294313.72486922.7232210.83741634.57
ST-b (ETEC)Low13193324624.6199215.1371928.197335.56537340.73
ST-b (ETEC)Med19807320516.18214110.81475624.0118299.23638932.26
Yersinia enterocoliticaLow19049197010.3417299.0816478.65232612.21387220.33
Yersinia enterocoliticaMed2002012466.2214747.36203610.17215610.77353817.67
stx2Low16291334720.54575735.34235314.44516131.68874753.69
stx2Med21117297714.1693132.8200559326.49939044.47
Campylobacter jejuniLow21865238410.9328215.01272712.47230210.53540224.71
Campylobacter jejuniMed26561299211.26387114.5825499.6290310.93623423.47
Rotavirus ALow4579242699.3232747.1529606.46506311.06795917.38
Rotavirus AMed45552467310.260041249.05762216.73984621.61
Cryptosporidium parvumLow25298375414.8422128.74331413.1363914.38657325.98
Cryptosporidium parvumMed2796626499.47312311.1726729.56322611.54585920.95

{34}------------------------------------------------

Premarket Notification 510(k)

{35}------------------------------------------------

LIMIT OF DETECTION (LoD)

A study was performed to assess the performance of the BioCode Gastrointestinal Pathogen Panel on the BioCode MDx 3000 at the Limit of Detection (LoD) for both unpreserved Stool and Cary- Blair specimens. In this study the BioCode GPP was tested with quantified bacteria, virus or parasite stocks (note Norovirus GI and Norovirus GII were tested by CDC). For initial screening, four replicates of each concentration in negative stool and Cary-Blair were extracted on the easyMAG System and tested in singlet with the BioCode GPP on the BioCode MDx 3000 system to estimate LoD. The LoD was confirmed by extracting 20 replicates of each sample type and testing each in singlet for a total of 20 replicates at or near presumptive LoD. LoD for each stock was defined as the lowest concentration with ≥95% detection of 20 replicates (19 out of 20), and was determined separately unpreserved stool and Cary-Blair preserved stool.

StrainSourceUnpreserved Stool LoDUnpreservedStoolDetectionCary-Blair Stool LoDCary-BlairStoolDetection
Campylobacter coliATCC 335595.6 x 101 CFU/mL20/205.6 x 101 CFU/mL20/20
Campylobacter jejunisubsp. jejuniATCC 332927.0 x 102 CFU/mL20/207.0 x 102 CFU/mL20/20
Clostridium difficile(toxinotype 0)ATCC 96891.9 x 102 CFU/mL20/201.9 x 102 CFU/mL20/20
Clostridium difficile(toxinotype III; Nap1)Zeptometrix0801619cf8.3 x 102 CFU/mL20/203.3 x 103 CFU/mL20/20
Enteroaggregative E. coli092:H33 (EAEC)STEC TW044401.4 x 103 CFU/mL20/201.4 x 103 CFU/mL20/20
Enteroinvasive E. coli029:NM (EIEC)ATCC 438923.6 x 102 CFU/mL20/207.5 x 102 CFU/mL20/20
Enterotoxigenic E. coli078:H11 H10407 (ETEC)ATCC 354015.6 x 102 CFU/mL20/205.6 x 102 CFU/mL20/20
Salmonella bongoriSGSC 49001.4 x 103 CFU/mL20/205.5 x 103 CFU/mL20/20
Salmonella enterica ssp.entericaATCC 140282.2 x 103 CFU/mL20/201.1 x 103 CFU/mL19/20
Shiga-like toxin producingE. coli (STEC)ATCC BAA-22172.5 x 103 CFU/mL20/202.5 x 103 CFU/mL20/20
E. coli 0157ATCC 7003763.3 x 103 CFU/mL20/203.3 x 103 CFU/mL20/20
Shigella sonneiATCC 299304.4 x 102 CFU/mL20/201.7 x 103 CFU/mL20/20
Vibrio choleraeATCC 258704.9 x 102 CFU/mL20/204.9 x 102 CFU/mL20/20
Vibrio parahaemolyticusATCC 178021.3 x 101 CFU/mL20/205.0 x 101 CFU/mL20/20

Limit of Detection (LoD) for BioCode GPP.

{36}------------------------------------------------

StrainSourceUnpreserved Stool LoDUnpreservedStoolDetectionCary-Blair Stool LoDCary-BlairStoolDetection
Yersinia enterocoliticaATCC 237151.5 × 103 CFU/mL20/201.5 × 103 CFU/mL20/20
Cryptosporidium hominisUKRC1.3 × 104oocysts/mL20/201.3 × 104oocysts/mL19/20
Cryptosporidium parvumwaterborneP102C3.1 × 103oocysts/mL20/203.1 × 103oocysts/mL20/20
Entamoeba histolyticaHB-301:NIHBEI NR-1783.1 × $10-1$ cysts/mL20/203.1 × $10-1$cysts/mL20/20
Giardia intestinalis(aka G. lamblia)waterborneP1011.8 × 103 cysts/mL20/201.8 × 103 cysts/mL20/20
Adenovirus 40 (dugan)Zeptometrix08100841.0 × $10-1$TCID50/mL20/201.0 × $10-1$TCID50/mL20/20
Adenovirus 41 (TAK)Zeptometrix08100859.4 × $10-2$TCID50/mL20/207.5 × $10-1$TCID50/mL20/20
Rotavirus AATCC VR-20182.5 × 103 TCID50/mL20/206.2 × 102TCID50/mL20/20
Norovirus GlaCDC6.4 × 105copies/gram20/206.5 × 105copies/gram20/20
Norovirus GllaCDC5.2 × 104copies/gram20/209.96 × 104copies/gram20/20

{37}------------------------------------------------

INTERFERING SUBSTANCES

A study was performed to demonstrate the accuracy of the BioCode GPP on the BioCode MDx 3000 in the presence of potentially inhibiting substances or microorganisms. Each member of the interfering substance panel (ISP) was added to prescreened negative stool sample spiked with representative members of the BioCode GPP at 3X LoD and a negative matrix only pre-screened negative stool. One parasite, one virus, one gram-positive bacterium, and one gram-negative bacterium were used as representatives for this study. Each sample was tested with and without potentially interfering substances. Each sample was extracted in triplicate on the easyMag System and tested in singlet with the BioCode GPP on the BioCode MDx 3000 system. Concentrations were determined by reviewing 510k summary results of previous GI panel clinical trials and FDA Guidance. No interfering substances or microorganisms were identified.

Interference Test Panel.

Sample NameOrganismSource
Sample AClostridium difficile (NAP1)Cryptosporidium parvumZeptometrix 0801619cfWaterborne P102C
Sample BHuman Rotavirus AEscherichia coli 10C-3114 (STEC)ATCC VR-2018ATCC BAA-2217
Sample CNegative Matrix OnlyN/A

Potential Microbial Interferents. No inhibition or unexpected results were observed in the presence of high titer for the organisms in the table below.

Microbial InterferentSourceConcentration(CFU/mL)Interference Yes(Y) or No (N)
No Interferent (Control)N/AN/AN
Bacteroides fragilisaZeptometrix 08015831 × 106N
Blastocystis hominisbATCC 507521 × 105N
Candida albicansATCC 140531 × 105N
Clostridium difficile non-toxigenicATCC 7000571 × 106N
Enterococcus faecalisATCC 512991 × 106N
Escherichia coli nonpathogenicATCC BAA-971 × 106N
Pseudomonas aeruginosaATCC 393241 × 106N
Saccharomyces boulardiiATCC MYA-7961 × 105N

a – 1/3 wells of B. fragilis in the negative stool only sample produced a false positive result for Adenovirus 40/41. An additional 5 extractions were then repeated with no false positives.

b - Blastocystis hominis is titered in cells/mL.

{38}------------------------------------------------

Potential Interfering Substances.

Substance InterferentBrand/SourceConcentrationInterference Yes(Y) or No (N)
No InterferentN/AN/AN
Blood (EDTA)Clinical Sample40% w/vN
AmpicillinAmpicillin152 umol/LN
Sodium hypochloriteBleach (10%)50% w/vN
CholesterolCholesterol5% w/vN
Mineral OilMineral oil, USP50% w/vN
HydrocortisoneHydrocortisone cream50% w/vN
Loperamide hydrochlorideImodium5% v/vN
SennosidesSenokot5% w/vN
Magnesium Hydroxide, Aluminum HydroxideMaalox5% w/vN
MetronidazoleMetronidazole14 mg/mLN
Benzalkonium chloride, EthanolMoist towelettes50% w/vN
Mono, di, triglycerides mixSupelco5% w/vN
MucinMucin3 mg/mlN
Naproxen sodiumNaproxen sodium14mg/mlN
Polymyxin B sulfate, bacitracin zinc, NeomycinNeosporin50% w/vN
NystatinNystatin1000 U/mLN
Bismuth subsalicylatePepto-Bismol5% v/vN
PetrolatumPreparation H5% v/vN
Calcium carbonateTums5% w/vN
VancomycinVancomycin12.5 mg/mLN

ANALYTICAL REACTIVITY/INCLUSIVITY

A study was performed to verify Analytical Reactivity/Inclusivity of the BioCode Gastrointestinal Pathogen Panel. Different strains were selected that represent various temporal, geographic, and genetic diversity for each analyte. This study tested a panel of titered stocks for relevant organisms diluted in Pre-Screened Negative Stools at 3X LoD. Stocks not detected at 3X LoD, if applicable, were tested at higher concentrations. Due to a lack of titered specimens, Adenovirus 40/41 clinical samples and Cryptosporidium DNA from the Cryptosporidium reference unit were used (Public Health England). Norovirus Gl and GII genotypes and the Rotavirus vaccine strain were tested by the CDC. All of the organisms were detected at the concentrations indicated.

Shiga-like toxin producing E. coli (STEC) stx1/stx2 and E. coli O157 Inclusivity results

OrganismSerotypeSourceConcentrationDetectedMultiplesof LoDDetected
E. coli 0157E. coli 0157:H45STEC SC373/2 TW07922$9.9 x 10^3$ CFU/mL3x
E. coli 0157:HNMSTEC DA-26 TW07952$9.9 x 10^3$ CFU/mL3x
E. coli 0157:H7STEC 93-111 TW04863$9.9 x 10^3$ CFU/mL3x
E. coli 0157: H7STEC MI06-19 TW14301$9.9 x 10^3$ CFU/mL3x
E. coli 0157:HNTSTEC DA-27 TW07953$9.9 x 10^3$ CFU/mL3x

{39}------------------------------------------------

OrganismSerotypeSourceConcentrationDetectedMultiplesof LoDDetected
Shiga toxinproducing E. coli(STEC)E. coli O157:H7 Strain EDL933BEI NR-11$9.9 x 10^3$ CFU/mL3x
E. coli O26:H11STEC 2332/00 TW08998$7.5 x 10^3$ CFU/mL3x
E. coli O113:H21STEC CL-15 TW02318$7.5 x 10^3$ CFU/mL3x
E. coli O45:H2STEC DEC11C DEC11c$7.5 x 10^3$ CFU/mL3x
E. coli O103:H2STEC 107-226 TW07881$7.5 x 10^3$ CFU/mL3x
E. coli O104:H21STEC G5506 TW04909$7.5 x 10^3$ CFU/mL3x
E. coli O111:H2STEC RD8 TW06296$7.5 x 10^3$ CFU/mL3x
E. coli O111:H8STEC DEC8B$7.5 x 10^3$ CFU/mL3x
E. coli O26:NMSTEC DA-22 TW07948$7.5 x 10^3$ CFU/mL3x
E. coli O145:NMATCC BAA-2192$7.5 x 10^3$ CFU/mL3x
E. coli O146:21STEC DEC16E TW01383$7.5 x 10^3$ CFU/mL3x
E. coli O45:H2STEC MI05-14 TW14003$7.5 x 10^3$ CFU/mL3x
E. coli O121:19STEC MDCH-4 TW07614$7.5 x 10^3$ CFU/mL3x
E. coli O121:NMSTEC DA-37 TW07972$7.5 x 10^3$ CFU/mL3x
E. coli *O104:H4ATCC BAA-2326$7.5 x 10^3$ CFU/mL3x
E. coli O45:H2ATCC BAA-2193$7.5 x 10^3$ CFU/mL3x
E. coli O26: H11BAA-2196$7.5 x 10^3$ CFU/mL3x
E. coli O121:H19BAA-2219$7.5 x 10^3$ CFU/mL3x

Enterotoxigenic E. coli (ETEC), Enteroaggregative E. coli (EAEC) Inclusivity results

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
EnteroaggregativeE. coli (EAEC)E. coli O44:H18STEC 042TW043934.08 x 103 CFU/mL3x
E. coli O111a,111b:K58:H21ATCC 295524.08 x 103 CFU/mL3x
E. coli O104:H4ATCC BAA-23264.08 x 103 CFU/mL3x
E. coli O3:K2aBEI NR-1024.08 x 103 CFU/mL3x

Shigella/Enteroinvasive E. coli (EIEC) Inclusivity results

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
EnteroinvasiveE. coli (EIEC)E. coli O121ATCC BAA-2190$1.08 x 10^3$ CFU/mL3x
E. coli O124:HNMSTEC 929-78TW16574$1.08 x 10^3$ CFU/mL3x
E. coli O136:HSTEC LT-41TW06139$1.08 x 10^3$ CFU/mL3x
E. coli O285A:HNMBEI NR-101$1.08 x 10^3$ CFU/mL3x
E. coli O15ATCC 49105$1.08 x 10^3$ CFU/mL3x
Shigella boydiiShigella boydii (Type 1)ATCC 9207$1.31 x 10^3$ CFU/mL3x
Shigella boydii (Type 2)BEI NR-521$1.31 x 10^3$ CFU/mL3x
Shigella boydii (Type 7)ATCC 9905$1.31 x 10^3$ CFU/mL3x
Shigella boydii (Type 20)ATCC BAA-1247$1.31 x 10^3$ CFU/mL3x
Shigella boydii (Type 3)ATCC 8702$1.31 x 10^3$ CFU/mL3x
Shigella dysenteriae (Type 1)BEI NR-520$1.31 x 10^3$ CFU/mL3x

{40}------------------------------------------------

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
ShigelladysenteriaeShigella dysenteriae (Type 3)ATCC 97511.31 x 103 CFU/mL3x
Shigella dysenteriae (Type 2)ATCC 97501.31 x 103 CFU/mL3x
Shigella dysenteriae (Type 5)ATCC 97641.31 x 103 CFU/mL3x
Shigella dysenteriae (Type 12)ATCC 495521.31 x 103 CFU/mL3x
Shigella flexneriShigella flexneri, strain 24570(Type 2a)BEI NR-5171.31 x 103 CFU/mL3x
Shigella flexneri, strain 2457T(Type 2a)BEI NR-5181.31 x 103 CFU/mL3x
Shigella flexneri (Type 2b)ATCC 120221.31 x 103 CFU/mL3x
Shigella flexneri (Type 6)ATCC 120251.31 x 103 CFU/mL3x
Shigella flexneri (Type 1b)ATCC 93801.31 x 103 CFU/mL3x
Shigella sonneiShigella sonneiATCC 259311.31 x 103 CFU/mL3x
Shigella sonneiATCC 110601.31 x 103 CFU/mL3x
Shigella sonneiATCC 92901.31 x 103 CFU/mL3x
Shigella sonneiATCC 290291.31 x 103 CFU/mL3x

Salmonella Inclusivity results

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
Salmonella entericasubsp. arizonaeSalmonella enterica subsp.EntericaATCC 133146.45 × 103 CFU/mL
Salmonella entericasubsp. salamaeserovar TranoroaATCC 7001486.45 × 103 CFU/mL
Salmonella entericasubsp. entericaserovar MontevideoATCC 7001BAA-7106.45 × 103 CFU/mL
serovar EnteritidisSGSC49016.45 × 103 CFU/mL
serovar EnteritidisATCC 49316.45 × 103 CFU/mL
serovar OranienburgSGSC40796.45 × 103 CFU/mL
serovar Paratyphi B var.L(+)tartrate+SGSC41506.45 × 103 CFU/mL
serovar TyphimuriumSGSC14126.45 × 103 CFU/mL
serovar SaintpaulSGSC25126.45 × 103 CFU/mL
serovar S. typhimurium LT2SGSC26666.45 × 103 CFU/mL
serovar NewportSGSC24936.45 × 103 CFU/mL
serovar NewportATCC 69626.45 × 103 CFU/mL
serovar MuenchenSGSC24906.45 × 103 CFU/mL
serovar AgonaSGSC24586.45 × 103 CFU/mL
serovar JavianaSGSC49176.45 × 103 CFU/mL
serovar SchwarzengrundSGSC25146.45 × 103 CFU/mL
serovar HeidelbergSGSC24806.45 × 103 CFU/mL
serovar InfantisSGSC24846.45 × 103 CFU/mL
serovar MontevideoSGSC24876.45 × 103 CFU/mL
serovar ThompsonSGSC 25196.45 × 103 CFU/mL
serovar HadarSGSC49656.45 × 103 CFU/mL
serovar MississippiSGSC40786.45 × 103 CFU/mL
serovar Paratyphi ASGSC24996.45 × 103 CFU/mL
serovar CholeraesuisSGSC47706.45 × 103 CFU/mL
serovar CholeraesuisATCC 133126.45 × 103 CFU/mL

{41}------------------------------------------------

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
serovar DublinSGSC41576.45 × 103 CFU/mL3x
serovar BraenderupATCC® 7001366.45 × 103 CFU/mL3x
serovar BareillyATCC® 91156.45 × 103 CFU/mL3x
serovar TyphiZeptometrix08019336.45 × 103 CFU/mL3x
Salmonella entericasubsp. IISalmonella enterica subs. IISGSC30396.45 × 103 CFU/mL3x
Salmonella entericasubsp. IIIaSalmonella enterica subs. IIIaSGSC30616.45 × 103 CFU/mL3x
Salmonella entericasubsp. IIIbSalmonella enterica subs. IIIbSGSC30686.45 × 103 CFU/mL3x
Salmonella entericasubsp. IVSalmonella enterica subs. IVSGSC30746.45 × 103 CFU/mL3x
Salmonella entericasubsp. VISalmonella enterica subs VISGSC31166.45 × 103 CFU/mL3x

Campylobacter inclusivity results.

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoDDetected
Campylobacter spp.Campylobacter jejuni subsp. jejuniBEI NR-3992.10 x 103 CFU/mL3x
Campylobacter jejuni subsp. jejuniBEI NR-4002.10 x 103 CFU/mL3x
Campylobacter jejuni subsp. doyleiATCC 493502.10 x 103 CFU/mL3x
Campylobacter jejuni subsp. doyleiATCC 493492.10 x 103 CFU/mL3x
Campylobacter coliATCC 434781.68 x 102 CFU/mL3x
Campylobacter coliATCC 434851.68 x 102 CFU/mL3x
Campylobacter coliBEI HM-2961.68 x 102 CFU/mL3x
Campylobacter coliATCC 434841.68 x 102 CFU/mL3x

Vibrio spp inclusivity results.

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoDDetected
Vibrio spp. (notparahaemolyticus)Vibrio cholerae (O:1 Inaba,Biotype E1 Tor)BEI NR-147$1.47 x 10^3$ CFU/mL3x
Vibrio cholerae (O:1 Inaba,Biotype E1 Tor)BEI NR-146$1.47 x 10^3$ CFU/mL3x
Vibrio cholerae(O:2)BEI NR-149$1.47 x 10^3$ CFU/mL3x
Vibrio cholerae (O:4)BEI NR-151$1.47 x 10^3$ CFU/mL3x
Vibrio cholerae (O:139)BEI NR-144$1.47 x 10^3$ CFU/mL3x
Vibrio cholerae (O:1 Ogawa)ATCC 14035$1.47 x 10^3$ CFU/mL3x
Vibrio vulnificusATCC 27562$1.47 x 10^3$ CFU/mL3x
Vibrio vulnificusATCC BAA-88$1.47 x 10^3$ CFU/mL3x
Vibrio vulnificusATCC 43382$4.9 x 10^3$ CFU/mL10X
Vibrio vulnificusATCC 29306$1.47 x 10^3$ CFU/mL3x
Vibrio vulnificusATCC 29307$1.47 x 10^3$ CFU/mL3x
VibrioparahaemolyticusVibrio parahaemolyticusBEI NR-21990$3.75 x 10^1$ CFU/mL3x
Vibrio parahaemolyticusBEI NR-21991$3.75 x 10^1$ CFU/mL3x

{42}------------------------------------------------

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoDDetected
Vibrio parahaemolyticusBEI NR-220023.75 x 101 CFU/mL3x
Vibrio parahaemolyticusZepto 08019033.75 x 101 CFU/mL3x
Vibrio parahaemolyticusBEI NR-220133.75 x 101 CFU/mL3x

Yersinia enterocolitica inclusivity results.

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
Yersinia enterocoliticaYersinia enterocolitica (O:8)ATCC 96104.43 x 103 CFU/mL3x
Yersinia enterocolitica (O:3)BEI NR-2124.43 x 103 CFU/mL3x
Yersinia enterocolitica (O:8)BEI NR-2064.43 x 103 CFU/mL3x
Yersinia enterocoliticaATCC 299134.43 x 103 CFU/mL3x
Yersinia enterocoliticaBEI NR-2134.43 x 103 CFU/mL3x

Clostridium difficile inclusivity results.

OrganismTOXINOTYPESourceConcentrationDetectedMultiple ofLoDDetected
Clostridium difficile0 A+B+ATCC 43255-FZ2.48 x 103 CFU/mL3x
0 A+B+ATCC 700792-FZ2.48 x 103 CFU/mL3x
0 A+B+ATCC BAA-1382-fz2.48 x 103 CFU/mL3x
0 A+B+ATCC 51695-fz2.48 x 103 CFU/mL3x
0 A+B+ATCC 43599-fz2.48 x 103 CFU/mL3x
0 A+B+ATCC 43596-fz2.48 x 103 CFU/mL3x
0 A+B+ATCC 17858-fz2.48 x 103 CFU/mL3x
0 A+B+ATCC 435942.48 x 103 CFU/mL3x
0 A+B+ATCC 436002.48 x 103 CFU/mL3x
0 A+B+ATCC 178572.48 x 103 CFU/mL3x
0 A+B+ATCC BAA-18712.48 x 103 CFU/mL3x
0 A+B+ATCC BAA-18722.48 x 103 CFU/mL3x
VIII A-B+ATCC 435982.48 x 103 CFU/mL3x
III A+B+ (Nap1)ATCC BAA-18052.48 x 103 CFU/mL3x
XXII A+B+ATCC BAA-18142.48 x 103 CFU/mL3x
III A+B+ATCC BAA-18702.48 x 103 CFU/mL3x
V A+B+ATCC BAA-18752.48 x 103 CFU/mL3x

{43}------------------------------------------------

OrganismSerotypeSourceConcentration DetectedMultiple ofLoDDetected
Adenovirus 40Human adenovirus 40 (dugan)lot #K1220A3.0E-01 TCID50/mL3
Adenovirus 40n/asample ID # SP143Clinical Sample unknown3
n/asample ID # SP258Clinical Sample unknown3
Adenovirus 41n/asample ID # SP170Clinical Sample unknown3
n/asample ID # SP174Clinical Sample unknown3
n/asample ID # SP 276Clinical Sample unknown3
n/asample ID # SP288Clinical Sample unknown3
n/asample ID # SP309Clinical Sample unknown3
n/asample ID # SP329Clinical Sample unknown3
n/asample ID # SP446Clinical Sample unknown3

Adenovirus 40/41 inclusivity results.

Norovirus GI/GII inclusivity results.

OrganismGenotypeSourceConcentrationDetectedMultiple ofLoD Detected
*Norovirus GIGI.1Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.2Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.3Clinical Specimen$1.93 x 10^8$ copies/gram300x
GI.4Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.5Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.6Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.7Clinical Specimen$1.93 x 10^6$ copies/gram3x
GI.8Clinical Specimen$1.93 x 10^6$ copies/gram3x
aNorovirus GIIGII.1Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.2Clinical Specimen$1.57 x 10^6$ copies/gram30x
GII.3Clinical Specimen$1.57 x 10^6$ copies/gram30x
GII.4 New OrleansClinical Specimen$1.57 x 10^6$ copies/gram30x
GII.4 SydneyClinical Specimen$1.57 x 10^5$ copies/gram3x
GII.5Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.6Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.7Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.8Clinical Specimen$1.57 x 10^6$ copies/gram30x
GII.12Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.13Clinical Specimen$1.57 x 10^5$ copies/gram3x
GII.14Clinical Specimen$1.57 x 10^6$ copies/gram30x
GII.16Clinical Specimen$1.57 x 10^6$ copies/gram30x
GII.17Clinical Specimen$1.57 x 10^5$ copies/gram3x

a — Assayed at CDC.

{44}------------------------------------------------

Norovirus GI/GII inclusivity results.

OrganismSerotypeSourceConcentrationDetectedMultiples ofLoDDetected
Rotavirus ARotavirus A (DS-1)ATCC VR-2550$7.44 x 10^3$ IU/mL3x
Rotavirus A (HRV 89-12C2)ATCC VR-2272$7.44 x 10^3$ IU/mL3x
Rotavirus A (WISC2)ATCC VR-2417$7.44 x 10^3$ IU/mL3x
Rotavirus A (HRV 408)ATCC VR-2273$7.44 x 10^3$ IU/mL3x
Rotavirus A (HRV 248)ATCC VR-2274$7.44 x 10^3$ IU/mL3x
Rotavirus A (HU)ATCC VR-1546$7.44 x 10^3$ IU/mL3x
Rotavirus A (HRV CJN)ATCC VR-2275$7.44 x 10^3$ IU/mL3x
Rotavirus ARotavirus A G1Clinical SampleCt= 20N/Aa
Rotavirus ARotavirus A G2Clinical SampleCt= 22
Rotavirus ARotavirus A G3Clinical SampleCt= 26
Rotavirus ARotavirus A G4Clinical SampleCt= 21
Rotavirus ARotavirus A G9Clinical SampleCt= 20
Rotavirus ARota vaccine strainRotaTeq vaccine$1.00 x 10^8$ pfu/mL
Rota vaccine strainRotarix vaccine$1.00 x 10^6$ CCID50/mL

a – Tested at CDC at indicated concentration at CDC, concentration cannot be related to LoD.

Cryptosporidium spp inclusivity results.

OrganismSerotypeSourceConcentration DetectedMultiple ofLoD Detected
C. parvumDNA subtype /IIaA17G1R1UKCR UK28Clinical Sample unknown3
DNA subtype /IIaA15G2R1UKCR UK29Clinical Sample unknown3
DNA subtype /IIaA19G1R1UKCR UK30Clinical Sample unknown3
DNA subtype /IIdA22G1UKCR UK31Clinical Sample unknown3
DNA subtype /IIdA15G1UKCR UK32Clinical Sample unknown3
C. hominisDNA subtype IaA14R3UKCR UKH14Clinical Sample unknown3
DNA subtype IdA18UKH12Clinical Sample unknown3
DNA subtype IbA10G2UKH13Clinical Sample unknown3
DNA E. coli O103:H2NR2520Clinical Sample unknown3

Entamoeba histolytica inclusivity results.

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
Entamoeba histolytica200:NIHBEI NR-1779.36 x 10-1 cysts/mL3x
Entamoeba histolyticaHB-301:NIHBEI NR-1769.36 x 10-1 cysts/mL3x
Entamoeba histolyticaRahmanBEI NR-1799.36 x 10-1 cysts/mL3x
Entamoeba histolyticaH-303:NIHBEI NR-1809.36 x 10-1 cysts/mL3x

{45}------------------------------------------------

OrganismSerotypeSourceConcentrationDetectedMultiple ofLoD Detected
Giardia lamblia/intestinalisEgypt-4BEI NR-92315.42 x 103 cysts/mL3x
MarioBEI NR-92325.42 x 103 cysts/mL3x
D.HallBEI NR-92345.42 x 103 cysts/mL3x
DANBEI NR-92355.42 x 103 cysts/mL3x

Giardia lamblia inclusivity results.

ANALYTICAL SPECIFICITY/CROSS REACTIVITY

A study was performed to verify that the BioCode GPP does not detect DNA or RNA from organisms commonly found in stool specimens or from organisms that can cause similar clinical symptoms. In addition, on-panel organisms were tested at high concentrations to insure there is no cross-reactivity with other panel targets. This study tested a panel of titered stocks and genomic DNA extracts for relevant organisms. Organisms that were not available for were analyzed in silico comparing the whole organism sequence against all primers to assess potential for cross reactivity. Analysis was conducted using BlastN and Primer Blast programs.

Cross-reactivity was not observed with microorganisms tested in this study except for the following.

Empirical testing and in silico sequence analysis indicate that the Vibrio spp assay may also react with some less common Vibrio species (i.e., V. alqinolyticus, and V. mimicus).

Empirical testing and in silico sequence analysis indicate a potential for cross-reactivity with Y. bercovieri, Y. frederiksenii, Y. intermedia and Y. mollaretii near the established LoD for Y. entericolitica (~1.5 x 10³ CFU/mL). Y. rohdei was also detected when present at high levels (>6.8 x 10° CFU/mL). These species are in the Y. enterocolitica group and are suspected human pathogens.

Shiga toxin (stx; identical to stx1 of STEC) is found in Shigella dysenteriae; therefore, a BioCode GPP report with positive test results for Shiga-like toxin-producing E. coli (STEC) and Shigella/Enteroinvasive E. coli (EIEC) in the same sample may indicate the presence of S. dysenteriae.

Empirical testing with a gene fragment construct and in silico sequence analysis do not predict cross reactivity with the closely related E. dispar.

Empirical testing has demonstrated that these assays will detect recombinant viruses included in Rotavirus vaccines.

Empirical testing indicates potential for cross reactivity with C. meleagridis with the Cryptosporidium assay.

{46}------------------------------------------------

No Cross reactivity was observed for the organisms tested below.

Bacteria
Aeromonas jandaeiEgglerthella lentaProteus penneri
Aeromonas mediaEnterobacter aerogenesProteus vulgaris
Aeromonas trotaEnterobacter cloacaeProvidencia alcalifaciens
Aeromonas caviaeEnterococcus faecalisProvidencia stuartii
Aeromonas hydrophilaEnterococcus faeciumPseudomonas aeruginosa
Acinetobacter baumanniiEnteropathogenic Escherichia coliPseudomonas fluorescens
Acinetobacter lwoffiiEscherichia coli Non pathogenicPseudomonas putida
Alcaligenes faecalisEscherichia coli Non pathogenicSaccharomyces boulardii
Bacillus cereusEscherichia coli Non pathogenicSerratia liquefaciens
Bacteroides fragilisEscherichia hermanniiSerratia marcescens
Bacteroides thetaiotaomicronEscherichia vulnerisShewanella algae
Bifidobacterium breveFusobacterium variumStaphylococcus aureus
Campylobacter fetusGardnerella vaginalisStaphylococcus epidermidis
Campylobacter hyointestinalisGemella morbillorumStenotrophomonas maltophilia
Campylobacter lariGrimontia hollisae (formerly vibrio)Streptococcus agalactiae
Campylobacter upsaliensisHaemophilus influenzaeStreptococcus intermedius
Candida albicansHafnia alveiStreptococcus pyogenes
Cedecea davisaeHelicobacter pyloriStreptococcus salivarius
Chlamydia trachomatisKlebsiella oxytocaStreptococcus suis
Citrobacter amalonaticusKlebsiella pneumoniaeTrabulsiella guamensis
Citrobacter freundiiLactobacillus acidophilusVeillonella parvula
Clostridium difficile non-toxigenicLactobacillus reuteriVibrio alginolyticusa
Clostridium difficile non-toxigenicLactococcus lactisVibrio fluvialis
Clostridium difficile non-toxigenicLeminorella grimontiiVibrio mimicusa
Clostridium histolyticumListeria monocytogenesYersinia bercovierib
Clostridium perfringensMorganella morganiiYersinia frederikseniib
Clostridium septicumPeptoniphilus asaccharolyticusYersinia intermediab
Clostridium sordelliiPlesiomonas shigelloidesYersinia mollaretiib
Clostridium sporogenesPorphyromonas asaccharolyticaYersinia pseudotuberculosis
Clostridium tetaniPrevotella melaninogenicaYersinia rohdeiᵇ
Edwardsiella tardaProteus mirabilis

a – Detected as Vibrio spp at high titers, see full submission for details.

b – Detected as Yersinia enterocolitica at high titers, see full submission for details.

{47}------------------------------------------------

No Cross reactivity was observed for the viruses and parasites tested below.

VirusesParasites
Adenovirus 3Enterovirus 68Cryptosporidium cuniculus (DNA)b
Adenovirus 4EnterovirusaCryptosporidium felis (DNA)
Adenovirus 7aEchovirus 11aCryptosporidium meleagridis b
Adenovirus 8HSV Type 2Cryptosporidium meleagridis (DNA)b
Adenovirus 14Norovirus GIVCryptosporidium muris
Adenovirus 37Rhinovirus 1ACryptosporidium ubiquitum (DNA)
Astrovirus type 1aSapovirus GlaEncephalitozoon cuniculi
Astrovirus type 4aSapovirus GIVaEncephalitozoon hellem
Coronavirus 229ESapovirus GVaEncephalitozoon intestinalis
Coronavirus NL63Giardia muris
Coxsackie virus A16aBlastocystis hominisaPentatrichomonas hominis
Coxsackievirus B3Blastocystis hominisToxoplasma gondii
Cytomegalovirus (CMV)Cryptosporidium canis (DNA)

a — tested at CDC

b – Detected as Cryptosporidium spp.

No Cross reactivity was observed for BioCode GPP targets.

BioCode GPP Targets
Campylobacter coliEnterotoxigenic E. coli O78:H11H10407 (ETEC)Vibrio parahaemolyticus
Campylobacter jejuni spp. jejuniShiga toxin producing E. coli (STEC)Yersinia enterocolitica
Clostridium difficile (toxinotype 0 )E. coli O157Cryptosporidium parvum
Clostridium difficile (toxinotype III;Nap1)Salmonella bongoriEntamoeba histolytica HB-301:NIH
Enteroaggregative E. coli O92:H33(EAEC)Salmonella enterica ssp. entericaGiardia intestinalis (aka G. lamblia)
Enteroinvasive E. coli O29:NM (EIEC)Shigella sonneiAdenovirus 40 (dugan)
Rotavirus A

No Cross reactivity was predicted by in silico analysis.

BacteriaBacteriaParasites
Anaerococcus tetradiusEubacterium cylindroidesAncylostoma duodenale
Bifidobacterium adolescentisEubacterium rectaleAscaris lumbricoides
Bifidobacterium longumMegamonas hypermegaleBalantidium coli
Campylobacter concisusMethanobrevibacter smithiiChilomastix mesnili
Campylobacter curvusPeptoniphilus asaccharolyticusCryptosporidium bovis
Campylobacter gracilisRuminococcus bromiiCryptosporidium canis
Campylobacter helveticusRuminococcus flavefaciensCryptosporidium cuniculusb
Campylobacter hominisRuminococcus obeumCryptosporidium felis
Campylobacter lariSelenomonas ruminantiumCryptosporidium fetus
Campylobacter mucosalisVibrio cincinnatiensisCryptosporidium meleagridisb

{48}------------------------------------------------

Campylobacter rectusVibrio furnissiiCryptosporidium muris
Campylobacter showaeVibrio metschnikoviiCryptosporidium ryanae
Campylobacter sputorumYersinia kristenseniiCryptosporidium xiaoi
Campylobacter upsaliensisVirusesDientamoeba fragilis
Campylobacter ureolyticusNorovirus GIVEndolimax nana
Clostridium acetobutylicumRotavirus BEntamoeba coli
Clostridium methylpentosumRotavirus CaEntamoeba dispar
Clostridium novyiRotavirus DEntamoeba hartmanni
Clostridium ramosumRotavirus EEntamoeba moshkovskii
Collinsella aerofaciensRotavirus FEntamoeba polecki
Desulfovibrio pigerSapovirus

a – Cross-reactivity predicted with Porcine Rotavirus C strains only, no cross-reactivity with human Rotavirus C.

b - C. Cuniculus and C. meleagridis cross reactivity for Cryptosporidium spp assay observed in lab testing was not predicted by in silico analysis (Primer Blast).

COMPETITIVE INHIBITION

A study was performed to evaluate the potential for inhibition in samples with mixed infections. Prescreened negative stool was spiked with one target at high concentration (≥10°CFU/mL for bacteria and ≥105 units/mL for viruses or parasites) and two targets at medium concentration (≤3x LoD). Common co-infections were determined by reviewing results of previous Gl Panel clinical trials from 510(k) summaries, publications/posters and internal clinical sample testing. Each sample was extracted in triplicate on the easyMag and each extraction tested in singlet with the Gastrointestinal Pathogen Panel on the BioCode MDx 3000 system. No competitive inhibition was observed.

Competitive inhibition testing results.

PanelDesignationViral/Bacteria StrainSourceLevelScreening TiterTargetProbeAverageMFIScreeningResult(n of 3)
CI-1Clostridium difficileZepto 801619High3.0x $10^6$ CFU/mLtcdB239693/3
Rotavirus AATCC VR-2018Medium7.44X $10^3$ TCID50/mLRota338993/3
Escherichia coliE2348/69 (EPEC)STEC TW06375Medium7.02x $10^3$ CFU/mLEPEC115853/3
CI-2092:H33 Escherichiacoli (EAEC)STEC JM221TW04440High3.0x $10^6$ CFU/mLEAEC265783/3
Escherichia coliE2348/69 (EPEC)STEC TW06375Medium7.02X $10^3$ CFU/mLEPEC106803/3
Clostridium difficileZepto 801619Medium5.7x $10^2$ CFU/mLtcdB87753/3
CI-3Escherichia coliE2348/69 (EPEC)STEC TW06375High3.0x $10^6$ CFU/mLEPEC276713/3
Clostridium difficileZepto 801619Medium5.7x $10^2$ CFU/mLtcdB88763/3
PanelDesignationViral/Bacteria StrainSourceLevelScreening TiterTargetProbeAverageMFIScreeningResult(n of 3)
Rotavirus AATCC VR-2018Medium$7.44X10^3$ TCID50/mLRota316173/3
CI-4Escherichia coliE2348/69 (EPEC)STEC TW06375High$3x10^6$ CFU/mLEPEC260593/3
092:H33 Escherichiacoli (EAEC)STEC JM221TW04440Medium$4.08X10^3$ CFU/mLEAEC221753/3
Campylobacter jejunisubsp .jejuniATCC 33292Medium$2.1X10^3$ CFU/mLCampy322863/3
Campylobacter jejunisubsp. DoyleiATCC 49349High$3.0x10^6$ CFU/mLCampy247383/3
CI-5Escherichia coliE2348/69 (EPEC)STEC TW06375Medium$7.02X10^3$ CFU/mLEPEC73823/3
092:H33 Escherichiacoli (EAEC)STEC JM221TW04440Medium$4.08X10^3$ CFU/mLEAEC224883/3
CI-6092:H33 Escherichiacoli (EAEC)STEC JM221TW04440High$3x10^6$ CFU/mLEAEC249683/3
Campylobacter jejunisub sp .jejuniATCC 33292Medium$2.1X10^3$ CFU/mLCampy332023/3
Escherichia coliE2348/69 (EPEC)STEC TW06375Medium$7.02X10^3$ CFU/mLEPEC103423/3
Shiga-toxin producingE. coli (STEC)ATCC BAA-2217High$3.0x10^6$ CFU/mLstx2349133/3
CI-7Giardia intestinaliswaterborne P101Medium$5.42X10^3$ cysts/mLG.lam372823/3
Shigella sonneiATCC 29930Medium$1.31X10^3$ CFU/mLShig105183/3
Cl-8Giardia intestinaliswaterborne P101High$3.0x10^5$ cysts/mLG.lam128323/3
Shigella sonneiATCC 29930Medium$1.31X10^3$ CFU/mLShig125303/3
Shiga-toxin producingE. coli (STEC)ATCC BAA-2217Medium$7.5X10^3$ CFU/mLstx2297853/3
CI-9Shigella sonneiATCC 29930High$3.0x10^6$ CFU/mLShig124833/3
Shiga-toxin producingE. coli (STEC)ATCC BAA-2217Medium$7.5X10^3$ CFU/mLstx2318903/3
Giardia intestinaliswaterborne P101Medium$5.42X10^3$ cysts/mLG.lam312273/3

{49}------------------------------------------------

CROSS CONTAMINATION/SAMPLE CARRYOVER

A study was performed to demonstrate the absence of carryover or cross-contamination when using the BioCode Gastrointestinal Pathogen Panel in conjunction with the easyMag. High-positive samples (EAEC at 10° CFU/mL) were tested alternating with no-template control samples in a

{50}------------------------------------------------

"checkerboard" pattern. Samples were extracted checkerboard and assayed in singlet. The study consisted of five complete runs from extraction to BioCode MDx 3000 results on one instrument. No cross contamination was observed.

FRESH VS. FROZEN STABILITY

Since many of the unpreserved samples in three of the four sites using 60 spiked unpreserved stool specimens to assess fresh vs frozen specimen stability. Specimens were contrived and tested at time 0 and after ≥2 freeze thaw cycles. Samples were tested at Low positive (1X-1.5X LoD) spiked into 7 negative stool samples. Cary-Blair specimens were not subjected to frozen storage during the clinical study, therefore there were not tested. Frozen samples displayed stability throughout the length of the study.

Fresh Vs. Frozen results.

Acceptance CriteriaResults
95% replicates for all positive targets in the contrived sample should beValid and Detected (≥60/63)63/63
95% of negative targets should be Valid and not detected (≥60/63)63a/63

a - Sample Target was detected, but was invalid for RNA IC.

Analysis of results from Fresh Vs Frozen study.

AnalysisDelta Fresh -FrozenPercentDecrease
Campylobacter jejuni subsp. jejuniATCC 33292FF1383514%
Escherichia coli 10C-3114 (STEC)ATCC BAA-2217FF2301912%
Human rotavirus AATCC VR-2018FF3443713%
Vibrio parahaemolyticusATCC 17802FF412607%
O78:H11 Escherichia coli strain H10407 (ETEC) STaATCC 35401FF525038%
Shigella sonneiATCC 29930FF6236520%
Salmonella enterica subsp. EntericaATCC 14028FF7-148-2%
Human adenovirus 40 (dugan)ZeptometrixFF86103%
Cryptosporidium parvumwaterborne P102MFF9313414%
NegativeNegativeFF1017196%

SPECIMEN STABILITY

A study was performed to assess the specimen stability limitations for the optimal performance of the BioCode Gastrointestinal Pathogen Panel on the BioCode MDx 3000. This study employed the use of spiked specimens to assess the following storage conditions:

Samples in Cary-Blair - 0, 2, 4 and 6 days at room temperature (20-25°C); 2, 4 and 6 days at 2-8°C Unpreserved Stool - 0, 2, 4 and 6 days at 2-8°C Fresh vs. Frozen (-90°C to -60°C) (2x freeze thaws) - 30, 60 and 90 days at -90°C to -60°C S.T.A.R. buffer after pretreatment (SK38 Tubes prior to extraction) - 24 hours at 2-8°C Fresh vs. Frozen (-90°C to -60°C) (2x freeze thaws) - 30, 60 and 90 days at -90°C to -60°C Extracted Nucleic Acid - 24 hours at 2-8°C Fresh vs. Frozen (-90°C to -60°C) (2x freeze thaws) - 30, 60 and 90 days at -90°C to -60°C

{51}------------------------------------------------

One parasite, one virus, one-gram positive bacterium, and one-gram negative bacterium was used for testing. Spiked and clinical samples were tested at ~3X LoD with two organisms in each sample. Concentrated organism stocks were serially diluted in Cary-Blair and combined with prescreened negative stool and each condition was assayed with 3 replicates at each time point. Samples displayed stability under the various conditions throughout the length of the study.

Image /page/51/Figure/2 description: The image contains four line graphs that show the MFI (Mean Fluorescence Intensity) values over time for different stool samples under different preservation conditions. The top two graphs show "Upreserved stool at 4°C" over 6 days, with lines representing SSA-Un tcdB, SSA-Un Crypto, SSB-Un Stx2, and SSB-Un Rota. The bottom left graph shows "Upreserved stool at -80°C" over 90 days, with lines for SSA-Un Crypto, SSB-Un Stx2, and SSB-Un Rota. The bottom right graph shows "SK38 tubes at 4°C" over 24 hours, with lines for SSA SK38 tcdB, SSA-SK38 Crypto, SSB-SK38 Stx2, and SSB-SK38 Rota.

{52}------------------------------------------------

Image /page/52/Figure/1 description: The image contains four line graphs that show the MFI (Mean Fluorescence Intensity) values over time in days for different specimens. The top two graphs show data for Nucleic Acid and SK38 tubes stored at -80°C, while the bottom two graphs show data for Cary-Blair Specimens stored at 4°C and at Room Temperature. Each graph plots four different lines representing SSA-Un tcdB, SSA-Un Crypto, SSB-Un Stx2, and SSB-Un Rota, with MFI values ranging from 0 to 50000 on the y-axis and time in days on the x-axis. The Cary-Blair Specimens graphs show data over a shorter time period of 0 to 6 days, while the Nucleic Acid and SK38 tubes graphs show data over a period of 0 to 90 days.

Image /page/52/Figure/2 description: The image is a title for a figure. The title states that the figure is a graphic display of MFI results at indicated storage temperatures and timepoints. The title is written in a clear, easy-to-read font.

Figure. Graphic display of MFI results at indicated storage temperatures and timepoints.

CONCLUSION

The intended use and fundamental scientific technology of the BioCode Gastrointestinal Pathogen Panel Assay is substantially equivalent to the predicate device. Clinical and non-clinical studies have established that the BioCode Gastrointestinal Pathogen Panel is substantially equivalent to the predicate device.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).