K Number
K173913
Manufacturer
Date Cleared
2018-05-04

(133 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MEDRAD® Imaging Bulk Package Transfer Set (IBP Transfer Set) is indicated for the transfer of ULTRAVIST® (Iopromide), ISOVUE® (Iopamidol), and OMNIPAQUE™ (Iohexal) contrast media as supplied in an approved Imaging Bulk Package (IBP) presentation to empty sterile syringes on single-use only syringe-based contrast power injection systems indicated for the controlled, automatic venous administration of contrast agents for CT procedures. The Transfer Set is to be discarded after one of the following conditions has occurred first; the contrast media container has been depleted, the contrast media use time has expired, or 10 hours has elapsed since the container was penetrated.

Device Description

The MEDRAD ® Imaging Bulk Package Transfer Set is a pre-administration filling device that is designed to transfer fluid from an imaging bulk package into multiple sterile syringes via a powered injector system prior to a CT procedure. There is no direct patient contact with the use of this device. It is intended to spike one bulk package of iodinated contrast media only. Each imaging bulk transfer set consists of a spike, flexible tubing, and a swabbable valve. The transfer set is provided sterile, individually packaged, and is not intended to be resterilized.

AI/ML Overview

The MEDRAD® Imaging Bulk Package Transfer Set is not an AI device, therefore, the requested information regarding AI device performance, sample sizes for training and test sets, expert involvement, and ground truth establishment is not applicable.

This device is an intravascular administration set designed for transferring contrast media. The provided document focuses on its substantial equivalence to a predicate device, as required for a 510(k) submission to the FDA.

Here's the relevant information provided in the document:

1. Table of Acceptance Criteria and Reported Device Performance (Non-AI related):

Acceptance Criteria (Standard / Test)Reported Device Performance (Compliance)
Sterilization: Validation to a sterility assurance level of 10-6 in accordance with ISO 11135-1:2014Complies with the standard
Shelf-Life: Packaging validated in accordance with ISO 11607-1:2006Complies with the standard (5-year shelf-life demonstrated, compared to 1-year for predicate)
Biocompatibility: Indirect patient contact materials verified in accordance with ISO 10993-1:2009 (R2013), including: Cytotoxicity, Sensitization, Irritation / Intracutaneous Reactivity, Acute Systemic Toxicity, Acute Systemic Injection Materials Mediated Pyrogen, Hemocompatibility (Hemolysis, Partial Thromboplastin Time, Platelet and Leukocyte)Materials comply with the standard
Performance - Bench: Tested in accordance with: - ISO 594-2:1998 (Conical fittings with 6% Luer taper) - ISO 8536-4:2007 (Infusion sets for single use, gravity feed)Complies with the standards
Additional Performance Testing: - Microbial ingress testing (Swabbable Valve, Bottle Septum, Spike, Syringe) at T=0, 4, 10, 14 hrs, and bottle depletion - Chemical compatibility and injectable particulate (Evaluation of conformance to approved release specifications of ULTRAVIST® and USP monographs for ISOVUE® and OMNIPAQUE™) - Material compatibility (Evaluation under worst-case conditions and simulated use with ULTRAVIST® 370 as solvent for leachable compounds)Complies with its predetermined specifications

2. Sample size used for the test set and the data provenance:

  • This device is not an AI/software device, so there isn't a "test set" in the traditional sense of a dataset for algorithm evaluation.
  • The performance "testing" refers to bench testing, sterilization validations, and biocompatibility assessments of the physical device components and its interaction with contrast media. The document does not specify the sample sizes for these physical and chemical tests, which is common for this type of submission.
  • Data provenance is from internal testing and validation by Bayer U.S. LLC, an American company. The tests are prospective in nature, as they are conducted to validate the new device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

  • Not applicable, as this is not an AI device relying on expert review of clinical data for ground truth. The "ground truth" for this medical device comes from established industry standards (e.g., ISO standards) and specified performance criteria for physical and chemical properties.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable, as this is not an AI device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable, as this is not an AI device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable, as this is not an AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For physical product testing, the "ground truth" is defined by established international standards (e.g., ISO, USP monographs), internal design specifications, and validated test methods. Compliance implies that the device meets these pre-defined, objectively measurable criteria (e.g., sterility level, material compatibility, functional performance).

8. The sample size for the training set:

  • Not applicable, as this is not an AI device.

9. How the ground truth for the training set was established:

  • Not applicable, as this is not an AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

May 4, 2018

Bayer U.S. LLC Leslie O'Nan Regulatory Affairs Manager 1 Bayer Drive Indianola, Pennsylvania 15051

Re: K173913

Trade/Device Name: MEDRAD® Imaging Bulk Package Transfer Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: POH Dated: April 13, 2018 Received: April 16, 2018

Dear Leslie O'Nan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina Kiang
-s

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173913

Device Name

MEDRAD® Imaging Bulk Package Transfer Set

Indications for Use (Describe)

The MEDRAD® Imaging Bulk Package Transfer Set (IBP Transfer Set) is indicated for the transfer of ULTRAVIST® (lopromide), ISOVUE® (lopamidol), and OMNIPAQUE™ (lohexal) contrast media as supplied in an approved Imaging Bulk Package (IBP) presentation to empty sterile syringes on single-use only syringe-based contrast power injection systems indicated for the controlled, automatic venous administration of contrast agents for CT procedures. The Transfer Set is to be discarded after one of the following conditions has occurred first; the contrast media container has been depleted, the contrast media use time has expired, or 10 hours has elapsed since the container was penetrated.

Type of Use ( Select one or both, as applicable )
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically, with each letter in a bold, sans-serif font. The letters are gray and outlined in black. The wordmark is encircled by a ring that transitions from green at the top to blue at the bottom, creating a gradient effect.

K173913 510(k) Summary

Date Prepared:April 13, 2018
Submitter:Bayer U.S. LLC1 Bayer DriveIndianola, PA 15051
Primary Contact:Leslie S. O'NanRegulatory Affairs ManagerPhone: (412) 406-3165Fax: (412) 767-2451Email: leslie.o’nan@bayer.com
Device Trade Name:MEDRAD ® Imaging Bulk Package Transfer Set
Common Name:Iodinated Contrast Media Transfer Tubing Set
Classification Name:Intravascular Administration Set
Regulation Number:21 CFR 880.5440
Product Code:PQH
Classification:Class II
Predicate Device:Bracco Injeneering Transfer SetBracco Injeneering S.A.K133147, June 20, 2014
Device Description:The MEDRAD ® Imaging Bulk Package Transfer Set is apre-administration filling device that is designed totransfer fluid from an imaging bulk package into multiplesterile syringes via a powered injector system prior to aCT procedure. There is no direct patient contact with

Bayer U.S. LLC
1 Bayer Drive
Indianola, PA 15051-0780
U.S.A.
(412) 767-2400
www.bayer.com

{4}------------------------------------------------

the use of this device. It is intended to spike one bulk package of iodinated contrast media only. Each imaging bulk transfer set consists of a spike, flexible tubing, and a swabbable valve. The transfer set is provided sterile, individually packaged, and is not intended to be resterilized. The MEDRAD® Imaging Bulk Package Transfer Set (IBP Indications for Use: Transfer Set) is indicated for the transfer of ULTRAVIST® (Iopromide), ISOVUE® (Iopamidol), and OMNIPAQUE™ (Iohexal) contrast media as supplied in an approved Imaging Bulk Package (IBP) presentation to empty sterile syringes on single-use only syringe-based contrast power injection systems indicated for the controlled, automatic venous administration of contrast agents for CT procedures. The Transfer Set is to be discarded after one of the following conditions has occurred first: the contrast media container has been depleted, the contrast media use time has expired, or 10 hours has elapsed since the container was penetrated. Performance Testing: The Imaging Bulk Package Transfer Set is ethylene oxide Sterilization: (EtO) sterilized and was validated to a sterility assurance level of 10-6 in accordance with the following standard: ISO 11135-1: 2014, Sterilization of health care . products Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. Verification results indicate that the MEDRAD® Imaging Bulk Package Transfer Set complies with the standard.

{5}------------------------------------------------

Shelf-Life:The Imaging Bulk Package Transfer Set is sterilized andits packaging was validated in accordance with thefollowing standard:ISO 11607-1: 2006 Packaging for terminallysterilized medical devices - Part 1: requirementsfor materials, sterile barrier systems andpackaging systems; andVerification results indicate that the MEDRAD® ImagingBulk Package Transfer Set complies with the standard.
Biocompatibility:The MEDRAD® Imaging Bulk Package Transfer Setindirect patient contact materials were verified inaccordance with the following standard:ISO 10993-1: 2009 (R2013) Biological evaluation ofmedical devices Part 1: Evaluation and testingwithin a risk management processThe following test program was selected for anexternally communicating, indirect blood path,limited contact (≤ 24 h) device: Cytotoxicity Sensitization Irritation / Intracutaneous Reactivity Acute Systemic Toxicity Acute Systemic Injection Materials Mediated Pyrogen Hemocompatibility Hemolysis Partial Thromboplastin Time Platelet and LeukocyteVerification results indicated that the materials complywith the standard.
Performance - Bench:The MEDRAD® Imaging Bulk Package Transfer Set wastested for performance and verified in accordance withthe following standards:

{6}------------------------------------------------

  • ISO 594-2:1998, Conical fittings with 6 % (Luer) . taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
  • . ISO 8536-4:2007, Infusion equipment for medical use – Part 4: Infusion sets for single use, gravity feed

Verification results indicate that the MEDRAD® Imaging Bulk Package Transfer Set complies with the standards.

Additional testing included:

  • Microbial ingress testing conducted at time . points, T = 0 hr, 4 hr, 10 hr, 14 hr, and bottle depletion for the following components:
    • Swabbable Valve
    • . Bottle Septum
    • Spike
    • Syringe
  • Chemical compatibility and injectable particulate ● – Through evaluation of conformance to the approved release specifications of ULTRAVIST® and USP monographs for ISOVUE® and OMNIPAQUE™
  • Material compatibility Through evaluation of the . transfer set under worst-case conditions and simulated use with ULTRAVIST® 370 as solvent for leachable compounds.

Verification results indicate that the MEDRAD® Imaging Bulk Package Transfer Set complies with its predetermined specifications

{7}------------------------------------------------

Predicate Device Comparison:

ItemPredicate Device:Bracco Injeneering TransferSetSubject Device:MEDRAD® Imaging BulkPackage Transfer SetComparison
Intended UseThe transfer set is intended for thetransfer of fluids from bulkcontainers to empty sterile syringeson syringe based contrast deliverysystems (injectors).The MEDRAD® Imaging Bulk PackageTransfer Set is intended for the transferof fluids from bulk containers to emptysterile syringes on syringe basedcontrast delivery systems (injectors).Same
The Bracco Injeneering Transfer Setis a component of a contrastmanagement system and isindicated for the transfer of Isovue®(Iopamidol Injection) contrast mediaas supplied in an Imaging BulkPackage to empty sterile syringes onsingle-use only syringe-basedcontrast injection systems indicatedfor the controlled, automaticadministration on the venous side,of contrast agents for CTprocedures. The Transfer Set is to bediscarded after the contrast mediacontainer has been depleted or 10hours has elapsed since thecontainer was penetrated,whichever occurs first.The MEDRAD® Imaging Bulk PackageTransfer Set (IBP Transfer Set) isindicated for the transfer ofULTRAVIST® (Iopromide), ISOVUE®(Iopamidol), and OMNIPAQUE™(Iohexal) contrast media as supplied inan approved Imaging Bulk Package (IBP)presentation to empty sterile syringeson single-use only syringe-basedcontrast power injection systemsindicated for the controlled, automaticvenous administration of contrastagents for CT procedures.The Transfer Set is to be discarded afterone of the following conditions hasoccurred first: the contrast mediacontainer has been depleted, thecontrast media use time has expired, or10 hours has elapsed since thecontainer was penetrated.The predicate device is indicatedfor use with ISOVUE® (Iopamidol)contrast media as supplied in anImaging Bulk Package; whereas,the subject device is indicated foruse with ULTRAVIST®(Iopromide), ISOVUE®(Iopamidol), and OMNIPAQUE™(Iohexal) contrast media assupplied in an Imaging BulkPackage. The MEDRAD® ImagingBulk Package Transfer Set hasbeen tested with each of thelisted contrast agents and testinghas demonstrated that thedifferences do not raise newquestions of safety and efficacy.
Indications forUse
ItemPredicate Device:Bracco Injeneering TransferSetSubject Device:MEDRAD® Imaging BulkPackage Transfer SetComparison
Tubing Length20"20"Same
Single UseYesYesSame
BiocompatibleYesYesSame
SterileYesYesSame
SterilityAssurance Level10-610-6Same
SterilizationMethodEthylene Oxide (EtO)Ethylene Oxide (EtO)Same
DisposableYesYesSame
Shelf Lifeone (1) year shelf lifefive (5) year shelf lifeShelf life testing hasdemonstrated that the differencedoes not raise new questions ofsafety and efficacy.
PackagingIndividually packaged in a TyvekpouchIndividually packaged in a Tyvek pouchSame
Microbial IngressTestingYesYesSame
ChemicalCompatibilityTestingYesYesSame
MaterialCompatibilityTestingYesYesSame

{8}------------------------------------------------

Bayer U.S., LLC

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged vertically in a circle. The circle has a gradient, transitioning from green at the top to blue at the bottom. The letters are in a sans-serif font and are colored gray.

Conclusion:

Bayer considers the MEDRAD® Imaging Bulk Package Transfer Set to be substantially equivalent to the predicate device listed above. This conclusion is based upon device similarities in indications for use, functional design, the technological characteristics comparison and testing that demonstrates that the MEDRAD® Imaging Bulk Package Transfer Set is substantially equivalent to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.