(29 days)
The Stryker AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the Stryker AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
This is a Special 510(k) submission for a proposed labeling change that impacts the Stryker® IRF Light Source and SafeLight Cable (referred to as Stryker AIM Light Source hereafter) which was cleared under K142310.
The Stryker AIM Light Source is part of the Stryker® Infrared Fluorescence (IRF) Imaging System, which is an endoscopic illumination and imaging system for real-time high definition (HD) visible light and near-infrared dye fluorescence imaging of indocvanine green (ICG) during minimally invasive surgery. The Stryker® Infrared Fluorescence (IRF) Imaging System consists of the following main components:
- A light source console and a light cable for outputting light within a visible light spectrum as well as near-infrared light spectrum (focus of this submission).
- A camera control unit for processing near-infrared and visible light images, and a coupler that is attached to the laparoscope and a camera head (cleared under K132785).
- A laparoscope for visible light and near-infrared light illumination and imaging (cleared under K142310)
- An imaging agent kit containing ICG (Kit containing IC-Green® drug and Aqueous Solvents, manufactured by Akorn (NDA 011525), cleared under K142310)
There are no proposed changes to any other components of the Stryker® Infrared Fluorescence (IRF) system.
The Stryker AIM Light Source uses an illuminator with a laser light source to illuminate the area of surgery. The ICG is administered intravenously prior to image obtainment. A laser light is illuminated to the site of the surgery using a laparoscope. Upon the absorption of laser light, the ICG in the vessels is excited and emits infrared light at a different wave length. The camera system captures the infrared emission, processes the image and displays it on a surgical display.
The proposed labeling change is to allow the Stryker AIM Light Source to work with two additional kit configurations of the same imaging agent (ICG). Specifically, both device user manual and ICG kit Instruction for Use are updated to adequately cross-label and cross-reference the compatible device/ICG kit(s) per FDA guidance. It is to be noted that Stryker Endoscopy is not the manufacturer of the Imaging agent kits or the ICG drug; the proposed change in labeling does not necessitate any change in the imaging drug formulation, dosage, rate, or route of administration,
This 510(k) premarket notification describes a labeling change for the Stryker AIM Light Source and SafeLight Cable (originally cleared under K142310) to confirm its compatibility with two additional Indocyanine Green (ICG) kits. Therefore, the information provided primarily focuses on demonstrating the substantial equivalence of the modified device (with updated labeling) to its predicate device, rather than presenting a new study for the core performance of the device itself.
Based on the provided text, a detailed answer to your request is limited:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that the modification is primarily a labeling change to allow compatibility with two additional ICG kits. It explicitly mentions: "The modification proposed within this submission for the Stryker AIM Light source does not impact the performance characteristics that were cleared in the previous submissions." Therefore, no new acceptance criteria or device performance metrics for the current submission are detailed, as the core performance of the device was accepted under K142310.
However, the document does state that "Bench testing was conducted to demonstrate the compatibility of the Stryker AIM Light Source with the PINPOINT PAQ® AIM." The specific acceptance criteria for this compatibility testing are not explicitly described in the provided text. It can be inferred that the acceptance criteria would involve successful functional operation of the light source with the new ICG kits without adverse effects or changes in imaging quality, consistent with the performance previously established.
| Acceptance Criteria (Inferred from "compatibility" testing) | Reported Device Performance |
|---|---|
| Successful functional operation with PINPOINT PAQ® AIM | Demonstrated compatibility |
| No adverse impact on current performance characteristics | Confirmed no impact |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: Not explicitly stated. The "bench testing" mentioned for compatibility generally doesn't involve a "test set" in the sense of patient data. It would likely involve a limited number of physical devices and the specified ICG kits.
- Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable for this specific modification as it involves bench testing for compatibility rather than clinical data. Previous clinical studies for the original device (K142310) are mentioned but not detailed here.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This submission focuses on a labeling change and bench testing for compatibility, not on establishing ground truth for a clinical test set.
4. Adjudication Method for the Test Set:
Not applicable.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The device is a light source and cable for fluorescence imaging, not an AI-assisted diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is hardware (light source and cable), not an algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):
Not applicable for this submission. The "ground truth" for the compatibility testing would be the successful functional output of the integrated system.
8. The Sample Size for the Training Set:
Not applicable. This is a hardware device with a labeling change, not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set was Established:
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
January 18, 2018
Stryker Jay Nayar Senior Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138
Re: K173866
Trade/Device Name: Stryker AIM Light Source and SafeLight cable Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: December 18, 2017 Received: December 20, 2017
Dear Jay Nayar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
Stryker AIM Light Source and SafeLight cable
Indications for Use (Describe)
The Stryker AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the Stryker AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 7. 510(k) Summary
General Information 1.
| 510(k) Sponsor | Stryker Endoscopy |
|---|---|
| Address | 5900 Optical CourtSan Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Correspondence Person | Jay NayarSr. Regulatory Affairs SpecialistStryker Endoscopy |
| Contact Information | Email: jay.nayar@stryker.comPhone: 408-754-2729 |
| Date Prepared | December 18, 2017 |
2. Proposed Device
Proposed Device:
| Proprietary Name | Stryker® AIM Light Source and SafeLight Cable |
|---|---|
| Common Name | Light Source, Illuminator |
| Classification Name | Confocal optical imaging |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | OWN |
| Regulatory Class | II |
Predicate Device: 3.
| Proprietary Name | Stryker® Infrared Fluorescence (IRF) Light Source and SafeLightCable |
|---|---|
| Premarket Notification | K142310 |
| Common Name | Light Source, Illuminator |
| Classification Name | Confocal Optical Imaging |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | OWN |
| Regulatory Class | II |
| This device has not been subjected to a design-related recall |
Reference Device 4.
| Proprietary Name | Stryker® Infrared Illuminating System (IRIS)[Consisting of AIM LightSource and IRIS Ureteral Kit ] |
|---|---|
| Premarket Notification | K151243 |
| Common Name | Light Source, Illuminator |
| Classification Name | Fiberoptic light ureteral catheter, Confocal optical imaging |
| Regulation Number | 21 CFR 876.4020, 21 CFR 876.1500, |
| Product Code | FCS, OWN |
| Regulatory Class | II |
| This device has not been subjected to a design-related recall |
{4}------------------------------------------------
Device Description ട്.
This is a Special 510(k) submission for a proposed labeling change that impacts the Stryker® IRF Light Source and SafeLight Cable (referred to as Stryker AIM Light Source hereafter) which was cleared under K142310.
The Stryker AIM Light Source is part of the Stryker® Infrared Fluorescence (IRF) Imaging System, which is an endoscopic illumination and imaging system for real-time high definition (HD) visible light and near-infrared dye fluorescence imaging of indocvanine green (ICG) during minimally invasive surgery. The Stryker® Infrared Fluorescence (IRF) Imaging System consists of the following main components:
- A light source console and a light cable for outputting light within a visible light spectrum ● as well as near-infrared light spectrum (focus of this submission).
- A camera control unit for processing near-infrared and visible light images, and a coupler ● that is attached to the laparoscope and a camera head (cleared under K132785).
- . A laparoscope for visible light and near-infrared light illumination and imaging (cleared under K142310)
- An imaging agent kit containing ICG (Kit containing IC-Green® drug and Aqueous . Solvents, manufactured by Akorn (NDA 011525), cleared under K142310)
There are no proposed changes to any other components of the Stryker® Infrared Fluorescence (IRF) system.
The Stryker AIM Light Source uses an illuminator with a laser light source to illuminate the area of surgery. The ICG is administered intravenously prior to image obtainment. A laser light is illuminated to the site of the surgery using a laparoscope. Upon the absorption of laser light, the ICG in the vessels is excited and emits infrared light at a different wave length. The camera system captures the infrared emission, processes the image and displays it on a surgical display.
The proposed labeling change is to allow the Stryker AIM Light Source to work with two additional kit configurations of the same imaging agent (ICG). Specifically, both device user manual and ICG kit Instruction for Use are updated to adequately cross-label and cross-reference the compatible device/ICG kit(s) per FDA guidance . It is to be noted that Stryker Endoscopy is not the manufacturer of the Imaging agent kits or the ICG drug; the proposed change in labeling does not necessitate any change in the imaging drug formulation, dosage, rate, or route of administration,
6. Imaging Agent Description
The proposed versions of ICG kits were previously cleared by Novadaq Technologies Inc under K150956 and K161792. There are two configurations of the ICG kits, one includes IC-Green from Akorn (NDA 011525), and the other one includes Indocyanine Green for Injection, USP from Diagnostic Green (formerly Pulsion, ANDA 040811). Indocyanine Green for Injection, USP is the
1 Guidance for Industry: New Contrast Indication Considerations for Devices and Approved Drug and Biological Products.
{5}------------------------------------------------
generic version of IC-Green.
The configurations are:
- . PINPOINT PAQ® AIM (containing IC-Green®, NDA 011525);
- PINPOINT PAQ® AIM (containing Indocyanine Green for Injection, USP, ANDA 040811);
Each PINPOINT PAQ® AIM contains sufficient amounts of ICG, diluents, syringes, and Instructions for Use for guidance for the end user. Each configuration of the PINPOINT PAQ® AIM has a unique part number assigned to it, and a corresponding Instructions for Use insert. The Instructions for Use also have unique part numbers for ease of assembly of the PAQs and guidance for the end user regarding compatible imaging devices.
7. Indications for Use
The Stryker® AIM Light Source and SafeLight Cable are indicated for use to provide realtime endoscopic visible and near-infrared fluorescence imaging. The Stryker® AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the Stryker® AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
Please note this is the same indications for use as the previously cleared one for the Stryker AIM Light Source under K142310.
8. Comparison of Technological Characteristics with the Predicate Device
The intent of this premarket submission is to modify the labeling of the Stryker® AIM Light Source to show compatibility with two additional ICG kits. The ICG kits consist of the imaging agent (indocyanine green or ICG), diluents, syringes and device specific Instruction For Use.
All proposed modifications have been assessed through risk analysis, and where necessary, verification and validation activities were performed. These activities did not raise new questions of safety or effectiveness.
Given this, the proposed Stryker® AIM Light Source is substantially equivalent to the respective legally marketed predicate device (cleared under K142310) in regard to intended use, fundamental scientific technology, operational principles, and performance attributes.
9. Performance Testing
The modification proposed within this submission for the Stryker AIM Light source does
{6}------------------------------------------------
not impact the performance characteristics that were cleared in the previous submissions. Bench testing was conducted to demonstrate the compatibility of the Stryker AIM Light Source with the PINPOINT PAQ® AIM .
Please note clinical studies, animal studies, human experience, bench testing and in vitro testing were conducted to evaluate the safety and performance of the Stryker® IRF Light Source and SafeLight Cable which was cleared under K142310. Additional clinical testing is not required to demonstrate substantial equivalence for this submission as the proposed modification is primarily to the labeling to satisfy the cross labeling requirements per FDA guidance2.
10. Conclusion
The information presented within this special premarket submission demonstrates that the Stryker AIM Light source is compatible with the PINPOINT PAQ® AIM, and is substantially equivalent to the Stryker® Infrared Fluorescence (IRF) Light Source and SafeLight Cable (cleared under K142310).
2 Guidance for Industry: New Contrast Indication Considerations for Devices and Approved Drug and Biological Products.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.