K Number
K150956

Validate with FDA (Live)

Date Cleared
2016-05-23

(410 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Device Description

The PINPOINT Endoscopic Fluorescence Imaging System (PINPOINT, PINPOINT System) is comprised of an endoscopic video processor/ illuminator (VPI) which is capable of providing visible and near-infrared illumination to a surgical laparoscope, surgical laparoscopes optimized for visible (VIS) and near-infrared (NIR) illumination and imaging, a camera head that is also optimized for visible and near-infrared imaging, and a flexible light guide cable. The following laparoscope models are included as part of the PINPOINT System: SC9100, SC9104, SC9130, SC9131, SC9134 and SC9144. These are the major components of the PINPOINT System.

During surgical procedures, PINPOINT may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the illuminator and the resulting reflected light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

To provide NIR fluorescence imaging, PINPOINT is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the VPI, and the fluorescence response is then imaged with the camera, processed and displayed on an HD video monitor.

When used with a VIS/NIR laparoscope, PINPOINT can operate to provide illumination and imaging of both visible light and ICG fluorescence images simultaneously.

AI/ML Overview

The provided text includes a 510(k) summary for the PINPOINT Endoscopic Fluorescence Imaging System, which describes performance data. However, it does not include specific acceptance criteria or a detailed study proving the device meets those criteria in the format requested.

Here's what can be extracted and what information is missing based on your request:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated in the document. The document primarily focuses on demonstrating substantial equivalence to predicate devices and conformance to general safety and performance standards.
  • Reported Device Performance:
    • Non-Clinical Performance:
      • Conforms to IEC 60601-1:2012 (Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance).
      • Conforms to IEC 60601-1-2:2007 (Electromagnetic compatibility).
      • Conforms to IEC 60601-2-18:2009 (Particular requirements for the basic safety and essential performance of endoscopic equipment).
      • Assessed as a Class 3R laser device with internal maximum Class 4 laser radiation according to IEC 60825:2007.
      • Animal testing (porcine model) validated in vivo fluorescence imaging capability.
    • Clinical Performance:
      • Clinical evaluation based on literature search and post-market surveillance adverse event information found continued clinical safety and performance.
      • Clinical study and case reports (Journal of Gastrointestinal Surgery, Surgical Innovation, Int! J Sur Case Reports) using the PINPOINT System support expanded Indications for Use.
      • No new concerns related to safety and effectiveness were introduced.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified for clinical studies in this document. Animal testing used a "porcine model" (singular, but likely implying multiple animals or instances).
  • Data Provenance:
    • Non-Clinical: Novadaq Technologies Inc. and accredited third-party laboratories.
    • Clinical: Clinical literature search, post-market surveillance (adverse event information), a clinical study, and case reports published in medical journals. The country of origin for the clinical data is not specified. It is likely retrospective for the literature review and post-market surveillance, and potentially prospective for the mentioned "clinical study," but this is not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not specified. The document refers to clinical studies and case reports, which would typically involve medical experts, but details about ground truth establishment and expert qualifications are absent.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study mentioned. This device appears to be an imaging system intended for direct visualization by surgeons, not an AI-assisted diagnostic tool for interpretation by readers. The document highlights its use "to provide real time endoscopic visible and near-infrared fluorescence imaging" and "enables surgeons to perform minimally invasive surgery" and "visual assessment."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is an imaging system for direct surgeon use, not an algorithm. The "device is not intended for standalone use for biliary duct visualization" refers to its use in conjunction with white light and cholangiography, not to an algorithm's standalone performance.

7. The type of ground truth used

  • Animal Testing: "In vivo fluorescence imaging capability" validated in a porcine model. This implies comparison against expected fluorescence patterns.
  • Clinical Evaluation: Clinical literature and adverse event information. For the clinical study and case reports, the ground truth would typically be surgical outcomes, direct visualization during surgery, or other clinical assessments, but this is not explicitly detailed.

8. The sample size for the training set

  • Not applicable as this is not an AI/machine learning device with a distinct training set. The device was "designed and developed."

9. How the ground truth for the training set was established

  • Not applicable (see above).

Summary of Missing Information:

The provided text is a 510(k) summary demonstrating substantial equivalence to predicate devices and conformance to general safety and performance standards. It lacks the detailed specifications of acceptance criteria, specific clinical study designs (e.g., sample sizes, ground truth establishment methods, expert qualifications, adjudication, MRMC studies) that would be present for a novel AI/algorithm-based diagnostic device. The performance data presented focuses on regulatory compliance, technological characteristics, and general safety/performance based on existing literature and animal models, rather than detailed quantitative performance metrics from controlled clinical trials against specific acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 23, 2016

Novadaq Technologies Inc. Ms. Jen Pendlebury Director of Regulatory Affairs 8329 Eastlake Drive Unit 101 Burnaby, British Columbia Canada V5A 4W2

Re: K150956

Trade/Device Name: Pinpoint Endoscopic Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: May 13, 2016 Received: May 16, 2016

Dear Ms. Pendlebury:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150956

Device Name

PINPOINT Endoscopic Fluorescence Imaging System (PC9000)

Indications for Use (Describe)

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 - 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92.

Trade Name:PINPOINT Endoscopic Fluorescence Imaging System
Device Model Number:PC9000
Common Name:Endoscope Video Imaging System
Classification:21 CFR § 876.1500
Classification Name:Laparoscope, General & Plastic Surgery; Angiographic X-raySystem
Product Code:GCJ; IZI
Classification:Class II
Manufacturer:Novadaq Technologies Inc.8329 Eastlake Drive, Unit 101Burnaby, British ColumbiaCanada V5A 4W2
Contact Name:Jen PendleburyDirector of Regulatory AffairsTel: (905) 629-3822 ext. 205Fax: (905) 249-0656
Date 510(k) Summary Prepared:May 12, 2016
Predicate Devices:SPY Scope Intra-Operative Imaging System/ PINPOINT(K091515)(Novadaq Technologies Inc.)Da Vinci Firefly Imaging System (K141077) (Intuitive SurgicalInc.)Stryker Infrared Light Source and SafeLight Cable (K142310)(Stryker Endoscopy)

Device Description:

The PINPOINT Endoscopic Fluorescence Imaging System (PINPOINT, PINPOINT System) is comprised of an endoscopic video processor/ illuminator (VPI) which is capable of providing visible and near-infrared illumination to a surgical laparoscope, surgical laparoscopes optimized for visible (VIS) and near-infrared (NIR) illumination and imaging, a camera head that is also optimized for visible and near-infrared imaging, and a flexible light guide cable. The following

{4}------------------------------------------------

laparoscope models are included as part of the PINPOINT System: SC9100, SC9104, SC9130, SC9131, SC9134 and SC9144. These are the major components of the PINPOINT System.

During surgical procedures, PINPOINT may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the illuminator and the resulting reflected light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

To provide NIR fluorescence imaging, PINPOINT is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the VPI, and the fluorescence response is then imaged with the camera, processed and displayed on an HD video monitor.

When used with a VIS/NIR laparoscope, PINPOINT can operate to provide illumination and imaging of both visible light and ICG fluorescence images simultaneously.

Proposed Indications for Use of the PINPOINT System:

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Summary of Technological Characteristics of the PINPOINT System and Predicate Devices

Based on the technological characteristics and fundamental scientific premise, the PINPOINT System (PC9000) has been determined to be substantially equivalent to the predicate devices. the SPY Intra-Operative Imaging System/PINPOINT (K091515), as well as the da Vinci Firefly Imaging System (K141077) and the Stryker IRF Light Source and SafeLight Cable (K142310). PINPOINT is substantially equivalent to the SPY Scope Intra-Operative Imaging System (SC8000), as both systems use the same technology and mode of imaging to provide real-time endoscopic visible and NIR fluorescence imaging during minimally invasive surgical procedures. PINPOINT is also substantially equivalent to the da Vinci Firefly Imaging System (K141077) and to the Stryker IRF Light Source and SafeLight Cable (K142310). Both systems and PINPOINT utilize very similar technology and rely on the same scientific premise – they are endoscopic video imaging systems which are used for high definition visible light and NIR fluorescence imaging during minimally invasive surgery. All of these systems utilize the same mode of imaging – visible and near infrared fluorescence imaging, with ICG as the imaging agent, used in the hospital operating room. All of these systems have the same integral components – a light source console and a light cable for outputting light, a camera control unit for processing NIR and VIS light images, a coupler attached to the laparoscope and a camera head, and a laparoscope for VIS and NIR light illumination and imaging.

{5}------------------------------------------------

Performance Data:

Non-Clinical Performance Testing of the PINPOINT System

The PINPOINT System was designed and developed by Novadag Technologies Inc. (NOVADAQ), in accordance with the applicable requirements and standards to establish performance and safety of the device. Device safety and performance were verified by tests conducted by NOVADAQ and accredited third party laboratories.

The PINPOINT System was tested in accordance with IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2007 Medical Electrical Equipment – Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility – Requirements and testing was also conducted on the PINPOINT System and test results showed that PINPOINT conforms to the applicable requirements. PINPOINT was also tested in accordance with IEC 60601-2-18:2009 Medical Electrical Equipment – Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment. and has demonstrated conformance to the standard. Conformance of PINPOINT with IEC 60825:2007 Safety of laser products - Part 1: Equipment classification and requirements was assessed by Underwriters Laboratories Inc. (UL) and showed that PINPOINT is a Class 3R laser device with internal maximum Class 4 laser radiation.

Animal testing using a porcine model validated the in vivo fluorescence imaging capability of the PINPOINT System.

Clinical Performance Data

A clinical evaluation based on a clinical literature search and reported clinical adverse event information from post-market surveillance was considered appropriate to evaluate the performance and safety of the PINPOINT System. It was concluded to objectively verify the continued clinical safety and performance of NOVADAQ'S Fluorescence Imaging devices, including the PINPOINT System.

Data presented in the clinical evaluation report provides evidence to support the safety and performance claims for the PINPOINT System. Additionally, results from a clinical study and case reports utilizing the PINPOINT System and published in the Journal of Gastrointestinal Surgerv and in Surgical Innovation and Int! J Sur Case Reports, respectively, together with findings from the most recent literature search are enclosed to support the expanded Indications for Use of PINPOINT (see section 20, Performance Testing - Clinical, for the summary of the relevant clinical literature reports). Findings demonstrated that no new concerns related to the safety and effectiveness of the device have been introduced.

Conclusions

It has been demonstrated in this 510(k) submission that the PINPOINT System, with the expanded Indications for Use, is substantially equivalent to the predicate devices discussed herein and there are no issues related to its safety or effectiveness. The additional performance data and scientific literature from a recent literature search supports the expanded Indications of the device, and further validates the well-established safety profile of NOVADAQ'S imaging systems.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.