(219 days)
The RHYTHMIA HDx™ Mapping System (with software Version 2.0) and accessories are indicated for catheter-based atrial and ventricular mapping. The mapping system allows real-time visualization of intracardiac catheters as well as display of cardiac maps in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be recorded and displayed on the system's display screen.
The RHYTHMIA HDx™ Mapping System is a catheter-based atrial and ventricular mapping system designed to display 3D anatomical and electroanatomical maps of the human heart in real-time. The mapping system is intended to be used during electrophysiology procedures for cardiac mapping and storage of cardiac electrophysiological data. In addition, based on user-defined criteria, the system is able to acquire data over multiple cardiac beats. It is capable of real-time display through construction of maps using catheter magnetic and impedance localization technology and data from intracardiac and surface electrograms.
The Signal Station and related accessories provide data connection pathways for external input/output devices (e.g. catheters and recording systems) and serve as the data conduit to the system Workstation computer and software.
The subject device is reusable capital equipment that is intended to be used in an electrophysiology laboratory by physicians fully trained in cardiac electrophysiology.
The provided FDA 510(k) summary for the RHYTHMIA HDx™ Mapping System (with software Version 2.0) describes the device and its performance testing. However, it does not include detailed acceptance criteria in a quantitative table format nor does it directly present a "study that proves the device meets the acceptance criteria" in the way one might expect for a robust AI/ML performance evaluation. The document focuses on showing substantial equivalence to a predicate device and software validation.
Here's a breakdown of the requested information based on the provided text, and where information is not available:
1. Table of Acceptance Criteria and Reported Device Performance
The document states: "The subject RHYTHMIA HDx Mapping System software Version 2.0 passed all tests in accordance with appropriate test acceptance criteria and standards." however, it does not provide a specific table of acceptance criteria or quantitative performance metrics.
The testing described is functional and safety-oriented, rather than a performance study with specific metrics like sensitivity, specificity, or accuracy.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Functional Performance | |
| Accurately create 3D maps of cardiac chambers | Confirmed that incorporation of new software features does not negatively impact this ability. |
| Support visualization of magnetic and impedance tracked catheters | Confirmed that incorporation of new software features does not negatively impact this ability. |
| New software features meet functional requirements | Confirmed. |
| System level functionality under expected and worst case use conditions | Confirmed. |
| Safety | |
| No new safety concerns associated with use | Demonstrated in animal study. |
| Software verification and validation based on 'major' level of concern is complete | Passed. |
| Device performs as intended | Demonstrated in V&V, bench, and animal studies. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated in terms of patient data or cases. The "bench performance testing" implies internal testing, but the size of data sets (e.g., number of mapping runs, number of anatomical models) is not provided. The animal study involved an unspecified number of animals to demonstrate safety, not a test set for performance metrics.
- Data Provenance: Not explicitly stated. The animal study is a pre-clinical study. For the bench testing, the origin of data is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided as the document does not describe a performance study involving expert-established ground truth for a test set. The evaluation is focused on functional verification and validation, as well as safety.
4. Adjudication method for the test set
This information is not provided as the document does not describe a performance study involving expert-established ground truth.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. The device is a mapping system, not specifically an AI diagnostic tool designed to augment human reader performance in a comparative study.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The RHYTHMIA HDx™ Mapping System is described as a "catheter-based atrial and ventricular mapping system" that provides "real-time visualization" and "display of cardiac maps." While it uses advanced software, it is not presented as a standalone AI algorithm (without human-in-the-loop performance) in the context of this document. Its function is to assist physicians during electrophysiology procedures.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the functional testing, the "ground truth" would likely be based on established engineering specifications and validated simulation models or physical phantoms. For the animal study, the ground truth for safety would be physiological measurements and observation of the animals. No mention of expert consensus, pathology, or outcomes data as ground truth for a performance study.
8. The sample size for the training set
This information is not applicable/provided. The document describes a software update for a medical device mapping system. It does not indicate that the system employs machine learning or a neural network that would require a "training set" in the conventional sense for AI/ML development. The software enhancements are described as "DirectSense™ feature provides real-time display of local impedance," "Enhanced CT and MRI capabilities," "Improved initial map setup," and "Support for multiple monitor displays," along with "Minor bug fixes." These are features unlikely to involve a separate training set.
9. How the ground truth for the training set was established
This information is not applicable/provided as there is no mention of an AI/ML training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 25, 2018
Boston Scientific Corporation Gregory Neal Senior Regulatory Affairs Specialist 125 Cambridgepark Drive Cambridge, Massachusetts 02140
Re: K173837
Trade/Device Name: RHYTHMIA HDx™ Mapping System (with software Version 2.0) Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: June 21, 2018 Received: June 25, 2018
Dear Gregory Neal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
(1-800-638-2041 or 301-796-7100).
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website
Sincerely,
(http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K173837
Device Name
RHYTHMIA HDx™ Mapping System (with software Version 2.0)
Indications for Use (Describe)
The RHYTHMIA HDx™ Mapping System (with software Version 2.0) and accessories are indicated for catheter-based atrial and ventricular mapping. The mapping system allows real-time visualization of intracardiac catheters as well as display of cardiac maps in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be recorded and displayed on the system's display screen.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY 510(K) SUMMARY COMPLYING WITH 21 CFR 807.92
I. SUBMITTER
Boston Scientific Corporation 125 Cambridgepark Drive, Suite 600 Cambridge, MA 02140
Phone: 617-218-3813 Fax: 617-218-3850
Contact Person: Gregory Neal Date Prepared: December 18, 2017
II. DEVICE
Name of Device: RHYTHMIA HDx™ Mapping System (with software Version 2.0) Common or Usual Name: Cardiac Mapping System Classification Name: Programmable diagnostic computer (21 CFR 870.1425) Regulatory Class: Class II Product Code: DQK
III. PREDICATE DEVICES
Predicate: RHYTHMIA HDx™ Mapping System Boston Scientific Corporation (K162793, S.E. 03/10/2017) Predicate: CARTO® 3 EP Navigation System version 4.2, Biosense Webster, Inc. (K133916, S.E. 07/01/2014)
IV. DEVICE DESCRIPTION
The RHYTHMIA HDx™ Mapping System is a catheter-based atrial and ventricular mapping system designed to display 3D anatomical and electroanatomical maps of the human heart in real-time. The mapping system is intended to be used during electrophysiology procedures for cardiac mapping and storage of cardiac electrophysiological data. In addition, based on user-defined criteria, the system is able to acquire data over multiple cardiac beats. It is capable of real-time display through construction of maps using catheter magnetic and impedance localization technology and data from intracardiac and surface electrograms.
The Signal Station and related accessories provide data connection pathways for external input/output devices (e.g. catheters and recording systems) and serve as the data conduit to the system Workstation computer and software.
The subject device is reusable capital equipment that is intended to be used in an electrophysiology laboratory by physicians fully trained in cardiac electrophysiology.
{4}------------------------------------------------
V. INDICATIONS FOR USE
The RHYTHMIA HDxTM Mapping System (with software Version 2.0) and accessories are indicated for catheter-based atrial and ventricular mapping system allows realtime visualization of intracardiac catheters as well as display of cardiac maps in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be recorded and displayed on the system's display screen.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Substantial equivalence of the RHYTHMIA HDx™ Mapping System with software Version 2.0 was established in terms of design features, performance testing, and indications for use as compared to the predicates RHYTHMIA HDx™ Mapping System and CARTO® 3 EP Navigation System. The subject device system has a similar design to the primary predicate device to include the same primary system components, operating principle, catheter tracking methods, and same intended use while incorporating the same feature functionality as the secondary predicate feature.
The technological differences between the subject device and the predicate RHYTHMIA HDx™ Mapping System include the following software enhancements:
- DirectSense™ feature provides real-time display of local impedance which . complements other diagnostic measures for the purpose of determining catheter location within the heart chamber.
- Enhanced CT and MRI capabilities, which allow import of pre-segmented file types. ●
- Improved initial map setup through enhanced mapping workflow. ●
- Support for multiple monitor displays. ●
- . Minor bug fixes.
VII. PERFORMANCE DATA
Testing was performed to verify functional performance, safety, and to support substantial equivalence determination. The subject RHYTHMIA HDx Mapping System software Version 2.0 passed all tests in accordance with appropriate test acceptance criteria and standards. No new questions of safety and effectiveness were raised.
Software Verification and Validation
Software verification and validation testing was conducted based on 'major' level of concern in accordance with FDA Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005.
{5}------------------------------------------------
Bench Performance Testing
- Mapping and Visualization - Confirmed that incorporation of the new software features does not negatively impact the ability to use the subject mapping system to: (1) accurately create 3D maps of the cardiac chambers of the heart; and (2) support visualization of magnetic and impedance tracked catheters.
- Software Performance Confirmed that the new software features meet the functional ● requirements.
- System Level Performance - Confirmed system level functionality under expected and worst case use conditions.
Animal Study
A pre-clinical animal study conducted with 21 CFR § 58 GLP regulations was performed and demonstrated that the subject device software does not present new questions of safety associated with its use.
VIII. CONCLUSIONS
The software verification and validation, bench performance testing and animal study demonstrated that the RHYTHMIA HDx™ Mapping System software Version 2.0 performs as intended. This data supports that the subject device is substantially equivalent to identified predicate devices.
§ 870.1425 Programmable diagnostic computer.
(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).