K Number
K173718
Date Cleared
2018-05-01

(147 days)

Product Code
Regulation Number
876.5630
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Liberty Cycler Sets are designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis.

Device Description

The Cycler Sets are sterile, single-use devices designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis. Each Cycler Set contains a cassette portion and seven (7) fluid lines. The cassette is composed of a rigid molded plastic body covered with a flexible film (membrane). The cassette contains molded features, such as fluid channels, valve domes, and pumping chambers. There are 7 fluid lines connected to the cassette body:

  • One (1) drain line (yellow clamp)
  • One (1) patient connection line with two (2) stay •safe® patient connectors (blue clamp)
  • Five (5) Dialysate Solution Lines (with Safe-Lock® connectors):
  • One (1) heater bag (red clamp)
  • One (1) last dialysate bag/'last fill option' (green clamp)
  • Three (3) additional solution bags (white clamps)
    The flow of peritoneal dialysate solution to and from the patient's peritoneal cavity is directed by interaction between the Liberty Cycler and the cassette.
AI/ML Overview

The Fresenius Liberty Cycler Set is an accessory for the Liberty Cycler, used for acute and chronic peritoneal dialysis. The 510(k) summary provides performance data and acceptance criteria for various aspects of the device.

Here's an overview of the acceptance criteria and supporting studies:

Key Takeaways:

  • No AI Component: The provided document is for a medical device accessory (cycler set) that does not contain software or electrical components, and therefore no AI. Consequently, many of the typical AI-related study components (like expert ground truth, MRMC studies, training set details) are not applicable here.
  • Focus on Physical and Biocompatibility Performance: The studies conducted focus on the physical performance of the cycler set (e.g., fill/drain accuracy, structural integrity, leak tests) and its biological safety (biocompatibility).
  • Comparison to Predicate Device: The primary method for proving substantial equivalence to a legally marketed predicate device (K043363) relies on demonstrating similar technological characteristics and meeting established performance and safety standards.

1. Table of Acceptance Criteria and Reported Device Performance

Test CategorySpecific TestAcceptance CriteriaReported Device Performance
PerformanceCycler Set fill and drain accuracyNot explicitly stated, but implies meeting functional specifications for fluid delivery and removal.Conducted (implies meeting criteria)
Load to close tubing clamps testNot explicitly stated, but implies meeting force requirements for secure clamping.Conducted (implies meeting criteria)
Load to push the stay•safe patient connector testNot explicitly stated, but implies meeting force requirements for secure connection.Conducted (implies meeting criteria)
Compression tests on cassette portsImplies meeting break-force limits to ensure structural integrity during use.Conducted (implies meeting criteria)
Structural IntegrityAbrasion testNot explicitly stated, but implies resistance to wear and tear.Conducted (implies meeting criteria)
Positive and negative pressure leak test (Cycler Set)Not explicitly stated, but implies no leaks under specified pressure ranges.Conducted (implies meeting criteria)
Bond strength testing (Tensile test)Not explicitly stated, but implies adequate strength of bonded components.Conducted (implies meeting criteria)
Positive pressure leak test (connectors/clamps)Not explicitly stated, but implies no leaks from patient connectors and tubing clamps under pressure.Conducted (implies meeting criteria)
Film seal integrity test (Endurance test)Not explicitly stated, but implies seals maintain integrity over device lifespan/during use.Conducted (implies meeting criteria)
Shipping and packaging testNot explicitly stated, but implies device maintains integrity and sterility after shipping.Conducted (implies meeting criteria)
Biological SafetySimulated use LeachablesImplies that leachable substances are within safe limits.Conducted (implies meeting criteria)
Cytotoxicity, ISO Elution Method with MEMImplies non-toxic to cells.Conducted (implies meeting criteria)
Sensitization, Guinea Pig MaximizationImplies a low risk of causing allergic sensitization.Conducted (implies meeting criteria)
Intracutaneous IrritationImplies materials do not cause significant skin irritation.Conducted (implies meeting criteria)
Material-Mediated PyrogenicityImplies materials do not cause fever.Conducted (implies meeting criteria)
Hemocompatibility, ASTM Hemolysis (Indirect) ExtractImplies materials do not cause significant red blood cell lysis.Conducted (implies meeting criteria)
OtherHuman Factors (HF) validationNot explicitly stated, but implies usability and safety are validated for intended users and environments.Conducted (implies meeting criteria)
Maintenance of sterilityNot explicitly stated, but implies sterility is maintained until point of use.Conducted (implies meeting criteria)

Note: The document generally states that testing "was conducted or leveraged to support the determination of substantial equivalence" and "implies meeting criteria" without providing specific numerical acceptance limits or raw performance results. This is common in 510(k) summaries where detailed results are part of the full submission but not always publicly accessible.

2. Sample size used for the test set and the data provenance

  • The document does not specify the exact sample sizes for each test in the test set.
  • The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature. However, these are performance tests and biocompatibility tests conducted on the physical device and its materials, typically under controlled laboratory conditions rather than with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This device is a cycler set (physical accessory), not an AI/software device that generates outputs requiring expert interpretation for ground truth. Ground truth for these tests would be established by objective measurements against engineering specifications and international standards (e.g., ISO, ASTM).

4. Adjudication method for the test set

  • Not applicable for the same reasons as point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device does not have an AI component and therefore no MRMC study was performed in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. The device does not contain a standalone algorithm.

7. The type of ground truth used

  • The ground truth for the performance and structural integrity tests are quantitative engineering specifications and established material testing standards.
  • For biocompatibility, the ground truth is based on established international standards (e.g., ISO 10993-1) for evaluating biological responses to medical device materials.

8. The sample size for the training set

  • Not applicable. This device is a physical accessory and does not involve machine learning or a training set.

9. How the ground truth for the training set was established

  • Not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 1, 2018

Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451

Re: K173718

Trade/Device Name: Liberty Cycler Set Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal Dialysis System and Accessories Regulatory Class: II Product Code: FKX Dated: March 29, 2018 Received: March 30, 2018

Dear Denise Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173718

Device Name Liberty Cycler Set

Indications for Use (Describe)

The Liberty Cycler Sets are designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue symbol consisting of three downward-pointing chevron shapes stacked vertically. To the right of the symbol are the words "FRESENIUS" and "MEDICAL CARE" stacked on top of each other, also in blue. The logo is simple and clean, with a focus on the company name and a recognizable symbol.

510(K) SUMMARY 5.

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) Summary is provided in conformance with 21 CFR §807.92.

5.1. Submitter's Information

Name:Fresenius Medical Care Renal Therapies Group, LLC
Address:920 Winter Street
Waltham, MA
02451-1457
Phone:(781) 699-4479
Fax:(781) 699-9635
Contact Person:Denise Oppermann
Senior Director Regulatory Affairs – Devices
Preparation Date:December 4, 2017

5.2. Device Name

Trade Name:Liberty Cycler Set
Common Name:Cycler Set
Regulation Name:Peritoneal dialysis system and accessories
Regulatory Class:Class II per 21 CFR §876.5630
Product Code:FKX
Product Code Name:System, Peritoneal, Automatic Delivery
Classification Panel:Gastroenterology/Urology

5.3. Legally Marketed Predicate Device

The legally marketed predicate device is the Fresenius Liberty Cycler and Disposable Cycler Sets (K043363). This device has not been subject to a design related recall.

The Liberty PDx Cycler Set (K141145) is used as a reference device.

5.4. Device Description

5.4.1. Device Identification

The Liberty Cycler Set (hereinafter referred to as "Cycler Set") is available in three (3) configurations. The sets differ only in the length of the patient and drain lines or in the number of patient connectors (Table 1).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that looks like three chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top of "MEDICAL CARE".

Cycler Set Product CodesProduct Name
050-87212Liberty Cycler Set, Two patient connectors
050-87215Liberty Cycler Set, One patient connector
050-87216Liberty Cycler Set, One patient connector withextended patient and drain lines

Table 1: Liberty Cycler Sets

5.4.2. Device Characteristics

The Cycler Sets are single-use Class II devices designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis. The Cycler Sets are provided sterile and nonpyrogenic. The Cycler Sets are sterilized using ethylene oxide (EO).

5.4.3. Environment of Use

The Cycler Sets are used in both healthcare and home environments.

5.4.4. Brief Written Description of the Device

The Cycler Sets are sterile, single-use devices designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis. Each Cycler Set contains a cassette portion and seven (7) fluid lines. The cassette is composed of a rigid molded plastic body covered with a flexible film (membrane). The cassette contains molded features, such as fluid channels, valve domes, and pumping chambers. There are 7 fluid lines connected to the cassette body:

  • One (1) drain line (yellow clamp) -
  • One (1) patient connection line with two (2) stay •safe® patient connectors (blue clamp)
  • -Five (5) Dialysate Solution Lines (with Safe-Lock® connectors):
    • One (1) heater bag (red clamp) -
    • One (1) last dialysate bag/'last fill option' (green clamp) ।
    • -Three (3) additional solution bags (white clamps)

The flow of peritoneal dialysate solution to and from the patient's peritoneal cavity is directed by interaction between the Liberty Cycler and the cassette.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue geometric shape on the left and the text "FRESENIUS MEDICAL CARE" on the right. The geometric shape is made up of three chevrons pointing downwards. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

5.4.5. Materials of Use

The Cycler Sets are classified as externally communicating, blood path indirect, prolonged contact (>24 hours to 30 days) duration. Class II (Category B) devices in accordance with FDA guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (16 June 2016).

The 3 Cycler Sets designed for use with the Liberty Cycler are constructed from materials identical to the predicate device with the exception of the new film.

Materials used in the manufacture of the Cycler Sets include:

  • . Polypropylene (PP)
  • . Polyvinyl chloride (PVC)
  • Polycarbonate (PC)
  • . High-density polyethylene (HDPE)
  • . Acrylic polymer
  • . Silicone
  • Thermoplastic elastomer copolymer

Key Performance Characteristics 5.4.6.

The Cycler Sets are used as accessories to the Liberty Cycler. When used in conjunction with the Liberty Cycler, they provide a sterile fluid path with secure connections (stay safe-Lock®, and Luer lock) to alternately draw the prescribed dialysate solution to and from the patient's peritoneal cavity.

5.5. Intended Use

The Liberty Cycler Sets are intended to be used with the Liberty Cycler.

5.6. Indications for Use

The Liberty Cycler Sets are designed to operate with the Liberty Cycler to perform acute and chronic peritoneal dialysis.

5.7. Comparison of Technological Characteristics with the Predicate Device

The following technological characteristics of the Cycler Sets are equivalent to the predicate Liberty Cycler Set (K043363).

  • . Intended use
  • Principle of operation ●
  • Design characteristics

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue emblem on the left and the text "FRESENIUS MEDICAL CARE" on the right. The emblem is made up of three downward-pointing chevron shapes stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

  • Sterilization method, packaging, and sterility label claims

5.8. Performance Data

Testing was either conducted or leveraged to support the determination of substantial equivalence.

  • . Performance
    • Cycler Set fill and drain accuracy ।
    • Load to close tubing clamps test |
    • Load to push the stay safe patient connector test |
    • -Compression tests to evaluate the break-force on the cassette ports
  • Structural integrity ●
    • । Abrasion test
    • Positive and negative pressure leak test for the Cycler Set ।
    • Bond strength testing (Tensile test) -
    • -Positive pressure leak test for stay•safe patient connector and tubing clamps
    • Film seal integrity test (Endurance test) -
    • Shipping and packaging test -
  • Biological safety (Biocompatibility) ●
  • Maintenance of sterility
  • Human Factors (HF) validation

5.8.1. Biocompatibility Testing

Testing was conducted to support the biological safety of the Cycler Sets.

  • Simulated use Leachables
  • Cytotoxicity, ISO Elution Method with MEM ●
  • . Sensitization, Guinea Pig Maximization
  • . Intracutaneous Irritation
  • Material-Mediated Pyrogenicity ●
  • Hemocompatibility, ASTM Hemolysis (Indirect) Extract ●

A toxicological risk assessment was also performed.

5.8.2. Electrical Safety and Electromagnetic Compatibility (EMC)

Not applicable. The Cycler Sets do not contain electrical components.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the words "FRESENIUS MEDICAL CARE" in blue on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The words "FRESENIUS" and "MEDICAL CARE" are stacked on top of each other.

5.8.3. Software Verification and Validation Testing

Not applicable. The Cycler Sets do not contain software.

5.8.4. Mechanical and Acoustic Testing

No mechanical or acoustic tests were performed.

Animal Studies 5.8.5.

No animal studies were performed.

Clinical Studies 5.8.6.

No clinical studies were performed.

5.9. Conclusion

The intended use, principle of operation, design characteristics, and sterilization method of the Liberty Cycler Sets are the same as those of the predicate device. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the Liberty Cycler Sets are safe and effective for their intended use.

§ 876.5630 Peritoneal dialysis system and accessories.

(a)
Identification. (1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments.
(3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles.
(4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
(b)
Classification. Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.