K Number
K173670
Manufacturer
Date Cleared
2018-02-12

(74 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use the Celerity 20 Steam Incubator to incubate and automatically read Celerity 20 Steam Biological Indicators at 57 °C for a fluorescent result within 20 minutes.

Device Description

The Celerity 20 Steam Incubator (Incubator) is an incubator/reader designed for use specifically with the Celerity 20 Steam Biological Indicator (BI), subject of a concurrent 510(k) currently under review. The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the BI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

AI/ML Overview

The provided text describes the Celerity 20 Steam Incubator, which is an incubator/reader for biological indicators, not an AI-powered diagnostic device. Therefore, many of the requested fields are not applicable or cannot be extracted from this type of regulatory submission (510(k)). This document focuses on the incubator's performance in reading biological indicators, not on interpreting medical images or data.

However, I can extract the relevant information regarding the acceptance criteria and the studies performed for this specific device.


1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Maintenance of Incubation TemperatureMaintain 55-60 °C for a minimum of 20 minutes (incubation time of Celerity 20 Steam BI)PASS
Qualification testing with Celerity 20 Steam BI (Fluorescent Read Alignment)20-minute fluorescent read meets >97% alignment with 7-day growth results per FDA guidance on reduced incubation time.PASS
Qualification testing with Celerity 20 Steam BI (Full Cycle Exposure)Pass testing with BI exposed to full cycle exposure and negative growth result in incubator.PASS
Qualification testing with Celerity 20 Steam BI (Abbreviated Cycle Exposure)Fail testing with BI exposed in abbreviated cycle exposure and positive result in incubator.PASS
Alarm, LED and Print function TestDemonstrate proper function of alarms, LED and print outputs.PASS

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes (number of Biological Indicators or test runs) used for each qualification test. It generally refers to "testing" performed for "Qualification testing with Celerity 20 Steam BI".

The data provenance (country of origin, retrospective/prospective) is not mentioned.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device is an incubator/reader for biological indicators. The "ground truth" for the biological indicators is established by their response to sterilization cycles and subsequent growth, determined by laboratory methods and compared to 7-day growth results. There are no human experts interpreting images or complex data in the way a radiologist would.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a human-in-the-loop diagnostic device requiring adjudication of expert interpretations. The "ground truth" for the biological indicators is determined by objective biological growth or lack thereof.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-powered diagnostic device, and no human readers are involved in interpreting results from this incubator. Its function is to automatically read the fluorescent signal from a biological indicator.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is inherently "standalone" in its reading function as it automatically detects fluorescence. The "Qualification testing with Celerity 20 Steam BI" section demonstrates its standalone performance in reading the biological indicators.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the biological indicator (BI) performance is based on:

  • 7-day growth results: For the "20-minute fluorescent read meets >97% alignment with 7-day growth results." This implies comparison to a gold standard biological growth assay over a longer incubation period.
  • Biological response to sterilization cycles: BIs exposed to "full cycle exposure" should show negative growth (no fluorescence), and BIs exposed to "abbreviated cycle exposure" should show positive growth (fluorescence).

8. The sample size for the training set

Not applicable. This device is not an AI/machine learning algorithm that requires a "training set" in the conventional sense. It is a hardware device with embedded logic for temperature control and fluorescence detection.


9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI/machine learning algorithm for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 12, 2018

STERIS Corporation Anthony Piotrkowski Senior Manager, Regulatory Affairs 5976 Heisley Road Mentor, Ohio 44060

Re: K173670

Trade/Device Name: CELERITY 20 Steam Incubator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 29, 2017 Received: November 30, 2017

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K173670

Device Name Celerity 20 Steam Incubator

Indications for Use (Describe) Use the Celerity 20 Steam Incubate and automatically read Celerity 20 Steam Biological Indicators at 57 °C for a fluorescent result within 20 minutes

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines in blue.

510(k) Summary For K173670 Celerity 20 Steam Incubator

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Manufacturing Facility

STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896

Contact: Anthony Piotrkowski

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tony_piotrkowski@steris.com

Submission Date: January 17, 2018

Premarket Notification Number: K173670

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION K173670 Celerity 20 Steam Incubator

Device Name 1.

Trade Name:Celerity 20 Steam Incubator
Common/usual Name:Incubator/Reader (accessory to Biological Indicator)
Device Classification:Class II
Classification Name:Indicator, Biological Sterilization Process(21 CFR 880.2800, FRC)

2. Predicate Device

K171587 - VERIFY Incubator for Assert VH2O2 Self-Contained Biological Indicator, (since renamed Celerity 20 HP Incubator)

3. Description of Device

The Celerity 20 Steam Incubator (Incubator) is an incubator/reader designed for use specifically with the Celerity 20 Steam Biological Indicator (BI), subject of a concurrent 510(k) currently under review. The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the BI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

4. Intended Use/ Indications for Use

Use the Celerity 20 Steam Incubator to incubate and automatically read Celerity 20 Steam Biological Indicators at 57 °C for a fluorescent result within 20 minutes.

5. Summary of Technical Characteristics

A comparison of technical characteristics are summarized in Table 5-1.

{5}------------------------------------------------

STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION K173670 Celerity 20 Steam Incubator

FeatureCelerity Steam Incubator(proposed)VERIFY Incubator(K71587) PredicateComparison
Intended UseUse the Celerity 20Steam Incubator toincubate andautomatically readCelerity 20 SteamBiological Indicatorsat 57 °C for afluorescent resultwithin 20 minutes.Use the VERIFYIncubator for AssertVH2O2 SelfContained BiologicalIndicators (Incubator)to incubate andautomatically readVERIFY AssertVH2O2 Self-Contained BiologicalIndicators at 57 °C fora fluorescent resultIdentical except forthe BI they are meantto read.Testing with theCelerity 20 Steam BI,included in thesubmission,demonstrates properperformance.
Basis ofReadoutPhotodiode detectsfluorescence producedby enzymatic activitythat results fromgrowing biologicalindicator organismswithin 20 minutes.Photodiode detectsfluorescence producedby enzymatic activitythat results fromgrowing biologicalindicator organismsSame
IncubationTemperatureRange55 - 60 °C55 - 60 °CSame
VoltageRange90 to 264 VAC with12 VDC conversion.90 to 264 VAC with12 VDC conversion.Same

Table 5-1 Summary of Incubator Physical Description and Technological Properties

6. Summary of Nonclinical Tests

Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below.

Table 5-2. Summary of Non-clinical Testing

TestAcceptance CriteriaConclusion
Maintenance ofIncubationTemperatureMaintain 55-60 °C for a minimum of 20 minutes(incubation time of Celerity 20 Steam BI)PASS
Qualification testingwith Celerity 20Steam BI20-minute fluorescent read meets >97%alignment with 7-day growth results per FDAguidance on reduced incubation time.PASS

{6}------------------------------------------------

STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION K173670 Celerity 20 Steam Incubator

TestAcceptance CriteriaConclusion
Qualification testingwith Celerity 20Steam BI• Pass testing with BI exposed to full cycleexposure and negative growth result inincubator• Fail testing with BI exposed in abbreviatedcycle exposure and positive result in incubatorPASSPASS
Alarm, LED andPrint function TestDemonstrate proper function of alarms, LED andprint outputsPASS

7. Conclusion

The Celerity 20 Steam Incubator has met the established performance criteria. The conclusions drawn from the nonclinical tests performed demonstrate the subject device is as safe, as effective, and performs as well or better than the legally marketed predicate device VERIFY Incubator for Assert VH2O2 Self-Contained Biological Indicator cleared in K171587 (21 CFR 880.2800, Product code FRC).

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).