K Number
K173471
Date Cleared
2018-01-08

(61 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-raginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel, urology and transesophageal (Cardiac) exams.

Device Description

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-mode, M-mode, PW-mode, CW mode, Color-mode, Color m-Mode, Power/Dirpower mode, TDI mode, 3D/4D mode, Elastography or the combined mode (i.e. B/M-mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array.

AI/ML Overview

The provided text is a 510(k) summary for the Shenzhen Mindray Bio-medical Electronics Co., LTD DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System. This document focuses on demonstrating substantial equivalence to predicate devices and adherence to relevant standards rather than detailing acceptance criteria and performance data for a novel algorithm or AI-driven feature.

Based on the provided information, the device is a general-purpose diagnostic ultrasound system, and the submission does not describe a study involving specific acceptance criteria for a novel AI-driven feature or algorithm. The document explicitly states "Clinical Tests: Not Applicable." and relies on non-clinical tests to demonstrate substantial equivalence to predicate devices.

Therefore, the requested information regarding algorithm-specific acceptance criteria, reported performance, sample sizes for test/training sets, data provenance, ground truth establishment, expert adjudication, or MRMC comparative effectiveness studies with AI assistance is not present in the provided text.

The document indicates that the device meets safety and performance standards through non-clinical testing for compliance with:

  • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, c1:2009/(r)2012 and a2:2010/(r)2012 (Medical Electrical Equipment - General Requirements for Basic Safety and Essential Performance)
  • IEC 60601-1-2 Edition 3: 2007-03 (Electromagnetic Compatibility)
  • IEC 60601-2-37 Edition 2.0 2007 (Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment)
  • AAMI / ANSI / IEC 62304:2006 (Medical Device Software - Software Life Cycle Processes)
  • ISO 14971 Second edition 2007-03-01 (Medical Devices - Application of Risk Management)
  • NEMA UD 2-2004 (R2009) (Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment)
  • AAMI / ANSI / ISO 10993-1:2009/(R)2013 (Biological Evaluation of Medical Devices)

The performance of the device is implicitly expected to be "substantially equivalent" to the predicate devices (Mindray DC-8 (K170277), Resona 7 (K171233), DC-70 (K163690), and M9 (K171034)) based on shared technology, intended uses, basic operating modes, materials, acoustic power levels, and compliance with safety standards.

In summary, the provided text does not contain detailed information on acceptance criteria and specific study results for a novel AI or algorithm performance, as it is a 510(k) submission focused on demonstrating substantial equivalence through adherence to general safety and performance standards for an ultrasound system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shenzhen Mindray Bio-medical Electronics Co., LTD. % Yang Zhaohui Engineer of Technical Regulation Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan ShenZhen, GangDong 518057 P.R. CHINA

January 8, 2018

Re: K173471

Trade/Device Name: DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 1, 2017 Received: November 8, 2017

Dear Yang Zhaohui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Yang Zhaohui

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D'Hara For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K173471

Device Name

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System

Indications for Use (Describe)

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-raginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel, urology and transesophageal (Cardiac) exams.

Type of Use (Select one or both, as applicable)
✔ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

PSC Publishing Services (301) 443-6740

EF

{3}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application Mode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPPNote 1, 2, 3, 4,6,7
AbdominalPPPPPPPNote 1, 2, 3, 4, 6, 7, 10, 11, 12
Intra-operative (Specify*)PPPPPPNote 1,2,4
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2, 3,4,6,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPPNote 1,2, 4,6,7
Adult CephalicPPPPPPPNote 1, 2,4, 6,7
Trans-rectalPPPPPPNote 1, 2, 3,4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 3,4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPPPPNote 1, 2, 4,6,7
Musculo-skeletal (Superficial)PPPPPPNote 1,2, 4,7
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6,7,9
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPPPPNote 1, 5,6
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
** Small organ-breast, thyroid, testes.
***Other use includes Urology.
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:C5-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2, 4,6,7
AbdominalPPPPPPNote 1, 2, 4, 6, 7,10, 11, 12
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,6,7
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes--B+M PW+B Color+B Power + B PW+Color+B Power + PW +B.*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L14-5WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPNote 1,2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1,2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletalNote 1,2, 4,7
(Conventional)PPPPPP
Musculo-skeletalNote 1,2, 4,7
(Superficial)PPPPPP
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPbPPNote 1,2, 4,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** * * * Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L9-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPNote 1, 2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1, 2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletalNote 1, 2, 4,7
(Conventional)PPPPPP
Musculo-skeletalNote 1, 2, 4,7
(Superficial)PPPPPP
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPPPPNote 1, 2, 4,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** * * * Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:V11-3HE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote 1, 2, 4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vessel
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:D8-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3, 4, 6,7
AbdominalPPPPPPNote 1, 3, 4, 6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:D7-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3, 4, 6
AbdominalPPPPPPNote 1, 3, 4, 6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K163690); E=added under Appendix E
Additional comments: Combined modes--B+M, PW+B, Color + B, Power + B, PW+Color+B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:C7-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined (specify)Other (specify)
1 Only)BMPWDCWDColor DopplerAmplitude Doppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetalPPPPPPNote 1, 2, 4,6,7
& OtherAbdominalPPPPPPNote 1, 2, 4,6,7,10
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPPPPNote 1, 2, 4,6,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
System:
DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L12-3EDiagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Intended Use:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)BMPWDCWDColorAmplitudeCombined Other (specify)
1 Only)DopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPNote 1, 2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPbNote 1, 2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletalNote 1, 2, 4,7
(Conventional)PPPPPP
Musculo-skeletalNote 1, 2, 4,7
PPPPPP
(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPPPPNote 1, 2, 4,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
** Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:DE11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationSpecific (Track 1 & 3)Mode of Operation
General (Track1 Only)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3,4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 3,4,6,7,8
Trans-vaginalPPPPPPNote 1, 3,4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:SP5-1E
Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPPNote 1, 2,4, 6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4, 6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicPPPPPPPNote 1, 2,4, 6,7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6,7,9
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vessel
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、PW +Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:SC5-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetalPPPPPPNote 1, 2, 4,6,7
& OtherAbdominalPPPPPPNote 1, 2,4,6,7,10,11,12
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalTrans-vaginalTrans-urethralTrans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,6,7
Musculo-skeletal(Superficial)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPPPPNote 1, 2, 4,6,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix EAdditional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.*Intraoperative includes abdominal, thoracic, and vascular**Small organ-breast, thyroid, testes.***Other use includes Urology.Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.Note 2: Smart3DNote 3:4D(Real-time 3D)Note 4: iScapeNote 5: TDINote 6: Color MNote 7: Biopsy GuidanceNote 8: ElastographyNote 9: Contrast imaging(contrast agent for LVO)Note 10: Contrast imaging (Contrast agent for Liver)Note 11: STENote 12: STQ
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:D6-2NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetalPPPPPPNote1,3, 4,6
& OtherAbdominalPPPPPPNote1,3, 4,6
Intra-operative (Specify*
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote1,3, 4,6
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)Musculo-skeletal(Conventional)Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)Intra-cardiac
Peripheral
Peripheral vesselOther (Specify***)
vessel
N=new indication; P=previously cleared by FDA(K163690); E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、Power + PW+B.*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:C6-2GE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of OperationOther (specify)
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude Doppler
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote1,2,4,6,7
AbdominalPPPPPPNote1,2,4,6,7,10
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote1,2,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K171233); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{17}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:C11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPNote 1, 2, 4,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPbNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal CephalicPPPPPPNote 1, 2, 4,6,7
Adult Cephalic
Trans-rectal
Trans-vaginalTrans-urethralTrans-esoph. (non-Card.)Musculo-skeletal(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPNote 1, 2, 4,6,7
Cardiac PediatricPPPPPPNote 1, 2, 4,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vesselPPPPPPNote 1, 2, 4,6,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{18}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L14-6NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1,2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2, 4,7
Musculo-skeletal(Superficial)PPPPPPNote 1,2, 4,7
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{19}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L14-6WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1,2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2, 4,7
Musculo-skeletal(Superficial)PPPPPPNote 1,2, 4,7
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{20}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:V11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2, 4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STENote 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{21}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:V11-3BE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)ColorAmplitudeCombined Other (specify)
1 Only)BMPWDCWDDopplerDoppler(specify)
OphthalmicOphthalmic
Fetal ImagingFetalPPPPPPPNote 1, 2, 4,6,7
& OtherAbdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPPNote 1, 2, 4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vessel
vesselOther (Specify***)PPPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
*** Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{22}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:DE10-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3,4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 3,4,6,7,8
Trans-vaginalPPPPPPNote 1, 3,4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:P4-2NE
Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPPNote 1, 2,4,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicPPPPPPPNote 1, 2,4,6,7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6,7,9
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:P7-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (TrackSpecific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
1 Only)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPPNote 1, 2,4,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4,6
Small Organ (Specify**)
Neonatal CephalicPPPPPPPNote 1, 2,4,6
Adult CephalicPPPPPPPNote 1, 2,4,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1, 2,4,6
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M 、PW+B 、Color + B 、Power + B 、PW+Color+B 、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{23}------------------------------------------------

{24}------------------------------------------------

{25}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:P10-4E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPPNote 1, 2,4,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4,6
Small Organ (Specify**)
Neonatal CephalicPPPPPPPNote 1, 2,4,6
Adult CephalicPPPPPPPNote 1, 2,4,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:CB10-4E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application Mode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPPNote 1, 2, 4,6,7
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:L16-4HE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationSpecific (Track 1 & 3)Mode of OperationBMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
General (Track 1 Only)
OphthalmicOphthalmic
Fetal Imaging & OtherFetal
AbdominalPPPPPPNote1,2,4
Intra-operative (Specify*)PPPPPPNote1,2,4
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote1,2,4
Small Organ (Specify**)PPPPPPNote1,2,4
Neonatal CephalicPPPPPPNote1,2,4
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPPPNote1,2,4
Musculo-skeletal (Superficial)PPPPPPNote1,2,4
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPNote1,2,4
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDINote 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:P7-3TE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPPPPNote 1, 5,6
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:CW2s
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricP
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System
Transducer:CW5s
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricP
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K170277); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Concurrence of CDRH, Office of Device Evaluation(ODE)

{26}------------------------------------------------

{27}------------------------------------------------

{28}------------------------------------------------

{29}------------------------------------------------

{30}------------------------------------------------

{31}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

1. Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Tel: +86 755 8188 5640 Fax: +86 755 2658 2680

Contact Person:

Yang ZhaoHui Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

November 1, 2017 Date Prepared:

  1. Device Name: DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System

Classification

Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

3. Device Description:

{32}------------------------------------------------

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-mode, M-mode, PW-mode, CW mode, Color-mode, Color m-Mode, Power/Dirpower mode, TDI mode, 3D/4D mode, Elastography or the combined mode (i.e. B/M-mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array.

4. Intended Use:

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel, urology and transesophageal (Cardiac) exams.

6. Comparison with Predicate Devices:

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices:

Predicate DeviceManufacturerModel510(k) Control Number
1. Primary predicate deviceMindrayDC-8K170277
2. Reference deviceMindrayResona 7K171233
3. Reference deviceMindrayDC-70K163690
4. Reference deviceMindrayM9K171034

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System employs the same technology as the predicate devices. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations. The subject device also has the same intended uses and basic operating modes as the predicate devices.

Subject device DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 has the same intended .

{33}------------------------------------------------

uses as the predicated device DC-8(K170277)

  • . The materials of probes and Needle-guided brackets of DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 are the same to the probe of predicate device.
  • . The acoustic power levels of DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 are below the limits of FDA, which is the same as the predicated device DC-8(K170277).
  • DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 is designed in compliance with the FDA recognized electrical and physical safety standards, which are the same as the predicated device DC-8(K170277).

7. Non-clinical Tests:

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical safety standards.

Non-clinical tests relied on in this premarket notification submission for a determination of substantial equivalence include testing showing compliance with the following standards:

  • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, c1:2009/(r)2012 and a2:2010/(r)2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (iec 60601-1:2005, mod).
  • IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
  • . IEC 60601-2-37 Edition 2.0 2007, medical electrical equipment - part 2-37: particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
  • . AAMI / ANSI / IEC 62304:2006, medical device software - software life cycle processes.
  • ISO 14971 Second edition 2007-03-01, medical devices application of risk . management to medical devices.

{34}------------------------------------------------

  • NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic . ultrasound equipment revision 3.
  • AAMI / ANSI / ISO 10993-1:2009/(R)2013, biological evaluation of medical I devices - part 1: evaluation and testing within a risk management process.

8. Clinical Tests:

Not Applicable.

Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.