K Number
K173362
Date Cleared
2018-04-03

(159 days)

Product Code
Regulation Number
888.3045
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The DSM Biomedical Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The DSM Biomedical Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.

Device Description

DSM Biomedical Calcium Phosphate Cement as an injectable, fast self-setting bone substitute. DSM Biomedical Calcium Phosphate Cement is composed of calcium phosphate which converts to hydroxyapatite in vivo, and bovine collagen powder. The device can also be used to augment provisional hardware to help support bone fragments during the surgical procedure. The cement is provided in a powder form packaged in a mixing syringe. The mixing syringe allows for the combination of saline, the patient's blood, or the patient's bone marrow at the required powder-to-liquid ratio. DSM Biomedical Calcium Phosphate Cement is supplied sterile by gamma irradiation and is non-pyrogenic.

AI/ML Overview

The provided text is a 510(k) summary for the DSM Biomedical Calcium Phosphate Cement. It states that the device is substantially equivalent to a predicate device based on various criteria, including performance data. However, the document does not contain the level of detail requested in the prompt regarding acceptance criteria, specific reported device performance values, sample sizes for test or training sets, expert qualifications, or adjudication methods for studies.

The document primarily focuses on establishing substantial equivalence to a predicate device (Stryker® Injectable Cement K060061) through comparisons of:

  • Indications for Use
  • Material Composition
  • Technological Characteristics
  • Performance Characteristics

It mentions that "material characterization, including chemical and material physical characterization, bench testing, biocompatibility testing, viral inactivation, and an animal study have been conducted to evaluate the performance characteristics and biological safety." It also states these were completed in accordance with FDA Guidance Document, "Resorbable Calcium Salt Bone Void Filler Device and ASTM F1185 Standard Specification for Composition of Hydroxylapatite for Surgical Implants."

An "ovine bilateral femoral defect animal study" was performed to evaluate bone healing.

Therefore, many of the specific details requested cannot be extracted from this document.

However, I can provide a summary of what is mentioned concerning performance and the basis for substantial equivalence:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with quantitative acceptance criteria and corresponding reported device performance values. It states that tests were conducted according to FDA guidance and ASTM standards, implying that the device met the requirements of those standards.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not explicitly stated for any of the tests.
  • Data Provenance: The animal study was an "ovine bilateral femoral defect animal study." This suggests an animal model study, not human subject data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as this is related to a medical image analysis or diagnostic device, which this bone void filler is not. The ground truth for performance would be based on the physical, chemical, and biological measurements from the lab and animal studies.

4. Adjudication method for the test set

Not applicable for this type of device and study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/imaging device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI/imaging device.

7. The type of ground truth used

For the performance of the material:

  • Physical and chemical characterization data (e.g., setting time, mechanical strength, composition, resorption rate – though specific values are not listed).
  • Biocompatibility test results (cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, hemocompatibility, subacute toxicity, implantation).
  • Viral inactivation study results.
  • Residual chemical assessment results.
  • Bone healing evaluation from the ovine animal study (comparison to predicate).

8. The sample size for the training set

Not applicable, as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/machine learning device.


Summary of Device Performance (as described in the document, without specific values):

CharacteristicDSM Biomedical Calcium Phosphate Cement (Proposed)Assessment in Document
Material CharacterizationCalcium phosphate powder with bovine collagenCompleted in accordance with FDA Guidance and ASTM F1185.
Physical CharacteristicsInjectable, sculptable, fast setting, converts to hydroxyapatiteCompared to predicate device, deemed similar.
BiocompatibilityBiocompatibleTesting completed per ISO 10993-1. Results indicate acceptable biocompatibility.
Viral InactivationConductedStudy performed.
Animal StudyBone healingEvaluated bone healing in an ovine bilateral femoral defect model compared to the predicate device.
SterilizationSterile, SAL 10-6Confirmed (stated in table).
ReusableSingle Use DeviceConfirmed (stated in table).

The document concludes that "The bench, animal, and biocompatibility testing demonstrates that DSM Biomedical Calcium Phosphate Cement is substantially equivalent to the predicate device." This implies that the device met the performance requirements necessary to establish substantial equivalence based on the completed tests.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.

Kensey Nash Corporation dba DSM Biomedical Susan Pileggi Senior Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19431

April 3, 2018

Re: K173362

Trade/Device Name: DSM Biomedical Calcium Phosphate Cement Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, OIS Dated: February 26, 2018 Received: February 28, 2018

Dear Ms. Pileggi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K173362

Device Name DSM Biomedical Calcium Phosphate Cement

Indications for Use (Describe)

DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The DSM Biomedical Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The DSM Biomedical Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping, translucent shapes in various colors, including blue, green, orange, and purple, creating a sense of dynamism and innovation.

510(k) Summary

Submitted By:DSM Biomedical735 Pennsylvania DriveExton, PA 19341
Contact Person:Susan Pileggisusan.pileggi@dsm.com(P) 484-713-2100(F) 484-713-2903
Date Prepared:February 23, 2018

Device:

Trade Name:DSM Biomedical Calcium Phosphate Cement
Common/Usual Name:Bone Void Filler
Classification Name:Resorbable Calcium Salt Bone Void Filler Device
Classification Regulation:21 CFR 880.3045
Device Class:II
Device Code:MQV: Filler, Bone Void, Calcium Compound
OIS: Calcium Salt Bone Void Filler, Drillable, Non-screw Augmentation
Advisory Panel:Orthopedic

Predicate Device: Stryker® Injectable Cement (K060061) [Stryker® CMF]

Device Description

DSM Biomedical Calcium Phosphate Cement as an injectable, fast self-setting bone substitute. DSM Biomedical Calcium Phosphate Cement is composed of calcium phosphate which converts to hydroxyapatite in vivo, and bovine collagen powder. The device can also be used to augment provisional hardware to help support bone fragments during the surgical procedure. The cement is provided in a powder form packaged in a mixing syringe. The mixing syringe allows for the combination of saline, the patient's blood, or the patient's bone marrow at the required powder-to-liquid ratio. DSM Biomedical Calcium Phosphate Cement is supplied sterile by gamma irradiation and is non-pyrogenic.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping curved shapes in various colors, including blue, green, purple, and orange. The overall design is modern and corporate.

Indications For Use:

DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the sketal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.

Basis for Substantial Equivalence:

The DSM Biomedical Calcium Phosphate Cement is substantially equivalent in terms of indications for use, material composition, technological characteristics and performance characteristics to that of the predicate device, Stryker® Injectable Cement (K060061).

DSM Biomedical Calcium Phosphate Cement and the predicate device have comparable indications for use. Both are calcium phosphate powders that when mixed with a mixing liquid form a fast setting cement. The addition of the collagen to the subject device does not raise questions of substantial equivalence as the safety of the collagen has been shown through biocompatibility, an animal study, and a viral inactivation The subject and predicate devices are both injectable, and fast setting calcium study. phosphate cement that converts to hydroxyapatite over time. Both the subject and predicate devices are packaged with accessories to aid in the mixing and delivery of the subject and predicate devices are provided sterile and are for single use only. The subject and predicate devices are biocompatible and non-pyrogenic.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the DSM company logo. The logo consists of a colorful abstract design on the left and the letters "DSM" in blue on the right. The abstract design is made up of overlapping shapes in various colors, including blue, green, yellow, and purple. The letters "DSM" are in a bold, sans-serif font.

CharactesticDSM Biomedical Calcium PhosphateCement (Proposed)Stryker® Injectable Cement(K060061)
Indications ForUseDSM Biomedical Calcium PhosphateCement is indicated to fill bony voids orgaps of the skeletal system (i.e.extremities and pelvis). These defectsmay be surgically created or osseousdefects created from traumatic injuryto the bone. The Calcium PhosphateCement is indicated only for bony voidsor gaps that are not intrinsic to thestability of the bony structure. TheCalcium Phosphate Cement cured in situprovides an open void/gap filler thatcan augment provisional hardware (e.g.K-Wires, plates, screws) to help supportbone fragments during the surgicalprocedure. The cured cement acts onlyas a temporary support media and is notintended to provide structural supportduring the healing process. TheCalcium Phosphate Cement resorbs andis replaced by bone during the healingprocess.Stryker Injectable Cement is a self-settingcalcium phosphate cement indicated to fillbony voids or gaps of the skeletal system(i.e. extremities, craniofacial, spine, andpelvis.) These defects may be surgicallycreated or osseous defects created fromtraumatic injury to the bone. The StrykerInjectable Cement is indicated only forbony voids or gaps that are not intrinsic tothe stability of the bony structure. StrykerInjectable Cement cured in situ provides an openvoid/gap filler that can augmentprovisional hardware (e.g. K-Wires,plates,screws) to help support bone fragmentsduring the surgical procedure. The curedcement acts only as a temporary supportmedia and is not intended to providestructural support during the healingprocess.
MaterialCalcium phosphate powder with bovinecollagen mixed with saline, patient'sblood, or patient's bone marrow to formhydroxyapatiteCalcium phosphate powder mixed withmixing liquid to form hydroxyapatite
FormInjectable, sculptable, and fast settingcalcium phosphate cement thatconverts to hydroxyapatiteInjectable, sculptable, and fast settingcalcium phosphate cement that converts tohydroxyapatite
PackagingPowder prepackaged in a mixing anddelivery syringe, packaged withaccessories to aid in the mixing anddelivery of cementCalcium phosphate powder in a bowl,packaged with accessories to aid in themixing and delivery of the cement
SterilizationSterile, SAL 10-6Sterile, SAL 10-6
ReusableSingle Use DeviceSingle Use Device
BiocompatibleYesYes

Performance Data:

Material characterization, including chemical and material physical characterization, bench testing, biocompatibility testing, viral inactivation, and an animal study have been conducted to evaluate the performance characteristics and biological safety of DSM Biomedical Calcium Phosphate Cement.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping shapes in various colors, including blue, green, purple, and orange, forming a circular design with a hexagonal shape in the center.

Material characterization and performance testing of the DSM Biomedical Calcium Phosphate Cement was completed in accordance with the FDA Guidance Document, Resorbable Calcium Salt Bone Void Filler Device and ASTM F1185 Standard Specification for Composition of Hydroxylapatite for Surgical Implants.

An ovine bilateral femoral defect animal study was performed to evaluate bone healing following treatment with the DSM Biomedical Calcium Phosphate Cement compared to the predicate device.

Biocompatibility testing was completed in accordance with the requirements of ISO 10993-1: 2009, Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing within a Risk Management Process for a permanent implant device with tissue/bone contact. Testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, hemocompatibility, subacute toxicity, and implantation. Other safety testing included a viral inactivation study and residual chemical assessment. Results indicate that the device's biocompatibility profile is acceptable.

Substantial Equivalence:

DSM Biomedical Calcium Phosphate Cement has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device. The minor technological differences that exist between the DSM Biomedical Calcium Phosphate Cement do not raise any new issues of safety and effectiveness. The bench, animal, and biocompatibility testing demonstrates that DSM Biomedical Calcium Phosphate Cement is substantially equivalent to the predicate device.

Conclusion:

The DSM Biomedical Calcium Phosphate Cement is substantially equivalent in terms of indications for use, material composition, technological characteristics, and performance characteristics to the predicate device, Stryker® Injectable Cement (K060061) [Stryker® CMF].

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.