K Number
K173194

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-10-10

(373 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Orantech Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A. These devices are indicated for used by qualified medical personnel only.

Device Description

Reusable temperature probes are used during patient temperature measurement for multi-patient use. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. Temperature probes measure temperature by a resistor that is sensitive to temperature changes. These probes are connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.

These temperature probes are typically used with legacy Philips monitor model MP30 and Nihon Kohden monitor model BSM-6301A.

AI/ML Overview

The medical device in question is the Orantech Inc. Temperature Probe, models TS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30, and TS-PH-AS30.

Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Set by Standard or Performance Goal)Reported Device PerformanceOutcome
Biocompatibility
Cytotoxicity (ISO 10993 series)No cytotoxicity detectedPass
Sensitization (ISO 10993 series)No sensitization detectedPass
Irritation (ISO 10993 series)Negligible irritationPass
Safety
Electrical Safety (IEC 60601-1)Meets requirements of IEC 60601-1Pass
Performance (Continuous Clinical Thermometer)
Laboratory Accuracy (25-45°C range)Not greater than 0.3°C (Accuracy: ± 0.1°C stated below)Pass
Time ResponseHeating transient time < 150sPass
Use Life TestingMeets requirements for expected service lifePass
Cleaning and Disinfection EffectivenessCleaning and disinfection method is effectivePass

Note on Accuracy: The 'Product Specification' table explicitly states an accuracy of "± 0.1°C" for all models within the 25-45°C range. The "Laboratory accuracy" test result states "Not greater than 0.3°C", which is a broader allowance than the stated ± 0.1°C. Assuming the ± 0.1°C is the target, 'Not greater than 0.3°C' still indicates compliance.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the sample size used for the test set for any of the performance studies (biocompatibility, safety, accuracy, time response, use life, cleaning/disinfection).

The document does not specify the data provenance (e.g., country of origin of the data, retrospective or prospective). The testing was conducted by or for Orantech Inc., which is based in China.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The type of device (temperature probe) and the nature of the tests (laboratory accuracy, electrical safety, biocompatibility) typically do not involve establishing ground truth through human experts in the same way, for example, an AI diagnostic imaging device would. The "ground truth" for these tests would be derived from calibrated equipment and established scientific methods, rather than expert consensus on interpretation.

4. Adjudication Method for the Test Set

This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies in ground truth establishment. For the types of tests conducted for a temperature probe, such methods are not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This type of study is relevant for AI-powered diagnostic devices that assist human readers (e.g., radiologists interpreting images). The Orantech Temperature Probe is a physical medical device for direct measurement, not an AI software.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable to the Orantech Temperature Probe. It is a physical device, not an algorithm. The performance tests (laboratory accuracy, time response, etc.) represent its "standalone" performance in a controlled environment, but it's not an algorithm whose performance can be evaluated independently of human interaction in the clinical use context that AI devices are.

7. The Type of Ground Truth Used

The ground truth for the various tests was established through:

  • Biocompatibility: Adherence to established international standards (ISO 10993 series) using laboratory testing for cytotoxicity, sensitization, and irritation. The "ground truth" is the scientific determination of biological response to the materials.
  • Safety: Compliance with electrical safety standard (IEC 60601-1) through testing. The "ground truth" is the objective measurement of electrical parameters against the standard's limits.
  • Performance (Laboratory Accuracy, Time Response, Use Life): Measurement against calibrated reference instruments and established performance criteria defined by medical device standards or internal specifications. For accuracy, the "ground truth" is the actual temperature measured by a highly accurate reference thermometer.
  • Cleaning and Disinfection: Validation against AAMI TIR documents and FDA guidance, likely involving microbiological testing to confirm the reduction/elimination of pathogens. The "ground truth" is the demonstrably effective reduction of microbial load.

8. The Sample Size for the Training Set

This information is not applicable as the device is a physical temperature probe and not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above (not an AI/ML algorithm).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 10, 2018

Orantech Inc. % Kevin Wang Chonconn Medical Device Consulting Co., Ltd. 22A, Haijing Square, No. 18, Taizi Road Nanshan District, Shenzhen, 518067 China

Re: K173194

Trade/Device Name: Temperature Probe (Models: TS-Y400-AG30. TS-Y400-AS30. TS-PH-AG30 and TS-PH-AS30) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 13, 2018 Received: September 13, 2018

Dear Mr. Kevin Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173194

Device Name

Temperature Probe (Model: TS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30 and TS-PH-AS30)

Indications for Use (Describe)

Orantech Temperature Probes are intended to be used for monitoring temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A. These devices are indicated for used by qualified medical personnel only.

Type of Use (Select one or both, as applicable)
✔ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2018/02/05

1. Submission sponsor

Name: Orantech Inc.

Address: Zone#A,4F,1st Bld,7th Industrial Zone, Yulv Community, GongMing, Guangming New District, Shenzhen, P.R. China 518106 Contact person: Hsin Xiong Title: General manager E-mail: Yunxi.xiong@orantech.cn Tel: +86-755 23699939

2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd. Address: 22A, Haijing Square, No. 18, Taizi Road, Nanshan District, Shenzhen, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin(@chonconn.com Tel: +86-755 33941160

Trade/Device NameTemperature Probe
ModelTS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30 and TS-PH-AS30
Common NameTemperature Probe
Regulatory ClassClass II
Classification21CFR 880.2910 / Clinical electronic thermometer / FLL
Submission typeTraditional 510(K)

3. Subject Device Information

4. Predicate Device

    1. Unimed Medical Supplies Inc. Temperature Probe under K121427.

5. Device Description

Reusable temperature probes are used during patient temperature measurement for multi-patient use. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. Temperature probes measure temperature by a resistor that is sensitive to temperature changes. These probes are connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.

These temperature probes are typically used with legacy Philips monitor model MP30 and Nihon Kohden monitor model BSM-6301A.

{4}------------------------------------------------

Products are packed individually into a plastic bag in non-sterile condition. Package label describes product REF codes, manufacturing date, CE-mark, legal entity information and a caution "Rx Only ".

Product Specification

ModelApplicationsitelengthCompatiblemonitorsMeasurementrangeAccuracy
TS-PH-AG30Rectum3.0mPhilipsModel: MP3025-45°C$\pm$ 0.1°C
TS-PH-AS30Skin3.0mPhilipsModel: MP3025-45°C$\pm$ 0.1°C
TS-Y400-AG30Rectum3.0mNihon KohdenModel: BSM-6301A25-45°C$\pm$ 0.1°C
TS-Y400-AS30Skin3.0mNihon KohdenModel: BSM-6301 A25-45°C$\pm$ 0.1°C

Indication for use 6.

Orantech Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A.

These devices are indicated for used by qualified medical personnel only.

Comparison to the Predicate Device 7.

Orantech ModelPredicate ModelDescription
TS-PH-AG30THP-AGReusable Temperature Probe, adult rectum, 3.0M
TS-PH-AS30THP-ASReusable Temperature Probe, adult skin, 3.0M
TS-Y400-AG30T2252-AGReusable Temperature Probe, adult rectum, 3.0M
TS-Y400-AS30T2252-ASReusable Temperature Probe, adult skin, 3.0M
FeaturesSubject DeviceOrantech TemperatureProbePredicate DeviceUnimed TemperatureProbeK121427Comparison
ClassificationNameTemperature ProbeTemperature ProbeSame
Product CodeFLLFLLSame
Regulation Number880.2910880.2910Same
PanelGeneral HospitalGeneral HospitalSame
ClassIIIISame
ThermistorNTC resistanceNTC resistanceSame
(2252 Ohms in 25°C)(2252 Ohms in 25°C)
Accuracy range25-45°C25-45°CSame
Accuracy+0.1℃C+0.1℃CSame
Measure siteSkin & RectumSkin & RectumSame
Patient populationAdultAdultSame
Material contact toThermistor Sensor Tip:Thermistor Sensor Tip:Different (1)
bodyEpoxy & Stainless steelStainless steel
Length3m1.5m & 3mSame
Cable materialPVC & TPUPVC & TPUSame
Plug materialPA66ABSDifferent (2)
BiocompatibilityCytotoxicityCytotoxicitySame
SensitizationSensitization
IrritationIrritation
OperationalContinualContinualSame
Principles
Indication for UseOrantech TemperatureUnimed Temperature ProbesDifferent (3)
Probes are intended to beare intended to be used for
used for monitoringmonitoring temperature. The
temperature. Thetemperature probes are
temperature probes aredesigned for use with
reusable and designed formonitors of Philips,
use with monitors ofMarquette, Mindray,
Philips Model MP30 andSpacelabs, Siemens,
Nihon Kohden ModelArtema/S&W and other
BSM-6301A. Thesemonitors compatible with
devices are indicated forYSI 400 series temperature
used by qualified medicalprobes.
personnel only.These devices are indicated
for used by qualified
medical personnel only.

{5}------------------------------------------------

Justifications for differences between Orantech Temperature Probe and the predicate device are shown as below:

Different (1): It's different with the patient contacted material of predicate device. However, it's conformed with ISO 10993 series standard. According to analysis results, all the Patient Contacting Materials shows no cytotoxicity, irritation response was negligible and no sensitization. Thus, different material will not raise any safety or biocompatibility issues.

Different (2): It's different with the plug material of predicate device. However, it's conformed with ISO 10993 series standard. According to analysis results, all the materials shows no cytotoxicity, irritation

{6}------------------------------------------------

response was negligible and no sensitization. Thus, different material of plug will not raise any safety or biocompatibility issues.

Different (3): The compatible monitors of proposed devices are different from the predicate device. The proposed devices are intended to use with Philips and Nihon Kohden monitors and the predicate devices are intended to use with Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors. The core component of temperature probe is Negative Coefficient (NTC) which is identical to the NTC used in predicate device. The NTC determines the accuracy and range of temperature measurement. Therefore, the indication for use of proposed devices is essentially the same as predicate device.

Performance Data 8.

Biocompatibility testing

The biocompatibility evaluation for the Orantech Temperature Probe was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation

The Orantech Temperature Probe is considered surface contacting for a duration of exceed 24 hours but not 30 days.

Non-clinical data

The Orantech Temperature Probe has been tested according to the following standards:

TestDescriptionResult
SafetyEnsures the temperature probes meet the requirements of IEC 60601-1 electrical safetyPass
Laboratory accuracyNot greater than 0.3 °C for a continuous clinical thermometer that is not an adjusted mode clinical thermometerPass
Time responseHeating transient time < 150sPass
Use life testingEnsures the temperature probes meet the requirements of the expected service life of the probe.Pass
Cleaning and DisinfectionValidate the cleaning and disinfection method is effective. The study was conducted based on following documents:1) AAMI TIR 30: 2011 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices;2) AAMI TIR 12: 2010 Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers;Pass

{7}------------------------------------------------

Oral ntech

3) FDA Guidance, Reprocessing Medical Devices in Health CareSettings: Validation Methods and Labelling document issued on:March 17, 2015.
-----------------------------------------------------------------------------------------------------------------------------------------------------

9. Conclusion

Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the predicate device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.