(167 days)
The Mega Plus Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Mega Plus Spine System is a top-loading multiple component, posterior spinal fixation systems which consist of pedicle screws (multi-axial, multi-axial long-arm, mono and long-arm), iliac screws, rods, locking bolts, cross-links and connectors. The Mega Plus Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. The Mega Plus Spine System implant components are supplied non-sterile, single use and fabricated from titanium alloy (Ti-6Al-4V ELI) that conforms to the ASTM F136. Various sizes of these implants are available. Specialized instruments are also available for the application and removal of the Mega Plus Spine System.
This document is a 510(k) summary for the "Mega Plus Spine System," a medical device intended for spinal fusion. The FDA has determined it is substantially equivalent to legally marketed predicate devices.
Here's an analysis of the provided information, focusing on the absence of information rather than its presence, as this is a mechanical device, not an AI/software device.
1. Table of acceptance criteria and the reported device performance
The document lists performance tests conducted for the Mega Plus Spine System, all based on ASTM standards. However, it does not explicitly state specific quantitative acceptance criteria or detailed numerical results for these tests. It only makes a general concluding statement about the sufficiency of the device's strength.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly stated (e.g., minimum load, cycles) | Static compression bending per ASTM F1717-15: "results...show that the strength of the Mega Plus Spine System is sufficient for its intended use" |
| Not explicitly stated (e.g., minimum load, cycles) | Static tension bending per ASTM F1717-15: "results...show that the strength of the Mega Plus Spine System is sufficient for its intended use" |
| Not explicitly stated (e.g., minimum torque, cycles) | Static torsion per ASTM F1717-15: "results...show that the strength of the Mega Plus Spine System is sufficient for its intended use" |
| Not explicitly stated (e.g., minimum load, cycles) | Dynamic axial compression bending fatigue per ASTM F1717-15: "results...show that the strength of the Mega Plus Spine System is sufficient for its intended use" |
| Not explicitly stated (e.g., specific component failure thresholds) | Component testing per ASTM F1798-13: "results...show that the strength of the Mega Plus Spine System is sufficient for its intended use" |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify the sample size for any of the performance tests.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether the tests were retrospective or prospective. It only mentions that BK MEDITECH Co., Ltd. is based in the Republic of Korea.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable to this document. The Mega Plus Spine System is a mechanical spinal implant, not a diagnostic or AI device that requires expert-established ground truth from medical imaging or clinical assessment for its performance evaluation (beyond internal engineering validation). The "ground truth" for this type of device is its mechanical integrity and compliance with engineering standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable as there is no mention of human observation or assessment of the performance data that would require an adjudication method. The performance data seems to be derived from objective mechanical testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. The Mega Plus Spine System is a mechanical implant, not an AI-assisted diagnostic tool or an imaging device. Therefore, MRMC comparative effectiveness studies involving human readers and AI assistance are irrelevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The Mega Plus Spine System is a mechanical implant; there is no algorithm or software component that operates in a standalone manner.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by conformance to recognized engineering and material standards (ASTM F1717-15 and ASTM F1798-13) and the mechanical properties of its constituent materials (titanium alloy Ti-6Al-4V ELI). These standards implicitly define what constitutes acceptable mechanical performance for spinal fixation devices.
8. The sample size for the training set
This question is not applicable. There is no "training set" in the context of mechanical device testing for a spinal implant. Training sets are relevant for machine learning algorithms or AI models.
9. How the ground truth for the training set was established
This question is not applicable for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The overall design is clean and professional, reflecting the organization's role in public health and safety.
March 15, 2018
BK MEDITECH Co., Ltd. % Meredith L. May, MS, RAC Vice President, Empirical Consulting Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K173180
Trade/Device Name: Mega Plus Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: February 23, 2018 Received: February 28, 2018
Dear Ms. May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|---|---|
| Food and Drug AdministrationIndications for Use | Expiration Date: January 31, 2017See PRA Statement on last page. |
| 510(k) Number ( if known ) | K173180 |
|---|---|
| Device Name | Mega Plus Spine System |
| Indications for Use ( Describe ) | The Mega Plus Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). |
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
| FORM FDA 3881 (9/13) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EE |
|---|---|---|
| ---------------------- | ------------- | ------------------------------------------- |
{3}------------------------------------------------
510(K) SUMMARY
| Submitter's Name: | BK MEDITECH CO., LTD. |
|---|---|
| Submitter's Address: | 58, Eunhaengnamu-ro, Yanggam-myeon,Hwaseong-si, Gyeonggi-do, 18633, KOREA, REPUBLIC OF |
| Submitter's Telephone: | 82-31-352-9135 |
| Contact Person: | Meredith L. May MS, RACEmpirical Consulting719-337-7579 |
| Date Summary was Prepared: | 27 September 2017 |
| Trade or Proprietary Name: | Mega Plus Spine System |
| Common or Usual Name: | Thoracolumbosacral Pedicle Screw System |
| Classification: | Class II per 21 CFR §888.3070 Device Classification |
| Product Code: | NKB |
| Classification Panel: | Division of Orthopedic Devices |
DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Mega Plus Spine System is a top-loading multiple component, posterior spinal fixation systems which consist of pedicle screws (multi-axial, multi-axial long-arm, mono and long-arm), iliac screws, rods, locking bolts, cross-links and connectors. The Mega Plus Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. The Mega Plus Spine System implant components are supplied non-sterile, single use and fabricated from titanium alloy (Ti-6Al-4V ELI) that conforms to the ASTM F136. Various sizes of these implants are available. Specialized instruments are also available for the application and removal of the Mega Plus Spine System.
INDICATIONS FOR USE
The Mega Plus Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).
TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:
- . Indications for Use
{4}------------------------------------------------
- Principles of Operation ●
- Materials of Manufacture .
- Implant Sizes ●
Table 5-1: Predicate Devices
| 510k | Trade or Proprietary or | Manufacturer | Predicate |
|---|---|---|---|
| Number | Model Name | Type | |
| K000236 | International Synergy™ VLS | INTERPORE CROSS | Primary |
| K050979 | S4 Spinal System | Aesculap®, Inc. | Additional |
| K111940 | S 100 Pedicle Screw System | Renovis | Additional |
| K130176 | Orthofix | Firebird Spinal Fixation System | Additional |
| K031175 | Click'X | Synthes | Additional |
| K123549 | Polaris Spinal System | Biomet | Additional |
| K113666 | XIA 3 Spinal System | Stryker Spine | Additional |
PERFORMANCE DATA
The Mega Plus Spine System has been tested in the following test modes:
- Static compression bending per ASTM F1717-15 ●
- Static tension bending per ASTM F1717-15 ●
- Static torsion per ASTM F1717-15 .
- Dynamic axial compression bending fatigue per ASTM F1717-15 ●
- Component testing per ASTM F1798-13 ●
The results of this non-clinical testing show that the strength of the Mega Plus Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the Mega Plus Spine System is substantially equivalent to the predicate device.
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.