K Number
K173130
Date Cleared
2017-12-19

(81 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reform® Midline Cortical Screw System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Device Description

The modular system titanium bone screws have a dual-lead cortical-cancellous thread form and are offered in both cannulated and non-cannulated versions with diameters ranging from Ø4.5 -9.5mm. The cobalt chrome tulips are offered in both standard and reduction style. The cobaltchrome rods are offered in multiple lengths with a Ø4.75mm. Various connectors are also provided with this system.

AI/ML Overview

This document describes the performance testing for the Reform® Midline Cortical Screw System (K173130), a medical device intended for posterior, non-cervical spinal fixation. The study evaluates the device's mechanical strength and equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

For this device, the "acceptance criteria" are implied by the comparison to predicate devices and the ASTM standards without explicit quantitative thresholds given in the provided text. The performance is reported as meeting these standards and being substantially equivalent.

Acceptance Criteria (Implied)Reported Device Performance
Strength sufficient for intended use"The results of this non-clinical testing show that the strength of the Reform® Midline Cortical Screw System is sufficient for its intended use..."
Substantially equivalent to legally marketed predicate devices"...and is substantially equivalent to legally marketed predicate devices." Also, "The overall technology characteristics and mechanical performance data lead to the conclusion that the Reform® Midline Cortical Screw System is substantially equivalent to the predicate device."
Conformance to ASTM F1717 (Static axial compression bending)The device was tested for "Static axial compression bending per ASTM F1717". Results support substantial equivalence.
Conformance to modified ASTM F1717 (Static torsion)The device was tested for "Static torsion per modified ASTM F1717". Results support substantial equivalence.
Conformance to ASTM F1717 (Dynamic axial compression bending)The device was tested for "Dynamic axial compression bending per ASTM F1717". Results support substantial equivalence.
Conformance to ASTM F1798 (Static tulip-shank dissociation)"Additionally, static tulip-shank dissociation testing was performed on the polyaxial screw per ASTM F1798." Results support substantial equivalence.
Materials conform to ASTM standards"Reform® Midline Cortical Screw System is made from materials that conform to ASTM standards."
Technological characteristics are identical to predicates"Specifically the following characteristics are identical between the subject and predicates: Indications for Use, Materials of manufacture, Structural support mechanism." "the minor differences do not raise any new issues of safety and effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not specify the exact number of devices or components tested for each mechanical test (Static axial compression bending, Static torsion, Dynamic axial compression bending, Static tulip-shank dissociation). It only states that the testing was performed.
  • Data Provenance: The data is produced from non-clinical mechanical testing. The origin is not country-specific in the provided text as it is laboratory testing against international standards (ASTM). It is inherently prospective in nature, as the tests are conducted to evaluate the device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • This type of study does not involve human experts establishing "ground truth" for the test set in the way a clinical study or AI performance evaluation would. Instead, the "ground truth" or standard for performance is defined by the ASTM consensus standards (ASTM F1717, ASTM F1798) and the performance of legally marketed predicate devices. The evaluation is based on mechanical properties and engineering principles rather than expert human interpretation.

4. Adjudication Method for the Test Set

  • Not applicable. This is a non-clinical mechanical testing study based on established engineering standards and comparison to predicate devices, not a study requiring human adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This is a non-clinical mechanical performance study of a spinal implant system, not an imaging or diagnostic AI-assisted workflow study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • No, a standalone algorithm performance study was not done. This device is a physical medical implant, not a software algorithm.

7. The Type of Ground Truth Used

  • The "ground truth" in this context is established by:
    • Consensus Engineering Standards: Specifically, ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model) and ASTM F1798 (Standard Test Method for Static and Fatigue Shear/Compression Loading of Intervertebral Body Spinal Implants). These standards define the methodologies and expected performance characteristics for spinal implants.
    • Performance of Predicate Devices: The key aspect of the substantial equivalence determination is demonstrating that the Reform® Midline Cortical Screw System performs equivalently or better than the identified predicate devices (K150856 Reform Pedicle Screw System and K171082 JANUS Midline) under the same test conditions.

8. The Sample Size for the Training Set

  • Not applicable. This is a non-clinical mechanical testing study of a physical device. There is no "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this type of mechanical device study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 19, 2017

Precision Spine, Inc. % Meredith L. May, MS, RAC Empirical Consulting, LLC 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K173130

Trade/Device Name: Reform® Midline Cortical Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: September 28, 2017 Received: September 29, 2017

Dear Ms. May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean –S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: January 31, 2017
See PRA Statement on last page.
Indications for Use
510(k) Number (if known)K173130
Device NameReform® Midline Cortical Screw System
Indications for Use (Describe)The Reform® Midline Cortical Screw System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13) the user requests the same thing as the previous turn, so I will output the same thing.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Submitter's Name:Precision Spine, Inc.
Submitter's Address:2050 Executive DrivePearl, MS 39208
Submitter's Telephone:973-455-7150
Contact Person:Meredith L May MS RACEmpirical Consulting LLC719-337-7579
Date Summary was Prepared:October 27th, 2017
Trade or Proprietary Name:Reform® Midline Cortical Screw System
Common or Usual Name:Thoracolumbosacral Pedicle Screw System
Classification:Class II per 21 CFR §888.3070
Product Code:NKB

510(K) SUMMARY

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The modular system titanium bone screws have a dual-lead cortical-cancellous thread form and are offered in both cannulated and non-cannulated versions with diameters ranging from Ø4.5 -9.5mm. The cobalt chrome tulips are offered in both standard and reduction style. The cobaltchrome rods are offered in multiple lengths with a Ø4.75mm. Various connectors are also provided with this system.

Indications for Use

The Reform® Midline Cortical Screw System is intended for posterior, non-cervical fixation as an adjuncted to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiograpahic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis and/ or failed previous fusion.

TECHNOLOGICAL CHARACTERISTICS

Reform® Midline Cortical Screw System is made from materials that conform to ASTM standards. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

  • Indications for Use
  • Materials of manufacture ●
  • Structural support mechanism

{5}------------------------------------------------

510kTrade or Proprietary or ModelManufacturerPredicate Type
NumberName
K150856 Reform Pedicle Screw SystemPrecision SpinePrimary
K171082 JANUS MidlineOrthofixAdditional

Table 5-1: Predicate Devices

PERFORMANCE DATA

The Reform® Midline Cortical Screw System has been tested in the following test modes:

  • · Static axial compression bending per ASTM F1717
  • · Static torsion per modified ASTM F1717
  • · Dynamic axial compression bending per ASTM F1717

Additionally, static tulip-shank dissociation testing was performed on the polyaxial screw per ASTM F1798.

The results of this non-clinical testing show that the strength of the Reform® Midline Cortical Screw System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Reform® Midline Cortical Screw System is substantially equivalent to the predicate device.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.