K Number
K173075

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-03-08

(160 days)

Product Code
Regulation Number
884.6160
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.

The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.

Device Description

The ZyMōt ICSI Sperm Separation Device and ZyMōt Multi Sperm Separation Device (850μl, 3ml) are used to prepare motile sperm for assisted reproductive technology (ART) procedures. They separate motile sperm from the semen based on the mobility (i.e., swimup nature) of motile sperm.

The ZyMōt ICSI Sperm Separation Device has five micro-channels, each accommodating 2ul of semen and processes up to five separations from one semen sample. Each channel has an inlet port for applying the semen sample and an outlet port for collecting the motile sperm. The ports are connected by a fluid-filled micro-channel in which separation occurs.

The ZyMōt Multi Sperm Separation Device has two processing volumes, 850µl and 3ml. Each version has an inlet port that communicates with the lower sample chamber. The sample chamber is separated from the upper collection chamber by a microporous filter. Untreated semen is added through the inlet port. After incubation, the separated motile sperm are collected from the upper chamber through the outlet port.

The subject devices are radiation-sterilized devices with a sterility assurance level (SAL) of 10-6. They are individually packaged and for single-use only.

AI/ML Overview

The document describes the non-clinical performance testing for the ZyMōt ICSI Sperm Separation Device and ZyMōt Multi Sperm Separation Device (850µl, 3ml).

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriterionReported Device Performance
Human sperm survival assay (HSSA)≥80% of control motility at 24 hours after 30 minutes exposureNot explicitly stated, but implies met (study supports SE)
Performance Testing (Motile Sperm)Not explicitly stated as a numerical acceptance criterion. Implied acceptable separation of motile sperm.
ZyMōt ICSI (motile sperm separation)N/A (implied improvement)29.7% / 87% (before/after separation)
ZyMōt Multi 850µl (motile sperm separation)N/A (implied improvement)67% / 83.6% (before/after separation)
ZyMōt Multi 3ml (motile sperm separation)N/A (implied improvement)67% / 86.5% (before/after separation)
Performance Testing (Progressively Motile Sperm)Not explicitly stated as a numerical acceptance criterion. Implied acceptable separation of progressively motile sperm.
ZyMōt ICSI (progressively motile sperm separation)N/A (implied improvement)26.3% / 74.2% (before/after separation)
ZyMōt Multi 850µl (progressively motile sperm separation)N/A (implied improvement)44.7% / 68.6% (before/after separation)
ZyMōt Multi 3ml (progressively motile sperm separation)N/A (implied improvement)44.7% / 75.4% (before/after separation)

2. Sample size used for the test set and the data provenance:

  • Human sperm survival assay (HSSA): "Donor sperm" were used. The exact number of donors or samples is not specified. Data provenance is not explicitly stated (e.g., country of origin), nor is it specified if it was retrospective or prospective.
  • Performance testing: "Each version of the subject device was used to separate motile sperm from donor semen samples." The exact number of donor semen samples or individual tests is not specified. Data provenance is not explicitly stated, nor is it specified if it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The tests performed are laboratory-based assays (HSSA and performance testing determining sperm motility percentage), which typically rely on standardized laboratory protocols and instrumentation rather than expert consensus for ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable as the tests are laboratory assays, not reliant on expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The device is a sperm separation device, not an AI-powered diagnostic tool requiring human reader interpretation. No MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The device is a physical sperm separation device, not an algorithm.

7. The type of ground truth used:

  • Human sperm survival assay (HSSA): The ground truth was established by comparing the rate of motile sperm after exposure to the device and incubation to that of a control sample (no exposure to the device). This relies on a direct measurement of motility.
  • Performance testing: The ground truth for motility and progressively motile sperm percentages was established through direct measurement of these parameters in the input (before separation) and output (after separation) semen samples.

8. The sample size for the training set:

This information is not applicable. The device is a physical sperm separation device, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established:

This information is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 8, 2018

DxNow, Inc. Kevin Sly Senior Advisor to DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, Maryland 20879-3429

Re: K173075

Trade/Device Name: ZyMot ICSI Sperm Separation Device, ZyMot Multi Sperm Separation Device (850µl, 3ml) Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: Class II Product Code: MQK Dated: February 6, 2018 Received: February 6, 2018

Dear Kevin Sly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173075

Device Name ZyMōt ICSI Sperm Separation Device ZyMot Multi Sperm Separation Device (850ul, 3ml)

Indications for Use (Describe)

The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.

The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (IVF) and intrauterine insemination (IUI) procedures.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K173075

1. Submitter

DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, MD 20879-3429 Contact: Bill Sharp, CEO Email: bsharp@dxnowinc.com Phone: 240-454-9893

2. Correspondent

Kevin Sly Senior Advisor, DxNow, Inc. Email: ksly@dxnowinc.com Phone: 240-751-6521

3. Date Prepared: March 6, 2018

4. Device Identification:

Trade name:ZyMōt ICSI Sperm Separation DeviceZyMōt Multi Sperm Separation Device (850µl, 3ml)
Common name:Sperm Separation Device
Classification name:Assisted Reproduction Labware (21 CFR 884.6160)
Product code:MQK (Labware, Assisted Reproduction)
Regulatory class:II

ട. Predicate Device

Qualis (K133295). This predicate device has not been subject to any design related recalls.

6. Device Description

The ZyMōt ICSI Sperm Separation Device and ZyMōt Multi Sperm Separation Device (850μl, 3ml) are used to prepare motile sperm for assisted reproductive technology (ART) procedures. They separate motile sperm from the semen based on the mobility (i.e., swimup nature) of motile sperm.

{4}------------------------------------------------

The ZyMōt ICSI Sperm Separation Device has five micro-channels, each accommodating 2ul of semen and processes up to five separations from one semen sample. Each channel has an inlet port for applying the semen sample and an outlet port for collecting the motile sperm. The ports are connected by a fluid-filled micro-channel in which separation occurs.

The ZyMōt Multi Sperm Separation Device has two processing volumes, 850µl and 3ml. Each version has an inlet port that communicates with the lower sample chamber. The sample chamber is separated from the upper collection chamber by a microporous filter. Untreated semen is added through the inlet port. After incubation, the separated motile sperm are collected from the upper chamber through the outlet port.

The subject devices are radiation-sterilized devices with a sterility assurance level (SAL) of 10-6. They are individually packaged and for single-use only.

7. Indications for Use

The ZyMōt ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.

The ZyMōt Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.

Device/PredicateDevice(s):K173075 (subject devices)K133295 (predicate device)
Indications forUseThe ZyMōt ICSI is intended for preparingmotile sperm from semen for use in thetreatment of infertile couples byintracytoplasmic sperm injection (ICSI)procedures.The ZyMōt Multi is intended for preparingmotile sperm from semen for use in thetreatment of infertile couples byintracytoplasmic sperm injection (ICSI), invitro fertilization (IVF) and intrauterineinsemination (IUI) procedures.The Qualis is intended for preparingmotile human sperm for use in thetreatment of infertile couples byintracytoplasmic sperm insertion(ICSI) fertilization.
DesignZyMōt ICSI – A disposable culture dishcontaining five micro-channels, eachconnected to an inlet port and an outletportZyMōt Multi – A disposable culture dishcontaining a separation chamber and aninlet port. The separation chamber has alower sample chamber and an upperA disposable culture dish with fourchambers connected by micro-channels

8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

{5}------------------------------------------------

collection chamber. The two chambersare separated by a microporous filter.
Mechanism ofactionZyMōt ICSI — The inlet port is preloadedwith sperm separation medium and thenloaded with semen. After incubation at37°C for 30 min, the motile sperm arecollected from the outlet port.ZyMōt Multi – The semen is added to theinlet port to fill the lower samplechamber; then, the separation medium isadded to the upper collection chamber.The loaded device is incubated at 37°C for30 min to allow motile sperm to swim upand cross the filter to migrate into theover-laying separation medium in theupper collection chamber.The semen sample is placed inChamber A and separation mediumis placed in Chamber B. Fluids fromboth chambers flow via the micro-channels into the central micro-channel where the two fluids passside-by-side in laminar flow. Motilesperm can swim across the interfaceof the laminar flow streams andpass into the separation mediumstream but non-motile sperm anddebris cannot. Motile sperm thatcross into the separation mediumflow are carried into Chamber Cwhere they are collected. Non-motile sperm and debris remain inthe semen sample flow fromChamber A into Chamber D.
MaterialZyMōt ICSI – Polymethylmethacrylate,borosilicate glass, flash-spun high-densitypolyethylene fibersZyMōt Multi - Polymethylmethacrylate,polycarbonate, flash-spun high-densitypolyethylene fibersCyclo-olefin polymer

The subject and predicate devices have different Indications for Use statements; however, they have the same intended use – preparing motile sperm from semen for use in ART procedures.

The subject and predicate devices have the same fundamental design incorporating a culture dish with an inlet port/chamber for loading of semen and an outlet port/chamber for collection of motile sperm. Both devices rely on sperm motility to separate motile sperm from non-motile sperm or non-sperm cells. The predicate device incorporates microfluidic technology, whereas the subject devices do not. However, this difference does not raise different questions of safety and effectiveness as compared to the predicate device, as both devices (and other cleared devices) have the same fundamental mode of action (i.e., using motility of sperm to prepare pools of motile sperm for use in ART procedures). In addition, the subject and predicate devices have different materials. However, differences in materials do not raise different questions of safety and effectiveness as compared to the predicate device.

g. Summary of Non-Clinical Performance Testing:

The following studies have been performed to support substantial equivalence to the predicate devices:

{6}------------------------------------------------

  • . Sterilization validation study per ISO 11137-2:2012
  • . Transportation simulation study per ASTM D5276-98
  • . Package integrity testing following accelerated aging per ASTM F1980-07 to support a 12-month shelf-life:
      • Dye penetration testing per ASTM F1929-98
      • Seal strength testing per ASTM F88/F88M-15
  • Endotoxin testing per USP<85>: ≤2.15 EU/device .
  • . Human sperm survival assay (HSSA) before and after accelerated aging to support a 12-month shelf-life:

Donor sperm were exposed to the subject devices for 30 minutes and then, incubated at room temperature for 24 hours. The rate of motile sperm after incubation was compared to that of the control (no exposure to the subject device). The acceptance criterion is ≥80% of the control motility at 24 hours after exposure for 30 minutes.

  • Performance testing: .
    Each version of the subject device was used to separate motile sperm from donor semen samples. The separation procedures followed the Instructions for Use, and the percentage of motile sperm and progressively motile sperm in the output samples were compared to those of the input samples. A summary of the results is provided in the table below:
DevicePercentage of motile sperm(before/after separation)Percentage of progressively motilesperm (before/after separation)
ZyMōt ICSI29.7% / 87%26.3% / 74.2%
ZyMōt Multi 850µl67% / 83.6%44.7% / 68.6%
ZyMōt Multi 3ml67% / 86.5%44.7% / 75.4%

10. Conclusion

The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.

§ 884.6160 Assisted reproduction labware.

(a)
Identification. Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.