(64 days)
The device is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid), Cephalic (neonatal and adult), Trans-rectal, Transvaginal, Peripheral Vascular, Cerebral Vasculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
This SonoScape P10 series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 1.0 MHz to 25.0 MHz. This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.
This FDA 510(k) summary does not contain performance acceptance criteria or detailed results from a study proving device efficacy. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a standalone efficacy study with concrete performance metrics.
Specifically, the document states:
- "No clinical testing was required." This indicates that no human clinical trials were performed to establish the device's diagnostic performance for its intended use cases against a ground truth.
- "Laboratory tests (including Phantom tests) were conducted to verify that the P10 system met all design specifications and the P10 system conformed to applicable medical device standards." This refers to technical performance verification against engineering specifications and regulatory standards (electrical safety, biocompatibility, acoustic output), not diagnostic accuracy or clinical effectiveness.
Therefore, the requested information regarding acceptance criteria and performance against those criteria through a study with specific sample sizes, ground truth establishment, expert adjudication, or MRMC studies for this product is not available in the provided text. The document is primarily a comparison against a predicate device to establish substantial equivalence.
{0}------------------------------------------------
December 1, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Sonoscape Medical Corp. % Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building, Yuquan road, Nanshan Shenzhen 518051, Guangdong CHINA
Re: K173058
Trade/Device Name: P10 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 15, 2017 Received: November 20, 2017
Dear Toki Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
P10 Series Digital Color Doppler Ultrasound System
Type of Use (Select one or both, as applicable)
Indications for Use (Describe)
The device is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid), Cephalic (neonatal and adult), Trans-rectal, Transvaginal, Peripheral Vascular, Cerebral Vasculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
| Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| CONTINUE ON A SEPARATE PAGE IF NEEDED. |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. |
| DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. |
| The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to: |
| Department of Health and Human ServicesFood and Druq AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff(@fda.hhs.gov |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
SonoScape P10 Exp,P10,P10Pro,P11,M10,P9 System: Diagnostic Ultrasound Pulsed Echo System
Diagnostic Ultrasound Pulsed Doppler Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Pediatric | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,5,6,7 | ||
| Neonatal Cephalic | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| FetalImaging&Other | Adult Cephalic | P | P | P | P | P | P | Note 1 | Notes 2,3,4 |
| Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intravascular | |||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Cardiac Adult | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| Cardiac Pediatric | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph. (Cardiac) | P | P | P | P | P | P | Note 1 | Notes 2,3,4 |
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Cerebral vascular | P |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{4}------------------------------------------------
Transducer: C322 Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Other (Urology) | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{5}------------------------------------------------
Transducer: 3C-A Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| FetalImaging&Other | Intra-operative Specify | ||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Other (Urology) | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D
Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{6}------------------------------------------------
Transducer: L741 Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | N | N | N | N | N | Note 1 | Notes2,4,6,7 | ||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral vessel | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Cerebral vascular |
P = previously cleared by FDA; N = new indication; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{7}------------------------------------------------
Transducer: C1-6 Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Other (Urology) | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA;
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 5: 4D
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{8}------------------------------------------------
Transducer: VC6-2 Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4,5 | ||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{9}------------------------------------------------
Transducer: C613 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | ||
| FetalImaging&Other | Adult Cephalic | |||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | ||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| PeripheralVessel | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{10}------------------------------------------------
Transducer: 2P1 Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | N | N | N | N | N | Note 1 | Notes 2,4 | |||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | ||
| FetalImaging&Other | Adult Cephalic | |||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | ||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 5: 4D
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{11}------------------------------------------------
Transducer: S1-5 Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| Fetal | Adult Cephalic | P | P | P | P | P | P | Note 1 | Notes 2,3,4 |
| Imaging& | Trans-rectal | ||||||||
| Other | Trans-vaginal | ||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| Cardiac Pediatric | P | P | P | P | P | P | Note 1 | Notes 2,3,4 | |
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D
Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{12}------------------------------------------------
Transducer: 7P-B Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | Other*Combined | Other*Specify | ||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | Note 1 | Notes 2 | ||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2 | |
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2 | |
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;
B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{13}------------------------------------------------
Transducer: L746 Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| FetalImaging&Other | Fetal | |||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | N | N | N | N | N | Note 1 | Notes 2,6,7 | |||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | Note 1 | Notes 2 | |||
| Musculo-skeletal(Superficial) | N | N | N | N | N | Note 1 | Notes 2 | |||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac | Cardiac Adult | |||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Trans-esoph.(Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| PeripheralVessel | Peripheral vessel | N | N | N | N | N | Note 1 | Notes 2 | ||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;
B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{14}------------------------------------------------
Transducer: L742 Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,6,7 | |||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{15}------------------------------------------------
Transducer: 1012 Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationGeneral(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes 2,4,6 | ||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Cerebral vascular |
P = previously cleared by FDA; E = added under this appendix N = new indication;
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{16}------------------------------------------------
Transducer: 9L-A Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | ||
| Ophthalmic | Ophthalmic | ||||||||||
| Fetal | |||||||||||
| Abdominal | |||||||||||
| Intra-operative Specify | |||||||||||
| Intra-operative Neuro | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,6,7 | ||||
| Neonatal Cephalic | |||||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||||
| Trans-rectal | |||||||||||
| Trans-vaginal | |||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph.(non-Card) | |||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | ||||
| Intravascular | |||||||||||
| Other (Ob/GYN) | |||||||||||
| Other (Urology) | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Cardiac | Intravascular(Cardiac) | ||||||||||
| Trans-esoph.(Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Other (specify) | |||||||||||
| PeripheralVessel | Peripheral vessel | N | N | N | N | N | Note 1 | Notes 2,4 | |||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{17}------------------------------------------------
Transducer: 12L-B Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,6,7 | ||
| FetalImaging&Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | ||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI
Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{18}------------------------------------------------
Transducer: 6V1 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Fetal | Adult Cephalic | ||||||||
| Imaging& | Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Other | Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{19}------------------------------------------------
Transducer: 6V3 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{20}------------------------------------------------
Transducer: 6V7 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | Other* | |||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Combined | Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{21}------------------------------------------------
Transducer: EC9-5 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D
Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{22}------------------------------------------------
Transducer: BCC9-5 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Other | Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{23}------------------------------------------------
Transducer: BCL10-5 Biplane Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Other | Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{24}------------------------------------------------
Transducer: MPTEE Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | P | P | P | P | P | P | Note 1 | Notes2,3,4 |
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{25}------------------------------------------------
Transducer: MPTEE mini Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | P | P | P | P | P | P | Note 1 | Notes2,3,4 |
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{26}------------------------------------------------
Transducer: LAP7 Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Note 7: Elastography
Note 5: 4D
{27}------------------------------------------------
Transducer: PWD2.0
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Fetal | Adult Cephalic | P | |||||||
| Imaging& | Trans-rectal | ||||||||
| Other | Trans-vaginal | ||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular | P |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{28}------------------------------------------------
Transducer: 13L-A Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,6,7 | ||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular(Cardiac) | ||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{29}------------------------------------------------
Transducer: 18L-A Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | ||
| Ophthalmic | Ophthalmic | ||||||||||
| Fetal | |||||||||||
| Abdominal | |||||||||||
| Intra-operative Specify | |||||||||||
| Intra-operative Neuro | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes2,4,6,7 | ||||
| Neonatal Cephalic | |||||||||||
| Fetal | Adult Cephalic | ||||||||||
| Imaging& | Trans-rectal | ||||||||||
| Other | Trans-vaginal | ||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph.(non-Card) | |||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||||
| Musculo-skeletal(Superficial) | P | P | P | P | P | Note 1 | Notes 2,4 | ||||
| Intravascular | |||||||||||
| Other (Ob/GYN) | |||||||||||
| Other (Urology) | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Cardiac | Intravascular(Cardiac) | ||||||||||
| Trans-esoph.(Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Other (specify) | |||||||||||
| Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{30}------------------------------------------------
Transducer: 10L-I Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes 2,4,6 | ||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | |
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{31}------------------------------------------------
Transducer: 6V3A Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| FetalImaging&Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral vessel | |||||||||
| PeripheralVessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{32}------------------------------------------------
Transducer: 12LT-A Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes 2,4,6 | |||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{33}------------------------------------------------
Transducer: 12LI-A Linear Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | P | P | P | P | P | Note 1 | Notes 2,4,6 | |||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | ||||||||||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 2,4 | ||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{34}------------------------------------------------
Transducer: 6CI-A Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph.(non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{35}------------------------------------------------
Transducer: 6CT-A Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Abdominal | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Intra-operative Specify | ||||||||||
| Intra-operative Neuro | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Fetal | Adult Cephalic | |||||||||
| Imaging& | Trans-rectal | |||||||||
| Other | Trans-vaginal | |||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Ob/GYN) | P | P | P | P | P | Note 1 | Notes 2,4 | |||
| Other (Urology) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular(Cardiac) | ||||||||||
| Cardiac | Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | ||||||||||
| Other (specify) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{36}------------------------------------------------
Transducer: CWD2.0
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| FetalImaging&Other | Fetal | ||||||||
| Abdominal | |||||||||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac | Cardiac Adult | P | |||||||
| Cardiac Pediatric | P | ||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{37}------------------------------------------------
Transducer: 3P-A Phased Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| FetalImaging&Other | Adult Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD;
B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 5: 4D
Note 3: TDI Note 4: 3D Note 6: Small Organ: breast, thyroid, testes
{38}------------------------------------------------
Transducer: C542 Micro-curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | Note 1 | Notes 2,4 | ||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Fetal | Adult Cephalic | ||||||||
| Imaging& | Trans-rectal | ||||||||
| Other | Trans-vaginal | ||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | |||||||||
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Intravascular(Cardiac) | |||||||||
| Cardiac | Trans-esoph.(Cardiac) | ||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
{39}------------------------------------------------
Transducer: C361 Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | N | N | N | N | N | N | Note 1 | Notes 2,4 | |
| Abdominal | N | N | N | N | N | N | Note 1 | Notes 2,4 | |
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Fetal | Adult Cephalic | ||||||||
| Imaging& | Trans-rectal | ||||||||
| Other | Trans-vaginal | ||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | N | N | N | N | N | N | Note 1 | Notes 2,4 | |
| Other (Urology) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular(Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| Vessel | Cerebral vascular |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
Note 4: 3D
Note 7: Elastography
Note 3: TDI
{40}------------------------------------------------
Transducer: C351 Curved Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(TRACK 1ONLY) | Specific(TRACKS 1 & 3) | B | M | PWD | CWD | ColorDoppler | Power(Amplitude)Doppler | Other*Combined | Other*Specify |
| Ophthalmic | Ophthalmic | ||||||||
| FetalImaging&Other | Fetal | N | N | N | N | N | N | Note 1 | Notes 2,4 |
| Abdominal | N | N | N | N | N | N | Note 1 | Notes 2,4 | |
| Intra-operative Specify | |||||||||
| Intra-operative Neuro | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (specify) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph.(non-Card) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Ob/GYN) | N | N | N | N | N | N | Note 1 | Notes 2,4 | |
| Other (Urology) | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (specify) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Cerebral vascular |
P = previously cleared by FDA; E = added under this appendix N = new indication;
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
{41}------------------------------------------------
510(k) Summary
1. Submitter [21 CFR807.92 (a) (1)]
| Submitter: | SONOSCAPE MEDICAL CORP. |
|---|---|
| Address: | 4/f, 5/f, 8/f, 9/f & 10/f, Yizhe Building, Yuquan Road, |
| Nanshan, Shenzhen 518051, Guangdong, China | |
| Contact Person: | Toki Wu |
| Tel: | +86 755 26722890 |
| Fax: | +86 755 26722850 |
| Email: | ra@sonoscape.net |
| Date Prepared | Sep 25, 2017 |
2. Device [21 CFR807.92 (a) (2)]
| Trade Name: | P10 Series Digital Color Doppler Ultrasound System |
|---|---|
| Common Name: | Diagnostic Ultrasound System and Transducers |
Classification Regulatory:
| FR Number | Product Code | |
|---|---|---|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
Classification Panel: Radiology Device Class: ll
3. Predicate Device(s) [21 CFR 807.92(a) (3)]
The identified predicate device within this submission is as follows: SonoScape P50 Digital Color Doppler Ultrasound System K170999
4. Device Description [21 CFR 807.92(a) (4)]
This SonoScape P10 series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of
{42}------------------------------------------------
anatomic structures and fluid flow within the body.
This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 1.0 MHz to 25.0 MHz.
This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.
This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.
5. Intended Use [21 CFR 807.92(a) (5)]
The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]
P10 Series Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate device:
SonoScape P50 Digital Color Doppler Ultrasound System K170999
Intended Use Comparison:
Compared with the predicate device SonoScape P50 (K170999), the Subject Device P10 Series has the same intended use.
Technical Characteristics Comparison:
The basic and key technical features of the Subject Device P10 Series are the same as the predicate device SonoScape P50 (K170999), including Design, Operation Controls, Display Modes, Operation Modes, Measurement Items, Cine Loop, Operating and Storage Condition.
The detailed technical features can be found in General Device Descriptions of this submission.
Probes Comparison:
Subject device P10 Series has the similar probes as the predicate device SonoScape P50 (K170999).
{43}------------------------------------------------
| ID | ComparisonItems | Predicate deviceSonoScape P50 Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30 Pro/P30S/P30/P27 | Subject deviceSonoScape P10 Exp, P10, P10Pro, P11, M10, P9 | Remark |
|---|---|---|---|---|
| 10 | Probes | C322 Micro-curved Array3C-A Curved ArrayC1-6 Curved ArrayC1-6A Curved Array6CT-A Curved Array6CI-A Curved ArrayC2-9 Curved ArrayVC6-2 Curved ArrayC613 Micro-curved ArrayVC2-9A Curved ArrayVC2-9 Curved Array | C322 Micro-curved ArrayC351 Micro-curved ArrayC361 Micro-curved Array3C-A Curved ArrayC1-6 Curved Array6CT-A Curved Array6CI-A Curved ArrayVC6-2 Curved ArrayC613 Micro-curved ArrayC542 Micro-curved Array | SEAnalysis1 |
| 8P1 Phased Array4P-A Phased ArrayS1-5 Phased Array7P-A Phased Array | 2P1 Phased ArrayS1-5 Phased Array7P-B Phased Array3P-A Phased Array | SEAnalysis2 | ||
| L742 Linear Array10I2 Linear Array9L-A Linear Array12L-A Linear ArrayVL12-5 Linear Array12L-B Linear Array13L-A Linear Array18L-A Linear Array10L-I Linear ArrayML3-18 Linear Array12LT-A Linear Array12LI-A Linear Array | L742 Linear ArrayL741 Linear ArrayL746 Linear Array10I2 Linear Array9L-A Linear Array12L-B Linear Array13L-A Linear Array18L-A Linear Array10L-I Linear Array12LT-A Linear Array12LI-A Linear Array | SEAnalysis3 |
Table 1 Probes Comparison
{44}------------------------------------------------
| 6V1 Micro-curved Array | 6V1 Micro-curved Array | ||
|---|---|---|---|
| 6V3 Micro-curved Array | 6V3 Micro-curved Array | ||
| 6V3A Micro-curved Array | 6V3A Micro-curved Array | ||
| 6V7 Micro-curved Array | 6V7 Micro-curved Array | ||
| EC9-5 Micro-curved Array | EC9-5 Micro-curved Array | ||
| BCC9-5 Micro-curved Array | BCC9-5 Micro-curved Array | ||
| BCL10-5 Biplane (Micro-curved+ Linear Array) | BCL10-5 Biplane (Micro-curved+ Linear Array) | ||
| VE9-5 Micro-curved Array | |||
| 12C-ER Micro-curved Array | |||
| C3-10V Micro-curved Array | |||
| VE3-10 Micro-curved Array | |||
| PWD2.0 Doppler | PWD2.0 Doppler | ||
| CWD5.0 CW | CWD2.0 CW | ||
| CWD8.0 CW | MPTEE Phased Array | ||
| CWD2.0 CW | MPTEE mini Phased Array | ||
| MPTEE Phased Array | LAP7 Linear Array | ||
| MPTEE mini Phased Array | |||
| LAP7 Linear Array | |||
SE Analysis 1 :
Compared with the predicate device, there are three new Curved Array probes as followed:
(C542), which is similar with the probe (C613) cleared with predicate device SonoScape P50 (K170999). (C351, C361), which is similar with the probe (C613) cleared with predicate device SonoScape P50 (K170999).
SE Analysis 2:
Compared with the predicate device, there are three new Phase probes as followed:
(3P-A, 7P-B), which is similar with the probe (4P-A) cleared with predicate device SonoScape P50 (K170999). (2P1), which is similar with the probe (8P1) cleared with predicate device SonoScape P50 (K170999).
SE Analysis 3:
Compared with the predicate device, there are two new Linear Array probes as followed: (L741, L746), which is similar with the probe (L742) cleared with predicate device SonoScape P50 (K170999).
Biopsy Bracket Comparison:
Subject device P10 Series has the similar Biopsy Brackets as the predicate device SonoScape P50 (K170999).
{45}------------------------------------------------
| ID | ComparisonItems | Subject DeviceSonoScape P10 Exp, P10, P10Pro, P11, M10, P9 | Predicate DeviceSonoScape P50Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30Pro/P30S/P30/P27(K170999) | Remark | |
|---|---|---|---|---|---|
| 11 | BiopsyBracket | NGBC613NGB3C-ANGBC322NGBC1-6NGBVC6-2NGBC542NGBC351NGBC361NGB6CT-ANGB6CI-ANGBL742-2NGBL741NGB9L-ANGB12L-BNGB13L-ANGB18L-ANGB10I2NGB10L-INGBL746NGB 12LT-ANGB 12LI-ANGB4P-ANGBS1-5NGB7P-BNGB2P1NGB6V1NGB6V3NGB 6V3ANGB6V7NGBBCL10-5NGBVE9-5NGBBCC9-5 | NGBC613NGB3C-ANGBC322NGBC1-6NGBC1-6ANGBC2-9NGBVC6-2NGBVC2-9NGBVC2-9ANGBL742-2NGB12L-ANGB9L-ANGB12L-BNGB13L-ANGB18L-ANGB10L-INGBML3-18NGBVL12-5NGB4P-ANGBS1-5NGB7P-ANGB6V1NGB6V3-2NGB6V7NGBBCL10-5NGBVE9-5NGBVE3-10NGBC3-10VNGBBCC9-5 | SEAnalysis 4 | |
| ID | ComparisonItems | Subject DeviceSonoScape P10 Exp, P10, P10Pro, P11, M10, P9 | Predicate DeviceSonoScape P50Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30Pro/P30S/P30/P27(K170999) | Remark | |
| NGBEC9-5 |
Table 2 Biopsy Bracket Comparison
{46}------------------------------------------------
SE Analysis 4:
Compared with predicate device, there are seven new biopsy brackets NGB2P1, NGBL741, NGB7P-B, NGBC542, NGBL746, NGBC351, NGBC361, the seven biopsy brackets are not cleared with predicate device (P50 series), but they have been cleared with other device. The detail information of all the seven biopsy brackets are described in table 6 as followed.
| BiopsybracketsModel | Availableprobemodel | PatientContactingMaterial | Type | Application | Knumber | Device |
|---|---|---|---|---|---|---|
| NGBL741 | L741 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K150045 | S30 |
| NGB7P-B | 7P-B | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K150045 | S30 |
| NGBC542 | C542 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K150045 | S30 |
| NGBL746 | L746 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K123616 | A6 |
| NGBC351 | C351 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K123616 | A6 |
| NGB2P1 | 2P1 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K150045 | S30 |
| NGBC361 | C361 | Stainlesssteel(SUS 304) | Reusable | Bodysurface | K102642 | SSI-8000 |
| Table 3 Technical specifications of new Biopsy brackets | ||
|---|---|---|
510(k) Summary
{47}------------------------------------------------
| BiopsybracketsModel | Availableprobemodel | PatientContactingMaterial | Type | Application | Knumber | Device |
|---|---|---|---|---|---|---|
| steel(SUS 304) | surface |
7. Non-Clinical Tests [21 CFR 807.92(b) (1)]
The P10 Series Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.
Laboratory tests (including Phantom tests) were conducted to verify that the P10 system met all design specifications and the P10 system conformed to applicable medical device standards.
The P10 Series system has been designed and manufactured to meet the following standards:
IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance [08/20/2012];
IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests [03/30/2007];
IEC 60601-2-37:2007, Medical Electrical Equipment-Part 2-37: Particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [08/09/2007];
ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity [06/01/2009];
ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for irritation and skin sensitization [08/01/2010];
AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for diagnostic ultrasound equipment [08/21/2009]; and
AIUM/NEMA UD 3:2004 (R2009), Standard for real-time display of thermal and mechanical acoustic output indices on diagnostic ultrasound equipment [08/13/2010].
8. Clinical Test [21 CFR 807.92(b) (2)]
No clinical testing was required.
{48}------------------------------------------------
9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]
In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SONOSCAPE MEDICAL CORP. concludes that P10 Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.