K Number
K173044
Date Cleared
2017-11-20

(53 days)

Product Code
Regulation Number
870.1120
Panel
CV
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The REUSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes. The DISPOSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.

Device Description

The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, pediatric and adult sizes. The proposed device includes disposable blood pressure cuff and reusable blood pressure, both disposable and reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube. The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics &PVC (Cuff) and PVC (Air Hose), thereinto, the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient. The disposable blood pressure cuff has 26 models with different size for different population with different arm size, and the 13 models of 26 are use single tube of air hose and other 13 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system. The reusable blood pressure cuff is reusable device, and which is made of 210D Nylon TPU (Cuff) and PVC (Air Hose), thereinto, the 210D Nylon TPU (Cuff) is the material used to contacting with the patient. The reusable blood pressure cuff has 16 models with different size for different population with different arm size, and the 8 models of 16 are use single tube of air hose and other 8 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.

AI/ML Overview

This is a 510(k) premarket notification for Reusable Blood Pressure Cuffs and Disposable Blood Pressure Cuffs. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

Here's an analysis of the provided information concerning acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by compliance with specific international standards. The performance is reported as meeting these standards.

Acceptance Criteria (Implied by Standards Compliance)Reported Device Performance
Biocompatibility:
Non-cytotoxicUnder the conditions of the study, not cytotoxic (Complies with ISO 10993-5:2009/(R) 2014)
Non-irritantUnder the conditions of the study, not an irritant (Complies with ISO 10993-10:2010)
Non-sensitizerUnder the conditions of the study, not a sensitizer (Complies with ISO 10993-10:2010)
Physical/Performance:
Pressure Range: 0-300 mmHg0-300 mmHg
Max. Pressure: 400 mmHg400 mmHg
Max. Leakage:

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).