(254 days)
Not Found
No
The description details a standard oximetry sensor and its function, relying on light absorption and a fixed algorithm for calculation. There is no mention of AI, ML, or any adaptive or learning components.
No.
The device is described as an accessory to oximeters intended for monitoring, not for treating any condition. Its purpose is to measure physiological parameters (oxygen saturation and pulse rate).
Yes
The device monitors physiological parameters (functional arterial oxygen saturation and pulse rate) for "spot checking or continuous monitoring" which provides data for medical assessment, indicating its use in diagnosing or managing conditions related to oxygen levels and heart rate.
No
The device description explicitly states that the proposed device, SpO2 Sensors, are accessories to oximeters and have physical components including LEDs, a photodiode, cable, and sensor. These are hardware components, not solely software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
- Device Function: The described device, SpO2 Sensors, measures functional arterial oxygen saturation and pulse rate non-invasively by emitting and receiving light through the human finger. It does not analyze samples taken from the body.
- Measurement Method: The method is based on light absorption, not on the analysis of biological or chemical components within a sample.
Therefore, while this device is a medical device used in healthcare settings for monitoring physiological parameters, it does not fit the definition of an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
Product codes
DQA
Device Description
The proposed device, SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in noninvasive U.S. legally marketed oximeters or patient monitors. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models.
The proposed devices measure, along with the corresponding oximeters, non-invasively, the arterial oxygen saturation of blood. The measurement method is based on the red and infrared light absorption of hemoglobin and oxyhemoglobin. Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. The light is emitted through human finger and received by a photodiode. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
The patient contact components include cable and sensor. The contact level is surface device and contact duration is limited contact (less than 24h).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult, pediatric or infant, depending on models.
Intended User / Care Setting
healthcare settings.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non clinical tests were conducted to verify that the proposed device met all design specifications and to demonstrate Substantially Equivalence(SE) to the predicate device. The test include following items
The performance tests performed on proposed devices include following items
- Pulse and SpO2 Accuracy test
- Pulse and SpO2 Accuracy test after High and Low Temperature
- Pulse and SpO2 Accuracy test after Maximum Times Disinfection
Biocompatibility tests were conducted per ISO 10993-1, the test items include
- Cytotoxicity Test
- Skin Irritation Test
- Skin Sensitization Test
Electrical safety and electromagnetic compatibility test were conducted on proposed devices per IEC 60601-1, IEC 60601-1-2 and ISO 80601-2-61.
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of proposed device versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
SpO2 Accuracy: +/- 3% @ 70-100%
PR Accuracy: +/- 3 bpm
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
June 8, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
APK Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K172981
Trade/Device Name: Reusable SpO2 Sensors, Disposable SpO2 Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: April 26, 2018 Received: April 30, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
1
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K172981
Device Name
Reusable SpO2 Sensors, Disposable SpO2 Sensors
Indications for Use (Describe)
The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
Type of Use (Select one or both, as applicable) |
---|
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
3
Exhibit #2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K172981
-
- Date of Preparation: 03/26/2018
-
- Sponsor
APK Technology Co., Ltd.
6Floor, B2 Building, Industry of Hengfeng, Hezhou, Xixiang, Bao'an Distric, Shenzhen, China
Establishment Registration Number: 3007699081
Contact Person: Caifang Wang Position: Management Representative Tel: +86-755-27325581 Fax: +86-755-27325585 Email: caifang@apk-technology.com
-
- Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
- Designated Submission Correspondent
Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
-
- Identification of Proposed Device
Trade Name: Reusable SpO2 Sensors Disposable SpO2 Sensors
- Identification of Proposed Device
4
Common Name: Oximeter (Accessory-sensor)
Regulatory Information Classification Name: Oximeter Classification: Class II Product Code: DQA Regulation Number: 21 CFR 870.2700 Review Panel: Anesthesiology
Intended Use Statement:
The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
Device Description
The proposed device, SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in noninvasive U.S. legally marketed oximeters or patient monitors. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models.
The proposed devices measure, along with the corresponding oximeters, non-invasively, the arterial oxygen saturation of blood. The measurement method is based on the red and infrared light absorption of hemoglobin and oxyhemoglobin. Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. The light is emitted through human finger and received by a photodiode. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
The patient contact components include cable and sensor. The contact level is surface device and contact duration is limited contact (less than 24h).
న్. Identification of Predicate Device
510(k) Number: K153184 Device Name: Caremed Reusable & Disposable SpO2 Sensors
-
- Non-Clinical Test Conclusion
5
Non clinical tests were conducted to verify that the proposed device met all design specifications and to demonstrate Substantially Equivalence(SE) to the predicate device. The test include following items
The performance tests performed on proposed devices include following items
-
Pulse and SpO2 Accuracy test
-
Pulse and SpO2 Accuracy test after High and Low Temperature
-
Pulse and SpO2 Accuracy test after Maximum Times Disinfection
Biocompatibility tests were conducted per ISO 10993-1, the test items include
- ▲ Cytotoxicity Test
-
Skin Irritation Test
-
Skin Sensitization Test
Electrical safety and electromagnetic compatibility test were conducted on proposed devices per IEC 60601-1, IEC 60601-1-2 and ISO 80601-2-61.
-
- Clinical Test Conclusion
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of proposed device versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
- Clinical Test Conclusion
-
- Substantially Equivalent (SE) Comparison
ITEM | Proposed device | Predicate Device |
---|---|---|
K153184 | ||
Product Code | DQA | DQA |
Regulation | ||
Number | 21 CFR 880.2910 | 21 CFR 880.2910 |
Intended Use | The devices are intended for spot | |
checking or continuous monitoring of | ||
functional arterial oxygen saturation | ||
and pulse rate in non-invasive | ||
oximeter equipment. The sensors are | ||
reusable or disposable depending on | ||
models, and are intended for adult, | ||
pediatric or infant, depending on | Caremed Reusable & Disposable | |
SPO2 Sensors are indicated for | ||
continuous non-invasive monitoring of | ||
functional oxygen saturation of arterial | ||
hemoglobin (SpO2) and pulse rate(PR) | ||
for adult patients weighing greater than | ||
40 kg and pediatric patients weighing | ||
10-50 kg at hospital facilities |
Table 1 Comparison of Technology Characteristics
6
| | models. They shall be used
in healthcare settings. | |
|----------------------------|-------------------------------------------------------|---------------------------------------------|
| Principle of
Operation | 2-wavelength Relative Optical
Absorption | 2-wavelength Relative Optical
Absorption |
| Light Emitting | Red: 660nm
Ired: 940nm | Red:660-666nm
Ired:880-950nm |
| Signal Detection
Method | Photodetector | Photodetector |
| SpO2 Range | 70-100% | 70-100% |
| SpO2 Accuracy | $\pm$ 3% @ 70-100% | $\pm$ 3% @ 70-100% |
| PR Range | 30-245 bpm | 30-250 bpm |
| PR Accuracy | $\pm$ 3 bpm | $\pm$ 3 bpm |
| Sterile | No | No |
| Usage | Reusable & Disposable | Reusable & Disposable |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 |
| Performance | Complied with ISO 80601-2-61 | Complied with ISO 80601-2-61 |
| Biocompatibility | | |
| Cytotoxicity | Complied with ISO 10993-5 | Complied with ISO 10993-5 |
| Skin Irritation | Complied with ISO 10993-10 | Complied with ISO 10993-10 |
| Sensitization | Complied with ISO 10993-10 | Complied with ISO 10993-10 |
-
- Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
- Substantially Equivalent (SE) Conclusion