K Number
K172957
Date Cleared
2018-09-28

(367 days)

Product Code
Regulation Number
864.9875
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Blood Drawing Kit with Antineedle Stick Protector is a single use, sterile transfer set designed for vascular access for blood withdrawal. Use of the Needleless Access Connector allows the clinician needle-free blood withdrawal when blood sampling is required. The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles. This device is not intended for blood transfusion.

Device Description

The Kawasumi Laboratories Blood Drawing Kit (BDK) System is a therapeutic device used to access a patient's vein for blood removal from the patient to a blood bag reservoir to aid in the treatment of a disease or other condition. The device is not intended for blood transfusion. The Kawasumi Laboratories Blood Drawing Kit (BDK) System consists of a Blood Drawing Kit device that is designed with an integral Antineedle Stick Protector that provides a safety feature intended to minimize accidental needle stick injuries when the device is activated during removal from the patient's vein and may incorporate a needleless access connector site used for blood sampling. The Blood Drawing Kit System is a single use, disposable Ethylene Oxide Sterilized medical device. The Kawasumi Laboratories Blood Drawing Kit (BDK) System is a sterile, single-use device that consists of the following components: Needle Protector, Cannula, Hub, Wing, Antineedle Stick Protector "K-Shield", Tubing, Needleless Access Connector (NAC) Y-Site, Clamp, Blood Bag Unit Reservoir.

AI/ML Overview

The provided text describes a 510(k) summary for the Kawasumi Laboratories Blood Drawing Kit (BDK) System. It lists various non-clinical performance tests conducted to demonstrate substantial equivalence to a predicate device, but it does not include acceptance criteria or reported device performance in a table format, nor does it detail a study that proves the device meets specific acceptance criteria in the way typically expected for an AI/ML device.

The document focuses on demonstrating that the new device is substantially equivalent to a previously cleared device (K001043) by showing similar technological characteristics and by conducting standard performance and biocompatibility tests for medical devices. It does not describe a study that uses a 'test set' with 'ground truth' established by experts, as would be relevant for an AI/ML device's performance evaluation.

Therefore, many of the requested details (like sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, training set size, and ground truth establishment for training) are not applicable to the information provided in this document, which pertains to a traditional medical device (a blood drawing kit) rather than an AI/ML-powered one.

However, I can extract the information provided about the non-clinical performance testing which serves as the "study" for this device, and list the categories of tests that were performed.


1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria with specific numerical targets and corresponding reported device performance values. Instead, it lists categories of non-clinical tests that were performed to demonstrate substantial equivalence. The implication is that the device passed these tests, thereby meeting the implicit acceptance criteria for each test type, which are typically defined by the referenced standards.

Acceptance Criteria Category (Implicit, based on standards)Reported Device Performance (Implicit: Passed all tests)
Biocompatibility (Cytotoxicity, Skin Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen, Hemolysis)Passed
Functional Performance (Visual Inspection, Leakage, Flow Regulator, Volume Marking Accuracy, Blockage, K-Shield Activation, NAC Function, Pull Strength, Sharpness)Passed
Chemical Performance (Reducing Matter, Metal Ions, Titration Acidity/Alkalinity, Non-Volatile Residue, Absorbance)Passed
Sterilization/Sterility (Residual Gas, Material Mediated Pyrogen, Particulate, Sterilization Validation, Packaging)Passed

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify
sample sizes for the individual non-clinical performance tests nor the data provenance. These tests are typically conducted in a laboratory setting by the manufacturer or a contract research organization.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The device is a traditional physical medical device (blood drawing kit), not an AI/ML device that requires expert-established ground truth for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an AI/ML device for which adjudication methods for expert disagreement would be relevant.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a traditional medical device; MRMC studies relate to AI/ML image interpretation or diagnostic tools.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a traditional medical device; standalone algorithm performance is relevant for AI/ML devices.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For a blood drawing kit, "ground truth" would relate to meeting physical, chemical, and biological performance specifications outlined in recognized standards. This is demonstrated through objective laboratory testing rather than expert consensus on diagnostic outcomes. The "ground truth" is adherence to established engineering and medical device standards.

8. The sample size for the training set

Not applicable. This is a traditional medical device, not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is a traditional medical device, not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 28, 2018

Kawasumi Laboratories, Inc. % Lisa Michels Regulatory Compliance Associates, Inc. 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158

Re: K172957

Trade/Device Name: Kawasumi Laboratories Blood Drawing Kit (BDK) System Regulation Number: 21 CFR 864.9875 Regulation Name: Transfer set Regulatory Class: Class II Product Code: KSB, LHI, FMI Dated: August 21, 2018 Received: August 28, 2018

Dear Lisa Michels:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172957

Device Name

Kawasumi Laboratories Blood Drawing Kit (BDK) System

Indications for Use (Describe)

The Blood Drawing Kit with Antineedle Stick Protector is a single use, sterile transfer set designed for vascular access for blood withdrawal. Use of the Needleless Access Connector allows the clinician needle-free blood withdrawal when blood sampling is required. The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles. This device is not intended for blood transfusion.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY (K172957)

l. SUBMITTER

Owner/Manufacturer:

Kawasumi Laboratories, Inc. Shinagawa Intercity Tower B, 9th Floor 2-15-2, Konan, Minato-Ku, Tokyo, 108-6109, Japan Phone: (81)-3-5769-2663 Contact Person: Katsu Furuya

Date Prepared: September 12, 2017

ll. DEVICE

Trade Name of Device: Kawasumi Laboratories Blood Drawing Kit (BDK) System Common or Usual Name of Device: Blood Drawing Kit System Classification Name: Set, Transfer (Blood Plasma)

Primary Classification Product Code and Classification Regulation:

  • KSB (Set, Transfer Blood/Plasma -21 CFR 864.9875) ●

Subsequent Classification Product Code(s) and Classification Regulation(s):

  • . LHI (Set I.V Fluid Transfer - 21 CFR 880.5440)
  • FMI (Needle, Hypodermic, Single Lumen 21 CFR 880.5570) .

Device Class: Class II

lll. PREDICATE DEVICE(S)

The Blood Drawing Kit (BDK) System is substantially equivalent to Kawasumi Laboratories' own predicate device(s) including the following Primary Predicate Device:

. Kawasumi Laboratories Blood Drawing Kit - cleared under (K001043)

IV. DEVICE DESCRIPTION

The Kawasumi Laboratories Blood Drawing Kit (BDK) System is a therapeutic device used to access a patient's vein for blood removal from the patient to a blood bag reservoir to aid in the treatment of a disease or other condition. The device is not intended for blood transfusion.

The Kawasumi Laboratories Blood Drawing Kit (BDK) System consists of a Blood Drawing Kit device that is designed with an integral Antineedle Stick Protector that provides a safety feature intended to minimize accidental needle stick injuries when the device is activated during removal from the patient's vein and may incorporate a needleless access connector site used for blood sampling.

{4}------------------------------------------------

The Blood Drawing Kit System is a single use, disposable Ethylene Oxide Sterilized medical device. Sterilization Validation was conducted per ISO 11135-1: 2007 – Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirement, Validation, and Routine Control of a Sterilization Process for Medical Devices. The Blood Drawing Kit System is considered blood path, indirect, limited contact (< 24 hours) per section 5.2.2 (a) of ISO 10993-1: 2009 - Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a Risk Management Process.

The Kawasumi Laboratories Blood Drawing Kit (BDK) System is a sterile, single-use device that consists of the following components:

*(Y) means the fluid contact components

  • Needle Protector ●
  • . Cannula (Y)
  • . Hub (Y)
  • Wing (Y skin contact only) .
  • Antineedle Stick Protector "K-Shield"
  • . Tubing (Y)
  • . Needleless Access Connector (NAC) Y-Site (Y)
  • . Clamp
  • . Blood Bag Unit Reservoir (Y)

The device is intended for use at healthcare facilities or in hospitals.

V. INDICATIONS FOR USE

The Blood Drawing Kit with Antineedle Stick Protector is a single use, sterile transfer set designed for vascular access for blood withdrawal. Use of the Needleless Access Connector allows the clinician needlefree blood withdrawal when blood sampling is required. The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles. The device is not intended for blood transfusion.

VI. COMPARISON OF TECHNOLOGICAL CHARAGTERISTICS WITH THE PRIMARY PREDICATE DEVICE

The Comparison of Technological Characteristics of the Subject Device and the Primary Predicate Device (K001043) are shown in the table below:

Kawasumi Laboratories Blood Drawing Kit (BDK) SystemTable of Substantial Equivalence
REGULATORY INFORMATION
Device Name[Proposed Subject Device][Primary Predicate Device]Similarities/Differences
Kawasumi LaboratoriesBlood Drawing Kit (BDK) SystemKawasumi LaboratoriesBlood Drawing Kit (K001043)
510(k) StatusPending - K172957K001043Same
DeviceClassificationNameSet, Transfer(Blood, Plasma)Set, Transfer(Blood, Plasma)Same
DeviceClassificationClass IIClass IISame
ProductCode andRegulationNumber(KSB)21 CFR 864.9875(KSB)21 CFR 864.9875Same
SubsequentProduct Code(s)and RegulationNumber(s)(LHI)21 CFR 880.5440(FMI)21 CFR 880.5570(LHI)21 CFR 880.5440Same
DESIGN FEATURES AND CAPABILITIES OF THE DEVICE
IntendedUse /Indicationfor UseThe Blood Drawing Kit with Antineedle StickProtector is a single use, sterile setdesigned for vascular access for bloodwithdrawal. Use of the Needleless AccessConnector allows the clinician needle-freeblood withdrawal when blood sampling isrequired. The Antineedle Stick Protector isan integral, active safety device intended tominimize accidental needle stick injurieswhen used to shield needles. This device isnot intended for blood transfusion.The Blood Drawing Kit is a sterile, single-usedevice used for therapeutic purposes to collectpatient's blood into a blood bag reservoir. Thecollected blood is properly discarded and notintended for transfusion.This 510(k) Submission did not originallyinclude the Antineedle Stick Protector. It wascleared for use with the BDK under K073257and K102994.BDK same asK001043; AntineedleStick Protector sameas K073257 andK102994
DESIGN FEATURES
Dimensions,Gauge, Length,WidthNeedle Gauge: 16Length: 1450 mm including needle cap)Needle Gauge: 16Length: 1450 mm including needle cap)Same
VenipunctureNeedleOption with the pre-attached venipunctureneedle or option without the pre-attachedvenipuncture needleOption with the pre-attached venipunctureneedle or option without the pre-attachedvenipuncture needleSame
AntineedleStickProtector(ANSP)OptionAntineedle Stick Protector (ANSP) Option.NoneThe Antineedle StickProtector Option wascleared for use with theBlood Drawing Kitunder K073257 andK102994.
InjectionOptionNeedleless Access Connector (NAC) Y-SiteOptionNoneNAC Y-Site Option isthe Same as K073257and K102994.The Needleless AccesConnector (NAC) Y-Site Option wascleared for use with theBlood Drawing Kitunder K073257 andK102994. Thisdifference does notraise new questions ofsafety andeffectiveness.
SterilizationMethodEthylene Oxide Gas (ETO)Autoclave (AC)Different Method.SAL is identical.Differences do notraise new questions ofsafety andeffectiveness
MATERIALS
NeedleProtectorPolyethylene (PE)Polyethylene (PE)Polyvinyl Chloride (PVC)Poly-Propylene (PP)Differences do notraise new questions ofsafety andeffectiveness
CannulaStainless SteelStainless SteelSame
HubPolycarbonate (PC)Polyvinyl Chloride (PVC)Differences do notraise new questions ofsafety andeffectiveness
WingPolyvinyl Chloride (PVC)NoneSame as K073257 andK102994
K-ShieldPolypropylene (PP)NoneSame as K073257 andK102994
TubingPolyvinyl Chloride (PVC)Polyvinyl Chloride (PVC)Same
InjectionSite(NAC-Y site)Polycarbonate /Synthetic PolyisopreneNoneSame as K073257 andK102994
ClampPolypropylene (PP)Polypropylene (PP)Same
Bag UnitPolyvinyl Chloride (PVC)Polyvinyl Chloride (PVC)Same
DISCUSSION OF TECHNOLOGICAL SIMILARITIES AND DIFFERENCES

{5}------------------------------------------------

{6}------------------------------------------------

Technological Characteristics and Substantial Equivalence

The Blood Drawing Kit (BDK) System is substantially equivalent in both function and use to the following Primary Predicate Device – Kawasumi Laboratories Blood Drawing Kit (cleared under K001043) as noted below:

  • Kawasumi Laboratories Blood Drawing Kit (Primary Predicate Device cleared under K001043); and the ●
  • Components including the K-Shield Antineedle Stick Protector (ANSP Option) and the Needleless Access Connector Site ● (NAC- Y Injector Site Option) used for blood sampling cleared under (K073257) and (K102994) for use with the Blood Drawing Kit (Primary Predicate Device cleared under K001043)]

These differences do not raise new questions of safety and effectiveness over the Primary Predicate Device. Therefore, the Blood Drawing Kit (BDK) System is substantially equivalent to the Primary Predicate Device based on the intended use of the device.

Summary ofNonclinicalPerformanceTestingThe determination of substantial equivalence was based on an assessment of non-clinical performance data includingthe following nonclinical tests that have been submitted, referenced, and relied on in this premarket notificationsubmission for a determination of substantial equivalence.
[Biocompatibility Testing]
The biocompatibility evaluation for the Blood Drawing Kit System was conducted in accordance with the followingguidance documents and standards as recognized by the FDA:
FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, "Biological Evaluation ofMedical Devices Part 1: Evaluation and Testing," issued May 1, 1995;
ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a RiskManagement Process"
ISO 10993-2 Biological Evaluation of Medical Devices - Part 2: Animal Welfare Requirements
ISO 10993-4 "Biological Evaluation of Medical Devices – Part 4: Selection of Tests for Interaction with Blood[Including: Amendment 1 (2006)]
ISO 10993-5 "Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
ISO 10993-10 "Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization
ISO 10993-11 "Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity
ISO 10993-12 "Biological Evaluation of Medical Devices – Part 12: Sample Preparation and ReferenceMaterials
The following non-clinical performance testing data were provided in support of the substantial equivalencedetermination:
Cytotoxicity Testing
Skin Sensitization Testing
Intracutaneous Reactivity Testing
Acute Systemic Toxicity Testing

{7}------------------------------------------------

  • Pyrogen Testing
  • . Hemolysis Testing

[Functional Performance Testing]

The functional performance testing for the Blood Drawing Kit System was conducted in accordance with the following guidance documents and standards as recognized by the FDA:

  • FDA Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features, issued on ● August 9, 2005
  • . ISO 3826-1:2003 - Plastic Collapsible Containers for Blood and Blood Components
  • ISO 1135-3:1986 Transfusion Equipment for Medical Use Part 3: Blood-Taking Sets .
  • ISO 1135-4:2012 Transfusion Equipment for Medical Use Part 4: Transfusion Sets for Single Use .
  • ISO 594-2: Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Medical . Equipment - Part 2: Lock Fittings

The following non-clinical performance testing data were provided in support of the substantial equivalence determination:

  • Physical Test for BDK (Non-Aging, 6 Months Aging and 3 Years Aging)
    • Visual Inspection of Set and Packing —
    • Test for Leakage
    • Flow Regulator (Clamp Test)
    • Volume Marking Accuracy Test & Blockage Test
    • Function of Large K-Shield Antineedle Stick Protector (Activation Test)
    • -Function Test of NAC Needleless Access Connector (NAC) Y-Site Option
    • Pull Strength Test
    • Sharpness Test
  • Chemical Test for BDK (Non-Aging, 6 Months Aging and 3 Years Aging)
    • Test for Reducing (Oxidizable) Matter
    • Test for Metal Ions
    • Test for Titration Acidity or Alkalinity
    • Test for Non-Volatile Residue ー
    • -Test for Absorbance

[Sterilization Testing / Sterility Testing / Packaging Testing]

The sterilization testing / sterility testing for the Blood Drawing Kit System was conducted in accordance with the following guidance documents and standards as recognized by the FDA:

  • ISO 10993-7 "Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals . [Including Technical Corrigendum 1 (2009)]
  • . ISO 11135: 2014: Sterilization of Health Care Products – Ethylene Oxide – Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices
  • ISO 11138-1: Sterilization of Health Care Products – Biological Indicators – Part 1: General Requirements
  • ISO 11607-1: Packaging for Terminally Sterilized Medical Devices Part 1: Requirements for Sterile Barrier . Systems and Packaging Systems [including Amendment 1 (2014)]
  • . ISO 11607-2: Packaging for Terminally Sterilized Medical Devices – Part 2: Validation Requirements for Forming Sealing and Assembly Processes [including Amendment 1 (2014)]
  • . ISO 11737-1: Sterilization of Medical Devices – Microbiological Methods – Part 1: Determination of a Population of Microorganisms on Products [including Technical Corrigendum 1 (2007)]

{8}------------------------------------------------

ISO 14644-1: Cleanrooms and Associated Controlled Environments – Part 1: Classification of Air Cleanlinessby Particle Concentration ISO 14622-2: Cleanrooms and Associated Controlled Environments – Part 2: Monitoring to Provide Evidenceof Cleanroom Performance related to Air Cleanliness by Particle Concentration The following non-clinical performance testing data were provided in support of the substantial equivalence determination: Material Mediated Pyrogen Testing Particulate Testing Sterilization Validation Testing Residual Gas Analysis Testing Packaging Testing
ConclusionsThe non-clinical performance data demonstrate that the Blood Drawing Kit (BDK) System is substantially equivalent to the Primary Predicate Device.

§ 864.9875 Transfer set.

(a)
Identification. A transfer set is a device intended for medical purposes that consists of a piece of tubing with suitable adaptors used to transfer blood or plasma from one container to another.(b)
Classification. Class II (performance standards).