K Number
K172895

Validate with FDA (Live)

Date Cleared
2018-06-18

(269 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AGE Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual.

Device Description

AGE Automatic Upper Arm Blood Pressure Monitor is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at arm within its claimed range and accuracy via the Oscillometry technique. The device also has low voltage indication, which will be triggered when the battery is low.

AI/ML Overview

The provided text describes a 510(k) submission for an Automatic Upper Arm Blood Pressure Monitor. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria based on a clinical trial.

Therefore, many of the requested details about acceptance criteria, study design, expert ground truth, and human-in-the-loop performance are not available in this document, as they are typically associated with performance validation studies for AI/ML devices or new diagnostic methods, not blood pressure monitors seeking 510(k) clearance based on substantial equivalence to an existing device.

However, I can extract the relevant information regarding performance standards and comparisons that are mentioned:

1. A table of acceptance criteria and the reported device performance

The document references the standard AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type as the key performance standard.

While explicit acceptance criteria are not listed in a table, the standard itself defines accuracy requirements for blood pressure monitors. The device claims the following accuracy, which would implicitly be the performance criteria it aims to meet based on the standard:

CriterionAcceptance / TargetReported Performance
Pressure Accuracy±3 mmHg±3 mmHg
Pulse Accuracy±5%±5%

2. Sample size used for the test set and the data provenance

The document states that the device was evaluated for "safety and performance by lab bench testing" and refers to "Clinical Validation Of Automated Measurement Type" according to ISO 81060-2.

  • Sample Size: The document does not explicitly state the sample size (number of subjects/measurements) used for the clinical validation. ISO 81060-2 typically specifies minimum patient numbers (e.g., 85 subjects for validation). Without the full study report, the exact sample size cannot be determined from this document.
  • Data Provenance: Not explicitly stated. Clinical validation studies for such devices typically involve prospective data collection from human subjects. The document does not specify the country of origin.
  • Retrospective/Prospective: Clinical validation according to ISO 81060-2 is inherently prospective, involving live measurements on human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

For blood pressure monitor validation (ISO 81060-2), ground truth is established by simultaneous auscultatory measurements performed by trained observers (commonly two, with a third if disagreements arise) using a reference sphygmomanometer. These are not "experts" in the sense of radiologists interpreting images, but rather trained technicians or clinicians. The document does not specify the exact number of observers or their specific qualifications, but the standard mandates specific training and performance for these observers.

4. Adjudication method for the test set

The ISO 81060-2 standard typically employs an adjudication method for reconciling reference blood pressure measurements. Commonly, two trained observers simultaneously measure, and if their readings differ by more than a predefined threshold, a third observer might be used, or the measurements are discarded and re-taken. This is inherent to the standard's methodology, though not explicitly detailed in this summary document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC study was not done. This type of study is relevant for diagnostic devices, particularly those involving human interpretation, often with AI assistance (e.g., radiology AI). This device is an automatic blood pressure monitor, not an interpretive diagnostic system.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, in essence. The primary performance evaluation for an automatic blood pressure monitor is its standalone accuracy against a reference method (auscultation). The device itself is designed to operate automatically, without continuous human intervention during the measurement cycle for interpretation. The reported accuracies (±3 mmHg for pressure, ±5% for pulse) reflect this standalone performance.

7. The type of ground truth used

The ground truth for non-invasive blood pressure monitors typically involves simultaneous auscultatory measurements performed by trained observers using a mercury or an equivalent validated reference sphygmomanometer. This is the "gold standard" for clinical validation of automated blood pressure devices as per ISO 81060-2.

8. The sample size for the training set

This document does not specify a separate "training set" sample size. For traditional medical devices like blood pressure monitors, there isn't typically a distinct "training set" in the machine learning sense. The device's algorithm (oscillometric method) is developed based on physiological principles and engineering, not "trained" on a large dataset of patient readings in the way an AI algorithm learns. The validation is done on a "test set" or clinical study population.

9. How the ground truth for the training set was established

As there isn't a "training set" in the AI/ML sense, this question is not applicable. The device's internal algorithms are based on established oscillometric principles for blood pressure determination. Their accuracy is then validated (tested) against the clinical ground truth described in point 7.

{0}------------------------------------------------

June 18, 2018

Ageless Health Industrial Co., Ltd. % Cassie Lee Manager Guangzhou Glomed Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, 510000 China

Re: K172895

Trade/Device Name: AGE Automatic Upper Arm Blood Pressure Monitor with Models BA-815, BA-816, BA-818 and BA-819 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 8, 2018 Received: May 15, 2018

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M. A. Wilhelm

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172895

Device Name

AGE Automatic Upper Arm Blood Pressure Monitor with Models BA-815, BA-816, BA-819

Indications for Use (Describe)

AGE Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Sponsor:Ageless Health Industrial Co., Ltd
Subject Device:AGE Automatic Upper Arm Blood Pressure Monitor Model: BA-815, BA-816, BA-818, BA-819
File No.:

File No.:

Chapter 5. 510(k) Summary

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: Ageless Health Industrial Ltd. Establishment Registration Number: Applying Address: 3/F, A1 Bldg, Dongshen Sima Industrial Area, No.33 Shenbei Road, Sima Village, Changping Town, Dongquan, Guangdong, China Contact Person: Victor Wan (Vice-president) Email: victor@agelh.com

Application Correspondent:

Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com

2. Subject Device Information

Common Name: System, Measurement, Blood-Pressure, Non-Invasive Trade Name: AGE Automatic Upper Arm Blood Pressure Monitor Models: BA-815, BA-816, BA-818, BA-819 Review Panel: Cardiovascular Product Code: DXN Regulation Number: 21 CFR 870.1130 Regulation Class: 2

3. Predicate Device Information

SponsorAgeless Health Industrial Ltd.
Device NameAGE Automatic Upper Arm Blood Pressure Monitor
510(k) NumberK153552
Product CodeDXN
Regulation Number21 CFR 870.1130
Regulation ClassII

{4}------------------------------------------------

4. Device Description

AGE Automatic Upper Arm Blood Pressure Monitor is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at arm within its claimed range and accuracy via the Oscillometry technique. The device also has low voltage indication, which will be triggered when the battery is low.

5. Intended Use / Indications for Use

AGE Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual.

6. Test Summary

effectiveness.

AGE Automatic Upper Arm Blood Pressure Monitor has been evaluated the safety and performance by lab bench testing according to the following standards:

  • ◆ IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 2005 + A1:2012
  • ◆ IEC 60601-1-2. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements and tests, 2014
  • ◆ IEC 60601-1-11, Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment [Including: Technical Corrigendum 1 (2011)]
  • ◆ IEC 80601-2-30, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers
  • � ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
  • ◆ ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
  • ◆ AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type. (Cardiovascular)

7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, and intended use is substantially equivalent to the predicate devices quoted above. The differences between the subject devices do not raise new issues of safety or

{5}------------------------------------------------

Sponsor:Ageless Health Industrial Co., Ltd
Subject Device:AGE Automatic Upper Arm Blood Pressure Monitor Model: BA-815, BA-816, BA-818, BA-819
File No.:K172895
Elements ofComparisonSubject DevicePredicate DeviceVerdict
ProductNameModel NameAGE Automatic Upper Arm BloodPressure MonitorBA-815, BA-816, BA-818, BA-819AGE Automatic Upper Arm BloodPressure MonitorBA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X--
Intended UseAGE Automatic Upper Arm BloodPressure Monitor is intended for useby medical professionals or at home tomonitor and display diastolic, systolicblood pressure and pulse rate on adulteach time, with the cuff around the leftupper arm according to the instructionin the user's guide manual.AGE Automatic Upper Arm BloodPressure Monitor is intended foruse by medical professionals or athome to monitor and displaydiastolic, systolic blood pressureand pulse rate on adult each time,with the cuff around the left upperarm according to the instruction inthe user's guide manual.SE
PowerSupplyDC 6V (4 X AA 1.5V alkline batteries)6Vdc (4 "AA" batteries)SE
MeasuringMethodOscillometryOscillometrySE
MeasuringRangePressure: 0280 mmHgPulse: 40199 beats/minuteSYS(systolic pressure ): 60-255mmHgDIA (diastolic pressure): 40-200mmHgPressure: 0280 mmHgPulse: 40199 beats/minuteSYS(systolic pressure ): 60-255mmHgDIA (diastolic pressure): 40-200mmHgSE
Pressureresolution1mmHg or 0.1kPa1mmHg or 0.1kPaSE
AccuracyPressure: ±3 mmHgPulse: ±5%Pressure: ±3mmHgPulse: ±5%SE
PatientPopulationAdultAdultSE
Measurement Site of BodyUpper ArmUpper ArmSE
Inflation andDeflationAutomaticAutomaticSE
Memory Size2 x 90 sets record2 x 90 sets recordSE
SoftwareVersionV01V01SE
IndicatorsBlood Pressure (Systolic andDiastolic), Pulse, Date, Time, WHO BPClassification Indicating Bar, LowBattery Icon, Heart Icon, MemoryBlood Pressure (Systolic andDiastolic), Pulse, Date, Time, WHOBP Classification Indicating Bar,Low Battery Icon, Heart Icon,MemorySE

{6}------------------------------------------------

Sponsor:Ageless Health Industrial Co., Ltd
Subject Device:AGE Automatic Upper Arm Blood Pressure Monitor Model: BA-815, BA-816, BA-818, BA-819
File No.:K172895
Elements ofComparisonSubject DevicePredicate DeviceVerdict
CuffCircumferenceFor BA-815: 22-34 cm;For BA-816: 28-42 cm;For BA-818 and BA-819, there are 6size:size A: 17cm--22cm (SMALL ADULT CUFF)size B: 22cm--30cm (ADULT CUFF-1)size C: 24cm--34cm (ADULT CUFF-2)size D: 22cm--42cm (L-LARGE ADULT CUFF)size E: 30cm--42cm (LARGE ADULT CUFF)size F: 42cm--50cm (EXTRA LARGE ADULT CUFF)Size A: 17cm--22cm (SMALL ADULT CUFF)Size B: 22cm--30cm (ADULT CUFF-1)Size C: 24cm--34cm (ADULT CUFF-2)Size D: 22cm--42cm (L-LARGE ADULT CUFF)Size E: 30cm--42cm (LARGE ADULT CUFF)Size F: 42cm--50cm (EXTRA LARGE ADULT CUFF)SENote 1
OPERATING & STORAGE CONDITIONS
StorageEnvironmentTemperature: -20°C ~ +65°CHumidity: 1095%RHAtmospheric Pressure:86 kPa106 kPaTemperature: -20°C ~ +65°CHumidity: 1095%RHAtmospheric Pressure:86 kPa106 kPaSE
WorkingEnvironmentTemperature: 5°C ~ 40°CHumidity: 1590%RHAtmospheric Pressure:86 kPa106 kPaTemperature: 5°C ~ 40°CHumidity: 1590%RHAtmospheric Pressure:86 kPa106 kPaSE
COMPLIANCE STANDARDS
Electrical,Mechanicaland ThermalEvaluationIEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 80601-2-30IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 80601-2-30SE
Biocompatibility EvaluationAll the patient contracting materials areevaluated by the biocompatibilitystandard ISO 10993 - 5, -10.All the patient contracting materialsare evaluated by thebiocompatibility standard ISO10993 - 5, -10.SE
PerformanceISO 81060-2-30ISO 81060-2-30SE

Comparison in Detail(s):

Note 1:

Although there is difference for measurement cuff circumference of subject device and predicate device, both of them are complied with ISO 81060-2. This difference does not affect the safety and effectiveness.

Conclusion:

The subject device AGE Automatic Upper Arm Blood Pressure Monitor has all features of the predicate device. The differences between them do not affect the safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device.

{7}------------------------------------------------

Sponsor:Ageless Health Industrial Co., Ltd
Subject Device:AGE Automatic Upper Arm Blood Pressure Monitor Model: BA-815, BA-816, BA-818, BA-819
File No.:K172895

8. Date of the summary prepared: September 16, 2017

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).