K Number
K172509
Date Cleared
2018-02-01

(164 days)

Product Code
Regulation Number
866.3305
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sentosa® SA201 HSV-1/2 PCR Test is a real-time PCR-based qualitative in vitro diagnostic test for detection and differentiation of Herpes Simplex Virus (HSV-1 and HSV-2) DNA from male and female skin lesions from anogenital or oral sites. The test is intended for use as an aid in diagnosis of herpes infection in symptomatic patients.

Warning: The Sentosa® SA201 HSV-1/2 PCR Test is not FDA cleared for use with cerebrospinal fluid (CSF). The test is not intended to be used for prenatal screening.

Device Description

The Sentosa® SA201 HSV-1/2 PCR Test is a (4x24) configuration contains reagents and enzymes for specific amplification of a 104 bp (base-pair) fragment of the UL30 gene common to both HSV1 and HSV2, and specific probes for the direct detection and differentiation of HSV1 and HSV2 amplicons, respectively. Pathogen detection by PCR is based on the amplification of specific regions of the pathogen genome. In real-time PCR, the amplified product is detected via fluorescent dyes, which are usually linked to oligonucleotide probes that bind specifically to the target sequences. Real-time monitoring of the fluorescence intensities during a PCR run allows the detection of the accumulating product. Amplification of the targets occurs in three channels: green, orange and red on the Sentosa® SA201. Output is recorded as the increase of fluorescence over time in comparison to background signal. Monitoring the fluorescence intensities during the PCR run allows the detection of the accumulating product without having to re-open the reaction tubes after the PCR run.

The Sentosa® SA201 HSV-1/2 PCR Test workflow starts with extraction of nucleic acids from samples (anogenital or oral swabs) using the Sentosa® SX Virus Total Nucleic Acid Kit on the Sentosa SX101 instrument. Following extraction, the instrument will automatically set up the PCR with the extracted nucleic acids in a 96-well PCR plate. Subsequently, the 96-well PCR plate is sealed and transferred to the Sentosa® SA201 for PCR amplification, followed by data analysis.

The Sentosa® Link facilitates data transfer between the Sentosa® SX101, the Sentosa® SA201 Reporter and existing LIS/LIMS (laboratory information systems) in the clinical lab. The Sentosa SX101 instrument communicates with Sentosa® SA201 thermocycler. This creates a user environment that links the SX101 and the Sentosa® SA201 to facilitate automated workflow to export results in a LIS/LIMS-compatible format.

AI/ML Overview

The provided document describes the analytical and clinical performance of the Sentosa SA201 HSV-1/2 PCR Test, a qualitative in vitro diagnostic test for the detection and differentiation of Herpes Simplex Virus (HSV-1 and HSV-2) DNA. It is not an AI/ML-based device. Therefore, the questions related to AI/ML specific aspects (e.g., number of experts, adjudication, MRMC study, training set, ground truth for training set) are not applicable. The information focuses on the device's ability to accurately detect and differentiate HSV-1 and HSV-2 DNA in patient samples.

Here's an analysis of the provided information, focusing on the device's acceptance criteria and the studies proving it meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a single table, but the performance studies demonstrate implied criteria. For diagnostic devices like this, the key performance metrics are sensitivity, specificity, limit of detection (LoD), precision, and freedom from interference.

Study/Performance MetricImplied Acceptance CriterionReported Device Performance
Limit of Detection (LoD)LoD should be low enough for clinical utility (e.g., detected with ≥95% probability).HSV-1 MacIntyre & KOS: 40 TCID50/mL (detected with 95% probability).HSV-2 MS & G: 4 TCID50/mL (detected with 95% probability).
PrecisionConsistent results across runs, operators, and instruments (e.g., agreement 100%, %CV < 10%).Agreement (all tested samples): 100% for all concentration levels (1.5x LoD, 3x LoD, NC, PC, Negative sample) across channels (Green, Orange, Red). One invalid sample excluded for 3x LoD HSV-2b reading. %CV of Ct values: All %CV values were less than 10%. (e.g., 1.5x LoD HSV-1a: 0.81 SD, 3x LoD HSV-1a: 0.90 SD, 1.5x LoD HSV-2b: 1.09 SD, 3x LoD HSV-2b: 0.54 SD, NC: 2.24 SD, PC Green: 1.10 SD, PC Orange: 0.69 SD, Negative Red: 2.03 SD).
ReproducibilityConsistent results across sites, operators, and lots (e.g., agreement 100%, %CV < 10%).Agreement (all tested samples): 100% for all concentration levels (1.5x LoD, 3x LoD, NC, PC, Negative sample, Blank) across channels (Green, Orange, Red). %CV of Ct values: All %CV values were less than 10% (e.g., 1.5x LoD HSV-1a: 5.25%, 3x LoD HSV-1a: 3.34%, 1.5x LoD HSV-2b: 1.60%, 3x LoD HSV-2b: 4.21%, NC: 2.23%, PC Green: 1.01%, PC Orange: 1.13%, Negative Red: 4.20%, Blank: 2.05%).
Analytical Reactivity / Cross-reactivity (Specificity)No cross-reactivity with closely related organisms or common flora/pathogens; no interference with HSV-1/2 detection.Cross-reactivity: No cross-reactivity observed with 55 tested organisms (closely related to HSV-1/2 or present in oral/genital swabs). No cross-reactivity within the multiplex panel (HSV-1 and HSV-2) in presence of high concentration HSV-1 or HSV-2. Interference: C. glabrata, S. aureus, S. epidermidis, P. melaninogenica and HHV6 might lead to slight competitive inhibition from HSV-1 to HSV-2 (this finding implies a potential area for further characterization, but the main finding is no cross-reactivity for the 55 organisms).
Interfering SubstancesNo interference on performance from common substances at elevated concentrations.No interfering effects at concentrations 5-10 times higher than normal active concentration for 31 common substances in genital/oral specimens. HSV-1 detection not interfered by HSV-2 up to 50xLoD of HSV-1; HSV-2 detection not interfered by HSV-1 up to 10xLoD of HSV-1. No mutual interference observed up to 500xLoD of each virus at equal concentrations.
Carry-over and Cross-contaminationContamination rate should be minimal (e.g., 0%).Overall contamination rate of 0% in 12 runs with high positive HSV-1 samples (1x10^5 TCID/mL). All 96 positive samples detected as positive, and all 183 negative samples detected as negative (no amplification signal).
Clinical Sensitivity & SpecificityHigh agreement with a legally marketed predicate device (ELVIS® HSV ID and D3 Typing Test System) for detection of HSV-1 and HSV-2 in anogenital and oral lesion samples.Anogenital Lesions (N=1581 for HSV-1, N=1978 for HSV-2):- HSV-1: Sensitivity 96.90% (95% CI: 94.21%-98.36%), Specificity 95.82% (95% CI: 94.58%-96.78%). - HSV-2: Sensitivity 98.49% (95% CI: 96.74%-99.31%), Specificity 90.70% (95% CI: 89.17%-92.04%).Oral Lesions (N=314 for HSV-1, N=317 for HSV-2):- HSV-1: Sensitivity 100.00% (95% CI: 95.36%-100.00%), Specificity 86.38% (95% CI: 81.41%-90.19%).- HSV-2: Sensitivity 66.67% (95% CI: 20.77%-93.85%), Specificity 99.68% (95% CI: 98.22%-99.94%).HSV-2 Oral Lesion Contrived Specimen Study: 100% detection of 30 HSV-2 contrived samples; 100% correct identification of 15 HSV-1 positive and 15 HSV-1/2 negative samples.

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Study):

    • Total Samples Enrolled: 2684
    • Samples in Final Analysis: 2295 (389 excluded due to unspecified reasons)
    • Oral Lesions: 317
    • Anogenital Lesions: 1978
    • Data Provenance: Residual anogenital lesion or oral lesion samples from male and female patients with signs and symptoms of HSV infections. Collected from 8 sites in the USA. Samples were tested either at the same facility they were obtained or shipped to a different testing site.
    • Retrospective/Prospective: The text does not explicitly state retrospective or prospective. However, "residual samples" suggest a retrospective collection. "Samples were either tested at the same facility at which they were obtained, or were shipped to a different testing site" could imply some forward-looking testing strategy, but it's not clearly defined as a prospective study. Given the clinical study results are compared against a "reference test," it most likely involves previously collected and banked samples.
  • Test Set (Analytical Studies):

    • Limit of Detection: Determined by Probit analysis; specific sample numbers not explicitly given beyond "multiple replicates" across concentrations.
    • Precision: 60 replicates per sample type (NC, PC, 3xLoD HSV-1, 1.5xLoD HSV-1, 3xLoD HSV-2, 1.5xLoD HSV-2).
    • Reproducibility: 90 replicates per sample type (same as precision).
    • Analytical Reactivity / Cross-reactivity: 55 organisms tested, plus high and low concentration HSV-1/2 samples with competing organisms. Specific replicate numbers not given for all reactivity tests.
    • Interfering Substances: Each of 31 substances assayed in triplicate. Competitive interference study used 20 replicates for each combination.
    • Carry-over and Cross-contamination: 12 runs; 96 positive samples, 183 negative samples.
  • Training Set: Not applicable for this type of PCR diagnostic device. This device is based on a well-defined molecular biology assay (real-time PCR) and does not involve AI/ML models that require dedicated training data sets in the typical sense. Performance is assessed through analytical validation (LoD, precision, specificity) and clinical concordance with a reference method.

3. Number of Experts and Qualifications to Establish Ground Truth

Not applicable. This is a molecular diagnostic test, not an image-based AI/ML device relying on expert human interpretation for ground truth.

4. Adjudication Method for the Test Set

Not applicable. Ground truth for the clinical study was established by a "reference test" (ELVIS® HSV ID and D3 Typing Test System) and supplemented by "bi-directional sequencing analysis" for discordant samples. This is a confirmatory molecular method, not human adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a diagnostic test; it is not described as providing AI assistance to human readers.

6. Standalone (Algorithm Only) Performance

The entire performance evaluation of the Sentosa SA201 HSV-1/2 PCR Test represents its "standalone" performance, as it is a fully automated system from nucleic acid extraction to result reporting. There is no human interpretative step described for the final result.

7. Type of Ground Truth Used

  • Clinical Study: The primary ground truth for the clinical study was the ELVIS® HSV ID and D3 Typing Test System. For samples with discordant results between the Sentosa test and the ELVIS system, bi-directional sequencing analysis was used as a confirmatory method to resolve the discrepancy. This is a molecular gold standard.
  • Analytical Studies: Ground truth for analytical studies was established by known concentrations of well-characterized viral strains (TCID50/mL) for LoD, precision, and reproducibility. For analytical specificity and interference studies, known panels of organisms and substances were used.

8. Sample Size for the Training Set

Not applicable, as this is not an AI/ML device requiring a training set in that context.

9. How the Ground Truth for the Training Set was Established

Not applicable.

{0}------------------------------------------------

February 1, 2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Vela Diagnostics USA Inc. Donald Henton Dir. Regulatory Affairs North America 353C US Route 46 West Suite 250 Fairfield, New Jersey 07004

Re: K172509

Trade/Device Name: Sentosa SA201 HSV 1/2 Oualitative PCR Test Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes simplex virus serological assays Regulatory Class: Class II Product Code: 000 Dated: December 29, 2017 Received: January 4, 2018

Dear Donald Henton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Uwe Scherf -S

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172509

Device Name

Sentosa® SA201 HSV-1/2 PCR Test

Indications for Use (Describe)

Intended Use

The Sentosa SA201 HSV-1/2 PCR Test is a real-time PCR-based qualitative in vitro diagnostic test for detection and differentiation of Herpes Simplex Virus (HSV-1 and HSV-2) DNA from male and female skin lesions from anogenital or oral sites. The test is intended for use as an aid in diagnosis of herpes infection in symptomatic patients.

Warning: The Sentosa® SA201 HSV-1/2 PCR Test is not FDA cleared for use with cerebrospinal fluid (CSF). The test is not intended to be used for prenatal screening.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Vela Diagnostics. The logo features a stylized human figure with arms raised, connected to a DNA helix in blue and green. To the right of the figure, the word "VELA" is written in large, bold, blue letters. Below "VELA", the word "DIAGNOSTICS" is written in smaller, green letters.

510(k) Summary

Date summary prepared: 1/31/2018

510(k) Submitter/Holder

Vela Diagnostics USA, Inc. 353C US Route 46 West Suite 250 Fairfield, NJ 07004

Contact

Donald Henton Director Regulatory Affairs North America Telephone: 973-369-3578 Fax: 973-521-7077 Email: donald.henton@veladx.com

Name of Device

Trade Name:Sentosa® SA201 HSV-1/2 PCR Test
Catalog Numbers:300216
Common Name:Herpes Simplex Virus Nucleic Acid Amplification Assay
Classification Name:Herpes simplex virus serological assays (21 CFR § 866.3305, Class II, OQO).

Predicate Device

The Sentosa® SA201 HSV-1/2 PCR Test was compared to and found to be substantially equivalent to the following product of comparable type in commercial distribution:

Trade Name:IMDx HSV-1/2 for Abbott m2000 Assay
Common Name:Herpes Simplex Virus Nucleic Acid Amplification Assay
510(k) Number:K140198 (cleared 05/13/2014)
Manufacturer:Intelligent Medical Devices, Inc.

{4}------------------------------------------------

Table 1. Sentosa® SA201 HSV-1/2 PCR Test and Predicate Similarities Comparison
ItemDevicePredicateSimilarities /Differences
CharacteristicsSentosa® SA201 HSV1/2PCR Test (K172509)IMDx HSV-1/2 forAbbott m2000 Assay(K140198)
Regulation21 CFR 866.330521 CFR 866.3305Same
Product CodeOQOOQOSame
Device ClassClass IIClass IISame
Intended useThe Sentosa® SA201HSV-1/2 PCR Test is areal-time PCR-basedqualitative in vitrodiagnostic test fordetection anddifferentiation of HerpesSimplex Virus (HSV-1and HSV-2) DNA frommale and female skinlesions from anogenitalor oral sites. The test isintended for use as an aidin diagnosis of herpesinfection in symptomaticpatients.Warning: The Sentosa®SA201 HSV-1/2 PCRTest is not FDA clearedfor use withcerebrospinal fluid(CSF). The test is notintended to be used forprenatal screening.The IMDx HSV-1/2for Abbott m2000assay is an in vitrodiagnostic test for thedirect, qualitativedetection anddifferentiation ofHerpes Simplex Virustype 1 (HSV-1) andtype 2 (HSV-2) DNAfrom male and femaleskin lesions fromanogenital or oralsites. The test isintended for use as anaid in the diagnosis ofHSV infection insymptomatic patients.The assay is intendedto be run on the Abbottm2000 instrumentsystem.Warning: The IMDxHSV-1/2 for Abbottm2000 assay is notFDA-cleared for usewith cerebrospinalfluid (CSF). The assayis not intended for pre-natal screening.The intended useis the sameexcept the notedsystems they aredesigned to berun on.The warning isthe same.
Test PrincipleReal-time PCR DNAamplificationReal-time PCR DNAamplificationSame
Assay ResultsQualitative detectionand differentiation ofHSV-1 and HSV-2DNAQualitative detectionand differentiation ofHSV-1 and HSV-2DNASame
Sample typeMale and female skinlesions from anogenitalor oral sitesMale and female skinlesions fromanogenital or oral sitesSame

{5}------------------------------------------------

Device Description

The Sentosa® SA201 HSV-1/2 PCR Test is a (4x24) configuration contains reagents and enzymes for specific amplification of a 104 bp (base-pair) fragment of the UL30 gene common to both HSV1 and HSV2, and specific probes for the direct detection and differentiation of HSV1 and HSV2 amplicons, respectively. Pathogen detection by PCR is based on the amplification of specific regions of the pathogen genome. In real-time PCR, the amplified product is detected via fluorescent dyes, which are usually linked to oligonucleotide probes that bind specifically to the target sequences. Real-time monitoring of the fluorescence intensities during a PCR run allows the detection of the accumulating product. Amplification of the targets occurs in three channels: green, orange and red on the Sentosa® SA201. Output is recorded as the increase of fluorescence over time in comparison to background signal. Monitoring the fluorescence intensities during the PCR run allows the detection of the accumulating product without having to re-open the reaction tubes after the PCR run.

The Sentosa® SA201 HSV-1/2 PCR Test workflow starts with extraction of nucleic acids from samples (anogenital or oral swabs) using the Sentosa® SX Virus Total Nucleic Acid Kit on the Sentosa SX101 instrument. Following extraction, the instrument will automatically set up the PCR with the extracted nucleic acids in a 96-well PCR plate. Subsequently, the 96-well PCR plate is sealed and transferred to the Sentosa® SA201 for PCR amplification, followed by data analysis.

The Sentosa® Link facilitates data transfer between the Sentosa® SX101, the Sentosa® SA201 Reporter and existing LIS/LIMS (laboratory information systems) in the clinical lab. The Sentosa SX101 instrument communicates with Sentosa® SA201 thermocycler. This creates a user environment that links the SX101 and the Sentosa® SA201 to facilitate automated workflow to export results in a LIS/LIMS-compatible format.

Intended Use

The Sentosa® SA201 HSV-1/2 PCR Test is a real-time PCR-based qualitative in vitro diagnostic test for detection and differentiation of Herpes Simplex Virus (HSV-2) DNA from male and female skin lesions from anogenital or oral sites. The test is intended for use as an aid in diagnosis of herpes infection in symptomatic patients.

Warning: The Sentosa® SA201 HSV-1/2 PCR Test is not FDA cleared for use with cerebrospinal fluid (CSF). The test is not intended to be used for prenatal screening.

{6}------------------------------------------------

Technological Characteristics

The Sentosa® SA201 HSV-1/2 PCR Test contains reagents and enzymes for specific amplification of a 104 bp fragment of UL30 gene common to both HSV1 and HSV2, and specific probes for the direct detection and differentiation of HSV1 and HSV2 amplicons, respectively. Pathogen detection by PCR is based on the amplification of specific regions of the pathogen genome. In real-time PCR, the amplified product is detected via fluorescent dyes, which are usually linked to oligonucleotide probes that bind specifically to the target sequences. Real-time monitoring of the fluorescence intensities during a PCR run allows the detection of the accumulating product. Amplification of the targets occurs in three channels: green, orange and red on the Sentosa" SA201. Output is recorded as the increase of fluorescence over time in comparison to background signal. Monitoring the fluorescence intensities during the PCR run allows the accumulating product without having to re-open the reaction tubes after the PCR run.

The Sentosa SA201 HSV-1/2 PCR Test workflow starts with extraction of nucleic acids from samples (anogenital or oral swabs) using the Sentosa® SX Virus Total Nucleic Acid Kit on the Sentosa® SX101 instrument. Following extraction, the instrument will automatically set up the PCR with the extracted nucleic acids in a 96-well PCR plate. Subsequently, the 96-well PCR plate is sealed and transferred to the Sentosa" SA201 for PCR amplification, followed by data analysis.

Drivers are installed on the Sentosa Link to connect the Sentosa SX101 instrument and the Sentosa SA201 thermocycler. This creates a user environment that links the SX101 and the Sentosa SA201 to facilitate automated workflow to export results in a LIS/LIMS-compatible format.

The Sentosa® SA201 HSV-1/2 PCR Test uses the Sentosa® SX101 hardware and Sentosa® SA201 hardware, along with the associated consumables to operate as an automated sample extraction, PCR setup, amplification and reporting system (referred to as workflow for short).

{7}------------------------------------------------

Image /page/7/Figure/0 description: This image describes the sample workflow for the Sentosa SA201 HSV1/2 PCR Test. The workflow begins with an oral or anogenital swab, followed by off-board preparation. The reagents used are the Sentosa SX Virus Total Nucleic Acid Kit (4x24) and the Sentosa SA201 HSV1/2 PCR Test (4x24). The instruments used are the Sentosa SX101 and the Sentosa SA201, and the software used is the Sentosa SX101 software, Sentosa Link, and the Sentosa SA201 Reporter.

Figure 1. Sentosa® SA201 HSV-1/2 PCR Test Workflow Overview.

{8}------------------------------------------------

Image /page/8/Figure/0 description: This image is a flowchart describing the steps of a laboratory procedure. The procedure starts with sample preparation, followed by scanning samples using the Sentosa Link software. The Sentosa SX101 software is then launched, and consumables, reagents, controls, and samples are loaded onto the Sentosa SX101. The procedure ends with the Sentosa SA201 Reporter software interpreting the results.

Sentosa® Link reads and displays the results

Figure 2. Sentosa® SA201 HSV-1/2 PCR Test Workflow Flowchart.

{9}------------------------------------------------

Performance

Analytical Studies:

  1. Analytical sensitivity: The limit of detection (LoD) was assessed for the Sentosa® SA201 HSV-1/2 PCR Test using two strains of HSV-1 (MacIntyre and KOS) and two strains of HSV-2 (MS and G). The LoD is defined as the HSV titer (TCID50/mL) detected with a probability of 95% or greater and was determined by Probit analysis using HSV-1 MacIntyre and HSV-2 MS. The results of the analytical sensitivity of the Sentosa SA201 HSV-1/2 PCR Test are summarized in the following table.
StrainLoD(TCID50/mL)
HSV-1 MacIntyre40
HSV-1 KOS40
HSV-2 MS4
HSV-2 G4
Table 2. Analytical sensitivity - Limit of Detection (LoD) HSV-1 and HSV-2
--------------------------------------------------------------------------------
  1. Precision: The study was conducted over a period of five (5) days with three (3) reagent lots. five (5) operators, and four (4) instrument systems at Vela Research Singapore. The test materials used in the study included NC, PC, and negative sample, 3xLoD HSV1, 1.5xLoD HSV1, 3xLoD HSV2 and 1.5xLoD HSV2. Each concentration was assayed three (3) times per run for 20 runs, resulting in 60 replicates (three (3) replicates/run x four (4) run/day x five (5) days x one (1) site = 60 replicates). The precision analysis was based on %CV of Ct values, and is presented in Table 3 below. The quality control analysis was based on the mean and standard deviations (SD) of the Ct values. The demonstrated agreement with the results being 100% for all tested types, and the %CV being (less than) < 10%.
Sample typeChannelAgreement(%)95%CI*Mean CtSD
1.5x LoD HSV-1aGreen60/60(100%)93.98% -100%31.790.81
3x LoD HSV-1aGreen60/60(100%)93.98% -100%30.720.90
1.5x LoD HSV-2bOrange60/60(100%)93.98% -100%30.251.09
3x LoD HSV-2bOrange59/59(100%)*93.89% -100%29.350.54
NCRed60/60(100%)93.98% -100%29.082.24
PCGreen60/60(100%)93.98% -100%26.091.10
Table 3. Precision

{10}------------------------------------------------

Sample typeChannelAgreement(%)95%CI*Mean CtSD
Orange60/60(100%)93.98% -100%28.510.69
Negative sampleRed60/60(100%)93.98% -100%28.982.03

41.5x LoD HSV-1 is 60 TCID50/mL and 3x LoD HSV-1 is 120 TCID50/mL

91.5x LoD HSV-2 is 6 TCID50/mL and 3x LoD HSV-2 is 12 TCID50/mL

  • Total number of samples is less than 60 due to exclusion of 1 invalid sample
  1. Reproducibility: The Reproducibility Study was to demonstrate the performance of the Sentosa SA201 HSV-1/2 PCR Test using simulated samples (with spiked-in HSV1/2 virus) using multiple sites, multiple instruments, multiple operators, and multiple lots using the matrix. The study demonstrated excellent agreement across the sites, operators, and systems with the results meeting the acceptance for all the tested types, and the %CV being (less than) < 10%. The results are summarized in the following Table 4.
Sample typeChannelAgreement(%)95% CI*Mean Ct ± SD% CV
1.5x LoD HSV-1aGreen90/90 (100%)95.91% - 100%31.45 ±1.655.25%
3x LoD HSV-1aGreen90/90 (100%)95.91% - 100%30.23 ±1.013.34%
1.5x LoD HSV-2bOrange90/90 (100%)95.91% - 100%29.29 ±0.471.60%
3x LoD HSV-2bOrange90/90 (100%)95.91% - 100%28.01 ±1.184.21%
NCRed90/90 (100%)95.91% - 100%26.93 ±0.602.23%
PCGreen90/90 (100%)95.91% - 100%25.80 ±0.261.01%
PCOrange90/90 (100%)95.91% - 100%28.32 ±0.321.13%
Negative sampleRed90/90 (100%)95.91% - 100%27.17 ±1.144.20%
BlankRed90/90 (100%)95.91% - 100%26.81 ±0.552.05%

Table 4. Reproducibility

  1. Analytical Reactivity / Cross-reactivity: The Analytical Reactivity and Specificity Study was to demonstrate the analytical reactivity and cross-reactivity (analytical specificity) performance of the Sentosa SA201 HSV-1/2 PCR Test with the Sentosa® workflow using high and low concentration HSV1 and HSV2 samples (with competing organisms). The results in the study showed that the 55 organisms, which are closely related to HSV1/2 or present in oral or genital swab samples at high concentrations showed no cross reactivity to HSV1/2 detection using the Sentosa® SA201 HSV-1/2 PCR Test. It also showed no interference with the detection of low concentration levels of HSV1 or HSV2. The results do suggest that C. glabrata, S. aureus, S. epidermidis, P. melaninogenica and HHV6 might lead to slight competitive inhibition from HSV1 to HSV2. The results also showed no cross-reactivity within the multiplex panel (HSV1 and HSV2) in the presence of either high concentration HSV1 or high concentration HSV2 strains.

{11}------------------------------------------------

Potential interfering organisms
Candida glabrata, NCYC 388Human papillomavirus type 16
Acinetobacter baumannii, 2208Human papillomavirus type 18
Acinetobacter calcoaceticusHuman DNA
Acinetobacter IwofiiHuman herpesvirus (HHV6)
Actinomyces isralii, serotype 1Human herpesvirus 4, B95-8
Adenovirus 1, strain Adenoid 71Klebsiella pneumonia, NCTC 9633
Adenovirus Type 7, strain GomenLactobacillus acidophilus
Bacteroides fragilis, VPI2553Maraxella catarrhalis, strain 20
Candida albicans, strain 132Mobiluncus curtisii, BV 345-16
Candida krusei, NRRL Y-6Mobiluncus muleris, BV 64-5
Candida parapsilosis, NRRL-Y-12969Mycoplasma hominis, PG21
Candida tropicalis, PK233Neisseria gonorrhea, B-585
Chlamydia trachomatis, UW-57/CxNeisseria meningitides, serogroup B
Clostridium difficile, strain 4118Prevotella melaninogenica, B282
CMV, AD-169Rubella virus
Corynebacterium genitalium, 392-1Staphylococcus aureus, F-182
Cryptococcus neoformans, 52Staphylococcus aureus, FDA 209
Enterobacter cloacae, QC strainStaphylococcus epidermidis, 255-01B
Enterococcus faecalis, AGR329Staphylococcus saprophyticus, LRA27.02
Enterovirus type 71, BrCrStreptococcus mitis, NCTC 12261
Epstein-Barr virus/Human herpesvirus 4,strain P-3Streptococcus mutans, UA159
Escherichia coli O103, strain NCDCH515bStreptococcus pneumonia, CIP104225
Fusobacterium necrophorum subspnecrophorum, VPI2553Streptococcus pyogenes
Fusobacterium nucleatum subspnucleatum, 1612ASV40 (Simian virus 40)
Gardnerella/Haemophilus vaginalis, 317Toxoplasma gondii
Haemophilus ducreyi, CIP 542Trichomonas vaginalis
Hepatitis A virus, strain PA21Varicella-Zoster Virus (VZV)
HIV-1, Group M Subtype C

Table 5. The Analytical Reactivity / Cross-reactivity tested organisms

  1. Interfering Substances: The Interfering Substances and Competitive Interference Study to evaluate the effect of potentially interfering substances and competitive interference on the performance of the Sentosa® SA201 HSV1/2 PCR Test was conducted over a period of ten (10) days with one (1) reagent lot, five (5) operators, and four (4) instrument workflow systems. The test materials used in this study included 31 substances commonly found in genital and oral specimens, which were dissolved into virus transport media (VTM) at approximately 3xLoD for both HSV1 and HSV2 to approximately 10 times of the normal active concentration. Each

{12}------------------------------------------------

material was assayed in triplicate. The highest concentration for each substance with no interfering effect on device performance was reported. The test materials used in the competitive interference study were HSV1 and HSV2 spiked-in samples with different combinations of HSV1 and HSV2 concentrations. Each combination was tested in 20 replicates. The virus with the highest concentration that resulted in 100% detection of another virus was recorded, as well as the highest equal concentration of the two viruses. No interfering effects on the performance at concentration levels higher (5 to 10 times) than the normal active concentration of the interference substances tested. The HSV1 concentration that resulted in 100% detection rate of 3xLoD HSV2 was at least 50xLoD. The HSV2 concentration that resulted in 100% detection rate of 3xLoD HSV1 was at least 10xLoD but less than 50xLoD. No mutual interference was observed in HSV1 and HSV2 equal concentration combinations to at least 500xLoD of each virus.

SubstanceActive IngredientsTested concentration
AcyclovirAcycloguanosine (10%)7mg/mL
Whole blood with EDTAN/A7% (v/v)
Female urineN/A7% (v/v)
Male urineN/A7% (v/v)
AlbuminAlbumin3.3 mg/mL
SalivaN/A7% (v/v)
K.Y. Jelly lubricantN/A7% (w/v)
Feminine washN/A5% (v/v)
Xylocaine 5%Lidocaine (50mg)7% (w/v)
ToothpasteStannous fluoride (0.454%)0.53% (w/v)
Desitin maximum originalpasteZinc Oxide (40%)7% (w/v)
Mentholatum lip balmMenthol (0.7%), Camphor(1.7%)7% (w/v)
Listerine anti-bacterialmouthwashEucalyptol (0.092%), Menthol(0.042%), Methy Salicylate(0.060%), Thymol (0.064%)7% (v/v)
CaseinCasein7 mg/mL
DoucheProvidone-iodine (10% w/v)7% (v/v)
YeastGardCandida albicans 27X* HPUS.Candida parapsilosis 27X*HPUSPulsatilla 27* HPUS7% (w/v)
Vaginal contraceptive gelNonoxynol-9 (4%)7% (w/v)
Vaginal contraceptive gelNonoxynol-9 (3%)7% (w/v)
Monistat-7Miconazole nitrate vaginalcream (2%)7% (w/v)
FleetBenzalkonium (0.06%w/w)/EDTA7% (w/v)
Gyno-TrosydTioconazole (100mg)7% (w/v)
Clotrimazole 1% CreamClotrimazole (50mg, 1%)7% (w/v)
Anti-Inch CreamLidocaine ph. Eur. (2% w/w)7% (w/v)

Table 6. Interfering Substances and Competitive Interference Study tested substances

{13}------------------------------------------------

SubstanceActive IngredientsTested concentration
AbrevaDocosanol (10%)7% (w/v)
Buffy coatWhite blood cell7% (w/v)
Talcum powderN/A7% (w/v)
Seminal fluidSeminal fluid7% (v/v)
FecesFeces7% (w/v)
Corn starchCorn starch1.25 mg/mL
ParacetamolAcetamidophenol5 mg/mL
AspirinAcetylsalicyclic acid10 mg/mL
  1. Carry-over and Cross-contamination: The Cross-contamination and Carry-over studies were to evaluate the potential cross-contamination during extraction and on subsequent runs on the performance of the Sentosa® SA201 HSV1/2 Test with the Sentosa® SX101 workflow instrument in accordance with CLSI MM17-A (2008). These carry-over and crosscontamination studies were conducted over a period of four (4) days with one (1) reagent lot, two (2) operators, and two (2) Sentosa workflow systems. The test materials used in the study included negative samples and 1x10^5 TCIDs(/2500xLoD) for the crosscontamination study, a NC, a PC, and negative samples for the carry-over contamination study. Three (3) run matrices were used, two (2) for the cross-contamination study and one (1) for the carry-over contamination study. The results showed that the overall contamination rate of the twelve runs was 0% at a HSV1 positive sample concentration of 1x10^5 TCID«/mL. All 96 positive samples were detected as positive, and the 183 negative samples were detected as negative (no amplification signal noted).

{14}------------------------------------------------

Clinical Studies

  1. Clinical Study: This study utilized residual anogenital lesion or oral lesion samples from male and female patients with signs and symptoms of HSV infections collected from 8 sites in the USA. Samples were either tested at the same facility at which thev were obtained, or were shipped to a different testing site. Each sample was tested with Sentosa® SA201 HSV-1/2 PCR Test (4 test sites) and the ELVIS® HSV ID and D3 Typing Test System reference test (3 test sites). A total of 2684 samples were enrolled in the study. 389 samples were excluded, leaving a total of 2295 samples in the final analysis. Out of these, 317 were oral lesions and 1978 were anogenital lesions.

The clinical performance of Sentosa® SA201 HSV-1/2 PCR Test on oral samples compared to the ELVIS® HSV ID and D3 Typing Test System (ELVIS) for sensitivity and specificity is summarized in the following tables. Discordant samples were tested using bi-directional sequencing analysis.

SentosaSA201 HSV-1/2 PCRTestELVIS Reference Method
PositiveNegativeTotal
Positive28154b335
Negative9a12371246
Total29012911581
ValueLower95% CIUpper95% CI
Sensitivity96.90%94.21%98.36%
Specificity95.82%94.58%96.78%

Table 6. HSV-1 results for anogenital lesions

a From sequencing analysis, 4 discordant samples (HSV-1 positive by ELVIS and HSV-1 negative by Sentosa®) were in agreement with Sentosa® results, 2 were in agreement with ELVIS results and 2 were not in agreement with Sentosa® nor ELVIS results. 1 discordant sample was not tested due to insufficient volume.

b From sequencing analysis, 44 discordant samples (HSV-1 negative by ELVIS and HSV-1 positive by Sentosa®) were in agreement with Sentosa® results and 10 were in agreement with ELVIS results.

{15}------------------------------------------------

SentosaSA201 HSV-1/2 PCRTestELVIS Reference Method
PositiveNegativeTotal
Positive391147d538
Negative6c14341440
Total39715811978
ValueLower95% CIUpper95% CI
Sensitivity98.49%96.74%99.31%
Specificity90.70%89.17%92.04%

Table 7. HSV-2 results for anogenital lesions

6 From sequencing analysis, 4 discordant samples (HSV-2 positive by ELVIS and HSV-2 negative by Sentosa ) were in agreement with Sentosa results and 2 were in agreement with ELVIS results.

d Out of 147 discrepant samples, 142 were unique samples (3 samples were ELVIS HSV-1 positive/HSV-2 negative and Sentosa® HSV-1 negative/HSV-2 positive and were doublecounted in the 9 samples that are ELVIS HSV-1 positive and Sentosa® HSV-1 negative. 2 samples were ELVIS HSV-1 negative and Sentosa® HSV-1 positive/HSV-1 positive/HSV-2 positive. These were double-counted in the 54 samples that are ELVIS HSV-1 negative and Sentosa HSV-1 positive.) From sequencing analysis, 119 discordant samples (HSV-2 negative by ELVIS and HSV-2 positive by Sentosa®) were in agreement with Sentosa results and 20 were in agreement with ELVIS results. Three discordant samples was not tested due to insufficient volume.

HSV-1 results for anogenital lesions (N=1581, 397 samples tested as HSV-2 positive by EL VIS were excluded from the 1978 anogenital lesion samples).

SentosaELVIS Reference Method
SA201
HSV-1/2PositiveNegativeTotal
PCR Test
Positive7932e111
Negative0203203
Total79235314
ValueLowerUpper
95% CI95% CI
Sensitivity100.00%95.36%100.00%
Specificity86.38%c81.41%90.19%

Table 8. HSV-1 results for oral lesions

6 From sequencing analysis, 27 discordant samples (HSV-1 negative by ELVIS and HSV-1 positive by Sentosa ) were in agreement with Sentosa results and 4 were in agreement with ELVIS results. One discordant sample was not tested due to insufficient volume.

{16}------------------------------------------------

SentosaSA201 HSV-1/2 PCRTestELVIS Reference Method
PositiveNegativeTotal
Positive21g3
Negative1f313314
Total3314317
ValueLower95% CIUpper95% CI
Sensitivity66.67%20.77%93.85%
Specificity99.68%98.22%99.94%

Table 9. HSV-2 results for oral lesions

4 From sequencing analysis, one sample (HSV-2 positive/HSV-1 negative by ELVIS and HSV-2 negative/HSV-1 positive by Sentosa®) was in agreement with Sentosa® results.

8 One discordant sample (HSV-1/2 negative by ELVIS and HSV-2 positive by Sentosa®) was not tested due to insufficient volume.

HSV-1 results for oral lesions (N=314, 3 samples tested as HSV-2 positive by ELVIS were excluded from the 317 oral lesion samples)

{17}------------------------------------------------

AgeHSV-1HSV-2
(years)FemaleMaleCombinedFemaleMaleCombined
0 - 102/271/373/640/270/370/64
7.4%2.7%4.7%0.0%0.0%0.0%
11 - 2059/1888/4067/22851/2326/4357/275
31.4%20.0%29.4%22.0%14.0%20.7%
21 - 30126/40019/103145/503173/53855/146228/684
31.5%18.4%28.8%32.2%37.7%33.3%
31 - 4050/2547/7057/32471/30014/7885/378
19.7%10.0%17.6%23.7%17.9%22.5%
41 - 5027/1854/2831/21358/22111/3669/257
14.6%14.3%14.6%26.2%30.6%26.8%
51 - 6022/1051/2723/13241/13912/3153/170
21.0%3.7%17.4%29.5%38.7%31.2%
61 - 705/612/127/7321/774/1425/91
8.2%16.7%9.6%27.3%28.6%27.5%
71 - 802/220/92/3112/303/1115/41
9.1%0.0%6.5%40.0%27.3%36.6%
81 - 900/60/40/103/92/65/15
0.0%0.0%0.0%33.3%33.3%33.3%
>900/00/30/30/01/31/3
0.0%0.0%0.0%0.0%33.3%33.3%
TOTAL293/124842/333335/1581430/1573108/405538/1978
23.5%12.6%21.2%27.3%26.7%27.2%

Table 10. Distribution of samples according to demographics for anogenital lesions as tested by Sentosa® SA201 HSV-1/2 PCR Test

{18}------------------------------------------------

Age(years)HSV-1HSV-2
FemaleMaleCombinedFemaleMaleCombined
0 - 1011/2510/2821/530/250/290/54
44.0%35.7%39.6%0.0%0.0%0.0%
11 - 207/135/1812/310/130/180/31
53.8%27.8%38.7%0.0%0.0%0.0%
21 - 308/406/2614/660/400/260/66
20.0%23.1%21.2%0.0%0.0%0.0%
31 - 405/235/1610/391/240/161/40
21.7%31.3%25.6%4.2%0.0%2.5%
41 - 509/192/611/250/190/60/25
47.4%33.3%44.0%0.0%0.0%0.0%
51 - 607/255/912/340/250/90/34
28.0%55.6%35.3%0.0%0.0%0.0%
61 - 704/187/1411/321/180/141/32
22.2%50.0%34.4%5.6%0.0%3.1%
71 - 809/137/1416/270/131/151/28
69.2%50.0%59.3%0.0%6.7%3.6%
81 - 901/32/43/70/30/40/7
33.3%50.0%42.9%0.0%0.0%0.0%
TOTAL61/17949/135110/3142/1801/1373/317
34.1%36.3%35.0%1.1%0.7%0.9%

Table 11. Distribution of samples according to demographics for oral lesions as tested by Sentosa® SA201 HSV-1/2 PCR Test

Positive and Negative Predictive Value: Hypothetical positive and negative predictive values (PPV & NPV) for the Sentosa® SA201 HSV-1/2 PCR Test are shown below. These calculations for hypothetical prevalence are based on overall sensitivity and specificity per sample type from the clinical study results. For HSV-1, these calculations are based upon an overall sensitivity and specificity of 96.90% and 95.82%, respectively, for anogenital swabs and 100.0% and 86.38%, respectively, for oral swabs. For HSV-2, these calculations are based upon an overall sensitivity and specificity of 98.49% and 90.70%, respectively, for anogenital swabs and 66.67% and 99.68%, respectively, for oral swabs.

AnogenitalOral
Prevalence(%)HSV-1HSV-2HSV-1HSV-2
PPVNPVPPVNPVPPVNPVPPVNPV
232.12%99.93%17.77%99.97%13.03%100.00%80.96%99.32%
554.96%99.83%35.79%99.91%27.87%100.00%91.64%98.27%
1072.03%99.64%54.06%99.82%44.93%100.00%95.86%96.42%
2085.28%99.20%72.58%99.59%64.73%100.00%98.12%92.29%
3090.86%98.63%81.95%99.29%75.88%100.00%98.89%87.47%
4093.92%97.89%87.59%98.90%83.04%100.00%99.29%81.77%
5095.86%96.87%91.37%98.36%88.01%100.00%99.52%74.94%

Table 12. Prevalence vs hypothetical Predictive Values

{19}------------------------------------------------

  1. HSV-2 oral lesion contrived specimen study: A contrived specimen study was performed to provide additional performance data for detection of HSV-2 in oral samples. A study was performed to test 30 HSV-2-contrived oral lesion samples, along with 15 HSV-1 positive oral lesion samples and 15 HSV-1/HSV-2 negative samples. In the test, all 30 HSV-2-contrived oral lesion samples were identified as HSV-2 positive, and all 15 HSV-1 positive and 15 HSV-1/2 negative samples were correctly identified.

Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3305 Herpes simplex virus serological assays.

(a)
Identification. Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.(b)
Classification. Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).