K Number
K172402
Date Cleared
2017-10-30

(82 days)

Product Code
Regulation Number
866.2680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARIES® Group A Strep Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of Streptococus pyogenes (Group A beta-hemolytic Streptococcus) in throat swab specimens from patients with signs and symptoms of pharyngitis.

The ARIES® Group A Strep Assay can be used as an aid in the diagnosis of Group A Streptococcal pharyngitis. The assay is not intended to monitor treatment for Group A Streptococcus infections.

The ARIES® Group A Strep Assay is indicated for use with ARIES® Systems.

Device Description

The ARIES® Group A Strep Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system that consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, an assay-specific cassette, and an assay-specific protocol file. The ARIES® Group A Strep Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® Group A Strep Assay cassette directly detects Streptococcus pyoqenes (Group A ß-hemolytic Streptococcus) in throat swab specimens collected from the surface of human tonsils and posterior pharyngeal wall.

Throat swab specimens are collected from patients using a commercially available Liquid Amies based transport system (Nylon Flocked Swab with 1 mL modified Liquid Amies (ESwab™). The specimen is then transported to the laboratory for testing.

The specimen is lysed and nucleic acid is extracted using an ARIES® System. An extractable sample processing control (SPC) target is present in the ARIES® Group A Strep Assay cassette and is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, the presence of inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the target Tm.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® Group A Strep Assay lyophilized PCR reagents in the PCR tube that contains primer pairs specific to the S. pyogenes DNaseB (sdaB) gene and the SPC sequence. Each of the primer pairs is labeled with a distinct fluorophore and detected in distinct channels of the ARIES® Systems. PCR amplification is performed and assay fluorescence is monitored. Incorporation of a quencher-labeled nucleotide results in a decrease in fluorescence for the associated primer pair. Following amplification, the reaction is slowly heated to separate the fluorescent-labeled strand from the quencher-labeled strand, a process that results in an increase in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the T ,, of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® Group A Strep Assay protocol and run files. ARIES® Group A Strep Assay results may be reported from the ARIES® Software or from the optional SYNCT® Software.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the ARIES® Group A Strep Assay, based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (ARIES Group A Strep Assay)
Clinical SensitivityHigh, to ensure detection of true positives97.5% (156/160), 95% CI: 93.7% - 99.0%
Clinical SpecificityHigh, to minimize false positives97.8% (448/458), 95% CI: 96.0% - 98.8%
Positive Predictive Value (PPV)High, indicating reliability of positive results94.0% (156/166), 95% CI: 89.3% - 96.7%
Negative Predictive Value (NPV)High, indicating reliability of negative results99.1% (448/452), 95% CI: 97.7% - 99.7%
Reproducibility (Overall Moderate Positive - 3X LoD)Consistent detection across sites98.9% (89/90)
Reproducibility (Overall Low Positive - 1X LoD)Consistent detection across sites for low concentrations96.7% (87/90)
Reproducibility (Overall Negative)Consistent negative results across sites1.1% (1/90) false positive, indicating high specificity
Lot-to-Lot Reproducibility (Overall Moderate Positive - 3X LoD)Consistent detection across lots100% (45/45)
Lot-to-Lot Reproducibility (Overall Low Positive - 1X LoD)Consistent detection across lots for low concentrations93.3% (42/45)*
Lot-to-Lot Reproducibility (Overall Negative)Consistent negative results across lots0.0% (0/45)
Within-Laboratory Precision/Repeatability (Moderate Positive - 3X LoD)Consistent detection within lab100% (30/30)
Within-Laboratory Precision/Repeatability (Low Positive - 1X LoD)Consistent detection within lab for low concentrations93.3% (28/30)*
Within-Laboratory Precision/Repeatability (Negative)Consistent negative results within lab0.0% (0/30)
Limit of Detection (LoD) - Bruno strainLowest concentration detected with ≥ 95% positivity2.58E+03 CFU/mL (95%)
Limit of Detection (LoD) - SF370 strainLowest concentration detected with ≥ 95% positivity4.13E+03 CFU/mL (100%)
Analytical Reactivity (Inclusivity)Detection of various S. pyogenes strains8/9 strains detected with 100% positivity at 3x LoD, 1 strain at 5x LoD
Interfering SubstancesNo inhibition/false results for tested substancesExpected results generally obtained, except for NyQuil (false negatives) and Mucin (invalid results)
Cross-Reactivity/Microbial InterferenceNo cross-reactivity with common throat microorganisms34/35 tested microorganisms showed no cross-reactivity or interference with GAS positivity (1 Treponema denticola interference at high concentrations)
Carry-Over/Cross-ContaminationNo carry-over between samples100% agreement with expected results for high positive and negative samples

Note: For Lot-to-Lot Reproducibility (Low Positive) and Within-Laboratory Precision/Repeatability (Low Positive), additional replicates were tested to achieve an overall positivity >94%, indicating acceptable performance at the LoD.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Performance (Test Set): A total of 618 unique specimens were available for analysis in the clinical performance study. Initially, 735 specimens were collected, but 112 were excluded due to various reasons (protocol non-compliance, insufficient volume, incorrect device, etc.).
  • Data Provenance: The data was prospectively collected from patients suspected of pharyngitis in four geographically distinct clinical sites within the United States.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the "number of experts" or their specific "qualifications" (e.g., "Radiologist with 10 years of experience") for establishing the ground truth of the clinical test set.

However, the ground truth for the clinical test set was established by a "reference method (bacterial culture followed by organism identification by Matrix-Assisted Laser Desorption/Ionization - Time-of-Flight Mass Spectrometry (MALDI-TOF MS))". This work was "performed at a centralized testing facility." It is implied that trained laboratory personnel and microbiologists conducted these reference tests, as is standard practice for such methods, but their specific experience levels are not detailed.

4. Adjudication Method for the Test Set

The document describes a robust "ground truth" establishment process using a reference laboratory technique (bacterial culture + MALDI-TOF MS). For specimens where the ARIES assay and the reference method disagreed, an adjudication method was employed:

  • Bidirectional sequencing analysis using analytically validated primers that targeted genomic regions distinct from the ARIES Group A Strep Assay was used.
    • For 2 false negative ARIES results (ARIES negative, culture positive), sequencing showed them to be GAS negative.
    • For 7 false positive ARIES results (ARIES positive, culture negative), sequencing showed them to be GAS positive.
      This indicates a method similar to "Truth by Consensus" with an expert-level tie-breaker (sequencing), effectively refining the ground truth against discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay that machine-interprets results (qualitative PCR). It does not involve human "readers" assessing images or data in a way that would be "assisted" by AI, hence the concept of a human reader improvement effect size does not apply.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The ARIES® Group A Strep Assay is an automated qualitative in vitro diagnostic test system. The clinical performance study directly assessed the agreement of the ARIES® Group A Strep Assay (algorithm only within the ARIES® Systems) against the established reference method, without human interpretation of the assay's raw output. The results (Positive, Negative, Invalid) are determined by the ARIES® System software.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

The ground truth used for the clinical performance study was primarily bacterial culture followed by organism identification by Matrix-Assisted Laser Desorption/Ionization - Time-of-Flight Mass Spectrometry (MALDI-TOF MS). In cases of discordance with the assay, bidirectional sequencing analysis was used as a confirmatory "higher truth" method. This can be considered a form of expert consensus/gold standard laboratory method.

8. The Sample Size for the Training Set

The document does not explicitly state a sample size for a training set. The ARIES® Group A Strep Assay is a PCR-based diagnostic test, where the "training" typically involves empirical optimization of assay parameters (primers, probes, reaction conditions, and cut-offs) during product development and internal verification studies, rather than machine learning on a distinct "training set" of patient data. The "Initial Assay Protocol File parameters were set during internal optimization studies" and "The final Assay Protocol File parameters were then established during internal verification studies." So, the optimization and verification process served a similar function to "training" for machine learning algorithms.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, a traditional "training set" with separate ground truth establishment isn't described in the context of a PCR assay. Instead, the "ground truth" for optimizing the assay protocol file parameters (Ct cut-off, Tm window, Tm Peak Threshold) was established through internal optimization and verification studies. These studies likely involved testing known positive and negative S. pyogenes samples (e.g., well-characterized strains, spiked samples) to define the optimal performance characteristics of the assay before clinical validation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 30, 2017

LUMINEX CORPORATION JENNIFER GRIMES INTERIM MANAGER, REGULATORY AFFAIRS 12212 TECHNOLOGY BLVD. AUSTIN TX 78727

Re: K172402

Trade/Device Name: ARIES Group A Strep Assay, ARIES Group A Strep Assay Protocol File Kit Regulation Number: 21 CFR 866.2680 Regulation Name: Streptococcus spp. nucleic acid-based assay Regulatory Class: II Product Code: PGX, OOI Dated: August 8, 2017 Received: August 9, 2017

Dear Ms. Grimes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Steven R. Gitterman -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172402

Device Name ARIES Group A Strep Assay

Indications for Use (Describe)

The ARIES® Group A Strep Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of Streptococus pyogenes (Group A beta-hemolytic Streptococcus) in throat swab specimens from patients with signs and symptoms of pharyngitis.

The ARIES® Group A Strep Assay can be used as an aid in the diagnosis of Group A Streptococcal pharyngitis. The assay is not intended to monitor treatment for Group A Streptococcus infections.

The ARIES® Group A Strep Assay is indicated for use with ARIES® Systems.

Type of Use (Select one or both, as applicable)X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

11.0 510(k) Summary

This Executive Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

A. 510(k) Number:

K172402

B. Purpose for Submission:

Traditional 510(k), New Device

C. Measurand:

Streptococcus pyogenes (Group A β-hemolytic Streptococcus)

D. Type of Test:

Qualitative Real Time Polymerase Chain Reaction (PCR)

E. Applicant:

Jennifer Grimes Luminex Corporation 12212 Technology Blvd Austin, TX 78727

Tel: (416) 593-4323

Date Prepared F.

October, 2017

G. Proprietary and Established Names:

ARIES® Group A Strep Assay

H. Regulatory Information:

ProductCodeClassificationRegulation SectionPanel
PGXClass II21 CFR 866.2680 - Streptococcus spp. nucleic acid based assayMicrobiology(83)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, filled circle above the "i" in "Luminex". A small registration mark is present to the right of the "x".

Intended Use: 1.

1. Intended use(s):

The ARIES® Group A Strep Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of Streptoccus pyogenes (Group A ß-hemolytic Streptococcus) in throat swab specimens from patients with signs and symptoms of pharyngitis.

The ARIES® Group A Strep Assay can be used as an aid in the diagnosis of Group A Streptococcal pharyngitis. The assay is not intended to monitor treatment for Group A Streptococcus infections.

The ARIES® Group A Strep Assay is indicated for use with ARIES® Systems.

    1. Indication(s) for use:
      Same as intended use.
    1. Special conditions for use statement(s):
      For prescription use only.
    1. Special instrument requirements:
      For use with the ARIES® Systems.

Device Description: J.

The ARIES® Group A Strep Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system that consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, an assay-specific cassette, and an assay-specific protocol file. The ARIES® Group A Strep Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® Group A Strep Assay cassette directly detects Streptococcus pyoqenes (Group A ß-hemolytic Streptococcus) in throat swab specimens collected from the surface of human tonsils and posterior pharyngeal wall.

Throat swab specimens are collected from patients using a commercially available Liquid Amies based transport system (Nylon Flocked Swab with 1 mL modified Liquid Amies (ESwab™). The specimen is then transported to the laboratory for testing.

{5}------------------------------------------------

The specimen is lysed and nucleic acid is extracted using an ARIES® System. An extractable sample processing control (SPC) target is present in the ARIES® Group A Strep Assay cassette and is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, the presence of inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the target Tm.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® Group A Strep Assay lyophilized PCR reagents in the PCR tube that contains primer pairs specific to the S. pyogenes DNaseB (sdaB) gene and the SPC sequence. Each of the primer pairs is labeled with a distinct fluorophore and detected in distinct channels of the ARIES® Systems. PCR amplification is performed and assay fluorescence is monitored. Incorporation of a quencher-labeled nucleotide results in a decrease in fluorescence for the associated primer pair. Following amplification, the reaction is slowly heated to separate the fluorescent-labeled strand from the quencher-labeled strand, a process that results in an increase in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the T ,, of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® Group A Strep Assay protocol and run files. ARIES® Group A Strep Assay results may be reported from the ARIES® Software or from the optional SYNCT® Software.

K. Substantial Equivalence Information:

    1. Predicate device name(s):
      Liat™ Strep A Assay (Manufactured by IQuum, Inc.)
    1. Predicate 510(k) number(s):
      K141338
    1. Comparison with predicate:
      The following tables compares the ARIES® Group A Strep Assay to IQuum, Inc.'s Liat™ Group A Strep Assay (K141338). Table 11-1 shows similarities between the new device and the predicate, while Table 11-2 shows the differences.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Luminex. The logo is in black and white, with the word "Luminex" in a bold, sans-serif font. There is a small, gray circle above the "i" in Luminex. The logo is simple and modern.

Similarities
AttributeNew DevicePredicate Device (K141338)
Intended UseThe ARIES® Group A Strep Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of Streptococcus pyogenes (Group A β-hemolytic Streptococcus) in throat swab specimens from patients with signs and symptoms of pharyngitis.The ARIES® Group A Strep Assay can be used as an aid in the diagnosis of Group A Streptococcal pharyngitis. The assay is not intended to monitor treatment for Group A Streptococcus infections.The ARIES® Group A Strep Assay is indicated for use with ARIES® Systems.The Liat™ Strep A Assay, performed on the Liat™ Analyzer, is a qualitative in vitro diagnostic test for the detection of Streptococcus pyogenes (Group A β-hemolytic Streptococcus, Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis.The Liat™ Strep A Assay utilizes nucleic acid purification and polymerase chain reaction (PCR) technology to detect Streptococcus pyogenes by targeting a segment of the Streptococcus pyogenes genome.
Sample typeThroat swabs (TS)Throat swabs (TS)
Assay resultsQualitativeQualitative
Assay formatReal-time PCRReal-time PCR
Strep A. TargetConserved region of Group A Streptococcus genomeConserved region of Group A Streptococcus genome
Extraction MethodAutomated with the ARIES® SystemsAutomated with the Liat™ Analyzer

Table 11-1: Similarities between New Device and Predicate

Table 11-2: Differences between New Device and Predicate

Differences
AttributeNew DevicePredicate Device (K141338)
Assay instrumentARIES System or ARIES M1 SystemLiat™ Analyzer
ControlsExtractable Sample Processing Control (SPC)Inactivated bacterial Internal Process Control (IPC)
Detection chemistryFluorescently labeled primers withquencher-labeled nucleotides; decreasein fluorescence over timeFluorescently-labeled hydrolysis probes;increase in fluorescence over time
Result InterpretationCt values coupled with melt curveanalysisCt and endpoint fluorescence values
Time-to-result~2 hours~15 minutes
Run size1-12 samples with the ARIES System1-6 samples with the ARIES M1 System1 sample with the Liat™ Analyzer

L. Standards/Guidance Documents Referenced:

Not applicable.

M. Test Principle:

The ARIES Group A Strep Assay is based on an expanded genetic alphabet technology, consisting of synthetic DNA base pair 2'-deoxy-5-methyl-isocytidine (iC): 2'deoxyisoguanosine (iG). The isobases (iC and iG) pair specifically with each other and

{7}------------------------------------------------

not with natural nucleotides. In addition, isobases are efficiently incorporated during PCR. During PCR amplification, a quencher-modified iGTP is incorporated by the polymerase opposite an iC and a fluorophore reporter attached to a PCR primer. If target is present and is amplified, assay fluorescence decreases with every cycle as amplification product accumulates. The decrease in assay fluorescence is monitored in real time using the ARIES System. Following PCR, the amplification products are thermally denatured and assay fluorescence is monitored. The strands of the amplification products are separated and assay fluorescence increases, thus enabling determination of the melting temperature (Tm) of the amplicon.

N. Performance Characteristics:

    1. Analytical performance:
    • a. Reproducibility/Precision/Repeatability:

Reproducibility of the ARIES® Group A Strep (GAS) Assay was evaluated by testing one lot of ARIES GAS Assay cassettes by two operators at each of three sites on five non-consecutive days. A blinded reproducibility panel, consisting of a GAS low positive (1X LoD), a GAS moderate positive (3X LoD) and a negative sample, was prepared by an independent operator. The reproducibility panel was tested in triplicate by each operator on each day of testing. The results of the reproducibility study are shown in Table 11-3.

Positive/Number (%)
LevelSite 1Site 2Site 3Overall
Moderate Positive30/30129/3030/3089/90
3X LoD(100)(96.7)(100)(98.9)
Low Positive29/3030/30128/30287/90
1X LoD(96.7)(100)(93.3)(96.7)
Negative0/301/300/301/90
(0.0)(3.3)(0.0)(1.1)

Table 11-3. ARIES® Group A Strep Assay Site-to-Site Reproducibility

1/30 samples was reported as Invalid on initial testing; reported as Positive upon repeat

2 All of 6 additional replicates that were tested were reported as Positive (overall 34/36 replicates, 94.4% were reported Positive at 1X LoD)

Lot-to-lot reproducibility of the ARIES® Group A Strep Assay was evaluated by one operator using a single ARIES system to test three lots of ARIES GAS Assay cassettes. A reproducibility panel was prepared containing moderate positive (3X LoD), low positive (1X LoD) and negative samples. A minimum of three replicates of each sample concentration were run a minimal five times on each cassette lot. The identity of these samples was blinded to the operator. The results of the study are shown in Table 11-4.

{8}------------------------------------------------

Positive/Tested (%)
LevelLot 1Lot 2Lot 3Overall
Moderate Positive3X LoD15/15(100)15/15(100)15/15(100)45/45(100)
Low Positive1X LoD14/15(93.3)15/15(100)13/15(86.7)42/45(93.3) 1
Negative0/15(0.0)0/15(0.0)0/15(0.0)0/45(0.0)
Table 11-4: ARIES Group A Strep Assay Lot-to-Lot Reproducibility Results
-------------------------------------------------------------------------------

All of 18 additional replicates that were tested (6 per reagent lot) were reported as Positive (overall 60/63 replicates, 95.2% reported Positive at 1X LoD)

Within-laboratory precision/repeatability of the ARIES® Group A Strep Assay was evaluated by two operators performing testing on a single ARIES System using a single lot of ARIES GAS Assay cassettes. A reproducibility panel was prepared containing moderate positive (3X LoD), low positive (1X LoD) and negative samples, which were blinded to operators with respect to expected GAS concentration. A minimum of three replicates of each sample concentration were run a minimal five times by each operator. The results of the study are shown in Table 11-5.

Table 11-5: ARIES® Group A Strep Assay Within-Laboratory Precision/Repeatability Results

LevelPositive/Tested (%)
Moderate Positive3X LoD30/30(100)
Low Positive1X LoD28/301(93.3)
Negative0/30(0.0)

All of 12 additional replicates that were tested were reported as Positive (overall 40/42 replicates, 95.2% were reported Positive at 1X LoD)

  • b. Linearity/assay reportable range:
    Not applicable. The ARIES® Group A Strep Assay is a qualitative assay.

  • C. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Stability:

Specimen Stability

{9}------------------------------------------------

Fresh specimen stability was assessed by testing specimens prepared at three times the Limit of Detection (3x LoD) in GAS-Negative Clinical Matrix (NCM) and stored at two different temperatures: ambient (20-25°C) and refrigerated (4-8°C). Following TO testing (the same day of specimen preparation), specimens stored at ambient temperature were tested at 24 and 48 hours; and specimens stored at 4-8°C were tested at 24 and 48 hours, 4 days and 7 days.

Data showed that GAS positive and GAS negative specimens stored at both temperatures (ambient and 4-8°C) generated expected 100% GAS positive and 100% GAS negative results, respectively, for all time points tested (Table 11-6). These results indicate that GAS specimens are stable for up to 48 hours when stored at 20-25°C or for up to seven days at 4°C to 8°C.

Target typeTime pointAgreement with Expected Resultsa, b
Ambient4-8°C
GAS in NCM(3x LoD)T0100% (20/20)c
24 hours100% (20/20)100% (20/20)
48 hours100% (20/20)100% (20/20)
4 days100% (20/20)
7 days100% (20/20)
NCM onlyT0100% (2/2)c
24 hours100% (2/2)100% (2/2)
48 hours100% (2/2)100% (2/2)
4 days100% (2/2)
7 days100% (2/2)

Table 11-6: ARIES Group A Strep Assay Fresh Specimen Stability Results

a Expected result for GAS 3x LoD diluted in NCM is 100% GAS positive; expected result for the NCM only is 100% GAS negative.

b An overall invalid rate of 0.6% (1/176) was observed from fresh specimen testing results.

^ TO is defined as the same day that the contrived specimens were prepared, aliquoted and tested before storing at the appropriate temperatures.

Frozen specimen stability was assessed by testing ARIES® Group A Strep Assay culture prepared at three times the Limit of Detection (3x LoD) in GAS-Negative Clinical Matrix (NCM) and GAS-Negative Simulated Matrix (NSM) and stored at ≤ -70°C. Following TO testing for specimens prepared in NCM or NSM, specimens stored at ≤ -70°C were tested at 1 month, 3 months and 6 months, with remaining time points to be tested at 9 and 12 months.

Data collected up to 6 months showed that GAS positive and GAS negative specimens stored at ≤ -70°C generated 100% GAS positive and 100% GAS negative results, respectively, for all time points tested. Results from frozen specimen

{10}------------------------------------------------

testing with the ARIES GAS assay are summarized in Table 11-7. Test results obtained to-date indicate that frozen specimens are stable for up to 6 months when stored at ≤ -70°C.

Target typeTime pointAgreement with Expecteda, bResults
GAS in NCM(3x LoD)T0100% (20/20)c
GAS in NCM(3x LoD)1 month100% (20/20)
GAS in NCM(3x LoD)3 month100% (20/20)
GAS in NCM(3x LoD)6 month100% (20/20)
GAS in NSM(3x LoD)T0100% (20/20)c
GAS in NSM(3x LoD)1 month100% (20/20)
GAS in NSM(3x LoD)3 month100% (20/20)
GAS in NSM(3x LoD)6 month100% (20/20)
NCM onlyT0100% (2/2)c
NCM only1 month100% (2/2)
NCM only3 month100% (2/2)
NCM only6 month100% (2/2)
NSM onlyT0100% (2/2)c
NSM only1 month100% (2/2)
NSM only3 month100% (2/2)
NSM only6 month100% (2/2)

Table 11-7: ARIES Group A Strep Assay Frozen Specimen Stability Results

Expected result for GAS 3x LoD diluted in NCM or NSM is 100% GAS positive; expected result for the NCM or NSM only is 100% GAS negative.

b An overall invalid rate of 0.6% (1/177) was observed for this study for data collected up to the 6 month time point.

& T0 is defined as the same day the contrived specimens were prepared, aliquoted and tested before storing at the appropriate temperature.

Shelf-Life Stability

A real time stability study was performed to evaluate the shelf life of ARIES GAS cassettes. Stability was assessed by testing 6 replicates of ARIES Extractable GAS Control (100X LoD) and negative targets (Copan Liquid Amies Medium) on three different lots of ARIES GAS cassettes stored at 2 different temperatures, 4°C (2 – 8°C) and room temperature (15 – 30°C), at 10 different time points extending out to 19 months. The study was designed following guidelines listed in EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved. Acceptance criteria for stability at each time point and temperature was established as 100% positivity for all GAS replicates and 100% negativity for all negative replicates. Data collected up to 7 months gave expected results indicating stability of the ARIES GAS cassettes up to 6 months

{11}------------------------------------------------

under the recommended storage conditions (with a 1 month safety margin). This study is on-going and will be updated as more data are collected.

Controls:

Process Control

Each ARIES® Group A Strep Assay cassette contains a Sample Process Control (SPC), which is processed with the sample and analyzed during the amplification reaction. The SPC verifies nucleic acid extraction, and proper reagent, cassette, ARIES® System, and assay protocol performance. The SPC has a known melting temperature (Tm) range and Ct range. Each time an assay is run, the system measures the temperature and fluorescence intensity of the SPC control to ensure the thermal and optical subsystems have remained in calibration.

External Controls

External controls should be tested according to guidelines or requirements of local, provincial and/or federal regulations or accreditation organizations. A reference Streptococcus pyogenes (S. pyogenes) strain or well characterized S. pyogenes clinical isolate may be used as a Positive Control. Liquid Amies medium may be used as a Negative Control. Alternatively, clinical specimens known to be positive or negative for S. pyogenes may be used as Positive and Negative External Controls, respectively. The ARIES® Group A Strep Assay Cassette Kit does not include external positive and negative controls.

  • d. Detection Limit:
    A Limit of Detection (LoD) study was performed to evaluate the analytical sensitivity of the ARIES® Group A Strep Assay using two strains of S. pyogenes diluted in negative clinical matrix (NCM) (pool of Group A Streptococcus negative clinical specimens - throat swabs in Liquid Amies). The LoD for each strain was determined as the lowest concentration that had a positivity rate of ≥ 95%. Preliminary LoD concentrations were determined using serial dilutions of each strain. All S. pyogenes strain concentrations were determined by plating and colony counting (CFU/mL). The preliminary LoD concentrations were confirmed by testing twenty (20) replicates of each strain. The final LoD concentrations are shown in Table 11-8.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". The word is in black and the background is white. There is a small copyright symbol after the "x" in "Luminex".

Assay TargetStrainLoD Concentration(CFU/mL)Positivity95% ConfidenceInterval
StreptococcuspyogenesBruno [CIP 104226](ATCC® 19615™)2.58E+0319/20 (95%)75.1% - 96.8%
StreptococcuspyogenesSF370; M1 GAS(ATCC® 700294™)4.13E+0320/20 (100%)83.2% - 100.0%

Table 11-8: ARIES Group A Strep Assay Limit of Detection Results

e. Analytical Reactivity (Inclusivity)

The analytical reactivity (inclusivity) of the ARIES® Group A Strep Assay was evaluated against nine (9) Streptococcus pyogenes strains that are different from those included in the Limit of Detection (LoD) study. Each strain was diluted to an initial concentration of three times (3X) the confirmed LoD in GAS-negative simulated matrix (NSM) and tested in triplicate. The results showed that 8 out of 9 strains were detected with 100% positivity (3/3 replicates) in the initial testing at 3x LoD; and the remaining 1 strain achieved 100% positivity (3/3 replicates) in testing at 5x LoD of the assay. The study results are shown in Table 11-9.

S. pyogenes StrainSourceCatalog #100% PositivityConcentration(CFU/mL)
Z018ZeptoMetrix08015121.24E+04
M-4, MGAS 10750 [FL01-86]ATCCBAA-10661.24E+04
M-6, MGAS 10394ATCCBAA-9461.24E+04
QC A62ATCC493991.24E+04
M-38, Typing strain C94 [13RS1]ATCC123701.24E+04
M-89, CDC-SS-1397 [NCTC 12067, PT-4245,R81/1352]ATCC7009491.24E+04
Typing strain T11ATCC123521.24E+04
M-1, Typing strain T1 [NCIB 11841, SF 130]ATCC123441.24E+04
M-3, Typing strain C203 [Dochez 1708]ATCC123842.07E+04

Table 11-9: ARIES® Group A Strep Assay Analytical Reactivity (Inclusivity) Results

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". A small registration mark is located to the right of the "x".

  • Analytical specificity: f.

Interfering Substances:

The potential inhibitory effect of non-microbial substances expected to be found in human throat swab specimens was evaluated for the ARIES® Group A Strep Assay by testing three replicates of GAS culture near the assay limit of detection (LoD) in simulated throat swab matrix that was spiked with the highest potential concentration of each substance. The expected results were obtained in all cases except in the presence of NyQuil and mucin. With NyQuil at 0.5% (v/v), 2/3 samples produced false negative results on initial testing, although repeat analysis under the same conditions produced the expected results. Invalid results were observed in the presence of mucin at >4mg/mL.

Interfering SubstanceTest Concentration
Advil®25 µg/mL
Amoxicillin25 µg/mL
Benadryl®350 ng/mL
Blood5% v/v
Cepacol®5 mg/mL
Chloraseptic® Sore throat (lozenges)5 mg/mL
Chloraseptic® Sore throat (spray)5% v/v
Chlor-Tripolon®25 ng/mL
Dequadin®12.5 µg/mL
Erythromycin15 µg/mL
Listerine® (mouth wash)5% v/v
Scope® (mouth wash)5% v/v
Purified Mucin Protein14 mg/mL
NyQuil® COMPLETE 20.5% v/v
Penicillin1.2 mg/mL
Ricola®5 mg/mL
Saline nasal spray5% v/v
Saliva5% v/v
Strepsils® extra5 mg/mL
Sucrets® Complete5 mg/mL
Toothpaste0.1 mg/mL
Tylenol®100 µg/mL
Zinc Lozenges0.1 mg/mL

Table 11-10. Interfering Substances Tested

Invalid results may be obtained in the presence of mucin >4g/mL

2 False negative results may be obtained in the presence of NyQuil® (0.5% v/v).

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, gray circle above the "i" in "Luminex". The word is in black and the background is white. The registered trademark symbol is present after the "x".

Cross Reactivity/Microbial Interference:

Analytical specificity for the ARIES® Group A Strep Assay was assessed with 35 microorganisms (Table 11-11) which may be present in throat swab specimens.

Cross reactivity was evaluated by spiking each potentially cross-reactive organism (CRO) in GAS-negative simulated matrix (NSM) to a final concentration of 10° CFU/mL (or the highest available concentration) and by testing in triplicate (n=3) on the ARIES System. All 35 microorganisms tested for cross reactivity yielded GAS negative results and are therefore considered non-reactive with the ARIES GAS Assay. Microbial interference was evaluated by preparing each CRO to a final concentration of 10° CFU/mL (or the highest available concentration) in NSM containing GAS culture prepared to 3 times the Limit of Detection (3x LoD). In the initial testing, 34 CROs tested in the presence of GAS yielded the expected positivity for GAS while one CRO, Treponema denticola, generated 2/3 GAS negative results. Repeat testing of this CRO at ~50% lower concentration gave 100% (3/3) GAS positivity (Table 11-11). Therefore, among the potential CROs tested with the ARIES GAS Assay, interference was only observed with high concentrations of T. denticola.

In addition to the 35 microorganisms tested on the ARIES System, 16 microorganisms and viruses were tested for potential cross reactivity by in silico analysis (Table 11-12). Two Pseudomonas spp., P. fluorescens LBUM223 and P. fluorescens KENGFT3, were at the borderline of the cut-off criteria for potential cross reactivity. Culture materials for these two organisms were not commercially available to evaluate with the ARIES GAS Assay; however, Pseudomonas aeruginosa, a species of Pseudomonas associated with respiratory infections in humans, showed no cross reactivity when tested with the ARIES GAS assay. Overall, the results of the in silico analysis showed that the potential for cross-reaction or interference with the ARIES GAS Assay is low.

Table 11-11. ARIES® Group A Strep Assay Cross Reactivity and Microbial
Interference Microorganisms Tested
#Cross Reacting Organisms
1Arcanobacterium haemolyticum
2Bacillus cereus
3Bordetella pertussis
4Burkholderia cepacia
5Campylobacter rectus*
6Candida albicans

{15}------------------------------------------------

#Cross Reacting Organisms
7Corynebacterium diphtheria
8Enterococcus faecalis
9Escherichia coli
10Fusobacterium necrophorum
11Haemophilus influenza
12Klebsiella pneumoniae
13Lactobacillus acidophilus
14Moraxella catarrhalis
15Neisseria gonorrhoeae
16Parvimonas micra
17Prevotella oralis
18Pseudomonas aeruginosa
19Saccharomyces cerevisiae
20Staphylococcus aureus
21Staphylococcus epidermidis
22Streptococcus agalactiae
23Streptococcus anginosus
24Streptococcus canis
25Streptococcus constellatus subsp. pharyngis
26Streptococcus dysgalactiae subsp. equisimilis
27Streptococcus gallolyticus
28Streptococcus intermedius
29Streptococcus mitis
30Streptococcus mutans
31Streptococcus pneumoniae
32Streptococcus salivarius
33Streptococcus sanguinus
34Treponema denticola ‡,‡
35Veillonella parvula

ٌ The final testing concentration for Campylobacter rectus was ≥ 4.55 x 103 CFU/mL.

¹ No titer information available. Growth in transparent film, no colony formation. *During initial testing of Treponema denticola in the presence of GAS at 3x LoD, 2 out of 3 replicates were GAS Negative. Repeat testing at ~50% of the initial concentration gave 100% (3/3)

Table 11-12. Microorganisms tested for Cross Reactivity/Microbial Interference in silico
--------------------------------------------------------------------------------------------
Genus and Species/Names
1Candida spp.9Adenovirus Type 1
2Enterococcus spp.10Adenovirus Type 7
3Klebsiella spp.11Human influenza virus A
4Lactococcus lactis12Human influenza virus B
5Legionella spp.13Human parainfluenza virus
6Mycoplasma pneumoniae14Human metapneumovirus
7Pseudomonas spp.15Respiratory syncytial virus Type B
8Stenotrophomonas maltophilia16Rhinovirus

GAS positivity.

{16}------------------------------------------------

Carry-Over/Cross-Contamination:

Carry-over and cross contamination for the ARIES® Group A Strep Assay was evaluated by testing thirty (30) high GAS positive samples in series, alternating with thirty (30) GAS negative samples consisting of GAS-Negative Simulated Matrix (NSM) only. The high positive samples were run adjacent to negative samples in an alternating pattern across ten (10) consecutive runs using one ARIES system. The results of this study show that no carry-over or cross contamination was observed, and the overall percent agreement with expected results was 100% for both high positive and negative samples.

  • g. Assay cut-off:
    For the ARIES® Group A Strep Assay, the target (sdaB) has a Ct cut-off, Tm window, and Tm Peak Threshold. In addition, the internal sample process control (SPC) also has a corresponding Ct cut-off, Tm window, and Tm Peak Threshold. Collectively, the cut-off values compose the assay protocol file parameters, which are used to determine the assay result for the detection target as POSTIVE, NEGATIVE, or INVALID. These values are hard-coded into the ARIES® Group A Strep Assay protocol file and are not modifiable. The Assay Protocol File parameters were determined, and their performance in the ARIES® Group A Strep Assay was evaluated according to the following general procedure:

  • . Initial Assay Protocol File parameters were set during internal optimization studies

  • The final Assay Protocol File parameters were then established during internal verification studies

  • . The selected Assay Protocol File parameter values were utilized in the determination of assay performance in the multi-site clinical trial conducted for the ARIES® Group A Strep Assay

The assay parameters for the ARIES® Group A Strep Assay are considered confidential and proprietary.

    1. Comparison Studies:
    • a. Method comparison with predicate device:

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The word is black, and there is a small, dark gray circle above the "i". A small registered trademark symbol is located to the right of the "x".

Not applicable.

  • b. Collection Media Comparison:
    Not applicable.

  • c. Swab Comparison:
    Not applicable.

    1. Clinical Performance:
      The clinical performance of ARIES® Group A Strep Assay was evaluated using deidentified, throat swab specimens prospectively collected from patients suspected of having respiratory infection attributable to Streptococcus pyoqenes (Group A B-hemolytic Streptococcus). Specimen collection was performed using the Copan or BD Liquid Amies Elution Swab (ESwab) Collection and Transport System.

Four (4) geographically distinct clinical sites within the United States prospectively collected specimens from January 2017 to May 2017 using the ARIES® System. All eligible clinical specimens were tested by both the reference method (bacterial followed by organism identification by Matrix-Assisted Laser culture Desorption/lonization - Time-of-Flight Mass Spectrometry (MALDI-TOF MS)) and the ARIES Group A Strep Assay and the results compared. Reference method testing was performed at a centralized testing facility while ARIES Group A Strep Assay testing was performed at each clinical site on their own clinical specimens.

A total of 735 throat swab specimens from subjects with signs and symptoms of pharyngitis were collected. Of these, 112 were excluded from the analysis of performance due to failure to comply with the reference culture protocol or delay in reference culture (67), inclusion criteria not met or confirmed (30), insufficient specimen volume (7), use of an incorrect collection and transport device or eligibility not confirmed (3), lack of a pure reference isolate (2), testing performed by an ineligible operator (2) or prior enrollment of the subject (1), leaving a total of 623 unique specimens available for analysis. All 623 specimens were tested for Group A Strep by both the reference method and the ARIES Group A Strep Assay. There were 6 specimens (6/623, 1.0%) that were re-tested with ARIES Group A Strep Assay because they yielded initial invalid results due to run failure or instrument error. All six (6) specimens that were re-run generated valid ARIES Group A Strep Assay results (i.e. positive or negative) after re-test. In addition, five (5) specimens generated inconclusive results by the comparator culture method (MALDI-TOF MS log(score

{18}------------------------------------------------

<2.00). All five (5) specimens with inconclusive reference results were excluded from the device performance calculations.

The ARIES® Group A Strep Assay Performance compared to bacterial culture followed by identification with MALDI-TOF MS is summarized in Table 11-13 below. Clinical sensitivity of the ARIES Group A Strep Assay for S. pyogenes against bacterial culture followed by identification with MALDI-TOF MS was 97.5% (156/160) with a lower bound 95% confidence interval of 93.7%. Clinical specificity of the ARIES Group A Strep Assay for S. pyoqenes against bacterial culture followed by identification with MALDI-TOF MS was 97.8% (448/458) with a lower bound 95% confidence interval of 96.0%. Positive Predictive Value (PPV) of the ARIES Group A Strep Assay for S. pyogenes against bacterial culture followed by identification with MALDI-TOF MS was 94.0% (156/166) with a lower bound 95% confidence interval of 89.3%. Negative Predictive Value (NPV) of the ARIES "Group A Strep Assay for S. pyogenes against bacterial culture followed by identification with MALDI-TOF MS was 99.1% (448/452) with a lower bound 95% confidence interval of 97.7%.

ARIES Group A StrepAssayBacterial Culture
PositiveNegativeTOTAL
Positive156102166
Negative41448452
TOTAL1604586183
95% CI
Sensitivity97.5%93.7% - 99.0%
Specificity97.8%96.0% - 98.8%
PPV94.0%89.3% - 96.7%
NPV99.1%97.7% - 99.7%

Table 11-13: ARIES Group A Strep Assay Performance Compared to Bacterial Culture followed by identification with MALDI-TOF MS

Two (2) of the ARIES Group A Strep Assay negative specimens that were positive by bacterial culture followed by identification with MALDI-TOF MS (i.e. False Negative) were Group A Strep negative by bidirectional sequencing analytically validated primers that targeted genomic regions distinct from the ARIES Group A Strep Assay.

2 Seven (7) of the ARIES Group A Strep Assay positive specimens that were negative by bacterial culture followed by identification with MALDI-TOF MS (i.e. False Positive) were positive by bi-directional sequencing analysis using analytically validated primers that targeted genomic regions distinct from the ARIES Group A Strep Assay.

Five (5) specimens generated inconclusive results by comparator culture followed by identification with MALDI-TOF MS. All five (5) specimens were excluded from the device performance calculations.

{19}------------------------------------------------

The study results demonstrate that the diagnostic accuracy of the ARIES Group A Strep Assay is acceptable for the detection of Streptoccus pyogenes in throat swab specimens from patients with signs and symptoms of pharyngitis.

    1. Expected values:
      During clinical evaluation of the assay, 26.8% (167/623) of throat swab specimens were reported as positive for Group A Strep by the ARIES "Group A Strep Assay. The prevalence of Group A Strep with the ARIES Group A Strep Assay summarized by age groups and by gender in Table 11-14 and Table 11-15 below.
Age (yrs)Total Positive by ARIES®Expected Value
<2325.0% (3/12)
2 - 1113532.5% (135/416)
12 - 212112.9% (21/163)
22 - 59828.6% (8/28)
≥6000.0% (0/4)
Overall16726.8% (167/623)1

Table 11-14: ARIES Group A Strep Assay Expected Values by Age Group

1 Includes five (5) subjects whose culture results were inconclusive but who had valid ARIES test results

Table 11-15: ARIES Group A Strep Assay Expected Values by Gender
--------------------------------------------------------------------------
GenderTotal Positive by ARIESExpected Value
Male8228.8% (82/285)
Female8525.1% (85/338)
Overall16726.8% (167/623)1

Includes five (5) subjects whose culture results were inconclusive but who had valid ARIES test results

O. Proposed Labeling:

The labeling provided in the submission satisfies the requirements of 21 CFR 809.10.

P. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.2680

Streptococcus spp. nucleic acid-based assay.(a)
Identification. AStreptococcus spp. nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify variousStreptococcus spp. nucleic acids extracted directly from clinical specimens. The device detects specific nucleic acid sequences for organism identification. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols, as appropriate.
(6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.