(29 days)
Not Found
No
The description focuses on the mechanical design and materials of a steerable guide catheter and a dilator. There is no mention of software, algorithms, or data processing that would indicate the use of AI or ML. The change described is a physical modification to a tip ring component.
No
The device is described as a guide catheter used for introducing other cardiovascular catheters, acting as a conduit rather than providing direct therapeutic action.
No
The device is a guide catheter used for introducing other cardiovascular catheters, not for making a diagnosis. While it allows visualization under fluoroscopy, this is for guidance during the procedure, not for diagnostic imaging.
No
The device description clearly details physical components like a catheter shaft, handle, steering knob, valves, and a dilator, indicating it is a hardware device. The 510(k) pertains to a design change to a physical component (tip ring).
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for "introducing various cardiovascular catheters into the left side of the heart through the interatrial septum." This describes a surgical or interventional procedure, not a diagnostic test performed on samples outside the body.
- Device Description: The description details a physical catheter designed for insertion into the body to facilitate other procedures. It does not mention any components or functions related to analyzing biological samples (like blood, urine, tissue, etc.).
- Lack of IVD Characteristics: There is no mention of reagents, assays, sample collection, or any other elements typically associated with in vitro diagnostics.
In summary, the device is a tool used during a medical procedure within the body, not a diagnostic test performed on samples from the body.
N/A
Intended Use / Indications for Use
The Steerable Guide Catheter is used for introducing various cardiovascular catheters into the left side of the heart through the interatrial septum.
Product codes (comma separated list FDA assigned to the subject device)
DRA
Device Description
The Steerable Guide Catheter consists of a Guide and a Dilator provided EtO sterile and for single use only. The Steerable Guide Catheter consists of a distal and proximal catheter shaft, a radiopaque tip ring, a handle with a steering knob, a hemostasis valve with a luer lock flush port, an atraumatic distal tip, and a Dilator with a single central lumen. The central lumen of the Guide allows for aspiration of air and infusion of fluids such as saline, and serves as a conduit during introduction and or exchange of the Dilator and ancillary devices (e.g. catheters) that have a maximum diameter of .204". The atraumatic distal tip of the Steerable Guide Catheter is radiopaque to allow visualization under fluoroscopy. The Dilator consists of a radiopaque shaft, an echogenic feature at the distal tip, a hemostasis valve with a flush port and an internal lumen designed to accept ancillary devices that have a maximum diameter of 0.035" (e.g. needles or guidewires). The Steerable Guide Catheter, Dilator and accessories are packaged in a tray that is individually pouched in a Tyvek/Nylon corner peel pouch, heat-sealed one time, and the single sealed pouch is placed into a cardboard nest and top-loading box.
The subject 510(k) pertains to a design change to the Steerable Guide Catheter. The proposed change to the Steerable Guide Catheter consists of a design and processing modification to the internal feature of the tip ring component at the distal tip.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
left side of the heart through the interatrial septum
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Tensile strength evaluation of the catheter tip was conducted on predicate (unmodified) and subject (modified tip) devices to demonstrate that the subject device meets performance specifications and is substantially equivalent to the predicate Steerable Guide Catheter.
Process validation was conducted to demonstrate that the manufacturing process consistently produces subject devices that meet performance specifications.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K083793, K091596, K093866, K100789, K112239, K161985
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1280 Steerable catheter.
(a)
Identification. A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.(b)
Classification. Class II (performance standards).
0
510(k) Summary
The 510(k) Summary is submitted in accordance with 21 CFR §807.92(c) and the requirements of the Safe Medical Device Act (SMDA) of 1990.
1. | Submitter's Name | Abbott Vascular |
---|---|---|
2. | Submitter's Address | 3200 Lakeside Drive Santa Clara, CA 95054 |
3. | Telephone | (408) 845-0937 |
4. | Fax | (408) 845-3734 |
5. | Contact Person | Makena Mc Gowan |
6. | Date Prepared | August 7, 2017 |
7. | Device Trade Name | Steerable Guide Catheter |
8. | Device Common Name | Steerable Catheter |
9. | Device Classification Name | Catheter, Steerable |
10. | Predicate Device Name | K083793 Steerable Guide Catheter |
K091596 Steerable Guide Catheter | ||
K093866 Steerable Guide Catheter | ||
K100789 Steerable Guide Catheter | ||
K112239 Steerable Guide Catheter | ||
K161985 Steerable Guide Catheter |
11. Device Description
The Steerable Guide Catheter consists of a Guide and a Dilator provided EtO sterile and for single use only. The Steerable Guide Catheter consists of a distal and proximal catheter shaft, a radiopaque tip ring, a handle with a steering knob, a hemostasis valve with a luer lock flush port, an atraumatic distal tip, and a Dilator with a single central lumen. The central lumen of the Guide allows for aspiration of air and infusion of fluids such as saline, and serves as a conduit during introduction and or exchange of the Dilator and ancillary devices (e.g. catheters) that have a maximum diameter of .204". The atraumatic distal tip of the Steerable Guide Catheter is radiopaque to allow visualization under fluoroscopy. The Dilator consists of a radiopaque shaft, an echogenic feature at the distal tip, a hemostasis valve with a flush port and an internal lumen designed to accept ancillary devices that have a maximum diameter of 0.035" (e.g. needles or guidewires). The Steerable Guide Catheter, Dilator and accessories are packaged in a tray that is individually pouched in a Tyvek/Nylon corner peel pouch, heat-sealed one time, and the single sealed pouch is placed into a cardboard nest and top-loading box.
The subject 510(k) pertains to a design change to the Steerable Guide Catheter. The proposed change to the Steerable Guide Catheter consists of a design and processing modification to the internal feature of the tip ring component at the distal tip.
1
12. Indication for Use
The Steerable Guide Catheter is used for introducing various cardiovascular catheters into the left side of the heart through the interatrial septum.
13. Technological Characteristics
Comparison of the new device and predicate device demonstrate that the technological characteristics such as the fundamental scientific technology, material, and indications for use are substantially equivalent to the current marketed predicate device.
14. Performance Data
Tensile strength evaluation of the catheter tip was conducted on predicate (unmodified) and subject (modified tip) devices to demonstrate that the subject device meets performance specifications and is substantially equivalent to the predicate Steerable Guide Catheter.
Process validation was conducted to demonstrate that the manufacturing process consistently produces subject devices that meet performance specifications.
15. Conclusions
The Steerable Guide Catheter has the same indications for use and performs as well as the predicate device.
2
Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
3
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Abbott Vascular Makena Mc Gowan Regulatory Affairs Specialist 3200 Lakeside Drive Santa Clara, California 95054
Re: K172394
Trade/Device Name: Steerable Guide Catheter Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: August 7, 2017 Received: August 8, 2017
Dear Ms. Mc Gowan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
4
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure