AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AESKUSLIDES® nDNA (Crithidia luciliae) is an indirect immunofluorescence assay utilizing Crithidia luciliae coated slides as a substrate for the qualitative and/or semi-quantitative determination of antibodies to native double stranded DNA (dsDNA) in human serum. This in vitro diagnostic assay is used as an aid for the diagnosis of Systemic Lupus Erythematosus (SLE) in conjunction with other clinical and laboratory findings. The assay can be processed manually and analyzed at the microscope or processed and analyzed with HELIOS® AUTOMATED IFA SYSTEM. All suggested results obtained with the HELIOS® AUTOMATED IFA SYSTEM must be confirmed by trained personnel.

Device Description

AESKUSLIDES® nDNA (Crithidia luciliae) is an indirect immunofluorescence assay utilizing Crithidia luciliae coated slides as a substrate for the qualitative and/or semiquantitative determination of antibodies to native double stranded DNA (dsDNA) in human serum.

Each kit contains (Quantity depends on product variant):

  • -Slides, each containing 10 wells coated with Crithidia Luciliae cells
  • 4.0 ml vial containing Fluorescein (FITC) labelled Anti-human Antibody lgG conjugate in a solution of BSA, ready for use
  • -0.5 ml vial of positive control containing human serum (diluted), ready for Use
  • 0.5 ml vial of negative control containing diluted human serum, ready for use -
  • -8.0 ml vial of mounting medium containing a solution of glycerol and PBS, ready for use
  • 70 ml bottle of sample buffer, containing BSA, PBS and ready for use -
  • -100 ml bottle of wash buffer, concentrated buffer 1:10 in distilled water, containing BSA, PBS.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the AESKUSLIDES® nDNA (Crithidia luciliae) device, based on the provided document:


Acceptance Criteria and Device Performance Study for AESKUSLIDES® nDNA (Crithidia luciliae)

1. Table of Acceptance Criteria and Reported Device Performance

The document presents several studies with specific acceptance criteria. Below is a summary for the Between-Lab Precision Study and the Lot-to-Lot Precision Study, as these provide clear, quantitative acceptance criteria and corresponding results. Other studies, like Serum Stability and Carryover, also met their respective criteria (e.g., 100% agreement, no significant deviations).

Between-Lab Precision Study (Excluding Borderline Samples - Method B: Reader Confirmation)

Type of AgreementAcceptance CriteriaReported Device Performance (All Sites, excluding borderline)
Positive Agreement> 90%97.9% (96.9 - 98.5)% CI
Negative Agreement> 90%99.4% (98 - 99.8)% CI
Overall Agreement> 90%98.2% (97.4 - 98.8)% CI

Between-Lab Precision Study (Excluding Borderline Samples - Method C: Manual)

Type of AgreementAcceptance CriteriaReported Device Performance (All Sites, excluding borderline)
Positive Agreement> 90%99.5% (99 - 99.8)% CI
Negative Agreement> 90%99.7% (98.4 - 100)% CI
Overall Agreement> 90%99.6% (99.1 - 99.8)% CI
Fluorescence Intensity Agreement> 90%97.5% (96.6 - 98.1)% CI

Lot-to-Lot Precision Study (Combined Readers)

Type of AgreementAcceptance CriteriaReported Device Performance (Combined Readers)
Positive Agreement> 90%100% (99.3 - 100)% CI
Negative Agreement> 90%100% (96.9 - 100)% CI
Overall Agreement> 90%100% (99.4 - 100)% CI
Fluorescence Intensity Agreement> 90%100% (99.4 - 100)% CI

2. Sample Size Used for the Test Set and Data Provenance

The primary clinical evaluation and method comparison studies used 776 clinical samples.

  • Test Set Size: 776 samples.
  • Data Provenance:
    • 746 samples were obtained from 10 US BioBanks (BioChain, BioReclamationIVT, Bioserve, ConversantBio, Cureline, DiscoveryLifeSciences, iSpecimen, Precision for Medicine, ProMedDx, and Vitrologic).
    • 30 serum samples were from a German University Hospital (used to complement rare diagnoses like Vasculitis).
  • Retrospective/Prospective: The samples were collected from BioBanks, indicating they are retrospective samples with pre-existing diagnoses.
  • Sample Characteristics: The samples were selected to reflect various diagnoses relevant to the study (e.g., 297 SLE, 479 other diseases) and different ethnic groups in the US population (White, Black/Black African, Asian, Hispanic). Serum stability over multiple freeze-thaw cycles and long-term storage was also evaluated and deemed acceptable.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The term "ground truth" for the diagnosis of the clinical samples (used in the method comparison studies) was established based on diagnostic standards used in the U.S. and Germany, such as ACR criteria. A written statement from different serum suppliers confirmed this.

For evaluation of the device performance within the studies (e.g., precision studies, clinical study), the ground truth for individual results was typically established by:

  • Two independent readers.
  • Qualifications of experts: The document consistently refers to "trained personnel" and "trained readers" to perform manual microscopy and confirm automated results. Specific certifications or years of experience (e.g., "Radiologist with 10 years of experience") are not explicitly provided in this document.

4. Adjudication Method for the Test Set

For studies involving human readers:

  • Results were analyzed by two independent readers.
  • The document implies a consensus-based approach or separate reporting of each reader's results for evaluation, but a formal adjudication method (like "2+1" or "3+1") where a third reader resolves discrepancies is not explicitly stated. For instance, in the Between-Lab and Within-Lab Precision studies, results for two readers were calculated separately and then combined. In cases of discrepancy (e.g., sample S1 in Within-Lab precision for Method B, where Reader 1 found 30 Negatives and Reader 2 found 6 Negatives and 24 Positives), the combined percentage reflects this disagreement rather than a formal adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not explicitly describe a formal MRMC comparative effectiveness study in the sense of comparing human reader performance with and without AI assistance to quantify an "effect size" of improvement.

Instead, it evaluates the agreement between:

  • Method C (Manual): Manual processing and manual reading by human readers.
  • Method B (Reader Confirmation): Automated processing and automated imaging, followed by manual reading of digital images by human readers.
  • Method A (HELIOS): Automated processing, automated imaging, and automated software interpretation (with required human confirmation).

The study compared the performance and agreement between these methods. It highlights that Method A (software-only interpretation) had lower agreement and sensitivity compared to human interpretation (Method B and C), underscoring the need for human confirmation. However, it does not quantify how much human readers improve when using AI as an assistance tool, but rather assesses the standalone performance of the AI component and the human element with and without automated processing.


6. Standalone (Algorithm Only) Performance

Yes, standalone performance was done:

  • Method A (HELIOS): This method represents the algorithm's standalone performance, where the "HELIOS DNA Pattern Plus Software" performs positive/negative classification.
  • Results: For Method A (HELIOS suggestion, excluding borderline samples) in the Between-Lab Precision Study:
    • Positive Agreement: 72.7% (69.1 - 76)% CI (across all sites)
    • Negative Agreement: 90% (84.7 - 93.6)% CI (across all sites)
    • Overall Agreement: 76.5% (73.5 - 79.3)% CI (across all sites)
      The document notes that acceptance criteria for Method A were met for negative and overall agreement, but positive agreement was sometimes below 70% for individual site comparisons, explaining that this was due to out-of-focus images that the software couldn't interpret as positive. This reinforces the requirement for human confirmation.

7. Type of Ground Truth Used (Clinical Samples)

For the 776 clinical samples used in the method comparison studies, the "ground truth" for patient diagnosis (e.g., SLE, other rheumatic diseases, autoimmune liver diseases, infections, leukemia) was established based on diagnostic standards used in the U.S. and Germany (e.g., ACR criteria). This implies a clinical diagnosis based on a combination of clinical findings and existing laboratory tests, rather than a single definitive gold standard like pathology or outcome data.

For validation within the performance studies (e.g., precision, stability), the "expected result" (positive/negative, fluorescence intensity) of specific control or serum samples was pre-defined or derived from manual readings by trained personnel, which can be considered an expert consensus type of ground truth for analytical performance.


8. Sample Size for the Training Set

The document does not provide information regarding the sample size used for training the HELIOS DNA Pattern Plus Software (Method A). It mainly focuses on the performance evaluation of the device in various settings.


9. How the Ground Truth for the Training Set Was Established

Since information on the training set sample size is absent, how its ground truth was established is also not specified in this document.

{0}------------------------------------------------

Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{1}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY AND INSTRUMENT COMBINATION TEMPLATE

  • A. 510(k) Number: K172348
  • B. Purpose for Submission: New Device
  • C. Measurand: native double helix DNA (dsDNA)
  • D. Type of Test: Qualitative and/or semi-quantitative, indirect immunofluorescence
  • E. Applicant: AESKU.Diagnostics GmbH & Co. KG
  • F. Proprietary and Established Names: AESKUSLIDES® nDNA (Crithidia luciliae)

G. Regulatory Information:

  1. Regulation section: §CFR 866.5100 - Antinuclear antibody immunological test system

  2. Classification: Class II

  3. Product code: LSW, Anti-DNA Antibody, Antigen and Control

  4. Clinical use Immunology (82)

{2}------------------------------------------------

H. Intended Use:

    1. Intended use(s):
      AESKUSLIDES® nDNA (Crithidia luciliae) is an indirect immunofluorescence assay utilizing Crithidia luciliae coated slides as a substrate for the qualitative and/or semi-quantitative determination of antibodies to native double stranded DNA (dsDNA) in human serum. This in vitro diagnostic assay is used as an aid for the diagnosis of Systemic Lupus Erythematosus (SLE) in conjunction with other clinical and laboratory findings. The assay can be processed manually and analyzed at the microscope or processed and analyzed with HELIOS® AUTOMATED IFA SYSTEM. All suggested results obtained with the HELIOS® AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
    1. Indication(s) for use:
      Same as intended use
    1. Special conditions for use statement(s):
      1. For prescription use only
  1. This device is only for use with reagents that are indicated for use with the device.

  2. The device is for use by a trained operator in a clinical laboratory setting. 4. All software-aided results must be confirmed by the trained operator. 5.For use only by manual microscopy or with HELIOS® AUTOMATED IFA SYSTEM.

l. Device Description:

AESKUSLIDES® nDNA (Crithidia luciliae) is an indirect immunofluorescence assay utilizing Crithidia luciliae coated slides as a substrate for the qualitative and/or semiquantitative determination of antibodies to native double stranded DNA (dsDNA) in human serum.

Each kit contains (Quantity depends on product variant):

  • -Slides, each containing 10 wells coated with Crithidia Luciliae cells
  • 4.0 ml vial containing Fluorescein (FITC) labelled Anti-human Antibody lgG conjugate in a solution of BSA, ready for use
  • -0.5 ml vial of positive control containing human serum (diluted), ready for Use
  • 0.5 ml vial of negative control containing diluted human serum, ready for use -
  • -8.0 ml vial of mounting medium containing a solution of glycerol and PBS, ready for use
  • 70 ml bottle of sample buffer, containing BSA, PBS and ready for use -
  • -100 ml bottle of wash buffer, concentrated buffer 1:10 in distilled water, containing BSA, PBS.

{3}------------------------------------------------

Standard Ref.DescriptionTests
53.100.USnDNA (Crithidia luciliae) (10wells)100
53.100.US.DemonDNA (Crithidia luciliae) (10wells) Demo kit20
53.100.US.Bulk5nDNA (Crithidia luciliae) (10wells) bulk kit x5500
53.100.US.Bulk10nDNA (Crithidia luciliae) (10wells) bulk kit x101000

Not provided in the kit:

Ref.ReagentQuantity/ VolumeDescriptionReadyto use
EBIFAEvansBlue0.2%1x3mlCapped white: Blue colouredsolutionContaining: PBS, EvansBlue.Dilute the Evans Blue 0.2%1:3000 in 1x WBIFANO

HELIOS AUTOMATED IFA SYSTEM (K153117) or equivalent manual microscope.

J. Substantial Equivalence Information:

    1. Predicate device name(s): NOVA Lite dsDNA Crithidia luciliae
    1. Predicate 510(k) number(s): K880742

{4}------------------------------------------------

July 28, 2017

3. Comparison with predicate device:

ItemPredicateNOVA Lite dsDNA Crithidia luciliaeAESKUSLIDES® nDNA (Crithidia luciliae)
SimilaritiesIntended UseNOVA Lite® dsDNA Crithidia luciliae is anindirect immunofluorescent assay for thescreening and semi-quantitativedetermination of anti-double strandedDNA (dsDNA) in human serum. Thepresence of anti-double stranded DNA canbe used in conjunction with otherserological tests and clinical findings to aidin the diagnosis of systemic lupuserythematosus (SLE)AESKUSLIDES® nDNA (Crithidia luciliae) is anindirect immunofluorescence assay utilizing Crithidialuciliae coated slides as a substrate for the qualitativeand/or semi-quantitative determination of antibodies tonative double stranded DNA (dsDNA) in human serum.This in vitro diagnostic assay is used as an aid for thediagnosis of Systemic Lupus Erythematosus (SLE) inconjunction with other clinical and laboratory findings.The assay can be processed manually and analyzed atthe microscope or processed and analyzed withHELIOS® AUTOMATED IFA SYSTEM. All suggestedresults obtained with the HELIOS® AUTOMATED IFASYSTEM must be confirmed by trained personnel.
MethologyImmunofluoreszenz assay (IFA)same as Predicate
ProcedureStandard IFA techniquesame as Predicate
Resultsqualitative and semi-quantitative titersame as Predicate
SamplesMatrixSerumsame as Predicate
AnalytedsDNAsame as Predicate
AntigendsDNA Crithidia luciliae cellssame as Predicate
FluorescenceMarkerFITCsame as Predicate
Controlsone positive control, one negative controlsame as Predicate
conjugateScreeninganti-human IgGsame as Predicate
DifferencesScreeningdilution1:10same as Predicate
Storage2 - 8 °Csame as Predicate
Shelf life24 monthssame as Predicate
Slides12 wells coated with antigen10 wells coated with antigen
ManualInterpretationof resultManual fluorescense microscopyManual fluorescense microscopy or with HELIOS w/trained operator verification
Automatedinterpretationof resultsN/AHELIOS w/ trained operator verification

Table 6: comparison with predicate device

{5}------------------------------------------------

K. Standard/ Guidance Document Referenced (if applicable):

Table 7: List of Standards / Guidance Documents

#Standards TitleVersion
1ISO 14971 - Medical Devices - Application of risk management to medical devicesSecond Edition
2IEC 62366 - Medical Devices - Part 1: Application Of Usability Engineering To Medical Devices [Including CORRIGENDUM 1 (2016)]Edition 1.0
315223-1- Medical Devices - Symbols To Be Used With Medical Device Labels, Labelling, and Information to be supplied - Part 1: General RequirementsSecond Edition
4Interference Testing in Clinical ChemistryEP07-A2
5Evaluation of Detection Capability for Clinical Laboratory Measurement ProceduresEP17-A2
6Evaluation of Stability of In Vitro Diagnostic ReagentsEP25-A
7Defining, Establishing and Verifying Reference Intervals in the Clinical LaboratoryEP28-A3c
8Evaluation of the Linearity of Quantitative Measurement Procedures A Statistical ApproachEP06-A
9Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline-Fourth EditionGP44-A4
10Guidance for Industry and FDA Staff: Recommendations for Anti-Nuclear Antibody (ANA) Test System Premarket (510k) Submissions-
11Factors to Consider Regarding Benefit-Risk in Medical Device Product Availability, Compliance, and Enforcement Decisions-
12Applying Human Factors and Usability Engineering to Medical Devices-

L. Test Principle

The AESKUSLIDES nDNA assay utilizes indirect immunofluorescent antibody assay techniques. Patient sera are diluted in wash/sample buffer and applied to a well on the slide. nDNA antibodies, if present, will bind to antigens coated on the slide. After washing with wash/sample buffer, a conjugate specific for human IgG is applied which binds to the nDNA antibodies immobilized on the slide surface. After a final wash to remove excess conjugate, the slide is mounted and read as soon as possible using a fluorescence microscope for manual microscopy or HELIOS AUTOMATED IFA SYSTEM.

Manual interpretation of test results

The fluorescence intensity level is the intensity of the specific fluorescence expressed as a numeric value. These values, if present, are reported as a number between "O" (no specific flurescence) and "4+" (very strong visible reaction).

{6}------------------------------------------------

July 28, 2017

IntensityInterpretation
4+high positivemaximal fluorescence, very strongvisible reaction; brilliant yellow-green
3+positivestrong visible reaction; less brilliant as4+; yellow-green fluorescence
2+positivemoderate visible reaction; definite butdull yellow-green fluorescence
1+positiveweak visible reaction, very dimsubdued fluorescence
0negativeno specific fluorescence

AESKU recommends a screening dilution of 1:10, followed by serial dilution for semiquantitative determinations and suggests each laboratory establish its own screening dilution and titration scheme based on its population and instrumentation.

Qualitative evaluation

A serum dilution is considered negative for nDNA antibodies if the cells exhibit < 1+ fluorescence of the kinetoplast. Likewise, a serum dilution is considered positive for nDNA antibodies if the cells exhibit ≥ 1+ fluorescence of the kinetoplast. A sample is considered positive for nDNA antibodies if it exhibits exhibit ≥ 1+ fluorescence of the kinetoplast at a sample dilution of 1:10 or greater. Operators should report all titers and specific fluorescence staining seen.

Semi-quantitative evaluation

The endpoint titer is defined as the highest sample dilution factor for which specific fluorescence of the kinetoplast is identifiable.

The titers are classified as:

  • 1:10 are considered low titers
  • 1:20 and 1:40 are considered medium titers
  • 1:80 and greater are considered high titers

{7}------------------------------------------------

Automated instrument interpretation of test results

The HELIOS DEVICE SOFTWARE examines the fluorescence intensity and uses an analysis algorithm which takes exposure and pixel frequency into account. It examines relevant regions of the captured images, such as the fluorescence cell area and subtracts the background, to provide an assessment of positive or negative results.

IntensityInterpretation
+nDNA positive
-nDNA negative

Trained operators must confirm all suggestions. Intensity and End Point Titers are identified by the HELIOS DEVICE SOFTWARE and pattern suggestions are made by the HELIOS PATTERN RECOGNITION tool.

AESKU recommends a screening dilution of 1:10, followed by serial dilution for semiquantitative determinations and suggests each laboratory establish its own screening dilution and titration scheme based on its population.

M. Performance Characteristics:

1. Analytical performance

During the study 4 different methods have been applied:

MethodProcessingImagingReading/Evaluation odSlideAlternate Name ofMethod
AAutomatedAutomatedAutomated (SoftwareInterpretation)HELIOS
BAutomatedAutomatedManual (read of digitalimage)HELIOS UserEvaluation
CManualManualManual (read ofmicroscope field)ManualAESKUSLIDESnDNA
DManualManualManual (read ofmicroscope field)Manual - PredicateNOVA Lite dsDNA

Method D is the predicate for method C

In the clinical study the method comparison was: Method D vs Method C Method C vs. Method A, B Method B vs Method A Nomenclature and acronyms used in studies

{8}------------------------------------------------

July 28, 2017

NomenclatureDescription/ Also known as
AESKUSLIDES nDNAnDNA
HELIOS AUTOMATED IFA SYSTEMHELIOS
Manual or Manual ReadingResults obtained by the operator after reading and interpreting the slides with a traditional fluorescence microscope
Method AHELIOS + AESKUSLIDES nDNA w/slides read by HELIOS
Method BHELIOS + AESKUSLIDES nDNA w/slides read by User
Method CManual AESKUSLIDES nDNA w/slides read by User
Method DManual Nova Lite dsDNA w/slides read by User
Negative IFA score/result, negativeneg, negative
Positive IFA score/resultpos, psoitive
Diseases AbbreviationDiseases Name
SLESystemic lupus erythematosus
APSAntiphospholipid Syndrome
MCTDMixed Connective Tissue Disease
UCTDUndifferentiated connective tissue disease
SScSystemic sclerosis
PMPolymyositis
DMDermatomyositis
RARheumatoid Arthritis
AIHAutoimmune Hepatitis
PBCPrimary biliary cholangitis
PSCPrimary sclerosing cholangitis
AAVANCA-associated vasculitides
HCVHepatitis C Virus
HBVHepatitis B Virus

Table 8: Nomenclature and acronyms

{9}------------------------------------------------

Description of study sample set:

776 Clinical samples have been used for the Clinical Evaluation and Method Comparisons studies described herein.

An exhaustive search for US Clinical Samples from 10 BioBanks (BioChain, BioReclamationIVT, Bioserve, ConversantBio, Cureline, DiscoveryLifeSciences, iSpecimen, Precision for Medicine, ProMedDx, and Vitrologic) has been performed and 746 Clinical Samples have been found. The samples have been selected according to their diagnosis, to reflect all important diagnosis for this study (see table). The sample set has been completed with 30 serum samples from a German University Hospital to complement some rare diagnosis (Vasculitis).

Study Sample Set
Sample DefinitionDiagnosisn
TargetDiagnosisSLESLE277
DifferentialDiagnosisAPSAPS19
Other rheumaticdiseasesMixed connective tissue disease (MCTD)31
Undifferentiated connective tissue disease (UCTD)21
Systemic sclerosis (SSc)44
Sjögren's syndrome30
Dermatomyositis (DM) and polymyositis (PM)37
Rheumatoid arthritis (RA)58
Autoimmune LiverDiseasesAutoimmune Hepatitis21
Primary biliary cholangitis (PBC)20
Primary sclerosing cholangitis (PSC)10
VasculitisAAVs44*
FibromyalgiaFibromyalgia30
InfectionsHCV44
HBV34
EBV25
LeukemiaLymphoma, Myeloma31
Sum776

Table: 9 Serum samples set for clinical and method comparison studies

*30samples from Germany

The diagnosis criteria of the different samples have been made in agreement with diagnostic standards used in the U.S and Germany. A written statement from different serum suppliers is available on request. Standard criteria are for example ACR criteria.

The US sample set was also selected to contain different ethnic groups (e.g. White, Black/Black African, Asian, Hispanic) to reflect the ethnic composition of the US population as good as possible.

{10}------------------------------------------------

The samples have been checked for purity and volume by visual inspection and for further contaminations by an in house prescreening before inclusion into the study. All 776 sera passed and were determined suitable for the study.

All serum samples which have been used for this studies have been shipped and stored at -20°C. Upon arrival from the Biobanks they were thawed once to aliquot them for the different studies. Subsequently, samples were frozen again. Therefore, samples underwent 2-3 freezing-thawing cycles before they were used in the studies.

To confirm that repeated thawing and freezing and also long term storage at -20°C has no influence on the samples we have performed the following two studies:

Serum Stability AESKUSLIDES nDNA (Crithidia luciliae)

Procedure:

Serum Stability was shown by testing 8 serum samples (7 positive, 1 negative). For each serum, one aliquot has been prepared that serves as control (no freezethawing). A second aliquot for each serum has been prepared that underwent 4 freeze-thaw cycles. Then, all samples and controls have been analyzed on AESKUSLIDES nDNA according to the IFU. Slides were read by two independent readers. Each freeze/thaw sample was compared to the respective control sample.

SampleIDAESKUSLIDES nDNAResultGrading
1Poslow positive
2Poslow positive
3Poslow positive
4Posmedium positive
5Posmedium positive
6Poshigh positive
7Poshigh positive
8NegNeg

{11}------------------------------------------------

Acceptance criteria:

  • -Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
  • -Overall Agreement: > 85%.
  • -FI is allowed to differ maximum + or - 1 from the expected value.
  • -FI Agreement: > 85%.

Results

  • All samples fulfilled the above criteria. No differences were observed between sera thawed a single time and sera that had undergone 4 cycles of freeze/thawing:
  • -Positive/Negative/Overall Agreement was 100% for AESKUSLIDES nDNA. All positive samples have been found positive, all negative samples have been found negative.
  • FI agreement was 100% for AESKUSLIDES nDNA. No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.

Raw data are in Attachment 10_Section 12_Raw data serum stability

Conclusions

All criteria are fulfilled in each test. The results show that multiple freeze-thaw cycles have no effect on test results.

Long Term Serum Stability AESKUSLIDES nDNA (Crithidia luciliae)

To show that long term storage of serum samples at -20°C has no effect on performance and results of AESKUSLIDES nDNA.

Procedure:

Seven different serum samples (4 positive and 3 negative) have been aliquoted and stored at -20°C for a time period of at least 14 months. Frozen aliquots of the different sera have been thawed and assayed on AESKUSLIDES nDNA at the indicated time points. All tests have been performed manually according to the IFU and subsequently analyzed at the microscope

SampleIDExpected Result
Pos/NegFI
Sample 1Pos3
Sample 2Pos2
Sample 3Pos4
Sample 4Neg0
Sample 5Neg0
Sample 6Neg0
Sample 7Pos2

{12}------------------------------------------------

Acceptance Criteria:

Positive sera have to be found positive and negative sera have to be found negative throughout the whole testing period. The fluorescence intensity (FI) is allowed to differ maximum +/- 1 level from the expected value at each test time point.

Results:

Positive samples have been found positive, negative samples have been found negative at all test time points. FI did not differ more than +/- 1 level from the expected values.

Sample IDmonth 0month 4month 8month 10month 12month 14
Pos/NegFIPos/NegFIPos/NegFIPos/NegFIPos/NegFIPos/NegFI
Sample 1Pos3Pos3Pos4Pos3Pos3Pos3
Sample 2Pos2Pos2Pos2Pos2Pos1Pos2
Sample 3Pos4Pos3Pos4Pos3Pos4Pos3
Sample 4Neg0Neg0Neg0Neg0Neg0Neg0
Sample 5Neg0Neg0Neg0Neg0Neg0Neg0
Sample 6Neg0Neg0Neg0Neg0Neg0Neg0
Sample 7Pos2Pos2Pos2Pos2Pos2Pos2

Table 10: Results long term serum stability

Conclusion:

All acceptance criteria have been fulfilled for the tested sera. The results show that long term storage of sera at -20°C over a time period of at least 14 months has no effect on performance and test results of AESKUSLIDES nDNA.

a) Precision/Reproducibilty

Within-Lab Precision AESKUSLIDES nDNA (Crithida luciliae)

To assess Within-Lab Precision of AESKUSLIDES nDNA (Crithidia luciliae) processed on HELIOS (Method A - HELIOS suggestions and Method B - Reader Confirmation of HELIOS images) and processed manually (Method C), based on CLSI Guideline EP12-A2.

Sample Set

11 samples were assayed on AESKUSLIDES nDNA. The sample set includes borderline, low, medium, high positive, and negative samples. The samples are characterized in the following table:

Sample IDResultGrading
1Posborderline
2Poslow positive
3Poslow positive
4Posmedium positive
5Posmedium positive
6Posmedium positive

Table 11: Sample table for Within-Lab Precision

{13}------------------------------------------------

July 28, 2017

7Poshigh positive
8Poshigh positive
9Poshigh positive
10Negnegative
11Negnegative

Procedure

11 samples have been tested on five days, two runs per day, three replicates per sample per run, resulting in 30 data points for each sample.

This Within-lab precision study was done for three different Methods:

Method A (HELIOS)Processing of slides and image recording at HELIOS +HELIOS result suggestion
Method B (ReaderConfirmation)HELIOS images recorded in Method A + result analysisby two independent readers
Method C (Manual)Manual processing of slides and result analysis at themicroscope by two independent readers

All runs have been performed according to the respective IFUs. 1 positive and 1 negative control (kit controls) were included in each run. Samples were tested at 1:10 dilution. Results were analyzed by two independent readers.

Data Analysis

Positive/Negative Classification was recorded for each sample in each run and for each Method. Pos/Neg Classification in Method A (HELIOS) is provided by the nDNA Pattern Plus software tool.

For Method C (Manual performance) fluorescence intensity was also reported.

As an estimate of the imprecision of the three methods A, B and C the percentages of positive and neqative results will be calculated for each sample:

  • % Positive (number of positive calls divided by the total number of data points for each sample)
    • % Negative (number of negative calls divided by the total number of data points per sample)

Results of the two readers (Method B/C) will be calculated and displayed separately as well as combined.

Acceptance Criteria

· Positive sera have to be found positive, and negative sera have to be found negative*.

· Reported FI is allowed to differ max. ± 1 level within the study

{14}------------------------------------------------

  • borderline samples are samples with analyte concentration near the cutoff. These samples are very low positive. They can be evaluated also negative in certain cases, depending on the reader, the subjective manner of result analysis and normal test variances.

For Method A we accept lower percentage of positive/negative results for positive and negative samples, respectively, because we state in the IFU that all results generated by the HELIOS (Method A) must be confirmed by trained personnel.

Results

The results are presented in % Positive and % Negative tables of each sample. Percentages were calculated for the two readers separately (30 replicates for each sample) and for both readers combined (60 data points per sample).

Raw Data can be found in 2 Report Within-Lab Precision re-evaluation nDNA

Table 12: Results - Within Lab Precision AESKUSLIDES nDNA - Method A, B, and C number of positive and negative results

Sample IDNMethod AMethod BMethod C
n Negn Posn Negn Posn Negn Pos
Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2
S130300306024452625
S2301020003030003030
S3301713033027003030
S4301416412629003030
S530030003030003030
S630228003030003030
S730030003030003030
S830030003030003030
S930129003030003030
S1030255303000293010
S1130219303000303000

{15}------------------------------------------------

Sample IDNMethod AMethod BMethod C
% Negative% Positive% Negative% Positive% Negative% Positive
Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2
S130100010020.0080.013.316.786.783.3
S23033.366.70010010000100100
S33056.743.30101009000100100
S43046.753.313.33.386.796.700100100
S53001000010010000100100
S6306.793.30010010000100100
S73001000010010000100100
S83001000010010000100100
S9303.396.70010010000100100
S103083.316.71001000096.71003.30
S113070.030.01001000010010000

Table 13: Results – Within Lab Precision AESKUSLIDES nDNA – Method A, B and C – percentages of positive and negative results

Table 14: Results – Within Lab Precision AESKUSLIDES nDNA – Method A, B and C – percentages of positive and negative results for both readers combined

Sample IDN(Method A)Method AN(Method B/C)Method BMethod C
%Negative%Positive%Negative%Positive%Negative%Positive
S13010006060.040.015.085.0
S23033.366.76001000100
S33056.743.3605.095.00100
S43046.753.3608.391.70100
S53001006001000100
S6306.793.36001000100
S73001006001000100
S83001006001000100
S9303.396.76001000100
S103083.316.760100098.31.7
S113070.030.06010001000

For manual testing (Method C) fluorescence intensities differed at maximum ±1 level within one run and between the runs.

Conclusion

AESKUSLIDES nDNA (Crithidia luciliae)

510(k) Traditional Submission

All pre-determined acceptance criteria were met.

{16}------------------------------------------------

Between-Lab Precision AESKUSLIDES nDNA (Crithida luciliae)

To assess Between-Lab Precision of AESKUSLIDES nDNA (Crithidia luciliae) processed on HELIOS (Method A - HELIOS suggestions and Method B - Reader Confirmation of HELIOS images) and processed manually (Method C).

Sample Set

11 samples were assayed on AESKUSLIDES nDNA. The sample set includes borderline, low, medium, high positive, and negative samples. The samples are characterized in the following table:

Sample IDAESKUSLIDES nDNA
ResultGradingFluorescence Intensity (FI)
1Posborderline positive1
2Poslow positive2
3Posborderline positive1
4Posmedium positive3
5Posmedium positive3
6Posmedium positive3
7Poshigh positive4
8Poshigh positive4
9Poshigh positive4
10Negnegative0
11Negnegative0
Table 15: Sample table for Precision study

Procedure

11 samples have been tested on five days, two runs per day, three replicates per sample per run, on three different study sites. Two study sites were in the US, one study site was in Germany, with one HELIOS device per study site. Results were analyzed by two different readers per study site, resulting in a total of 90 data points per sample per reader.

This Between-lab precision study was done for three different Methods:

Method A (HELIOS)Processing of slides and image recording at HELIOS +HELIOS result suggestion
Method B (ReaderConfirmation)HELIOS images recorded in Method A + result analysisby two independent readers
Method C (Manual)Manual processing of slides and result analysis at themicroscope by two independent readers

All runs have been performed according to the respective IFUs. 1 positive and 1 negative control (kit controls) were included in each run. Samples were tested at 1:10 dilution. Results were analyzed by two independent readers.

{17}------------------------------------------------

Data Analysis

Positive/Negative Classification was recorded for each sample in each run and for each Method. Pos/Neg Classification in Method A (HELIOS) is provided by the nDNA Pattern Plus software tool.

For Method C (Manual performance) fluorescence intensity was also reported.

The following % Agreements including 95% CI (confidence intervals) will be calculated for each Method A, B and C:

  • Positive % Agreement (across all positive samples: number of correctly found samples divided through number of total positive samples)
  • -Negative % Agreement (across all negative samples: number of correctly found samples divided through number of total negative samples)
  • -Overall % Agreement (across all positive and negative samples: number of correctly found samples divided through number of total samples)
  • -Fluorescence Intensity % Agreement (across all samples, only for Method C: number of correctly found samples divided through number of total samples)

Results of the two readers will be calculated separately as well as combined.

Acceptance Criteria

  • -Positive sera have to be found positive, and negative sera have to be found negative.
  • Reported FI is allowed to differ max. ± 1 level from the expected value. -
  • -Agreements should meet the following criteria:

Table 16: Acceptance criteria for between-lab precision study

Type of Agreement (borderline positivesamples excluded*)Method AMethod BMethod C
Positive Agreement> 70%> 90%> 90%
Negative Agreement> 70%> 90%> 90%
Overall Agreement> 70%> 90%> 90%
Fluorescence Intensity Agreement> 90%
  • borderline samples are very low positive samples that can be evaluated also negative in certain cases, depending on the reader, the subjective manner of result analysis and normal test variances.

For Method A we accept a lower agreement (>70%), because we state in the IFU that all results generated by the HELIOS (Method A) must be confirmed by trained personnel.

From the results of those study, the following will be calculated:

  1. Between-Lab Precisions, 2. Between-Operator Precisions, 3. Single-Operator Precisions, and 4. Instrument Precision

"The e-Copy is an exact duplicate of the paper copy"

{18}------------------------------------------------

Results

Results are presented in the tables below. Agreements were calculated 1. including the results of the two borderline positive samples, 2. excluding the results of those two samples.

All predetermined acceptance criteria were met.

Between-Site Agreement

Manual (Method C) Between-Site Agreement (Calculation including borderline samples):

Table 17: Between-Site Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method C

Number ofSamplesSite 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement107191080106911108010602010801600201620
NegativeAgreement2391240240024023912403591360
OverallAgreement1310101320130911132012992113201959211980
FI Agreement1303171320128931132012863413201939411980

Table 18: Between-Site Agreement (Calculation including borderline samples) % Agreements (95% CI) Method C

% Agreement(95% CI)Site 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
PositiveAgreement99.2(98.4 - 99.6)99(98.2 - 99.4)98.1(97.2 - 98.8)98.8(98.1 - 99.2)
NegativeAgreement99.6(97.7 - 99.9)100(98.4 - 100)99.6(97.7 - 99.9)99.7(98.4 - 100)
OverallAgreement99.2(98.6 - 99.6)99.2(98.5 - 99.5)98.4(97.6 - 99)98.9(98.4 - 99.3)
FI Agreement98.7(97.9 - 99.2)97.7(96.7 - 98.3)97.4(96.4 - 98.2)97.9(97.2 - 98.5)

Manual (Method C) Between-Site Agreement (Calculation excluding borderline samples):

{19}------------------------------------------------


Table 19: Between-Site Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method C

Number ofSamplesSite 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement840084083468408346840125461260
NegativeAgreement2391240240024023912403591360
OverallAgreement107911080107461080107371080161371620
FI Agreement1063171080104931108010463410801579411620

Table 20: Between-Site Agreement (Calculation excluding borderline samples) % Agreements (95% Cl) Method C

% Agreement(95% CI)Site 1 vs Site2Site 2 vs Site3Site 1 vs Site3All Sites
PositiveAgreement100(99.5 - 100)99.3(98.5 - 99.7)99.3(98.5 - 99.7)99.5(99 - 99.8)
NegativeAgreement99.6(97.7 - 99.9)100(98.4 - 100)99.6(97.7 - 99.9)99.7(98.4 - 100)
OverallAgreement99.9(99.5 - 100)99.4(98.8 - 99.7)99.4(98.7 - 99.7)99.6(99.1 - 99.8)
FI Agreement98.4(97.5 - 99)97.1(96 - 98)96.9(95.6 - 97.7)97.5(96.6 - 98.1)

{20}------------------------------------------------

July 26, 201

Reader Confirmation (Method B) Between-Site Agreement (Calculation including borderline samples):

Table 21: Between-Site Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method B

Number ofSamplesSite 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement9958510809121681080909171108014082121620
NegativeAgreement2382240238224024002403582360
OverallAgreement1233871320115017013201149171132017662141980
FI Agreementnanananananananananana

Table 22: Between-Site Agreement (Calculation including borderline samples) % Agreements (95% Cl) Method B

% Agreement(95% CI)Site 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
PositiveAgreement92.1(90.4 - 93.6)84.4(82.2 - 86.5)84.2(81.9 - 86.2)86.9(85.2 - 88.5)
NegativeAgreement99.2(97 - 99.8)99.2(97 - 99.8)100(98.4 - 100)99.4(98 - 99.8)
OverallAgreement93.4(91.9 - 94.6)87.1(85.2 - 88.8)87(85.1 - 88.8)89.2(87.7 - 90.5)
FI Agreementnananana

{21}------------------------------------------------

Reader Confirmation (Method B) Between-Site Agreement (Calculation excluding borderline samples):

Table 23: Between-Site Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method B

Number ofSamplesSite 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement833784081822840815258401233271260
NegativeAgreement2382240238224024002403582360
OverallAgreement107191080105624108010552510801591291620
FI Agreementnananananananananananana

Table 24: Between-Site Agreement (Calculation excluding borderline samples) % Agreements (95% Cl) Method B

% Agreement(95% CI)Site 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
PositiveAgreement99.2(98.3 - 99.6)97.4(96.1 - 98.3)97(95.6 - 98)97.9(96.9 - 98.5)
NegativeAgreement99.2(97 - 99.8)99.2(97 - 99.8)100(98.4 - 100)99.4(98 - 99.8)
OverallAgreement99.2(98.4 - 99.6)97.8(96.7 - 98.5)97.7(96.6 - 98.4)98.2(97.4 - 98.8)
FI Agreementnananana

{22}------------------------------------------------

July 28, 2017

HELIOS (Method A) Between-Site Agreement (Calclation including borderline samples):

Table 25: Between-Site Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method A

Number ofSamplesSite 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement341199540282258540321219540472338810
NegativeAgreement1051512011461201051512016218180
OverallAgreement446214660396264660426234660634356990
FI Agreementnanananananananananana

Table 26: Between-Site Agreement (Calculation including borderline samples) % Agreements (95% Cl) Method A

% Agreement(95% CI)Site 1 vs Site2Site 2 vs Site3Site 1 vs Site3All Sites
PositiveAgreement63.1(59 - 67.1)52.2(48 - 56.4)59.4(55.3 - 63.5)58.3(54.8 - 61.6)
NegativeAgreement87.5(80.4 - 92.3)95(89.5 - 97.7)87.5(80.4 - 92.3)90(84.7 - 93.6)
OverallAgreement67.6(63.9 - 71)60(56.2 - 63.7)64.5(60.8 - 68.1)64(61 - 67)
FI Agreementnananana

{23}------------------------------------------------

HELIOS (Method A) Between-Site Agreement (Calclation excluding borderline samples):

Table 27: Between-Site Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method A

Site 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
Number ofSamplesFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement32892420278142420310110420458172630
NegativeAgreement1051512011461201051512016218180
OverallAgreement433107540392148540415125540620190810
FI Agreementnananananananananananana

Table 28: Between-Site Agreement (Calculation excluding borderline samples) % Agreements (95% Cl) Method A

% Agreement (95% CI)Site 1 vs Site 2Site 2 vs Site 3Site 1 vs Site 3All Sites
Positive Agreement78.1 (73.9 - 81.8)66.2 (61.5 - 70.5)73.8 (69.4 - 77.8)72.7 (69.1 - 76)
Negative Agreement87.5 (80.4 - 92.3)95 (89.5 - 97.7)87.5 (80.4 - 92.3)90 (84.7 - 93.6)
Overall Agreement80.2 (76.6 - 83.3)72.6 (68.7 - 76.2)76.9 (73.1 - 80.2)76.5 (73.5 - 79.3)
FI Agreementnananana

All acceptance critieria were met for Method C (Manual) and Method B (Reader Confirmation). All agreements were >90%. For Method A (HELIOS suggestion) acceptance criteria were met for negative and overall agreement for all study sites (>70%). Positive agreement for Method A was also >70% for site to site precision of site 1 vs site 2 and site 1 vs 3, but not for site 2 vs site 3, where a slightly lower agreement could be observed (66.2%). However, positive agreement met acceptance criteria when calculated across all three sites (72.7%).

Overall-Between-Lab agreements ranged from 72.6% to 80.2% for Method A (HELIOS), from 97.7% to 99.2% for Method B (Reader Confirmation), and from 96.9% to 98.4% for Method C (Manual).

{24}------------------------------------------------

Between-Operator Agreement

Manual (Method C) Between-Operator Agreement (Calculation including borderline samples):

Table 29: Between-Operator Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method C

Site 1Site 2Site 3
Number of SamplesFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected Numbers
Positive Agreement5319540540054052911540
Negative Agreement119112012001201200120
Overall Agreement65010660660066064911660
FI Agreement65010660653766063624660

Table 30: Between-Operator Agreement (Calculation including borderline samples) % Agreements (95% CI) Method C

% Agreement(95% CI)Site 1Site 2Site 3
PositiveAgreement98.3(96.9 - 99.1)100(99.3 - 100)98(96.4 - 98.9)
NegativeAgreement99.2(95.4 - 99.9)100(96.9 - 100)100(96.9 - 100)
OverallAgreement98.5(97.2 - 99.2)100(99.4 - 100)98.3(97 - 99.1)
FI Agreement98.5(97.2 - 99.2)98.9(97.8 - 99.5)96.4(94.6 - 97.5)

{25}------------------------------------------------

Manual (Method C) Between-Operator Agreement (Calculation excluding borderline samples):

Table 31: Between-Operator Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method C

Site 1Site 2Site 3
Number ofSamplesFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement420042042004204146420
NegativeAgreement119112012001201200120
OverallAgreement539154054005405346540
FI Agreement53010540533754051624540

Table 32: Between-Operator Agreement (Calculation excluding borderline samples) % Agreements (95% Cl) Method C

% Agreement(95% CI)Site 1Site 2Site 3
Positive Agreement100(99.1 - 100)100(99.1 - 100)98.6(96.9 - 99.3)
Negative Agreement99.2(95.4 - 99.9)100(96.9 - 100)100(96.9 - 100)
Overall Agreement99.8(99 - 100)100(99.3 - 100)98.9(97.6 - 99.5)
FI Agreement98.1(96.6 - 99)98.7(97.3 - 99.4)95.6(93.5 - 97)

{26}------------------------------------------------

borderline samples):

Reader Confirmation (Method B) Between-Operator Agreement (Calculation including

Table 33: Between-Operator Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method B

Number ofSamplesSite 1Site 2Site 3
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement4964454049941540413127540
NegativeAgreement120012011821201200120
OverallAgreement6164466061743660533127660
FI Agreementnanananananananana

Table 34: Between-Operator Agreement (Calculation including borderline samples) % Agreements (95% CI) Method B

% Agreement (95% CI)Site 1Site 2Site 3
Positive Agreement91.9(89.2 - 93.9)92.4(89.9 - 94.4)76.5(72.7 - 79.9)
Negative Agreement100(96.9 - 100)98.3(94.1 - 99.5)100(96.9 - 100)
Overall Agreement93.3(91.2 - 95)93.5(91.3 - 95.1)80.8(77.6 - 83.6)
FI Agreementnanana

{27}------------------------------------------------

Reader Confirmation (Method B) Between-Operator Agreement (Calculation excluding borderline samples):

Table 35: Between-Operator Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method B

Number ofSamplesSite 1Site 2Site 3
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement4155420418242040020420
NegativeAgreement120012011821201200120
OverallAgreement5355540536454052020540
FI Agreementnananananananana

Table 36: Between-Operator Agreement (Calculation excluding borderline samples) % Agreements (95% CI) Method B

% Agreement (95% CI)Site 1Site 2Site 3
Positive Agreement98.8(97.2 - 99.5)99.5(98.3 - 99.9)95.2(92.8 - 96.9)
Negative Agreement100(96.9 - 100)98.3(94.1 - 99.5)100(96.9 - 100)
Overall Agreement99.1(97.9 - 99.6)99.3(98.1 - 99.7)96.3(94.3 - 97.6)
FI Agreementnanana

Overall-Between-Operator agreements ranged from 96.3% to 99.3% for Method B (Reader Confirmation), and from 98.9% to 100% for Method C (Manual).

{28}------------------------------------------------

Single-Operator Agreement

Manual (Method C) Single-Operator Agreement (Calculation including borderline samples):

Table 37: Single-Operator Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method C

Site 1Site 2Site 3
Number of SamplesReader 1Reader 2Reader 3Reader 4Reader 5Reader 6
Found CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected NumbersFound CorrectlyFound IncorrectlyExpected Numbers
Positive Agreement266427026552702700270270027026642702637270
Negative Agreement591606006060060600606006060060
Overall Agreement325533032553303300330330033032643303237330
FI Agreement3291330321933032913303246330322833031416330

Table 38: Single-Operator Agreement (Calculation including borderline samples) % Agreements (95% CI) Method C

% Agreement(95% CI)Site 1Site 2Site 3
Reader 1Reader 2Reader 3Reader 4Reader 5Reader 6
PositiveAgreement98.5(96.3 - 99.4)98.1(95.7 - 99.2)100(98.6 - 100)100(98.6 - 100)98.5(96.3 - 99.4)97.4(94.7 - 98.7)
NegativeAgreement98.3(91.1 - 99.7)100(94 - 100)100(94 - 100)100(94 - 100)100(94 - 100)100(94 - 100)
OverallAgreement98.5(96.5 - 99.4)98.5(96.5 - 99.4)100(98.8 - 100)100(98.8 - 100)98.8(96.9 - 99.5)97.9(95.7 - 99)
FI Agreement99.7(98.3 - 99.9)97.3(94.9 - 98.6)99.7(98.3 - 99.9)98.2(96.1 - 99.2)97.6(95.3 - 98.8)95.2(92.3 - 97)

{29}------------------------------------------------

Manual (Method C) Single-Operator Agreement (Calculation excluding borderline samples):

Table 39: Single-Operator Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method C

Site 1Site 2Site 3
Number ofSamplesReader 1Reader 2Reader 3Reader 4Reader 5Reader 6
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement210021021002102091210
NegativeAgreement591606006060060
OverallAgreement269127027002702691270
FIAgreement269127026192702628270
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement210021021002102055210
NegativeAgreement600606006060060
OverallAgreement270027027002702655270
FIAgreement2691270264627025416270

Table 40: Single-Operator Agreement (Calculation excluding borderline samples) % Agreements (95% CI) Method C

% Agreement(95% CI)Site 1Site 2Site 3
Reader 1Reader 2Reader 3Reader 4Reader 5Reader 6
PositiveAgreement100(98.2 - 100)100(98.2 - 100)100(98.2 - 100)100(98.2 - 100)99.5(97.4 - 99.9)97.6(94.5 - 99)
NegativeAgreement98.3(91.1 - 99.7)100(94 - 100)100(94 - 100)100(94 - 100)100(94 - 100)100(94 - 100)
OverallAgreement99.6(97.9 - 99.9)100(98.6 - 100)100(98.6 - 100)100(98.6 - 100)99.6(97.9 - 99.9)98.1(95.7 - 99.2)
FI Agreement99.6(97.9 - 99.9)96.7(93.8 - 98.2)99.6(97.9 - 99.9)97.8(95.2 - 99)97(94.3 - 98.5)94.1(90.6 - 96.3)

{30}------------------------------------------------

July 28, 2017

Reader Confirmation (Method B) Single-Operator Agreement (Calculation including borderline samples):

Table 41: Single-Operator Agreement (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method B

Site 1Site 2Site 3
Number ofSamplesReader 1Reader 2Reader 3Reader 4Reader 5Reader 6
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement236342702601027025713270242282702096127020466270
NegativeAgreement600606006060060582606006060060
OverallAgreement296343303201033031713330300303302696133026466330
FIAgreementnananananananananananananananananana

Table 42: Single-Operator Agreement (Calculation including borderline samples) % Agreements (95% CI) Method B

% Agreement(95% CI)Site 1Site 2Site 3
Reader 1Reader 2Reader 3Reader 4Reader 5Reader 6
PositiveAgreement87.4(82.9 - 90.8)96.3(93.3 - 98)95.2(91.9 - 97.2)89.6(85.4 - 92.7)77.4(72.1 - 82)75.6(70.1 - 80.3)
NegativeAgreement100(94 - 100)100(94 - 100)100(94 - 100)96.7(88.6 - 99.1)100(94 - 100)100(94 - 100)
OverallAgreement89.7(85.9 - 92.5)97(94.5 - 98.3)96.1(93.4 - 97.7)90.9(87.3 - 93.6)81.5(77 - 85.3)80(75.4 - 84)
FI Agreementnananananana

{31}------------------------------------------------

July 28, 2017

Reader Confirmation (Method B) Single-Operator Agreement (Calculation excluding borderline samples):

Table 43: Single-Operator Agreement (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method B

Number ofSamplesSite 1Site 2Site 3
Reader 1Reader 2Reader 3Reader 4Reader 5Reader 6
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement2064210209121021002102082210203721019713210
NegativeAgreement600606006060060582606006060060
OverallAgreement2664270269127027002702664270263727025713270
FIAgreementnananananananananananananananananana

Table 44: Single-Operator Agreement (Calculation excluding borderline samples) % Agreements (95% CI) Method B

% Agreement(95% CI)Site 1Site 2Site 3
Reader 1Reader 2Reader 3Reader 4Reader 5Reader 6
PositiveAgreement98.199.51009996.793.8
(95.2 - 99.3)(97.4 - 99.9)(98.2 - 100)(96.6 - 99.7)(93.3 - 98.4)(89.7 - 96.3)
NegativeAgreement10010010096.7100100
(94 - 100)(94 - 100)(94 - 100)(88.6 - 99.1)(94 - 100)(94 - 100)
OverallAgreement98.599.610098.597.495.2
(96.3 - 99.4)(97.9 - 99.9)(98.6 - 100)(96.3 - 99.4)(94.7 - 98.7)(91.9 - 97.2)
Fl Agreementnananananana

Overall-Single-Operator agreements ranged from 95.2% to 100% for Method B (Reader Confirmation), and from 98.1% to 100% for Method C (Manual).

{32}------------------------------------------------

Instrument Precision

HELIOS (Method A) Instrument Agreement (Calculation including borderline samples):

Table 45: Instrument Precision (Calculation including borderline samples) Number of Correctly/Incorrectly found samples Method A

Number ofSamplesSite 1Site 2Site 3
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement19080270151119270131139270
NegativeAgreement4812605736057360
OverallAgreement23892330208122330188142330
FI Agreementnananananananana

Table 46: Instrument Precision (Calculation including borderline samples) % Agreements (95% Cl) Method A

% Agreement(95% CI)Site 1Site 2Site 3
PositiveAgreement70.4(64.7 - 75.5)55.9(50 - 61.7)48.5(42.6 - 54.5)
NegativeAgreement80(68.2 - 88.2)95(86.3 - 98.3)95(86.3 - 98.3)
OverallAgreement72.1(67 - 76.7)63(57.7 - 68.1)57(51.6 - 62.2)
FI Agreementnanana

{33}------------------------------------------------

HELIOS (Method A) Instrument Agreement (Calculation excluding borderline samples):

Number ofSamplesSite 1Site 2Site 3
FoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement180302101486221013080210
NegativeAgreement4812605736057360
OverallAgreement228422702056527018783270
FI Agreementnanananananananana

Table 47: Instrument Precision (Calculation excluding borderline samples) Number of Correctly/Incorrectly found samples Method A

Table 48: Instrument Precision (Calculation excluding borderline samples) % Agreements
(95% Cl) Method A
% Agreement(95% CI)Site 1Site 2Site 3
PositiveAgreement85.7(80.3 - 89.8)70.5(64 - 76.2)61.9(55.2 - 68.2)
NegativeAgreement80(68.2 - 88.2)95(86.3 - 98.3)95(86.3 - 98.3)
OverallAgreement84.4(79.6 - 88.3)75.9(70.5 - 80.6)69.3(63.5 - 74.5)
FI Agreementnanana

Acceptance criteria were met for site 1 and site 2. Positive, Neqative and Overall Agreements were >70%, At site 3, positive agreement was slightly below 70% which impacts also on the overall agreement at this site. Upon rechecking the respective HELIOS images we found that some of the images (approx. 11% of all images) that had been suggested false negative by HELIOS were out of focus. Therefore, the DNA Pattern Plus Software was not able to detect the sample as positive, although the readers still could find the sample as positive. This explains the observed lower positive agreement at site 3. This is also a reason why we clearly state that all results suggested by HELIOS have to be confirmed by a trained reader.

Overall, the three different HELIOS devices gave comparable results. Overall-Instrument agreements ranged from 69.3% to 84.4%.

Raw data see Attachment 12 Section 12 Raw data between lab Precision Study

{34}------------------------------------------------

Conclusions

Overall-Between-Lab agreements ranged from 72.6% to 80.2% for Method A (HELIOS), from 97.7% to 99.2% for Method B (Reader Confirmation), and from 96.9% to 98.4% for Method C (Manual).

Overall-Between-Operator aqreements ranged from 96.3% to 99.3% for Method B (Reader Confirmation), and from 98.9% to 100% for Method C (Manual).

Overall-Single-Operator agreements ranged from 95.2% to 100% for Method B (Reader Confirmation), and from 98.1% to 100% for Method C (Manual).

Overall-Instrument agreements ranged from 69.3% to 84.4% (Method A, HELIOS).

The results show that all results suggested by HELIOs have to be confirmed by a trained reader.

Lot to Lot Precision Study AESKUSLIDES nDNA (Crithidia luciliae)

A lot to lot reproducibility study was performed using three reagent lots each for AESKUSLIDES nDNA. 11 sera were tested:

The samples used for this study are the same as for Within Lab and Between Lab Precision Study AESKUSLIDES nDNA (Crithidia luciliae)

Each of the 11 serum samples was assayed 10 times on each reagent lot to give a total of 30 replicates per serum sample.

Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.

Agreements were calculated across all 3 reagent lots, for individual as well as for combined readers.

Acceptance Criteria:

  • -Positive sera have to be found positive, and negative sera have to be found negative.
  • -Correct pattern has to be found.
  • -Reported FI is allowed to differ max. ± 1 level from the expected value.
  • -Agreements should meet the following criteria:
Type of Agreement% Agreement
Positive Agreement> 90%
Negative Agreement> 90%
Overall Agreement> 90%
Fluorescence Intensity Agreement> 90%

{35}------------------------------------------------

Results:

Table 49: Lot to lot agreement (across all 3 lots) AESKUSLIDES nDNA – Number of Samples
-----------------------------------------------------------------------------------------------
Number ofReader 1Reader 2Combined Readers
SamplesFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbersFoundCorrectlyFoundIncorrectlyExpectedNumbers
PositiveAgreement270O27027002705400540
NegativeAgreement60O60600601200120
OverallAgreement330O33033003306600660
FAgreement330O33033003306600660

FI = Fluorescence Intensity

Table 50: Lot to lot agreement (across all 3 lots) AESKUSLIDES nDNA – % Agreement:
% Agreement(95% CI)Reader 1Reader 2CombinedReaders
Positive Agreement100(98.6 - 100)100(98.6 - 100)100(99.3 - 100)
Negative Agreement100(94 - 100)100(94 - 100)100(96.9 - 100)
Overall Agreement100(98.8 - 100)100(98.8 - 100)100(99.4 - 100)
FI Agreement100(98.8 - 100)100(98.8 - 100)100(99.4 - 100)

FI = Fluorescence Intensity

All agreements fulfilled the above criteria.

  • -Positive agreement was 100% for AESKUSLIDES nDNA for both readers.
  • -Negative agreement for nDNA was 100% for both readers.
  • Overall agreement for nDNA was 100% for both readers. -
  • FI agreement for nDNA was 100% for both readers. -
  • No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.

Raw data see Attachment 13_Section 12_ Raw data Lot to Lot Precision Study

Conclusion:

Overall lot to lot agreement (across all 3 lots) for AESKUSLIDES nDNA was 100%. All acceptance criteria have been fulfilled.

{36}------------------------------------------------

b) Carryover:

To investigate and evaluate the carry over of samples from one reaction well to the following reaction well when slides are processed by HELIOS.

Three high positive samples were run on HELIOS in alternate with a negative serum sample. Results were confirmed by a reader. Tests were performed in accordance with the respective IFUs.

SampleIDAESKUSLIDES nDNAPos/NegGrading
1Poshigh positive
2Poshigh positive
3Poshigh positive
4Negnegative

Acceptance criteria:

  • Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
  • FI is allowed to differ maximum + or 1 from the expected value. -

Results:

Table 51: Results Carry Over

Sample IDAESKUSLIDES nDNA
ResultFI
1Pos3
4Neg0
4Neg0
2Pos3
4Neg0
4Neg0
3Pos3
4Neg0
4Neg0

All samples fulfilled the above criteria. No carry over has been observed from well to well. All positive samples were identified as positive. All negative samples were identified as negative.

Conclusions

No carry over could be observed.

c) Time Extension Study

To investigate the stability of AESKUSLIDES nDNA performance upon shortening and elongation of the recommended incubation times for samples and conjugate.

Procedure:

AESKUSLIDES nDNA was assayed with 10 samples on HELIOS with the following incubation times:

{37}------------------------------------------------

    1. 15 min sera incubation + 15min incubation of conjugate
    1. 30 min sera incubation + 30min incubation of conjugate
    1. 45 min sera incubation + 45min incubation of conjugate

The following samples were used:

Sample IDExpected Results
ResultGrading
Sample 1Poslow positive
Sample 2Posmedium positive
Sample 3Poslow positive
Sample 4Poslow positive
Sample 5Posmedium positive
Sample 6Posmedium positive
Sample 7Poshigh positive
Sample 8Poshigh positive
Sample 9Negnegative
Sample 10NegNeg

Tests were performed in accordance with the respective IFUs. Results have been confirmed by a trained reader.

Results of the 30 min protocol will be compared to the results of the 15 min and 45 min protocol.

Acceptance criteria:

  • Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
  • -FI is allowed to differ maximum + or - 1 from the expected value.

Results

IncubationTimes2 x 15 Minutes2 x 30 Minutes2 x 45 Minutes
Sample IDResultFluorescenceIntensityResultFluorescenceIntensityResultFluorescenceIntensity
PCPos3Pos3Pos3
NCNegNegNegNegNegNeg
Sample 1Pos1Pos1Pos1
Sample 2Pos2Pos2Pos2
Sample 3Pos1Pos1Pos2
Sample 4Pos1Pos1Pos2
Sample 5Pos2Pos2Pos2
Sample 6Pos2Pos2Pos3
Sample 7Pos3Pos3Pos3
Sample 8Pos3Pos3Pos3
Sample 9NegNegNegNegNegNeg
Sample 10NegNegNegNegNegNeg

Table 52: Results Time Extention

{38}------------------------------------------------

All acceptance criteria were fulfilled in each assay. All positive samples were identified as positive. All negative samples were identified as negative. Fluorescence intensities did not deviate more than +/-1 level.

Conclusions

The results demonstrate stability of the AESKUSLIDES nDNA during the tested incubation times of conjugate and sample.

d) Detection limit

Not applicable

e) Analytical specificity

Interfering substances:

The interference study was performed according to CLSI EP07-A2. Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Interference by 8 different substances was assessed by testing 8 serum samples:

Sample IDAESKUSLIDES nDNA
Pos/NegGrading
1Poslow positive
2Poslow positive
3Posmedium positive
4Posmedium positive
5Poshigh positive
6Poshigh positive
7Negnegative
8Negnegative

The following substances have been tested with the indicated concentrations:

Table 53: Interfering substances with the indicated concentrations

Interfering SubstanceMinimum finalConcentrationtestedMaximum finalConcentrationtested
Bilirubin conjugated0.1 mg/mL0.4 mg/mL
Bilirubin unconjugated0.1 mg/mL0.4 mg/mL
Hemoglobin2.5 mg/mL5.0 mg/mL
Triglycerides5 mg/mL20 mg/mL
RF IgM200 U/mL400 U/mL
Rituximab0.5 mg/mL2.0 mg/mL
Methylprednisolone0.2 mg/mL0.8 mg/mL
Cyclophosphamide1.0 mg/mL4.0 mg/mL
Methotrexate0.025 mg/mL0.1 mg/mL
Azathioprine0.0075 mg/mL0.03 mg/mL
Belimumab2 mg/mL8 mg/mL
Hydroxychloroquine0.006 mg/mL0.024 mg/mL
Mycophenolat0.012 mg/mL0.048 mg/mL
Ibuprofen0.5 mg/mL2 mg/mL
Naproxen0.5 mg/mL2 mg/mL

"The e-Copy is an exact duplicate of the paper copy"

{39}------------------------------------------------

Interfering substances were spiked in each of the 8 samples in two different concentrations. Controls were prepared for each serum sample by spiking in only the respective amount of diluent without interfering substances. Spiked serum samples and their controls have been tested in triplicates for each concentration of interferent with AESKUSLIDES nDNA. All tests have been performed manually according to the IFU. Results have been analyzed by two independent readers at the microscope.

The results (positive/negative classification. fluorescence intensity (FI)) of the spiked samples were compared to the results of the respective controls. Agreements have been calculated by dividing the number of correctly found results through the number of total results for spiked samples.

Acceptance criteria:

  • -Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
  • -Overall Agreement: > 90%.
  • FI is allowed to differ maximum + or 1 from the expected value. -
  • FI Agreement: > 90%. -

Results

All samples fulfilled the above criteria. No significant differences were observed between sera spiked with different concentrations of the interferents and their controls:

  • Positive Agreement was 100% for all tested substances and concentrations. except for 20 mg/ml Triglycerides with Reader 1 (97%). There. one out of three repetitions of sample 2 was found negative instead of low positive (FI 1). Reader 2 found the sample correctly. Therefore. we consider this single discrepancy as a result of normal test variances and the subjective nature of the result analysis.
  • Negative Agreement was 100% for all tested interferents and concentrations.
  • Overall Agreement ranged from 97.9% to 100%. It was 100% for all tested substances and concentrations. except for 20 mg/ml Triglycerides with Reader 1 (97.9%).
  • -FI Agreement was 100% for all tested interferents and concentrations.

Raw Data see Attachment 14_Section 12_Raw data Interfering Substances

Conclusions

All acceptance criteria are fulfilled in each test. No interference was detected with the tested substances. up to the maximal concentrations indicated above.

{40}------------------------------------------------

f) Assay cut-off:

AESKU recommends a screening dilution of 1:10. followed by serial dilutions for semi-quantitative determinations. but suggests each laboratory establish its own screening dilution and titration scheme based on its population. The titers of 1:10 are considered low titers. 1:20 and 1:40 are considered medium titers, and 1:80 and greater are considered high titers.

2. Comparison studies

a. Method comparison with predicate device:

To show comparable performance of AESKUSLIDES nDNA and the predicate assay NOVA Lite dsDNA (Crithidia luciliae) of INOVA.

Procedure:

A clinical study with the same 776 serum samples (comprising 297 serum samples from patients with SLE and 479 samples from patients with other diseases) was performed with AESKUSLIDES nDNA and the predicate assay. A detailed description of the study sample can be found on page 12-50.

For each assay the sample set was processed manually and analyzed (positive/negative classification) at the microscope (Method C).

All assays were performed according to the respective IFUs. Sample were screened at a dilution of 1:10.

From the results of both assays the relative sensitivity and specificity / % positive, negative and overall agreements were calculated.

Furthermore, diagnostic sensitivity and specificity was calculated for both assays, and the cross reactivities in different diagnosis will be shown in a table.

Results:

Table 54: Method comparison with predicate device Diagnostic Sensitivity and Specificity Number of positive and negative results and % agreements:

Diagnostic Sensitivity &SpecificityDiagnosis
SLEOther DiseasesTotal
AESKUSLIDESnDNAPos732598
Neg224454678
Total297479776
Predicate AssayPos11068178
Neg187411598
Total297479776

{41}------------------------------------------------

% DiagnosticSensitivity &Specificity (95%CI)% Sensitivity(95% CI)% Specificity(95% CI)Predictive Valueof a Positive TestResult (PPV)Predictive Valueof a NegativeTest Result(NPV)
SLE (n= 297)OD (n=479)
AESKUSLIDESnDNA24.6(20 - 29.8)94.8(92.4 - 96.4)74.5%67.0%
Predicate Assay37(31.7 - 42.7)85.8(82.4 - 88.6)61.8%68.7%

Sensitivity of AESKUSLIDES nDNA is a little bit lower (24%.6) than that calculated for the predicate assay (37%) in this study. On the other hand, specificity of AESKU is higher (94.8%) than that of the predicate (85%.8). Those values for sensitivity correlate well with values reported in the literature, where sensitivities of nDNA IFA assays for SLE are described from 13% up to 67%. Anti dsDNA antibodies are associated with disease activity and also have been shown to disappear upon treatment. The SLE cohort tested in this study was not further selected for inactive or active patients, or for treatment naïve patients. Treatment naïve patients are very rare, therefore we expect to have a greater proportion of patients under treatment and also inactive patients in the cohort. This explains why in this study the sensitivities are at the lower end of the reported sensitivities in the literature.

The higher specificity and lower sensitivity of AESKUSLIDES nDNA goes along with a higher PPV (74.5% compared to 61.8% for the predicate). NPVs of both assays are comparable (67% vs 68.7%).

The results for both assays are shown in detail in the crossreactivity table below. With AESKUSLIDES nDNA fewer crossereactivities are seen. Crossreactivities could be detected in APS (8.3%), in Dermato-/Polymyositis (16.1%), RA (17.2%), MCTD (12.5%), AIH (9.5%), PBC (11.1%), and Vasculitis (4.5%). The predicate assay detected positives in almost all diagnosis tested and also higher numbers of positives.

{42}------------------------------------------------

Image /page/42/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font size and lighter shade of gray.

CONFIDENTIAL

AESKUSLIDES nDNA (Crithidia luciliae) 510(k) Traditional Submission

July 28, 2017

DiagnosisnAESKUSLIDES nDNAPredicate
n Pos%Posn Neg%Negn Pos%Posn Neg%Neg
Target diseaseSystemic Lupus Erythematosus2977324.6%22475.4%11037.0%18763.0%
Antiphospholipid Syndrome1218.3%1191.7%433.3%866.7%
Sjoegren's Syndrome3000.0%30100.0%516.7%2583.3%
Systemic Sclerosis4400.0%44100.0%511.4%3988.6%
Other RheumaticDiseasesDermatomyositis/Polymyositis31516.1%2683.9%39.7%2890.3%
Rheumatoid Arthritis581017.2%4882.8%1525.9%4374.1%
Mixed Connective Tissue Diseases24312.5%2187.5%520.8%1979.2%
Undifferentiated Connective Tissue Diseases2100.0%21100.0%00.0%21100.0%
ControldiseasesAutoimmune LiverDiseasesAutoimmune Hepatitis2129.5%1990.5%733.3%1466.7%
Primary Biliary Cirrhosis20211.1%1888.9%544.4%1555.6%
Primary Sclerosing Cholangitis1000.0%10100.0%00.0%10100.0%
VasculitisANCA associated Vasculitis4424.5%4295.5%12.3%4397.7%
Epstein-Barr Virus Infection2500.0%25100.0%416.0%2184.0%
InfectiousDiseasesHepatitis B Virus Infection3400.0%34100.0%411.8%3088.2%
Hepatitis C Virus Infection4400.0%44100.0%24.5%4295.5%
LeukemiaLymphoma3100.0%31100.0%722.6%2477.4%
OtherFibromyalgia3000.0%30100.0%13.3%2996.7%
Total7769812.6%67887.4%17822.9%59877.1%

Table 55: Method comparison with predicate device Number and percent positive samples in the different tested diagnosis

{43}------------------------------------------------

Image /page/43/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.

Table 56: Method comparison with predicate device, Relative sensitivity and specificity Number of positive and negative results and % agreements:

Method ComparisonPredicate
PosNegTotal
AESKUSLIDESnDNAPos801393
Neg98585683
Total178598776
Method ComparisonPredicate vs AESKU% Agreement(95% CI)
Positive Agreement44.9(37.8 - 52.3)
Negative Agreement97.8(96.3 - 98.7)
Overall Agreement85.7(83.1 - 88)

Overall Agreement between both assays was good (85.7%), and negative agreement was very good with 97.8%. The rather low positive agreement of 44.9% is due to the fact that the predicate assay detects more positive results.

For the interpretation of the discrepant results we also tested the respective samples by ds DNA ELISA (IgG). The results are shown in the following tables.

Table 57: Method comparison with predicate device, Discrepant samples in the Non-Healthy Control group:

SampleIDDiagnosisPredicateAESKUSLIDESnDNAAESKULISAdsDNA-G
ResultFIResultFIU/ml
688DM/PMNeg0Pos34,9
690DM/PMNeg0Pos37,4
691DM/PMNeg0Pos28,7
67RANeg0Pos11,8
694RANeg0Pos14,3
596VasculitisNeg0Pos10,7
632APSPos1Neg00,9
758APSPos1Neg028,3
759APSPos1Neg03,1
243AIHPos1Neg04,3
246AIHPos1Neg03,1
322AIHPos1Neg09,9
361AIHPos1Neg03,4
234PBCPos1Neg06,5
237PBCPos1Neg046,6
240PBCPos1Neg04,8
209EBVPos1Neg03,2
212EBVPos1Neg05,3
220EBVPos1Neg012,2
221EBVPos1Neg0<Min
542FibromyalgiaPos1Neg017,2
331HBVPos1Neg013,2
338HBVPos1Neg02,1
339HBVPos1Neg08,4
341HBVPos1Neg08,3
78HCVPos1Neg05,9
79HCVPos1Neg01,1
114LeukemiaPos1Neg01,5
116LeukemiaPos1Neg02,2
119LeukemiaPos1Neg02,8
121LeukemiaPos1Neg00,7
124LeukemiaPos1Neg01,3
127LeukemiaPos1Neg01,6
129LeukemiaPos1Neg01,5
770MCTDPos1Neg02,2
771MCTDPos1Neg01,3
767DM/PMPos1Neg04,2
178RAPos1Neg013,5
180RAPos1Neg05,6
185RAPos1Neg01,7
188RAPos1Neg01,3
191RAPos3Neg04,6
193RAPos1Neg02,1
698RAPos1Neg012,5
271Sjoegren'sPos1Neg06,3
281Sjoegren'sPos1Neg012,5
397Sjoegren'sPos1Neg07,7
422Sjoegren'sPos1Neg01,6
423Sjoegren'sPos1Neg03,4
393SScPos1Neg03,6
400SScPos1Neg0109,7
402SScPos1Neg01,9
579SScPos1Neg071,3
587SScPos1Neg06,2
609VasculitisPos1Neg00,9

{44}------------------------------------------------

Image /page/44/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray.

July 28, 2017

"The e-Copy is an exact duplicate of the paper copy"

{45}------------------------------------------------

July 28, 2017

55 of 97 discrepant samples have been found in the Non-Healthy Control group. Of those 55 samples, 6 samples are reported positive with the AESKU assay, and negative with the predicate assay. The positive results could not be confirmed by ELISA. Therefore, we think, that for those 6 samples the predicate outcompetes AESKU.

49 of those 55 samples were reported positive with the predicate assay and negative with the AESKU assay. Of these 49 positive samples only 4 could be confirmed by ELISA. Furthermore, all positive results had a low intensity (Fl=1) in the predicate assay. Therefore, we think that for this 49 samples, AESKU outcompeted the predicate, by being more specific than the predicate.

Table 58: Method comparison with predicate device, Discrepant samples in the SLE group.

SampleIDDiagnosisPredicateAESKUSLIDESnDNAAESKULISAdsDNA-G
ResultFIResultFIU/ml
28SLENeg0Pos26,5
56SLENeg0Pos13,6
22SLEPos1Neg06,1
27SLEPos1Neg02,9
32SLEPos1Neg015,3
138SLEPos1Neg01,4
144SLEPos1Neg0108,4
148SLEPos1Neg06,5
150SLEPos1Neg00,6
151SLEPos1Neg02,0
153SLEPos1Neg04,6
154SLEPos1Neg02,1
155SLEPos2Neg04,3
156SLEPos3Neg0296,9
161SLEPos1Neg0180,4
164SLEPos1Neg0>Max
200SLEPos1Neg01,9
201SLEPos1Neg02,4
207SLEPos1Neg020,3
317SLEPos1Neg028,8
323SLEPos1Neg08,5
324SLEPos1Neg062,3
328SLEPos1Neg03,9
371SLEPos1Neg04,8
381SLEPos1Neg0301,0
383SLEPos1Neg02,1
386SLEPos1Neg014,0
432SLEPos2Neg011,3
433SLEPos1Neg013,1
434SLEPos1Neg04,6
447SLEPos1Neg01,8
451SLEPos1Neg052,5
456SLEPos1Neg01,8
461SLEPos1Neg06,2
466SLEPos1Neg057,0
467SLEPos1Neg04,3
482SLEPos1Neg061,5
499SLEPos1Neg07,1
505SLEPos1Neg08,3
513SLEPos1Neg03,6
514SLEPos1Neg014,9
672SLEPos2Neg084,7

{46}------------------------------------------------

Image /page/46/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white medical symbol inside. To the right of the square is the company name, "AESKU.DIAGNOSTICS," in gray text. Below the company name is the tagline, "THE DIAGNOSTIC TOOL THAT WORKS," in a lighter gray font.

July 28, 2017

{47}------------------------------------------------

42 of the 97 of the discrepant samples were found in the SLE cohort. Of these 42 samples, 2 have been reported positive by the AESKU assay and negative by the predicate. However, the positive results could not be confirmed by ELISA.

40 of these 42 samples have been reported positive by the predicate assay and negative by the AESKU assay. Almost all positive samples had a low fluorescence intensity (FI=1). Only 12 samples of those 40 could be confirmed positive by ELISA. Therefore, we think that the predicate outcompetes AESKU in those 12 cases. This is only 4% of the total SLE samples. 28 of the 40 samples could not be confirmed by ELISA which is generally more sensitive than the nDNA immunofluorescence assays.

Overall, the predicate assay detected more positives in the SLE cohort, but most of those results could not be confirmed by ELISA. This matches the observation that INOVA has a lower specificity than AESKU.

Generally, Crithidia luciliae immunofluorescence tests provided by different manufactures on the market are known to deliver different results. This could be explained by differences in the manufacturing process, e.g. cultivation medium, cultivation period, and the fixation method.

Conclusions

AESKUSLIDES nDNA showed a lower diagnostic sensitivity (24.6%) compared to the predicate assay NOVA Lite provided by INOVA (37%). However, the AESKU assay is characterized by a higher specificity (94.8% vs. 85.8%), because of less crossreactivities with other disease groups.

Raw data see Attachment 15_Section 12_Raw data Method comparison with predicate device

b. Method Comparison of Method A, B, and C and clinical study To show comparable performance of AESKUSLIDES nDNA when processed manually and on HELIOS.

Procedure:

A clinical study with the same 776 serum samples (comprising 297 serum samples from patients with SLE and 479 samples from patients with other diseases) was performed on 3 different study sites (2 US sites, one German site). A detailed description of the study sample can be found on page 12-50.

At each study site the sample set was processed

    1. manually and analyzed (positive/negative classification) at the microscope by two independent readers (Method C)
    1. on HELIOS and HELIOS images were analyzed by two independent readers at the PC (Method B)

"The e-Copy is an exact duplicate of the paper copy"

{48}------------------------------------------------

    1. on HELIOS and positive/negative classification was done by the HELIOS DNA Pattern Plus Software (Method A)
      All assays were performed according to the IFU.

For each study site, the results of the single and combined readers with the different Methods were compared to each other and % agreements were calculated:

  • Method C (Manual) vs Method B (Reader Confirmation)
  • Method B (Reader Confirmation) vs Method A (HELIOS)
  • Method C (Manual) vs Method A (HELIOS).

Furthermore, the number of positive samples found for each tested diagnosis for all study sites and all Methods will be compared as well as the calculated diagnostic sensitivities and specificities.

Acceptance Criteria:

Positive, Negative, and Overall Agreements between the different Methods should be >85%. The analysis of diagnostic immunofluorescence assays and especially of nDNA assays depends strongly on the readers experience and is very subjective. Therefore, we expect and accept variances up to 15% in this clinical study.

Results:

% Agreement(95% CI)Method C vs BMethod B vs AMethod C vs A
Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2
Site1PositiveAgreement85.1(76.1 - 91.1)85.5(75.9 - 91.7)51.8(41.3 - 62.1)50.6(40 - 61.2)48.3(38.1 - 58.6)51.3(40.3 - 62.2)
NegativeAgreement98.4(97.2 - 99.1)97.7(96.3 - 98.6)84.4(81.5 - 86.9)84(81.1 - 86.6)84(81.1 - 86.6)83.9(80.9 - 86.4)
TotalAgreement96.9(95.4 - 97.9)96.5(95 - 97.6)80.8(77.9 - 83.4)80.5(77.6 - 83.2)80(77.1 - 82.7)80.7(77.7 - 83.3)
Site2PositiveAgreement91.3(83 - 95.7)90.7(82 - 95.4)54.8(44.1 - 65)56.1(45.3 - 66.3)53.8(42.9 - 64.3)54.7(43.4 - 65.4)
NegativeAgreement98.4(97.2 - 99.1)98(96.7 - 98.8)86.3(83.5 - 88.6)86.3(83.6 - 88.7)85.9(83.1 - 88.3)85.7(83 - 88.1)
TotalAgreement97.7(96.4 - 98.5)97.3(95.9 - 98.2)82.9(80 - 85.3)83.1(80.3 - 85.6)82.6(79.8 - 85.1)82.7(79.9 - 85.2)
Site3PositiveAgreement85.5(75.9 - 91.7)87(77 - 93)48(37.1 - 59.1)47.3(36.3 - 58.5)43.4(32.9 - 54.6)46.4(35.1 - 58)
NegativeAgreement98.6(97.4 - 99.2)98(96.7 - 98.8)90.6(88.2 - 92.5)90.5(88.1 - 92.4)90.1(87.7 - 92.1)90.1(87.7 - 92.1)
TotalAgreement97.3(95.9 - 98.2)97(95.6 - 98)86.5(83.9 - 88.7)86.3(83.7 - 88.6)85.6(82.9 - 87.9)86.2(83.6 - 88.5)

Table 59: Method Comparison of Method A, B and C - Single Reader Agreements

{49}------------------------------------------------

Image /page/49/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The image is a professional and modern design.

Table 60: Method Comparison of Method A, B and C – Combined Reader Agreements

% Agreement (95%CI)Method C vsBMethod B vsAMethod C vsA
Site 1PositiveAgreement85.3(79 - 89.9)51.2(43.7 - 58.7)49.7(42.1 - 57.3)
NegativeAgreement98.1(97.2 - 98.7)84.2(82.2 - 86)83.9(81.9 - 85.8)
TotalAgreement96.7(95.7 - 97.5)80.7(78.6 - 82.6)80.3(78.3 - 82.2)
Site 2PositiveAgreement91(85.4 - 94.5)55.4(47.8 - 62.8)54.2(46.3 - 61.8)
NegativeAgreement98.2(97.4 - 98.8)86.3(84.4 - 88)85.8(83.9 - 87.6)
TotalAgreement97.5(96.6 - 98.2)83(81 - 84.8)82.7(80.7 - 84.5)
Site 3PositiveAgreement86.2(79.7 - 90.9)47.7(39.8 - 55.6)44.8(37 - 53)
NegativeAgreement98.3(97.5 - 98.9)90.5(88.9 - 91.9)90.1(88.5 - 91.6)
TotalAgreement97.2(96.2 - 97.9)86.4(84.6 - 88)85.9(84.1 - 87.5)

{50}------------------------------------------------

Image /page/50/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo appears to be for a company that specializes in diagnostic tools.

July 28, 2017

Table 61: Method comparison Number and percent positive samples in the different tested diagnosis

Cross ReactivitySite 1Site 2Site 3
Manual PerformanceReader Confirmation at HELIOSHELIOS suggestionManual PerformanceReader Confirmation at HELIOSHELIOS suggestionManual PerformanceReader Confirmation at HELIOSHELIOS suggestion
DiagnosisnReader 1n Pos% PosReader 2n Pos% PosReader 1n Pos% PosReader 2n Pos% Posn Pos% PosReader 3n Pos% PosReader 4n Pos% PosReader 3n Pos% PosReader 4n Pos% Posn Pos% PosReader 5n Pos% PosReader 6n Pos% PosReader 5n Pos% PosReader 6n Pos% Posn Pos% Pos
SLE2976220.95919.96321.26020.26120.56722.66220.96321.26722.66521.96521.95618.95719.25719.24515.2
APS1200.000.018.300.0541.700.000.018.300.0541.700.000.000.000.000.0
Sjögren3000.000.013.313.31033.300.000.013.300.0413.300.000.013.313.3413.3
SSc4412.300.000.000.0613.600.000.000.000.0715.900.000.012.312.3715.9
Myositis3113.213.213.213.2619.413.213.213.226.5412.900.000.000.000.013.2
RA58813.8712.1712.1813.81525.9610.3610.3712.1610.31220.758.6813.8712.1610.3712.1
MCTD2428.328.3312.528.31041.728.328.3312.528.3833.328.314.228.328.3416.7
UCTD2100.000.000.000.029.500.000.000.000.0314.300.000.000.000.029.5
AIH2129.529.529.529.5523.829.529.529.514.8314.314.800.029.529.5314.3
AIH/PBC900.000.0111.100.0222.200.000.0111.1111.1111.100.0111.100.0111.1111.1
PBC11327.300.0327.3327.3436.400.000.0327.319.1436.400.000.000.000.019.1
PSC1000.000.000.000.000.000.000.000.000.0330.000.000.000.000.000.0
Vasculitis44511.4511.424.524.51125.012.312.324.512.349.124.512.324.524.5613.6
EBV2500.000.000.000.0312.000.000.000.000.014.000.000.000.000.0312.0
HBV3425.900.012.912.912.912.912.912.912.9411.812.900.025.912.9514.7
HCV4400.000.000.000.036.800.000.000.000.0511.400.000.000.000.024.5
Fibromyalgia3000.000.000.000.0310.000.000.000.000.000.000.013.300.000.0413.3
Leukemia3113.200.000.013.2516.100.000.000.000.0825.800.013.213.213.2722.6

{51}------------------------------------------------

% Diagnostic Sensitivity &Specificity (95% CI)% Sensitivity(95% CI)% Specificity(95% CI)% Sensitivity(95% CI)% Specificity(95% CI)
SLE (n= 297)OD (n=479)SLE (n= 297)OD (n=479)
ManualReader 120.9(16.6 - 25.9)94.8(92.4 - 96.4)CombinedReaders20.4(17.3 - 23.8)95.6(94.1 - 96.7)
Reader 219.9(15.7 - 24.8)96.5(94.4 - 97.8)
Site 1ReaderConfirmationat HELIOSReader 121.2(16.9 - 26.2)95.4(93.1 - 96.9)CombinedReaders20.7(17.6 - 24.1)95.5(94 - 96.7)
Reader 220.2(16 - 25.1)95.6(93.4 - 97.1)
HELIOS20.5(16.3 - 25.5)81(77.2 - 84.3)
Site 2ManualReader 322.6(18.2 - 27.6)97.3(95.4 - 98.4)CombinedReaders21.7(18.6 - 25.2)97.3(96.1 - 98.1)
Reader 420.9(16.6 - 25.9)97.3(95.4 - 98.4)
ReaderConfirmationat HELIOSReader 321.2(16.9 - 26.2)95.4(93.1 - 96.9)CombinedReaders21.9(18.7 - 25.4)96.1(94.7 - 97.2)
Reader 422.6(18.2 - 27.6)96.9(94.9 - 98.1)
HELIOS21.9(17.6 - 26.9)84.1(80.6 - 87.1)
Site 3ManualReader 521.9(17.6 - 26.9)97.7(95.9 - 98.7)CombinedReaders20.4(17.3 - 23.8)97.5(96.3 - 98.3)
Reader 618.9(14.8 - 23.7)97.3(95.4 - 98.4)
ReaderConfirmationat HELIOSReader 519.2(15.1 - 24.1)97.3(95.4 - 98.4)CombinedReaders19.2(16.2 - 22.6)96.9(95.6 - 97.8)
Reader 619.2(15.1 - 24.1)96.5(94.4 - 97.8)
HELIOS15.2(11.5 - 19.7)88.1(84.9 - 90.7)

Table 62: Diagnostic Sensitivity and Specificity

Summary of Results:

Agreements for Method Comparison between Method C (Manual) and Method B (Reader Confirmation of HELIOS images) were > 85% and acceptance criteria were met.

Positive Agreement of Method C vs B ranged from 85.3% to 91%.

Negative Agreement of Method C vs B ranged 98.1% to 98.3%.

Overall Agreement of Method C vs B ranged from 96.7% to 97.5%.

The results show that Methods C (Manual) and B (Reader Confirmation of HELIOS images) give equivalent results. Agreements were > 85%. Furthermore, result of the 3 study site were comparable with each other. We accept an agreement of 85% for AESKUSLIDES nDNA in this study, since the analysis of nDNA slides is very subjective and strongly depends on the experience of the reader.

{52}------------------------------------------------

Image /page/52/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.

Agreements for Comparison between Method B (Reader Confirmation of HELIOS images) and Method A (HELIOS positive/negative classification by the DNA Pattern Plus software) were as follows:

Positive Agreement of Method B vs A ranged from 47.7% to 55.4%.

Negative Agreement of Method B vs A ranged from 84.2% to 90.5%.

Overall Agreement of Method B vs A ranged from 80.3% to 85.9%.

According to this acceptance criteria are not met for comparison of Method B vs A in this study. The same holds true for the comparison of Method C (Manual) vs A (HELIOS suggested results).

Positive Agreement of Method C vs A ranged from 44.8% to 54.2%.

Negative Agreement of Method C vs A ranged from 83.9% to 90.1%.

Overall Agreement of Method C vs A ranged from 80.3% to 85.9%.

The reason why acceptance criteria were not met for comparison of Methods B vs A and C vs A is due to the fact that Method A (HELIOS DNA Pattern Plus software) detected only around 50% of the positive samples that had been detected by the readers at the microscope and at the screen.

However, if one takes into account the results of the calculated sensitivities and specificities for the different study sites and methods it can be seen, that the results for sensitivities are very similar between the different methods and study sites:

At site 1, sensitivities ranged from 20.4% to 20.7%.

At site 2. sensitivities ranged from 21.7% to 21.9%.

At site 2, sensitivities ranged from 15.2% to 20.4%.

All sensitivities were around 20% except for site 3, Method A. where only 15.2% were achieved.

Overall these results show, that Method A, B, and C give comparable results for the diagnostic sensitivity. The reason why the positive agreement between Method B vs A and Method C vs A is only around 50% is that in total, the DNA Pattern Plus software detected a similar number of positive samples in the SLE group, but 50% of those were different samples than those detected by the readers in Method B and C.

Furthermore, the calculated specificities for Method C and B were >95% for all three study sites. The specificity for Method A was slightly lower, it ranged from 81% to 88.1%. This is due to the fact that the DNA Pattern Plus Software detected more "false positives" in the non-healthy control group.

The described results are also reflected by the Cross Reactivity Table. Overall, for Method C and B there were only few crossreactivities which is also reflected by the high specificity of this methods (>95%). The most crossreactivities could be seen in RA (up to

"The e-Copy is an exact duplicate of the paper copy"

{53}------------------------------------------------

Image /page/53/Picture/2 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo is clean and professional, with a focus on the company's name and its diagnostic tools.

13.8%), MCTD (up to 12.5%), in the Autoimmune Liver Disease Group (AIH, PBC, and PCS: up to 11.8%), and in Vasculitis (up to 11.4%). These results reflect values reported in the literature. Method A has a lower specificity as described and therefore, higher crossreactivities can be seen.

Conclusion:

Method C and Method B show substantial equivalence. AESKUSLIDES nDNA performs substantially equivalent when processed manually and on HELIOS with Reader Confirmation of HELIOS images, as shown by the method comparison study, the calculated diagnostic sensitivities and specificities, and the cross reactivity table.

For Method A in summary, when single samples were compared to each other like has been done in the method comparison described, we found a lower agreement of Method A, when compared to Method B and C. But when diagnostic sensitivities are compared, results are comparable between all three Methods. It has been shown that with Method A additional samples could be detected in the SLE group, compared to Method B and C, with a slightly lower specificity.

Our results show that results generated at HELIOS must always be confirmed by trained readers. The DNA Pattern Plus software can support the reader by providing additional information, and might encourage to recheck further samples.

Raw data see Attachment 16 Section 12 Raw data Method comparison of Method A, B, and C and Clinical Study

c. Endpoint Titer Comparison AESKUSLIDES nDNA (Crithidia luciliae)

To show that titration of samples results in the same endpoint titer when assayed manually and on HELIOS.

Procedure:

Five serum samples were assayed in duplicates and analyzed on HELIOS (Method B, Reader Confirmation) and manually at the microscope (Method C) by two independent readers:

Sample IDPos/NegExpected EndpointTiter
1Pos1280
2Pos640
3Pos40
4Pos320
5Pos80

{54}------------------------------------------------

Image /page/54/Picture/1 description: The image contains text that describes a submission related to AESKUSLIDES nDNA (Crithidia luciliae). The text indicates that it is a 510(k) Traditional Submission. The text is written in a clear, sans-serif font and is horizontally aligned.

Image /page/54/Picture/2 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, resembling a stylized "Y" with horizontal lines. To the right of the square, the text "AESKU.DIAGNOSTICS" is displayed in gray, with the word "AESKU" being larger and bolder than "DIAGNOSTICS". Below the company name, the tagline "THE DIAGNOSTIC TOOL THAT WORKS" is written in a smaller, lighter gray font.

All assays were performed according to the respective IFUs.

Determination of endpoint titer:

The reciprocal of the highest dilution in which both replicates were reported as positive was defined as the endpoint titer of the sample.

Example: Highest dilution in which both replicates are detected as positive = 1:160

=> Endpoint Titer = 160

Acceptance Criteria:

The reported endpoint titer of a sample is allowed to differ maximum 1 titer level between Method B and C and Reader 1 and 2.

Results:

Table 63: Endpoint titer comparison

Sample IDMethod BMethod C
Reader 1Reader 2Reader 1Reader 2
Serum 11280128012801280
Serum 2640128012801280
Serum 380804040
Serum 4160160320320
Serum 580804080

The reported endpoint titers for the tested sera differ maximally 1 titer level between the different methods and readers.

Raw data see Attachment 17_Section 12_ Raw data Endpoint titer comparison

Conclusion:

Acceptance criteria were met. Sera titrated on HELIOS and manually result in comparable endpoint titers.

3. Clinical cut-off:

See assay cut-off

{55}------------------------------------------------

Image /page/55/Picture/1 description: The image contains text that describes a submission related to AESKUSLIDES nDNA (Crithidia luciliae). The text indicates that it is a 510(k) Traditional Submission. The text is left-aligned and appears to be part of a document or presentation slide.

Image /page/55/Picture/2 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside of what appears to be a vest with a rope around it. To the right of the square is the company name in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below.

4. Expected values/Reference range

Normal Range Study AESKUSLIDES nDNA (Crithidia luciliae)

Expected values for nDNA were analyzed with a panel of 164 sera from healthy donors: 114 from Germany (63 females, 51 males, mean age 32 years, range of 18-58 years, Ethnicity: German) and 50 from the US (17 females, 33 males, mean age 34 years, range of 19-62 years, Ethnicity: 30 Black, 10 Hispanic, 10 White).

Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.

Results:

With AESKUSLIDES nDNA, there was 1 positive result found by reader 1 and no positive result found by reader 2.

Table 64: Normal Range Study Number and Percent positive and negative samples:

Normal RangeStudyPos (%)Neg (%)Total
Reader 11 (0.6)163 (99.4)164
Reader 20 (0)164 (100)164

Table 65: Normal Range Study Positive samples

Normal Range StudyReader 1Reader 2
Sample IDSample OriginResultFIResultFI
NC 072GermanyPos1Neg0

FI = Fluorescence Intensitiy

Only one sample (NC 072) has been found positive in the AESKUSLIDES nDNA assay (FI = 1) by reader 1.

Raw data see Attachment 18_Section 12_Raw data Normal Range Study

Conclusion:

The low number of positive samples we found with AESKUSLIDES nDNA in the tested healthy population correlate well with numbers reported in the literature.

{56}------------------------------------------------

Image /page/56/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the company name, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.

N. Instrument Name

HELIOS AUTOMATED IFA SYSTEM

O. System Description

The HELIOS AUTOMATED IFA SYSTEM is an automated system including pipetting unit with microscope and software that acquires. interprets. stores and displays digital images of stained indirect immunofluorescence slides. The HELIOS DEVICE SOFTWARE is designed to support input of results from the AESKUSLIDES into electronic laboratory data management systems.

The HELIOS AUTOMATED IFA SYSTEM should only be used with AESKUSLIDES assays that are cleared or approved for use on the instrument. All suggested results obtained with the HELIOS DEVICE SOFTWARE must be confirmed by trained personnel.

1. Modes of Operation:

HELIOS: automated processing with automated imaging and automated reading HELIOS User Eval: automated processing with automated imaging and manual reading HELMED Mode: automated processing with manual imaging and manual reading Manual Mode: manual processing with manual imaging and manual reading

Does the applicant's device contain the ability to transmit data to a computer. webserver. or mobile device?

Yes

Does the applicant's device transmit data to a computer. webserver. or mobile device using wireless transmission?

No

2. Software:

General

The HELIOS AUTOMATED IFA SYSTEM is an automated system for immunofluorescence processing with an integrated camera with an optic (microscope) and software for routine laboratory use by professional users under controlled environmental conditions.

The HELIOS camera with an optic (microscope) automatically focuses on the areas of biological material on the slides. Pictures are stored and evaluated by the HELIOS DEVICE SOFTWARE. The software performs a positive/negative of the cells. The software suggests the result as a qualitative result (positive. negative). All suggested

{57}------------------------------------------------

Image /page/57/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.

results obtained with the HELIOS AUTOMATED IFA SYSTEM and the HELIOS DEVICE SOFTWARE must be confirmed by trained personnel before the result is finalized.

Level of Concern

The Level of concern was determined according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005. Following the recommendations of this standard. the HELIOS DEVICE SOFTWARE is assigned moderate level of concern.

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types.

    1. Specimen Identification: Manual sample identification and/or Barcode
    1. Specimen Sampling and Handling: Not applicable
    1. Calibration: There is no calibration of the instrument by the user.
    1. Quality Control:

Positive and negative controls are supplied with the assay reagents (see assay description above).

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable

Q. Proposed Labeling:

    1. AESKUSLIDES nDNA Package Labels: See Section 13
    1. AESKUSLIDES nDNA Package Insert: See Section 13
  1. HELIOS Label - "For Use with AESKUSLIDES nDNA (Crithidia luciliae)" See Section 13
    1. HELIOS Instruction Manual Draft Section 13
    1. All other Labels see Section 13

{58}------------------------------------------------

February 16, 2018

Image /page/58/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Aesku Diagnostics GmbH & Co. KG Sandra Reuter Regulatory Affairs Manager Mikro-Forum-Ring 2 Wendeslheim, 55234 DE

Re: K172348

Trade/Device Name: AESKUSLIDES nDNA (Crithidia luciliae), AESKUSLIDES nDNA (Crithidia luciliae) Demo Kit, AESKUSLIDES nDNA (Crithidia luciliae) Bulk kit x5, AESKUSLIDES nDNA (Crithidia luciliae) Bulk kit x10 Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LSW, PIV Dated: July 28, 2017 Received: August 3, 2017

Dear Sandra Reuter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{59}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).