(254 days)
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
The Straub Endovascular System consists of:
- Rotarex®S catheter set which includes a Rotarex®S catheter, guidewire, collecting bag and drape. ● All are provided sterile and are for single use
- . Drive System, consisting of the Control Unit, Motor and Foot switch. The Drive System serves to drive and control the rotational catheters.
Rotarex S is an over-the-wire, percutaneous catheter. The catheter consists of a flexible braided shaft, a rotating head, and a rotating helix which runs the length of the catheter. A lumen for the passage of the supplied guidewire runs the entire length of the helix and through the head of the catheter. The catheter head is made up of two overlying metal cylinders, with two side openings.
The outer cylinder is connected to the rotating helix, and the inner cylinder to the catheter shaft. The helix and the catheter head rotate at approximately 40.000-60.000 rpm depending on the model, by means of a gear box in the catheter housing and a motor contained within the catheter handle driven by the Drive System. The rotating outer cylinder is fitted with facets at its foremost tip, which when rotating, serve to abrade occluding material lying in front of it. Concomitantly, the rotation of the catheter head creates a vortex in the blood which assists to further erode occluding material from the vessel lumen. The detached material is aspirated into the side openings of the overlying cylinders where it is further broken down within the head and then carried through the inner lumen into a collecting bag outside of the body.
The provided text is a 510(k) Summary for the Straub Endovascular System, which is an atherectomy device. The document details the device's indications for use, its comparison to predicate devices, and a summary of performance testing and clinical evaluations.
However, the information provided does not contain the specifics required to answer your request about acceptance criteria and a study proving the device meets those criteria for an AI/ML medical device. The document describes a traditional medical device (an atherectomy system), and the testing outlined (bench validation, biocompatibility, sterilization, electrical safety, animal study, and a meta-analysis of clinical studies) are standard for such devices.
There is no mention of any AI/ML components, no acceptance criteria related to AI/ML performance metrics (like sensitivity, specificity, AUC), no test set details for AI/ML evaluation, no expert ground truth establishment, no MRMC study for AI assistance, and no standalone AI performance analysis.
Therefore, I cannot extract the information requested as it is not present in the provided text.
To answer your request, the document would need to describe a study that directly evaluates an AI/ML component of the Straub Endovascular System, including:
- A table of acceptance criteria (e.g., minimum sensitivity, specificity, AUC, or other relevant AI performance metrics) and the reported device performance against these.
- Sample size and data provenance for an AI/ML test set.
- Number and qualifications of experts used to establish ground truth for AI/ML.
- Adjudication method for ground truth in an AI/ML context.
- MRMC study details if human readers were assisted by AI, including effect size.
- Standalone AI performance (algorithm only).
- Type of ground truth (e.g., expert consensus on images, pathology results, patient outcomes) specifically for AI evaluation.
- Training set sample size for the AI model.
- How ground truth for the training set was established.
Based solely on the provided text, the device is evaluated through engineering verification tests, animal studies, and a meta-analysis of existing clinical data. These are not AI/ML specific evaluations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
April 12, 2018
Straub Medical AG % Ms. Barbara Atzenhoefer Principal Medical Research Scientist, Regulatory NAMSA 400 Hwy 169 South, Suite 500 Minneapolis, MN 55426
Re: K172315
Trade/Device Name: Straub Endovascular System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, DQX Dated: March 9, 2018 Received: March 12, 2018
Dear Ms. Atzenhoefer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
1000-1050.
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K172315
Device Name Straub Endovascular System
Indications for Use (Describe)
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Straub Medical. The word "straub" is written in a dark blue, sans-serif font, with a registered trademark symbol next to it. Below the word "straub" is a teal rectangle with the word "MEDICAL" written in a white, sans-serif font. There are also three horizontal teal lines above the teal rectangle.
510(k) Summary
Applicant Information
| Date Prepared: | April 10, 2018 |
|---|---|
| Applicant: | Straub Medical AGStraubstrasse 12CH 7323 Wangs, Switzerland |
| Contact Person: | Amy Ehlenfeldt, Manager Quality and Regulatory Compliance |
| Telephone: | +41 (0) 81 720 04 60 |
| Fax: | +41 (0) 81 720 04 61 |
Subject Device
| Name of Device: | Straub Endovascular System |
|---|---|
| Common/Usual Name: | Peripheral Atherectomy Catheter |
| Classification Name: | Intraluminal Artery Stripper |
| Regulation: | 21 CFR §870.4875 |
| Regulatory Class: | Class II |
| Product Codes: | MCW and DQX |
Predicate
MEDRAD Jetstream Atherectomy System K133023
Asahi Intecc Co. Ltd. Asahi Asato 30 guidewire K071721
Device Description
The Straub Endovascular System consists of:
- Rotarex®S catheter set which includes a Rotarex®S catheter, guidewire, collecting bag and drape. ● All are provided sterile and are for single use
- . Drive System, consisting of the Control Unit, Motor and Foot switch. The Drive System serves to drive and control the rotational catheters.
Rotarex S is an over-the-wire, percutaneous catheter. The catheter consists of a flexible braided shaft, a rotating head, and a rotating helix which runs the length of the catheter. A lumen for the passage of the supplied guidewire runs the entire length of the helix and through the head of the catheter. The catheter head is made up of two overlying metal cylinders, with two side openings.
The outer cylinder is connected to the rotating helix, and the inner cylinder to the catheter shaft. The helix and the catheter head rotate at approximately 40.000-60.000 rpm depending on the model, by means of a gear box in the catheter housing and a motor contained within the catheter handle driven by the Drive System. The rotating outer cylinder is fitted with facets at its foremost tip, which when rotating, serve to abrade occluding material lying in front of it. Concomitantly, the rotation of the catheter head creates a vortex in the blood which assists to further erode occluding material from the vessel lumen. The detached material is aspirated into the side openings of the overlying cylinders where it is further broken down within the head and then carried through the inner lumen into a collecting bag outside of the body.
{4}------------------------------------------------
Indications for Use
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
Comparison of Technological Characteristics
As compared to the predicate devices, the Straub Endovascular System has similar components, indications for use, and identical classification codes as the selected predicate devices. The Straub Endovascular System is comprised of similar patient contact materials and has undergone appropriate biocompatibility testing for its intended use. As with the predicate devices, the Rotarex®S Catheter Set components are provided sterile and are intended for single-use.
With respect to general device design and technological characteristics, the Straub Endovascular System is substantially equivalent to the selected predicate device, since it similarly uses a cutting tip controlled by an external control unit and allows for the removal of tissue, thrombus, and fluid via aspiration. A summary of the similarities and differences are shown below.
| Catheter Predicate - Similarities and Differences | ||
|---|---|---|
| Characteristic | Straub Endovascular System | Predicate |
| Intended Use | Atherectomy and thrombus removal | Atherectomy and thrombus removal |
| System Components | Catheter, console/control unit, guidewire,Collecting bag, sterile drape for motor | Guidewire, Collecting bag, console /control unit |
| Mechanism of Action | Front cutting / rotational | Front cutting / rotational |
| Catheter Diameter | 6F: 2.00 mm8F: 2.67 mm | XC 2.1/3.0: 2.1mm (3.0mm blades out)XC 2.4/3.4: 2.4mm (3.4mm blades out) |
| Catheter Length | 6F: 110 or 135 cm8F: 85 or 110 cm | XC 2.1/3.0: 135 cmXC 2.4/3.4: 120 cm |
| Max. Guidewire Dia. | 0.018" | 0.014" |
| Min. Introducer Size | 6F Rotarex: 6-7 F8F Rotarex: 8-9 F | XC 2.1/3.0: 7 FXC 2.4/3.4: 7 F |
| Enclosed CuttingBlades | Yes | No |
| Target speed | 6F: 60,000 rpm8F: 40,000 rpm | XC 2.1/3.0: 70,000 rpmXC 2.4/3.4: 70,000 rpm |
| Aspiration | Yes | Yes |
| Infusion | No | Yes |
| Sterilization Method | EtO | EtO |
There are no significant technological differences among the subject device and the predicate device. Therefore, the Straub Endovascular System is substantially equivalent to the Jetstream Atherectomy System.
Performance Testing Summary
The following testing was performed to demonstrate that the system and its components perform as intended:
Design verification and bench validation testing:
Rotarex®S system: Visual and Dimensional Inspections, Dynamic (Durability) Test Under Simulated Use Conditions, Tensile Test, Torsional Strength, Kink Test, Corrosion Resistance, Temperature Rise (heat generation) Testing, Aspiration Capacity, FAS (Helix) Integrity, Stator Coating Integrity, In vitro Calcified Plaque and Thrombus Removal Studies, Rotational Speed stability). Guidewire: Visual and Dimensional Inspections, Corrosion resistance, Fracture Testing, Bending, Tensile Strength, Tip Flexibility, Torque Response, Torque Strength, Radiopacity
K172315 510(k) Summary
{5}------------------------------------------------
Shelf-life was established at 3-years based on accelerated aging studies.
Biocompatibility testing was conducted on the catheter and guidewire in accordance with ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process (and related sub-parts). Testing and included Cytotoxicity, Acute Systemic Toxicity, Sensitization, Intracutaneous Reactivity, Complement Activation, Hemolysis and Material-mediated Pyrogenicity.
Sterilization validations of the catheter and guidewire were conducted in accordance with ISO 11135-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process Sterilization of health-care products - Ethylene oxide - Requirements for the development. validation and routine control of a sterilization process for medical devices.
Electrical safety and Electromagnetic compatibility testing was conducted and the system determined to be compliant with the requirements of IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
A GLP animal study was conducted to assess the safety of the Straub Endovascular System in a porcine model. The following were assessed: procedure and device handling data, clinical pathology, adverse events, gross necropsy, histopathology and vascular injury score.
Clinical Testing
A meta-analysis of 8 clinical studies comprising data obtained from 2,107 patients studied from 2002 to 2015 was performed to establish Rotarex® S clinical performance. The statistical summary of the literature data was to provide a quantitative review and synthesis of related, but independent, studies of the Rotarex device. The analyses allowed for the estimation of various clinical outcomes including, but not limited to technical and clinical success, target lesion revascularization, restenosis rates, ankle-brachial index, Rutherford score and procedure-related adverse events.
Conclusions
The data and information presented within this submission and the similarities between the subject and predicate device support a determination of substantial equivalence, and therefore market clearance of the subject Straub Endovascular System through this 510(k) Premarket Notification.
§ 870.4875 Intraluminal artery stripper.
(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).