K Number
K172254
Manufacturer
Date Cleared
2017-10-23

(89 days)

Product Code
Regulation Number
866.3830
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lyme B. burgdorferi (IgM) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgM antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2-EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

Device Description

The kit is an immunoblot method to detect IgM antibodies against B. burqdorferi antigens. The test kit contains:
. Nitrocellulose Test Strips with purified B. burgdorferi antigens (3) and quality control lines (3) present in specific positions
. Sample Diluent. Provided for specimen dilutions. Contains BSA and PBS
Positive Control derived from human serum positive for Lyme disease. Contains <0.1% sodium azide
Negative Control derived from human serum negative for Lyme disease. Contains <0.1% sodium azide
Conjugate. Antihuman IgM-HRP Conjugate binds reactive antibodies to the Substrate
TMB Substrate. Provides colorimetric reaction for visual read of bound antibodies
PBS Wash Buffer concentrate. Removes reagents and unbound antibodies after incubation steps. Must be reconstituted to 1L with distilled or deionized water.

To perform the test, serum or plasma is individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diluted, ready-to-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 2 (or more) of the 3 test lines are considered positive for specific IgM antibody to B. buradorferi.

AI/ML Overview

The provided document describes the Lyme B. burgdorferi (IgM) MarStripe Test, an immunoblot assay for detecting IgM antibodies to Borrelia burgdorferi. The submission aims to establish substantial equivalence to a legally marketed predicate device, the MarDx B. burgdorferi IgM MarBlot Strip Test System (K951709).

Here's an analysis of the acceptance criteria and the study data:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for performance metrics. Instead, the study aims to demonstrate substantial equivalence to a predicate device. Therefore, the "acceptance criteria" can be inferred as achieving comparable performance to the predicate device and demonstrating acceptable analytical characteristics.

Performance MetricImplied Acceptance Criteria (based on predicate equivalence)Reported Device Performance (Lyme B. burgdorferi (IgM) MarStripe Test)
Method Comparison (vs. Predicate IgM WB for EIA-positive/equivocal samples)Substantial equivalence to predicate device's agreement percentages.Positive % Agreement: 93.5% (95% CI: 90.1% - 95.8%) Negative % Agreement: 93.5% (95% CI: 90.2% - 95.7%)
Analytical Specificity (Normal Individuals)High specificity (e.g., >95%)99.5% (95% CI: 97.1% - 100%)
Cross-reactivityAll positive results should be confirmed by predicate device. Low overall false positivity rate.All 7 positive specimens (out of 246 potentially cross-reactive) by MarStripe Test were confirmed positive by predicate IgM Western blot device. Overall positive specimens in cross-reactive cohort: 7 (2.8%)
Sensitivity Comparison (Well-characterized Lyme disease specimens)Comparable sensitivity to predicate device.18.4% (16/87) (95% CI: 11.2% - 28.4%)
Sensitivity Comparison (CDC Panels)Comparable sensitivity to predicate device.Overall: 23.8% (10/42)
Precision (Intra-laboratory)High qualitative agreement, especially for negative and strongly positive samples.Low Negative and Moderate Positive samples: 100% agreement. Low Positive: 99.0% positive agreement. Cutoff: 81.3% positive agreement.
Reproducibility (Inter-laboratory)High qualitative agreement across sites and operators.Low Negative and one Moderate Positive samples: 100% final positive/negative agreement. One High Negative: 99.7% negative agreement. One Low Positive: 99.3% positive agreement. Cutoff: 68.4% positive agreement.
Interference100% qualitative agreement with and without interfering agents.100% qualitative agreement for all tested specimens.
Serum vs. Plasma Matrix Comparison100% qualitative agreement between serum and plasma samples.100% qualitative agreement for all 20 pairs.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison: 676 specimens tested (those found positive or equivocal on an FDA-cleared first-step EIA). The study was a prospective study performed at three geographically distinct study sites. The country of origin for the data is not explicitly stated but implied to be the USA given the FDA context and reference to CDC.
  • Analytical Specificity: 220 sera from normal individuals (blood bank donors) representing endemic geographic regions of the United States. Data provenance is from US blood bank donors, retrospective or prospective not specified but typical for such collections.
  • Cross-reactivity: 246 potentially cross-reactive specimens from individuals with various other infectious conditions. Data provenance is not specified.
  • Sensitivity: 87 well-characterized Lyme disease clinical specimens. Samples included early, early disseminated, and late phases. Data provenance is not specified but implied to be clinical.
  • CDC Panels: 10 samples from the Lyme Disease Validation Panel and 32 samples from the Lyme Disease Basic Research Panel (Total 42 samples). These are reference panels from the Centers for Disease Control and Prevention (CDC).
  • Precision: 8 specimens, each tested in 4 replicates, two runs per day over 12 days (total 96 tests per specimen).
  • Reproducibility: 8 specimens, each tested in 4 replicates, two runs per day over 12 days, at each of three laboratory sites (total 192 tests per specimen per site, or 576 read-outs in total).
  • Interference: 2 Lyme IgM negative and 3 IgM positive sera.
  • Serum vs. Plasma Matrix Comparison: 20 pairs of sera/plasma (total 40 samples).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test sets.

  • For the Method Comparison study, the ground truth was established by comparison to results from an "FDA cleared immunoblot results" (the predicate device) which followed "recommended criteria described by the Centers for Disease Control (CDC) and the Second National Conference on Serological Diagnosis of Lyme Disease". This implies standardized interpretation criteria rather than individual expert consensus.
  • For the Sensitivity study, "87 well characterized Lyme disease clinical specimens" were used, implying their disease status was previously determined, likely through a combination of clinical diagnosis and laboratory testing, but specific expert involvement is not detailed.
  • For the CDC Panels, the ground truth is the CDC's characterization of these reference panels.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set interpretation by human readers. For the reproducibility study, results were read by "2 human operators at each site," but it does not specify if or how discrepancies between these operators were resolved (e.g., 2+1, 3+1). The "ground truth" for the overall studies appears to be based on established reference methods or clinical characterization rather than adjudicated interpretations of the MarStripe test itself.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described in the context of human readers improving with AI vs. without AI assistance. The studies performed were primarily focused on the device's analytical performance and its agreement with a predicate device and established reference panels.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

The device itself (Lyme B. burgdorferi (IgM) MarStripe Test) is an immunoblot assay that requires visual reading. The results show "Positives", "Negatives", "Band Type" (Pos., Neg., Cut., Wpos.), implying visual interpretation by a human. For example, the reproducibility study explicitly mentions "Results were read by 2 human operators at each site". Therefore, this is not a standalone algorithm without human-in-the-loop performance; human visual interpretation is an integral part of the test.

7. The Type of Ground Truth Used

The ground truth varied depending on the study:

  • Method Comparison: Ground truth was the results from an FDA-cleared predicate IgM Western blot device, interpreted according to CDC and Second National Conference on Serological Diagnosis of Lyme Disease criteria.
  • Analytical Specificity: Implied ground truth is "normal individuals" from blood banks, presumed negative for Lyme disease.
  • Cross-reactivity: Ground truth was the clinical diagnosis of other infectious conditions, with validation against the predicate IgM Western blot device for any MarStripe positive results.
  • Sensitivity: "Well characterized Lyme disease clinical specimens," implying clinical diagnosis of Lyme disease stages.
  • CDC Panels: Characterization provided by the CDC for their reference panels.
  • Precision/Reproducibility/Interference/Serum vs. Plasma: The ground truth for these analytical studies was related to the known characteristics of the samples (e.g., known negative, known positive, spiked with interferents, paired serum/plasma) as determined by prior testing (e.g., FDA-cleared B. burgdorferi ELISA results).

8. The Sample Size for the Training Set

The document describes performance studies for the Lyme B. burgdorferi (IgM) MarStripe Test. This device is a diagnostic assay, not an AI/ML model that requires a "training set" in the computational sense. The studies presented are validation studies of the finished device. There is no mention of an algorithm development phase with a dedicated training set.

9. How the Ground Truth for the Training Set Was Established

Since the device is a diagnostic assay (not an AI/ML model), there is no "training set" as understood in machine learning. Therefore, this question is not applicable to the information provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected by flowing lines that resemble a ribbon or a wave.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 23, 2017

TRINITY BIOTECH KEVIN LAWSON REGULATORY OFFICER 60 PINEVIEW DR. BUFFALO NY 14031

Re: K172254

Trade/Device Name: Lyme B. Burgdorferi (igm) Marstripe Test Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: II Product Code: LSR Dated: July 11, 2017 Received: July 26, 2017

Dear Mr. Lawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kristian M. Roth -S

For:

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172254

Device Name

Lyme B. burgdorferi (IgM) MarStripe Test

Indications for Use (Describe)

Lyme B. burgdorferi (IgM) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgM antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2-EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green dots on the left side, followed by the word "immco" in blue, sans-serif font. Below "immco" is the word "DIAGNOSTICS" in green, sans-serif font. Underneath that is the text "A Trinity Biotech Company" in a smaller, sans-serif font.

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

1.Submitter Name:Immco Diagnostics, Inc. a Trinity Biotech Company
Address:60 Pineview Dr., Buffalo, NY 14228
Phone Number:716-691-0091 ext. 110
Contact Person:Kevin Lawson
Summary Prepared:7-24-2017
  • Device Name: Lyme B. burgdorferi (IgM) MarStripe Test 2. Common Name: Lyme B. burgdorferi (IgM) Immunoblot Product Code: LSR
    1. Substantially equivalent to: MarDx B. burqdorferi IgM MarBlot Strip Test System (K951709)
    1. Description of device: The kit is an immunoblot method to detect IgM antibodies against B. burqdorferi antigens. The test kit contains:
    • . Nitrocellulose Test Strips with purified B. burgdorferi antigens (3) and quality control lines (3) present in specific positions
    • . Sample Diluent. Provided for specimen dilutions. Contains BSA and PBS
    • Positive Control derived from human serum positive for Lyme disease. Contains <0.1% sodium azide
    • Negative Control derived from human serum negative for Lyme disease. Contains <0.1% sodium azide
    • Conjugate. Antihuman IgM-HRP Conjugate binds reactive antibodies to the Substrate
    • TMB Substrate. Provides colorimetric reaction for visual read of bound antibodies
    • PBS Wash Buffer concentrate. Removes reagents and unbound antibodies after incubation steps. Must be reconstituted to 1L with distilled or deionized water.

To perform the test, serum or plasma is individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diluted, ready-to-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 2 (or more) of the 3 test lines are considered positive for specific IgM antibody to B. buradorferi.

  • Intended Use: Lyme B. burqdorferi (JgM) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human 5. lgM antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K.- EDTA) in samples which have been found positive or equivocal using an ElA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.
    1. Similarities and Differences: The Lyme B. burgdorferi (IgM) MarStripe Test was compared to a commercially marketed kit by Trinity Biotech the B. burgdorferi (JgM) MarBlot Strip Test System (K951709). Both kits have the same intended use and use the same methodology except that the MarStripe Test has been validated in plasma (K- EDTA) as well as serum. Both immunoblot kits detect antibodies to B. burgdorferi antigens using the standard immunoblot IgM algorithm. The MarStripe uses a Horseradish Peroxidase Conjugate and Tetramethylbenzidine (TMB) Substrate/Chromogen in comparison to The MarBlot Alkaline Phosphatase Conjugate and BCIP/NBT Substrate/Chromogen. The MarStripe uses a Cutoff Control line incorporated in the strip in comparison to the 41kD band of Weakly Reactive Control.
  • Non-clinical Tests: 7. Precision

60 Pineview Drive Buffalo, NY 14228-2120 Toll free (800) 537-8378 tel. (716) 691-0091

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a series of blue and green dots on the left side, followed by the word "immco" in blue, with the word "DIAGNOSTICS" in green underneath. Below that, it says "A Trinity Biotech Company" in a smaller font.

Eight specimens were tested by Lyme B. burqdorferi (JgM) MarStripe Test in 4 replicates, two runs per day over 12 days for a total of 96 tests for each specimen. Samples were selected based on FDA cleared B. burqdorferi ELISA results, including 2 low negative samples, 2 high negative samples, 1 cutoff sample and 2 moderate positive samples. Qualitative agreement was 100% for the low negative and moderate positive samples. The low positive specimen produced 99.0% positive agreement. The cutoff specimen produced 81.3% positive agreement.

Samplen=96p41p39p23
1Low Negative
Band TypeNeg.Neg.Neg.
Positives000
Negatives969696
% Positive0.0%0.0%0.0%
2Low Negative
Band TypeNeg.Neg.Neg.
Positives003
Negatives969693
% Positive0.0%0.0%3.1%
3High Negative
Band TypeNeg.Neg.Neg.
Positives000
Negatives969696
% Positive0.0%0.0%0.0%
4High Negative
Band TypeNeg.Neg.Cut.
Positives003
Negatives969693
% Positive0.0%0.0%3.1%
5Cutoff
Band TypeCut.Neg.Wpos.
Positives88079
Negatives89617
% Positive91.7%0.0%82.3%
6Low Positive
Band TypePos.Neg.Pos.
Positives96096
Negatives0960
% Positive100.0%0.0%100.0%
7Moderate Positive
Band TypePos.Neg.Pos.
Positives96096
Negatives0960
% Positive100.0%0.0%100.0%
8Moderate Positive
Band TypePos.Pos.Pos.
Positives969696
Negatives000
% Positive100.0%100.0%100.0%

Pos = positive band. Neg = negative band. Wpos = weak positive band. Cut = equivocal band.

Reproducibility

Eight specimens were tested by Lyme B. burgdorferi (JgM) MarStripe Test in 4 replicates, two runs per day over 12 days for a total of 192 tests for each specimen at each of three laboratory sites. Results were read by 2 human operators at each site, equaling a total of 576 read-outs. Samples were selected based on FDA cleared B. burgdorferi ELISA results, including 2 low negative samples, 2 high negative samples, 1 cutoff sample and 2 moderate positive samples. Final positive or negative agreement was 100% for both low negative and one moderate positive sample. One high negative sample produced 99.7% negative arreement, one low positive sample produced 99.3% positive agreement. The cutoff sample produced 68.4% positive agreement.

60 Pineview Drive ● Buffalo, NY 14228-2120 ● USA Toll free (800) 537-8378 ● tel. (716) 691-0091 ● fax (716) 691-0466

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a series of blue and green dots arranged in a circular pattern on the left side. To the right of the dots is the company name "immco" in blue, with "DIAGNOSTICS" in green below it. Underneath the company name is the text "A Trinity Biotech Company" in a smaller font size.

Samplen=576p41p39p23
1Low Negative
Band TypeNeg.Neg.Neg.
Positives001
Negatives576576575
% Positive0.0%0.0%0.2%
2Low Negative
Band TypeNeg.Neg.Neg.
Positives0026
Negatives576576550
% Positive0.0%0.0%4.5%
3High Negative
Band TypeNeg.Neg.Neg.
Positives000
Negatives576576576
% Positive0.0%0.0%0.0%
4High Negative
Band TypeNeg.Neg.Cut.
Positives1258
Negatives575574518
% Positive0.2%0.3%10.1%
5Cutoff
Band TypeCut.Neg.WPos.
Positives4420504
Negatives13457672
% Positive76.7%0.0%87.5%
6Low Positive
Band TypePos.Neg.Pos.
Positives5720576
Negatives45760
% Positive99.3%0.0%100.0%
7Moderate Positive
Band TypePos.Neg.Pos.
Positives5730572
Negatives15742
% Positive99.8%0.0%99.7%
8Moderate Positive
Band TypePos.Pos.Pos.
Positives576572576
Negatives040
% Positive100.0%99.3%100.0%

Pos = positive band. Neg = negative band. Cut = equivocal band. Note: two replicates of one run are missing due to lack of sample 7.

Analytical Specificity: 220 sera from normal individuals (blood bank donors) representing endemic geographic regions of the United States were tested with Lyme B. burgdorferi (JgM) MarStripe Test. Analytical specificity was determined to be 99.5% (95% CI: 97.1% - 100%).

Normal Individuals
Lyme IgM MarStripeTestPositive1
Negative219
Total220

Cross-reactivity: A total of 246 potentially cross-reactive specimens from individuals with other or infectious conditions were tested on Lyme B. burgdorferi (lgM) MarStripe Test. All positive specimens by the Lyme B. burgdorferi (IgM) MarStripe Test were confirmed positive when tested by the predicate IgM Western blot device.

60 Pineview Drive ● Buffalo, NY 14228-2120 ● USA Toll free (800) 537-8378  •  tel. (716) 691-0091  •  fax (716) 691-0466

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a cluster of blue and green dots on the left, followed by the word "immco" in blue, and "DIAGNOSTICS" in green. Below the word "DIAGNOSTICS" is the text "A Trinity Biotech Company".

Positive specimens/reactive antibody lines - n(%)
PopulationnPositivep41p39p23
E. chafeensis103 (30)3 (30)5 (50)4 (40)
B. microti102 (20)2 (20)1 (10)1 (10)
L. interogans101 (10)1 (10)1 (10)1 (10)
H. pylori1001 (1)1 (1)1 (1)1 (1)
Syphilis100 (0)0 (0)0 (0)0 (0)
Influenza100 (0)0 (0)1 (10)0 (0)
Epstein-Barr Virus220 (0)1 (4.5)0 (0)0 (0)
Rocky Mountain Spotted fever100 (0)0 (0)0 (0)0 (0)
Parvovirus B1990 (0)0 (0)2 (22.2)1 (11.1)
Systmatic lupus erythematosus150 (0)0 (0)0 (0)0 (0)
Cytomegalovirus100 (0)0 (0)0 (0)0 (0)
Rheumatoid arthritis150 (0)0 (0)0 (0)0 (0)
Celiac150 (0)0 (0)0 (0)0 (0)
Total2467 (2.8)8 (3.2)11 (4.4)8 (3.2)

Interference: Two Lyme IgM negative and three IgM positive sera were spiked with hemoglobin (2g/L), unconjugated bilirubin (342 µmo//L), RF (100 IU/ml), triglycerides (3.7 mmol/L) and total cholesterol (13 mmol/L) and tested using this assay. Samples were tested with and without interfering agents. Qualitative agreement was 100% for all specimens.

Serum vs. Plasma Matrix Comparison Studies: To establish equivalence of serum vs. plasma matrix, 20 pairs of sera/plasma (samples A-J below) were sourced from specimens tested on an FDA cleared Lyme EIA assay. These specimens included 3 Western Blot IgM positives and 17 negatives. These samples were assayed on the Lyme B. burgdorferi (IgM) MarStripe Test. Qualitative agreement for all pairs was 100%.

SampleTypep41p39p23ResultBand % Agrmt
ASerum000NEG100
APlasma000NEG
BSerum101POS100
BPlasma101POS
CSerum000NEG100
CPlasma000NEG
DSerum001NEG100
DPlasma001NEG
ESerum101POS100
EPlasma101POS
FSerum001NEG100
FPlasma001NEG
GSerum001NEG100
GPlasma001NEG
HSerum000NEG100
HPlasma000NEG
ISerum000NEG100
IPlasma000NEG
JSerum101POS100
JPlasma101POS
KSerum000NEG100
KPlasma000NEG
LSerum000NEG100
LPlasma000NEG
MSerum000NEG100
MPlasma000NEG
NSerum000NEG100

60 Pineview Drive USA Toll free (800) 537-8378 ● tel. (716) 691-0091 ● fax (716) 691-0466

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a cluster of blue and green dots on the left, followed by the word "immco" in blue, with a small dot above the "o". Below "immco" is the word "DIAGNOSTICS" in green. Underneath that is the text "A Trinity Biotech Company" in a smaller font.

NPlasma000NEG
OSerum000NEG100
OPlasma000NEG
PSerum001NEG100
PPlasma001NEG
QSerum000NEG100
QPlasma000NEG
RSerum000NEG100
RPlasma000NEG
SSerum001NEG100
SPlasma001NEG
TSerum000NEG100
TPlasma000NEG
  1. 0 = negative band result. 1 = positive band result.

8. Clinical Tests:

Method Comparison: A prospective study of FDA cleared first-step ElA specimens was performed at three geographically distinct study sites. The specimens testing positive (n=676) on a FDA cleared first-step ElA were tested with Lyme B. burgdorferi (lgM) MarStripe Test and an FDA cleared immunoblot results followed the recommended criteria described by the Centers for Disease Control (CDC) and the Second National Conference on Serological Diagnosis of Lyme Disease 22. The results are summarized below.

Predicate IgM WB
PositiveNegativeTotal
Lyme IgMMarStripe TestPositive30223325
Lyme IgMMarStripe TestNegative21330351
Total323553676

Positive % Agreement: Negative % Agreement:

93.5% (95% CI: 90.1% - 95.8%) 93.5% (95% Cl: 90.2% - 95.7%)

Sensitivity: 87 well characterized Lyme disease clinical specimens were tested with the Lyme B. burgdorferi (bgM) MarStripe Test. Specimens included samples from early, early diseminated, and late phases of the sensitivity obtained was compared with that of the predicate device.

IntervalnLyme IgM MarStripe TestPredicate IgM WB
Positive%Positive%
Early Lyme (stage 1)19842.1947.4
Early disseminated (stage 2)43511.6614
Late Lyme (stage 3)2531228
Overall871618.41719.5

Sensitivity Comparison:

Lyme B. burgdorferi (IgM) MarStripe Test: 18.4% (16/87) (95% CI: 11.2% - 28.4%) Predicate device: 19.5% (17/87) (95% CI: 12.1% - 29.7%) Difference in proportion: 1.1%

CDC Panels: Reference panels from the Center for Disease Control and Prevention (Lyme Disease Validation Panel n=10, Lyme Disease Basic Research Panel n=32) were tested on the Lyme B. burgdorferi (JgM) MarStripe Test and the predicate device.

60 Pineview Drive ● Buffalo, NY 14228-2120 ● USA Toll free (800) 537-8378 ● tel. (716) 691-0091 ● fax (716) 691-0466

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green dots on the left side, followed by the word "immco" in blue, sans-serif font. Below "immco" is the word "DIAGNOSTICS" in green, sans-serif font. Underneath that is the text "A Trinity Biotech Company" in a smaller, sans-serif font.

IntervalnLyme IgMMarStripe TestPredicate IgM WB
positive%positive%
Controls250028
Early Lyme (stage 1)10660660
Early disseminated (stage 2)331003100
Late Lyme (stage 3)4125250
Overall421023.81331.0

Note: The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.

  • Conclusion: From the performance data and kit comparison above, it is our conclusion that the Lyme B. burgdorferi (IgM) 9. MarStripe Test is substantially equivalent to the B. burgdorferi (JgM) MarBlot Strip Test System (K951709) commercially marketed by Trinity Biotech
    Kus lausa

Kevin J. Lawson VP Regulatory Affairs

60 Pineview Drive ● USA Toll free (800) 537-8378 ● tel. (716) 691-0091

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).