K Number
K172238
Date Cleared
2017-12-14

(141 days)

Product Code
Regulation Number
876.5820
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CAREline Twister: The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the responsibility of the physician. The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. The Fresenius Access Flow Reversing Connector (AFRC) is for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure. Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K2 and T Hemodialysis Machines. CAREline Standard: The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the resonsiblity of the physician. The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K.2 and T Hemodialysis Machines.

Device Description

The CAREline bloodlines are intended for use in acute and chronic hemodialysis therapy. The bloodline is part of the hemodialysis extracorporeal circuit, which transports arterial blood from the patient's arterial access (e.g., fistula or catheter) through a hemodialyzer and back to the patient's venous access.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the CAREline Airless Hemodialysis Blood Tubing Sets. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study to prove device performance against specific acceptance criteria.

Therefore, much of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment is not explicitly available in the provided document. The document outlines performance testing conducted to support substantial equivalence.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of explicit acceptance criteria with numerical targets. Instead, it lists various performance tests conducted to support substantial equivalence. The "reported device performance" is implied to have met the requirements of these tests, as the FDA concluded substantial equivalence.

Performance Test TypeReported Device Performance
Performance testing of connectors (hemodialyzer, vascular access device, and ancillary components)Results support determination of substantial equivalence (specific outcomes not detailed).
Structural integrity testingResults support determination of substantial equivalence (specific outcomes not detailed).
Biological safety testing (biocompatibility)Passed: Chemical analysis (extractables and leachables), Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Material Mediated Pyrogenicity, Genotoxicity, Hemocompatibility. Toxicological risk assessment performed.
Sterility and pyrogenicity testingResults support determination of substantial equivalence (specific outcomes not detailed).
Structural integrity - Pressure and pull testingResults support determination of substantial equivalence (specific outcomes not detailed).
Pressure testing Access ports and pressure output devices (PODs)Results support determination of substantial equivalence (specific outcomes not detailed).
Endurance testingConducted with Fresenius Medical Care 2008® Series K, K2, and T Hemodialysis Machines. Results support determination of substantial equivalence (specific outcomes not detailed).
Clamp testing (Ability to occlude and effects of repeated use)Results support determination of substantial equivalence (specific outcomes not detailed).
Usability evaluationResults support determination of substantial equivalence (specific outcomes not detailed).

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for each of the performance tests. It also does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. The testing was conducted in accordance with ISO 8638:2010.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The performance tests described are primarily engineering and biological safety tests, which typically rely on standardized protocols and instrumental measurements rather than human expert interpretation for "ground truth" in the way a diagnostic AI might.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which are not detailed here for these performance tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was conducted. The device is a physical medical device (blood tubing set) and does not involve AI or human readers in an interpretative capacity.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this is a physical medical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the listed performance tests, the "ground truth" would be established by the physical and chemical properties measured according to international standards (e.g., ISO 8638:2010) and established biological safety testing methods. For instance:

  • Biocompatibility: Ground truth is based on the results of validated assays (e.g., cytotoxicity, sensitization, genotoxicity tests) against predetermined pass/fail criteria specific to those tests, often compared to controls.
  • Structural Integrity: Ground truth is based on whether the device withstands specified pressures, pulls, or cycles without failure, as defined by engineering specifications.

8. The sample size for the training set

Not applicable. This is a physical medical device undergoing performance testing, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 14, 2017

Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451

Re: K172238

Trade/Device Name: CAREline Airless Hemodialysis Blood Tubing Sets Regulation Number: 21 CFR& 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: KOC, FJK Dated: November 10. 2017 Received: November 13, 2017

Dear Denise Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Charles Viviano -S

Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172238

Device Name

CAREline Airless Hemodialysis Blood Tubing Sets

Indications for Use (Describe)

CAREline Twister:

The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the responsibility of the physician.

· The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.

• The Fresenius Access Flow Reversing Connector (AFRC) is for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure.

· Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K2 and T Hemodialysis Machines.

CAREline Standard:

The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the resonsiblity of the physician.

• The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.

· Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K.2 and T Hemodialysis Machines.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top and "MEDICAL CARE" on the bottom.

K172238 Page 1 of 6

5. 510(K) SUMMARY

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR 807.92.

5.1. Submitter's Information

Name:Fresenius Medical Care Renal Therapies Group, LLC
Address:920 Winter StreetWaltham, MA02451-1457
Phone:(781) 699-4479
Fax:(781) 699-9635
Contact Person:Denise Oppermann, Senior DirectorRegulatory Affairs – Devices
Preparation Date:24 July 2017

5.2. Device Name

Trade Name:CAREline Airless Hemodialysis Blood Tubing Set
Common Name:KOC – Accessories, Blood Circuit,HemodialysisFJK - Set, Tubing, Blood, With and WithoutAnti-regurgitation Valve
Classification Name :Hemodialysis system and accessories
Regulatory Class:Class II per 21 CFR 876.5820
Product Code:KOC / FJK
Classification Panel:Gastroenterology-Urology

5.3. Legally Marketed Predicate Device

Primary Predicate - CombiSet® Blood Tubing with Access Flow 5.3.1. Reversing Connector (Twister®)

CombiSet Blood Tubing with Access Flow Reversing Connector (Twister), K022536, is the primary predicate for the CAREline Airless Hemodialysis Blood Tubing Sets (hereafter referred to as "CAREline bloodlines").

Secondary Predicate - Streamline Airless System Set with Locksite® 5.3.2. Needleless Access Site

Streamline Airless System Set with Locksite Needleless Access Site, K080807, is the secondary predicate for the CAREline bloodlines.

{4}------------------------------------------------

K172238 Page 2 of 6

5.4. Device Description

The CAREline bloodlines are intended for use in acute and chronic hemodialysis therapy. The bloodline is part of the hemodialysis extracorporeal circuit, which transports arterial blood from the patient's arterial access (e.g., fistula or catheter) through a hemodialyzer and back to the patient's venous access.

5.4.1. Device Identification

The proposed CAREline bloodlines include the following devices (Table 1):

Table 1: Proposed CAREline Bloodlines

ProductAlternate name
CAREline Airless Hemodialysis Blood Tubing Set withAttached Priming SetCAREline Standard bloodline
CAREline Airless Hemodialysis Blood Tubing Set withAccess Flow Reversing Connector (Twister) and AttachedPriming SetCAREline Twister® bloodline

5.4.2. Environment of Use

The CAREline bloodlines are used in environments where acute and chronic hemodialysis are performed.

5.4.3. Brief Written Description of the Device

The CAREline bloodlines are intended for use in acute and chronic hemodialysis therapy. The bloodline is part of the hemodialysis extracorporeal circuit, which transports arterial blood from the patient's arterial access (e.g., fistula or catheter) through a hemodialyzer and back to the patient's venous access.

5.4.4. Materials of Use

The CAREline bloodlines are classified as an external communicating medical device with prolonged exposure (>24 hours < 30 days) to circulating blood in accordance with ISO 10993- 1:2009, FDA G95-1, and FDA Guidance on Hemodialysis blood tubing sets (2008).

Materials used in the manufacture of the CAREline bloodlines include:

  • Tubing and Components:
    • -Polyvinylchloride (PVC)
    • -Polycarbonate (PC)
    • Polypropylene (PP) -
    • -Polyethylene (PE)
    • Methyl Isobutyl Ketone (TetraMEK) -

{5}------------------------------------------------

  • । Cyclohexanone
  • Pressure Output Device (POD): ●
    • -Methyl Methacrylate Acrylonitrile Butadiene Styrene (MABS)
    • Thermoplastic Elastomer (TPE) made of Medalist® MD |
  • Twister (Twister model only): ●
    • Acrylic-based polymer (Cryo XT) —
    • Polyisoprene Rubber -
    • -Siloxanes
    • Silicones -
    • Dimethyl alcohol -
    • Isopropyl alcohol |

Key Performance Specifications/Characteristic 5.4.5.

The CAREline bloodlines are single-use devices sterilized by ethylene oxide (EO).

The CAREline bloodlines have the following performance characteristics:

  • Transmission of arterial and venous pressures to the hemodialysis (HD) machine's pressure transducer via a Pressure Output Device (POD)
  • . Collection of blood samples and administration of prescribed medications via a needleless access port (T-port)
  • Prevention of stretched or kinked tubing via a dialyzer holder lock sleeve (a ● class I exempt accessory)
  • Measurement of a patient's access flow via the Twister component, which is used to reverse blood flow to and from the arterial and venous vascular access sites.

5.5. Intended Use

The CAREline Airless Hemodialysis Blood Tubing Sets are indicated for use in acute and chronic hemodialysis therapy.

Indications for Use 5.6.

CAREline Twister

The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/hemodialyzer configuration is the responsibility of the physician.

  • The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. ●
  • The Fresenius Access Flow Reversing Connector (AFRC) is for use during

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other.

hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure.

  • Blood Tubing Set is intended to be used with Fresenius Medical Care ● 2008® Series K, K2 and T Hemodialysis Machines.

CAREline Standard

The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/hemodialyzer configuration is the responsibility of the physician.

  • The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. .
  • Blood Tubing Set is intended to be used with Fresenius Medical Care 2008® Series K, K2 and T Hemodialysis Machines.

5.7. Comparison of Technological Characteristics with the Predicate Device

The CAREline bloodlines and the predicate devices share the following characteristics:

  • Similar intended use including similar indications for use
  • Similar design and configuration .
  • . Same scientific technology and principles of operation
  • . Same sterilization method, packaging, and sterility label claims
  • Same materials - Polyvinylchloride (PVC), Polypropylene (PP), Polyethylene (PE)

5.8. Performance Data

Performance testing was conducted in accordance with ISO 8638:2010. Results of the testing listed below support the determination of substantial equivalence.

  • Performance testing of connectors (hemodialyzer, vascular access device, and ancillary components)
  • . Structural integrity testing
  • Biological safety testing (biocompatibility) .
  • . Sterility and pyrogenicity testing
  • . Structural integrity - Pressure and pull testing
  • Pressure testing Access ports and pressure output devices (PODs) ●

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

K172238

  • Endurance testing Conducted with Fresenius Medical Care 2008® Series K, K2, . and T Hemodialysis Machines
  • Clamp testing Ability to occlude and effects of repeated use .
  • . Usability evaluation

5.8.1. Biocompatibility Testing

The following testing was performed to support the biological safety of the CAREline bloodlines:

  • Chemical analysis Extractables and leachables ●
  • Cytotoxicity
  • Sensitization ●
  • Irritation ●
  • Systemic Toxicity ●
  • Material Mediated Pyrogenicity
  • Genotoxicity
  • Hemocompatibility .

A toxicological risk assessment was also performed.

5.8.2. Electrical Safety and Electromagnetic Compatibility (EMC)

No electrical safety and electromagnetic compatibility (EMC) tests were performed.

5.8.3. Software Verification and Validation Testing

No software verification and validation tests were performed.

5.8.4. Mechanical and Acoustic Testing

No mechanical or acoustic tests were performed.

5.8.5. Animal Studies

No animal studies were performed.

5.8.6. Clinical Studies

No clinical studies were performed.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font and is also blue.

5.9. Conclusions

Based on the information and data provided in this Traditional 510(k) submission, the CAREline bloodlines are substantially equivalent in intended use, indications for use, design, principle of operation, technology, materials, and performance to the predicate devices (K022536, CombiSet Blood Tubing with Access Flow Reversing connector (Twister) and K080807, Streamline Airless System Set with Locksite Needleless Access Site). The devices are safe and effective for their intended use.

§ 876.5820 Hemodialysis system and accessories.

(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.