K Number
K172173
Date Cleared
2018-04-06

(261 days)

Product Code
Regulation Number
864.5620
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DiaSpect Tm system is intended for the in vitro quantitative measurement of total hemoglobin in non-anticoagulated capillary whole blood and venous whole blood drawn in K2EDTA or lithium heparin tubes. The DiaSpect Tm system consists of the DiaSpect Tm analyzer and specifically designed disposable cuvettes, the DiaSpect Tm Cuvettes. The device is intended for use in point-of-care settings. The DiaSpect Tm analyzer is only to be used with DiaSpect Tm Cuvettes.

Device Description

The DiaSpect Tm system consists of an analyzer and cuvettes. The DiaSpect Tm analyzer is a spectrophotometric instrument for the total hemoglobin concentration in unaltered human blood. The DiaSpect Tm Cuvette is injection-molded of poly methyl methacrylate (PMMA) and contains a cavity of 10 uL volume. The cavity is empty.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the DiaSpect Tm system meets those criteria, based on the provided document:

Acceptance Criteria and Reported Device Performance

Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Precision/Reproducibility20-Day Precision: Total CV < 7% for external controls.20-Day Precision Summary (Total %CV):- Level 1 (Mean 7.99 g/dL): 1.38%- Level 2 (Mean 12.58 g/dL): 1.09%- Level 3 (Mean 15.82 g/dL): 1.41%All met criteria.
Single-Day Precision: Total CV < 7% for venous whole blood.Single-Day Precision Summary (Total %CV):- Level 1 (Mean 4.87 g/dL): 2.57%- Level 2 (Mean 10.19 g/dL): 2.29%- Level 3 (Mean 13.75 g/dL): 3.19%- Level 4 (Mean 17.46 g/dL): 4.5%- Level 5 (Mean 22.93 g/dL): 6.47%All met criteria.
LinearityCorrelation coefficient > 0.95 and slope of linear equation within 1.0 ± 0.1 compared to reference method.Linearity: Correlation coefficient > 0.95 and slope (0.9837) was within 1.0 ± 0.1.Met criteria.
Detection LimitsLoQ (Limit of Quantitation): To be determined based on specified Total Error.LoB (Limit of Blank): 0.0 g/dLLoD (Limit of Detection): 0.3 g/dLLoQ (Limit of Quantitation): 1.2 g/dL
Analytical Specificity(Interference)Bias < 7% for various potential interferents (pharmaceuticals, endogenous substances like bilirubin, and disease states).All potential interferents (Bilirubin, Cholesterol, Creatinine, Protein, Triglyceride, Urea, Uric Acid, Acetaminophen, Ascorbic Acid, Dopamine, Ibuprofen, Tetracycline, Ferrous Sulfate, Ammonium Ferric Citrate, Ferrous Fumarate, Iron Dextran, Folic Acid, Vitamin B12, Lithium Carbonate, Immunoglobin, Methyldopa, Salicylic Acid, 5x EDTA, Hypochromia, High WBC Count, Polycythemia, Sickle Cell) were tested and passed the acceptance criterion of < 7%.
Method ComparisonDemonstrated substantial equivalence to predicate device (HemoPoint® H2 Hemoglobin Analyzer).Overall Results Summary, Natural and Contrived, Samples All Sites:- EDTA Venous (N=349 clinical, 35 contrived): Slope: 0.9858 (95% CI: 0.969 - 1.002) Intercept: 0.2130 (95% CI: -0.029 - 0.455) R: 0.986- Capillary (N=363 clinical): Slope: 0.9903 (95% CI: 0.963 - 1.018) Intercept: 0.1164 (95% CI: -0.276 - 0.509) R: 0.963- Passing-Bablok Regression: Capillary: y = -0.1198 + 1.011x (R=0.965) EDTA Venous: Y = 0.4867 + 0.9637x (R=0.986)The data supports substantial equivalence.
Matrix ComparisonAgreement between different sample types (EDTA vs. Heparin, Venous vs. Capillary).DiaSpect EDTA vs Heparin: Slope: 0.9981, Intercept: 0.0644, R: 0.9812DiaSpect, Venous vs Capillary: Slope: 0.9702, Intercept: 0.4634, R: 0.8839Agreement between capillary and EDTA samples was favorable, supporting interchangeability and substantial equivalence.
StabilityCuvette shelf-life demonstrating consistent performance.At least 2.5 years shelf life demonstrated with a mean Hb of 12.59 g/dL and total CV of 0.75% for control material across 12 lots over time.
CalibrationDevice is factory calibrated and not user adjustable.Device is factory calibrated and not user adjustable.
Quality ControlSystem includes control solutions for quality control.The system includes 3 levels of control solutions (DiaSpect Control HBT) with known hemoglobin concentrations.

Study Details

  1. Sample sizes used for the test set and data provenance:

    • Precision Studies:
      • 20-Day Precision: 240 measurements per level (3 levels), across 3 sites, with 6 operators, using 3 lots and 3 instruments.
      • Single-Day Precision: 54 measurements per level (5 levels), at a single site, using 3 instruments, 3 lots, and 3 self-trained operators.
    • Linearity Study: 11 hemoglobin levels, tested in triplicate on one DiaSpect Tm analyzer.
    • Detection Limit Studies:
      • LoB: Plasma samples from 5 individual whole blood donors, quadruplicate tests (x4) for 3 days on one meter using 2 cuvette lots.
      • LoD: 4 whole blood samples (K2EDTA), 60 tests each on one device using 2 cuvette lots for 3 days.
      • LoQ: 9 unique donor whole blood samples, at least 10 tests per lot for 3 days using 2 cuvette lots on one analyzer.
    • Analytical Specificity (Interference) Studies: Whole venous blood (K2EDTA) at 3 hemoglobin concentrations (11.0, 14.0, 18.0 g/dL), spiked with potential interferent at specified concentrations. For disease conditions, K2EDTA venous blood specimens from up to five (5) donors with specific conditions were collected.
    • Method Comparison Studies:
      • Total of 399 subjects across four point-of-care clinical sites.
      • Clinical Samples: 364 male and female subjects (9 months to 89 years of age) provided:
        • 363 capillary samples (ranges 8.5 to 20.1 g/dL)
        • 349 K2EDTA venous samples (ranges 6.5 to 19.9 g/dL)
        • 120 lithium heparin venous samples (ranges 10.4 to 20.0 g/dL)
      • Contrived Samples: An additional 35 contrived EDTA venous samples used to challenge the full measuring range.
      • Data Provenance: The studies were conducted at multiple clinical sites, suggesting a U.S. context (given FDA submission). The samples were described as "natural and manipulated K2EDTA venous blood" and "natural and contrived samples," implying a mix of retrospective (manipulated/contrived) and prospective (clinical site collections) data. IRB approval was obtained prior to initiating the clinical studies.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the Method Comparison study, the predicate device, HemoPoint® H2 Hemoglobin Analyzer (K081719), was used as the comparative method, effectively serving as the "ground truth" reference. The performance of the predicate device is established through its own FDA clearance.
    • For the Analytical Specificity studies, the predicate device HemoPoint® H2 was also used as the reference method.
    • For Linearity, a HemoPoint® H2 analyzer was also used as the reference method.
    • The document does not explicitly state the qualifications of the individuals who performed the tests on the predicate device or established its performance. However, for the method comparison study, "untrained, intended use operators" (full-time employees of Certificate of Waiver test sites who normally perform phlebotomy and/or patient testing) operated the DiaSpect Tm system, and the HemoPoint H2 was used as the comparative method alongside these measurements.
  3. Adjudication method for the test set:

    • Not applicable in this context. The study compares the new device's measurements against a recognized predicate device and its own inherent precision and analytical performance, rather than requiring human expert adjudication of results.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, this is not an MRMC study. This device is an automated hemoglobin measurement system, not an imaging device for which human readers would interpret results, nor does it incorporate AI assistance for human interpretation. The study is a comparison of a new device against a predicate device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The DiaSpect Tm system is an automated spectrophotometric instrument. Its performance is inherently "algorithm only" in terms of its ability to measure hemoglobin concentration. The precision, linearity, detection limits, and interference studies are all evaluations of this standalone performance. The method comparison study evaluates its standalone performance against another standalone device.
  6. The type of ground truth used:

    • Comparative method / Reference Standard: The predicate device, HemoPoint® H2 Hemoglobin Analyzer (K081719), served as the primary reference method for method comparison and some analytical specificity studies. Its results were considered the "ground truth" for comparison.
    • Defined Samples: For precision and linearity, samples with pre-defined or manipulated hemoglobin levels were used. For detection limits, plasma "blank" and low-concentration whole blood samples were prepared.
    • Clinical Samples: Natural patient samples (capillary, K2EDTA venous, lithium heparin venous) spanning a range of hemoglobin values were collected and tested.
  7. The sample size for the training set:

    • This document describes a pre-market notification (510(k)) for a medical device that measures hemoglobin. It does not refer to a machine learning or AI algorithm that requires a "training set" in the conventional sense. The device's underlying "algorithm" is spectrophometric measurement, calibrated at the factory. Therefore, there's no "training set" as one would discuss for, e.g., a deep learning model.
  8. How the ground truth for the training set was established:

    • As there is no "training set" in the AI/ML sense, this question is not applicable. The device relies on principles of spectrophotometry and is factory calibrated. Quality control materials and established reference methods are used to verify its accuracy and performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 06, 2018

EKF-diagnostic GmbH Mick Fenton Head of Global QA/RA Ebendorfer Chaussee 3 39179 Barleben Germany

Re: K172173

Trade/Device Name: DiaSpect Tm system Regulation Number: 21 CFR 864.5620 Regulation Name: Automated hemoglobin system Regulatory Class: Class II Product Code: GKR Dated: February 20, 2018 Received: March 05, 2018

Dear Mick Fenton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172173

Device Name DiaSpect Tm system

The DiaSpect Tm system is intended for the in vitro quantitative measurement of total hemoglobin in non-anticoagulated capillary whole blood and venous whole blood drawn in K2EDTA or lithium heparin tubes. The DiaSpect Tm system consists of the DiaSpect Tm analyzer and specifically designed disposable cuvettes, the DiaSpect Tm Cuvettes. The device is intended for use in point-of-care settings. The DiaSpect Tm analyzer is only to be used with DiaSpect Tm Cuvettes.

Type of Use (Select one or both, as applicable)
☑Prescription Use (Part 21 CFR 801 Subpart D) ☐Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to a collection of"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

510k Number

K172173

Introduction

In accordance with the requirements of 21CFR807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

I. Submitter

EKF-diagnostic GmbH Ebendorfer Chaussee 3 39179 Barleben Germany

Tel.: +49 39203 511 0 Fax: +49 39203 511 171

Contact Person: Mick Fenton Phone: +44 0290705274 Email: mickfenton@ekfdiagnostics.com

Secondary Contact: Karen Golomb Phone: +1 (830) 249-072 Fax: +1 (830) 249-0851 Email: karengolomb@ekfdiagnostics.com

Date prepared: 10th July 2017

II. Device Name

Proprietary names:DiaSpect Tm system
Common names:As above
Classification:21 CFR 864.5620 – Automated hemoglobin systemClass IIHematology

Product Code: GKR

III. Predicate Device

The DiaSpect Tm system is substantially equivalent to the currently marketed HemoPoint® H2 Measurement System (K081719).

{4}------------------------------------------------

IV. Device Description

The DiaSpect Tm system consists of an analyzer and cuvettes. The DiaSpect Tm analyzer is a spectrophotometric instrument for the total hemoglobin concentration in unaltered human blood. The DiaSpect Tm Cuvette is injection-molded of poly methyl methacrylate (PMMA) and contains a cavity of 10 uL volume. The cavity is empty.

V. Indications for Use

The DiaSpect Tm system is intended for the in vitro quantitative measurement of total hemoglobin in non-anticoagulated capillary whole blood and venous whole blood drawn in K2EDTA or lithium heparin tubes. The DiaSpect Tm system consists of the DiaSpect Tm analyzer and specifically designed disposable cuvettes, the DiaSpect Tm Cuvettes. The device is intended for use in point-of-care settings. The DiaSpect Tm analyzer is only to be used with DiaSpect Tm Cuvettes.

Rx only.

VI. Comparison with Predicate

    1. Predicate device name(s): HemoPoint® H2 Measurement System
    1. Predicate 510(k) number(s): K081719

Similarities compared to the chosen (FDA cleared; marketed) predicate device (K081719)

PerformancePredicate DeviceHemoPoint H2 MeasurementSystem(K081719)Candidate DeviceDiaSpect Tm system(K172173)
Indications for useDetermine hemoglobin content ofwhole bloodSame
AnalyteHemoglobinSame
Sample preparation(pre-treatment)NoneSame
AutomationFully automated assaySame
CalibrationprocedureFactory calibratedSame
Built in QualityControlAuto self-check betweenmeasurementsSame
Differences compared to the chosen (FDA cleared; marketed) predicate device (K081719)
PerformancePredicate DeviceHemoPoint H2 MeasurementSystem(K081719)Candidate DeviceDiaSpect Tm system(K172173)
Intended UseThe HemoPoint H2 system isintended for the quantitativedetermination of hemoglobin(Hgb) in whole blood of adults,infants, and children in aprofessional point-of-caresetting. It consists of a dedicatedphotometer and individual,single-use microcuvettes filledwith reagents.The DiaSpect Tm system isintended for the in vitroquantitative measurement of totalhemoglobin in non-anticoagulated capillary wholeblood and venous whole blooddrawn in K2EDTA or lithiumheparin tubes. The DiaSpect Tmsystem consists of the DiaSpectTm analyzer and specificallydesigned disposable cuvettes, theDiaSpect Tm Cuvettes. Thedevice is intended for use inpoint-of-care settings. TheDiaSpect Tm analyzer is only tobe used with DiaSpect TmCuvettes. RX only
Method of detection(Test methodology)Azide methemoglobinOptical absorbance
Sample typeCapillary, arterial or venousCapillary or venous
Sample volume8 µL< 10μL
Cuvette reagentcomponentsAzide methemoglobin reagentNone
Cuvette storage15 - 30°C0 – 50°C
Cuvettecomponents2 canisters of 50 or 4 canisters of505 packs of 100
Control KitcomponentsTwo concentration levels ofcontrols (3 vials of each)Three concentration levels ofcontrols (1 vial of each)
Quality ControlRequires two buffer basedcontrols to validate thecalibrationRequires three buffer basedcontrols to validate thecalibration
MeasurementRange0.0-25.6 g/dL1.2 - 25.5 g/dL
Measuring Time30-60 seconds1 second

{5}------------------------------------------------

he chosen (EDA alaarad; marketad) prodicata daviga (kf)

VII. Performance Characteristics

1. Analytical Performance

a. Precision/Reproducibility:

Precision was determined in accordance with CLSI EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods". Two precision studies were performed at the clinical sites per protocol EKF-PS-01. One study was a 20-Day multi-site study using

{6}------------------------------------------------

three levels of external control material, and the other study was a single-day, single-site precision study with 5 levels of natural and manipulated K2EDTA venous blood.

20-Day Precision

The 20-Day precision was performed by running external controls in duplicate twice per day by two operators at each of three sites (a 20 x 2 x 2 x 3 format with 6 operators, 3 lots, 3 instruments and 3 sites). Lots and instruments were nested with sites. The study was performed by six of the same operators who performed the Method Comparison studies. The 20-Day precision data met the criteria of <7% CV.

SampleNMeanWithin-Run (SD,%CV)Between-Run (SD,%CV)Between-Day(SD,%CV)Between-Operator(SD, %CV)Between-Site(SD, %CV)Total (SD,%CV)
Level 12407.99(0.085,1.06%)(0.05, 0.59%)(0.04,0.47%)(0, 0%)(0.04,0.45%)(0.11,1.38%)
Level 224012.58(0.11,0.88%)(0.05, 0.38%)(0.03,0.22%)(0, 0%)(0.06,0.47%)(0.14,1.09%)
Level 324015.82(0.15,0.92%)(0.06, 0.36%)(0.04,0.27%)(0, 0%)(0.15,0.97%)(0.22,1.41%)

20-Day Precision Summary

Single-Day Precision

Single-Day precision was performed at a single site using K2EDTA venous whole blood. Five donors provided 5 samples, some of which were manipulated by removing plasma to increase hemoglobin, or diluting with the same donor's plasma to decrease hemoglobin. The study was performed at a new single site using 3 instruments, 3 lots and 3 untrained operators who had no prior experience running the DiaSpect Tm, and self-trained on the system. Each of 3 operators ran duplicate tests on all 5 levels with 3 lots of cuvettes in each of 3 instruments, providing 54 measurements for each level. The higher %CV found for Level 5 may be due to the increased viscosity of the abnormally high hematocrit for the samples. The Single-Day precision study met the criteria of <7% CV.

Single-Day Precision Summary

SampleNMeanWithin-Run (SD,%CV)Between-lot (SD,%CV)Between-Operator(SD,%CV)Between-Analyzer (SD,%CV)Total (SD,%CV)
Level 1544.87(0.05,(0.04, 0.82%)(0.02,(0.11,(0.13,
0.97%)0.42%)2.19%)2.57%)
Level 25410.19(0.19,(0, 0%)(0.06, 0.6%)(0.12,(0.23,
1.87%)1.17%)2.29%)
Level 35413.75(0.38,(0.07, 0.48%)(0.21, 1.5%)(0, 0%)(0.44,
2.77%)3.19%)

{7}------------------------------------------------

Level 45417.46(0.69,3.93%)(0, 0%)(0.08,0.48%)(0.37,2.14%)(0.79, 4.5%)
Level 55422.93(1.45,6.33%)(0, 0%)(0.3, 1.32%)(0, 0%)(1.48,6.47%)

b. Linearity

A linearity study was conducted based on CLSI EP6-A. Specimens having eleven hemoglobin levels were prepared from venous whole blood. Each level was tested in triplicate on one DiaSpect Tm analyzer. Average results were plotted against results obtained from a HemoPoint® H2 analyzer, which was used as the reference method. Summarized results are presented in the following diagram. The correlation coefficient was greater than 0.95 and the slope of the linear equation was within 1.0±0.1 which met acceptance criteria.

Image /page/7/Figure/3 description: The image is a scatter plot titled "Target Hgb Conc vs Diaspect Values". The x-axis is labeled "DiaSpect Values" and ranges from -5 to 30. The y-axis is labeled "Target Hgb Conc" and ranges from -5 to 30. The plot shows a linear relationship between the two variables, with a linear fit equation of (0.1365 +0.9837x). There is also a 95% confidence interval and a 95% prediction interval shown on the plot.

d. Detection Limit:

LoB

Plasma samples were obtained from (5) individual whole blood donors. The five (5) plasma "blank" samples were tested for 3 days on one (1) DiaSpect Tm meter in quadruplicate (x4) using two (2) different lots of DiaSpect Tm cuvettes.

LoD

Four (4) whole blood samples (K2EDTA) were collected from different donors. Four (4) independent samples covering Hgb concentrations 0.1 to 0.4 g/dL were prepared. Each sample was tested sixty (60) times on (1) DiaSpect Tm device using two (2) different lots of DiaSpect Tm cuvettes for 3 days.

{8}------------------------------------------------

LoQ

Nine (9) unique donor whole blood samples were used to prepare concentrations from 0.7-1.5 g/dL. Each sample was tested using two lots of DiaSpect Tm cuvettes on one DiaSpect Tm analyzer at least ten times for each lot for three days to achieve variability. The mean and SD for each sample concentration for each cuvette lot was determined and the bias for each of the nine levels calculated. LoQ was determined from the specified Total Error to be 1.2g\dL.

SensitivityConcentration
LoB0.0 g/dL
LoD0.3 g/dL
LoQ1.2 g/dL

e. Analytical specificity:

Interference and cross-reactivity studies were performed in accordance with the CLSI guidance EP7-A2 "Interference testing in clinical chemistry".

Whole venous blood was collected with K2EDTA tubes and was manipulated to create 3 hemoglobin concentrations of 11.0, 14.0, and 18.0 g/dL. Each hemoglobin concentration was spiked with potential interferent at the test concentrations listed below.

For Disease Conditions K2EDTA venous blood specimens from up to five (5) donors with the following conditions were collected: Polycythemia, Hypochromia, High WBC Count and Sickle Cell. The predicate device HemoPoint(H2) was used as the reference method to obtain the Hgb concentration. Each sample was tested in duplicate on the H2 and five times on the DiaSpect Tm device. The predicate HemoPoint(H2) average was used to calculate the bias against each individual DiaSpect Tm result.

Potential interferentTest ConcentrationDesignacceptancecriteriaResult
Bilirubin20.0 mg/dL< 7%Passed
Cholesterol500 mg/dL< 7%Passed
Creatinine5 mg/dL< 7%Passed
Protein12 mg/dL< 7%Passed
Triglyceride1000 mg/dL< 7%Passed
Urea258 mg/dL< 7%Passed
Uric Acid24 mg/dL< 7%Passed
Acetaminophen2 mg/dL< 7%Passed
Ascorbic Acid6 mg/dL< 7%Passed
Dopamine0.1 mg/dL< 7%Passed
Ibuprofen55 mg/dL< 7%Passed
Tetracycline1.5 mg/dL< 7%Passed
Ferrous Sulfate22 mg/dL< 7%Passed
Ammonium Ferric Citrate30 mg/dL< 7%Passed
Ferrous Fumarate30 mg/dL< 7%Passed

{9}------------------------------------------------

Iron Dextran284 mg/dL< 7%Passed
Folic Acid1000 ng/mL< 7%Passed
Vitamin B122500 ng/mL< 7%Passed
Lithium Carbonate23 mg/dL< 7%Passed
Immunoglobin500 mg/dL< 7%Passed
Methyldopa1.7 mg/dL< 7%Passed
Salicylic Acid100 mg/dL< 7%Passed
5x EDTATube filled to 1/5 volume< 7%Passed
HypochromiaDisease state< 7%Passed
High WBC CountDisease state< 7%Passed
PolycythemiaDisease state< 7%Passed
Sickle CellDisease state< 7%Passed

All potential interferants tested within the acceptance criteria.

f. Assay cut-off:

Not applicable

  • g. Stability
    Cuvettes from 12 different lots produced between December 2008 and July 2011 were tested in July 2011. Four cuvettes from each lot were tested by measuring the medium level of the DiaSpect Control HB, target value 12.6 g/dL, on the DiaSpect device.

The mean value of the measurements from all lots is 12.59 g/dL, with a total CV 0.75%. The results of measurements are summarized in the following Table.

Cuvette LotProduction DateMean Hb g/dL
D0802052008-12-1812.53
D0900012009-01-2012.40
D0900022009-05-2612.50
D0900072009-08-3112.50
D0900202009-12-0812.60
D1000092010-03-3112.70
D1000192010-05-0312.65
D1000342010-08-0512.70
D1000452010-11-2812.60
D1100072011-02-2812.58
D1100132011-05-0212.70
D1100222011-07-1112.60
Mean Hb g/dL12.59

SD 0.09 0.75 CV % Min - Max, Hb g/dL - 12.4 - 12.7

Conclusion: The DiaSpect Tm cuvette has a shelf life of at least 2.5 years.

{10}------------------------------------------------

h. Calibration

The DiaSpect Tm analyzer is factory calibrated and is not user adjustable.

i. Quality Control

The DiaSpect Tm system includes 3 levels of control solutions with known hemoglobin concentration. The DiaSpect Control HBT is produced in three concentrations that correspond to three levels of human hemoglobin. Each vial contains 1.9 mL of a solution of a red dye (Rhodamine) in purified water. The reagent does not contain any material of human or animal origin.

The DiaSpect Control HBT is 510(k) exempt.

2. Method Comparison Studies

Method comparison:

Method comparison studies were performed across four sites to support the substantial equivalence of the DiaSpect Tm Hemoglobin System with the predicate device. Finger stick capillary, K2EDTA and lithium heparin venous whole blood specimens from a total of 399 subjects across four point-of-care clinical sites were tested on-site with both the DiaSpect Tm system (in singlicate) and in the HemoPoint® H2 Hemoglobin Analyzer (comparative method, in duplicate). IRB approval was obtained prior to initiating the studies.

A total of 364 male and female subjects, ranging from 9 months to 89 years of age, provided 363 capillary samples ranging from 8.5 to 20.1 g/dL, 349 EDTA venous samples ranging from 6.5 to 19.9 g/dL and 120 heparin venous samples ranging from 10.4 to 20.0 g/dL when tested in the DiaSpect Tm system. An additional 35 contrived EDTA venous samples were divided between three sites and tested to challenge the full measuring range of the DiaSpect Tm system. Of the 349 EDTA venous samples, five samples with hemoglobin values of 6-7 g/dL were tested in comparison with the predicate at one additional internal site to demonstrate the performance of the device around these levels.

Testing was performed over a 5-week period by 9 operators using four DiaSpect Tm analyzers and three lots of DiaSpect Tm Cuvettes. A single lot of tri-level DiaSpect Control HBT external controls were run each day specimens were tested. Clinical Study sites were required to have a current Certificate of Waiver and all operators were untrained, intended use operators. Operators were full time employees of the Certificate of Waiver test sites used in the study, who normally perform phlebotomy and/or participate in patient testing. Capillary whole blood and K2EDTA venous whole blood were tested at the four sites by 9 operators. Lithium heparin was tested at one site by two operators. The results for all natural and contrived samples combined are summarized in Table below.

{11}------------------------------------------------

Site #BloodTypeN(Clinical)N(Contrived)Slope95% CISlopeIntercept95% CIInterceptR
1EDTAVenous100120.98050.947 -1.0140.56070.099 -1.0220.984
Capillary101N/A0.96280.928 -1.0080.3985-0.167 -0.9640.977
2EDTAVenous120111.00670.980 -1.032-0.1795-0.558 -0.1990.989
HeparinVenous120111.02291.004 -1.042-0.3606-0.643 -0.0790.994
Capillary119N/A1.01290.951 -1.0750.0958-0.989 -0.7980.943
3EDTAVenous119120.99440.963 -1.011-0.0547-0.295 -0.3820.992
Capillary120N/A0.98890.952 -1.0260.0772-0.454 -0.6980.978
4EDTAVenous5N/A0.94870.642 -1.2550.7755-3.518 -5.0690.985
Capillary23N/A1.01450.743 -1.286-0.3496-4.109 -3.4100.861
TOTALEDTAVenous349*350.98580.969 -1.0020.2130-0.029 -0.4550.986
Capillary363N/A0.99030.963 -1.0180.1164-0.276 -0.5090.963

Table Overall Results Summary, Natural and Contrived, Samples All Sites

  • See introductory study description.

Passing-Bablok Regression All Sites

Passing-BablokRegressionR
All resultsCapillary$y = -0.1198 + 1.011x$0.965
EDTAVenous$Y = 0.4867 + 0.9637x$0.986

Matrix Comparison

120 paired natural EDTA and heparin samples, ranging from 10.4 - 19.9 g/dL (EDTA) or 10.4 - 20.0 g/dL (heparin) Hgb, agreed closely and can be used interchangeably in the DiaSpect Tm system. Agreement between capillary and EDTA samples in the DiaSpect was favourable to that of the predicate system, and thus the DiaSpect Tm and predicate are substantially equivalent.

ComparisonSlopeInterceptR
DiaSpect EDTA vs Heparin0.99810.06440.9812
DiaSpect, Venous vs Capillary0.97020.46340.8839
HemoPoint (predicate), Venous vs Capillary0.89641.46950.8926

{12}------------------------------------------------

Expected Values/reference Range

Reference interval calculations were performed to verify that the results obtained for a normal reference population using the DiaSpect Tm Hemoglobin Measuring System are equivalent to published ranges (Dacie and Lewis, Practical Haematology, Twelfth Edition, Elsevier Limited 2017). Prospective fingerstick and venous EDTA DiaSpect Tm values from normal, healthy individuals enrolled in the method comparison studies were pooled and used. The interval between the 2.5th and 97.5th percentiles was taken as the reference interval. Results for subjects with no medications, capillary and EDTA venous samples combined.

PopulationAge RangeCited Reference Range
Adult Male$\geq$ 22 years13.0 - 17.0 g/dL
Adult Female$\geq$ 22 years12.0 - 15.0 g/dL
Child/Adolescent> 2 years to 21 years11.0 - 15.5 g/dL
Infant1 month to 2 years9.4 - 16.5 g/dL

VII. Conclusion

The DiaSpect Tm system data presented and provided is complete and supports the basis for substantial equivalence to the predicate device.

§ 864.5620 Automated hemoglobin system.

(a)
Identification. An automated hemoglobin system is a fully automated or semi-automated device which may or may not be part of a larger system. The generic type of device consists of the reagents, calibrators, controls, and instrumentation used to determine the hemoglobin content of human blood.(b)
Classification. Class II (performance standards).