K Number
K172078
Date Cleared
2018-03-30

(263 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

An enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of anti-RNA POL III IgG antibodies in human serum as an aid in diagnosis of systemic scleroderma) in conjunction with other laboratory and clinical findings.

Device Description

An enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of anti-RNA POL III IgG antibodies in human serum as an aid in diagnosis (scleroderma) in conjunction with other laboratory and clinical findings. This test is performed as a solid phase immunoassay. Microwells are coated with recombinant RNA POLIII antigen. Controls, calibrators and patient sera are incubated in the antigen coated wells to allow specific antibodies present in the RNA POL III antigen. Bound antibodies are detected by adding an enzyme labeled anti-human IgG conjugate. Specific enzyme substrate (TMB) is then added and the presence of antibodies is detected by a spectrophotometer at 450 nm. Results are expressed in ELISA units per milliliter (EU/ml) and reported as positive or negative.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a formal table format with target performance values. Instead, it presents various performance characteristics and their corresponding results. I will infer the acceptance criteria from the reported performance, implying that these results met the manufacturer's internal criteria for substantial equivalence.

Performance CharacteristicAcceptance Criterion (Inferred)Reported Device Performance
Method Comparison (Qualitative - Borderline considered positive)Sufficient agreement with predicate device (INOVA QUANTA Lite™ RNA POL III ELISA) for Positive, Negative, and Overall.Positive Percent Agreement: 96.2% (95% Cl 85.7% - 99.3%)Negative Percent Agreement: 99.4% (95% CI 97.8% - 99.9%)Overall Agreement: 99.0% (95% Cl 97.8% - 99.7%)
Method Comparison (Qualitative - Borderline considered negative)Sufficient agreement with predicate device (INOVA QUANTA Lite™ RNA POL III ELISA) for Positive, Negative, and Overall.Positive Percent Agreement: 78.8% (95% Cl 64.9% - 88.4%)Negative Percent Agreement: 99.7% (95% Cl 98.2% - 100.0%)Overall Agreement: 97.1% (95% Cl 95.0% - 98.3%)
Cross-Reactivity (Specificity)Low positivity rate in various autoimmune and infectious disease control populations and cancer patients. Minimal reactivity outside of Systemic Sclerosis (SSc).Percent Positive for various control populations detailed in the "Cross Reactivity" table (e.g., SLE: 0.0%, Sjögren's: 0.0%, HSV-1: 0.0%, etc.). Some minor reactivity noted (e.g., Myositis: 2.5%, Rheumatoid arthritis: 2.5%, Hepatitis C: 5.0%, Breast Cancer: 14.3%, Psoriasis: 11.1%).
Precision (Total Imprecision CV%)Acceptable variability (CV%) across different concentrations.Ranged from 4.4% to 7.8% (for EU/ml values from 10.0 to 159.0)
Reproducibility (Qualitative Agreement)High qualitative agreement, especially for samples not near the cutoff.Cutoff specimen: 60% agreement~20% above cutoff: 98% agreementAll other specimens: 100% agreement
Limit of Detection (LoD)Low enough to detect relevant low levels of analyte.3.2 EU/ml
Linearity and RecoveryDemonstrated linearity and acceptable recovery across the assay's reportable range.Test Range 2.5 to 33.5: Slope 1.02, R^2 0.9943, Recovery 96% to 117%Test Range 8.0 to 66.1: Slope 0.98, R^2 0.9987, Recovery 97% to 109%Test Range 31.8 to 169.5: Slope 0.99, R^2 0.9890, Recovery 91% to 103%
InterferenceNo significant interference from common substances at specified levels.No significant interference demonstrated for a list of 16 substances.
Clinical Sensitivity (Systemic Sclerosis)Clinically relevant sensitivity for Systemic Sclerosis.23.1% (95% C.I. 18.4-28.6%)
Clinical Specificity (Systemic Sclerosis)High clinical specificity for Systemic Sclerosis.98.2% (95% C.I. 96.6-99.1%)

2. Sample sizes used for the test set and the data provenance:

  • Method Comparison Test Set: 413 samples (52 positive, 361 negative, based on the predicate device).
    • Data Provenance: "well-characterized systemic sclerosis subjects and disease controls." The country of origin is not specified but implicitly North America (given the US FDA submission). Retrospective, as these were "well-characterized" samples.
  • Cross-Reactivity Test Set:
    • Systemic Sclerosis (SSc): 281 samples
    • Diffuse cutaneous SSc (dcSSc): 105 samples
    • Limited cutaneous SSc (lcSSc): 176 samples
    • Various autoimmune and infectious disease controls, and cancer patients: 549 samples in total across 22 different categories (sample sizes per category detailed in the table, e.g., SLE: 40, Sjögren's: 41, etc.).
    • Data Provenance: Clinical samples. Country of origin not specified. Retrospective, as they were "sets of clinical samples" collected and tested.
  • Clinical Study Test Set:
    • 281 systemic sclerosis samples
    • 549 autoimmune and infectious disease controls (the same breakdown as the Cross-Reactivity section).
    • Data Provenance: Clinical samples. Country of origin not specified. Retrospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not specify the number of experts or their qualifications for establishing the ground truth for any of the test sets. For the "Method Comparison," it refers to "well-characterized systemic sclerosis subjects and disease controls," implying a clinical diagnosis as the ground truth, likely established by medical professionals. For "Cross Reactivity" and "Clinical Study," it refers to "clinical samples" and "autoimmune and infectious disease controls," which implicitly means ground truth was based on established clinical diagnoses.

4. Adjudication method for the test set:

Not explicitly stated. Given that the ground truth for method comparison relied on a predicate device and for cross-reactivity/clinical study on "clinical samples" and "disease controls," it's highly probable that the ground truth was established by clinical diagnosis, which often involves a consensus among treating physicians or established diagnostic criteria rather than a dedicated adjudication process for the purpose of this study.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an in-vitro diagnostic (IVD) ELISA test, not an AI-powered image analysis or diagnostic support tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable to this type of device and was not performed or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This device is a standalone in-vitro diagnostic test. Its performance metrics (sensitivity, specificity, agreement with predicate, precision, etc.) are all based on its own intrinsic performance, without any human-in-the-loop interpretation impacting these reported analytical and clinical performance characteristics. So, yes, a standalone performance was done for the device itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Method Comparison Test Set: The ground truth was based on the performance of a legally marketed predicate device, the INOVA QUANTA Lite™ RNA POL III ELISA, and "well-characterized systemic sclerosis subjects and disease controls," implying clinical diagnosis.
  • Cross-Reactivity and Clinical Study Test Sets: The ground truth was based on the clinical diagnosis of the patients from whom the samples were obtained ("Systemic sclerosis," "Systemic lupus erythematosus," "Sjögren's syndrome," etc.), and the characterization of "autoimmune and infectious disease controls." This points to clinical diagnosis (and possibly expert medical consensus) as the ground truth.

8. The sample size for the training set:

The document does not explicitly mention a "training set" in the context of device development. Given that this is an ELISA kit, which is a chemical assay, it does not typically involve machine learning algorithms that require distinct training and test sets in the same way an AI-driven device would. The development samples and internal validation studies would likely fall under what could be considered "training" or optimization, but a specific "training set" size is not reported as it would be for an AI model.

9. How the ground truth for the training set was established:

As mentioned above, a formal "training set" as understood in AI/ML contexts is not directly applicable here. The development and optimization of the ELISA assay would typically involve extensive testing with known positive and negative controls, and clinical samples previously characterized by established diagnostic methods (e.g., other validated tests, clinical diagnosis) to refine the assay parameters (e.g., antibody concentrations, incubation times, cutoff values). This characterization would rely on standard clinical diagnostic practices and existing reference methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

IMMCO Diagnostics, Inc. Mr. Kevin Lawson Chief Regulatory Officer 60 Pineview Dr. Buffalo, New York 14228 March 30, 2018

Re: K172078

Trade/Device Name: ImmuLisa Enhanced RNA POL III Antibody ELISA Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: NYO Dated: July 7, 2017 Received: July 10, 2017

Dear Kevin Lawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Kelly Oliner-S" in a large, sans-serif font. The text is black and stands out against a light blue background. The name is written in a clear and legible manner.

For

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172078

Device Name ImmuLisa™ Enhanced RNA POL III Antibody ELISA

Indications for Use (Describe)

An enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of anti-RNA POL III IgG antibodies in human serum as an aid in diagnosis of systemic scleroderma) in conjunction with other laboratory and clinical findings.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green circles on the left, followed by the word "immco" in blue, with a small circle above the "o". Below "immco" is the word "DIAGNOSTICS" in green, and below that is the text "A Trinity Biotech Company" in a smaller font size.

ImmuLisa™ Enhanced RNA POL III Antibody ELISA 510(k) Summary

Submitter:Immco Diagnostics, Inc.
Address:60 Pineview Dr., Buffalo, NY 14228
Phone Number:716-691-0091 ext. 110
Contact:Kevin Lawson
Summary Prepared:3-22-2018
Device Name:ImmuLisa Enhanced™ Anti-RNA POL III Antibody ELISA
Common Name:Anti-RNA POL III Antibody ELISA
Product Code:Autoantibodies, anti-ribonucleic acid polymerase (RNAP) III antibody [NYO]
Substantially Equivalent to:INOVA QUANTA Lite™ RNA POL III ELISA

General Description: Antinuclear antibodies (ANA) occur in sera of patients with various connective tissue disorders such as systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), systemic sclerosis / scleroderma (SSc), polymyositis and Siögren's syndrome. These ANA are directed against nucleding DNA, nucleohistone and various extractable nuclear antigens (ENA) such as RNP, Sm, SS-A (Ro), SS-B (La), Centromerase I), Jo-1 and RNA polymerase III (RNA POL III). These antigens are macromolecular complexes of protein and RNA.

Systemic sclerosis is associated with many autoantibodies. The antibody has been observed in a large percentage of patients with limited cutaneous systemic sclerosis (ICSC) or CREST synaud's phenomenon, esophageal motility abnormalities, sclerodactylia and telangietasia) variant of systemic sclerosis or scleroderma. The anti-Scl-70 antibodies are associated with systemic sclerosis with the risk of diffuse skin involvement. The anti-RNA Pol III antibodies are also predictive of diffuse skin involvement, as well as a predictor for the development of hypertensive renal crisis. Anti-RNA Pol III antibodies are frequently found without the presence of anti-Scl-70 or anti-centromere antibodies in patients with systemic sclerosis.

This test is performed as a solid phase immunoassay. Microwells are coated with recombinant RNA POLIII antigen. Controls, calibrators and patient sera are incubated in the antigen coated wells to allow specific antibodies present in the RNA POL III antigen. Bound antibodies are detected by adding an enzyme labeled anti-human IgG conjugate. Specific enzyme substrate (TMB) is then added and the presence of antibodies is detected by a spectrophotometer at 450 nm. Results are expressed in ELISA units per milliliter (EU/ml) and reported as positive or negative.

Intended Use: An enzyme linked immunoassay (EUSA) for the qualitative detection of anti-RNA POL III IgG antibodies in human serum as an aid in diagnosis (scleroderma) in conjunction with other laboratory and clinical findings.

Similarities and Differences: Both kits use recombinant RNA POL III coated on 96 well plates to detect IgG RNA POL III antibody with HRP anti-human lgG conjugate and TMB substrate. The IMMCO kit utilizes a 5 point calibrator curve with a borderline/indeterminate range of 20-25 EU/ml; while the INOVA kit uses a single as a calibrator and has a cutoff of 20 units with no borderline range.

Non-clinical Tests:

Method Comparison: Both kits were tested with well-characterized systemic sclerosis subjects and disease controls. Semiquantitative results are presented below. Qualitative results are the borderline considered positive results.

Other RNA POL III Ab ELISA
PositiveNegativeTotal
IMMCORNA POL III AbELISAPositive41142
Borderline9110
Negative2359361
Total52361413

Borderline considered positive

60 Pineview Drive Buffalo, NY 14228-2120 Toll free (800) 537-8378

www.immco.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green dots on the left, followed by the word "immco" in blue, and the word "DIAGNOSTICS" in green. Below that, the words "A Trinity Biotech Company" are written in a smaller font.

Positive Percent Agreement: 96.2% (95% Cl 85.7% - 99.3%) Negative Percent Agreement: 99.4% (95% CI 97.8% - 99.9%) 99.0% (95% Cl 97.8% - 99.7%) Overall Agreement:

Borderline considered negative

Positive Percent Agreement: 78.8% (95% Cl 64.9% - 88.4%) Negative Percent Agreement: 99.7% (95% Cl 98.2% - 100.0%)

Overall Agreement: 97.1% (95% Cl 95.0% - 98.3%)

Cross Reactivity

Sets of clinical samples were tested to on the Immulisa™ RNA POLIII EUSA. Populations included systemic sclerosis sera as well as autoimmune and infectious disease controls.

Populationnn Positive*% Positive
Systemic sclerosis (SSC)2816523.1%
Diffuse cutaneous SSc (dcSSC)1056158.1%
Limited cutaneous SSc (IcSSC)17642.3%
Systemic lupus erythematosus4000.0%
Sjögren's syndrome4100.0%
Myositis4012.5%
Rheumatoid arthritis4012.5%
HSV-12000.0%
HSV-22000.0%
Lyme2000.0%
Toxoplasmosis2000.0%
CMV2000.0%
Rubella2000.0%
Syphilis2000.0%
Hepatitis C2015.0%
Autoimmune hepatitis1300.0%
Acute renal failure2900.0%
Breast Cancer7114.3%
Colorectal Cancer1000.0%
Ovarian Cancer1000.0%
Pancreatic Cancer1000.0%
Esophageal reflux3113.2%
Hypertension2713.7%
MCTD1915.3%
Morphea200.0%
Pulmonary Hypertension2900.0%
Psoriasis27311.1%
Raynauds1300.0%
Vitiligo100.0%

Precision

Precision was tested with positive specimens selected throughout the range of the assay. Seven patients were run in duplicate, twice per day for 20 days (n=80 replicates per sample). Assays were run with two different operators / equipment sets.

S#MeanEU/mlTotalImprecisionSDTotalImprecisionCV%Within daySDWithin dayCV%Between daysSDBetween daysCV%Within run(Repeatability)SDWithin run(Repeatability)CV%
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

60 Pineview Drive ● Toll free (800) 537-8378 ● tel. (716) 691-0091 ● fax (716) 691-0466

www.immco.com

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green dots on the left, followed by the word "immco" in bold, blue letters. Below "immco" is the word "DIAGNOSTICS" in smaller, green letters, and below that is the text "A Trinity Biotech Company" in a smaller, gray font.

110.00.77.3%0.77.3%0.00.0%0.5 5.1%
218.20.84.4%0.84.4%0.00.0%0.42.2%
319.60.94.8%0.94.7%0.21.0%0.62.8%
422.01.35.8%1.25.4%0.52.3%1.14.9%
546.23.67.8%3.47.3%1.32.7%2.45.3%
682.85.26.2%5.06.1%1.21.4%3.84.6%
7159.09.45.9%9.35.8%1.20.7%5.63.5%

Reproducibility

Qualitative reproducibility was tested with 40 runs of samples in the negative range, ~20% below cutoff, in the moderate positive range of the assay and near the qualitative analysis method. Samples were tested in duplicate twice per day for 20 days by two different operators / equipment sets. Assay results for the cutoff specimen produced 60% qualitative (positive) agreement. The sample that was ~20% above cutoff produced 98% qualitative (positive) agreement. All other specimens produced 100% qualitative agreement.

Limit of Detection

The limit of detection (LoD) was determined based on 60 replicates of the blank and 10 replicates each of 6 low-level (NHS) samples on kits from two different lots. LoD was determined to be 3.2 EU/ml.

Linearity and Recovery

Linearity and recovery were tested by diluting positive specimens through the assay range in equidistant dilutions and comparing actual vs. expected results. The linear range of the assay was determined be 3.2 (LoD) – 160 EU/ml. Results are summarized below:

Test Range(EU/ml)Slope(95% CI)Y-intercept(95% CI)R2% recovery
2.5 to 33.51.02 (0.96 to 1.10)-0.15 (-1.72 to 1.43)0.994396% to 117%
8.0 to 66.10.98 (0.95 to 1.01)1.56 (0.13 to 2.98)0.998797% to 109%
31.8 to 169.50.99 (0.88 to 1.09)-2.68 (-14.56 to 9.20)0.989091% to 103%

Interference

Interference was studied by mixing sera with known RNA POL III antibody levels with negative serum samples spiked with potential interferents and studying deviation from expected results. No significant interference was demonstrated for the following substances at the levels indicated: hemoglobin (2 g/L), bilirubin (342 umol/L), rheumatoid factor (100 EU/ml), cholesterol (13 mmo/L), triglycerides (37 mmo/L), prednisone (0.84 umol /L), Naproxen (25 mg/M/L, Enalapril (0.86 umol/L, Sildenafil (12.9pmol/L), Amlodipine (245 umol/L), Cyclophosphamide (1437 umol/L), Omeprazole (17.4 umol/L), Metoclopramide (1.5 umol/L), Doxycycline (67.5 umol/L), Bosentan (30g/ml), Tocilizumab (549ug/ml), Mycophenolate mofetil (14.01 ug/ml) and Heparin (3000 U/L).

Clinical Study: Sets of clinical samples were tested on the IMMCO RNA POLIII ELISA as described in the Cross Reactivity section. This included 281 systemic sclerosis samples and 549 autoimmune and infectious disease controls.

Systemic sclerosis Clinical Sensitivity: 23.1% (95% C.I. 18.4-28.6%) Systemic sclerosis Clinical Specificity: 98.2% (95% C.I. 96.6-99.1%)

te samples for these studies were considered positive. NHS were excluded in sensitivity/specificity calculations.

Kein lam

Kevin J. Lawson Chief Regulatory Officer

60 Pineview Drive ● USA Toll free (800) 537-8378  •  tel. (716) 691-0091  •  fax (716) 691-0466

www.immco.com

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).