K Number
K171884

Validate with FDA (Live)

Date Cleared
2017-11-22

(152 days)

Product Code
Regulation Number
882.1460
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

I-Portal® Portable Assessment System™ - Nystagmograph (I-PAS™) is used in vestibular and neuro otologic diagnostic testing. The I-PAS provides stimuli to a patient through visual cues, monitors the patient's response, and presents the data for interpretation by qualified medical personnel trained in vestibular diagnosic testing. This device does not provide diagnoses nor does it provide diagnostic recommendations.

Device Description

The I-PAS is a head-mounted apparatus enclosing a digital display, infrared lights, dual mirrors, dual cameras, and optical focus knobs. It is attached to the head with straps. The head mounted goggle system displays independent stimuli separately to each eye on a small digital display while simultaneously recording motion from each eye independently with digital video cameras. The system also includes: a computer with Neuro Kinetics' I-Portal and VEST software packages: isolation transformer and connection cables; foot pedals and a hand-held control box with buttons that can be pressed to register motor responses. As with the I-Portal VNG, I-PAS includes the standard battery of VNG tests and can operate in conjunction with caloric test equipment.

AI/ML Overview

The provided text describes the I-Portal® Portable Assessment System™ - Nystagmograph (I-PAS™), a medical device used for vestibular and neuro otologic diagnostic testing. The document is a 510(k) summary, which aims to demonstrate that the I-PAS is substantially equivalent to a legally marketed predicate device.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't present a formal table of acceptance criteria with specific performance metrics and corresponding reported device performance values. Instead, it describes general compliance and functional equivalence.

The key performance claims are related to:

  • Functional Indistinguishability of Visual Cues: The visual cues provided by the I-PAS are functionally indistinguishable from those of the predicate device.
  • Safety (No Dizziness): The I-PAS stimuli do not cause dizziness.
  • Compliance with Standards: The device complies with ANSI-ASA S3.45-2009 and applicable IEC 60601 standards.
  • Intended Performance: Verification and validation testing confirm that the I-PAS performs as intended.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample size for the clinical test set used to demonstrate functional equivalence or lack of dizziness, nor does it mention the data provenance (country of origin, retrospective/prospective). It only states that "Clinical studies were also run and confirmed..."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not provide information on the number of experts used or their qualifications for establishing ground truth in any clinical testing. The device itself "presents the data for interpretation by qualified medical personnel trained in vestibular diagnostic testing," implying that human experts interpret the device's output, but this refers to its intended clinical use, not the validation study ground truth.

4. Adjudication Method for the Test Set:

The document does not describe any adjudication method (e.g., 2+1, 3+1) for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned. The study focuses on demonstrating the functional equivalence of the I-PAS to its predicate, not on the improvement of human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance:

The I-PAS is a nystagmograph that "monitors the patient's response, and presents the data for interpretation by qualified medical personnel." It explicitly states, "This device does not provide diagnoses nor does it provide diagnostic recommendations." Therefore, it is a data acquisition and presentation device, not an AI algorithm performing diagnostic interpretation. A standalone performance evaluation in the context of an "algorithm only" would not be applicable, as it requires human interpretation of its output.

7. Type of Ground Truth Used:

The type of "ground truth" for the clinical studies mentioned appears to be based on observations and comparisons related to the "functional indistinguishability" of visual cues and the absence of dizziness. This implies that the ground truth was likely established through direct comparison or expert assessment of the visual stimuli and patient comfort/symptoms, rather than comparison to a definitive diagnostic gold standard like pathology or long-term outcomes, as the device itself does not provide diagnoses.

8. Sample Size for the Training Set:

The device is not described as using machine learning or AI in a way that would involve a training set. It's a medical device for acquiring and presenting physiological data. Therefore, the concept of a "training set" is not applicable here.

9. How Ground Truth for the Training Set Was Established:

Since there is no mention of a training set for a machine learning algorithm, this question is not applicable.

Summary of Provided Information Regarding Acceptance Criteria and Study:

The document provides a high-level overview of validation activities primarily focused on demonstrating the I-PAS's equivalence to its predicate device and adherence to relevant standards.

  • Acceptance Criteria (Implied):
    • Functional indistinguishability of visual cues compared to predicate.
    • Absence of dizziness caused by I-PAS stimuli.
    • Compliance with ANSI-ASA S3.45-2009 and IEC 60601 standards.
    • Proper functioning of the device as intended.
  • Reported Performance: "The clinical testing performed confirms that the visual cues are functionally indistinguishable from the predicate, and that I-PAS stimuli do not cause dizziness. Additionally, bench testing confirms that the I-PAS complies with ANSI-ASA S3.45-2009 and verification and validation testing demonstrate that users can safely and effectively use the I-PAS."
  • Study Details:
    • Bench tests were performed for verification, validation, and standard compliance.
    • Clinical studies were run to confirm functional equivalence of visual cues and lack of dizziness.
    • No specific sample sizes, expert details, adjudication methods, MRMC studies, or AI training/ground truth information are provided in this summary. The device's function as a data acquisition and presentation tool means many of these AI-specific questions are not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2018

Neuro Kinetics, Inc. Brian Sullivan, COO 128 Gamma Drive Blawnox, PA 15238

Re: K171884

Trade/Device Name: I-Portal® Portable Assessment System™ - Nystagmograph (I-PAS™) Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: Class II Product Code: GWN Dated: October 19, 2017 Received: October 23, 2017

Dear Brian Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171884

Device Name

I-Portal® Portable Assessment System™ - Nystagmograph (I-PASTM)

Indications for Use (Describe)

I-Portal® Portable Assessment System™ - Nystagmograph (I-PAS™) is used in vestibular and neuro otologic diagnostic testing. The I-PAS provides stimuli to a patient through visual cues, monitors the patient's response, and presents the data for interpretation by qualified medical personnel trained in vestibular diagnosic testing. This device does not provide diagnoses nor does it provide diagnostic recommendations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

NKI's I-PAS™

SubmitterJ. Howison SchroederPresident & CEONeuro Kinetics, Inc.128 Gamma DrivePittsburgh, PA 15238Phone: (412) 963-6649Fax: (412) 963-6722
Date Prepared:June 20, 2017
Trade/Proprietary Name:I-Portal® Portable Assessment System™ Nystagmograph (I-PAS™)
Common Name:Nystagmograph
Classification Name:Nystagmograph (21 CFR 882.1460, Product Code GWN)
Predicate DeviceI-Portal® VNG (K083603 & K143607)

Device Description

All I-Portal® devices, including the proposed I-PAS, function as nystagmographs, defined by 21 CFR 882.1460 as "a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball." Through its nystagmograph functionality, the I-PAS is indicated for use as a diagnostic tool to assist trained physicians in their analysis of vestibular and neurotologic disorders, requiring a separation of central and peripheral nervous system deficits. The I-PAS and all other I-Portal devices are used in an institutional environment on the order of a physician.

The I-Portal devices use a battery of oculomotor and vestibular tests across a family of products: including the I-Portal NOTC, I-Portal VNG, and I-Portal VOG. This application seeks to add I-PAS to this family. The VNG testing battery included on the I-PAS and other I-Portal devices include ten (10) oculomotor and subjective visual tests.

{4}------------------------------------------------

The I-PAS is a head-mounted apparatus enclosing a digital display, infrared lights, dual mirrors, dual cameras, and optical focus knobs. It is attached to the head with straps. The head mounted goggle system displays independent stimuli separately to each eye on a small digital display while simultaneously recording motion from each eye independently with digital video cameras. The system also includes: a computer with Neuro Kinetics' I-Portal and VEST software packages: isolation transformer and connection cables; foot pedals and a hand-held control box with buttons that can be pressed to register motor responses. As with the I-Portal VNG, I-PAS includes the standard battery of VNG tests and can operate in conjunction with caloric test equipment.

Intended Use / Indications for Use

The modified I-Portal has the same intended use and indications for use as the predicate cleared per K083603 & K143607. Both the I-PAS and its predicate device are intended for use as a nystagmograph. The indications for use for the modified I-Portal device and the predicate device cleared per K083603 & K143607 are identical, and are as follows:

I-PASTM

I-Portal® Portable Assessment System Nystagmograph (I-PAS) is used in vestibular and neuro otologic diagnostic testing. The I-PAS provides stimuli to a patient through visual cues, monitors the patient's response, and presents the data for interpretation by qualified medical personnel trained in vestibular diagnostic testing. This device provides no diagnoses nor does it provide diagnostic recommendations.

Technological Characteristics

The I-PAS has identical technological characteristics as the previously cleared predicate (K083603 & K143607), with the exception of the following modifications:

Integrated head mounted display/monitor inches from the eyes replacing projection devices /monitors that display visual cues on a surface at a distance of feet.

A small screen or monitor (smart phone display) is divided into two halves, each providing visual cue(s) to a single eye displayed approximately 3 inches from each eve in the head mounted screen that create a perceived distance, or digital equivalents that subtend the same location, angles, and speeds of the visual stimuli, i.e., dots. The visual cues are functionally equivalent to predicate visual cues projected by the pursuit tracker laser and optokinetic (OKN) on to a surface between 2 to 3 feet away from the subiect. The embedded optics of the I-PAS are also adjustable to optimize the image for patients whose eyes require a prescription, since it is not possible to wear eyeglasses while wearing I-PAS.

Fully enclosed unit that keeps out ambient light

Unlike the predicate, I-PAS is portable and the stimulus presentation and eye tracking are completely self-contained within the head mounted system. The same light seal for all I-Portal devices provides a test environment without ambient light. The predicate device projects stimuli on a wall in a dark room while the subject viewed the stimuli while wearing eyetracking goggles.

Use of 6DOF sensor to provide head position measurements during vestibular tests, e.g. Subjective Visual tests

{5}------------------------------------------------

The I-PAS includes a 6DOF sensor to help make I-PAS portable. A patient's head may not always be stationary, i.e.. secured in a headrest. As a result, if the patient tilts their head away from earth vertical, tests such as the subjective visual tests can generate inaccurate results. The 6DOF provides an accurate measure of head and stimulus orientation to either enable the tester to orient the patient's head correctly, and or for a comparison of subjective visual results to earth ground.

The clinical testing performed confirms that the visual cues are functionally indistinguishable from the predicate, and that I-PAS stimuli do not cause dizziness. Additionally, bench testing confirms that the I-PAS complies with ANSI-ASA S3.45-2009 and verification and validation testing demonstrate that users can safely and effectively use the I-PAS.

Non-Clinical and Clinical Performance Data

Both bench and clinical testing provide significant support for this application.

Bench tests were executed to verify and validate the I-PAS, and to demonstrate its conformance with ANSI-ASA S3.45-2009, FDA's applicable consensus standard.

Compliance with applicable IEC 60601 standards is confirmed via 3d party testing.

Verification testing consisted of verification of the software requirements and hardware requirements.

Validation testing was performed to ensure that users were able to properly use I-PAS and that the user interface was operating as intended.

The verification and validation test results confirm that the I-PAS performs as intended.

Clinical studies were also run and confirmed that the I-PAS visual cues are functionally indistinguishable from the predicate visual cues and that I-PAS stimuli do not cause dizziness.

Substantial Equivalence

The I-PAS is substantially equivalent to its predicate. It has the same intended use, the same indications for use and equivalent technological and equivalent clinical characteristics and equivalent principles of operation. The technological differences do not present any new issues of safety or effectiveness, which is supported by performance data.

The I-PAS is substantially equivalent to its predicate (K083603 & K143607) device.

§ 882.1460 Nystagmograph.

(a)
Identification. A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.(b)
Classification. Class II (performance standards).