K Number
K171873
Date Cleared
2017-11-15

(145 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device in this 510(k) Notification is a Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs

The subject device is a patient examination glove made from nitrile compound, blue in color, powder free and non-sterile (as per 21 CFR 880.6250, class I).

The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

AI/ML Overview

This document describes the acceptance criteria and performance study for the Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs (K171873).

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 6319-10Meets
Physical PropertiesASTM standard D 6319-10Meets
Freedom from pinholes21 CFR 800.20, ASTM D5151-11Meets
Powder ResidualASTM D6319-10 and D6124-06 (Reapproved 2011)Meets
Biocompatibility (Skin Irritation)ISO 10993-10:2010 Primary Skin IrritationNot a primary skin irritant under the conditions of the study
Biocompatibility (Sensitization)ISO 10993-10:2010 Dermal sensitization in the guinea pigNot a contact sensitizer under the conditions of the study
Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2013) - Minimum Breakthrough Detection Time (drug specific)See table below for specific drug performance

Chemotherapy Drug Permeation Performance:

Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes (Acceptance Criteria per ASTM D6978-05)Reported Device Performance (Minimum Breakthrough Detection Time in Minute)
Carmustine (BCNU) 3.3 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.14.7
Cisplatin 1.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Cyclophosphamide (Cytoxan) 20 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Cytarabine 100 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Dacarbazine (DTIC) 10.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Doxorubicin Hydrochloride 2.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Etoposide (Toposar) 20.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Fluorouracil 50.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Ifosfamide 50.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Methotrexate 25 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Mitomycin C 0.5 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Mitoxantrone 2.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Paclitaxel (Taxol) 6.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
Thiotepa 10.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.58.8
Vincristine Sulfate 1.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240

Note: The document highlights that Carmustine (BCNU) and Thiotepa have "extremely low permeation times" and a warning statement is to be included in labeling for these specifically.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in terms of a dataset of items being classified or evaluated by an algorithm or human readers. Instead, it refers to bench tests conducted on the device itself.

  • Sample Size: Not explicitly stated for each specific test (e.g., number of gloves tested for pinholes, physical properties, or chemotherapy permeation). The general statement is "Bench tests were conducted to verify that the proposed device met all specifications."
  • Data Provenance: The tests are described as "Bench tests" which are typically laboratory-based and prospective in nature for a new device. The country of origin for the data is not specified, but the applicant company is Zhonghong Pulin Medical Products Co., Ltd. located in China.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable to this type of device submission. The "ground truth" for glove performance is established through adherence to recognized international standards (e.g., ASTM, ISO) and regulatory requirements (e.g., 21 CFR 800.20), not through expert consensus on a test set of data. The performance is objectively measured against predefined criteria in these standards.

4. Adjudication Method for the Test Set

Not applicable. As noted above, performance is measured against objective standards, not through consensus or adjudication of expert interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This type of study is not relevant for a medical device like an examination glove. MRMC studies are typically used for evaluating diagnostic imaging or AI algorithms where human interpretation plays a significant role.

6. Standalone Performance Study (Algorithm Only Without HIL)

No. This device is a physical medical glove and does not involve AI algorithms or human-in-the-loop performance evaluation in the context of an algorithm. The "device performance" refers to the physical and chemical properties of the glove itself.

7. Type of Ground Truth Used

The "ground truth" for evaluating the glove's performance is based on established industry standards and regulatory requirements. This includes:

  • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical application (Dimension, Physical Properties, Powder Residual).
  • 21 CFR 800.20 & ASTM D5151-06 (2011): Freedom from pinholes.
  • ASTM D6124-06 (2011): Standard Test Method for Residual Powder on Medical Gloves.
  • ISO 10993-10:2010: Biocompatibility (Primary Skin Irritation, Dermal Sensitization).
  • ASTM D6978-05 (Reapproved 2013): Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

8. Sample Size for the Training Set

Not applicable. This device is not an AI algorithm that requires a training set. Its performance is evaluated through physical and chemical testing.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.