(268 days)
Fixed Core Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.
The Fixed Core Wire Guides, subject of this submission, are Class II devices according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The subject device consists of a core mandril, a spring coil anchored over the mandril, and an inner safety wire. The Fixed Core Wire Guides are available with outside diameters ranging from 0.015 inches to 0.052 inches and lengths ranging from 15 centimeters to 480 centimeters. The flexible tip portion of the wire guides is either straight or J-tipped. The Fixed Core Wire Guides will be uncoated or coated with Polytetrafluoro-ethylene (PTFE). The Fixed Core Wire Guides are packaged, sterile devices intended for single use.
The document describes the premarket notification (510(k)) for the Fixed Core Wire Guides. However, it does not pertain to an AI/ML medical device. Therefore, much of the information requested in your prompt (e.g., patient data, training/test sets, expert consensus, MRMC studies, standalone performance, etc.) is not applicable to this submission, as it is for a physical medical device (wire guides).
The "study that proves the device meets the acceptance criteria" refers to the various engineering and biological tests conducted on the physical device.
Here's an attempt to answer the prompt based on the provided document, acknowledging the non-AI nature of the device:
Device: Fixed Core Wire Guides (K171764)
Device Type: Physical Medical Device (Catheter Guide Wire)
Regulation: 21 CFR 870.1330 (Catheter Guide Wire)
Given that this is a 510(k) submission for a physical medical device (wire guide) and not an AI/ML-driven device, many of the standard questions regarding AI/ML acceptance criteria and study design are not relevant or applicable. The "acceptance criteria" here refer to the passing thresholds for various physical and biological performance tests.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide specific quantitative acceptance criteria values or detailed performance results for each test. Instead, it states that "The pre-determined acceptance criteria were met" for all listed tests.
Test Category | Acceptance Criteria Description (as stated) | Reported Device Performance (as stated) |
---|---|---|
Biocompatibility | Tested in accordance with ISO 10993-1:2009. | The pre-determined acceptance criteria were met. |
Corrosion Testing | Tested in accordance with Annex B of ISO 11070:2014. | The pre-determined acceptance criteria were met. |
Flexing Test | Tested in accordance with Annex G of ISO 11070:2014. | The pre-determined acceptance criteria were met. |
Fracture Testing | Tested in accordance with Annex F of ISO 11070:2014. | The pre-determined acceptance criteria were met. |
Tensile Testing | Tested in accordance with ISO 11070:2014, Annex H. | The pre-determined acceptance criteria were met. |
Tip Flexibility | Characterization testing performed in accordance with the FDA Coronary and Cerebrovascular Guidewire Guidance (1995). | (No explicit statement of "met criteria," but characterization was successful) |
Torque Strength Testing | Characterization testing performed in accordance with the FDA Coronary and Cerebrovascular Guidewire Guidance (1995). | (No explicit statement of "met criteria," but characterization was successful) |
Acute Performance Evaluation | Evaluated in an animal model in accordance with an approved study protocol. | The pre-determined acceptance criteria were met. |
2. Sample size used for the test set and the data provenance
For a physical device, "sample size" typically refers to the number of device units tested. The document does not specify the exact sample sizes (number of units) used for each of the performance tests (e.g., biocompatibility samples, number of wires for tensile testing, etc.).
- Data Provenance: The tests are laboratory and animal studies, not human clinical data.
- Biocompatibility, Corrosion, Flexing, Fracture, Tensile, Tip Flexibility, Torque Strength: Performed in a laboratory setting.
- Acute Performance Evaluation: Performed in an animal model.
- Retrospective or Prospective: These would be considered prospective tests performed specifically for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable. For physical device performance testing, "ground truth" is established by the test procedures themselves (e.g., measuring force, observing breakage, biological reactions) and assessed by qualified laboratory personnel and engineers, not clinical experts for diagnostic accuracy.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This concept is relevant for human interpretation of data, not for material or mechanical performance testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is for AI/ML devices where human readers interact with AI. This device is a physical wire guide.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This pertains to AI/ML software performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this physical device, the "ground truth" for demonstrating performance is derived from:
- Standardized industry and international test methods (e.g., ISO, FDA guidance documents).
- Quantitative measurements (e.g., force, elongation, cycles to failure).
- Qualitative observations (e.g., absence of corrosion, successful access in animal model).
- Biocompatibility assessments based on established biological endpoints.
8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of AI/ML or statistical modeling from patient data. The "training" for such a device would be the design and manufacturing processes.
9. How the ground truth for the training set was established
Not applicable. See point 8.
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.