K Number
K171580
Date Cleared
2017-11-01

(154 days)

Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B125/B105 is not intended for use during MRI.

The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.

The Monitor B125/B105 monitors and displays : ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Tympani/Rectal/ Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration rate, CO2.

The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Couplet, Bigeminy, Trigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.

Device Description

The Monitor B125/B105 is a multi-parameter patient monitor that is developed based on the predicate Monitor B40 (K151063) platform. The Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650(K102239)), WLAN (FCC ID: OU5B1X501) and touch screen.

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 is a multiparameter patient monitor, utilizes 12inches /10inches LCD display with an integrated keypad and a pre-configuration (hemodynamic module) which provide basic parameters: ECG, RESP, NIBP, IBP, TEMP, SPO2.

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has optional CO2 parameter provided by the identical E-MiniC module (K052582).

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has a mounting plate on the bottom of the monitor. The monitor can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the GE Monitor B125/B105. While it details various performance data related to electrical safety, EMC, software verification, environmental, mechanical stress, and packaging tests, it does not include specific acceptance criteria or a study that directly proves the device meets such criteria for parameters like arrhythmia detection accuracy.

The document states:

  • "The proposed Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650 (K102239))"
  • "The proposed device uses the same ECG algorithm (EK-Pro V12) as predicate device B40 (K151063)"
  • "The proposed device enabled full arrhythmia analysis which disabled in the predicate device B40 (K151063)."
  • "Full arrhythmia analyses implemented with EK -ProV12 ECG algorithm was cleared in B650 (K102239)"

This indicates that the arrhythmia detection capabilities rely on a previously cleared algorithm (EK-Pro V12) and that the full arrhythmia analysis feature was enabled on the new device, having been previously cleared in the predicate device B650 (K102239). However, no new study data for arrhythmia detection performance or specific acceptance criteria for these functionalities are presented in this document.

Therefore, based solely on the provided text, I cannot complete the requested information regarding acceptance criteria and a study proving the device meets them for ECG arrhythmia detection. The document primarily focuses on demonstrating substantial equivalence to predicate devices through comparisons of technological characteristics and adherence to various safety and performance standards, but not on detailed performance metrics for specific clinical functions like arrhythmia detection.

No information is available for the following points in the provided document:

  • A table of acceptance criteria and the reported device performance.
  • Sample size used for the test set and the data provenance for arrhythmia detection.
  • Number of experts used to establish the ground truth for the test set and their qualifications for arrhythmia detection.
  • Adjudication method for the test set for arrhythmia detection.
  • Multi-reader multi-case (MRMC) comparative effectiveness study information.
  • Standalone (algorithm only) performance information for arrhythmia detection.
  • Type of ground truth used for arrhythmia detection.
  • Sample size for the training set for arrhythmia detection.
  • How the ground truth for the training set was established for arrhythmia detection.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 1. 2017

GE Medical Systems (China) Co., Ltd. % Mr. Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223

Re: K171580

Trade/Device Name: Monitor B125/B105 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, DSK, GWQ, FLL, NHO, NHP, NHQ Dated: May 30, 2017 Received: May 31, 2017

Dear Mr. Robert Casarsa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Robert Casarsa

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Wilhelmmen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171580

Device Name Monitor B125/B105

Indications for Use (Describe)

The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B125/B105 is not intended for use during MRI.

The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.

The Monitor B125/B105 monitors and displays : ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Tympani/Rectal/ Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration rate, CO2.

The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Couplet, Bigeminy, Trigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary

K171580

I. SUBMITTER

GE MEDICAL SYSTEMS (CHINA) CO., LTD. No. 19 Changjiang road National Hi-Tech Dev. Zone Wuxi, Jiangsu, China 214028 Sun YanLi Regulatory Affairs Leader

Primary Contact Person: Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. Telephone: 414-362-3063 Fax at 414-362-2585 E-mail: Robert.casarsa@ge.com

Secondary Contact Person: Douglas Kentz Regulatory Affairs Director GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 Phone: 414 362-2038 Fax: 414-262-2585 E-mail: Douglas.kentz@ge.com

II. DEVICE

Name of Device: Monitor B125/B105 Common or Usual Name: Multi-parameter Patient Monitor Classification Name: 21 CFR 870.1025 monitor, physiological, patient (with arrhythmia detection or alarms) 21 CFR 868.2375 monitor, breathing frequency 21 CFR 868.1700 analyzers, gas, nitrous-oxide, gaseous phase (anesthetic conc.) 21 CFR 868.1620 analyzers, gas, halothane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzers, gas, enflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1400 analyzers, gas, carbon-dioxide, gaseous-phase 21 CFR 868.1720 analyzers, gas, oxygen, gaseous-phase 21 CFR 870.1130 system, measurement, blood-pressure, non-invasive 21 CFR 870.2700 oximeters 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm)

{4}------------------------------------------------

  • 21 CFR 870.2770 plethysmographs, impedance 21 CFR 870.1110 computers, blood-pressure 21 CFR 882.1400 full-montage standard electroencephalograph 21 CFR 880.2910 thermometers, electronic, clinical 21 CFR 868.1500 analyzers, gas, desflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, sevoflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, isoflurane, gaseous-phase (anesthetic concentration)
    Regulatory Class: II Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, DSK, GWQ, FLL, NHO, NHP, NHQ

III. PREDICATE DEVICE

K151063 Monitor B40 and K102239 CARESCAPE Monitor B650 These predicates have not been subject to a design-related recall. No reference devices were used in this submission.

DEVICE DESCRIPTION IV.

The Monitor B125/B105 is a multi-parameter patient monitor that is developed based on the predicate Monitor B40 (K151063) platform. The Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650(K102239)), WLAN (FCC ID: OU5B1X501) and touch screen.

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 is a multiparameter patient monitor, utilizes 12inches /10inches LCD display with an integrated keypad and a pre-configuration (hemodynamic module) which provide basic parameters: ECG, RESP, NIBP, IBP, TEMP, SPO2.

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has optional CO2 parameter provided by the identical E-MiniC module (K052582).

As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has a mounting plate on the bottom of the monitor. The monitor can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.

{5}------------------------------------------------

V. INDICATONS FOR USE

The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner.

The Monitor B125/B105 is not intended for use during MRI.

The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.

The Monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/ Myocardial/Core/Surface temperature, impedance respiration, respiration rate, CO2.

The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia. VT>2. Ventricular Bradycardia. Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The proposed Monitor B125/B105 is a multi-parameter patient monitor that is developed based on the predicate Monitor B40 (K151063) platform. The proposed Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650 (K102239)), WLAN (FCC ID: OU5B1X501) and touch screen

The proposed Monitor B125/B105 uses the identical hemodynamic module (Hemo module) as the predicate Monitor B40 (K151063) in the monitoring of ECG, Resp, NIBP, IBP, SpO2 and Temp parameters data.

The proposed Monitor B125/B105 uses identical E-MiniC module (K052582) as the predicate Monitor B40 (K151063).

The fundamental technology of the proposed Monitor B125/B105 is the same as the predicate devices.

{6}------------------------------------------------

The proposed Monitors B125/B105 is as safe and effective as the predicate devices.

The following table includes comparisons of the main features of the device, and includes the features that are different from the predicate.

WLANSupport wireless communication.WLAN module is FCC qualified, FCC ID: OU5B1X501)
Touch screenSupport touch screen per customer needs
ECG arrhythmiaanalysisEnabled ECG full arrhythmia analysisThe proposed device uses the same ECG algorithm (EK-ProV12) as predicate device B40 (K151063)The proposed device enabled full arrhythmia analysis whichdisabled in the predicate device B40 (K151063).Full arrhythmia analyses implemented with EK -ProV12 ECGalgorithm was cleared in B650 (K102239)
CARESCAPECentral StationV2 (K162012)compatibilityAdded the compatibility with CARESCAPE Central StationV2 (K162012)CARESCAPE Central Station V2 (K162012) has the samecommunication protocol as CARESCAPE central station V1(K133882) which has been cleared in predicate device B40(K151063)
ImprovedoperatingworkflowAdd standby and night mode Up to 168 hours trending storage and 200 snapshot checks Add event view software feature to combine alarm history and snapshot
Peripheral I/OinterfaceAdd DVI-D connector to support slave display. Add USB connector to support USB disk for service purpose
Masimo SpO2accessorieschangeAdd Masimo RD sensors and related cables, the sensors arecompatible with MS-2011SB board (K053269)Two reusable sensors (K051212)MASIMO-RD-4050
Four adhesive sensors (K042346):
MASIMO-RD-4000
MASIMO-RD-4001
MASIMO-RD-4003
MASIMO-RD-4004
Two RD cables (K042536)
MASIMO-RD-4085
MASIMO-RD-4084

{7}------------------------------------------------

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

No biocompatibility testing was performed as the Monitor B125/B105 is not in contact with the patient. All supplies and accessories for use with the proposed Monitor B125/B105 that contact the patient have been previously cleared by FDA.

Electrical safety and electromagnetic compatibility (EMC)

The Monitor B125/B105 is designed and tested for compliance with the following performance standards:

  • . IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • . IEC60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-1-8:2006 + A1:2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • . IEC 60601-2-27: 2011 + C1:2012 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
  • IEC 80601-2-30:2009 + A1:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers
  • IEC 60601-2-34:2011 Medical electrical equipment - Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
  • . IEC 60601-2-49:2011 Medical electrical equipment – Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment

{8}------------------------------------------------

  • ISO 80601-2-55:2011 Medical electrical equipment -- Part 2-55: Particular ● requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 80601-2-56:2009 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • . ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
  • IEC 62366-1:2015 Medical devices-Part 1: Application of usability engineering to ● medical devices
  • IEC 60601-1-6:2010+A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability

Software Verification and Validation Testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Major" level of concern, where a failure could result in death or serious injury, either to a patient or to a user of the device. The following bench tests were performed:

Chapter 2: Vectors and Scalars

Software Verification include below Testing:

  • Unit Testing: Code reviews and/or static code analysis are the methods are adopted . for the software unit verification to ensure the unit acceptance criteria, as defined in the SDLC procedure, are met.
  • Software Integration Test: Performed to confirm that integrated (changed) SW units conform to the architecture and requirements, and other portions of the system have not been adversely affected.
  • . Software Regression Test: The purpose of this test is to verify the side effect of code changes. Software leader and engineering team shall review the side effect of all SPRs and it shall be mentioned in the SPR test description or individual test procedure.
  • System Testing: Performance testing included the system level testing and subsystem level testing. System level procedures were executed to confirmed that all product system requirements specified in the product specifications were met. Design Specifications specify design requirements for each subsystem of Patient Monitor B125/B105. Each Design Specification has one or more corresponding Verification Procedures that were used to verify that the design requirements have been met.

Software Validation is integrated with System Validation: System Validation will confirm that the product fulfills the user needs and intended uses under actual or simulated use conditions.

{9}------------------------------------------------

K171580 Page 7 Of 7

Environmental Testing:

Perform Rapid Change of Temperature, Change of Temperature, Dry heat for non-heatdissipating specimen with gradual change of temperature, Damp heat, Cold for non-heat dissipating specimen with gradual change of temperature, Cold for heat-dissipating specimen with gradual change of temperature and steady state testing to ensure the unit preference meet the environmental requirements.

Mechanical stress Testing:

Perform sinusoidal vibration, shock, free fall, altitude test and broadband random vibration testing to ensure the unit meet the Mechanical stress requirements.

Package Testing

Perform package transport testing to ensure the packaged product withstand the distribution environment and to show the serviceability of the transport packages.

Animal Study

i

No animal studies have been performed on the Monitor B125/B105.

Clinical Studies

No clinical studies have been performed on the Monitor B125/B105.

VIII. CONCLUSIONS

GE Healthcare considers the proposed Monitor B125/B105 to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.