K Number
K171547
Manufacturer
Date Cleared
2018-08-02

(433 days)

Product Code
Regulation Number
890.5360
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Qualisys Clinical System is a camera and computer system used to quantify and graphically display human movement patterns for adults and children. It is intended to be used for movement analysis in the fields of gait analysis, rehabilitation, sports medicine and ergonomics.

Device Description

The Qualisys Clinical System (QCS) is a camera and computer system used to quantify and graphically display human movement patterns for adults and children. It is intended to be used for movement analysis in the fields of gait analysis, rehabilitation, sports medicine and ergonomics.

In order to achieve this, the QCS utilizes data captured from image sensors (motion capture cameras) to triangulate the 3D-position of one or several reflective markers attached to the patient. This is done by two or more cameras calibrated to provide overlapping field of views from multiple angles. QCS produce data with 3 degrees of freedom for each marker, i.e. positional information. Rotational information (e.g. of a limb) may be retrieved from the relative orientation of three or more markers.

The tracking cameras may record images and identify the position of the markers with a high spatial and temporal resolution to generate high performance motion capture of the markers.

The markers are usually attached directly to the skin. The markers are specially designed to reflect the IR-light flashed from the tracking cameras.

In addition, subsystems may be added to record data in synchronization with the tracking cameras to facilitate the analysis, or for pure documentation purposes. The additional subsystems may be any one of, or a combination of: 1) Ordinary video, 2) EMG data and 3) Force plate data.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information based on the provided document:

Acceptance Criteria and Device Performance

The document does not explicitly present a table of acceptance criteria with specific performance metrics (e.g., accuracy, precision) that the Qualisys Clinical System (QCS) must meet for regulatory purposes. Instead, the acceptance criteria are implicitly tied to demonstrating substantial equivalence to a predicate device (SMART-D [K131660]) through technical and functional comparisons, along with performance data from non-clinical tests.

The reported device performance is described generally through the successful completion of software and hardware tests, and demonstration of metrological accuracy and reliability.

Table 1: Implicit Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Implicit)Reported Device Performance and Discussion
Pertaining to Substantial Equivalence: Technical Characteristics
Identical CFR Section, Product Code, Regulation Number, and Classification Name to PredicateMet: All are identical to the predicate (Table 5-1, items 1-4).
Equivalent Intended Use to PredicateMet: The QCS and predicate perform the same tasks for motion analysis (Table 5-1, item 5).
Equivalent System Components to PredicateMet: Both consist of Workstation + Software, Tracking Cameras (Table 5-1, item 6).
Software Tools enable similar functions (data acquisition, 3D marker tracking, 3D kinematics, reporting)Met with discussion: QCS software modules (QTM, Qualisys Report, Clinical Gait PAF Module) combined with third-party software (Visual3D) perform similar functions to the predicate's software (SMARTcapture, SMARTtracker, SMARTanalyzer, SMARTclinic) (Table 5-1, item 7; Table 5-3). The differences in specific software names and whether analysis protocols are customizable are discussed as not impacting safety or effectiveness.
Equivalent Contraindications, Target Population, Light Emission, Modulation/External Sync, and LensMet: All are identical to the predicate (Table 5-1, items 8, 9, 12, 15, 16).
Frame rate appropriate for clinical gait analysisMet with discussion: QCS operates at 100 fps. This is discussed as being appropriate and in line with scientific literature recommendations for clinical gait analysis, even though the predicate operates at 200 fps (Table 5-1, item 10; Table 5-3).
Acquisition Frequencies acceptableMet with discussion: QCS records analog data faster (500-2000 Hz) than the predicate (250 Hz, up to 500 Hz). This improves time-resolution, not reducing safety or effectiveness (Table 5-1, item 11; Table 5-3).
Light wavelength comparable to predicateMet with discussion: QCS (850 nm) and predicate (880 nm) have very similar IR wavelengths, behaving the same in terms of attenuation and scatter. Optimized for QCS cameras. No influence on intended use or safety/effectiveness (Table 5-1, item 13; Table 5-3).
Output Angle (FOV) sufficient for marker detectionMet with discussion: QCS has a wider FOV (61-42 degrees) than the predicate (40 degrees), which improves marker detectability, not reducing safety or effectiveness (Table 5-1, item 14; Table 5-3).
Number of Cameras supportedMet with discussion: QCS supports up to 100 cameras, more than the predicate's 16. This does not influence intended use or reduce safety/effectiveness (Table 5-1, item 17; Table 5-3).
Camera Resolution sufficient for precisionMet with discussion: QCS has higher resolutions (1216x800 - 4096x3072) than the predicate (800x600), improving precision of marker detection. Not influencing intended use or safety/effectiveness (Table 5-1, item 18; Table 5-3).
Reporting capability equivalentMet: Both provide reporting functionalities (Table 5-1, item 21).
Physical characteristics (Weight, Dimension, Max input power) not impacting safety or effectivenessMet with discussion: Differences in weight, dimension, and max input power are discussed as not influencing intended use or reducing safety/effectiveness (Table 5-1, items 22, 23, 25; Table 5-3).
Equivalent Power Supply and Electrical Safety/EMC standards complianceMet: Equivalent power supply and compliance with relevant electrical safety and EMC standards (Table 5-1, items 26, 27, 28).
Pertaining to Substantial Equivalence: Functional Characteristics
Real-time visualization of acquired dataMet: Both support real-time visualization (Table 5-2, item 1; Table 5-1, item 24).
3D marker trajectory and kinematics reconstructionMet: Both perform 3D marker trajectory evaluation and mapping to anatomical models using validated protocols (Table 5-2, item 2).
Management of acquired data (analysis protocols)Met with discussion: QCS uses predefined analysis protocols, while the predicate allows user customization. This is discussed as not influencing intended use or reducing safety/efficiency, as the QCS protocol is a crucial, verified part of the software (Table 5-2, item 3; Table 5-4).
Report drafting and data visualizationMet: Both offer visualization tools and reporting (Table 5-2, item 4).
Performance Data
Software meets specified requirementsReported: "All software tests passed successfully, demonstrating that the QCS meets all specified requirements." (Page 10)
Hardware complies with EMC and electrical safety standardsReported: Hardware tested successfully for EMC (IEC-60601-1-2) and electrical safety (ANSI AAMI ES60601-1) (Page 10).
Metrological accuracy and reliability of 3D position of reflective markersReported: Verified and validated. Calibration devices traceable to a national meter reference (Page 10).
Clinical relevance of gait assessments supportedReported: A Clinical Evaluation report has been provided (Page 10).

Study Information

The document describes non-clinical tests rather than a formal clinical study with human subjects for novel device performance claims. The primary "study" is a comparative analysis demonstrating substantial equivalence to a predicate device.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: Not explicitly stated as a number of patients/cases. The "test set" primarily refers to software and hardware testing.
  • Data Provenance: Not specified for any dataset related to the performance data. It is focused on demonstrating substantial equivalence through technical and functional comparisons and non-clinical testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts: Not applicable. The ground truth for the non-clinical tests and metrological accuracy would be established by engineering standards, calibrated instruments, and scientific literature rather than expert human interpretation of medical data.
  • Qualifications of Experts: N/A.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. This concept typically applies to studies where human readers are interpreting a test set and their discrepancies need to be resolved. The document focuses on technical verification and validation against specified requirements and predicate device characteristics.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device is a motion capture system, not an AI-assisted diagnostic tool that aids human readers in interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Yes, the described "Non-clinical tests performed on the QCS include software and hardware tests" and "Metrological accuracy and reliability of the raw output of the system (three-dimensional position of reflective markers) has been verified and validated" indicate standalone performance evaluations of the system's components and overall output. The system quantifies and displays movement patterns.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth:
    • For metrological accuracy, the ground truth is established by physical standards (e.g., "dimensions of the calibration devices supplied with the system are traceable to a national meter reference").
    • For software and hardware functioning, the ground truth is against predefined technical specifications and requirements.
    • For the relevance of gait assessments, it refers to a "Clinical Evaluation report," implying a review against established clinical understanding and literature on gait analysis.

8. The sample size for the training set

  • Training Set Sample Size: Not applicable. The document describes a medical device for motion capture, not a machine learning or artificial intelligence system that requires a "training set" in the conventional sense.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable, as there is no mention of a training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 2, 2018

Qualisys AB Nils Betzler, PhD Product Owner Kvarnbergsgatan 2 Göteborg, 411 05 Se

Re: K171547

Trade/Device Name: Qualisys Clinical System Regulation Number: 21 CFR 890.5360 Regulation Name: Measuring Exercise Equipment Regulatory Class: Class II Product Code: LXJ Dated: July 2, 2018 Received: July 5, 2018

Dear Dr. Betzler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Hoffmann -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171547

Device Name Qualisys Clinical System

Indications for Use (Describe)

The Qualisys Clinical System is a camera and computer system used to quantify and graphically display human movement patterns for adults and children. It is intended to be used for movement analysis in the fields of gait analysis, rehabilitation, sports medicine and ergonomics.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in bold, black letters. Above the "I" in "QUALISYS" are four circles, the first one is red and the other three are gray. Below the word "QUALISYS" is the text "Motion Capture Systems" in a smaller font.

5. 510(k) Summary

Submitter Name and Address Qualisys AB Kvarnbergsgatan 2, SE-411 05 Gothenburg, Sweden

Establishment registration No: N/A

Contact

Nils Betzler Qualisys AB, Kvarnbergsgatan 2, SE-411 05 Gothenburg, Sweden Telephone: +46 31 336 94 34 Fax: +46 31 336 94 20 Email: nils.betzler@qualisys.com

Date Summary Prepared August 2, 2018

Device Name Qualisys Clinical System

Common Name QCS

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Qualisys Motion Capture Systems. The word "QUALISYS" is written in large, bold, black letters. Above the "S" in Qualisys are four circles, the first one is red and the other three are gray. Below the word "QUALISYS" is the phrase "Motion Capture Systems" in a smaller font.

Classification
Product CodeLXJ
Device Class2
Regulation Number890.5360
Review PanelPhysical Medicine
Predicate Device
Name
Manufacturer
510(k) Number
Product Code
Device Class2
Regulation Number890.5360
Review PanelPhysical Medicine

Device Description

The Qualisys Clinical System (QCS) is a camera and computer system used to quantify and graphically display human movement patterns for adults and children. It is intended to be used for movement analysis in the fields of gait analysis, rehabilitation, sports medicine and ergonomics.

In order to achieve this, the QCS utilizes data captured from image sensors (motion capture cameras) to triangulate the 3D-position of one or several reflective markers attached to the patient. This is done by two or more cameras calibrated to provide overlapping field of views from multiple angles. QCS produce data with 3 degrees of freedom for each marker, i.e. positional information. Rotational information (e.g. of a limb) may be retrieved from the relative orientation of three or more markers.

The tracking cameras may record images and identify the position of the markers with a high spatial and temporal resolution to generate high performance motion capture of the markers.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in a bold, sans-serif font. Above the "I" and "S" are three circles, one red and two gray. Below the word "QUALISYS" is the phrase "Motion Capture Systems" in a smaller font.

The markers are usually attached directly to the skin. The markers are specially designed to reflect the IR-light flashed from the tracking cameras.

In addition, subsystems may be added to record data in synchronization with the tracking cameras to facilitate the analysis, or for pure documentation purposes. The additional subsystems may be any one of, or a combination of: 1) Ordinary video, 2) EMG data and 3) Force plate data.

Intended Use

The Qualisys Clinical System is a camera and computer system used to quantify and graphically display human movement patterns for adults and children. It is intended to be used for movement analysis in the fields of gait analysis, rehabilitation, sports medicine and ergonomics.

Substantial Equivalence: The Predicate

Name:Smart-D
510(k) Holder:BTS SPA, Via della Croce Rossa 11, PD Italy 35129
510(k) Number:K131660
Date Cleared:April 2014

Substantial Equivalence: Technical and functional comparison

The proposed device, Qualisys Clinical System (QCS), is claimed to be substantially equivalent to the predicate SMART-D[K131660].

The comparison of technical and functional characteristics is presented in table 5-1 and table

5-2, respectively. Identified differences are discussed in the next section.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in a bold, sans-serif font, with a red circle and three gray circles above the "S". Below the word "QUALISYS" is the text "Motion Capture Systems" in a smaller font.

Name: Qualisys Clinical SystemK number: 171547Summary: ---SMART-D SystemK number: 131660Summary: (Att AD)Eq.
1CFR Section2(*)2(*)Yes
2Product codeLXJLXJYes
3Regulation number890.5360890.5360Yes
4Classification nameSystem, optical position/movementrecordingSystem, opticalposition/movementrecordingYes
5Intended useThe Qualisys Clinical System is acamera and computer system usedto quantify and graphically displayhuman movement patterns foradults and children. It is intendedto be used for movement analysisin the fields of gait analysis,rehabilitation, sports medicine andergonomics.SMART-D is a system formotion analysis, intendedfor the recording andanalysis of humanmovement patterns in thefields of rehabilitation,sports medicine,ergonomics.Yes
6System ComponentsWorkstation + Software, TrackingCamerasWorkstation + Software,Tracking CamerasYes
7Software ToolsQCS Software modules (QTM,Qualisys Report, Clinical Gait PAFModule) combined with third party"off-the-shelf" software(Visual3D) to achieve:a) Data acquisitionb) 3-dimensional marker trackingc) 3D kinematics reconstruction byusing an international scientific andvalidated protocold) Real-time visualizatione) Creating MS Word or WebreportsSMARTcapture,SMARTtracker,SMARTanalyzer,SMARTclinic to achieve:a) Real-time visualization ofsignals of all integrateddevices,b) 3D kinematicsreconstruction validated byinternational scientificcommunity protocols,c) Tool for easyconstruction of analysisprotocols,d) Multimedia customizableand web reportingNo
8ContraindicationsNot neededNot neededYes
9Target populationAll populationAll populationYes
10Frame rate100 fps200 fpsNo
11Acquisition Frequencies500-1500 HZ(dependent on source)250 Hz (up to 500 Hz)No
12Light emissionInfraredInfraredYes
13Light wave length850 nm880 nmNo
14Output Angle61-42 degree (FOV)40 degreeNo
ો રModulation /External SyncYY(Square wave synchronizedwith the acquisitionfrequency)Yes
16LensC-mountC-mountYes
17Number of Camerasup to 100up to 16 digital TVCNo
18Camera resolution1216×800 - 4096×3072800 H x 600 VNo
193D TrajectoryReconstructionY (QTM)Y (SMARTtracker)Yes
20CustomizableAnalysis ProtocolN (PAF)Y (SMARTanalyzer)No
21ReportingY (Qualisys Report Generator)Y (SMARTclinic)Yes
22Weight1 kg / camera20 kgNo
23Dimension1408784 mm (camera)450330500 mmNo
24Realtime visualizationYYYes
25Max input powerMax 30W / camera.To be scaled accordingly tonumber of cameras used.950 WNo
26Power supply110 V, 60 Hz / 230 V, 50 HzDual power.110 V, 60 HzYes
27Electrical SafetyANSI AAMI ES60601-1:2005/(R)2012 and A1:2012,C1:2009/(R)2012 andA2:2010/(R)2012AAMI/ANSI ES60601-1:2005 and furtheramendmentsYes
28EMCIEC-60601-1-2IEC 60601-1-2Yes
29OptionsNoneNoneYes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Qualisys Motion Capture Systems. The word "QUALISYS" is written in large, bold, black letters. Above the "I" and "S" are three circles, the first is red and the other two are gray. Below the word "QUALISYS" is the phrase "Motion Capture Systems" in a smaller font.

Tabell 5-1 Technical Characteristic Comparison. Note Eq – Equivalent. * - previously unclassified

FunctionalityCharacteristicsQualisys Clinical SystemK number: 171547Summary: ---SMART-D SystemK number: 131660Summary: (Att AD)Eq.
1Calibration and SignalsacquisistionReal-time visualization of allacquired data using Qualisys QTMand "Clinical Gait PAF Module"SMARTcapture: a) Real-time visualization of signalsof all integrated devicesYes
2Trajectory reconstruction,signal elaboration3D marker trajectory evaluated bythe Qualisys QTM softwaremodule. Marker mapped to ananatomical human model by ascientific international validatedprotocolSMARTtracker: b) 3Dkinematics reconstructionvalidated by internationalscientific communityprotocolYes
3Management of acquireddata customizableanalysis protocolsPredefined analysis protocolSMARTanalyzer: c) Toolfor easy construction ofanalysis protocolsNo

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in a bold, sans-serif font. Above the "I" in "QUALISYS" is a red circle, and above the "S" are three gray circles. Below the word "QUALISYS" are the words "Motion Capture Systems" in a smaller font.

4Report drafting, data visualizationVisualization tool to verify the trajectories and anatomical mapping algorithm. Graphical presentation in ReportsSMARTclinic: d) Multimedia customizable and web reportingYes
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Tabell 5-2 Functional Characteristic Comparison. Note Eq - Equivalent.

Substantial Equivalence: Discussion of technical and functional differences

The identified technical and functional differences (table 5-2) are discussed below in table 5-3 and table 5-4, respectively.

TechnicalCharacteristicsQualisys Clinical SystemK number: 171547Summary: ---SMART-D SystemK number: 131660]Summary: (Att AD)
Software ToolsQCS Software modules (QTM, Qualisys Report, Clinical Gait PAF Module) combined with third party “off-the-shelf” software (Visual3D) to achieve:a) Data acquisitionb) 3-dimensional marker trackingc) Mapping marker movement to a body-model using a international scientific protocol.d) Creating MS Word or Web reportsSMARTcapture, SMARTtracker, SMARTanalyzer, SMARTclinic to achieve:a) Real-time visualization of signals of all integrated devices,b) 3D kinematics reconstruction validated by international scientific community protocols,c) Tool for easy construction of analysis protocols,d) Multimedia customizable and web reporting

The QCS and its predicate perform the same tasks. Acquire data, reconstruct 3D-marker positions/trajectories, mapping data to a human anatomical body to extract kinematic information and creating reports

Frame rate100 fps200 fps
------------------------------

Whilst all of the supported Qualisys motion capture camera models are capable of operating at higher frame rates, Qualisys recommends a frame rate of 100 fps for gait analysis. This is in line with recommendations from the scientific literature:

  • Baker (2013, p. 210): "Most current systems can capture at 100 120 Hz, which is more than ● adequate for clinical gait analysis (although faster sample rates may be required for running). '
  • Richards (2008, p. 108): "Cameras exist that can provide sampling frequencies up to 10 kHz, but it is well accepted that 50 Hz is adequate for studying many aspects of human walking."
  • . Stergiou (2004 p. 236): recommended sampling frequencies for walking are 50 to 100 Hz for motion capture.

This will not influence intended use or reduce safety or effectiveness of the OCS compared to the predicate.

Full references:

Baker, Richard (2013): Measuring Walking: A Handbook of Clinical Gait Analysis: 1st Edition. London: Mac Keith Press.

Qualisys AB • Kvarnbergsgatan 2, SE-411 05 Gothenburg • SWEDEN • www.qualisys.comsales@qualisys.com • +46 31 336 94 00

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Qualisys logo. The word "QUALISYS" is written in large, bold, black letters. Above the "ISYS" portion of the word are four circles, the first of which is red and the other three of which are gray. Below the word "QUALISYS" are the words "Motion Capture Systems" in a smaller font.

Richards, Jim (2008): Biomechanics in Clinic and Research. 1st Edition. London: Churchill Livingstone.Stergiou, Nicholas (2004): Innovative Analyses of Human Movement. Champaign, IL: Human Kinetics.
AcquisitionFrequencies500-2000 Hz(dependent on source)
QCS will record analog data faster than the predicate. This improves the time-resolution and will therefore not influence intended use or reduce the safety or effectiveness of the QCS compared to the predicate
250 Hz (up to 500 Hz)
Light wave length850 nm
880 nm
The IR wavelength is close to be the same and both are well within the range of IR-light. From a practicalpoint of view, the IR-light from the two systems behave the same in terms of attenuation and scatter. Furtherthe QCS-cameras are optimized to work at this specific wavelength. There is no influence on Intended use orreduction of safety and effectiveness due to the slight difference in IR wave length
Output Angle61-42 degree (FOV)
40 degree
The QCS has a wider Field of View (FOV) than the predicate. This allows the QCS to detect reflectivemarkers over a wider spatial region which in turn will improve marker detectability. There is no influenceon Intended use or reduction of safety and effectiveness due to the wider FOV
Number of Camerasup to 100
up to 16 digital TVC
The QCS supports more cameras than the predicate. Does not influence the Intended use or reduce thesafety or effectiveness
Camera resolution1216×800 - 4096×3072
800 H x 600 V
The QCS system has a higher image resolution than its predicate. This will improve the precision of themarker detection and/or improve the detection distance from the camera compared to the predicate. Will notinfluence the "Intended use" or reduce the safety or effectiveness
CustomizableAnalysis ProtocolN (PAF)
Y (SMARTanalyzer)
The QCS does not allow the user to modify the analysis protocol. The protocol is considered to be a crucial part of the software, requiring specific verification and validation. As such, this does not influence the "Intended use", impose any risk or reduced safety or efficiency.
Weight1 kg / camera
20 kg
The QCS does not require a base station, instead the sub-systems are directly connected to the PC workstation. The weight of a camera does not bring a point of non-substantial equivalence between the QCS system and its predicate. The difference in weight does no influence the "Intended use" or reduce the safety or the effectiveness
Dimension1408784 mm
450330500 mm
The dimensions of a camera do not constitute a point of non-substantial equivalence between the QCS system and its predicate. The difference in dimensions does no influence the "Intended use" or reduce the safety or the effectiveness

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in bold, black letters. Above the "I" in "QUALISYS" are four circles, the first one is red and the other three are gray. Below the word "QUALISYS" is the text "Motion Capture Systems" in a smaller, gray font.

Max input powerMax 30W / camera.To be scaled accordingly tonumber of cameras used.950 W
The max input power does not constitute a point of non-substantial equivalence between the SMART-DSystem and the predicate devices, as it does not influence the intended use, the performance, the safety andeffectiveness of the system than the predicate devices

Tabell 5-3 Technical Characteristic Differences Discussion.

FunctionalityCharacteristicQualisys Clinical SystemK number: 171547Summary: ---SMART-D SystemK number: 131660Summary: (Att AD)
Management ofacquired datacustomizableanalysis protocolsPredefined analysis protocolSMARTanalyzer: c) Tool for easy construction ofanalysis protocols

The QCS does not allow the user to change the analysis protocol. The reason for this is that the protocol is considered to be a crucial part of the software and the corresponding verification and validation of the software. The quality of its design is considered an integral part of the system. This does not influence the "Intended use", impose any risk or reduction on safety or on efficiency

Tabell 5-4 Functionality Characteristic Differences Discussion.

Performance Data

Non-clinical tests performed on the QCS include software and hardware tests. All software tests passed successfully, demonstrating that the QCS meets all specified requirements. Hardware has been tested successfully for EMC compatibility (IEC-60601-1-2), and electrical safety has been assessed and found to comply with a relevant standard (ANSI AAMI ES60601-1), which is equal to the standard used by the predicate. Metrological accuracy and reliability of the raw output of the system (three-dimensional position of reflective markers) has been verified and validated, and the dimensions of the calibration devices supplied with the system are traceable to a national meter reference. Further, a Clinical Evaluation report has been provided, supporting the clinical relevance of gait assessments.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the Qualisys logo. The logo consists of the word "QUALISYS" in bold, black letters. Above the "I" in "QUALISYS" are four circles, the first one is red and the other three are gray. Below the word "QUALISYS" is the phrase "Motion Capture Systems" in a smaller, lighter font.

Conclusion

Based on the comparison, discussion of identified differences and test results, it is concluded that the proposed device (Qualisys Clinical System) is substantially equivalent to its predicate (SMART-D[K131660]).

§ 890.5360 Measuring exercise equipment.

(a)
Identification. Measuring exercise equipment consist of manual devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. These devices also include instrumentation, such as the pulse rate monitor, that provide information used for physical evaluation and physical planning purposes., Examples include a therapeutic exercise bicycle with measuring instrumentation, a manually propelled treadmill with measuring instrumentation, and a rowing machine with measuring instrumentation.(b)
Classification. Class II (special controls). The device, when it is a measuring exerciser or an interactive rehabilitation exercise device for prescription use only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.