K Number
K171316

Validate with FDA (Live)

Device Name
O2Vent W
Date Cleared
2017-07-11

(68 days)

Product Code
Regulation Number
872.5570
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The O2Vent W is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The device is indicated for use during sleep to aid in the treatment of these conditions.

Device Description

The O2Vent W is an oral appliance and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The O2Vent W is made up of two parts: 1. The Upper Tray fitted over the upper teeth, with the breathing port at the front leading to the airways on each side to the rear of the appliance, is made from medical grade titanium (3D printed). The upper titanium tray has two side protrusions (wings) that extend vertically down to interface with the lower acrylic tray for titration of the device. The upper tray is lined with acrylic insert, and is customized to patient bite and impression. 2. The Lower Tray customized to the lower teeth, positions the lower jaw forward, preventing the soft tissue of the throat from collapsing and obstructing the airway. The lower tray features an adjuster assembly set into the acrylic, comprising of an adjuster block and also adjustment screws each located on either side of the lower arch. The adjuster block interfaces with the side protrusions on the upper tray to provide titration. The screws are adjusted with an Adjustment Key for mandibular adjustment to be set by the dentist and/or by the patient under the direction of the dentist.

AI/ML Overview

The provided text does not contain information about the acceptance criteria or a study proving that the O2Vent W device meets specific performance criteria. Instead, it is a 510(k) summary for the device, focusing on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study.

Here's what can be extracted from the document regarding testing, and why it doesn't align with the detailed request:

1. A table of acceptance criteria and the reported device performance
This information is not available in the provided text. The document focuses on comparing the O2Vent W to predicate devices based on design features, intended use, and materials.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
No clinical study with a test set is described. Non-clinical testing was performed, but sample sizes for these tests are not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no ground truth for a test set is described in the context of a clinical study for performance evaluation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an oral appliance, not an AI-assisted diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is a medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. The document states that "Human clinical studies were not deemed necessary to evaluate the substantial equivalence of the O2Vent W".

8. The sample size for the training set
Not applicable, as no training set for an algorithm is discussed.

9. How the ground truth for the training set was established
Not applicable.


Information that is present regarding testing:

  • Summary of Non-Clinical Tests (Page 8):

    • Scope: Risk assessment of physical properties and ability to achieve intended use.
    • Preliminary Testing: Performed to confirm design can withstand forces during normal use and reasonable abuse (e.g., bruxing).
    • Results: The O2Vent W "met the specifications set." (Specific specifications are not detailed).
    • Materials: New materials (adjustment screws) are made of special grade stainless steel, have prior 510(k) approval, and are widely used in dental appliances, matching those in the predicate SOMNODENT CLASSIC device.
    • Biocompatibility: An assessment was performed. No new materials were used; all materials are already in predicate devices (O2Vent T & OVENT). Biocompatibility testing of the adjustment screws was deemed not warranted due to their source, nature, and long history of safe use in the intraoral environment. The device is considered biocompatible based on similar materials to predicate devices (O2Vent T, OVENT & SOMNODENT CLASSIC).
  • Summary of Clinical Tests (Page 8):

    • Conclusion: Human clinical studies were not deemed necessary for substantial equivalence due to:
      • Designs not dissimilar from predicate devices.
      • No new technologies different from legally-marketed intra-mandibular repositioning devices.
      • No deviation from indications for use of predicate devices (O2Vent T, OVENT, SOMNODENT CLASSIC).
    • Risk Management: Risks and mitigating controls (e.g., potential for soreness) are addressed by advising patients and dentists in the "Instructions for Use," covering contraindications, warnings, precautions, storage, fitting, and adjustment.

In essence, the document confirms that the O2Vent W underwent non-clinical testing for usability and material safety, met its internal specifications for these tests, and was deemed not to require clinical studies for its 510(k) submission based on its substantial equivalence to existing devices. Specific quantitative acceptance criteria and detailed performance data from a clinical study are not provided.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 11, 2017

Oventus Manufacturing Pty Ltd % M.W. Andy Anderson, Ph.D., RAC Senior Principal Advisor Regulatory and Clinical Research Institute. Inc. 5353 Wayzata Boulevard, #505 Minneapolis, Minnesota 55416

Re: K171316

Trade/Device Name: O2Vent W Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: June 15, 2017 Received: June 16, 2017

Dear M.W. Andy Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mary S. Runner -A

Lori Wiggins Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171316

Device Name O2VENT W

Indications for Use (Describe)

The O2Vent W is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The device is indicated for use during sleep to aid in the treatment of these conditions.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)X
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Oventus Medical Ltd. The logo features a stylized letter "O" in a gradient of blue, with a wave-like shape running through the center. Above the "O" are several small blue circles, arranged in a cluster. Below the symbol is the word "OVENTUS" in a sans-serif font, with the letters "VEN" in a lighter shade of blue.

510(k) SUMMARY FOR O₂VENT W

prepared in accordance to 21CFR 807.92 K171316

Date PreparedJune 29, 2017
510 (k) Owner DetailsName:Oventus Manufacturing Pty Ltd
Address1 Swann Road, Indooroopilly, QLD 4068, Australia.
Phone:+61 7 3180 3196
Name of Contact PersonHemangi Malde
Device Trade nameO2Vent W
Common NameOral Appliance – Anti-snoring device
ClassificationClass II
Classification NameIntraoral devices for snoring and obstructive sleep apnea forsnoring and obstructive sleep apnea.
Classification CodeLRK
Classification Regulation21 CFR 872.5570
Device Model Number(s)O2VWUS
Predicate DeviceO2Vent T - Oventus Manufacturing Pty Ltd - PrimaryOVENT – Oventus Manufacturing Pty LtdSOMNODENT CLASSIC – Somnomed Inc.
Device DescriptionThe O2Vent W is an oral appliance and is intended to reduce oralleviate snoring and mild to moderate obstructive sleep apnea(OSA).The O2Vent W is made up of two parts:1. The Upper Tray fitted over the upper teeth, with the breathingport at the front leading to the airways on each side to the rear ofthe appliance, is made from medical grade titanium (3D printed).The upper titanium tray has two side protrusions (wings) thatextend vertically down to interface with the lower acrylic trayfor titration of the device. The upper tray is lined with acrylicinsert, and is customized to patient bite and impression.
2. The Lower Tray customized to the lower teeth, positions thelower jaw forward, preventing the soft tissue of the throat from
collapsing and obstructing the airway. The lower tray features an
adjuster assembly set into the acrylic, comprising of an adjusterblock and also adjustment screws each located on either side of
the lower arch.
The adjuster block interfaces with the side protrusions on theupper tray to provide titration. The screws are adjusted with anAdjustment Key for mandibular adjustment to be set by thedentist and/or by the patient under the direction of the dentist.
Indications for UseThe O 2 Vent W is a removable medical device that is fitted in thepatient's mouth and is intended to reduce or alleviate snoring andmild to moderate obstructive sleep apnea (OSA). The device isindicated for use during sleep to aid in the treatment of theseconditions.
Target PopulationAdult patients 18 years and older.
Environment of UseHome Use and Sleep laboratories.
Summary of Comparisonof TechnologicalCharacteristics of theDevice to Predicate DeviceAs mentioned in the following pages.

{4}------------------------------------------------

{5}------------------------------------------------

FeatureO2Vent WO2Vent TOVENTSOMNODENT CLASSIC
510k NumberK171316K161832K160234K050592
Device PhotoImage: O2Vent W deviceImage: O2Vent T deviceImage: OVENT deviceImage: SOMNODENT CLASSIC device
Indications for UseThe O2Vent W is aremovable medical devicethat is fitted in the patient'smouth and is intended toreduce or alleviate snoringand mild to moderateobstructive sleep apnea(OSA). The device isindicated for use duringsleep to aid in the treatmentof these conditions.The O2Vent T is aremovable medical devicethat is fitted in the patient'smouth and is intended toreduce or alleviate snoringand mild to moderateobstructive sleep apnea(OSA). The device isindicated for use duringsleep to aid in the treatmentof these conditions.The OVENT is a removablemedical device that is fittedin the patient's mouth and isintended to reduce oralleviate snoring and mildto moderate obstructivesleep apnea, OSA. Thedevice is indicated for useduring sleep to aid in thetreatment of theseconditions.To intended to reduce nighttime snoring and mild tomoderate obstructive sleepapnea (OSA) in adults.
Product CodesLRKLRKLRKLRK
Regulation21CFR 872.557021CFR 872.557021CFR 872.557021CFR 872.5570
Use of deviceRemovable intraoral device.Single patient multiple use.Prescription use only.Removable intraoral device.Single patient multiple use.Prescription use only.Removable intraoral device.Single patient multiple use.Prescription use only.Removable intraoral device.Single patient multiple use.Prescription use only.
Target PopulationPeople over 18 years of agewho snore and/or have sleepapnea.People over 18 years of agewho snore and/or have sleepapnea.People over 18 years of agewho snore and/or have sleepapnea.People over 18 years of agewho snore and/or have sleepapnea.
Environment of UseTo be used in the patient'shome or in sleeplaboratories.To be used in the patient'shome or in sleeplaboratories.To be used in the patient'shome or in sleeplaboratories.To be used in the patient'shome or in sleeplaboratories.
FeatureO2Vent WO2Vent TOVENTSOMNODENT CLASSIC
CleaningShould be cleaned daily inShould be cleaned daily inShould be cleaned daily inShould be cleaned daily in
Instructionssoap and water.soap and water.soap and water.soap and water.
Can be periodically cleanedCan be periodically cleanedCan be periodically cleanedCannot be cleaned in an
in an ultrasonic cleaner within an ultrasonic cleaner within an ultrasonic cleaner withultrasonic cleaner.
an effervescent denturean effervescent denturean effervescent denture
cleaning tablet.cleaning tablet.cleaning tablet.
DeviceRepositions the lower jawRepositions the lower jawRepositions the lower jawRepositions the lower jaw
Functionalityforward.forward.forward.forward.
Acts by increasing theActs by increasing theActs by increasing theActs by increasing the
pharyngeal space topharyngeal space topharyngeal space topharyngeal space to
improve the patient's abilityimprove the patient's abilityimprove the patient's abilityimprove the patient's ability
to exchange air.to exchange air.to exchange air.to exchange air.
Does not have a lingualDoes not have a lingualHas a lingual flange forDoes not have a lingual
flange for holding theflange for holding theholding the mandibleflange for holding the
mandible forward.mandible forward.forward1.mandible forward.
Permits patient to breathePermits patient to breathePermits patient to breathePermits patient to breathe
through their mouth2.through their mouth2.through their mouth2.through their mouth2.
Retains the top and bottomRetains the top and bottomRetains the top and bottomRetains the top and bottom
teeth using rigid trays.teeth using rigid trays.teeth with monobloc.teeth using rigid trays.
Device DesignCustomized for each patientCustomized for each patientCustomized for each patientCustomized for each patient
in a dental laboratory.in a dental laboratory.in a dental laboratory.in a dental laboratory.
Upper and lower trays areUpper and lower traysNo upper and lower trays.Upper and lower trays
separate.separate.separate.
The upper tray haveThe two trays are connectedThe device is a one pieceThe protrusions (wings) are
protrusions (wings) at thein the front of the deviceassembly.at the rear of the bottom
rear on the titanium part towith a screw assembly.tray to interface with the
interface with the adjusteradjuster block in the upper
block in the lower tray.tray.
FeatureO₂Vent WO₂Vent TOVENTSOMNODENT CLASSIC
AdjustmentCan be adjusted by the clinician and patient⁴.Lower jaw adjusted using titration of the adjustment screws equally on both sides with an adjustment key.Can be adjusted by the clinician and patient⁴.Lower jaw adjusted using titration of the adjustment screw with an adjustment key.Cannot be adjusted by the clinician or patient⁴.Lower jaw cannot be adjusted and there is no adjustment key.Can be adjusted by the clinician and patient⁴.Lower jaw adjusted using titration of the adjustment screws equally on both sides with an adjustment key.
SuppliedSterile/Non-SterileNon-SterileNon-sterileNon-sterileNon-sterile
Materials Used - 1Medical grade metal used⁵.Medical grade metal used⁵.Medical grade metals used⁵.Medical grade metals used⁵.
Materials Used - 2Dental plastic laminates and acrylics used for upper and lower trays which is in contact with the patient's teeth.Dental plastic laminates and acrylics used for upper and lower trays which is in contact with the patient's teeth.Dental plastic laminates and acrylics used for upper and lower trays which is in contact with the patient's teeth.Dental acrylics used for upper and lower trays which is in contact with the patient's teeth.
Biocompatibility*Not performed as the materials are identical as in the OVENT.Not performed as the materials identical as in the OVENT.Passes Part 5 and Part 10 of ISO10993.Information could not be verified.

{6}------------------------------------------------

{7}------------------------------------------------

  • This is assumed as the predicate devices are FDA 510(k) cleared and are currently on the OVENT device was tested as part of this submission. Biocompatibility was not performed on the O Vent W device as the materials used in the device were similar to that of OVENT.

Note (1) The OVENT has a choice of lingual flance (to be decided by the clinician) although for most cases adequate retention (in advanced position) is achieved by the molded polymer insert.

Note (2) The O2Vent W, Q2Vent T and OVENT all have a dedicated breathing port (delivering air to the mouth) to allow the user to breathe through their mouth if they want, the SOMNODENT CLASSIC allows users to open their mouth to a degree.

Note (4) The OVENT requires a new insert to be made for mandbular adjustment, the O2Vent W, O2Vent T and SOMNODENT CLASSIC device can be adjusted by the clinician or patient through the adjustment screws and the adjustment key.

Note (5) The OxVent W, O2Vent T and OVENT use 3D printed titanium trays and stainless steel screws (only for O2 Vent T) and the SOMNODENT CLASSIC uses stainless steel screws and ball clasps.

{8}------------------------------------------------

Summary of Non ClinicalTestsThe non-clinical testing included risk assessment of the physicalproperties of the O2Vent W and its ability to achieve its intendeduse.Preliminary testing was carried out to confirm that the designcan withstand forces applied during normal use and alsoreasonable abuse in cases such as bruxing. The O2Vent W metthe specifications set.
The only new material that was incorporated in the O2Vent Ware the adjustment screws which have been granted prior 510(k)approval, made of special grade stainless steel and have beenwidely used in manufacturing of dental appliances. Thespecifications of the adjuster screws are same as in the predicateSOMNODENT CLASSIC device. No other risks are introducedin the device. The O2Vent W met the specifications as set.
A biocompatibility assessment of the device was performed. Thepurpose of the biocompatibility assessment was to ensure thatbiocompatibility had been established for the device.No new materials are being used in the device; all materials arealready used in the predicate devices (O2Vent T & OVENT).No new risks are introduced in the new device apart from thetwo adjustment screws on the lower tray, which are not presentin the predicate O2Vent T & OVENT devices, however arepresent in the SOMNODENT CLASSIC predicate device. Thesescrews have been granted prior 510(k) approval, made of specialgrade stainless steel and have been widely used in themanufacturing of dental appliances.
Our conclusion is that biocompatibility testing of the adjustmentscrews is not warranted due to the source and nature of thecomponent, it has long been used in intraoral environment (i.e.mucosal, gingival, and palatal) with prolonged contact and hasbeen widely used in the manufacturing of dental appliances.
The device is deemed to be biocompatible, based on thesimilarity of the materials of constructions to the predicatedevices (O2Vent T, OVENT & SOMNODENT CLASSIC).
Summary of Clinical TestsHuman clinical studies were not deemed necessary to evaluatethe substantial equivalence of the O2Vent W, as the O2Vent Wdoes not:• Use designs dissimilar from the predicate device and otherpreviously-cleared devices under a 510(k);
• Use new technologies different from legally-marketed intra-mandibular repositioning devices for snoring and/or obstructivesleep apnea;
• Deviate from the indications for use identified in the predicatedevices: the O2Vent T, OVENT and SOMNODENT CLASSIC.In lieu of human clinical testing, the risks and mitigatingcontrols associated with the use of mandibular repositioningdevices, as identified by the FDA, have been addressed in the"Risk Assessment". The function of mandibular repositioningdevices required that the prescribing dentist be aware of thepotential for soreness caused by mandibular advancement.Management of these risks is achieved by advising the patientsand dentists in the "Instructions for Use". The contraindications,warnings, precautions, storage, fitting and adjustment directionsare written to avoid potential problems from arising or persistingcaused by Mandibular Advancement Devices.
Differences between theproposed and predicatedevice(s).As discussed in the comparison chart above, the main functionaldifference between the proposed O2Vent W and predicate devicesO2Vent T & OVENT is the method of adjustment:• O2Vent W allows protrusion of the mandible from theoriginal treatment position via use of adjuster assemblylocated on both sides of the lower arch, whereas O2Vent Tallows protrusion via a single screw mechanism in the frontof the device and additional impressions or laboratoryprocedures are required for OVENT.• The adjustments can be made by the clinician or by thepatient under the direction of the clinician for the O2VentW and O2Vent T device, however it is not possible in caseof the OVENT device.
• Additionally, in order to facilitate the titration of the device,the O2Vent W have a separate upper tray with sideprotrusions (wings) and lower tray with the adjusterassembly on both the side that interface with theprotrusions, while O2Vent T has separate upper and lowertrays connected via the screw mechanism in the front of thedevice, and the OVENT has a monobloc design.
No new risks are introduced in the new device apart from the wingsand two adjustment screws, which are not present in the predicateO2Vent T and OVENT device, however are present in theSOMNODENT CLASSIC predicate device.
The treatment outcomes for O2Vent W, O2Vent T and OVENT areexpected to be equivalent as the final levels of mandibularprotrusion and the provision for oral breathing via the customizedairway is same for each device.
The difference between the proposed O2Vent W and predicatedevice SOMNODENT CLASSIC is the location of the wings and
titration blocks, which is switched between the upper and lowertrays. The purpose of the wings and the titration blocks is tostabilize the mandible in a more advanced position and thus be ableto advance the mandible forward incrementally to reduce airwaycollapse during sleep. The fact that the wing is on the top part ofthe device as opposed to the bottom is inconsequential as itachieves the same purpose of mandibular advancement in thisregard.
Hence the differences mentioned are unlikely to adversely orotherwise affect the expected clinical outcomes for patients usingthe O2Vent W device as compared to those using the predicatedevices, thus not affecting the substantial equivalency of theproposed device.
Conclusion on Non Clinicaland Clinical TestsThe O2Vent W is considered to be substantially equivalent to thepredicate devices based on the following:• It has the same intended use and is indicated for the same userpopulation.· It has equivalent technological characteristics to the predicates.· The device is as safe, as effective and performs as well as orbetter than the legally marked devices identified above.

{9}------------------------------------------------

{10}------------------------------------------------

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”