K Number
K171072
Date Cleared
2017-05-12

(32 days)

Product Code
Regulation Number
862.1225
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

N Latex Cystatin C: N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases. N Protein Standard UY: N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of a1microglobulin and Cystatin C using the BN Systems.

Device Description

The N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium heparinized plasma by means of particle-enhanced immunonephelometry using the BN II and BN ProSpec® Systems. Used in conjunction with other clinical and laboratory findings, N Latex Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Used in conjunction with the assay reagents, N Protein Standard UY is for use in the establishment of reference curves for the determination of cystatin C on the BN II and BN ProSpec Systems. N Cystatin C Control 1 and 2 are included with N Latex Cystatin C kit and are for use as assayed accuracy controls and precision controls in the determination of cystatin C by immunonephelometry with the BN II and BN ProSpec Systems. The N Latex Cystatin C test system is based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with specific antibodies to human cystatin C are aggregated when mixed with samples containing human cystatin C. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

AI/ML Overview

Acceptance Criteria and Device Performance for N Latex Cystatin C

This document summarizes the acceptance criteria and performance study results for the N Latex Cystatin C device, as described in the provided FDA 510(k) summary (K171072).

1. Table of Acceptance Criteria and Reported Device Performance

The provided document details performance studies but does not explicitly define a "table of acceptance criteria". Instead, it describes established methodologies and states that "Results met the pre-established acceptance criteria." for the method comparison study. Based on the reported data, the following can be inferred as performance metrics that met the sponsor's internal acceptance criteria for substantial equivalence:

Performance CharacteristicAcceptance Criteria (Inferred from reported data)Reported Device Performance (N Latex Cystatin C)
PrecisionCVs within acceptable ranges for repeatability and within-device/lab precision, especially near the medical decision point.BN ProSpec: - CysC Control 1: Repeatability CV 1.78%, Within-device/lab CV 1.78% - CysC Control 2: Repeatability CV 1.00%, Within-device/lab CV 1.01% - Serum Pool 1: Repeatability CV 3.85%, Within-device/lab CV 3.85% - Serum Pool 2: Repeatability CV 1.39%, Within-device/lab CV 1.41% - Serum Pool 3: Repeatability CV 1.62%, Within-device/lab CV 1.62% - Plasma Pool 1: Repeatability CV 2.38%, Within-device/lab CV 2.65% - Plasma Pool 2: Repeatability CV 2.09%, Within-device/lab CV 2.09% - Plasma Pool 3: Repeatability CV 1.07%, Within-device/lab CV 1.26% BN II: - CysC Control 1: Repeatability CV 1.93%, Within-device/lab CV 2.24% - CysC Control 2: Repeatability CV 1.42%, Within-device/lab CV 1.54% - Serum Pool 1: Repeatability CV 2.54%, Within-device/lab CV 2.62% - Serum Pool 2: Repeatability CV 1.34%, Within-device/lab CV 1.55% - Serum Pool 3: Repeatability CV 1.69%, Within-device/lab CV 1.80% - Plasma Pool 1: Repeatability CV 2.59%, Within-device/lab CV 2.59% - Plasma Pool 2: Repeatability CV 1.62%, Within-device/lab CV 1.75% - Plasma Pool 3: Repeatability CV 1.98%, Within-device/lab CV 2.42%
Measuring Range (Linearity)Acceptable linearity over the specified range.Demonstrated linearity over the range of 0.27 to 10.3 mg/L.
Limit of Quantitation (LoQ)LoQ at a specified %CV (e.g., <8% CV).LoQ demonstrated to be 0.08 mg/L at < 8% CV.
High Dose Hook Effect (Antigen Excess)No high dose hook or prozone effects up to a specified concentration.No high dose hook or prozone effects up to 42.91 mg/L.
Method Comparison (Correlation Coefficient)High correlation (e.g., r close to 1).BN ProSpec vs. Tina-quant: r = 0.988 BN II vs. Tina-quant: r = 0.998
Method Comparison (Slope)Slope close to 1 (indicating proportional agreement with predicate).BN ProSpec vs. Tina-quant: Slope = 0.982 BN II vs. Tina-quant: Slope = 0.981
Method Comparison (Intercept)Intercept close to 0 (indicating lack of constant bias with predicate).BN ProSpec vs. Tina-quant: Intercept = -0.078 mg/L BN II vs. Tina-quant: Intercept = -0.079 mg/L

2. Sample Size Used for the Test Set and Data Provenance

  • Precision: The precision studies used 80 replicates for each control and serum/plasma pool (e.g., CysC Control 1, Serum Pool 1, Plasma Pool 1, etc.) on both the BN ProSpec and BN II instruments. The data provenance is Siemens Healthcare Diagnostics Products, Marburg, Germany (internal study).
  • Measuring Range (Linearity): Not explicitly stated, but "Each dilution was tested in replicates of five using an individual aliquot." The exact number of dilutions is not provided. Data provenance is Siemens Healthcare Diagnostics Products, Marburg, Germany (internal study).
  • Limit of Quantitation (LoQ): Not explicitly stated, but implies multiple measurements according to CLSI EP17-A2. Data provenance is Siemens Healthcare Diagnostics Products, Marburg, Germany (internal study).
  • High Dose Hook Effect: "Each dilution was run five times." The exact number of dilutions is not specified. Data provenance is Siemens Healthcare Diagnostics Products, Marburg, Germany (internal study).
  • Reference Interval Study: 203 serum samples from a US population. This data is prospective (samples measured specifically for this study).
  • Method Comparison: 186 serum samples. The study was conducted at one external site in the United States. This data is implicitly prospective (samples measured for the comparison).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not describe the use of "experts" to establish ground truth in the traditional sense for diagnostic device performance (e.g., for image interpretation). Instead, the performance studies, particularly method comparison, relied on comparison to a legally marketed predicate device (Tina-quant Cystatin C Gen.2) and established reference methods (traceability to ERM-DA471/IFCC).

4. Adjudication Method for the Test Set

Adjudication methods (e.g., 2+1, 3+1) are typically used for subjective assessments or when there's ambiguity in ground truth determination, often in scenarios involving human interpretative tasks. This is not applicable to the quantitative analytical performance studies described for the N Latex Cystatin C, where measurements are objective.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. The provided document describes the performance of a diagnostic assay (N Latex Cystatin C) used for quantitative determination of cystatin C. MRMC studies are typically for evaluating diagnostic imaging devices or other tools where human interpretation plays a significant role, and there's interaction between human readers and the AI/device output. This is not relevant for this type of in vitro diagnostic device.

6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the performance studies detailed (e.g., precision, linearity, LoQ, high dose hook effect, method comparison) represent the standalone performance of the N Latex Cystatin C assay on the BN II and BN ProSpec systems. These are automated systems that produce quantitative results based on the assay's chemical reactions, without direct human intervention in the result determination beyond instrument operation and sample loading.

7. The Type of Ground Truth Used

  • Precision, Linearity, LoQ, High Dose Hook Effect: The ground truth for these analytical performance studies is based on consensus reference values (e.g., value of controls, spiked samples) and the inherent performance characteristics of the instrument and assay method itself. The method relies on the principle of particle-enhanced immunonephelometry, where the intensity of scattered light is proportional to the concentration of cystatin C, which is the "ground truth" being measured by the method.
  • Reference Interval Study: The ground truth for establishing the reference interval was the measured Cystatin C concentrations in serum samples from a "US population of apparently healthy adults or patients with diseases not linked to renal disease or failure."
  • Method Comparison: The ground truth for the method comparison study was the results obtained from the legally marketed predicate device, Tina-quant Cystatin C Gen.2 assay, on Roche cobas systems. The new device's performance was evaluated against this established method.
  • Traceability: The ultimate ground truth/standardization for Cystatin C measurement for the device is the IFCC European Reference Material ERM-DA471/IFCC.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an in vitro diagnostic assay, not a software algorithm that learns from data in that manner. The development and optimization of the assay would have involved various internal studies and analytical evaluations, which could be considered analogous to "training" in a broader sense of product development, but not in the specific context of AI model training.

9. How the Ground Truth for the Training Set Was Established

As noted above, there is no "training set" in the common AI/machine learning sense described in this 510(k) summary. The "ground truth" for the development of the N Latex Cystatin C assay reagents and calibrators involves established chemical and immunological principles, and calibration traceable to a recognized international reference material (ERM-DA471/IFCC). The "ground truth" in this context refers to the accurate and precise measurement of cystatin C concentrations based on validated methodologies and standards.

{0}------------------------------------------------

November 7, 2017

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Siemens Healthcare Diagnostics Products GmbH Christine Perkins Regulatory Specialist Emil-von-Behring-Str. 76 Marburg, 35041 De

Re: K171072

Trade/Device Name: N Latex Cystatin C; N Protein Standard UY Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: NDY, JIX Dated: April 6, 2017 Received: April 10, 2017

Dear Christine Perkins:

This letter corrects our substantially equivalent letter of May 12, 2017.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of

Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171072

Device Name

N Latex Cystatin C; N Protein Standard UY

Indications for Use (Describe)

N Latex Cystatin C:

N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases.

N Protein Standard UY:

N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of a1microglobulin and Cystatin C using the BN Systems.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary per 21 CFR 807.92

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92 and follows the FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", issued July 28, 2014.

The assigned 510(k) number is: K171072

5.1 Submitter

Siemens Healthcare Diagnostics Products GmbH Emil-von-Behring-Str. 76 35041 Marburg, Germany

Contact Person:Christine Perkins
Email:christine.perkins@siemens.com
Phone:302-631-8811
Fax:302-631-6299
Date of Preparation:May 11, 2017

5.2 Device Information

Trade Name:N Latex Cystatin C
Common or Usual Name:Test, Cystatin C Creatinine test system
Classification Name:Creatinine test system (21CFR 862.1225)
Regulatory Class:Class II
Product Code:NDY
510(k) Review Panel:Clinical Chemistry
Trade Name:N Protein Standard UY
Common or Usual Name:Calibrator, Multi-Analyte Calibrator
Classification Name:Calibrator (21 CFR 862.1150)

sification Name: Class II Regulatory Class: Product Code: الا 510(k) Review Panel: Clinical Chemistry

Calibrator (21 CFR 862.1150)

5.3 Predicate Device:

The predicate device is the Tina-quant Cystatin C Gen.2 Test System, for use on Roche/Hitachi cobas c systems cleared by the FDA under K141143.

{4}------------------------------------------------

Assay:Tina-quant Cystatin C Gen.2
Calibrator:C.f.a.s. (Calibrator for automated systems) Cystatin C

5.4 Device Description / Test Principle

The N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium heparinized plasma by means of particle-enhanced immunonephelometry using the BN II and BN ProSpec® Systems. Used in conjunction with other clinical and laboratory findings, N Latex Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases.

Used in conjunction with the assay reagents, N Protein Standard UY is for use in the establishment of reference curves for the determination of cystatin C on the BN II and BN ProSpec Systems. N Cystatin C Control 1 and 2 are included with N Latex Cystatin C kit and are for use as assayed accuracy controls and precision controls in the determination of cystatin C by immunonephelometry with the BN II and BN ProSpec Systems.

The N Latex Cystatin C test system is based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with specific antibodies to human cystatin C are aggregated when mixed with samples containing human cystatin C. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

The N Latex Cystatin C subject of this 510(k) is materially the same as the assay cleared in K041878. The reason for the submission is for a new standardization to ERM-DA471/IFCC. This is being accomplished solely by changing the calibrator value assignment process.

5.5 Intended Use / Indications for Use

N Latex Cystatin C assay

N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases.

N Protein Standard UY

N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of α1-microglobulin and Cystatin C using the BN Systems.

{5}------------------------------------------------

Comparison of Technological Characteristics

Predicate Device:
AttributesProposed Device:Roche Tina-quant Cystatin
N Latex Cystatin C assayC Gen.2
(K141143)
Intended Use / Indicationsfor UseN Latex Cystatin C is an invitro diagnostics kit containingreagents for the quantitativedetermination of cystatin C inhuman serum and lithium-heparinized plasma by meansTina-quant Cystatin C Gen. 2is an in vitro test for thequantitative determination ofcystatin C in human serumand plasma on Roche/Hitachicobas c systems.
of particle-enhancedimmunonephelometry usingthe BN Systems.Cystatin C measurements areused in the diagnosis andCystatin C measurements areused as an aid in thediagnosis and treatment ofrenal diseases.
treatment of renal diseases.
Analytecystatin CSame
Assay type/ TechnologyImmunonephelometryTurbidimetry
Sample TypeHuman serum and plasmaHuman serum and plasma
PermissibleLi-HeparinLi-Heparin,
anticoagulantsK2-EDTA, K3-EDTA
CompositionN Latex Cystatin C consists ofa suspension of polystyreneparticles coated withapproximately 0.016 g/L anti-human cystatin C from rabbitsReagent 2 is latex particles inqlycine buffer coated with anti-Cystatin C antibodies (rabbit);preservatives and stabilizers.
Ancillary ReagentsN Cystatin C SupplementaryReagent A contains rabbitimmunoglobulin (14 g/L) inbuffered solution.N Cystatin C SupplementaryReagent B consists of anaqueous solution ofpolyethylene glycol sorbitanmonolaureate (85 g/L) andpolyethylene glycol ether (27g/L).R1 contains solution ofpolymers in MOPS-bufferedsaline; preservative,stabilizers.
Unitsmg/LSame
Analytical measuring0.27 - 9.4 mg/L0.40 - 6.80 mg/L
range0.40 - 13.6 mg/L (1:2 dilution)
(Calibrator lot dependent)
Reference Interval0.49 - 1.19 mg/LAge 21-72Age 21 - 77 0.61 - 0.95 mg/L

{6}------------------------------------------------

Proposed Device:N Protein Standard UYPredicate Device:C.f.a.s. Cystatin C(K141143)
Intended UseN Protein Standard UY isused for preparing referencecurves for theimmunonephelometricdetermination of α1-microglobulin and Cystatin Cusing the BN Systems.Calibrator for automatedsystems Cystatin C is for usein the calibration ofquantitative Roche methodson Roche clinical chemistryanalyzers.
MatrixN Protein Standard UYconsists of lyophilizedpolygeline with urinaryproteins of human origin.Contains sodium azide (<1g/L) as a preservative.C.f.a.s. Cystatin C is based onpooled delipidated humanserum enriched withrecombinant human CystatinC produced in E. Coli.
How suppliedLyophilizedLiquid
Analyte Detectedcystatin C,α1-microglobulin*cystatin C
TraceabilityERM-DA417/IFCCSame

*N Protein Standard UY α1-microglobulin analyte was initially cleared in Κ991075 and has not been modified (in terms of composition, value assignment and stability) for the α1-microglobulin analyte since.

The differences between the predicate devices and proposed reagents, calibrators and controls do not result in a change to the intended use, the indications for use, or to safety and efficacy when used according to the product labeling.

5.7 Performance Data

Analytical performance studies of N Latex Cystatin C were performed at Siemens Healthcare Diagnostics Products, Marburg Germany on a BN II and a BN ProSpec analvzer.

Matrix comparison, interference, specificity and stability claims were not impacted by the new standardization. The claims are unchanged since clearance (K041878).

The following performance data are provided in support of a substantial equivalence determination.

5.7.1 Analytical Performance

5.7.1.1 Precision

Confirmation of precision studies was performed internally on one BN II and one BN ProSpec instrument for twenty days following the scheme of two runs per day, with two

{7}------------------------------------------------

replicates per run according to CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. The order of the analysis of parameter samples and quality control samples, between morning and afternoon runs, was varied so as not to add an inherent bias to the study.

Precision samples included serum and plasma pools and N Cystatin C Control 1 and 2, spanning the measuring range with sample pools close to the medical decision point (MDP) of 1.1 mg/L.

BN ProSpec Anova
IDUnitnMean valueRepeatabilityWithin-
CVdevice/lab
[%]CV
1%
CysC Control 1mg/L801.1551.781.78
CysC Control 2mg/L802.1761.001.01
Serum Pool 1mg/L800.4653.853.85
Serum Pool 2mg/L801.1421.391.41
Serum Pool 3mg/L808.4251.621.62
Plasma Pool 1mg/L800.4812.382.65
Plasma Pool 2mg/L801.1292.092.09
Plasma Pool 3mg/L808.8631.071.26
BN II Anova
IDUnitnMean valueRepeatabilityCV[%]Within-device/labCV[%]
CysC Control 1mg/L801.0701.932.24
CysC Control 2mg/L802.1101.421.54
Serum Pool 1mg/L800.4452.542.62
Serum Pool 2mg/L801.0781.341.55
Serum Pool 3mg/L808.6361.691.80
Plasma Pool 1mg/L800.4512.592.59
Plasma Pool 2mg/L801.0691.621.75
Plasma Pool 3mg/L808.7981.982.42

{8}------------------------------------------------

5.7.1.2 Measuring Range (Linearity and LoQ)

A linearity study was performed according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. One N Latex Cystatin C reagent lot, one BN II and one BN ProSpec instrument were used.

A concentrated Cystatin C stock solution was used to spike pooled human serum to a Cystatin C concentration between 10 ± 0.5 mg/L which is just above the upper limit of the measuring interval of 9.4 mg/L. This high pool was serially diluted with human serum containing undetectable levels of Cystatin C down to the minimum measuring range. Each dilution was tested in replicates of five using an individual aliquot. The N Latex Cystatin C assay demonstrates acceptable linearity over the range of 0.27 to 10.3 mg/L.

Limit of Quantitation (LoQ) testing was performed according to CLSI Guideline EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. LoQ was demonstrated to be 0.08 mg/L at < 8% CV.

The measuring range of the N Latex Cystatin C assay is 0.27 to 9.4 mg/L.

5.7.1.3 High Dose Hook Effect (Antigen Excess)

A serum sample was spiked with purified cystatin C to obtain a sample with a calculated concentration higher than 28 mg/L. A series of dilutions was prepared using cystatin C deficient serum. Each dilution was run five times. The analyte concentration of the high sample was determined using the mean of at least two manual dilutions that recovered within the measuring range of the assay. The results of the dilutions are compared to the upper limit of the calibration curve. Based on the results, Cystatin C concentrations up to 42.91 mg/L did not exhibit any high dose hook or prozone effects.

5.7.2 Clinical Studies

5.7.2.1 Reference Interval Study

A reference interval was performed according to CLSI EP28-A3C, ' Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory'.

Two hundred three (203) serum samples from a US population of apparently healthy adults or patients with diseases not linked to renal disease or failure, between the ages of 21 and 72 and evenly distributed across gender were measured. The data analysis within the 2.5% and 97.5% percentile resulted in a range of Cystatin C from 0.49 to 1.19 mg/L.

{9}------------------------------------------------

The sponsor recommends that each facility should determine its own reference intervals since values may vary depending on the population studied.

5.7.2.2 Method comparison

A method comparison study designed according to CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples: Approved Guideline-Third Edition was conducted at one external site in the United States.

Serum samples (186) were measured on both the predicate device. Roche Tina-quant Cystatin C Gen.2 assay as well as on the N Latex Cystatin C assay. Results were compared by Passing-Bablok regression analysis. Results met the pre-established acceptance criteria for slope, intercept and Pearson correlation coefficient (r). The following summary of Passing-Bablok regression analysis proves substantial equivalence between the N Latex Cystatin C assay and the predicate device.

MethodComparison(Devices)Correlation-coefficient(r)SlopeIntercept
N Latex Cystatin C on BNProSpecvs.Tina-quant Cystatin CGen.2 on Roche cobas0.9880.982-0.078mg/L
N Latex Cystatin C on BN IIvs.Tina-quant Cystatin CGen.2on Roche cobas0.9980.981-0.079mg/L

5.8 Traceability

The calibration of the assay is traceable to the IFCC European Reference Material ERM-DA471/IFCC, certified for cystatin C measurements.

5.8.1 Calibrator Traceability

N Protein Standard UY is traceable to Siemens internal Master Calibrator which is directly traceable to ERM-DA471/IFCC.

5.9 Conclusion

The N Latex Cystatin C assay including reagents and calibrator and the Roche Tinaquant Cystatin C Gen.2 assay and calibrator are substantially equivalent in principle and

{10}------------------------------------------------

performance based on intended use and indications for use and supported by the performance testing described in the 510(k) submission.

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.