(75 days)
Not Found
No
The description focuses on standard oscillometric blood pressure measurement and data transmission for analysis by separate software, with no mention of AI or ML algorithms.
No.
The device is strictly for measurement and monitoring of blood pressure and heart rate, and is not intended for the treatment or therapy of any condition.
Yes
The device measures and records blood pressure and heart rate, and its data can be transmitted for "further evaluation or archiving purposes by analysis software" to assist in assessing a patient's health condition. While it states that diagnosis "depends not alone from the measurement of the TONOPORT VI," it contributes to the diagnostic process by providing essential physiological data.
No
The device description explicitly mentions hardware components such as an electrical pump, cuff, and batteries, indicating it is a physical device with integrated software, not a software-only device.
Based on the provided information, the TONOPORT VI is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to provide information about a person's health. These tests are performed outside of the body (in vitro).
- TONOPORT VI Function: The TONOPORT VI is a non-invasive blood pressure monitor. It measures blood pressure directly on the patient's upper arm using an oscillometric method. This is a measurement taken on the body (in vivo), not on a sample taken from the body.
- Intended Use: The intended use clearly states it's for "ambulatory measurement of the blood pressure by the oscillometric method" and "measuring the systolic, diastolic, mean arterial blood pressure and the heart rate of human beings." This describes a physiological measurement, not a diagnostic test performed on a sample.
Therefore, the TONOPORT VI falls under the category of a non-invasive medical device for physiological monitoring, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The TONOPORT VI is a small-size, lightweight, portable, non-invasive blood pressure (NIBP) monitor for ambulatory measurement of the blood pressure by the oscillometric method. The cuff is borne on the upper arm and an electrical pump in the device generates the pressure in the cuff. The TONOPRT VI is powered by two AA size batteries (alkaline or rechargeable NiMH batteries). The device records up to 400 blood pressure measurements at predefined intervals. In order to assess the measurements, the stored data can be transmitted via a RS-232. USB 1.1 or USB 2.0 interface to a PC for further evaluation or archiving purposes by analysis software.
TONOPORT VI is intended to be used for measuring the systolic, diastolic, mean arterial blood pressure and the heart rate of human beings.
The intended paient populations are adults and children (but not neonates) with a circumference of the upper arm in the range of 17 to 42 cm.
TONOPORT VI is a prescription device in health care medicine and is intended for use following consultation and instruction by a physician (family doctor, specialist or hospital). It can be used, if the physical condition of the patient allows an automatic, non-invasive blood pressurement. Preparation and application of the TONOPORT VI is done by medically trained staff. The patient does not operate with the TONOPORT VI by himself. TONOPORT VI is not intended to be used in intensive care medicine or for alarming of life-threatening conditions.
A measurement with TONOPORT VI can be combined with other measurements and medical examinations at the patient that a diagnosis about the patient's health condition depends not alone from the measurement of the TONOPORT VI.
Product codes (comma separated list FDA assigned to the subject device)
DXN
Device Description
The TONOPORT VI measures the blood pressure non-invasively and provides the systolic, diastolic, mean arterial blood pressure and the heart rate of human beings. The cuff is borne on the upper arm of the patient and an electrical pump in the device generates the pressure in the cuff. The TONOPRT VI is powered by two AA size batteries (alkaline or rechargeable NiMH batteries). The device records up to 400 blood pressure measurements at predefined intervals. In order to assess the measurements, the stored data can be transmitted via a RS-232, USB 1.1 or USB 2.0 interface to a PC for further evaluation or archiving purposes by analysis software. It is intended to get used by physicians (primary care physician, specialists and hospitals).
The blood pressure is measured by the oscillometric method. The criteria for this method are the pressure pulsations superimposed with every systole on the air pressure in the cuff. In order to measure the blood pressure, a blood pressure cuff wrapped around the upper arm needs to be inflated and subsequently deflated.
The TONOPORT VI is a small-sized, light weighted and portable NIBP monitor with an easy to read display.
Used materials: Polycarbonate (PC) and acrylonitrile butadiene styrene (ABS) for the case and plastic foil for the labels of the device. Tested biocompatible materials (Nylon and PVC) are used for the cuffs.
Physical properties:
Size: width: 73.0 mm / height: 27.0 mm / depth: 108.0 mm
Weight: under 190 g, incl. batteries
Connectors: cuff, USB and RS-232 connector
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Upper arm
Indicated Patient Age Range
Adults and children (but not neonates)
Intended User / Care Setting
Physician (family doctor, specialist or hospital)
Medically trained staff
Health care medicine
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Quality assurance measures were applied to the development of the TONOPORT VI including:
- Requirements specification review
- Software and hardware testing
- Safety testing
- Environmental testing
- Final verification and validation
- Compliance with performance standards
The device complies with the following standards:
- IEC 60601-1 for Safety (electrical, mechanical, chemical, thermal and radiation) and essential performance - Result: Pass
- IEC 60601-1-11 for Safety (electrical, mechanical, chemical, thermal and radiation) and essential performance - Result: Pass
- IEC 80601-2-30 for Safety (electrical, mechanical, chemical, thermal and radiation) and essential performance - Result: Pass
- IEC 60601-1-6 for Usability - Result: Pass
- IEC 62304 for Software - Result: Pass
- ISO 14971 for Risk analysis - Result: Pass
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
TONOPORT V (510(k) -No.: K012647)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be part of a single, continuous line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2017
PAR Medizintechnik GmbH & Co. KG Thomas Fischer Sachsendamm 6 10829 Berlin Germany
Re: K170966
Trade/Device Name: TONOPORT VI Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 12, 2017 Received: May 15, 2017
Dear Thomas Fischer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
1
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hillebrenner
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
2
Statement of Indications for Use 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K170966
Device Name TONOPORT VI
Indications for Use (Describe)
The TONOPORT VI is a small-size, lightweight, portable, non-invasive blood pressure (NIBP) monitor for ambulatory measurement of the blood pressure by the oscillometric method. The cuff is borne on the upper arm and an electrical pump in the device generates the pressure in the cuff. The TONOPRT VI is powered by two AA size batteries (alkaline or rechargeable NiMH batteries). The device records up to 400 blood pressurements at predefined intervals. In order to assess the measurements, the stored data can be transmitted via a RS-232. USB 1.1 or USB 2.0 interface to a PC for further evaluation or archiving purposes by analysis software.
TONOPORT VI is intended to be used for measuring the systolic, diastolic, mean arterial blood pressure and the heart rate of human beings.
The intended paient populations are adults and children (but not neonates) with a circumference of the upper arm in the range of 17 to 42 cm.
TONOPORT VI is a prescription device in health care medicine and is intended for use following consultation and instruction by a physician (family doctor, specialist or hospital). It can be used, if the physical condition of the patient allows an automatic, non-invasive blood pressurement. Preparation and application of the TONOPORT VI is done by medically trained staff. The patient does not operate with the TONOPORT VI by himself. TONOPORT VI is not intended to be used in intensive care medicine or for alarming of life-threatening conditions.
A measurement with TONOPORT VI can be combined with other measurements and medical examinations at the patient that a diagnosis about the patient's health condition depends not alone from the measurement of the TONOPORT VI.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
3
Special 510(k) TONOPORT VI
5 510(k) Summary
Table 2: Submitter information | |
---|---|
Submitter: | PAR Medizintechnik GmbH & Co. KG |
Sachsendamm 6 | |
10829 Berlin | |
Germany | |
Contact person: | Mr. Thomas Fischer |
Research and development department | |
PAR Medizintechnik GmbH & Co. KG | |
E-mail: info@par-berlin.com | |
Phone: +49 30 235070 10 | |
Fax: +49 30 2138542 | |
Date: | March 29th, 2017 |
Table 3: Device information
Table 3: Device information | |||
---|---|---|---|
Device: | Trade name: | TONOPORT VI | |
Common name: | Non-Invasive Blood-Pressure Monitor | ||
Classification name: | Non-invasive blood pressure measurement | ||
system | |||
(21 CFR 870.1130, Product Code DXN, | |||
Class II) | |||
Legally marked | |||
device: | Device name: | TONOPORT V | |
510(k) number: | K012647 |
Device Description 5.1
Table 4: Device description
| Device functions: | The TONOPORT VI measures the blood pressure non-invasively and
provides the systolic, diastolic, mean arterial blood pressure and the
heart rate of human beings. The cuff is borne on the upper arm of the
patient and an electrical pump in the device generates the pressure in
the cuff. The TONOPRT VI is powered by two AA size batteries
(alkaline or rechargeable NiMH batteries). The device records up to
400 blood pressure measurements at predefined intervals. In order to
assess the measurements, the stored data can be transmitted via a RS-
232, USB 1.1 or USB 2.0 interface to a PC for further evaluation or
archiving purposes by analysis software. It is intended to get used by
physicians (primary care physician, specialists and hospitals). |
------------------- | --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
---|
4
| PAR Medizintechnik
GmbH & Co. KG | Special 510(k) TONOPORT VI | |
---|---|---|
Basic scientific | ||
concept: | The blood pressure is measured by the oscillometric method. The | |
criteria for this method are the pressure pulsations superimposed | ||
with every systole on the air pressure in the cuff. In order to measure | ||
the blood pressure, a blood pressure cuff wrapped around the upper | ||
arm needs to be inflated and subsequently deflated. | ||
Device design: | The TONOPORT VI is a small-sized, light weighted and portable | |
NIBP monitor with an easy to read display. | ||
Used materials: | Polycarbonate (PC) and acrylonitrile butadiene styrene (ABS) for the | |
case and plastic foil for the labels of the device. Tested biocompatible | ||
materials (Nylon and PVC) are used for the cuffs. | ||
Physical | ||
properties: | Size: width: 73.0 mm / height: 27.0 mm / depth: 108.0 mm | |
Weight: under 190 g, incl. batteries | ||
Connectors: cuff, USB and RS-232 connector |
5.2 Indications for Use
The TONOPORT VI is a small-size, lightweight, portable, non-invasive blood pressure (NIBP) monitor for ambulatory measurement of the blood pressure by the oscillometric method. The cuff is borne on the upper arm and an electrical pump in the device generates the pressure in the cuff. The TONOPRT VI is powered by two AA size batteries (alkaline or rechargeable NiMH batteries). The device records up to 400 blood pressure measurements at predefined intervals. In order to assess the measurements, the stored data can be transmitted via a RS-232, USB 1.1 or USB 2.0 interface to a PC for further evaluation or archiving purposes by analysis software.
TONOPORT VI is intended to be used for measuring the systolic, diastolic, mean arterial blood pressure and the heart rate of human beings.
The intended patient populations are adults and children (but not neonates) with a circumference of the upper arm in the range of 17 to 42 cm.
TONOPORT VI is a prescription device in health care medicine and is intended for use following consultation and instruction by a physician (family doctor, specialist or hospital). It can be used, if the physical condition of the patient allows an automatic, noninvasive blood pressure measurement. Preparation and application of the TONOPORT VI is done by medically trained staff. The patient does not operate with the TONOPORT VI
5
by himself. TONOPORT VI is not intended to be used in intensive care medicine or for alarming of life-threatening conditions.
A measurement with TONOPORT VI can be combined with other measurements and medical examinations at the patient that a diagnosis about the patient's health condition depends not alone from the measurement of the TONOPORT VI.
6
AR Medizintechni
GmbH & Co. KG
oecial 510(k) TONOPORT
Comparison of the TONOPORT VI and the predicate dev 5.3
The TONOPORT VI is substantially equivalent and employs the same functional technology as the predicate device TONOPORT
Table 5: Device compariso
| Characteristic | Subject device
TONOPORT VI | Primary predicate device
TONOPORT V (510(k) -No.: K012647) | Explanation of the
differences | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | DXN | DXN | Identical | |
| Classification
number | 21 CFR 870.1130 | 21 CFR 870.1130 | Identical | |
| Common name | Non-Invasive Blood-Pressure Monitor | Non-Invasive Blood-Pressure Monitor | Identical | |
| Class | Class II | Class II | Identical | |
| Intended use | The TONOPORT VI measures and stores
the blood pressure data (systolic pressure,
diastolic pressure and heart rate) at human
beings. The blood pressure can be
measured several times at the physician or
up to 400 times during a long-term
monitoring with one set of fully charged
batteries. Medically trained staff has to
apply the device and the physician has to
analyze the results. The patient does not
operate with the TONOPORT VI by
himself. The stored data can be
transmitted to a PC based application via
serial interface for analysis and
evaluation. | The TONOPORT V measures and stores
the blood pressure data (systolic pressure,
diastolic pressure and heart rate) at human
beings. The blood pressure can be
measured several times at the physician or
up to 200 times during a long-term
monitoring with one set of fully charged
batteries. Medically trained staff has to
apply the device and the physician has to
analyze the results. The patient does not
operate with the device by himself. The
stored data can be printed directly on a
printer and/or transmitted to a PC-based
application via serial interface for analysis
and evaluation. | Equivalent
The storage capacity is
increased and the
transmission interfaces are
state of the art for the
clearance times. These
changes are documented
and do not influence the
indications for use. | |
| PAR Medizintechnik
GmbH & Co. KG | Special 510(k) TONOPORT VI | | | |
| Characteristic | Subject device
TONOPORT VI | Primary predicate device
TONOPORT V (510(k) -No.: K012647) | Explanation of the
differences | |
| Indications for use | Target population:
Adults and children, but not neonates
Anatomical site:
Upper arm (see section 0)
Where used:
In hospital and at home (following consultation and instruction by a physician) | Target population:
Adults and children, but not neonates
Anatomical site:
Upper arm
Where used:
In hospital and at home (following consultation and instruction by a physician) | Identical | |
| Design | New and good looking ergonomic design
of the case and the user interface | Functional design of the case and the user
interface | Equivalent
The changes of the case
ergonomics and the user
interface that enhance the
operator's and patient's
convenience are
documented and are not
relevant for the
indications for use. | |
| PAR Medizintechnik
GmbH & Co. KG | | Special 510(k) TONOPORT VI | | |
| Characteristic | Subject device
TONOPORT VI | Primary predicate device
TONOPORT V (510(k) -No.: K012647) | Explanation of the
differences | |
| Display | 27.0 mm x 58.0 mm
(width to height)
Image: [display] | 30.0 mm x 10.5 mm
(width to height)
Image: [display] | Equivalent
The change of the
ergonomics of the user
interface into a clear and
easy to read display that
increases the safety of the
device and enhance the
convenience are
documented. | |
| Dimensions and
weight | Smaller with lower weight compared to
TONOPORT V | Small size and low weight | Equivalent
The changes in
dimensional specifications
are documented, do not
affect the indications for
use and enhance the
operator's and patient's
convenience. | |
| Width | 73.0 mm | 80.0 mm | Identical | |
| Height | 27.0 mm | 27.0 mm | | |
| Depth | 108.0 mm | 100.0 mm | | |
| Weight | Approximately 190 g with batteries | Approximately 199 g with batteries | | |
| Communication | Measurement settings of the device and
the read out of the stored data can be done
via serial interface | Measurement settings of the device and
the read out of the stored data can be done
via serial interface | Identical | |
| Event marking | By pressing the START STOP Button | By pressing the START STOP Button | Identical | |
| Energy source | Powered by two AA size batteries
(alkaline or NiMH batteries). | Powered by two AA size batteries
(alkaline or NiMH batteries). | Identical | |
| GmbH & Co. KG | Characteristic | Subject device
TONOPORT VI | Primary predicate device
TONOPORT V (510(k) -No.: K012647) | Explanation of the
differences |
| | Storage capacity | Up to 400 measurements | Up to 200 measurements | Equivalent
The evaluation of user needs showed that more
storage capacity is need.
The change is documented
and do not influence the
indications for use. |
| | Measuring method | Oscillometric measuring method during
deflation of the cuff | Oscillometric measuring method during
deflation of the cuff | Identical |
| | Measuring range | 60 to 260 mmHg
40 to 220 mmHg
35 to 240 min-1
systolic pressure
diastolic pressure
pulse rate (HR) | 60 to 260 mmHg
40 to 220 mmHg
35 to 240 min-1
systolic pressure
diastolic pressure
pulse rate (HR) | Identical |
| | Max cuff pressure | 300 mmHg | 300 mmHg | Identical |
| | Materials | PC and ABS for the case and plastic foil
for the labels of the device. Biocompatible
materials are used for the applied parts
(cuff). | PC for the case and plastic foil for the
labels of the device. Biocompatible
materials are used for the applied parts
(cuff). | Equivalent
The change of the material
formulation to increase
the durability of the
device is documented. The
changed materials are not
in contact with the body
of the patient. |
5-4
7
ecial 510(k) TONOPORT VI - Doc.-Rev. E from June 14th, 201
K170966
Page 5 of 9
8
9
AR Medizintechni
Special 510(k) TONOPORT VI
10
AR Medizintechnil
GmbH & Co. KG
pecial 510(k) TONOPORT VR
| | Characteristic | Subject device
TONOPORT VI | | Primary predicate device
TONOPORT V (510(k) -No.: K012647) | | Explanation of the
differences |
|----------------------------------------------------------------|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------------------------------------------------------------------------------|
| Environmental
conditions
for
Operation | Temperature | +5 to +40 °C | | +10 to +40 °C | Temperature | Equivalent |
| | Humidity | 15 to 93 % | Atmospheric pressure | 30 to 75 % | Humidity | The environmental
conditions are changed |
| | Atmospheric pressure | 700 to 1060 hPa | | 700 to 1060 hPa | Atmospheric pressure | according to the |
| Environmental
conditions
for
transport and
storage | Temperature | -25 to +70 °C | | -20 to +70 °C | Temperature | requirements of the |
| | Humidity | 10 to 93 % | | 10 to 90 % | Humidity | standards IEC 80601-2-30
and 60601-1-11. These |
| | Atmospheric pressure | 500 to 1060 hPa | | 500 to 1060 hPa | Atmospheric pressure | changes do not affect the
indications for use. |
| Safety | Increased safety | An additional pressure sensor and a
pressure release valve are added.
Two controllers (main and supervisor
controller) that monitor each other and the
safety related components are used. | | Standard safety
One pressure sensor and one valve for the
deflation of the cuff are used.
One controller (main controller) is used. | | Equivalent
The improvement of the
safety properties of the
device is documented. |
| Sterilization | Not applicable | Not applicable | | Not applicable | | Identical |
| Expiration date | Not applicable | Not applicable | | Not applicable | | Identical |
11
Summary of performance testing 5.4
Characteristic | Standard/Test/FDA Guidance | Result |
---|---|---|
Safety (electrical, | ||
mechanical, chemical, | ||
thermal and radiation) and | ||
essential performance | IEC 60601-1 | |
IEC 60601-1-11 | ||
IEC 80601-2-30 | Pass | |
Pass | ||
Pass | ||
Usability | IEC 60601-1-6 | Pass |
Software | IEC 62304 | Pass |
Risk Management | ||
Risk analysis | ISO 14971 | Pass |
Table 6: Performance testing to demonstrate equivalence
The TONOPORT VI complies with the voluntary and mandatory standards. The following quality assurance measures were applied to the development of the TONOPORT VI to ensure that design inputs, safety requirements, new features, preexisting requirements were met and re-tested:
- Requirements specification review
- o Software and hardware testing
- Safety testing
- Environmental testing
- Final verification and validation
- o Compliance with performance standards
An assessment of clinical data to show safety and performance of the device was not made, because the TONOPORT VI uses identical algorithms for the blood pressure measurement like the predicate device TONOPORT V. Consequently the form FDA-3674 'Certification of compliance with requirements of ClinicalTrails.gov data bank' is not completed and submitted.
5.5 Conclusion
The TONOPORT VI is substantially equivalent to the TONOPORT V with respect to the indications for use and the basic scientific concept.