K Number
K170941
Manufacturer
Date Cleared
2017-07-07

(99 days)

Product Code
Regulation Number
870.5100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RX NC Takeru PTCA Balloon Dilatation Catheter is indication of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.

This product is also indicated for the post-delivery expansion of balloon expandable stents.

Device Description

RX NC Takeru PTCA Balloon Dilatation Catheter (RX NC Takeru) is a rapid exchange type of balloon dilation catheter, which consists of a distal tube, guidewire transition tube, balloon, radiopaque markers, mid tube, proximal tube, core wire, hub, and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. RX NC Takeru has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the guidewire transition tube to the guidewire port for the exit of a guidewire.

The maximum compatible diameter of a guidewire used together with RX NC Takeru in a PTCA procedure is 0.014 inches. Additionally, guiding catheters with a diameter of 5 or 6 Fr have been deemed to be compatible with RX NC Takeru. The nominal inflated balloon diameters range from 2.0 mm to 5.0 mm with balloon working lengths of 6 mm to 30 mm. The catheter working length is 1450 mm.

AI/ML Overview

This document describes the premarket notification (510(k)) for the RX NC Takeru PTCA Balloon Dilatation Catheter. The submission aims to demonstrate substantial equivalence to a predicate device.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test PerformedAcceptance Criteria (Implied)Reported Device Performance
Dimensional VerificationNot explicitly stated, but implies meeting design specifications for dimensions."met all the predetermined acceptance criteria of design verification"
Balloon Preparation, Deployment and RetractionNot explicitly stated, but implies proper preparation, deployment, and retraction without failure."met all the predetermined acceptance criteria of design verification"
Balloon Rated Burst PressureNot explicitly stated, but implies meeting a minimum pressure without rupture."met all the predetermined acceptance criteria of design verification"
Balloon Fatigue (Repeat Balloon Inflations)Not explicitly stated, but implies demonstrating durability over repeated inflations."met all the predetermined acceptance criteria of design verification"
Balloon ComplianceNot explicitly stated, but implies meeting specified compliance characteristics."met all the predetermined acceptance criteria of design verification"
Balloon Inflation and Deflation TimeNot explicitly stated, but implies meeting clinical performance requirements for speed."met all the predetermined acceptance criteria of design verification"
Catheter Bond StrengthNot explicitly stated, but implies meeting minimum strength requirements to prevent detachment."met all the predetermined acceptance criteria of design verification"
Flexibility and Kink TestNot explicitly stated, but implies demonstrating sufficient flexibility without kinking."met all the predetermined acceptance criteria of design verification"
Torque StrengthNot explicitly stated, but implies meeting minimum strength requirements for torque transmission."met all the predetermined acceptance criteria of design verification"
RadiopacityNot explicitly stated, but implies sufficient visibility under fluoroscopy."met all the predetermined acceptance criteria of design verification"
Coating IntegrityNot explicitly stated, but implies the coating remains intact."met all the predetermined acceptance criteria of design verification"
Particulate EvaluationNot explicitly stated, but implies acceptable levels of particulate release."met all the predetermined acceptance criteria of design verification"
Balloon Rated Burst Pressure (in Stent)Not explicitly stated, but implies meeting a minimum pressure without rupture when deployed in a stent."met all the predetermined acceptance criteria of design verification"
Balloon Fatigue (Repeat Balloon Inflations; in Stent)Not explicitly stated, but implies demonstrating durability over repeated inflations when deployed in a stent."met all the predetermined acceptance criteria of design verification"
Transportation and Shelf Life TestingNot explicitly stated, but implies maintaining device integrity and performance after transport and over its shelf life."met all the predetermined acceptance criteria of design verification"
Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen, Hemocompatibility, Genotoxicity)Adherence to ISO 10993-1 standards for biological evaluation."The results from these tests demonstrate that the RX NC Takeru is biocompatible for its intended use similar to the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

The document states that the performance testing was conducted via "in vitro tests." However, it does not specify the sample size for the test set used for any of the performance tests. The provenance of this test data is not explicitly mentioned but is implied to be from internal Kaneka Corporation testing as part of their design verification and validation. Given it's a 510(k) submission, this is typically part of the manufacturer's own testing for regulatory clearance, not from an independent study with external data provenance in the way clinical trials would.

For biocompatibility, testing was performed on a "reference device (RX Takeru)." The sample size for this is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The performance testing described consists of bench tests (in vitro) and biocompatibility tests. These types of tests typically rely on established engineering principles, international standards (e.g., ISO), and laboratory procedures, rather than human expert consensus for "ground truth" in the way a diagnostic AI algorithm might. Therefore, the concept of "experts establishing ground truth" for these specific tests would not directly apply.

4. Adjudication Method for the Test Set

This information is not provided and is generally not applicable to the types of bench and biocompatibility tests described. Adjudication methods like 2+1 or 3+1 are relevant for clinical studies where human interpretation of data (e.g., imaging, clinical symptoms) is involved and discrepancies need to be resolved to establish ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document describes bench testing and biocompatibility testing, not a clinical study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a physical medical device (PTCA Balloon Dilatation Catheter), not an AI algorithm or software device.

7. The Type of Ground Truth Used

For the performance testing (in vitro tests), the "ground truth" is established by adherence to predefined engineering specifications, international standards (e.g., FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters" and ISO 10993-1), and internal test protocols. The acceptance criteria for these tests define what constitutes acceptable performance for the device's physical properties and safety.

8. The Sample Size for the Training Set

This question is not applicable. The device is a physical medical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as the device is a physical medical device and does not have a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2017

Kaneka Corporation c/o Christopher M. Sloan Quintiles Consulting 1801 Rockville Pike, Suite 300, Rockville, MD 20852 US

Re: K170941

Trade/Device Name: RX NC Takeru Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: June 19, 2017 Received: June 20, 2017

Dear Christopher Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Sincerely,

Fernando Aguel

Fernando Aguel
-S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K170941

Device Name

RX NC Takeru PTCA Balloon Dilatation Catheter

Indications for Use (Describe)

The RX NC Takeru PTCA Balloon Dilatation Catheter is indication of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.

This product is also indicated for the post-delivery expansion of balloon expandable stents.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

5.0 510(K) SUMMARY

RX NC Takeru PTCA Balloon Dilatation Catheter

510(k) Submitter

Kaneka Corporation Medical Device Division 1-12-32, Akasaka, Minato-ku Tokyo, Japan Contact Person: Toshihiko Motomine Telephone: +81-3-5574-8023 Email:

Official Correspondent

Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Phone: (301) 272-3114 Email: chris.sloan@quintilesims.com

Date Prepared: June 9, 2017

{4}------------------------------------------------

Subject Device Name:

Trade NameRX NC Takeru
Common or usual namePercutaneous Transluminal CoronaryAngioplasty (PTCA) catheter
Classification namePercutaneous Transluminal CoronaryAngioplasty (PTCA) catheter [21 CFR870.5100; product code LOX]
ClassII
Classification PanelCardiovascular (74)

Predicate Device:

  • NC Quantum Apex™ Monorail PTCA Dilatation Catheter or "NC Quantum Apex" ● [P860019/S241 (Boston Scientific Corporation)]

Reference Device:

  • RX Takeru PTCA Balloon Dilatation Catheter or "RX Takeru" [K163372 (Kaneka Corporation)]

Device Description:

RX NC Takeru PTCA Balloon Dilatation Catheter (RX NC Takeru) is a rapid exchange type of balloon dilation catheter, which consists of a distal tube, guidewire transition tube, balloon, radiopaque markers, mid tube, proximal tube, core wire, hub, and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. RX NC Takeru has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the guidewire transition tube to the guidewire port for the exit of a guidewire.

The maximum compatible diameter of a guidewire used together with RX NC Takeru in a PTCA procedure is 0.014 inches. Additionally, guiding catheters with a diameter of 5 or 6 Fr have been deemed to be compatible with RX NC Takeru. The nominal inflated balloon diameters range from 2.0 mm to 5.0 mm with balloon working lengths of 6 mm to 30 mm. The catheter working length is 1450 mm.

{5}------------------------------------------------

Indications for Use

The RX NC Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.

This product is also indicated for the post-delivery expansion of balloon expandable stents.

Comparison of Indications for Use to Predicate Device

The RX NC Takeru has the same intended use (percutaneous transluminal coronary angioplasty) as the NC Quantum Apex. The indications for use of the RX NC Takeru is similar to those of the NC Quantum Apex. T Therefore, the subject device, RX NC Takeru, may be considered substantially equivalent to the predicate device.

Comparison of Technological Characteristics to Predicate Device

Percutaneous Transluminal Coronary Angioplasty (PTCA) is the technological principle for both RX NC Takeru and the predicate device. PTCA is based on the use of Percutaneous Coronary Intervention (PCI) devices for the purpose of myocardial perfusion.

The RX NC Takeru and predicate device have following same technological elements:

  • Operating principle balloon dilatation of stenotic portion by pressurization of . inflation medium
  • . Fundamental catheter design - balloon, shaft, radiopaque marker, hub, hydrophilic coating
  • Shaft type - rapid exchange
  • Concomitantly used devices guidewire, guiding catheter, inflation device ●
  • . Sterilization - Ethylene oxide

There are following minor technological differences between RX NC Takeru and predicate device:

  • Combination of balloon diameter and balloon length .
  • Shaft diameters
  • Catheter effective length
  • . Nominal Pressure and Rated Burst Pressure of certain balloon sizes

Performance Testing

To demonstrate substantial equivalence of RX NC Takeru to the predicate device, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA)

{6}------------------------------------------------

Catheters" dated September 8, 2010. The following in vitro tests were performed on the subject device:

Dimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Radiopacity, Coating Integrity, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), Transportation and Shelf Life Testing.

The results from these tests demonstrate that the technological characteristics and performance criteria of the RX NC Takeru are adequate for the intended use of the device and that the device can perform in a manner equivalent to devices currently on the market with the same intended use.

Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the RX NC Takeru to the predicate device, the following biocompatibility testing, which was performed on the reference device (RX Takeru) in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff" (dated June 16, 2016), was leveraged:

Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen, Hemocompatibility (in vivo thromboresistance, hemolysis, and complement), and Genotoxicity (mouse lymphoma, bacterial reverse mutation and in vivo cytogenetics assay)

The results from these tests demonstrate that the RX NC Takeru is biocompatible for its intended use similar to the predicate device.

Conclusions:

The RX NC Takeru met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The RX NC Takeru PTCA Balloon Dilatation Catheter is substantially equivalent to legally marketed predicate devices.

§ 870.5100 Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter.

(a)
Standard PTCA Catheter —(1)Identification. A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion.(2)
Classification. Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
Cutting/scoring PTCA Catheter —(1)Identification. A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.