K Number
K170915

Validate with FDA (Live)

Date Cleared
2017-11-21

(238 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To facilitate the placement of devices during diagnostic or interventional procedures.

Device Description

The TXM Hydrophilic Guidewire is a hydrophilic coated device which is constructed with a nitinol inner core which is covered with a polyurethane outer layer coated with a lubricious coating to minimize friction during use. The TXM Hydrophilic Guidewire outer diameter is 0.035" and is available in a variety of lengths ranging from 80 cm to 260 cm. The TXM Hydrophilic Guidewire is also constructed in stiff and standard shaft configuration with straight or angled distal tip. The device is supplied sterile and is intended for single use only.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (TXM Hydrophilic Guidewire), not for an AI/ML powered medical device. Therefore, the requested information regarding acceptance criteria and studies proving the device meets them in the context of AI/ML performance metrics (like sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set information) is not applicable here.

This document describes the substantial equivalence of the TXM Hydrophilic Guidewire to a predicate device (Hydrophilic Coated Guidewire, K133155). The acceptance criteria and "study" are in the context of demonstrating this substantial equivalence through bench testing, biocompatibility testing, and sterilization testing.

Here's an analysis of the provided text based on the nature of the device:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list numerical "acceptance criteria" for each test. Instead, it states that "The TXM Hydrophilic Guidewire met the predetermined acceptance criteria." and then reports "Pass" for each test, indicating conformity. The applicable standards or internal test methods implicitly define the acceptance criteria.

Test NameApplicable Standard or Internal Test MethodReported Device Performance
Dimensional and Physical AttributesISO 11070:2014Pass
Torque StrengthFDA Coronary and Cerebrovascular Guidewire Guidance / Internal Test MethodPass
TorqueabilityFDA Coronary and Cerebrovascular Guidewire Guidance / Internal Test MethodPass
Guidewire compatibilityFDA Coronary and Cererovascular Guidewire Guidance / Internal Test MethodPass
RadiopacityASTM-F640-12Pass
Tensile StrengthFDA Coronary and Cerebrovascular Guidewire Guidance / ISO 11070:2014Pass
Coating Adherence and Integrity; Particulate QuantitationFDA Coronary and Cerebrovascular Guidewire Guidance / Internal Test MethodPass
Durability of Hydrophilic CoatingInternal Test MethodPass
Lubricity of Hydrophilic CoatingInternal Test MethodPass
Shaft stiffness and Tip FlexibilityFDA Coronary and Cerebrovascular Guidewire Guidance / ASTM D747-10Pass
FractureISO 11070:2014Pass
Tip ImpactFDA Coronary and Cerebrovascular Guidewire Guidance / Internal Test MethodPass
Corrosion ResistanceISO 11070:2014Pass
TrackabilityInternal Test MethodPass
Packing Dye PenetrationASTM F1929-15/F1886-09Pass
Seal StrengthASTM F88 / F88M - 15Pass
Shipping and TransitISTA 3APass
Accelerated AgingASTM-E1980-07Pass
Biocompatibility Tests:Pass (implied by "Compliant")
CytotoxicityISO10993-5Pass (Compliant)
IrritationISO10993-10Pass (Compliant)
SensitizationISO10993-10Pass (Compliant)
Systemic ToxicityISO10993-11Pass (Compliant)
Material Mediated Pyrogenicity (Rabbit)USP<151> / ISO10993-11Pass (Compliant)
Material Mediated Pyrogenicity (LAL)USP 38, NF 33, 2015Pass (Compliant)
Hemolysis-Complete (Direct and Indirect)ISO10993-4Pass (Compliant)
Complement ActivationISO10993-4Pass (Compliant)
In-Vivo ThrombogenicityISO10993-4Pass (Compliant)
Sterilization Testing:
Ethylene Oxide GasISO 11135Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size: The document does not specify the exact number of devices tested for each bench test or biocompatibility test. It only indicates that tests were performed on "non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2)." This implies at least two groups of samples for most tests.
  • Data Provenance: Not applicable in the context of an AI/ML device. For physical device testing, the data is generated internally by the manufacturer (Texas Medical Technologies, Inc.) through their testing processes. There's no mention of country of origin for test data, as it's product performance data. The tests are prospective, as they are performed on newly manufactured devices for regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. For a physical medical device, "ground truth" is established by direct measurement and observation against engineering and biological standards, not by expert consensus in the way it is for AI/ML image interpretation. The "experts" would be the engineers, technicians, and biologists performing and interpreting the standardized tests, but their specific number and qualifications are not detailed in this type of submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. Adjudication methods are relevant for human interpretation tasks, especially in AI/ML performance evaluation where human experts might disagree. For device bench testing, the results are objectively measured against predefined criteria.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable, as this is a physical medical device, not an AI-powered diagnostic or interpretive tool. MRMC studies are used to evaluate AI performance with human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable, as this is a physical medical device, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

For this device, the "ground truth" is defined by established engineering and biological standards (e.g., ISO, ASTM, FDA Guidance documents, USP). Deviations from these standards constitute a "failure." For example:

  • Bench Testing: Mechanical properties (e.g., tensile strength, flexibility, lubricity) are measured directly and compared against predefined ranges or thresholds specified in the applicable standards.
  • Biocompatibility Testing: Biological responses (e.g., cytotoxicity, irritation, sensitization) are assessed using standardized in-vitro and in-vivo assays, and the absence of unacceptable biological reactions indicates compliance.
  • Sterilization Testing: Sterility Assurance Level (SAL) of 10^-6 is the standard ground truth for terminal sterilization.

8. The sample size for the training set:

Not applicable. This is a physical device, not an AI/ML algorithm that requires a "training set."

9. How the ground truth for the training set was established:

Not applicable. As there is no AI/ML algorithm, there is no "training set" or ground truth establishment for it. The design specifications of the device are based on engineering principles and regulatory requirements, not learned from data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 21, 2017

Texas Medical Technologies, Inc. % Mr. E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114

Re: K170915

Trade/Device Name: TXM Hydrophilic Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: October 13, 2017 Received: October 16, 2017

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Mr. Smith

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170915

Device Name TXM Hydrophilic Guidewire

Indications for Use (Describe)

To facilitate the placement of devices during diagnostic or interventional procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.1 - Company Information & Contact Person

Company Name:Texas Medical Technologies Inc.
Company Address:9005 Montana Ave. Ste. A
El Paso, Texas 79925
Telephone:(915) 774-4321
Fax:(915) 774-4323
Contact Person:Cesar Rios, Quality Assurance & Regulatory Manager
Date Prepared:02/28/2017

5.2 - Device Name & Classification

Proprietary Name:TXM Hydrophilic Guidewire
Common Name:Catheter guide wire
Classification Name:Wire, Guide, Catheter
Regulation Number:21 CFR 870.1330
Product Code:DQX
Device Class:II

5.3 - Predicate Device

Legally Marketed Substantially Equivalent Predicate Device

Proprietary Name:Hydrophilic Coated Guidewire
Company Name:Lake Region Medical
Common Name:Guides, guidewires, or spring guidewires
Classification Name:Wire, Guide, Catheter
Regulation Number:21 CFR 870.1330
Product Code:DQX
Device Class:II
510(k) NumberK133155

5.4 - Device Description

The TXM Hydrophilic Guidewire is a hydrophilic coated device which is constructed with a nitinol inner core which is covered with a polyurethane outer layer coated with a lubricious coating to minimize friction during use. The TXM Hydrophilic Guidewire outer diameter is 0.035" and is available in a variety of lengths ranging from 80 cm to 260 cm. The TXM Hydrophilic Guidewire is also constructed in stiff and standard shaft configuration with straight or angled distal tip. The device is supplied sterile and is intended for single use only.

{4}------------------------------------------------

The following table lists the models and sizes available for TXM Hydrophilic Guidewire.

Product CodeOuter Diameter(inches)Length(cm)ShaftConfigurationTip Shape
ShaftTip
HGW-35080-FS0.0335"0.0305"80StandardStraight
HGW-35150-FS0.0335"0.0305"150StandardStraight
HGW-35180-FS0.0335"0.0305"180StandardStraight
HGW-35260-FS0.0335"0.0305"260StandardStraight
HGW-35080-FA0.0335"0.0305"80StandardAngled
HGW-35150-FA0.0335"0.0305"150StandardAngled
HGW-35180-FA0.0335"0.0305"180StandardAngled
HGW-35260-FA0.0335"0.0305"260StandardAngled
HGW-35080-SS0.0335"0.0305"80StiffStraight
HGW-35150-SS0.0335"0.0305"150StiffStraight
HGW-35180-SS0.0335"0.0305"180StiffStraight
HGW-35260-SS0.0335"0.0305"260StiffStraight
HGW-35080-SA0.0335"0.0305"80StiffAngled
HGW-35150-SA0.0335"0.0305"150StiffAngled
HGW-35180-SA0.0335"0.0305"180StiffAngled
HGW-35260-SA0.0335"0.0305"260StiffAngled

Table 5.4 TXM Hydrophilic Guidewire Models and Sizes

5.5 - Indications for Use

To facilitate the placement of devices used during diagnostic and interventional procedures.

The Indications for Use are identical, the devices are both intended for intravascular use and the Indications for Use do not change the intended use of the TXM Hydrophilic Guidewire when compared to the predicate device.

5.6 - Summary of Technological Characteristics Comparison

Based on a comparison of the indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging, it is determined that the TXM Hydrophilic Guidewire is substantially equivalent to the predicate device. Table 5.6 below provides a comparison of the TXM Hydrophilic Guidewire and the predicate.

{5}------------------------------------------------

TechnicalCharacteristics /Principle ofOperationTXM HydrophilicGuidewireHydrophilic CoatedGuidewireSubstantiallyEquivalent?
Length80cm-260cm80cm-260cmYes
Outer Diameter0.035"0.035"Yes
ShapeStraight or AngledStraight or AngledYes
Shaft ConfigurationStandard or StiffStandard or StiffYes
Inner Core MaterialNitinolNitinolYes
Outer JacketMaterialPolyurethane,TungstenPolyurethane, Radiopaquematerial unknownYes
Hydrophilic CoatedYes*YesYes
Anatomical Site UsePeripheral Vascular SystemPeripheral Vascular SystemYes
Delivery to SiteThrough Introducer SheathThrough Introducer SheathYes
PackagingTyvek PouchTyvek PouchYes
SterilizationEtO GasEtO GasYes

Table 5.6 Comparison of the TXM Hydrophilic Guidewire and the Predicate Device

  • Denotes a patient-contacting material.

5.7 - Testing Summary

The following bench tests were performed to evaluate the design elements and performance characteristics of the TXM Hydrophilic Guidewire and to demonstrate substantial equivalence to the predicate device. The TXM Hydrophilic Guidewire met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2). Tests results show that the TXM Hydrophilic Guidewire is substantially equivalent to the predicate device.

5.7.1- Bench Testing Table

Table 5.7.1 below provides a summary of the bench testing performed on the TXM Hydrophilic Guidewire.

{6}------------------------------------------------

TestNo.Test NameApplicable Standard or Internal TestMethodTest Results
1Dimensional and Physical AttributesISO 11070:2014T=0T=2Pass
2Torque StrengthFDA Coronary and CerebrovascularGuidewire Guidance / Internal Test MethodT=0T=2Pass
3TorqueabilityFDA Coronary and CerebrovascularGuidewire Guidance / Internal Test MethodT=0T=2Pass
4Guidewire compatibilityFDA Coronary and CerebrovascularGuidewire Guidance / Internal Test MethodT=0T=2Pass
5RadiopacityASTM-F640-12T=0T=2Pass
6Tensile StrengthFDA Coronary and CerebrovascularGuidewire Guidance / ISO 11070:2014T=0T=2Pass
7Coating Adherence and Integrity;Particulate QuantitationFDA Coronary and CerebrovascularGuidewire Guidance / Internal Test MethodT=0T=2Pass
8Durability of Hydrophilic CoatingInternal Test MethodT=0T=2Pass
9Lubricity of Hydrophilic CoatingInternal Test MethodT=0T=2Pass
10Shaft stiffness and Tip FlexibilityFDA Coronary and CerebrovascularGuidewire Guidance / ASTM D747-10T=0T=2Pass
11FractureISO 11070:2014T=0T=2Pass
12Tip ImpactFDA Coronary and CerebrovascularGuidewire Guidance / Internal Test MethodT=0T=2Pass
13Corrosion ResistanceISO 11070:2014T=0T=2Pass
14TrackabilityInternal Test MethodT=0T=2Pass
15Packing Dye PenetrationASTM F1929-15/F1886-09T=0T=2Pass
16Seal StrengthASTM F88 / F88M - 15T=0T=2Pass
17Shipping and TransitISTA 3AT=0T=2Pass
18Accelerated AgingASTM-E1980-07T=0T=2Pass

Table 5.7.1 Bench Testing Performed on TXM Hydrophilic Guidewire

5.7.2 - Biocompatibility

The TXM Hydrophilic Guidewire is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours). Biocompatibility testing was performed in accordance with ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" (2009).

Table 5.7.2 below describes the testing performed to determine biocompatibility.

{7}------------------------------------------------

Biological EffectTestComplianceStandard
CytotoxicityL929 MEM ElutionL929 Neutral Red Uptake (NRU) - ISOISO10993-5
IrritationIntracutaneous Injection - ISOISO10993-10
SensitizationKligman MaximizationMurine Local Lymph AssayISO10993-10
Systemic ToxicityISO Acute Systemic Toxicity TestISO10993-11
Material MediatedPyrogenicityPyrogen Test in RabbitUSP<151>ISO10993-11
Material MediatedPyrogenicityLimulus Amebocyte LysateUSP 38, NF 33, 2015
HemocompatibilityHemolysis-Complete (Direct and Indirect)ISO10993-4
HemocompatibilityComplement ActivationISO10993-4
HemocompatibilityIn-Vivo ThrombogenicityISO10993-4

Table 5.7.2 Biocompatibility Testing for the TXM Hydrophilic Guidewire

5.8 - Sterilization Testing Summary

Table 5.8 Sterilization Testing for the TXM Hydrophilic Guidewire
Validation SterilizationProcessCompliance StandardSterility AssuranceLevel (SAL)ValidationResult
Ethylene Oxide GasISO 1113510-6Pass

5.9 – Conclusion

The TXM Hydrophilic Guidewire is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.