K Number
K170722
Device Name
Moxxly Flow Collection System
Manufacturer
Date Cleared
2017-08-28

(172 days)

Product Code
Regulation Number
884.5160
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk
Device Description
Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk. The Moxxly Flow consists of two collection units designed to be compatible with select compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use. The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands. The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.
More Information

Not Found

No
The description focuses on the mechanical design and function of a breast milk collection system, with no mention of AI or ML capabilities.

No.
The device is a breast pump collection system, used for expressing human milk, which is a supportive rather than a therapeutic function.

No

The device description clearly states its purpose is "for the purpose of expressing human milk" and "to aid collection of breast milk," which are functional, not diagnostic, uses. There is no mention of disease detection, diagnosis, or assessment in the provided text.

No

The device description clearly outlines physical components such as collection units, flanges, necks, valve assemblies, and bottles, and the performance studies include testing of these physical attributes. There is no mention of software as a component of this device.

Based on the provided information, the Moxxly Flow breast pump collection system is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is "for the purpose of expressing human milk." This is a physical process of collecting a biological sample (milk), not a diagnostic test performed on a biological sample to determine a medical condition or state.
  • Device Description: The description focuses on the mechanical function of collecting milk and its design for comfort and compatibility with breast pumps. It does not mention any analysis or testing of the milk itself.
  • Lack of Diagnostic Elements: There are no mentions of analyzing the milk, detecting biomarkers, or providing any diagnostic information based on the milk collected.
  • Anatomical Site: The anatomical site is the breast, which is where the milk is produced and collected, not where a diagnostic test would be performed on the milk.

IVD devices are used to examine specimens (like blood, urine, or tissue) taken from the human body to provide information for diagnosis, monitoring, or screening. The Moxxly Flow simply facilitates the collection of a biological fluid.

N/A

Intended Use / Indications for Use

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk.

Product codes

HGX

Device Description

Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk.

The Moxxly Flow consists of two collection units designed to be compatible with select compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use.

The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands.

The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Breast

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Lactating Mothers

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Verification and Validation was conducted to demonstrate safety and effectiveness and for comparison to the predicate device. Testing included: flange design, compatible pumps, vacuum performance, capacity and from leakage, and ability to be supported, usability, and biocompatibility. Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device. Testing data demonstrate the device meets all of its specifications including compliance with the following:

Biocompatibility:

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: For In . vitro Cytotoxicity
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Test for Irritation and Skin Sensitization

Usability:

  • IEC 62366-1:2015 Medical Devices Part 1: Application of Usability . Engineering for Medical Devices

Key Metrics

Not Found

Predicate Device(s)

K130349

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2017

Moxxly, Inc. Rory A. Carrillo Ouality/Regulatory 1777 Yosemite Ave. Suite 235 San Francisco. CA 94124

K170722 Re:

Trade/Device Name: Moxxly Flow Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 18, 2017 Received: July 24, 2017

Dear Rory A. Carrillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

1

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K170722

Device Name Moxxly Flow

Indications for Use (Describe)

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

A. Submitter

Moxxly, Inc. 1777 Yosemite Avenue, Suite 235 San Francisco, CA 94124 Phone: (650) 416-6084 Email: cara@moxxly.com

B. Date Prepared

August 28, 2017

C. Contact Person

Rory A. Carrillo Quality/Regulatory Phone: Email:

| Device Proprietary

Name:Moxxly Flow
Device Common NamePowered Breast Pump
Device Classification:
Regulatory Class:
Product Code21 CFR§884.5160
Class II
HGX

D. Device Name and Classification

E. Predicate Device

The Moxxly Flow is substantially equivalent to the DAO Health Freemie Breast Pump Collection System, K130349.

The predicate devices have not been subject to a design-related recall.

F. Intended Use

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk.

G. Device Description

Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk.

The Moxxly Flow consists of two collection units designed to be compatible with select

4

compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use.

The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands.

The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.

MFRDAO HealthMoxxly, Inc.
Model
NameFreemie Breast Pump
Collection System (K130349)Moxxly Flow (K170722)
Indications
for UseThe Freemie breast pump
collection system is intended to
be used in conjunction with an
approved powered breast pump
for the purpose of expressing
human milk.The Moxxly Flow breast pump
collection system is intended to
be used in conjunction with a
compatible powered breast
pump for the purpose of
expressing human milk.
MFRDAO HealthMoxxly, Inc.
Model NameFreemie Breast Pump
Collection SystemMoxxly Flow
510(k)
NumberK130349K170722
Classification21 CFR§884.516021 CFR§884.5160
Device TypePump, Breast, PoweredPump, Breast, Powered
Product CodeHGXHGX
User PopulationLactating MothersLactating Mothers
Target AreaBreastBreast
How SuppliedNon-sterileNon-sterile
Single UserYesYes
Cleaning MethodsSoap and warm water
Dishwasher
BoilingSoap and warm water
Dishwasher
Boiling
Flange Size(s)25mm
28mm
32mm24mm (Medium)
27mm (Large)
Overflow ProtectionYesYes
Volume Capacity16 oz10 oz
Pump CompatibilityYesYes
• Pump in Style Advance
• Medela Symphony
• Lactina
• Ameda Purely Yours
• Twin Electric Breast Pump
• Spectra Dew 300
• Spectra Dew 350• Ameda Purely Yours
• Evenflo Advanced
• Freemie Freedom
• Hygeia Enjoye
• Medela Pump In Style
Advanced
• Medela Symphony

H. Comparison of Indications for Use with Predicate Device

The intended use is similar to that of the predicate device.

5

I. Comparison of Technical Characteristics with Predicate Device

The subject device and the predicate device do not have the same technological characteristics. The flange size, collection bottle size, material, and compatible pumps are different between the subject device and the predicate device.

The different technological characteristics of the subject device do not raise different safety or

6

effectiveness questions because all devices used for collection of breast milk in conjunction with breast must demonstrate that they are biocompatible and work in conjunction with the identified compatible breast pump.

J. Non-Clinical Performance Test Summary

Verification and Validation was conducted to demonstrate safety and effectiveness and for comparison to the predicate device. Testing included: flange design, compatible pumps, vacuum performance, capacity and from leakage, and ability to be supported, usability, and biocompatibility. Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device. Testing data demonstrate the device meets all of its specifications including compliance with the following:

Biocompatibility:

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: For In . vitro Cytotoxicity
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Test for Irritation and Skin Sensitization

Usability:

  • IEC 62366-1:2015 Medical Devices Part 1: Application of Usability . Engineering for Medical Devices

K. Conclusion

The Moxxly Flow is substantially equivalent to the predicate device.