K Number
K170722

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-08-28

(172 days)

Product Code
Regulation Number
884.5160
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk

Device Description

Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk.

The Moxxly Flow consists of two collection units designed to be compatible with select compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use.

The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands.

The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.

AI/ML Overview

The provided text is a 510(k) Summary for the Moxxly Flow breast pump collection system. It outlines the device's description, intended use, and comparison to a predicate device, as well as a summary of non-clinical performance tests.

However, the document does not describe a study that proves the device meets specific acceptance criteria related to AI/algorithm performance in medical imaging or diagnosis. The "acceptance criteria" and "study" described in the document pertain to the physical and functional aspects of a breast pump system, such as biocompatibility, vacuum performance, and usability, which are typical for mechanical medical devices.

Therefore, many of the specific points requested in your prompt regarding AI/algorithm performance (e.g., sample size for test/training sets, data provenance, number/qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth) are not applicable to the content provided.

Here's how to address the requested information based on the provided document:


Acceptance Criteria and Device Performance (Based on provided document for Physical Device)

The document primarily focuses on demonstrating the substantial equivalence of the Moxxly Flow breast pump system to a predicate device. The performance tests are geared towards ensuring the device functions safely and effectively as a breast pump, rather than providing diagnostic or AI-driven insights.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria Category (Derived from "Non-Clinical Performance Test Summary")Reported Device Performance (Summary from text)
Flange DesignDevice meets design requirements for flange.
Compatible PumpsDevice is functional with identified compatible breast pumps.
Vacuum PerformanceDevice demonstrates appropriate vacuum performance for breast milk expression.
Capacity and Freedom from LeakageDevice maintains intended capacity and does not leak during use.
Ability to be SupportedDevice can be supported/held in place as designed (e.g., in a bra).
UsabilityDevice passed usability testing per IEC 62366-1:2015.
BiocompatibilityDevice passed biocompatibility testing per ISO 10993-5:2009 (Cytotoxicity) and ISO 10993-10:2010 (Irritation and Skin Sensitization).
Overall Safety and EffectivenessDevice meets its design requirements and intended use, and is as safe, as effective, and performs as well as the predicate device.

The document notes that "Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device." However, specific numerical performance metrics for each criterion (e.g., exact vacuum pressure range, specific leakage rates, quantitative usability scores) are not provided in this summary.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the 510(k) summary. The document mentions "Verification and Validation" and "Testing," but does not specify sample sizes for any of the non-clinical performance tests (e.g., number of units tested for vacuum performance, number of individuals in usability study). The data provenance (country of origin, retrospective/prospective) is also not stated, as these tests are laboratory/simulated use tests, not clinical data collection.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This is not applicable as the device is a mechanical breast pump system, not an AI/diagnostic device. Ground truth, in the context of medical imaging interpretation or diagnosis, is not a concept applied to the testing described for this device. The testing primarily involves engineering and materials science validation.

4. Adjudication Method for the Test Set

This is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used in clinical studies, particularly for interpreting medical images or clinical outcomes, where human experts might disagree. The non-clinical tests described for this breast pump would have defined pass/fail criteria or quantitative measurements without a need for expert adjudication in this sense.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

This is not applicable. MRMC studies are specific to evaluating the diagnostic performance of human readers, often comparing performance with and without AI assistance, especially in medical imaging. The Moxxly Flow is a mechanical breast pump, not a diagnostic imaging device, and its evaluation does not involve human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. There is no AI algorithm being evaluated for standalone performance within the context of this breast pump.

7. The Type of Ground Truth Used

This is not applicable in the context of AI/diagnostic device evaluation. For the physical breast pump, "ground truth" would be established by engineering specifications, validated test methods, and regulatory standards (e.g., ISO standards for biocompatibility).

8. The Sample Size for the Training Set

This is not applicable. The device is a physical product, not an AI model that requires a "training set" of data.

9. How the Ground Truth for the Training Set was Established

This is not applicable for the same reason as point 8.


Summary of what is present in the document relevant to device acceptance:

The 510(k) summary for the Moxxly Flow emphasizes substantial equivalence to a predicate device (DAO Health Freemie Breast Pump Collection System, K130349). Acceptance criteria for this type of device are primarily based on:

  • Meeting design specifications (evidenced by vacuum performance, capacity, leakage, etc.).
  • Compliance with relevant international standards (e.g., ISO 10993 for biocompatibility, IEC 62366 for usability).
  • Demonstrating that any differences in technological characteristics (e.g., flange size, material) do not raise new questions of safety or effectiveness.

The study proving acceptance involves a series of non-clinical performance tests focused on the physical and functional attributes of the breast pump system. While the document asserts that "Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device," it does not provide detailed quantitative results or specific methodologies beyond the general categories of testing and the standards referenced.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2017

Moxxly, Inc. Rory A. Carrillo Ouality/Regulatory 1777 Yosemite Ave. Suite 235 San Francisco. CA 94124

K170722 Re:

Trade/Device Name: Moxxly Flow Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 18, 2017 Received: July 24, 2017

Dear Rory A. Carrillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170722

Device Name Moxxly Flow

Indications for Use (Describe)

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

A. Submitter

Moxxly, Inc. 1777 Yosemite Avenue, Suite 235 San Francisco, CA 94124 Phone: (650) 416-6084 Email: cara@moxxly.com

B. Date Prepared

August 28, 2017

C. Contact Person

Rory A. Carrillo Quality/Regulatory Phone: Email:

Device ProprietaryName:Moxxly Flow
Device Common NamePowered Breast Pump
Device Classification:Regulatory Class:Product Code21 CFR§884.5160Class IIHGX

D. Device Name and Classification

E. Predicate Device

The Moxxly Flow is substantially equivalent to the DAO Health Freemie Breast Pump Collection System, K130349.

The predicate devices have not been subject to a design-related recall.

F. Intended Use

The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk.

G. Device Description

Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk.

The Moxxly Flow consists of two collection units designed to be compatible with select

{4}------------------------------------------------

compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use.

The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands.

The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.

MFRDAO HealthMoxxly, Inc.
ModelNameFreemie Breast PumpCollection System (K130349)Moxxly Flow (K170722)
Indicationsfor UseThe Freemie breast pumpcollection system is intended tobe used in conjunction with anapproved powered breast pumpfor the purpose of expressinghuman milk.The Moxxly Flow breast pumpcollection system is intended tobe used in conjunction with acompatible powered breastpump for the purpose ofexpressing human milk.
MFRDAO HealthMoxxly, Inc.
Model NameFreemie Breast PumpCollection SystemMoxxly Flow
510(k)NumberK130349K170722
Classification21 CFR§884.516021 CFR§884.5160
Device TypePump, Breast, PoweredPump, Breast, Powered
Product CodeHGXHGX
User PopulationLactating MothersLactating Mothers
Target AreaBreastBreast
How SuppliedNon-sterileNon-sterile
Single UserYesYes
Cleaning MethodsSoap and warm waterDishwasherBoilingSoap and warm waterDishwasherBoiling
Flange Size(s)25mm28mm32mm24mm (Medium)27mm (Large)
Overflow ProtectionYesYes
Volume Capacity16 oz10 oz
Pump CompatibilityYesYes
• Pump in Style Advance• Medela Symphony• Lactina• Ameda Purely Yours• Twin Electric Breast Pump• Spectra Dew 300• Spectra Dew 350• Ameda Purely Yours• Evenflo Advanced• Freemie Freedom• Hygeia Enjoye• Medela Pump In StyleAdvanced• Medela Symphony

H. Comparison of Indications for Use with Predicate Device

The intended use is similar to that of the predicate device.

{5}------------------------------------------------

I. Comparison of Technical Characteristics with Predicate Device

The subject device and the predicate device do not have the same technological characteristics. The flange size, collection bottle size, material, and compatible pumps are different between the subject device and the predicate device.

The different technological characteristics of the subject device do not raise different safety or

{6}------------------------------------------------

effectiveness questions because all devices used for collection of breast milk in conjunction with breast must demonstrate that they are biocompatible and work in conjunction with the identified compatible breast pump.

J. Non-Clinical Performance Test Summary

Verification and Validation was conducted to demonstrate safety and effectiveness and for comparison to the predicate device. Testing included: flange design, compatible pumps, vacuum performance, capacity and from leakage, and ability to be supported, usability, and biocompatibility. Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device. Testing data demonstrate the device meets all of its specifications including compliance with the following:

Biocompatibility:

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: For In . vitro Cytotoxicity
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Test for Irritation and Skin Sensitization

Usability:

  • IEC 62366-1:2015 Medical Devices Part 1: Application of Usability . Engineering for Medical Devices

K. Conclusion

The Moxxly Flow is substantially equivalent to the predicate device.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).