K Number
K170680
Date Cleared
2017-05-16

(71 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TEVADAPTOR is a Closed System Drug Transfer Device (CSTD) that mechanically prohibits the release of the drug in vapor, aerosol or liquid form during preparation and administration, and prevents the introduction of microbial and airborne contaminants into the drug or fluid path, allowing the system to minimize exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

Device Description

The TEVADAPTOR® Closed Drug Reconstitution and Transfer System is a system of components that allows the reconstitution of liquid or pre-dissolved powder drugs into infusion bags, flexible bottles or syringes. Single, partial or multiple vials can be used for each infusion solution container. The TEVADAPTOR® Closed Drug Reconstitution and Transfer System prevents contamination of the user or the environment by the drug through the use of elastomeric seals and an active carbon filter. Sterility of the drug in the vial is maintained because any air entering the vial during pressure equalization enters through of a hydrophobic acrylic copolymer membrane with a pore size of 0.2 micron.

The purpose of this Special 510(k) is to add an additional design of one of the components, the TEVADAPTOR® Syringe Adaptor Lock.

The Syringe Adaptor (SA) Lock is one component of the system which is intended for connection to a standard Luer lock syringe. A non-latex elastomer protector covers the liquid dispensing needle tip. A clamp mechanism reversibly connects the Syringe Adaptor Lock to ther components of the TEVADAPTOR® system. The TEVADAPTOR® Syringe Adaptor Lock includes a stainless steel needle that is shielded at all times during preparation, use and disposal. The Syringe Adaptor Lock is supplied with a cap.

Disconnection of a syringe is not possible with the Syringe Adaptor Lock, because of a design change from the syringe adaptor that introduces a two-component distal hub. One of these components attaches, and rotates within the second component, together with the syringe, if disconnection is attempted.

The components of the TEVADAPTOR® system available separately are:

  • Vial Adaptor 20 mm with 13 mm Vial Converter
  • Vial Adaptor 28 mm
  • Syringe Adaptor
  • Syringe Adaptor Lock
  • Spike Port Adaptor
  • Connecting Set
  • Luer Lock Adaptor
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (TEVADAPTOR® Syringe Adaptor Lock)
Bidirectional flowDemonstrated equivalence with the Syringe Adaptor
Needle to Syringe Adapter Luer-lock disconnection testDemonstrated equivalence with the Syringe Adaptor
Breakage of SA Lock Luer retention teethDemonstrated equivalence with the Syringe Adaptor
Disconnection of male luer and SA LockDemonstrated equivalence with the Syringe Adaptor; specifically, disconnection of a syringe is not possible with the Syringe Adaptor Lock due to a two-component distal hub design.
Luer fitting compliance with ISO 594/1 & 2 (including Gauging test, Resistance for overriding, Unscrewing torque, Ease of assembly, Liquid leakage, Air leakage, Separation force of conical fitting assembly, Stress cracking)Demonstrated equivalence with the Syringe Adaptor
Visual inspection for absence of cracksDemonstrated equivalence with the Syringe Adaptor
Sterility Assurance Level (SAL)10⁻⁶ (supplied sterile, same as remainder of TEVADAPTOR® components cleared under K141448)
BiocompatibilityAll materials within the fluid path are unchanged from the Syringe Adaptor cleared under K141448 or K071741. No additional tests were carried out.
Risk MitigationNew risks identified from design changes (Syringe Adaptor Lock) were adequately mitigated and verified by bench testing.

Study that Proves the Device Meets Acceptance Criteria:

The study conducted was a bench testing evaluation to demonstrate equivalence of the new TEVADAPTOR® Syringe Adaptor Lock with the existing TEVADAPTOR® Syringe Adaptor (predicate device).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state a specific numerical sample size for the bench tests. It broadly mentions "bench testing has been carried out."
  • Data Provenance: The study was conducted by Teva Medical Ltd., which is based in Israel (North Industrial Zone, Kiryat Shmona 10258, ISRAEL). The data is retrospective in the sense that it's a submission for clearance of a modified device, and the testing was performed to demonstrate equivalence to a previously cleared predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable to this submission. The device is a Closed System Drug Transfer Device, and the performance evaluation relies on objective physical and mechanical tests (e.g., flow rates, leakage, disconnection forces, ISO compliance) rather than expert interpretation of medical images or clinical outcomes. The "ground truth" for these tests is defined by established engineering standards and specifications.

4. Adjudication Method for the Test Set

This is not applicable as the evaluation relies on objective measurements and compliance with engineering standards, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. The device is not an AI-powered diagnostic tool, but rather a medical device for drug transfer. Therefore, no MRMC study or AI-related effectiveness assessment was performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable. The device is not an algorithm or software; it is a physical medical device.

7. The Type of Ground Truth Used

The ground truth used for this evaluation consists of:

  • Established engineering specifications and standards, particularly ISO 594/1 & 2 for Luer fittings.
  • The performance characteristics of the predicate device (TEVADAPTOR® Syringe Adaptor cleared under K141448), as the goal was to demonstrate equivalence.

8. The Sample Size for the Training Set

This is not applicable. The device is a physical medical device, not a machine learning algorithm. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2017

Teva Medical Ltd., Mitgada Plant c/o Mr. Roger Gray Donawa Lifescience Consulting Piazza Albania 10 Rome 00153 ITALY

Re: K170680

Trade/Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: ONB Dated: April 12, 2017 Received: April 17, 2017

Dear Mr. Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely, Tina Kiang

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K170680

Device Name

TEVADAPTOR Closed Drug Reconstitution and Transfer System

Indications for Use (Describe)

TEVADAPTOR is a Closed System Drug Transfer Device (CSTD) that mechanically prohibits the release of the drug in vapor, aerosol or liquid form during preparation and administration, and prevents the introduction of microbial and airborne contaminants into the drug or fluid path, allowing the system to minimize exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is divided into two sections. The left section is a light gray color and contains the word "TEVA" in a stylized font. The right section is green and contains the word "Medical" in a cursive font, with the words "Teva Medical Ltd." below it in a smaller, sans-serif font. The image appears to be a logo for a medical company.

K170680

510(k) Summary in accordance with 21 CFR 807.92(c)

Device Name:TEVADAPTOR® Closed Drug Reconstitution and Transfer System
Type of 510(k) submission:Special
Date of Submission:May 12, 2017
Manufacturer:Teva Medical Ltd., MIGADA PlantNorth Industrial ZoneKiryat Shmona 10258ISRAEL
FDA Registration Number:9611423
Owner/Operator Number:9001925
510(k) Owner & Submitter:Teva Medical Ltd., MIGADA PlantNorth Industrial ZoneKiryat Shmona 10258ISRAEL
Phone:+972-4-6908826
Fax:+972-9-892-1659
510(k) Application Correspondent:Mr Roger GrayVP Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 RomeItaly
Phone: +39 06 578 2665Fax: +30 06 574 3786Email: rgray@donawa.com
FDA Product Code:ONB
FDA Regulation Number:880.5440
FDA Regulation Name:Set, Administration, Intravascular
Classification Panel:General Hospital
Common Name:Closed Drug Reconstitution and Transfer System
FDA Classification:Class II
Indications for Use:TEVADAPTOR® is a Closed System Drug Transfer Device (CSTD) that mechanicallyprohibits the release of the drug in vapor, aerosol or liquid form, duringpreparation and administration, and prevents the introduction of microbial andairborne contaminants into the drug or fluid path, allowing the system tominimize exposure of individuals, healthcare personnel, and the environment tohazardous drugs.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains two logos side by side. The first logo on the left has the word "TEVA" in a stylized font with a teal color. The second logo on the right has the word "Medical" in a white serif font on a green background, with the text "Teva Medical Ltd." below it in a smaller font.

Predicate Device:

The predicate device selected for comparison with the revised TEVADAPTOR® Closed Drug reconstitution and Transfer System is:

Predicate Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System

510(k) Sponsor:Teva Medical
510(k) Number:K141448
Clearance Date:23 January 2015
FDA Product Code:ONB
Regulation Name:Intravascular administration set
Regulation Number:880.5440

Reference Device:

Reference Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System
510(k) Sponsor:Teva Medical
510(k) Number:K071741
Clearance Date:24 September 2007
FDA Product Code:LHI
Regulation Name:Intravascular administration set
Regulation Number:880.5440

Device Description:

The TEVADAPTOR® Closed Drug Reconstitution and Transfer System is a system of components that allows the reconstitution of liquid or pre-dissolved powder drugs into infusion bags, flexible bottles or syringes. Single, partial or multiple vials can be used for each infusion solution container. The TEVADAPTOR® Closed Drug Reconstitution and Transfer System prevents contamination of the user or the environment by the drug through the use of elastomeric seals and an active carbon filter. Sterility of the drug in the vial is maintained because any air entering the vial during pressure equalization enters through of a hydrophobic acrylic copolymer membrane with a pore size of 0.2 micron.

The purpose of this Special 510(k) is to add an additional design of one of the components, the TEVADAPTOR® Syringe Adaptor Lock.

The Syringe Adaptor (SA) Lock is one component of the system which is intended for connection to a standard Luer lock syringe. A non-latex elastomer protector covers the liquid dispensing needle tip. A clamp mechanism reversibly connects the Syringe Adaptor Lock to ther components of the TEVADAPTOR® system. The TEVADAPTOR® Syringe Adaptor Lock includes a stainless steel needle that is shielded at all times during preparation, use and disposal. The Syringe Adaptor Lock is supplied with a сар.

Disconnection of a syringe is not possible with the Syringe Adaptor Lock, because of a design change from the syringe adaptor that introduces a two-component distal hub. One of these components attaches, and rotates within the second component, together with the syringe, if disconnection is attempted.

The components of the TEVADAPTOR® system available separately are:

  • Vial Adaptor 20 mm with 13 mm Vial Converter
  • Vial Adaptor 28 mm ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Teva Medical logo. The logo consists of the word "TEVA" in a stylized font on the left side of the image. To the right of the word "TEVA" is the word "Medical" in a serif font, followed by the words "Teva Medical Ltd." in a smaller font.

  • Syringe Adaptor
  • Syringe Adaptor Lock
  • Spike Port Adaptor
  • Connecting Set
  • Luer Lock Adaptor

The environment of use is unchanged from that of the predicate TEVADAPTOR® system cleared under K141448.

Device Performance Testing:

Bench testing has been carried out on the Syringe Adaptor Lock to demonstrate equivalence with the Syringe Adaptor, including:

  • Bidirectional flow
  • Needle to Syringe Adapter Luer-lock disconnection test
  • Breakage of SA Lock Luer retention teeth
  • Disconnection of male luer and SA Lock
  • Luer fitting compliance with ISO 594/1 & 2, including:
    • o Gauging test
    • O Resistance for overriding
    • Unscrewing torque O
    • O Ease of assembly
    • O Liquid leakage
    • O Air leakage
    • O Separation force of conical fitting assembly
    • O Stress cracking
  • Visual inspection for absence of cracks ●

The packaging materials and process for the Syringe Adaptor Lock are unchanged from those utilized for the Syringe Adaptor and other TEVADAPTOR® components cleared under K141448.

The subject devices are supplied sterile, with an SAL of 10 °. The devices are sterilized by Ethylene Oxide, with the same parameters as the remainder of the TEVADAPTOR® components, as described in K141448.

No additional biocompatibility tests have been carried out on the components of the Syringe Adaptor Lock, because all materials within the fluid path of the item are unchanged from the Syringe Adaptor cleared under K141448 or its previous design, cleared under K071741, included within this submission as a reference device. The material of the hub component of the syringe adaptor lock is identical to the material of the hub component of the syringe adaptor cleared under K071741.

A risk assessment of the changes in design introduced by the Syringe Adaptor Lock concluded that any new risks identified were adequately mitigated and verified by means of bench testing.

Comparison with Predicate Device:

The major difference between the Syringe Adaptor and Syringe Adaptor Lock is in the 'Luer Hub' component, which has been replaced by two components, the 'SA Luer-lock' and the 'SA Hub-lock'. The SA Hub-lock and End Cap for the Syringe Adaptor Lock have new colorants which are not used in the Syringe Adaptor components, for easy user-recognition purposes, however neither of these new components of the Syringe Adaptor Lock are in the fluid path, being external only components.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains two logos, one for TEVA and one for Teva Medical Ltd. The TEVA logo is on the left side of the image and is in a light blue color. The Teva Medical Ltd. logo is on the right side of the image and is in a green color. The text "Medical" is in a larger font than the text "Teva Medical Ltd."

The separation of the 'Luer Hub' into two components has allowed the design to permanently attach a Luer syringe to the Syringe Adaptor Lock, whereas a Luer syringe can be disconnected from the predicate Syringe Adaptor. When attaching a Syringe to the Syringe Adaptor Lock, the SA Luer-lock will rotate with the Syringe until two teeth on the SA Luer-lock component engage with two of the four cam upstands on the inner circumference of the SA Hub-lock component. Once the SA Luer-lock teeth have engaged with the SA Hub-lock, no further rotation of the SA Luer-lock is possible.

If disconnection of the Syringe Adaptor Lock is attempted, the SA Luer-lock will rotate with the Syringe. When a quarter rotation is reached, by means of a cam action, the SA Luer-lock teeth will slip over the upstands on the SA Hub-lock inner surface, and will thus continue to rotate with the Syringe, preventing disconnection of the Syringe from the Syringe Adaptor Lock.

The only other difference is that 'Main Body' component of the Syringe Adaptor has been shortened by 5.2 mm to accommodate the above new components within the Syringe Adaptor Lock.

All other components of the TEVADAPTOR® system remain unchanged from those cleared under K141448.

Substantial Equivalence Conclusion:

Based on the performance testing conducted on the subject device, the Syringe Adaptor Lock, does not raise new types of safety and effectiveness questions. It is concluded that the TEVADAPTOR® device that has been modified by the introduction of the Syringe Adaptor Lock is substantially equivalent to the identified predicate device cleared under K141448, which is already in interstate commerce within the USA.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.