(134 days)
The sensiFirm device is intended for the following medical conditions using non thermal RF combined with massage:
Temporary reduction in the appearance of cellulite.
The sensiFirm is an OTC, home use device, designed to deliver non-thermal RF energy and mechanical skin massaging. RF energy does not cause thermal damage to the treated skin and the subdermal adipose layer. The level of the non-thermal RF power, to achieve maximum comfort is user selectable. The ergonomic hand held device allows efficient treatment of the indicated body areas. It is a noninvasive, non-ablative device and it is supplied as non-sterile.
The sensiFirm device consists of power supply, RF power generator, Programmable Logic Controller (PLC, microcontroller) embedded in PCBA and user interface. The device also includes a vibration motor, four Bi-polar (2 pairs) electrodes and two redundant temperature sensors. The PLC (on the PCBA) is specially configured software combined with hardware circuits that provides the operational and safety function of the system.
The sensiFirm device is operated while continuously moving it over the treatment area. This ensures a uniform distribution of the non-thermal RF energy together with mechanical massage over the entire treatment area.
The document describes the acceptance criteria and the study that proves the device, sensiFirm, meets these criteria for the temporary reduction in the appearance of cellulite.
Here’s the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Safety - Skin Temperature | Max skin temperature of 41°C / 105.8°F (or below pain threshold of 44°C and burn threshold of 48°C) | RF delivery automatically stops when skin temperature exceeds 41°C, resuming at 40.5°C. No adverse events occurred during and after the clinical study. |
| Safety - Electrical/Mechanical | Compliance with IEC 60601 family of standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-2). | Bench testing conducted and device complies with all listed standards. |
| Safety - Software | Software verification and validation per FDA guidance. | Software verification and validation testing conducted and documentation provided, classified as "moderate" level of concern. |
| Safety - Human Factors (Label Comprehension) | Typical OTC users can understand device labeling and correctly choose the device for its intended use and self-selection (contraindications). Primary endpoint: 100% correct decision regarding intended use, 98% correct for treatment areas, 96% correct for contraindications. | 100% of participants made correct decision regarding intended use, 98% correctly identified treatment areas, and 96% reported correct answers for contraindications (K=0.91). Success criteria for primary endpoint met. |
| Safety - Human Factors (Usability/User Interface) | Intended OTC user can correctly use the device based on labeling materials for aesthetic purpose. Primary endpoint: 100% success rate per each task scenario. | All 21 tasks across four use scenarios were completed by 100% of participants, with a 100% success rate per each task scenario. Success criteria for primary endpoint met. |
| Effectiveness - Cellulite Appearance (Evaluator Assessment) | At least 2 out of 3 independent evaluators agree on at least one degree of improvement (score reduction) in the Modified Neurenberg-Muller Cellulite Grading Scale compared to baseline. | At T8 (end of active treatment): 93% of participants showed improvement. At FU1 (1 month post-treatment): 96% showed improvement. At FU2 (3 months post-treatment): 100% showed improvement. Average reduction: -1.28 (T8), -1.59 (FU1), -1.78 (FU2). Untreated side remained unchanged. |
| Effectiveness - User Satisfaction | High user satisfaction with the device and improvement in skin appearance, cellulite appearance, and skin texture. | 95% of participants were generally satisfied. 94% reported improvement in skin's appearance, 89% in cellulite appearance and skin tightening, 90% in skin texture. Device found easy to use, comfortable to hold and reach treatment areas (82-93% agreement). 96% found the treatment regimen reasonable. All found it painless/discomfort-less. |
| Performance - Power Accuracy | Measured total power within predefined error margin on 150 Ω load and other impedance values. | The device met the acceptance criteria. |
| Performance - Parameter Validation | Varied parameters (radio frequency, pulse cycle, waveform, pulse duration) within acceptance criteria. | All results were within the acceptance criteria. |
| Biocompatibility | All parts in contact with user comply with ISO-10993-1, ISO 10993-5:2009 (cytotoxicity), and ISO 10993-10:2010 (sensitization and irritation). | Body contact materials are biocompatible per listed ISO standards. |
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Study (Effectiveness):
- Sample Size: 33 women (age 29-63, average 46 ± 9 years).
- Data Provenance: Prospective, interventional, single-arm clinical study conducted in a home-use environment. Country of origin not explicitly stated but implied to be where the company is located (Israel) or a clinical site arranged by them.
- Human Factors Validation Testing (Usability/Label Comprehension):
- Labeling Comprehension/Self-Selection Study: 47 participants (5 men, 42 women, age 24-71, average 42 ± 12 years).
- Usability/User Interface Study: Not explicitly stated for this particular subsection, but it's part of the broader Human Factors testing which implies it leveraged the participants from the Human Factors study. (The document states "None of the participants required assistance from the moderator while performing the tasks," suggesting it refers to the same participant pool).
- Data Provenance: Not explicitly stated for Human Factors studies, but typically prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Clinical Study (Effectiveness):
- Number of Experts: 3 evaluators.
- Qualifications: 3 board-certified dermatologists and plastic surgeons.
4. Adjudication Method for the Test Set
- Clinical Study (Effectiveness):
- Adjudication Method: "A participant was considered a success if at least 2 out of the 3 independent evaluators agreed that the participant had shown at least one degree of improvement in the cellulite appearance." This is a form of 2-out-of-3 consensus or 2/3 agreement.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study was a single-arm clinical trial assessing the device's efficacy, with a comparison to an untreated control side of the body. There was no AI component to assist human readers described in this document.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. The sensiFirm device is a physical device that delivers RF energy and massage; it is not an AI algorithm. Its performance is evaluated through its direct physical effects and user interaction, not as a standalone AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Clinical Study (Effectiveness): Expert consensus of 3 board-certified dermatologists and plastic surgeons using the Modified Neurenberg-Muller Cellulite Grading Scale, combined with participant satisfaction questionnaires (outcomes data). The design included "subject as her own control" and a concurrent untreated control side for objective comparison.
- Human Factors Validation Testing: Expert observation and qualitative assessment by a moderator for correct task completion and labeling comprehension, along with participant self-reporting/satisfaction.
8. The sample size for the training set
- This information is not provided in the document. The document describes a clinical study and human factors testing for device validation, not the development or training of an AI model.
9. How the ground truth for the training set was established
- Not applicable, as no AI model training is described in the provided document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2017
El Global Trade Ltd Yael Liebes-Peer RA/QA Manager Tzoran 8th St, P.O. Box 8242 Netanya, 4250608, Israel
Re: K170637
Trade/Device Name: sensiFirm Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: May 21, 2017 Received: May 31, 2017
Dear Dr. Yael Liebes-Peer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Image /page/1/Picture/9 description: The image shows the text "Jennifer R. Stevenson -S3". The text is written in a clear, sans-serif font and is easy to read. The background of the image is a light blue color, and the text is black.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name sensiFirm
Indications for Use (Describe)
The sensiFirm device is intended for the following medical conditions using non thermal RF combined with massage:
Temporary reduction in the appearance of cellulite.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
RD-12062 A0
DATE PREPARED: MAY 215T, 2016
510(K) OWNER NAME 1.
EL Global Trade Ltd. Tzoran 8th st, P.O.Box 8242, Netanya 425068, Israel. Phone: +972-9-7889069, Fax: +972-9-7734831.
Contact person name: Dr. Yael Liebes-Peer, RA/ QA Manager (E mail: Yael@sensica.com) and Martin Gurovich, CEO (E mail: Martin@elglobalt.com). Phone: +972-9-7889069, Fax: +972-9-7734831.
2. DEVICE NAME
Common/Usual Name: OTC device for improvement in the appearance of cellulite
Proprietary/Trade name: sensiFirm
Classification: EL Global Trade Ltd.'s sensiFirm device has been classified as Class II device under the following classification names:
| Device Category | ProductCode | RegulationNumber | Panel |
|---|---|---|---|
| Massager, Vacuum, RadioFrequency Induced Heat | PBX | 878.4400 | General and PlasticSurgery |
| Electrosurgical, Cutting &Coagulation & Accessories | GEI | 878.4400 | General and PlasticSurgery |
3. PREDICATE DEVICES
EL Global Trade Ltd.'s sensiFirm device is substantially equivalent to the following Predicate Devices:
- 3.1 InMode System MiniFX Handpiece by InMode MD Ltd. device, cleared under Premarket Notification K160329, dated August 19th, 2016.
- 3.2 V10 System, V-Form Handpiece by Viora Ltd. device,
cleared under Premarket Notification K150035, dated May 1st, 2015.
| EL Global Trade Ltd. 510(k) | July 13, 2017 |
|---|---|
| sensiFirm – K170637 | Page 1 of 15 |
{4}------------------------------------------------
| EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, IsraelTel: +972-9-7889069 | Fax: +972-9-7734831 |
|---|---|---|
| conciFirm K170627/S001 | BD 12062 A0 |
RD-12062 A0
DEVICE DESCRIPTION 4.
The sensiFirm is an OTC, home use device, designed to deliver non-thermal RF energy and mechanical skin massaging. RF energy does not cause thermal damage to the treated skin and the subdermal adipose layer. The level of the non-thermal RF power, to achieve maximum comfort is user selectable. The ergonomic hand held device allows efficient treatment of the indicated body areas. It is a noninvasive, non-ablative device and it is supplied as non-sterile.
The sensiFirm device consists of power supply, RF power generator, Programmable Logic Controller (PLC, microcontroller) embedded in PCBA and user interface. The device also includes a vibration motor, four Bi-polar (2 pairs) electrodes and two redundant temperature sensors. The PLC (on the PCBA) is specially configured software combined with hardware circuits that provides the operational and safety function of the system.
The sensiFirm device is operated while continuously moving it over the treatment area. This ensures a uniform distribution of the non-thermal RF energy together with mechanical massage over the entire treatment area.
Device specifications:
- Maximal power output: 10±1 Watt. .
- . Skin level power: 6±1, 8±1 and 10±1 Watt (energy levels 1-3).
- Frequency: 1±0.05 MHz. .
- Maximal temperature allowed: 41℃/ 105.8ºF. ●
5. INTENDED USE/ INDICATIONS FOR USE
The sensiFirm device is intended for the treatment of the following medical conditions using non thermal RF combined with massage:
Temporary reduction in the appearance of cellulite.
{5}------------------------------------------------
RD-12062 A0
6. NON-CLINICAL (BENCH) PERFORMANCE DATA
The following performance data (bench tests) were provided in order to support the performance, safety and efficacy of the sensiFirm device as well as the substantial equivalence determination.
Safety Bench Tests and Verification & Validation (V&V) Summary
Bench testing was conducted according to IEC 60601 family of standards to demonstrate that the sensiFirm device performs as expected under anticipated conditions of use. This testing included safety performance verification (electrical and mechanical) according to IEC 60601-1, essential requirements according to IEC 60601-1-2 and particular requirements for home use medical device per IEC60601-1-11.
In addition, the following bench testing was conducted (among others) to demonstrate the device performance characteristics as part of the V&V (Verification and Validation):
- . Over-heating safety. The two redundant temperature sensors embedded in the device constantly measure the skin temperature and deactivate RF delivery when temperature exceeds 41ºC/ 105.8ºF. This temperature was determined to be acceptable, based on published information from the National Institute for for for Standard Technology (http://www.nist.gov/fire/fire_behavior.cfm), which states that human skin begins to feel pain at 44°C and may start to develop skin burns at 48°C.
- Power accuracy. The device was validated for the different power levels . on a 150 Ω load, which is appropriate as the reference of the average load of the user, as well as on other impedance values and in overall relevant environmental conditions. The measured total power was within the
{6}------------------------------------------------
RD-12062 A0
predefined error margin, indicating that the device met the acceptance criteria.
- Parameter validation. The device was tested for its varied parameters, . including radio frequency, pulse cycle, wave form and pulse duration. All results were within the acceptance criteria.
Electrical safety and electromagnetic compatibility (EMC)
The sensiFirm device complies with the following recognized standards for Electrical Safety and EMC testing:
- IEC 60601-1:2012/EN 60601-1:2013, General safety standard: safety . requirements for medical electrical systems.
- IEC 60601-1-2:2014, Medical electrical equipment Part 1-2 General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
- . IEC 60601-2-2: 2009. Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
- IEC 60601-1-11:2015, Medical electrical equipment Part 1-11 -. Requirements for the medical electrical equipment and medical electrical systems used in the home healthcare environment.
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since inadvertent software errors could result in skin burns to the user.
{7}------------------------------------------------
RD-12062 A0
Human Factors Validation Testing
A human Factors Validation Testing was done to evaluate the HFE/ UE (Human Engineering/ Usability Engineering) concerns, including Label Factors Comprehension/ Self-Selection and Usability/ User Interface studies (one arm, simulated use and summative evaluation). In both studies, participants had received the labeling materials of the device, and given enough time to read it according to their own wish.
Labeling Comprehension/ Self Selection Study goal was to demonstrate that the typical OTC users can understand the device labeling (mainly the outer package) and correctly choose the sensiFirm device for themselves for its intended use. The study included three main questions regarding the intended use of the device, treatment areas and self-selection (the latter was acording to the contraindications), with possible answers: "Yes"or "No", followed by explenation and further validation by the moderator. A total of 47 participants, 5 men and 42 women, from age 24 to 71 years old (average 42 ± 12 years old) were enrolled to this study. Participants had different educational levels, as obtained by REALM testing with 26% below 9th grade level. All participants had met the inclusion/ exclusion criteria and signed ICF. 100% of the participants had made the correct descion regarding the intended use of the device, 98% had correctly identified the treatment areas and 96% had reported correct answers, as further validated by the moderator (K=0.91). Therfore, the success criteria of the primary enpoint was met.
Usability/ User Interface Study goal was to demonstrate the intended OTC user can correctly use the device, based on the labeling material for the intended aesthetic puspose. The study included four use scenarios including 21 relevant tasks: "getting to know the device and device set up" (7 tasks). "self-preparation" (4 tasks), "device usage" (7 tasks) and "after usage" (3 tasks). None of the participants required assistance from the moderator while performing the tasks.
{8}------------------------------------------------
RD-12062 A0
Altogether, all tasks were completed by 100% of the participants, and the success rate (pass criteria) was 100% per each task scenario, demonstrating rhe sucess criteria of the primary enpoint was met.
Furthermore, 62% of the participants (16 out of 26) had additionally commented that the device instructions are clear and that the use is easy and comfortable.
Participants were also asked regarding their user experience and satisfaction after this study, and high satisfaction was obtained, with an average result of 1.95±0.51, using a 1-5 discrete scale (when 1 represent best and 5 worst).
The human Factors Validation Testing , including both Labeling Comprehension/ Self-Selection Study and Usability/ User Interface Study, demonstrates that the potential OTC users can understand the package labeling, correctly choose the device and use it for the indicated aesthetic use, based solely on reading the lableling materials.
7. PERFORMANCE TESTING - ANIMAL
No animal testing was performed with the subject device
8. CLEANING, STERILIZATION, SHELF LIFE AND BIOCOMPATIBILITY
The sensiFirm is a non-sterile, reusable device, intended for a single user. The device cleaning instructions are based on the cleaning instructions of the predicate devices due to the fact that both devices are made from the same materials are used similarly.
The shelf-life expectancy of the device is 5 years, similarly to the predicate device.
The biocompatibility evaluation for the sensiFirm device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International
{9}------------------------------------------------
RD-12062 A0
Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
EL Global has categorized its sensiFirm device as: "Surface device, Skin Contact for limited contact duration", with contact duration of less than 24h ("A", up to 24 hours). Therefore, the battery of testing included the following tests: cytotoxicity, sensitization and irritation tests. The body contact materials are biocompatible per:
- . ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10:2010, Biological evaluation of medical devices Part 10: . Tests for irritation and skin sensitization.
9. CLINICAL PERFORMANCE DATA
The sensiFirm safety and efficacy for non-invasive temporary reduction in cellulite appearance, was examined in an open label, prospective, interventional single arm clinical study, in a home use environment.
33 women, age ranging between 29 to 63 years (average age 46 ± 9 years), fulfilling the inclusion criteria, were enrolled to the study by the Principle Investigator (PI). Participants treated one to three treatment areas, according to participants own preference and the PI's approval: 27 participants chose to treat the cellulite (buttocks, including one participant treated the inner thigh area), and 8 and 4 chose to treat the abdomen and arms regions, respectively resulting in total of 39 areas that had completed 8 sensiFirm treatments according to the instructions for use and appeared to the 3 months after treatments last follow-up (FU2).
{10}------------------------------------------------
RD-12062 A0
The cellulite appearance was assessed using the Modified Neurenberg-Muller Cellulite Grading Scale and improvement was defined as decrease of at least one score in the Scale. This assessment was conducted by 3 evaluators (3 board certified dermatologists and plastic surgeons) at the end of the active treatment regimen (T8), as well as one and three months after the last treatment (follow-up regimen FU1 and FU2, respectively). A participant was considered a success if at least 2 out of the 3 independent evaluators agreed that the participant had shown at least one degree of improvement in the cellulite appearance (one score reduction in the Modified Neurenberg-Muller Cellulite Grading Scale) when comparing the score obtained at each time point (T8, FU1 and FU2) to the baseline score ("subject as her own control"). In addition, a concurrent control was used, and only one side of the body (e.g. left buttocks) was treated to allow comparison to the other untreated side (e.g. right buttocks).
No adverse events occurred during and after the study, meaning that the success criteria for primary safety endpoint have been met.
As for effectiveness, after the completion of the active treatment regimen, 93%, 96% and 100% of the study participants (according to agreement between at least 2 out of the 3 evaluators) have shown an improvement in overall cellulite appearance of the treated side at T8, FU1 and FU2, respectively. This improvement was manifested in an average reduction of -1.28. -1.59 and -1.78 in the Modified Neurenberg-Muller Cellulite Grading Scale, at T8, FU1 and FU2, respectively. In contrast, the untreated (control) side remained unchanged during the course of the study.
Effectiveness was also assessed according to the participants' satisfaction as recorded by the satisfaction questionnaire which was filled by all the participants at each follow-up visit. 95% of the participants (average value for after the last active treatment and follow-ups) were generally satisfied with the device. 94%
{11}------------------------------------------------
| EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, IsraelTel: +972-9-7889069 | Fax: +972-9-7734831 |
|---|---|---|
| --------------------- | ----------------------------------------------------------------------------- | --------------------- |
RD-12062 A0
had reported on improvement in their skin's appearance, 89% had reported on improvement in cellulite appearance and skin tightening and 90% had reported on skin texture improvement. In addition, the participants found the device easy to use, with 82% stating using sensiFirm was clear and simple, 93% and 89% stating that reaching the treatment areas and holding the device was comfortable, respectively, and 96% replied that a regime of total of 8 treatments was reasonable (all percentages refer to T8 visit). All participants found the device to be painless or discomfort less. At the last follow-up (FU2), participants were also asked whether they would like to continue using the device for the other untreated side and whether they would like to treat additional anatomical areas. All participants answered YES to these two questions.
To summarize, these results demonstrate that the device is safe and effective for the intended use under anticipated conditions of use (home environment).
10. SUBSTANTIAL EQUIVALENCE
The indications for use and technological characteristics of the sensiFirm device are substantially equivalent to the indications for use and technological characteristics of the predicate device. A detailed comparison between the sensiFirm by EL Global LTD. proposed device and the predicate device is presented at Table 1 below, including the similarities and differences and the related conclusion.
| Feature | EL GlobalsensiFirmNew Device | InMode MDMiniFX/ InMode[K160329]Predicate Device #1 | VioraV-FORM/ V10[K150035]Predicate Device #2 | Characteristicscomparison(similarity) |
|---|---|---|---|---|
| Regulatory classification | ||||
| DeviceClass | Class II | Class II | Class II | Same |
| Classification Panel | General and PlasticSurgery | General and PlasticSurgery | General and PlasticSurgery | Same |
| Feature | EL GlobalsensiFirmNew Device | InMode MDMiniFX/ InMode[K160329]Predicate Device #1 | VioraV-FORM/ V10[K150035]Predicate Device #2 | Characteristicscomparison(similarity) |
| Productcode | PBX, GEI | PBX, GEI | PBX, ISA | Same asPredicateDevice #1 |
| Regulationnumber | • 21 CFR 878.4400• 21 CFR 878.4400 | • 21 CFR 878.4400• 21 CFR 878.4400 | • 21 CFR 878.4400• 21 CFR 890.5660 | Same asPredicateDevice #1 |
| DeviceCategory | • Massager, Vacuum,Radio FrequencyInduced heat• Electrosurgical,Cutting & Coagulation& Accessories | • Massager, Vacuum,Radio FrequencyInduced heat• Electrosurgical,Cutting & Coagulation& Accessories | • Massager, Vacuum,Radio FrequencyInduced heat• Massager, Therapeutic,Electric | Same asPredicateDevice #1 |
| Regulationdescription | • Electrosurgical cuttingand coagulation deviceand accessories• Electrosurgical cuttingand coagulation deviceand accessories | • Electrosurgical cuttingand coagulation deviceand accessories• Electrosurgical cuttingand coagulation deviceand accessories | • Electrosurgical cuttingand coagulation deviceand accessories• Therapeutic massager | Same asPredicateDevice #1 |
| Intendeduse | The sensiFirm device isintended for deliveringnon thermal RFcombined with massagefor temporary reductionin the appearance ofcellulite | The InMode System(with MiniFXHandpiece) is intendedfor delivering nonthermal RF combinedwith massage fortemporary reduction inthe appearance ofcellulite | The V-Form Handpiece(with BC and FCapplicators) is indicatedfor delivering nonthermal RF combinedwith massage fortemporary reduction inthe appearance ofcellulite | Same |
| Device Characteristics | ||||
| Mode ofoperation | Mechanical (vibration)massage combinedwith RF energyemitted from Bi-polarelectrodes via a deliveryHandpiece | Mechanical vacuummassage combinedwith RF energyemitted from Bi-polarelectrodes via a deliveryHandpiece | Mechanical vacuummassage combinedwith RF energyemitted from Bipolarelectrodes via a deliveryHandpiece | Similar.Note No. 1. |
| EnergySource | RF (Bi-polar), 1 MHz | RF (Bi-polar), 1 MHz | RF (Bi-polar), 0.8, 1.7,and/ 2.45 MHz | Same asPredicateDevice #1.Note No. 2. |
| Number ofelectrodes | 4 electrodes (2 pairs) | 2 electrodes (1 pair) | 6 or 4 electrodes (3 or 2pairs, for BC or FCapplicators, respectively) | Note No. 2. |
| Massage | Vibrational massage(1000 rpm, 1.5 [g])assisted by manualmanipulation | Mechanical vacuummassage, with pulsedvacuum ( 300 - 500mbar) | Mechanical vacuummassage, with pulsedvacuum (up to 500 mbar) | Differentmechanism,since vacuumtechnology |
| Feature | EL GlobalsensiFirmNew Device | InMode MDMiniFX/InMode[K160329]Predicate Device #1 | VioraV-FORM/V10[K150035]Predicate Device #2 | Characteristicscomparison(similarity) |
| Treatmentareas | Cellulite areas (buttocks,thigh, upper legs) | Cellulite areas (buttocks,thigh, upper legs) | Cellulite areas (buttocks,thigh, upper legs) | Same |
| Treatmentregimen | 20 minutes per each area,1 time per week, for 8weeks | 1 time per week, for 6weeks.* Information accordingto reference #17, seeAppendix F, VOL 003. | 20minutes per each area,1 time per week for 8weeks.* Information accordingto reference #5, seeAppendix F, VOL 003. | Similar.Treatmentregimen wasvalidated viaclinical Trial.Note No.3. |
| Intendedpopulation | Adult people who desireto improve their bodyappearance | Adult people who desireto improve their bodyappearance | Adult people who desireto improve their bodyappearance | Same |
| UseEnvironment | Home Use, self-operationby an untrained user. | Aesthetic Clinic | Aesthetic Clinic | Validated in aHuman Factorstests and in aClinical Trial.Same asauxiliarydevice.Note No. 4. |
| Device Features | ||||
| Hand pieceDimensions | 146 X 74 X 63 mm3/5.7 X 2.9 X2.5 inch3 | Unknown | 230 X 230 X 95 mm3/9.1 X 9.1 X 3.7 inch3 or230 X 230 X 65 mm3/9.1 X 9.1 X 2.6 inch3,BC or FC applicators,respectively | Similar.Note No. 4. |
| Weight | 192 gr/ 0.42 lb | Unknown | 930 gr/ 2.05 lb or 750 gr/1.65 lb, BC or FCapplicators, respectively) | Similar.Note No. 4. |
| Electrodestreatmentarea [mm²] | ~2000 (calculatedaccording to: 2×33×30),where• 2- numbers of electrodespairs• 33 - average distancebetween electrodes is 33 mm• electrodes length 30 mm | Unknown.Distance betweenelectrodes is 25 mm. | 6200 or 450 (BC or FCapplicators, respectively) | Similar.Note No. 2. |
| Device maincomponents | Power adaptor, hand-heldhandpiece | Console, MiniFXhandpiece | Console, V-FORMhandpiece | Similar.Note No. 4. |
| Userinterface | • 4 LEDs:• 3 green LEDs forvisualizing the RFenergy level | • Touch screen (mainconsole)• Different audiblebeeping | • Touch screen (mainconsole)• Audible soundindications at various | Different.Note No. 4. |
| InMode MD | Viora | |||
| EL GlobalsensiFirm | MiniFX/ InMode[K160329] | V-FORM/ V10[K150035] | Characteristicscomparison(similarity) | |
| Feature | Predicate Device #1 | Predicate Device #2 | ||
| New Device | ||||
| • 1 green/ orange/ redLED reflecting thedevice status• Button for energy levelselection• The massage achievedby the vibration motormakes audible soundand the sensation ofmassage itself make theuser aware of treatmentbeing delivered. | sounds when thesuction is adequatelycoupled and the RF isbeing administered andaccording to differenttemperature measured | points of the treatments | ||
| Energylevels(maximaland userselectable) | Up to 10 Watts.3 energy levels (softwareand hardware limited): 6,8 or 10 Watts, dependingon the user selection. | 10-25 Watts | Up to 25 or 50 Watts (forFC or BC handpiece,respectively).4 energy levels | Similar.Note No. 2. |
| Safetymechanism | • 2 redundant thermalsensors for skintemperaturemeasurements.Automatically stopdelivering RF whentemperature of 41°C/105.8°F is detected,until temperaturelowers to 40.5°C/104.9°F (hysteresis).• Device automatic testfor proper skin contactprior to release a pulse(using a short and lowenergy pulse)• The vibrationalmassage beingdelivered reflects to theuser the device is active• A steady LED whenthe device is active | • Integrated IRthermometer, withmaximal temperature(skin surface 41°C/105.8°F ) | • Integrated IRthermometer• Default treatmentparametersautomaticallysuggested• A trigger requirespressing prior toemitting energy• Device automatic testfor proper couplingprior to release a pulse(using a short sub-pulse)• Proper connection ofthe handpiece,otherwise alert | Similar toauxiliarydevice.Usefulness ofall safetymechanismswas validated inclinical trial andusability study.Note No. 4. |
| Electricalrequirements | 100-240 V50-60 Hz | 100-240 V50-60 Hz | 100-240 V50-60 Hz | Same |
| Safety standards | ||||
| Safety/EMC | • IEC 60601-1:2012/EN60601-1:2013• IEC 60601-1-2: 2014 | • IEC 60601-1:1988(Am. 1 1991 + Am. 21995)• IEC 60601-1-2: 2001 | • IEC 60601-1• EC 60601-1-2 | Same, withadditionalsafety standard(60601-1-11) |
| Feature | EL GlobalsensiFirmNew Device | InMode MDMiniFX/ InMode[K160329]Predicate Device #1 | VioraV-FORM/ V10[K150035]Predicate Device #2 | Characteristicscomparison(similarity) |
| • IEC 60601-2-2: 2009• IEC 60601-1-11:2015 | • IEC 60601-2-2: 2006 | • IEC 60601-2-2 | for home useenvironment | |
| Biocompatibility | All parts that are incontact with user complywith the requirements ofISO-10993-1. | All parts that are incontact with user complywith the requirements ofISO-10993-1. | All parts that are incontact with user complywith the requirements ofISO-10993-1. | Same |
| Software | The software wasverified and validatedaccording to the FDAguidance. | The software wasverified and validatedaccording to the FDAguidance. | The software wasverified and validatedaccording to the FDAguidance. | Same |
Table 1. Substantial Equivalence of EL Global's sensiFirm with Predicate Device.
{12}------------------------------------------------
sensiFirm – K170637/ S001
RD-12062 A0
{13}------------------------------------------------
sensiFirm – K170637/ S001
RD-12062 A0
{14}------------------------------------------------
sensiFirm – K170637/ S001
RD-12062 A0
{15}------------------------------------------------
sensiFirm — K170637/ S001
RD-12062 A0
Note No. 1. The massage Type
There are several methods commonly used for massaging, including: manual massage, pressuring, vibrations, rollers assisted, vacuum assisted etc. Vibration assisted massage offers safe and localized treatment with similar therapeutic effect. Such devices are classified as class I devices and are 510(K) exempt (product code IRO or ISA).
Note No. 2 - The Energy Delivered and Electrodes Architecture
Treatment with all of the devices (sensiFirm and both predicate devices) utilizes the same technology of Bi-polar RF, on the same anatomical sites and for the same purpose. Generally, safety and efficacy of the device for the intended use is affected by the method of RF delivery, power energy output and the configuration/ area of the treatment electrodes. All these factors were taken into account when designing the treatment protocol, which has been validated in a clinical trial.
Note No. 3 - The Treatment Regimen
The treatment regimen of the sensiFirm was especially planned to achieve its intended use in a safe and effective manner. Both sensiFirm and its predicate devices are designed for the exact same anatomical sites, with similar treatment
{16}------------------------------------------------
RD-12062 A0
protocols. Though the treatment protocols are different, the total exposure to the sensiFirm device throughout the entire treatment regimen is identical to predicate device #2 and similar to predicate device #1. Therefore, no issues in means of safety for the user are raised. Furthermore, the efficacy was validated in a clinical trial.
Note No. 4 - Applicability to Home Use Environment
While the sensiFirm was designed for OTC home use, both predicate devices are prescription devices, to be used in a clinic by a professional operator. Auxiliary device (Newa by EndyMed Medical Ltd.. cleared under DEN 150005) supports the OTC home use environment. Several device's features, all related to its design and usability elements, were specially tailored for the home use environment, which was tested in a Human Factors Test study and further validated in a clinical trial.
11. CONCLUSIONS
Due to these identical clauses and high similarities, the sensiFirm device is at least as safe and as effective as the predicate device. The differences between the sensiFirm device and the predicate device do not raise new safety or effectiveness issues or questions, as detailed above.
The performance specifications of the sensiFirm (i.e., RF frequency and RF power) are substantially equivalent to those of the predicate devices. The safety features and compliance with safety standards of both are similar. The minor differences in the technological characteristics do not raise new safety or effectiveness concerns and are demonstrated to be substantially equivalent through relevant performance tests. Furthermore, the sensiFirm device has qualified with varied performance tests, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1, electromagnetic compatibility testing according to IEC 60601-1-2, compatibility as medical electrical equipment for home healthcare environment according to
{17}------------------------------------------------
RD-12062 A0
60601-1-11, and high frequency of surgical equipment according to IEC 60601-2-
- The performance tests demonstrated that the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Based on the performance testing and comparison to predicate devices, the sensiFirm device is substantially equivalent to the previously cleared predicate devices.
. . . . . . . . . . .
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.