K Number
K170604

Validate with FDA (Live)

Date Cleared
2017-05-30

(90 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FilmArray® Respiratory Panel 2 (RP2) is a multiplexed nucleic acid test intended for use with FilmArray® 2.0 or FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray® RP2:

  • · Adenovirus
  • Coronavirus 229E
  • · Coronavirus HKU1
  • · Coronavirus NL63
  • Coronavirus OC43
  • Human Metapneumovirus
  • Human Rhinovirus/Enterovirus
  • · Influenza A, including subtypes H1, H1-2009, and H3
  • Influenza B
  • Parainfluenza Virus 1
  • Parainfluenza Virus 2
  • Parainfluenza Virus 3
  • Parainfluenza Virus 4
  • · Respiratory Syncytial Virus
  • · Bordetella parapertussis (IS1001)
  • Bordetella pertussis (ptxP)
  • Chlamydia pneumoniae
  • Mycoplasma pneumoniae

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the FilmArray® RP2 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus and Enterovirus, the FilmArray® RP2 cannot reliably differentiate them. A positive FilmArray® RP2 Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

Performance characteristics for Influenza A were established when Influenza A H1-2009 and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The FilmArray Respiratory Panel 2 (RP2) is designed to simultaneously identify 21 different potential pathogens (see Table 1) of respiratory tract infection from a single NPS specimen in a time frame (~45 minutes) that allows the test results to be used in determining appropriate patient treatment and management. FilmArray RP2 is compatible with BioFire Diagnostics' (BioFire) PCR-based in vitro diagnostic FilmArray 2.0 and FilmArray Torch systems for infectious disease testing. A specific software module (i.e. FilmArray RP2 pouch module) is used to perform FilmArray RP2 testing on these systems.

A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and an NPS sample (in transport media) mixed with the provided Sample Buffer into the other port of the FilmArray RP2 pouch and placing it in a FilmArray instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

AI/ML Overview

Here's a breakdown of the requested information about the FilmArray® Respiratory Panel 2 (RP2) based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for each analyte are implicitly demonstrated by the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) values meeting a high standard, generally interpreted as near 100% agreement. While explicit numerical acceptance targets are not stated (e.g., "PPA must be >95%"), the reported performance in both prospective and archived studies consistently shows high agreement. For brevity, I've listed a representative sample of analytes.

Analyte (Category)Acceptance Criteria (Implied)Reported Performance (Overall Prospective Clinical Study)Reported Performance (Archived Specimen Study)
Viruses
AdenovirusHigh PPA and NPAPPA: 94.6% (70/74), NPA: 96.9% (1490/1538)NPA: 97.4% (189/194) (No PPA data for archived)
Coronavirus 229EHigh PPA and NPAPPA: 91.7% (11/12), NPA: 99.7% (1595/1600)PPA: 100% (15/15), NPA: 100% (175/175)
Coronavirus HKU1High PPA and NPAPPA: 100% (43/43), NPA: 99.2% (1557/1569)NPA: 100% (194/194) (No PPA data for archived)
Human Metapneumovirus (hMPV)High PPA and NPAPPA: 97.3% (73/75), NPA: 99.5% (1529/1537)PPA: 100% (1/1), NPA: 99.5% (192/193)
Human Rhinovirus/EnterovirusHigh PPA and NPAPPA: 97.5% (425/436), NPA: 93.5% (1099/1176)PPA: 94.7% (18/19), NPA: 96.0% (168/175)
Influenza AHigh PPA and NPAPPA: 100% (78/78), NPA: 100% (1531/1531)PPA: 100% (22/22), NPA: 100% (172/172)
RSVHigh PPA and NPAPPA: 99.4% (175/176), NPA: 98.3% (1412/1436)PPA: 100% (2/2), NPA: 99.5% (191/192)
Bacteria
Bordetella parapertussisHigh PPA and NPAPPA: 85.7% (6/7), NPA: 100% (1605/1605)PPA: 100% (16/16), NPA: 100% (4/4)
Bordetella pertussisHigh PPA and NPAPPA: 66.7% (2/3), NPA: 99.9% (1608/1609)PPA: 96.2% (25/26), NPA: 98.8% (160/162)
Chlamydia pneumoniaeHigh PPA and NPAPPA: 100% (5/5), NPA: 99.9% (1606/1607)PPA: 100% (17/17), NPA: 100% (176/176)
Mycoplasma pneumoniaeHigh PPA and NPAPPA: 95.8% (23/24), NPA: 99.7% (1583/1588)PPA: 100% (16/16), NPA: 98.8% (171/173)

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Clinical Study:
    • Sample Size: 1612 valid nasopharyngeal swab (NPS) specimens.
    • Data Provenance: Obtained from individuals suspected of respiratory tract infections at three geographically distinct U.S. study sites during portions of the 2015-2016 and 2016-2017 respiratory illness seasons. Specimens were both prospectively collected and immediately tested ("fresh," N=940) and prospectively collected and frozen for later testing ("archived/frozen," N=695, with a total of 1635 acquired, 23 excluded, leaving 1612 in the final dataset).
  • Archived Specimens Study:
    • Sample Size: 214 valid preselected archived retrospective clinical specimens. These were chosen to supplement low-prevalence analytes from the prospective study.
    • Data Provenance: Archived NPS in VTM specimens that had previously tested positive for specific analytes (Coronavirus 229E, Influenza A H1, H3, Influenza B, Parainfluenza Virus 1, 4, Bordetella parapertussis, B. pertussis, Chlamydia pneumoniae, Parainfluenza Virus 2, 3, Mycoplasma pneumoniae). These were sourced from a BioFire internal collection.
  • Contrived Specimens Study:
    • Sample Size: 50 contrived positive specimens (spiked with Influenza A H1 at various concentrations) and 50 un-spiked negative specimens, totaling 100 specimens.
    • Data Provenance: Prepared using individual unique residual NPS specimens that had previously tested negative by the FDA-cleared multiplexed respiratory pathogen panel at the source laboratory. Tested at one of the clinical testing sites participating in the prospective evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their specific qualifications for establishing the initial ground truth. However, the ground truth was established by:

  • Primary Comparator: An FDA-cleared multiplexed respiratory pathogen panel. This implies that the results from this predicate device served as the primary reference, which itself would have undergone rigorous validation.
  • Secondary Comparator (for B. parapertussis): Two analytically-validated PCR assays followed by bi-directional sequencing. This is a highly specific and expert-level molecular technique.
  • Discrepancy Resolution: Discrepancies were investigated by performing independent molecular methods with primers different from the FilmArray RP2 and/or comparator method retesting. This suggests review and confirmation by skilled laboratory personnel, capable of performing and interpreting advanced molecular diagnostics.

4. Adjudication Method

The adjudication method described for the prospective clinical study's discrepant results was:

  • "The discrepancy investigation was mainly conducted by performing independent molecular methods with primers that are different from that of the FilmArray RP2 and/or comparator method retesting."
  • For B. parapertussis, the bi-directional sequencing data "meeting pre-defined quality acceptance criteria that matched organism-specific sequences deposited in the NCBI GenBank database" was considered Positive.

This indicates a multi-methodological approach to resolve conflicting results between the investigational device and the primary comparator, rather than a purely expert consensus reading without additional testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was mentioned or performed. This device is an in vitro diagnostic (IVD) assay designed for qualitative detection of nucleic acids, not an imaging device requiring human interpretation, therefore a human-in-the-loop study with AI assistance is not applicable. The device provides automated results.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

  • Yes, standalone performance was done. The entire clinical and analytical performance evaluation describes the algorithm's (FilmArray RP2 test's) direct performance against comparator methods and ground truth. The device is intended for "simultaneous qualitative detection and identification" which is an automated process without human interpretation of raw data. The "FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel."

7. Type of Ground Truth Used

The ground truth for the test sets was primarily established using a combination of:

  • FDA-cleared multiplexed respiratory pathogen panel: For the majority of analytes in both prospective and archived studies.
  • Analytically-validated PCR assays followed by bi-directional sequencing: Specifically for Bordetella parapertussis.
  • Independent molecular methods and/or retesting: For discrepancy resolution.
  • Known concentrations of organisms: For analytical studies like Limit of Detection and Inclusivity.

This represents a high standard of ground truth, relying on established and highly sensitive molecular diagnostic techniques.

8. Sample Size for the Training Set

The document describes a clinical study evaluating the device's performance, but it does not mention a separate "training set" for the algorithm's development. This is typical for IVD submissions where the device's analytical and clinical performance is validated rather than its algorithm being developed and trained on a dataset. The development of the assays and their internal thresholds would have occurred during the R&D phase prior to these validation studies.

9. How the Ground Truth for the Training Set Was Established

As no specific "training set" for the algorithm was mentioned in the context of this submission, the method for establishing its ground truth is not provided. The performance data presented are for the validation of the already-developed device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, with three intertwined shapes representing people.

Public Health Service

May 30, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

BioFire Diagnostics, LLC Kristen J. Kanack, Ph.D. Vice President, Regulatory and Clinical Affairs 515 Colorow Drive Salt Lake City, UT 84108

Re: K170604

Trade/Device Name: FilmArray® Respiratory Panel 2 (RP2) Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: II Product Code: OCC, OEM, OEP, OOU, OTG, OZE, OZX, OZY, OZZ, OOI, NSU Dated: February 28, 2017 Received: March 1, 2017

Dear Dr. Kanack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Steven R. Gitterman -S

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170604

Device Name

The FilmArray® Respiratory Panel 2 (RP2)

Indications for Use (Describe)

The FilmArray® Respiratory Panel 2 (RP2) is a multiplexed nucleic acid test intended for use with FilmArray® 2.0 or FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray® RP2:

  • · Adenovirus
  • Coronavirus 229E
  • · Coronavirus HKU1
  • · Coronavirus NL63
  • Coronavirus OC43
  • Human Metapneumovirus
  • Human Rhinovirus/Enterovirus
  • · Influenza A, including subtypes H1, H1-2009, and H3
  • Influenza B
  • Parainfluenza Virus 1
  • Parainfluenza Virus 2
  • Parainfluenza Virus 3
  • Parainfluenza Virus 4
  • · Respiratory Syncytial Virus

{3}------------------------------------------------

  • · Bordetella parapertussis (IS1001)
  • Bordetella pertussis (ptxP)
  • Chlamydia pneumoniae
  • Mycoplasma pneumoniae

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the FilmArray® RP2 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus and Enterovirus, the FilmArray® RP2 cannot reliably differentiate them. A positive FilmArray® RP2 Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

Performance characteristics for Influenza A were established when Influenza A H1-2009 and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary BioFire Diagnostics, LLC

FilmArray Respiratory Panel 2 (RP2)

Introduction:

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

BioFire Diagnostics, LLC 515 Colorow Drive Salt Lake City, UT 84108

Telephone: 801-736-6354 Facsimile: 801-588-0507

Contact: Kristen J. Kanack, ext. 330

Date Submitted: February 28, 2017

Device Name and Classification:

Trade Name: FilmArray Respiratory Panel 2 (RP2)

Regulation Number: 21 CFR 866.3980

Classification Name: Respiratory viral panel multiplex nucleic acid assay

Predicate Device:

K160068 - FilmArray Respiratory Panel (RP)

Intended Use:

The FilmArray® Respiratory Panel 2 (RP2) is a multiplexed nucleic acid test intended for use with FilmArray® 2.0 or FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray RP2:

  • Adenovirus ●
  • Coronavirus 229E ●
  • Coronavirus HKU1
  • Coronavirus NL63
  • Coronavirus OC43

{6}------------------------------------------------

  • Human Metapneumovirus ●
  • Human Rhinovirus/Enterovirus
  • Influenza A, including subtypes H1, H1-2009, and H3
  • Influenza B
  • . Parainfluenza Virus 1
  • Parainfluenza Virus 2
  • Parainfluenza Virus 3
  • . Parainfluenza Virus 4
  • Respiratory Syncytial Virus
  • Bordetella parapertussis (IS 1001)
  • Bordetella pertussis (ptxP)
  • Chlamydia pneumoniae ●
  • Mycoplasma pneumoniae ●

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the FilmArray RP2 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus and Enterovirus, the FilmArray RP2 cannot reliably differentiate them. A positive FilmArray RP2 Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

Performance characteristics for Influenza A were established when Influenza A H1-2009, A H1, and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description:

The FilmArray Respiratory Panel 2 (RP2) is designed to simultaneously identify 21 different potential pathogens (see Table 1) of respiratory tract infection from a single NPS specimen in a

{7}------------------------------------------------

time frame (~45 minutes) that allows the test results to be used in determining appropriate patient treatment and management. FilmArray RP2 is compatible with BioFire Diagnostics' (BioFire) PCR-based in vitro diagnostic FilmArray 2.0 and FilmArray Torch systems for infectious disease testing. A specific software module (i.e. FilmArray RP2 pouch module) is used to perform FilmArray RP2 testing on these systems.

Viral Targets DetectedBacterial Targets Detected
AdenovirusCoronavirus 229ECoronavirus HKU1Coronavirus NL63Coronavirus OC43Human MetapneumovirusHuman Rhinovirus/EnterovirusInfluenza A, including subtypesH1, H3 and H1-2009Influenza BParainfluenza Virus 1Parainfluenza Virus 2Parainfluenza Virus 3Parainfluenza Virus 4Respiratory Syncytial VirusBordetella parapertussis (IS1001)Bordetella pertussis (ptxP)Chlamydia pneumoniaeMycoplasma pneumoniae

Table 1. Pathogens Detected by the FilmArray RP2

A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and an NPS sample (in transport media) mixed with the provided Sample Buffer into the other port of the FilmArray RP2 pouch and placing it in a FilmArray instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first

{8}------------------------------------------------

stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

Substantial Equivalence:

The FilmArray Respiratory Panel 2 is substantially equivalent to the FilmArray Respiratory Panel (K160068), which was most recently cleared on February 8, 2016 and determined to be a Class II device.

Table 2 compares the FilmArray Respiratory Panel 2 (RP2) to the FilmArray Respiratory Panel (RP) and outlines the similarities and differences between the two tests.

ElementNew Device:FilmArray Respiratory Panel 2 (RP2)Predicate:FilmArray Respiratory Panel (RP)(K160068)
Specimen TypesNPS in VTMSame
OrganismsDetectedAdenovirus, Coronavirus 229E, CoronavirusHKU1, Coronavirus NL63, Coronavirus OC43,Human Metapneumovirus, Influenza A,Influenza A subtype H1, Influenza A subtypeH3, Influenza A subtype H1-2009, Influenza B,Parainfluenza Virus 1, Parainfluenza Virus 2,Parainfluenza Virus 3, Parainfluenza Virus 4,Human Rhinovirus/Enterovirus, RespiratorySyncytial Virus, Bordetella parapertussis,Bordetella pertussis, Chlamydia pneumoniae,and Mycoplasma pneumoniaeDoes not detect Bordetella parapertussisReports Chlamydia pneumoniae by alternatename: Chlamydophila pneumoniae
AnalyteDNA/RNASame
TechnologicalPrinciplesMultiplex nucleic acidSame
InstrumentationFilmArray 2.0 or FilmArray TorchFilmArray, FilmArray 2.0, or FilmArray Torch
Time to resultAbout 45 minutesAbout 1 hour
Reagent StorageRoom temperatureSame
TestInterpretationAutomated test interpretation and reportgeneration. User cannot access raw data.Same

Table 2. Comparison of the FilmArray Respiratory Panel 2 (RP2) and the FilmArray Respiratory Panel (RP)

{9}------------------------------------------------

ControlsTwo controls are included in each reagent pouchto control for sample processing and both stagesof PCR and melt analysis.Same
User ComplexityModerate/LowSame

Summary of Performance Data

Clinical Performance

The clinical performance of the FilmArray RP2 was established during a multi-center study conducted at three geographically distinct U.S. study sites during portions of the 2015-2016 and 2016-2017 respiratory illness seasons. A total of 1635 residual NPS specimens in viral transport media (VTM) were acquired for the prospective clinical study. Between January and March 2016. specimens were prospectively collected from all comers meeting the study eligibility criteria and immediately frozen (N=695 specimens) for later testing as prospective archived/frozen (Category II) specimens. Between September 2016, specimens were prospectively collected from all comers meeting the study eligibility criteria and tested fresh (N=940 specimens) as prospective fresh (Category I) specimens. Category II specimens were distributed to study sites beginning in September 2016. Study sites also began testing Category I specimens at this time. At each site, Category II specimens were thawed and tested according to the study procedures as time permitted over the remaining duration of the clinical study. A total of 23 prospective specimens (Category I and II specimens) were excluded from the final performance data analysis due to incompliance with the study protocol. The most common reasons for specimen exclusion were that a valid external control was not completed on the day of testing, that specimens were tested outside the 3-day refrigerated storage window, or that the specimen was found to not meet the inclusion criteria after the specimen had been enrolled. The final data set consisted of 1612 prospective specimens. Table 3 provides a summary of demographic information for the 1612 specimens included in the prospective study.

OverallSite 1Site 2Site 3
Male867(54%)331(57%)271(51%)265(53%)
SexFemale745(46%)250(43%)256(49%)239(47%)
Age< 5 years885(55%)379(65%)170(32%)336(67%)
6 - 21 years331(21%)13289 (17%)(23%)110(22%)
22 - 49 years128 (8%)27 (5%)79 (15%)22 (4%)
50+ years268(17%)43 (7%)189(36%)36 (7%)
Outpatient329(20%)77 (13%)66 (13%)186(37%)
StatusHospitalized640(40%)229(39%)197(37%)214(42%)
Emergency643(40%)275(47%)264(50%)104(21%)
Total1612581527504

Table 3. Demographic Summary for Prospective FilmArray RP2 Clinical Evaluation

{10}------------------------------------------------

The performance of the FilmArray RP2 was evaluated by comparing the FilmArray RP2 test results with those from an FDA-cleared multiplexed respiratory pathogen panel (the main comparator method) as well as with results from two analytically-validated PCR assays followed by bi-directional sequencing for B. parapertussis (this analyte is not detected by the FDA-cleared multiplexed respiratory pathogen panel). The B. parapertussis comparator assays were designed to amplify a different sequence than that amplified by the FilmArray RP2. Any specimen that had bi-directional sequencing data meeting pre-defined quality acceptance criteria that matched organism-specific sequences deposited in the NCBI GenBank database (www.ncbi.nlm.nih.gov) with acceptable E-values was considered Positive. Any specimen that tested negative by both of the comparator assays was considered Negative.

Positive Percent Agreement (PPA) for each analyte was calculated as 100% x (TP / (TP + FN)). True positive (TP) indicates that both the FilmArray RP2 and the comparator method had a positive result for this specific analyte, and false negative (FN) indicates that the FilmArray RP2 result was negative while the comparator result was positive. Negative Percent Agreement (NPA) was calculated as 100% x (TN / (TN + FP)). True negative (TN) indicates that both the FilmArray RP2 and the comparator method had negative results, and a false positive (FP) indicates that the FilmArray RP2 result was positive but the comparator result was negative. The exact binomial two-sided 95% confidence interval was calculated. Samples for which false positive and/or false negative results (i.e., discrepant results) were obtained when comparing the FilmArray RP2 results to the comparator method results were further investigated. The discrepancy investigation was mainly conducted by performing independent molecular methods with primers that are different from that of the FilmArray RP2 and/or comparator method retesting. The prospective clinical study results are summarized in Table 4.

AnalytePositive Percent AgreementNegative Percent Agreement
TP/(TP + FN)%95%CITN/(TN + FP)%95%CI
Viruses
AdenovirusaFresh36/3894.782.7-98.5850/88096.695.2-97.6
Frozen34/3694.481.9-98.5640/65897.395.7-98.3
Overall70/7494.686.9-97.91490/153896.995.9-97.6
CoV-229EbFresh5/510056.6-100909/91399.698.9-99.8
Frozen6/785.748.7-97.4686/68799.999.2-100
Overall11/1291.764.6-98.51595/160099.799.3-99.9
CoV-HKU1cFresh1/1100-917/91710099.6-100
Frozen42/4210091.6-100640/65298.296.8-98.9
Overall43/4310091.8-1001557/156999.298.7-99.6
CoV-NL63dFresh0/0--917/91899.999.4-100
Frozen40/4010091.2-100645/65498.697.4-99.3
Overall40/4010091.2-1001562/157299.498.8-99.7
CoV-OC43eFresh11/1384.657.8-95.7904/90599.999.4-100
Frozen22/2878.660.5-89.8662/66699.498.5-99.8
Overall33/4180.566.0-89.81566/157199.799.3-99.9
hMPVfFresh5/510056.6-100913/91310099.6-100
AnalytePositive Percent AgreementNegative Percent Agreement
TP/(TP + FN)%95%CITN/(TN + FP)%95%CI
Frozen68/7097.190.2-99.2616/62498.797.5-99.3
Overall73/7597.390.8-99.31529/153799.599.0-99.7
Fresh320/32897.695.3-98.8532/59090.287.5-92.3
HRV/EVgFrozen105/10897.292.1-99.1567/58696.895.0-97.9
Overall425/43697.595.5-98.61099/117693.591.9-94.7
Fresh3/310043.9-100915/91510099.6-100
FluAhFrozen75/7510095.1-100616/61610099.4-100
Overall78/7810095.3-1001531/153110099.7-100
Fresh0/0--918/91810099.6-100
FluA H1Frozen0/0--691/69110099.4-100
Overall0/0--1609/160910099.8-100
Fresh0/0--918/91810099.6-100
FluA H1-2009Frozen74/7410095.1-100617/61710099.4-100
Overall74/7410095.1-1001535/153510099.8-100
Fresh3/310043.9-100915/91510099.6-100
FluA H3Frozen1/1100-690/69010099.4-100
Overall4/410051.0-1001605/160510099.8-100
Fresh0/0--918/91810099.6-100
FluBiFrozen14/1410078.5-100678/68099.798.9-99.9
Overall14/1410078.5-1001596/159899.999.5-100
Fresh0/0--918/91810099.6-100
MERS-CoVFrozen0/0--694/69410099.4-100
Overall0/0--1612/161210099.8-100
Fresh5/510056.6-100913/91310099.6-100
PIV1jFrozen4/410051.0-100689/69099.999.2-100
Overall9/910070.1-1001602/160399.999.6-100
Fresh46/4797.988.9-99.6863/87199.198.2-99.5
PIV2kFrozen0/0--694/69410099.4-100
Overall46/4797.988.9-99.61557/156599.599.0-99.7
Fresh40/4295.284.2-98.7867/87699.098.1-99.5
PIV3lFrozen3/310043.9-100690/69199.999.2-100
Overall43/4595.685.2-98.81557/156799.498.8-99.7
Fresh6/610061.0-100910/91299.899.2-99.9
PIV4mFrozen3/310043.9-100686/69199.398.3-99.7
Overall9/910070.1-1001596/160399.699.1-99.8
Fresh44/4597.888.4-99.6867/87399.398.5-99.7
RSVnFrozen131/13110097.2-100545/56396.895.0-98.0
Overall175/17699.496.9-99.91412/143698.397.5-98.9
Bacteria
B. parapertussis(IS1001)oFresh4/580.037.6-96.4913/91310099.6-100
Frozen2/210034.2-100692/69210099.4-100
Overall6/785.748.7-97.41605/160510099.8-100
B. pertussis(ptxP)pFresh2/210034.2-100915/91699.999.4-100
Frozen0/10.0-693/69310099.4-100
AnalytePositive Percent AgreementNegative Percent Agreement
TP/(TP + FN)%95%CITN/(TN + FP)%95%CI
C. pneumoniaeqOverall2/366.720.8-93.91608/160999.999.6-100
Fresh2/210034.2-100915/91699.999.4-100
Frozen3/310043.9-100691/69110099.4-100
Overall5/510056.6-1001606/160799.999.6-100
M. pneumoniaerFresh17/1710081.6-100897/90199.698.9-99.8
Frozen6/785.748.7-97.4686/68799.999.2-100
Overall23/2495.879.8-99.31583/158899.799.3-99.9

Table 4. FilmArray RP2 Prospective Clinical Performance Summary

{11}------------------------------------------------

BioFire Diagnostics 510(k)
FilmArray Respiratory Panel 2 (RP2)

{12}------------------------------------------------

Adenovirus was detected in 34 FN speciment molecular method. Adenovirus was detected in 38/48 FP specimens using an independent molecular method; an additional two FP specimens were collected from subjects with an acute history of adenovirus infection.

9 The single FN specimen was negative for CoV-229E when tested using an independent molecular method. All five FP specimens were negative for CoV-229E when tested using an independent molecular method.

CoV-HKU1 was detected in 3/12 FP specimens upon comparator method retest.

9 CoV-NL63 was detected in 3/10 FP specimens during discription; two were detected using an independent molecular method and one was detected upon comparator method retest.

° Of the eight FN specimens, six were TP for Col-HKU1. They were confirmed to be due to a known cross-neactivity with CoV-HKU1 by the comparator method; All six specimens were negative for Col-OC43 when tested with two independent PCR assays; the remaining two FN specimens were negative for CN-0043 when tested using an independent molecular method. CoV-0043 was detected in 2/5 FP specimens upon comparator method retest.

Both FN specimens were negative for hMPV when tested using an independent molecular method. MMPV was detected in 6/8 FP specimens during discrepancy investigation; one was detected using an ine were detected upon comparator method retest.

9 HRV/EV was detected in 5/11 FN specimens during discrepation; one was detected using an independent molecular method and four were detected upon FilmArray RP2 retest. HRV/EV was detected in 3377 FP specimens during discrepancy investigation; four were detected using an independent molecular method and 29 were detected upon comparator method retest.

" Three specimens were excluded from influenza A analysis: one with a comparator method result of Influenza A (No Subtype Detected) and two FilmArray RP2 Influenza A (Equivocal) detections.

' FluB was detected in both FP specimens during discripation; one was detected using an independent molecular method and one was detected upon comparator method retest.

1 The single FP specimen was negative for PIV1 when tested using an independent molecular method.

^ The single FN specimen was negative for PV2 when tested using an independent molecular method. PV2 FP specimens during discrepancy investigation; one was detected using an method and four were detected upon comparator method retest.

PV3 was detected in both FN specimens during discrepancy investigation; one was detected using an independent molecular method and one was detected upon FilmArray RP2 retest. PIV3 was detected in 4/10 FP specimens during discrepancy investigation; two were detected using an independent molecular method and two were detected upon comparator method retest.

"PIV4 was detected in 1/7 FP specimens using an independent molecular method.

" The single FN specimen was negative for RSV when tested using an independent molecular method. RSV was detected in 8/24 FP specimens during discrepancy investigation; three were detected using an independent mothod and five were detected upon comparator method retest.

· B. parapertussis was detected in the single FN specimen upon FilmArray RP2 retest.

° B. pertussis was detected in the FN and FP specimens using an independent molecular method.

9 C. pneumoniae was detected in the single FP specimen using an independent molecular method.

[ M. pneumoniae was detected in the single FN specimen upon FilmArray RP2 retest. M. pneumoniae was detected in all five FP specimens during discrepancy investigation: three were detected using an independent mothod and two were detected upon comparator method retest.

FilmArray RP2 reported a total of 245 specimens with discernible multiple organism detections (15.2% of all specimens, 245/1612; and 24.0% of positive specimens, 245/1020; Table 5). The majority of multiple detections (190/245; 77.6%) contained two organisms, while 20.0% (49/245) contained three organisms, 1.6% (4/245) contained four organisms, 0.4% (1/245) contained five organisms, and 0.4% (1/245) contained six organisms. Out of the 245 specimens with multiple detections, 124 specimens (50.6%; 124/245) were concordant with the comparator methods. One hundred twenty-one (121) specimens (49.4%; 121/245) contained one or more organisms that had not been detected by the comparator methods (i.e. false positive results).

{13}------------------------------------------------

The three organisms that were most prevalent in multiple detections were also the three most prevalent organisms in the study as a whole (i.e. HRV/EV, RSV, and adenovirus). The most prevalent multiple detections (≥5 instances) are shown in Table 6.

AnalytePrevalence in MultipleDetections (N=245)
Viruses
Adenovirus8534.7%
CoV-229E62.4%
CoV-HKU14116.7%
CoV-NL633112.7%
CoV-OC43197.8%
hMPV3313.5%
HRV/EV15061.2%
FluA H100%
FluA H1-200993.7%
FluA H320.8%
FluB62.4%
PIV152.0%
PIV2156.1%
PIV3218.6%
PIV4124.9%
RSV10542.9%
Bacteria
B. parapertussis (IS1001)62.4%
B. pertussis (ptxP)00%
C. pneumoniae10.4%
M. pneumoniae72.9%

Table 5. Prevalence of Analytes in Multiple Detections as determined by the FilmArray RP2

The most prevalent multiple detection was adenovirus with HRV/EV (1.9% of all specimens; 30/1612) followed by HRV/EV with RSV (1.4% of all specimens; 22/1612); as previously stated these were also the most prevalent organisms detected in the study.

Table 6. Multiple Detection Combinations (≥5 instances) as Determined by the FilmArray RP2

Distinct Co-Detection CombinationsTotalMultipleDetectionsNumber ofSpecimens withFalse PositiveResultsFalse Positive Analyte(s)a
Analyte 1Analyte 2Analyte 3
AdenovirusHRV/EV3015Adenovirus (15), HRV/EV (1)
HRV/EVRSV227HRV/EV (3), RSV (4)

{14}------------------------------------------------

CoV-HKU1RSV137CoV-HKU1 (4), RSV (3)
CoV-NL63RSV133CoV-NL63 (2), RSV (1)
HRV/EVPIV2117HRV/EV (6), PIV2 (2)
HRV/EVPIV3116HRV/EV (3), PIV3 (4)
AdenovirusRSV105Adenovirus (4), RSV (1)
AdenovirusHRV/EVRSV95Adenovirus (2), HRV/EV (3), RSV (1)
CoV-NL63HRV/EV82CoV-NL63 (2)
CoV-HKU1HRV/EV52CoV-HKU1 (1), HRV/EV (1)
CoV-OC43HRV/EV53HRV/EV (3)
hMPVHRV/EV51HRV/EV

ª Of the 67 discrepant analytes (out of 293 total analytes), 32 (47.8%) were observed as being present in the specimen during discrepancy investigation; 22/67 (32.8%) were observed using an method and 13/67 (19.4%) were observed upon comparator method retest.

The overall success rate for initial specimen tests in the prospective study was 99.3% (1611/1623) (95% CI: 98.7% - 99.6%); 12 tests were unsuccessful (one due to an incomplete test, one due to an instrument error, and ten due to control failures). Two tests (2/1623; 0.1%) did not complete on the initial run, resulting in an instrument success rate of 99.9% (1621/1623) (95% CI: 99.6% - 100%) for initial specimen tests. Both specimens were able to be retested and valid results were produced after a single retest. Ten tests (10/1621; 0.6%) did not produce valid pouch controls, resulting in a pouch control success rate of 99.4% (1611/1621) (95% CI: 98.9% -99.7%) for completed runs in the initial specimen tests. Nine of the 10 invalid specimens were able to be retested and produced valid control results after a single retest; one was not able to be retested due to insufficient specimen volume.

Testing of Preselected Archived Specimens

Some of the analytes on the FilmArray RP2 were of low prevalence and were not encountered in large enough numbers during the prospective study to adequately demonstrate system performance. To supplement the results of the prospective clinical study, an evaluation of preselected archived retrospective specimens was performed at BioFire. These specimens were archived NPS in VTM specimens that were selected because they had previously tested positive for one of the following analytes: coronavirus 229E, influenza A H1, influenza A H3, influenza B, parainfluenza virus 1, parainfluenza virus 4, Bordetella parapertussis, B. pertussis, and Chlamydia pneumoniae. Parainfluenza virus 2, parainfluenza virus 3, and Mycoplasma pneumoniae were also expected to be low prevalence based on BioFire data collected during the 2015-2016 respiratory season, therefore archived testing was performed for these analytes as well and included in the study data (although ultimately they were observed in larger numbers during the prospective clinical study).

A total of 217 preselected archived retrospective clinical specimens were initially received for testing in this retrospective study. Prior to testing with the FilmArray RP2, the composition/integrity of the specimens was first confirmed with confirmatory molecular methods (PCR followed by bi-directional sequencing for B. parapertussis or an FDA-cleared multiplexed respiratory pathogen panel.

{15}------------------------------------------------

The specimens were divided into two different groups for testing based on the method of confirmation testing performed: all specimens containing analytes on the FDA-cleared multiplexed respiratory pathogens panel comparator method were tested in Group 1 and specimens containing B. parapertussis were tested in Group 2. Negative NPS specimens were also included in each group for testing.

The FDA-cleared multiplexed respiratory pathogen panel comparator method was performed on 197 of the 217 preselected archived retrospective clinical specimens only (Group 1). One of the 197 specimens was excluded from performance analysis because of an invalid FilmArray RP2 run with insufficient volume to retest. Additionally, two of the 197 specimens were also excluded from performance analysis because a valid FDA-cleared multiplexed respiratory pathogens panel comparator method confirmation result was not obtained and there was insufficient specimen volume for retesting: one comparator run was incomplete and the other comparator run had a control failure. Valid comparator method and FilmArray RP2 results were obtained for 194 of these 197 archived specimens (Group 1).

The B. parapertussis PCR followed by bi-directional sequencing comparator assays were performed on 20 of the 217 preselected archived retrospective clinical specimens only (Group 2). The FDA-cleared multiplexed respiratory pathogens panel comparator method was not performed on Group 2 specimens. Valid comparator method and FilmArray RP2 results were obtained for 20 of these 20 archived specimens.

A summary of the available demographic information of these 214 valid archived specimens is provided in Table 7.

Total Specimens214
SexFemale (%)75 (35%)
Male (%)81 (38%)
Unknown58 (27%)
Age Range≤ 5 years78 (36%)
6 - 21 years46 (21%)
22 - 49 years13 (6%)
50+ years19 (9%)
Unknown58 (27%)

Table 7. Available Demographic Summary for All Valid Archived Specimens

All Group 1 and Group 2 positive archived specimens (as determined at the source laboratory) that were not confirmed by the respective comparator method were further excluded from the performance calculation for each of the respective analytes.

The FilmArray RP2 retrospective specimens testing performance data against the comparator methods are provided in Table 8 by analyte.

{16}------------------------------------------------

Positive Percent AgreementNegative Percent Agreement
AnalyteTP/(TP + FN)%95% CITN/(TN + FP)%95% CI
Viruses
Adenovirus0/00N/A189/19497.494.1-98.9
CoV- 229Ea15/1510079.6-100175/17510097.9-100
CoV-HKU10/00N/A194/19410098.1-100
CoV-NL632/210034.2-100192/19210098.0-100
CoV-OC430/00N/A194/19410098.1-100
hMPV1/110020.7-100192/19399.597.1-99.9
HRV/EV18/1994.775.4-99.1168/17596.092.0-98.0
Influenza A22/2210085.1-100172/17210097.8-100
Influenza A H13/310043.9-100191/19110098.0-100
Influenza A 2009-H11/110020.7-100193/19310098.0-100
Influenza A H318/1810082.4-100176/17610097.9-100
Influenza Bb16/1610080.6-100177/17710097.9-100
Parainfluenza Virus 116/1610080.6-100178/17810097.9-100
Parainfluenza Virus 2c16/1610080.6-100177/17710097.9-100
Parainfluenza Virus 317/1710081.6-100175/17798.996.0-99.7
Parainfluenza Virus 417/1710081.6-100174/17798.395.1-99.4
RSV2/210034.2-100191/19299.597.1-99.9
Bacteria
Bordetella parapertussis (IS1001)d16/1610080.6-1004/410051.0-100
Bordetella pertussis (ptxP)e25/2696.281.1-99.3160/16298.895.6-99.7
Chlamydia pneumoniaef17/1710081.6-100176/17610097.9-100
Mycoplasma pneumoniaeg16/1610080.6-100171/17398.895.9-99.7

Table 8, FilmArray RP2 Archived Specimen Performance Data Summary

ª Four of 19 CoV-229E positive archived specimens by the source laboratory were not confirmed by the comparator method and therefore were excluded from the performance calculation for CoV-229E

  • One of the 17 Influenza B positive archived specimens by the source laboratory was not confirmed by the comparator method and therefore was excluded from the performance calculation for Influenza B.

^ One of the 17 Parainfluenza Virus 2 positived specimens the source laboratory was not confirmed by the comparator method and therefore was excluded from the performance calculation for Parainfluenza Virus 2 .

4 The comparator B. parapertussis PCR followed by sequencing assays were performed on 20 achieved specimens only (Group 2). The comparator method for the other analytes was not performed on these 20 specimens.

Six of the 31 B. pertussis positive archived specifical by the source aboratory were not confirmed by the comparator were excluded from the performance calculation for B. pertussis.

f One of the 17 C. pneumoniae positived specimens by the souce laboratory was not confirmed by the comparator method and therefore was excluded from the performance calculation for C. pneumoniae.

s Five of the 21 M. pneumoniae positived specimens by the source laboratory were not confirmed by the comparator method and therefore were excluded from the performance calculation for M. pneumoniae.

Testing of Contrived Specimens

Influenza A H1 is of such rarity that both prospective and retrospective archived testing efforts were insufficient to demonstrate system performance. To supplement the prospective and retrospective data, an evaluation of contrived specimens was performed at one of the three clinical testing sites participating in the prospective evaluation. Contrived clinical specimens were prepared using individual unique residual NPS specimens that had previously tested

{17}------------------------------------------------

negative by the FDA-cleared multiplexed respiratory pathogen panel (i.e., the same test as the comparator method employed in the prospective and retrospective clinical evaluations) at the source laboratory. Spiking was performed using multiple quantified isolates of Influenza A H1. The spiking scheme was such that at least 25 of the contrived positive specimens had analyte concentrations at 2 × the limit of detection (LoD), while the remaining 25 contrived positive specimens were at additional concentrations that spanned the clinically relevant range which was based on FilmArray® RP2 Cp observations of influenza A (A H1, A H-2009, and H3) from the prospective and archived specimen studies. Contrived positive specimens were prepared and randomized along with 50 un-spiked influenza A H1 negative specimens such that the analyte status of each contrived specimen was unknown to the users performing the testing. The results of the FilmArray RP2 testing contrived specimens are presented in Table 9.

Positive Percent AgreementNegative Percent Agreement
Analyte× LoDTP/(TP + FN)%95% CITN/(TN + FP)%95% CI
Influenza A H1222/23 a95.7%79.0-99.250/5010092.9-100
1010/10100%72.3-100
505/5100%56.6-100
2005/5100%56.6-100
10005/5100%56.6-100
Combined47/48 a97.9%89.1-99.6

Table 9. FilmArray RP2 Performance Using Contrived Specimens

4 The FN specimen was spiked with influenza A/Weiss/43; this strain was detected at all other concentrations. Two specimens (also spiked with strain A/Weiss/43) had a result of Influenza A Equivocal and were excluded from Influenza A H1 performance calculation.

{18}------------------------------------------------

Selected Analytical Studies

Limit of Detection

The limit of detection (LoD) for FilmArray RP2 analytes was estimated by testing dilutions of contrived samples containing known concentrations of organisms. Confirmation of LoD was achieved by testing 20 replicates on FilmArray 2.0 and 20 replicates on FilmArray Torch. LoD was confirmed when the organism was detected in at least 19 of the 20 replicates tested (19/20 = 95%). The confirmed LoD for each FilmArray RP2 analyte is listed in Table 10 in viable units. An equivalent DNA or RNA copies/mL was calculated based on an alternate quantitative PCR assay for each analyte. LoD is equivalent when testing on FilmArray 2.0 and FilmArray Torch systems.

RP2 AnalyteIsolateLoDConcentrationFilmArray2.0FilmArray Torch
Viruses
Species A, Serotype 18ATCC VR-195.0E+00 TCID50/mL7.6E+03 copies/mL20/20100%19/2095%
Species B, Serotype 7AZeptometrix 0810021CF5.0E-02 TCID50/mL3.9E+01 copies/mL20/20100%20/20100%
AdenovirusaSpecies C, Serotype 2ATCC VR-8462.0E+00 TCID50/mL3.7E+01 copies/mL19/2095%20/20100%
Species D, Serotype 37Zeptometrix 0810119CF5.0E-02 TCID50/mL9.0E+00 copies/mL20/20100%20/20100%
Species E, Serotype 4aS. Carolina/2004, UIRF1.0E+01 TCID50/mL3.0E+03 copies/mL19/2095%19/2095%
Species F, Serotype 41Tak, ATCC VR-9301.0E+00 TCID50/mL1.2E+02 copies/mL20/20100%20/20100%
Coronavirus 229EATCC VR-7404.0E-01 TCID50/mL6.5E+01 copies/mL20/20100%20/20100%
Coronavirus HKU1Clinical specimen2.0E+03 RNA copies/mLb20/20100%20/20100%
Coronavirus NL63BEI NR-4702.5E-01 TCID50/mL5.4E+01 copies/mL20/20100%20/20100%
Coronavirus OC43ATCC VR-7593.0E+01 TCID50/mL5.6E+02 copies/mL19/2095%20/20100%
Human Metapneumovirus16, Type A1 IA10-2003Zeptometrix 0810161CF1.0E+01 TCID50/mL1.2E+03 copies/mL20/20100%20/20100%
Human Rhinovirus/EnterovirusaHuman RhinovirusType 1AZeptometrix 0810012CFN1.0E-01 TCID50/mL8E+01 copies/mL20/20100%20/20100%
Enterovirus D68ATCC VR-18233.0E+02 TCID50/mL2.6E+01 copies/mL20/20100%20/20100%
Influenza A H1Influenza A H1N1A/New Caledonia/20/99Zeptometrix 0810036CF1.0E+03 TCID50/mL1.4E+02 copies/mL20/20100%20/20100%
Influenza A H1-2009Influenza A H1N1 pdmA/Swine/NY/03/2009Zeptometrix 0810249CF5.0E-01 TCID50/mL3.3E+02 copies/mL20/20100%20/20100%
Influenza A H3Influenza H3N2A/Port Chalmers/1/73ATCC VR-8101.0E-01 TCID50/mL2.1E+01 copies/mL20/20100%20/20100%
Influenza BB/FL/04/06Zeptometrix 0810255CF5.0E+00 TCID50/mL3.4E+01 copies/mL20/20100%20/20100%

Table 10. Summary of Limit of Detection (LoD) for FilmArray RP2 Analytes

{19}------------------------------------------------

RP2 AnalyteIsolateLoDConcentrationFilmArray2.0FilmArray Torch
Parainfluenza Virus 1Type 1Zeptometrix 0810014CF$5.0E+00$ TCID50/mL$1.0E+03$ copies/mL20/20100%20/20100%
Parainfluenza Virus 2Type 2Zeptometrix 0810015CF$5.0E-01$ TCID50/mL$3.0E+01$ copies/mL20/20100%20/20100%
Parainfluenza Virus 3Type 3Zeptometrix 0810016CF$2.5E+00$ TCID50/mL$3.8E+01$ copies/mL19/2095%20/20100%
Parainfluenza Virus 4Type 4aZeptometrix 0810060CF$5.0E+01$ TCID50/mL$1.6E+03$ copies/mL20/20100%20/20100%
Respiratory Syncytial VirusType AZeptometrix 0810040ACF$2.0E-02$ TCID50/mL$9.0E+00$ copies/mL19/2095%20/20100%
Bacteria
Bordetella parapertussis( IS1001 )A747Zeptometrix 0801461$4.1E+01$ CFU/mL$6.0E+01$ IS1001 copies/mLc20/20100%19/2095%
Bordetella pertussis (ptxP)A639Zeptometrix 0801459$1.0E+03$ CFU/mL20/20100%20/20100%
Chlamydia pneumoniaeTW183ATCC VR-2282$1.0E-01$ TCID50/mL$6.6E+01$ copies/mL20/20100%19/2095%
Mycoplasma pneumoniaeM129Zeptometrix 0801579$1.0E+01$ TCID50/mL$4.6E+02$ copies/mLd20/20100%20/20100%

"The differences in LoD concentration between Adenovines species (5.0E-02 - 1.0E-01 TCD-J/mL) or between Human Rhinovirus (1.0E-01 - 3.0E+02TCID=ymL) may not indicate actual different species or serotypes, but ather the variability of the TCD-measurement method.

" A cultured isolate of Coronavirus HKU1 was not available for testing. LoD for Coronavirus HKU1 was theremined by testing dilutions of a clinical NPS specimen known to contain the virus. The anount of viral RNA copies/mL) was determined by real-ime RT-PCR against a standard curve.

  • IS/001 sequences can be present in more than one copy per cell, so the relationship between CFU/mL and copies/mL may vary from strain and culture to culture. LoD was determined based on the copy number of IS 001 measured by an independent quantitative real-ime PCR assay and for this culture.

The copy number of Mycoplasma pneumoniae was determined by a multicopy assay (16S rRNA) and may not accurately reflect the number of copies detected by the single-copy FilmArray RP2 gene target.

Note: LoD concentrations of the cultured viruses and obligate intracelleular bacteria (C. pneumoniae and M. pneumoniae) are provided in units of TCID-6 (50% Tissue Culture Infectious Dose). TCID50 is an indirect measure of viral or bacterial concentration based on infectivity and cytotoxicity and will therefore vary considerably depending on technique and methodology (including cell type, culture media and conditions, cytotoxicity of the virus, etc.). It is not appropriate to make determinations on relative sensitivity of different molecular assays for detection of viruses and bacteria based on LoD values measured in TCID50/mL

Inclusivity

Analytical reactivity (inclusivity) was evaluated with a collection of 177 isolates that represent temporal and geographic diversity of FilmArray RP2 analytes, including relevant species, strains, serotypes, or genotypes. All isolates tested were detected by the FilmArray RP2 at concentrations within 10× LoD. When possible, in silico analysis of sequence data was used to make predictions of assay reactivity for less common strains or serotypes that were not tested but that may be detected by the FilmArray RP2 (Table 11 - Table 22).

{20}------------------------------------------------

RP2 influenza A assays will react variably with non-human influenza A viruses and rarely encountered human influenza A viruses that are not H1, H1-2009 or H3; generally producing Influenza A Equivocal or Influenza A (no subtype detected) results.

Note: FilmArray RP2 Influenza A (subtype) and Influenza B assays are predicted to react with attenuated viruses used in vaccines.

SpeciesaSerotypeIsolate ID/Source[Strain/Location/Year]xLoDDectectedResult
A12ATCC VR-863[Huie/Massachusetts]3x
A18ATCC VR-19[Washington DC/1954]1x
A31Zeptometrix 0810073CF-3x
A3Zeptometrix 0810062CF-3x
A7AZeptometrix 0810021CF-1x
B7d/d2Univ of Iowa Research Foundation[Iowa/2001]3x
B7hUniv of Iowa Research Foundation[Iowa/1999]3x
B11Univ of Iowa Research Foundation[Wisconsin/2005]3x
B14Univ of Iowa Research Foundation[Missouri/2005]3x
B16ATCC VR-17[CH.79/Saudia Arabia/1955]3x
B21Univ of Iowa Research Foundation[Missouri/2005]3x
B34ATCC VR-716[Compton/1972]3x
B35ATCC VR-718[Holden]3x
B50ATCC VR-1602[Wan/Amsterdam/1988]3xAdenovirusDetected
C1Zeptometrix 0810050CF-3x
C2ATCC VR-846[Adenoid 6]1x
C5Zeptometrix 0810020CF-3x
C6ATCC VR-6[Tonsil 99/Washington DC]3x
D8Zeptometrix 0810069CF-3x
D20Zeptometrix 0810115CF-3x
D37Zeptometrix 0810119CF-1x
E4aUniv of Iowa Research Foundation[S Carolina/2004]1x
E4Zeptometrix 0810070CF-3x
E40Zeptometrix 0810084CFNCPV 0101141v-3x
F41ATCC VR-930[Tak/73-3544/Netherlands/1973]1x
F41Zeptometrix 0810085CF-3x

Table 11. Adenovirus Isolates Tested and Detected by FilmArray RP2

ª In silco analysis of available sequences predicts that the FilmArray RP2 will also react with Adenovirus B55, C57, species D serotypes, and G52.

Table 12. Coronavirus Isolates/Specimens Tested and Detected by FilmArray RP2
-------------------------------------------------------------------------------
CoronavirusTypeIsolate ID/Source[Location/Year]xLoDDectectedResult
229EATCC VR-740-1xCoronavirus 229E
229EZeptometrix 0810229CF-3xCoronavirus 229E
HKU1Clinical Specimen[Utah/2015]1x
HKU1Clinical Specimen[Utah/2015]3x
HKU1Clinical Specimen[Utah/2015]3xCoronavirus HKU1
HKU1Clinical Specimen[S. Carolina/2010]3xCoronavirus HKU1
HKU1Clinical Specimen[Detroit/2010]3xCoronavirus HKU1
NL63BEI NR-470a[Amsterdam/2003]1xCoronavirus NL63
NL63Zeptometrix 0810228CF-3xCoronavirus NL63
OC43ATCC VR-759b-1xCoronavirus OC43
OC43Zeptometrix 0810024CF-3xCoronavirus OC43

ª Organism obtained through the NIH Biodefense and Emerging Infections Research Resources Repository, NIAID, NIH: Human Coronavirus NL63, NR-470.

{21}------------------------------------------------

b Discontinued part number; see ATCC VR-1558.

GenotypeSerotypeIsolate ID/Source[Location/Year]xLoDDetectedResult
A116Zeptometrix 0810161CF[Iowa10/2003]1xHumanMetapneumovirus
A19Zeptometrix 0810160CF[Iowa3/2002]3x
A220Zeptometrix 0810163CF[Iowa14/2003]3x
A227Zeptometrix 0810164CF[Iowa27/2004]3x
B13Zeptometrix 0810156CF[Peru2/2002]3x
B15Zeptometrix 0810158CF[Peru3/2003]3x
B113Univ of Iowa Research Foundation[Iowa7/2003]3x
B24Zeptometrix 0810157CF[Peru1/2002]3x
8Zeptometrix 0810159CF[Peru6/2003]3x
18Zeptometrix 0810162CF[Iowa18/2003]3x
22Univ of Iowa Research Foundation[Iowa16/2003]3x

Table 13. Human Metapneumovirus Isolates Tested and Detected by FilmArray RP2

Table 14. Human Rhinovirus and Enterovirus Isolates Tested and Detected by FilmArray RP2

SpeciesSerotypeIsolate ID/Source[Strain/Location/Year]xLODDetectedResult
Human Rhinovirus
1Zeptometrix 0810012CFN[1A]1x
2ATCC VR-482[HGP]3x
7ATCC VR-1601[68-CV11]3x
A16ATCC VR-283[11757/WashingtonDC/1960]3x
34ATCC VR-507ª[137-3]3x
57ATCC VR-1600[Ch47]3xHumanRhinovirus/Enterovirus
77ATCC VR-1187[130-63]3x
85ATCC VR-1195[50-525-CV54]3x
3ATCC VR-483[FEB]3x
14ATCC VR-284[1059/S Carolina/1959]3x
B17ATCC VR-1663[33342/N Carolina/1959]3x
27ATCC VR-1137[5870]3x
42ATCC VR-338[56822]3x
83ATCC VR-1193[Baylor 7]3x
Enterovirus
ACoxsackievirus 10ATCC VR-168[NY/1950]3x
Enterovirus 71ATCC VR-1432[H]3x
Coxsackievirus A9Zeptometrix 0810017CF-3xHumanRhinovirus/Enterovirus
Coxsackievirus B3Zeptometrix 0810074CF-3x
Coxsackievirus B4Zeptometrix 0810075CF-3x
BEchovirus 6Zeptometrix 0810076CF-3x
Echovirus 9Zeptometrix 0810077CF-3x
Echovirus 11Zeptometrix 0810023CF-3x
CCoxsackievirus A21ATCC VR-850[Kuykendall/California/1952]3x
CCoxsackievirus A24ATCC VR-583[DN-19/Texas/1963]3x
D68ATCC VR-1823[US/MO/2014-18947]1x
TypeIsolate ID/Source[Strain/Location/Year]xLoDDetectedResult
Zeptometrix 0810036CF[New Caledonia/20/1999]1x
H1N1HumanATCC VR-219[NWS/1933]3xInfluenza AH1
ATCC VR-95[PR/8/1934]10xa
ATCC VR-96[Weiss/1943]3x
ATCC VR-97[FM/1/1947]3x
ATCC VR-98[Mal/302/1954]3x
ATCC VR-546[Denver/1/1957]3x
SwineZeptometrix 0810036CFN[Solomon Isl/03/2006]3x
Zeptometrix 0810244CF[Brisbane/59/2007]3x
ATCC VR-333[A/Swine/Iowa/15/1930]3x
SwineATCC VR-99[A/Swine/1976/1931]3x
ATCC VR-897[A/New Jersey/8/76 (Hsw1N1)]10xa
H1N2RecombinantBEI NR-9677b[Kilbourne F63, A/NWS/1934 (HA) xA/Rockefeller Institute/5/1957 (NA)]3x
H1N1pdm09HumanZeptometrix 0810249CFN[SwineNY/03/2009]1xInfluenza AH1-2009
Zeptometrix 0810248CFN[SwineNY/01/2009]3x
Zeptometrix 0810109CFN[SwineNY/02/2009]3x
Zeptometrix 0810109CFJ[Canada/6294/2009]3x
Zeptometrix 0810165CF[California/07/2009]3x
Zeptometrix 0810166CF[Mexico/4108/2009]3x
BEI NR-19823c[Netherlands/2629/2009]3x
BEI NR-44345d[Hong Kong/H090-761-V1(0)/2009]10xe
BEI NR-42938f[Georgia/F32551/2012]3x
HumanATCC VR-810[Port Chalmers/1/1973]1x
ATCC VR-776[Alice (live attenuated vaccine)]3x
Zeptometrix 0810238CF[Texas/50/2012]3x
ATCC VR-547[Aichi/2/1968]3xInfluenza AH3
H3N2gATCC VR-544[Hong Kong/8/1968]3x
ATCC VR-822[Victoria/3/1975]3x
Zeptometrix 0810252CF[Wisconsin/67/2005]3x
Zeptometrix 0810138CF[Brisbane/10/2007]3x
RecombinantATCC VR-777[MCR2(A/England/42/72xA/PR8/34)]3x
H3N2vHumanClinical Specimen[Ohio/2012]3x
H2N2HumanBEI NR-2775h[Japan/305/1957]10xeInfluenza A(no subtypedetected)
RecombinantBEI NR-9679i[Korea/426/1968xPuerto Rico/8/1934 ]10xeInfluenza A(no subtypedetected)
H2N3MRI Globalj[Mallard/Alberta/79/2003]3xInfluenza AEquivocal
H5N1AvianMRI Globalj[A/Chicken/Yunnan/1251/2003]3x
H5N2MRI Globalj[A/Northernpintail/Washington/40964/2014]3xInfluenza A
H5N3BEI NR-9682k[A/Duck/Singapore/645/1997]3x(no subtypedetected)
H5N8MRI Globalj[A/Gryfalcon/Washington/41088-6/2014]3x
H7N7MRI Globalj[A/Netherlands/219/2003]3x
H7N9MRI Globalj[A/Anhui/01/2013]3x
H10N7BEI NR-2765l[Chicken/Germany/N/49]3xInfluenza AEquivocal
LineageIsolate ID/Source[Strain/Location/Year]xLoDDetectedResult
N/AATCC VR-101[Lee/1940]3x
ATCC VR-102[Allen/1945]3x
ATCC VR-103[GL/1739/1954]3x
ATCC VR-296[1/Maryland/1959]3x
ATCC VR-295[2/Taiwan/1962]3x
ATCC VR-786[Brigit/Russia/1969]3x
VictoriaATCC VR-823[5/Hong Kong/1972]3xInfluenza B
Zeptometrix 0810258CF[2506/Malaysia/2004]3x
CDC 2005743348[1/Ohio/2005]3x
YamagataZeptometrix 0810256CF[07/Florida/2004]3x
Zeptometrix 0810255CF[04/Florida/2006]1x
Zeptometrix 0810241CF[1/Wisconsin/2010]3x
Zeptometrix 0810239CF[2/Massachusetts/2012]3x

ª Discontinued part number; see ATCC VR-1365.

Table 15. Influenza A Isolates Tested and Detected by FilmArray RP2

{22}------------------------------------------------

4 Reported as Influenza A (no subtype detected) at 3× LoD.

b Genomic RNA obtained through the NIH Biodefense and Emerging Infections Resources Respiratory NAID, NIH Kilbourne F63: A/NWS/1934 (HA) x A/Rockefeller Institute/5/1957 (NA) (H1N2), Reassortant NWS-F, NR-9677.

& Organism obtained through BEI Resources, NIAID, NIH: Influenza A Virus, A/Netherlands/2629/2009 (H1N1)pdm09, NR-19823.

4 Organism obtained through BEI Resources, NIAID, NIH: Influenza A Virus, A/Hong Kong/H090-761-V (0)/2009 (H1N1)pdm09, NR-44345. e Reported as Influenza A Equivocal or Influenza A (no subtype detected) at 3× LoD.

f Organism obtained through BEI Resources, NIAID, NIH: Influenza A Virus, A/Georgia/F32551/2012 (HIN1)pdm09, NR-42938.

8 Human isolates of recent swine variant H3N2v) were not available for testing, but reactivity near LoD is predicted by sequence analysis.

4 Genomic RNA obtained through the NIH Biodefense and Emerging Infections Resources Repository, NIAID, NIH: Genomic RNA from Influenza A Virus, A/Japan/305/1957 (H2N2), NR-2775.

{23}------------------------------------------------

1 Genomic RNA obtained through the NIH Biodefense and Emerging Infections Resources Repository, NIAID, NIH: Genomic RNA from Kilbourne F38: A/Korea/426/1968 (HA, NA) x A/Puerto Rico/8/1934 (H2N2), NR-9679.

J Isolate provided and tested by MRI Global, Kansas City, MO.

  • Genomic RNA obtained through the NIH Biodefense and Emerging Infections Research Resources Repository NIAID, NIH: Genomic RNA from Kilbourne F181: A/duck/Singapore/645/1997 (H5N3), Wild Type, NR-9682.

1 Genomic RNA obtained through the NIH Biodefense and Emerging Infections Resources Repository NIAID, NIH: Genomic RNA from Influenza A Virus, A/chicken/Germany/N/1949 (H10N7), NR-2765

Table 16. Influenza B Isolates Tested and Detected by FilmArray RP2

Table 17. Parainfluenza Virus Isolates Tested and Detected by FilmArray RP2

TypeSubtypeIsolate ID/Source[Strain/Location/Year]xLoDDetectedResult
1Zeptometrix 0810014CF-1xParainfluenza Virus1
ATCC VR-94[C-35/Washington DC/1957]3x
BEI NR-3226a[C39]3x
BEI NR-48680b[FRA/29221106/2009]3x
2Zeptometrix 0810015CF-1xParainfluenza Virus2
ATCC VR-92[Greer/Ohio/1955]3x
3Zeptometrix 0810016CF-1xParainfluenza Virus3
ATCC VR-93[C-243/Washington DC/1957]3x
BEI NR-3233c[NIH 47885, Wash/47885/57]3x
4AZeptometrix 0810060CF-1xParainfluenza Virus4
ATCC VR-1378[M-25/1958]3x
BZeptometrix 0810060BCF-3x
ATCC VR-1377[CH-19503/WashingtonDC/1962]3x

ª Discontinued part number.

  • Obtained through BEI Resources, NIAID, NIH: Human Parainfluenza Virus 1, HPV 1/FRA/29221106/2009, NR-48680.

º Obtained through BEI Resources, NIAID, NIH: Human Parainfluenza Virus 3, NIH 47885, NR-3233.

Table 18. Respiratory Syncytial Virus Isolates Tested and Detected by FilmArray RP2
TypeSource[Strain/Location/Year]xLoDDetectedResult
AZeptometrix 0810040ACF[2006]1x
AATCC VR-26[Long/Maryland/1956]3x
AATCC VR-1540[A2/Melbourne/1961]3xRespiratorySyncytial Virus
BZeptometrix 0810040CF[Ch-93 (18)-18]3x
BATCC VR-1400[WV/14617/1985]3x
BATCC VR-955[9320/Massachusetts/1977]3x
BATCC VR-1580[18537/Washington DC/1962]10x

{24}------------------------------------------------

SpeciesSource[Strain/Location/Year]xLoDDetectedResult
Bordetella parapertussisZeptometrix 0801461[A747]1x
Bordetella parapertussisZeptometrix 0801462[E595]3x
Bordetella parapertussisATCC 15237[NCTC 10853]3xBordetellaparapertussis(IS1001)
Bordetella parapertussisATCC 15311[NCTC 5952]3xBordetellaparapertussis(IS1001)
Bordetella parapertussisATCC BAA-587[12822/Germany/1993]3xBordetellaparapertussis(IS1001)
Bordetella bronchiseptica(containing IS1001)NRRL B-59909[MBORD849/Pig/Netherlands]3x

Table 19. Bordetella parapertussis (and Bordetella bronchiseptica) Isolates Tested and Detected by FilmArray RP2

2 Readivity with IS1001 sequences in B. bronchiseptica reactivity of the assay, however the analyte will be inacurately reported as B. parapertussis. The assay does not react with IS 1001-like sequences in B. holmesil (see Analytical Reactivity).

Table 20. Bordetella pertussis Isolates Tested and Detected by FilmArray RP2
Isolate ID/Source[Strain]xLoD DetectedResult
Zeptometrix 0801459[A639]1x
Zeptometrix 0801460[E431]3x
ATCC 8467[F]3x
ATCC 9340[5,17921]3x
ATCC 9797[18323/NCTC 10739]3xBordetella
ATCC 10380[10-536]3xpertussis (ptxP)
ATCC 51445[CNCTC Hp 12/63,623]3x
ATCC BAA-589[Tohama]3x
ATCC BAA-1335[MN2531]3x

Table 21. Chlamydia pneumoniae Isolates Tested and Detected by FilmArray RP2

Isolate ID/Source[Strain/Location/Year]xLoD DetectedResult
ATCC VR-2282[TW-183/Taiwan/1965]1xChlamydiapneumoniae
ATCC VR-1310[CWL-029]3x
ATCC VR-1360[CM-1/Georgia]3x
ATCC 53592[AR-39/Seattle/1983]3x
Table 22. Mycoplasma pneumoniae Isolates Tested and Detected by FilmArray RP2
--------------------------------------------------------------------------------------------
TypeIsolate ID/Source[Strain]xLoD DetectedResult
1Zeptometrix 0801579[M129]1xMycoplasmapneumoniae
ATCC 29342[M129-B7]3x
ATCC 29085[PI 1428]3x
2ATCC 15531[FH strain of Eaton Agent [NCTC 10119]3x
ATCC 15492[Mac]3x
unknownATCC 15293[M52]3x
ATCC 15377[Bru]3x
ATCC 39505[Mutant 22]3x
ATCC 49894[UTMB-10P]3x

Exclusivity

The potential for non-specific amplification and detection by the FilmArray RP2 assays was evaluated in silico by sequence analysis and testing of high concentrations of organisms in contrived samples. A total of 22 on-panel organisms were tested to assess the potential for intrapanel cross-reactivity and 50 off-panel organisms were tested to evaluate panel specificity. Offpanel organisms included normal respiratory flora and pathogens that may be present in NPS

{25}------------------------------------------------

specimens as well as near-neighbors or species genetically related to the organisms detected by the FilmArray RP2. The on-panel and off-panel organisms tested are shown in Table 23.

The final concentration of analyte in the sample (typically at least 1×106 CFU/mL for bacteria and fungi and at least 1×10 TCID<0/mL for viruses) represented levels ~70 - 400,000 fold higher than the LoD of the FilmArray RP2 assays. Some potential for intra-panel cross-reactivity with Bordetella species and Influenza A subtypes of swine origin was predicted by sequence analysis and observed when tested at high concentrations, as summarized in Table 24.

On-Panel VirusesOn-Panel Bacteria
AdenovirusHuman RhinovirusInfluenza BBordetella parapertussis
Coronavirus 229 EHuman MetapneumovirusParainfluenza Virus 1Bordetella pertussis
Coronavirus HKU1ªInfluenza A H1N1Parainfluenza Virus 2Chlamydia (Chlamydophila)pneumoniae
Coronavirus NL63Influenza A H3N2Parainfluenza Virus 3Mycoplasma pneumoniae
Coronavirus OC43Influenza A H1N1pdm09Parainfluenza Virus 4
Enterovirus (Echovirus 6)Influenza A Hsw1N1Respiratory Syncytial Virus
Off-Panel BacteriaOff-Panel Viruses
Acinetobacter calcoaceticusEnterobacter aerogenesNeisseria gonorrhoeaeBocavirus
Bordetella aviumEscherichia coliNeisseria meningitidisCytomegalovirus (CMV)
Bordetella bronchisepticaHaemophilus influenzaeProteus mirabilisEpstein-Barr Virus (EBV)
Bordetella hinziiKlebsiella oxytocaPseudomonas aeruginosaHerpes Simplex Virus 1
Bordetella holmesiiKlebsiella pneumoniaeSerratia marcescensMeasles Virus
Legionella bozemaniiLactobacillus acidophilusStaphylococcus aureusMumps
Legionella dumofiiLactobacillus plantarumStaphylococcus epidermidisMiddle East RespiratorySyndrome Coronavirusb
Legionella feeleiiMoraxella catarrhalisStenotrophomonas maltophilia
Legionella longbeacheaeMycoplasma genitaliumStreptococcus pneumoniaeOff-Panel Fungi/Yeast
Legionella micdadeiMycoplasma hominisStreptococcus agalactiaeCandida albicans
Legionella pneumophilaMycoplasma oraleStreptococcus pyogenesCryptococcus neoformans
Chlamydia trachomatisMycobacterium tuberculosisStreptococcus salivarusAspergillus fumigatus
Corynebacterium diphtheriaeNeisseria elongataUreaplasma urealyticumAspergillus flavus

Table 23. Organisms Tested for Evaluation of FilmArray RP2 Analytical Specificity Organisms with the potential for non-specific amplification by a FilmArray RP2 assay are shown in bold.

a Two different clinical specimens containing up to 8.9×10° RNA copies/mL of Coronavirus HKU1.

b Heat-inactivated viral culture obtained through BEI Resources, NIAID, NIH: Middle East Respiratory Syndrome Coronavirus (MERS-CoV), EMC/2012, Heat-Inactivated, NR-50171.

{26}------------------------------------------------

Cross-reactive Organism(s)FilmArrayRP2 ResultDescription
Non-pertussis Bordetella species(e.g., Bordetella parapertussisBordetella bronchisepticaa)Bordetella pertussis (ptxP) b,cThe Bordetella pertussis (ptxP) assay can amplifypertussis toxin pseudogene sequences in B.bronchiseptica and B. parapertussis primarilywhen present at high concentration ((≥1.2E+09CFU/mL).
Bordetella bronchisepticaa(with IS1001 sequences)Bordetella parapertussis (IS1001)Some strains of B. bronchiseptica carry IS1001insertion sequences identical to those carried byB. parapertussis. These sequences will beefficiently amplified by the IS1001 assay andreported by FilmArray RP2 as Bordetellaparapertussis (IS1001).
Bordetella pertussisBordetella parapertussisBordetella bronchisepticaHuman Rhinovirus/Enterovirusd,eThe Human Rhinovirus/Enterovirus assay mayamplify off-target sequences found in strains of B.pertussis, B. bronchiseptica, and B. parapertussiswhen present at high concentration. Cross-reactivitywith B. pertussis was observed at a concentration of4.5E+07 CFU/mL or higher.
Influenza A H1N1(swine origin)Influenza A H1-2009fThe Influenza A H1-2009 assay may react withH1 hemagglutinin gene sequences from viruses ofswine origin.FilmArray RP2 will report either Influenza A H1or Influenza A H1-2009, depending on the strainand concentration in the samplef.

Table 24. Predicted and Observed Cross-Reactivity of the FilmArray RP2

a B. bronchiseptica infection is rare in humans and more common in domesticated animals ('kennel cough').

b Cross-reactivity was observed only when tested at a high concentration (≥1.2E+09 CFU/mL).

6 Cross-reactivity between the Bordetella pertussis (ptxP) assay and B. parapertussis will be reported as a co-detection (Bordetella parapertussis (IS/001) Detected and Bordetella pertussis (pt.rP) Detected); while cross-reactivity with most strains of B. bronchissprice (that do not carry IS/00/) will be reported only as Bordetella pertussis (ptxP) Detected.

4 Cross-reactivity with B. pertussis was observed when tested at a concentration of 4.5E+07 CFU/mL and higher. Cross-reactivity with B. parapertussis and B. bronchiseptica is predicted based on in silico analysis, but was not observed when tested at a concentration of 1.2E+09 CFU/mL.

6 Cross-reactivity between the Human Rhinovirus assays and B. pertussis will be reported as a cedection (Bordetella pertussis (pt.P) Detected and Human Rhinovirus/Enterovins Detected parapertussis (15/00/) Detected and Human Rhinovirus Detected); while cross-reactivity with most strains of B. bronchiseptical (that do not carry IS/100) will be reported (falsely) only as Human Rhinovirus/Enterovirus Detected.

' Swine origin Hsw1N1 (A/New Jersey/8/1976 : ATCC VR-897) was detected as either Influenza A H1-2009 at a concentration of 8.9E+06 CEID50/mL.

Interference

Potentially interfering substances that could be present in NPS specimens or introduced during specimen collection and testing were evaluated for their effect on FilmArray RP2 performance. Substances included endogenous substances that may be found in specimens at normal or elevated levels (e.g. blood, mucus/mucin, human genomic DNA), various commensal or infectious microorganisms, medications, washes or topical applications for the nasal passage, various swabs and transport media for specimen collection, and substances used to clean, decontaminate, or disinfect work areas.

Each substance was added to contrived samples containing representative organisms at concentrations near (2-3×) LoD. The concentration of substance added to the samples (Table 25) was equal to or greater than the highest level expected to be in NPS specimens.

None of the substances were shown to interfere with the FilmArray RP2 function. However, it was observed that exposure of samples to bleach prior to testing could damage the

{27}------------------------------------------------

organisms/nucleic acids in the sample, leading to inaccurate FilmArray RP2 test results (lack of analyte detection). The effect of bleach was dependent on the concentration and/or length of time the bleach was allowed to interact with the sample.

Substance TestedConcentration TestedResult
Endogenous Substances
Human Whole Blood10% v/vNo Interference
Human Mucus (Sputum)1 swab/mL sampleNo Interference
Human Genomic DNA20 ng/µLNo Interference
Competitive Microorganisms
Coronavirus 229E1.7E+04 TCID50/mLNo Interference
Adenovirus A128.9E+05 TCID50/mLNo Interference
Parainfluenza Virus 36.6E+05 TCID50/mLNo Interference
Bordetella pertussis5.8E+08 CFU/mLNo Interference
Enterovirus D681.6E+07 TCID50/mLNo Interference
Echovirus 61.0E+07 TCID50/mLNo Interference
Respiratory Syncytial Virus4.2E+04 TCID50/mLNo Interference
Staphylococcus aureus2.5E+07 CFU/mLNo Interference
Streptococcus pneumoniae1.7E+07 CFU/mLNo Interference
Haemophilus influenzae6.2E+07 CFU/mLNo Interference
Candida albicans1.0E+06 CFU/mLNo Interference
Herpes Simplex Virus 11.6E+06 TCID50/mLNo Interference
Cytomegalovirus1.2E+06 TCID50/mLNo Interference
Exogenous Substancesa
Tobramycin (systemic antibiotic)0.6 mg/mLNo Interference
Mupirocin(active ingredient in anti-bacterial ointment)2% w/vNo Interference
Saline Nasal Spray with Preservatives(0.65% NaCl, Phenylcarbinol, Benzalkonium chloride)1% v/vNo Interference
Nasal Decongestant Spray(Oxymetazoline HCl 0.05%, Benzalkonium chloride, phosphate)1% v/vNo Interference
Analgesic ointment (Vicks®VapoRub®)1% w/vNo Interference
Petroleum Jelly (Vaseline®)1% w/vNo Interference
Snuff (Tobacco)1% w/vNo Interference
Disinfecting/Cleaning Substances
Bleach1% and 2% v/v[up to 1024 ppm chlorine]Interferenceb
Disinfecting wipes (ammonium chloride)½ in²No Interference
Ethanol7% v/vNo Interference
DNAZap (Ambion™ AM9891G & AM9892G)1% v/vNo Interference
RNaseZap (Ambion™ AM9782)1% v/vNo Interference
Specimen Collection Materials
Rayon Swabs (Copan 168C)N/ANo Interference
Nylon Flocked Swabs (Copan 553C)N/ANo Interference
Polyester Swabs (Copan 175KS01)N/ANo Interference
Calcium Alginate Swabs (Puritan 25-801 A 50)N/ANo Interference
Substance TestedConcentration TestedResult
M4® Transport Medium (Remel)100%No Interference
M4-RT® Transport Medium (Remel)100%No Interference
M5® Transport Medium (Remel)100%No Interference
M6™ Transport Medium (Remel)100%No Interference
Universal Viral Transport vial (BD)100%No Interference
Sigma-Virocult™ Viral Collection and Transport System(Swab and Transport Medium)100%No Interference
Copan ESwab™ Sample Collection and Delivery System(Swab and Liquid Amies Medium)100%No Interference

Table 25. Evaluation of Potentially Interfering Substances on the FilmArray RP2

{28}------------------------------------------------

4 Nasal influenza vaccines (e.g. FluMis) were not evaluated to be reactive with the FilmArray RP2 Influenza A (subtype) and Influenza B assays.

b Not Detected results were reported for several analytes after incubation of the sample with 2% bleach for 10 minutes or overnight. It was concluded that interference resulted primarily from danage to the sample, rather than inhibition or interference with pouch function(s).

Reproducibility

Reproducibility testing of contrived samples was performed at three test sites on a combination of FilmArray 2.0 and FilmArray Torch systems. The study incorporated a range of potential variation introduced by site, day, operator (at least two per site), system, instrument or Torch module (at least three per site/sample), and pouch lot (at least three). The samples contained various combinations of twelve different FilmArray RP2 analytes, each at three different concentrations (Negative, Low Positive (1×LoD), and Moderate Positive (3×LoD)). Frozen samples were repeatedly tested on five different days for 120 data points per sample (60 per system) over 480 total valid runs.

A summary of results (percent (%) agreement with the expected Detected or Not Detected result) for each analyte (by site and system) is provided in Table 26.

Agreement with Expected Result
AnalyteConcentrationTestedExpectedResultFilmArray TorchSite AFilmArray TorchSite CFilmArray TorchSystem Sub-TotalFilmArray 2.0Site BFilmArray 2.0Site CFilmArray 2.0SystemSub-TotalAll Sites/Systems(95% CI)
Viruses
AdenovirusC2ATCC VR-846Moderate Positive3× LoD6.0E+00 TCID50/mL(1.1E+02 copies/mL)Detected30/30100%29/3096.7%59/6098.3%29/3096.7%30/30100%59/6098.3%118/12098.3%(94.1%-99.8%)
AdenovirusC2ATCC VR-846Low Positive1× LoD2.1E+00 TCID50/mL(3.7E+01 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%29/3096.7%59/6098.3%119/12099.2%(95.4%-100%)
AdenovirusC2ATCC VR-846None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Coronavirus 229ENone(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
Coronavirus HKU1None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
AnalyteConcentration TestedExpected ResultFilmArray TorchFilmArray 2.0All Sites/Systems (95% CI)
Site ASite CSystem Sub-TotalSite BSite CSystem Sub-Total
Coronavirus OC43ATCC VR-759Moderate Positive3x LoD9.0E+01 TCID50/mL(1.7E+03 copies/mL)Detected29/3096.7%29/3096.7%58/6096.7%29/3096.7%30/30100%59/6098.3%117/12097.5%(92.9%-99.5%)
Low Positive1x LoD3.0E+01 TCID50/mL(5.6E+02 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%27/3090.0%57/6095.0%117/12097.5%(92.9%-99.5%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Coronavirus NL63None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
HumanMetapneumovirusType 16, A1IA10-2003Zeptometrix0810161CFModerate Positive3x LoD3.0E+01 TCID50/mL(3.6E+03 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD1.0E+01 TCID50/mL(1.2E+03 copies/mL)Detected30/30100%30/30100%60/60100%28/3093.3%30/30100%58/6096.7%118/12098.3%(94.1%-99.8%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
HumanRhinovirus/EnterovirusRhinovirus 1AZeptometrix0810012CFNModerate Positive3x LoD3.0E-01 TCID50/mL(1.1E+02 copies/mL)Detected30/30100%30/30100%60/60100%28/3093.3%30/30100%58/6096.7%118/12098.3%(94.1%-99.8%)
Low Positive1x LoD1.0E-01 TCID50/mL(3.8E+01 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Influenza A H3Influenza A H3N2A/PortChalmers/1/73ATCC VR-810Moderate Positive3x LoD3.0E-01 TCID50/mL(6.3E+01 copies/mL)Detected30/30100%30/30100%60/60100%29/3096.7%30/30100%59/6098.3%119/12099.2%(95.4%-100%)
Low Positive1x LoD1.0E-01 TCID50/mL(2.1E+01 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Influenza AH1-2009None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
ExpectedResultAgreement with Expected ResultAll Sites/Systems(95% CI)
AnalyteConcentrationTestedFilmArray TorchFilmArray 2.0
Site ASite CSystem Sub-TotalSite BSite CSystemSub-Total
Influenza A H1None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
Influenza BB/FL/04/06Zeptometrix0810255CFModerate Positive3x LoD1.5E+01 TCID50/mL(1.0E+02 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD5.0E+00 TCID50/mL(3.4E+01 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Parainfluenza Virus1None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
Parainfluenza Virus2Type 2Zeptometrix0810015CFModerate Positive3x LoD1.5E+00 TCID50/mL(9.0E+01 copies/mL)Detected30/30100%29/3096.7%59/6098.3%29/3096.7%30/30100%59/6098.3%118/12098.3%(94.1%-99.8%)
Low Positive1x LoD5.0E-01 TCID50/mL(3.0E+01 copies/mL)Detected30/30100%29/3096.7%59/6098.3%30/30100%27/3090.0%57/6095.0%116/12096.7%(91.7%-99.1%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Parainfluenza Virus3None(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
Parainfluenza Virus4Type 4aZeptometrix0810060CFModerate Positive3x LoD1.5E+02 TCID50/mL(4.8E+03 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD5.0E+01 TCID50/mL(1.6E+03 copies/mL)Detected30/30100%29/3096.7%59/6098.3%29/3096.7%30/30100%59/6098.3%118/12098.3%(94.1%-99.8%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
RespiratorySyncytial VirusType AZeptometrix0810040ACFModerate Positive3x LoD6.0E-02 TCID50/mL(2.7E+01 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD2.0E-02 TCID50/mL(9.0E+00 copies/mL)Detected29/3096.7%30/30100%59/6098.3%30/30100%29/3096.7%59/6098.3%118/12098.3%(94.1%-99.8%)
AnalyteConcentrationTestedExpectedResultFilmArray TorchFilmArray 2.0All Sites/Systems(95% CI)
Site ASite CSystem Sub-TotalSite BSite CSystemSub-Total
None(no analyte)None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Bacteria
Bordetellaparapertussis(IS1001)A747Zeptometrix0801461Moderate Positive3x LoD1.8E+02 IS1001copies/mL(1.2E+02 CFU/mL)Detected30/30100%30/30100%60/60100%29/3096.7%30/30100%59/6098.3%119/12099.2%(95.4%-100%)
Low Positive1x LoD6.0E+01 IS1001copies/mL (4.1E+01CFU/mL)Detected24/30a80.0%29/3096.7%53/60188.3%29/3096.7%30/30100%59/6098.3%112/12093.3%(87.3%-97.1%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Bordetella pertussis(ptxP)A639Zeptometrix0801459Moderate Positive3x LoD3.0E+03 CFU/mLDetected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD1.0E+03 CFU/mLDetected28/3093.3%30/30100%58/6096.7%30/30100%30/30100%60/60100%118/12098.3%(94.1%-99.8%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
Chlamydia(Chlamydophila)pneumoniaeTW183ATCC VR-2282Moderate Positive3x LoD3.0E-01 TCID50/mL(2.0E+02 copies/mL)Detected30/30100%30/30100%60/60100%30/30100%30/30100%60/60100%120/120100%(97.0%-100%)
Low Positive1x LoD1.0E-01 TCID50/mL(6.6E+01 copies/ml)Detected28/3093.3%30/30100%58/6096.7%29/3096.7%30/30100%59/6098.3%117/12097.5%(92.9%-99.5%)
None(no analyte)NotDetected60/60100%60/60100%120/120100%60/60100%60/60100%120/120100%240/240100%(98.5%-100%)
MycoplasmapneumoniaeNone(no analyte)NotDetected120/120100%120/120100%240/240100%120/120100%120/120100%240/240100%480/480100%(99.2%-100%)
Overall Agreement with the Expected ResultAll Analytes and All Test Levels(95% Confidence Interval)9,562/9,60099.6%(99.5% - 99.7%)

Table 26. Reproducibility of FilmArray RP2 Results on FilmArray Torch and FilmArray 2.0 Systems

{29}------------------------------------------------

{30}------------------------------------------------

{31}------------------------------------------------

  • Data from Site A were further reviewed by the uniables including test day, Toch module, and operator. No correlation could be found between the Not Detected results and any one of these variables. The Not Detected results at Site A were found to be statistically non-significant (p>0.05 by Fisher's exact test) and therefore do not increatent variance in precision of the FilmArray RP2 Bordeella parapertusis (IS1001) results.

The reproducibility (standard deviation) of melting temperature (Tm) for the amplification products generated by the FilmArray RP2 was also evaluated, with a Tm standard deviation for

{32}------------------------------------------------

each assay of ± 0.5℃ or less observed within and between the FilmArray 2.0 and FilmArray Torch systems (Table 27).

Site ASite CSite BSite CAll Sites/Systems
AnalyteAssayMeanStDevMeanStDevMeanStDevMeanStDevMeanStDev
ControlsyeastRNA82.3± 0.382.1± 0.282.2± 0.382.0± 0.282.1± 0.3
PCR276.1± 0.275.9± 0.276.0± 0.275.8± 0.275.9± 0.2
VIRUSES
Adenovirus C2Adeno289.0± 0.288.8± 0.188.8± 0.388.7± 0.288.8± 0.2
(ATCC VR-846)Adeno689.6± 0.289.4± 0.289.4± 0.389.2± 0.289.4± 0.3
Coronavirus OC43(ATCC VR-759)CoV-OC43-280.7± 0.280.6± 0.180.6± 0.380.5± 0.280.6± 0.2
Human Metapneumovirus(Zeptometrix 0810161CF)hMPV78.2± 0.378.0± 0.278.0± 0.377.8± 0.278.0± 0.3
Rhinovirus(Zeptometrix 0810012CFN)HRV/EV84.3± 0.284.2± 0.284.3± 0.384.1± 0.284.2± 0.2
Influenza A H3N2FluA-pan184.2± 0.284.0± 0.184.0± 0.383.8± 0.284.0± 0.2
(ATCC VR-810)FluA-pan278.9± 0.278.9± 0.178.9± 0.278.8± 0.278.9± 0.2
FluA-H382.1± 0.281.9± 0.282.0± 0.381.9± 0.282.0± 0.2
Influenza B(Zeptometrix 0810255CF)FluB80.4± 0.380.3± 0.280.4± 0.280.2± 0.280.3± 0.2
Parainfluenza virus 2(Zeptometrix 0810015CF)PIV283.2± 0.283.1± 0.283.2± 0.283.0± 0.283.1± 0.2
Parainfluenza virus 4(Zeptometrix 0810060CF)PIV477.1± 0.277.0± 0.377.2± 0.377.0± 0.277.1± 0.3
Respiratory Syncytial Virus(Zeptometrix 0810040ACF)RSV81.2± 0.281.1± 0.281.1± 0.281.0± 0.281.1± 0.2
BACTERIA
Bordetella parapertussis(Zeptometrix 0801461)IS 100187.7± 0.287.6± 0.287.6± 0.387.5± 0.287.6± 0.2
Bordetella pertussis(Zeptometrix 0801459)ptxP88.6± 0.288.5± 0.288.5± 0.388.2± 0.288.4± 0.3
Chlamydia pneumoniae(ATCC VR-2282)Cpne79.6± 0.379.5± 0.279.5± 0.379.3± 0.279.5± 0.3

Table 27. Reproducibility of Tm (°C) For Select FilmArray Torch and FilmArray Torch and FilmArray 2.0 Systems Mean Tm values are calculated from a combination of Tm values obtained at the 3x LoD and 1x LoD concentrations.

4 Calculated from results at the 3× LoD and 1× LoD test concentrations combined.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.