K Number
K170491
Manufacturer
Date Cleared
2017-05-11

(83 days)

Product Code
Regulation Number
866.3130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® C. difficile Assay is an in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile Toxin A gene (tcdA) in unformed stool specients of patients suspected of having Clostridium difficile-infection (CDI). The Solana C. difficile Assay is intended for use as an aid in diagnosis of CDI. The assay utilizes helicase-dependent amplification (HDA) for the amplification of a highly conserved fragment of the Toxin A gene sequence. The Solana C. difficile Assay is intended for use only with the Solana® instrument.

Device Description

The Solana C. difficile Assay combines simple sample processing and Helicase-Dependent Amplification (HDA) performed in Solana for the detection of toxigenic Clostridium difficile directly from CDI-suspected diarrheal specimens.

A small amount of specimen is transferred to a Lysis Tube using a swab. The Lysis Tube is then subjected to heat-treatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of target sequence. In Solana, the target sequence is amplified by specific primers and detected by a specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC is amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then report the test results to the user on its display screen, and it can print out the results via a printer.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Solana C. difficile Assay, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a separate section with predefined thresholds. However, the "Performance in Comparison to Enhanced toxigenic bacterial culture" section for sensitivity and specificity can be considered as the primary clinical performance evaluation against a recognized gold standard. The "Performance in Comparison to FDA-cleared molecular device" provides comparative performance.

Given this, I will present the clinical performance metrics as the reported device performance against what would typically be considered implicit acceptance targets for diagnostic devices of this type.

Table 1: Reported Device Performance Against Ground Truth (Enhanced Toxigenic Culture)

Performance MetricReported Performance (95% CI)
Sensitivity93.0% (86.9% - 96.4%)
Specificity99.2% (98.2% - 99.6%)

Table 2: Reported Device Performance Compared to an FDA-Cleared Molecular Device

Performance MetricReported Performance (95% CI)
Positive Percent Agreement97.0% (91.59% - 99.0%)
Negative Percent Agreement97.9% (96.6% - 98.7%)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 854 raw specimens were collected. After removing 2 invalid specimens, 852 specimens were used for analysis.
  • Data Provenance:
    • Country of Origin: United States. The study was conducted at "three distinct geographical sites across the United States."
    • Retrospective or Prospective: Prospective. The study documents state, "Performance characteristics of the Solana C. difficile Assay were established during a prospective study conducted November 2016 to February 2017."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts or their qualifications for establishing the ground truth (enhanced toxigenic culture). It describes the methodology for the enhanced toxigenic culture, which is a laboratory-based process, rather than an expert human interpretation.

4. Adjudication Method for the Test Set

The document does not explicitly detail an adjudication method beyond the methodology for the enhanced toxigenic culture. It mentions, "Three (3) of the six (6) specimens were positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device, three (3) were negative." and "Six (6) of eight (8) specimens were found positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device., and two (2) were negative", which suggests some level of discordant analysis against a second method, but no formal expert adjudication process is described for the ground truth establishment itself.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic test that provides a qualitative result (positive/negative) directly from the instrument. It is not designed to be interpreted by human readers, nor does it assist human readers in making a diagnosis. Therefore, the concept of human readers improving with or without AI assistance does not apply here.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire clinical performance evaluation described compares the "Solana C. difficile Assay" (the algorithm/device) directly against the enhanced toxigenic culture (ground truth) and an FDA-cleared molecular device. The results reported are for the device's performance without human intervention in the interpretation of the C. difficile Toxin A gene (tcdA) detection.

7. Type of Ground Truth Used

The primary ground truth used for evaluating the Solana C. difficile Assay was enhanced toxigenic bacterial culture.

An "alternate FDA cleared molecular device" was also used for comparison in cases of discordance, but the initial sensitivity and specificity calculations used the enhanced toxigenic culture as the reference.

8. Sample Size for the Training Set

The document does not provide information regarding a specific training set size or methodology for the Solana C. difficile Assay. The information provided focuses on the analytical and clinical validation of the final product. For an IVD like this, the "training" may happen during internal development and optimization phases, which are not typically detailed in 510(k) summaries.

9. How the Ground Truth for the Training Set Was Established

As no specific training set or its sample size is mentioned, the method for establishing ground truth for a training set (if one existed separately from analytical validation) is not described in the provided document. The ground truth for the clinical test set was established by enhanced toxigenic bacterial culture.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.

May 11, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Quidel Corporation Ronald Lollar Sr. Director, Clinical, Regulatory, Scientific Affairs 2005 East State Street. Suite 100 Athens. Ohio 45701

Re: K170491

Trade/Device Name: Solana C. difficile Assay Regulation Number: 21 CFR 866.3130 Regulation Name: Clostridium difficile toxin gene amplification assay Regulatory Class: Class II Product Code: OZN Dated: February 16, 2017 Received: February 17, 2017

Dear Ronald Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Image /page/1/Picture/9 description: The image shows the name "Steven R. Gitterman -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or presentation. The background is plain, ensuring the text is easily readable.

for Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170491

Device Name Solana® C. difficile Assay

Indications for Use (Describe)

The Solana® C. difficile Assay is an in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile Toxin A gene (tcdA) in unformed stool specients of patients suspected of having Clostridium difficile-infection (CDI). The Solana C. difficile Assay is intended for use as an aid in diagnosis of CDI. The assay utilizes helicase-dependent amplification (HDA) for the amplification of a highly conserved fragment of the Toxin A gene sequence. The Solana C. difficile Assay is intended for use only with the Solana® instrument.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical. Regulatory, Scientific Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 – Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

May 10, 2017

A. 510(k) Number:

K170491

B. Purpose for Submission:

To obtain substantial equivalence for the Solana® C. difficile Assay when performed on the Solana® instrument

C. Measurand:

A highly conserved region in the 5' end of the tcdA gene

D. Type of Test:

Helicase-dependent amplification (HDA)

E. Applicant:

{4}------------------------------------------------

Quidel Corporation

F. Proprietary and Established Names:

Solana® C. difficile Assay

G. Regulatory Information:

Table 1. Regulatory Information
Product CodeClassificationRegulation SectionPanel
OZN - Amplificationassay for thedetection ofClostridium difficiletoxin genes fromstool specimens ofsymptomaticpatientsClass II (SpecialControls)21 CFR 866.3130 - C. difficile NucleicAcid Amplification Test AssayMicrobiology (83)

H. Intended Use:

1. Intended Use(s):

The Solana® C. difficile Assay is an in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile Toxin A gene (tcdA) in unformed stool specimens of patients suspected of having Clostridium difficile-infection (CDI). The Solana C. difficile Assay is intended for use as an aid in diagnosis of CDI. The assay utilizes helicase-dependent amplification (HDA) for the amplification of a highly conserved fragment of the Toxin A gene sequence. The Solana C. difficile Assay is intended for use only with the Solana® instrument.

    1. Indication(s) for Use:
      Same as intended Use
    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only
    • For prescription use only

{5}------------------------------------------------

4. Special instrument requirements:

Solana® instrument

l. Device Description:

The Solana C. difficile Assay combines simple sample processing and Helicase-Dependent Amplification (HDA) performed in Solana for the detection of toxigenic Clostridium difficile directly from CDI-suspected diarrheal specimens.

A small amount of specimen is transferred to a Lysis Tube using a swab. The Lysis Tube is then subjected to heat-treatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of target sequence. In Solana, the target sequence is amplified by specific primers and detected by a specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC is amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then report the test results to the user on its display screen, and it can print out the results via a printer.

Materials Provided:

Cat. #M307 48 Tests per Kit

Table 2. Kit Components
ComponentQuantityStorage
Neonatal flocked swabs48 tubes/kit2°C to 30°C
Lysis Buffer48 tubes/kit 1.0 mL2°C to 8°C
Dilution Buffer48 tubes/kit 1.8 mL2°C to 8°C
Reaction Tubes48 tubes/kit2°C to 8°C

Materials required but not provided:

{6}------------------------------------------------

510(k) Summary

  • External controls for C. difficile (e.g. Quidel Molecular C. difficile Control Set, which contains positive and negative controls, serves as an external processing and extraction control)
  • Sterile DNase-free filter-blocked or positive displacement micropipettor tips
  • Micropipettor
  • Stopwatch or timer
  • Scissors or a blade ●
  • Heat block capable of 95° C ± 2° C temperature ●
  • Solana workflow tray and transfer rack
  • Solana Instrument
  • Thermometer

Substantial Equivalence Information: J.

    1. Predicate device name(s):
      Portrait Toxigenic C. difficile Assay (Great Basin Scientific)
    1. Predicate 510(k) number(s): K113358 (DEN120013)

3. Comparison with predicate:

Table 3.      Similarities
ItemSubject DeviceSolana® C. difficile AssayPredicate DeviceGreat Basin Scientific PortraitToxigenic C. Difficile AssayK113358 (DEN120013)
Intended UseThe Solana® C. difficile Assay is an in vitrodiagnostic test for the direct, qualitativedetection of the Clostridium difficileToxin A gene ( tcdA ) in unformed stoolspecimens of patients suspected ofhaving Clostridium difficile -infection(CDI). The Solana C. difficile Assay isintended for use as an aid in diagnosis ofCDI. The assay utilizes helicase-dependent amplification (HDA) for theamplification of a highly conservedPortrait Toxigenic C. difficileAssay, a prescription deviceunder 21 CFR Part 801.109that is indicated for thedetection of toxigenicClostridium difficile in humanfecal samples collected frompatients suspected of havingClostridium difficile infection
Table 3. Similarities
ItemSubject DeviceSolana® C. difficile AssayPredicate DeviceGreat Basin Scientific PortraitToxigenic C. Difficile AssayK113358 (DEN120013)
fragment of the Toxin A gene sequence.The Solana C. difficile Assay is intendedfor use only with the Solana® instrument.(CDI). The test utilizesautomated blocked primerenabled helicase-dependentamplification (bpHDA) todetect toxin gene sequencesassociated with toxinproducing C. difficile. ThePortrait Toxigenic C. difficileAssay is intended as an aid inthe diagnosis of CDI.
SpecimenUnformed stoolSame
TechnologicalprincipleNucleic acid amplificationSame
Assay techniqueIsothermal, helicase-dependentnucleic acid amplificationSame
Assay ResultsQualitativeSame
Extraction MethodsNot requiredSame
Detection methodAutomatedSame

{7}------------------------------------------------

510(k) Summary

  • Table 4. Differences

{8}------------------------------------------------

Subject Device Predicate Device ltem Solana C. difficile Assay Great Basin Scientific Portrait Toxigenic C. Difficile Assay K113358 Target Toxin A gene (tcdA) Toxin B gene (tcdB) Portrait Analyzer Amplification and Solana instrument Detection instruments

510(k) Summary

K. Standard/Guidance Document Referenced (if applicable):

Class II Special Controls Guideline Document: Toxin Gene Amplification Assays for the Detection of Clostridium difficile

L. Test Principle:

A small amount of specimen is transferred to a Lysis Tube using a swab. The Lysis Tube is then subjected to heat-treatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of target sequence. In Solana, the target sequence is amplified by specific primers and detected by a specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC is amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then report the test results to the user on its display screen, and it can print out the results via a printer.

M. Performance Characteristics:

    1. Analytical performance:

{9}------------------------------------------------

510(k) Summary

a. Precision/Reproducibility:

Reproducibility

The reproducibility of the Solana C. difficile Assay was evaluated at three laboratory sites. A blinded and randomized study panel containing Clostridium difficile negative and positive samples were tested at three (3) test sites (two (2) clinical sites). Each site tested a reproducibility panel and Assay Controls for five (5) days in triplicate. Testing was done by two operators at each site. Each operator ran the panel once a day. A total of 540 specimens were tested (including controls).

Table 5. Reproducibility Summary
CategorySITESITESITEOverall PercentAgreement95% ConfidenceInterval
Site #1Site #2Site #3
#expectedresults/#tested%Agreement#expectedresults/#tested%Agreement#expectedresults/#tested%Agreement
C. difficile HighNegative(4.8 X $10^2$CFU/mL)12/3040%19/3063.3%12/3040%43/9047.8%37.8% - 58.0%
C. difficile LowPositive(1.7 X $10^3$CFU/mL)30/30100%30/30100%29/3096.7%89/9098.9%94% - 99.8%
C. difficileModeratePositive(3.4 X $10^3$CFU/mL)30/30100%30/30100%30/30100%90/90100%95% - 100%
Negative30/30100%30/30100%30/30100%90/90100%95% - 100%
C. difficile PositiveControl30/30100%30/30100%30/30100%90/90100%95% - 100%
Assay NegativeControl30/30100%30/30100%30/30100%90/90100%95% - 100%

The results suggest that there are no significant differences between different users using different instruments at different sites on different days.

{10}------------------------------------------------

  • b. Linearity/assay reportable range:
    Not applicable - This assay is qualitative.

  • ﻥ Traceability, Stability, Expected values (controls, calibrators, or methods):
    Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

An analytical study was conducted to determine the potential specimen storage capabilities of the Solana® C. difficile Assay by testing contrived stool samples that have been stored to mimic what may happen in a clinical setting. Specimen stability was also assessed in the prospective clinical study that tested 854 specimens that were stored for up to 3 days at 2° to 8°C.The results of clinical and analytical studies supported the storage of stool specimens at 2 to 8°C for up to 3 days before testing.

Controls:

  • . External controls for C. difficile (e.g. Quidel Molecular C. difficile Control Set, which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana C. difficile Assay each day of testing. The external controls produced the expected results in all analytical and clinical testing.
  • . The process control monitored sample processing, HDA inhibitory specimens and confirmation of the integrity of assay reagents and Solana instrument functionality throughout the clinical and analytical testing. The process control identified two (2) specimens in the clinical testing as invalid (2/854, 0.2%). The process control was detected in all analytical testing.
  • d. Detection limit:

The analytical sensitivity (limit of detection or LOD) of the Solana C. difficile Assay was determined using serial dilutions of two (2) toxigenic C. difficile strains, ATCC BAA-1805 and CCUG 20309 spiked in negative matrix, and also quantified C. difficile genomic DNA, BAA-1382DQ™ spiked in lysis buffer. Analytical sensitivity (LOD) is defined as the lowest concentration at which 95% of all replicates tested positive.

{11}------------------------------------------------

Table 6. LOD Values
Stool MatrixC. difficile StrainsStrain LOD
Unpreserved StoolATCC BAA-18059.13E+03 CFU/mL
Unpreserved StoolCCUG 203094.90E+03 CFU/mL
N/AGenomic DNA: ATCC®BAA-1382DQ™15 copy/assay

e. Analytical specificity:

Cross Reactivity and Microbial Interference:

The analytical specificity of the Solana C. difficile Assay was evaluated by testing a panel consisting of sixty-eight (68) bacterial, viral and yeast microorganisms and human DNA representing common enteric pathogens, flora or nucleic acid commonly present in the intestine. Microorganisms or nucleic acid was mixed with pooled negative matrix and tested directly or in the presence of 1.83E+04 CFU/mL of C. difficile for cross-reactivity and microbial interference, respectively.

The table below lists the bacterial, viral and yeast microorganisms used in these studies. There was no evidence of cross reactivity or interference with any of the panel members and the Solana C. difficile Assay.

Microorganism
Abiotrophia defectiveATCC 49176
Acinetobacter baumanniiATCC 19606
Aeromonas hydrophilaATCC 7966
Alcaligenes faecalis subsp. faecalisATCC 15554
Bacillus cereusATCC 13472
Bacteroides fragilisATCC 25285
Campylobacter coliATCC 43479
Campylobacter jejuni sub sp .jejuniATCC 33292
Candida albicansATCC 10231
Citrobacter freundiiATCC 8090
Clostridium bifermentansATCC 638
Clostridium botulinumIn silico analysis
Clostridium butyricumCCRI-11128
Microorganism
Clostridium haemolyticumATCC 19398
Clostridium novyiATCC 19402
Clostridium orbiscindensATCC 49531
Clostridium perfringensATCC 13124
Clostridium scindensATCC 35704
Clostridium septicumATCC 12464
Clostridium sordelliiATCC 9714
Clostridium sordelliiZ077
Clostridium sordelliiCCUG 6329
Clostridium sordelliiCCUG 9284
Clostridium sordelliiCCUG 33098
Clostridium sordelliiCCUG 36938
Clostridium sordelliiCCUG 43123
Clostridium sordelliiCCUG 47545
Clostridium sordelliiCCUG 59819
Clostridium difficile (non-toxigenic)ATCC 43593
Clostridium difficile (non-toxigenic)ATCC 43601
Clostridium sporogenesATCC 15579
Edwardsiella tardaATCC 15947
Enterobacter aerogenesATCC 13048
Enterobacter cloacaeATCC 13047
Enterococcus faecalis vanBATCC 51299
Escherichia coliATCC 23511
Escherichia coli O157:H7ATCC 700927
Helicobacter pyloriATCC 43504
Klebsiella oxytocaATCC 33497
Lactobacillus acidophilusATCC 4356
Listeria monocytogenesATCC BAA-389
Peptostreptococcus anaerobiusATCC 27337
Plesiomonas shigelloidesATCC 14029
Porphyromonas asaccharolyticaATCC 25260
Prevotella melaninogenicaATCC 25845
Proteus mirabilisATCC 25933
Providencia alcalifaciensATCC 9886
Pseudomonas aeruginosaATCC 35554
Salmonella choleraesuis (typhimurium)ATCC 14028
Salmonella enterica subsp. arizonaeATCC 13314
Salmonella enterica subsp. entericaATCC 7001
Microorganism
Serratia liquefaciensATCC 27592
Serratia marcescensATCC 13880
Shigella boydiiATCC 9207
Shigella dysenteriaeATCC 11835
Shigella sonneiATCC 29930
Staphylococcus aureusATCC 43300
Staphylococcus epidermidisATCC 14990
Streptococcus agalactiaeATCC 12973
Vibrio parahaemolyticusATCC 17802
Adenovirus
Rotavirus
Norovirus
Enterovirus
Echovirus
Coxsackie virus
Cytomegalovirus
Human DNA

{12}------------------------------------------------

{13}------------------------------------------------

Interference:

The performance of Solana C. difficile Assay was evaluated with potentially interfering substances that may be present in stool specimens. A panel composed of thirty-two (32) substances was tested in the absence or presence of C. difficile at 1.83E+04 CFU/mL in the Solana C. difficile Assay. There was no evidence of interference caused by the substances tested at the concentrations shown below.

Substance NameActive IngredientsTest Concentration
NystatinNystatin1% (w/v)
Cortizone 10Hydrocortisone1% (w/v)
Fleet Glycerin SuppositoriesGlycerin1% (w/v)
DesitinZinc Oxide1% (w/v)
Anusol Pluspramoxine hydrochloride and zincsulfate monohydrate1% (w/v)
Preparation HPhenylephrine1% (w/v)
NystatinNystatin1% (w/v)
Cortizone 10Hydrocortisone1% (w/v)
Fleet Glycerin SuppositoriesGlycerin1% (w/v)
DesitinZinc Oxide1% (w/v)

{14}------------------------------------------------

Substance NameActive IngredientsTest Concentration
Anusol Pluspramoxine hydrochloride and zincsulfate monohydrate1% (w/v)
Preparation HPhenylephrine1% (w/v)
TumsCalcium Carbonate10% (w/v)
Equate Antacid Max StrengthAluminum hydroxide, Magnesiumhydroxide10% (w/v)
Mesalazine Rectal SuspensionEnemaMesalazine10% (w/v)
Fleet Mineral Oil EnemaMineral Oil10% (w/v)
Gynol II Vaginal ContraceptiveNonoxynol-91% (w/v)
Imodium ADLoperamide HCl10% (w/v)
Pepto BismolBismuth subsalicylate10% (w/v)
Ex-LaxSennosides1% (w/v)
MetronidazoleMetronidazole12.5 mg/ml
VancomycinVancomycin12.5 mg/ml
PolysporinBacitracin and Polymyxin B1% (w/v)
Naproxen sodiumNaproxen sodium12.5 mg/ml
Tucks personal cleaning padsWitch hazel10% (v/v)
Benzalkonium Chloride TowelettesBenzalkonium Chloride10% (v/v)
EthanolEthanol10% (v/v)
MucusImmunoglobulins, Lysozyme,Polymers, etc.3.5%
Whole BloodGlucose, Hormones, Enzymes, Ions,Iron, etc.10%
Palmitic acidPalmitic acid12.5 mg/ml
Steric AcidSteric Acid12.5 mg/ml
Triglyceride Mix (C2 - C10)Triglyceride10%

Analytical Reactivity (Inclusivity):

The inclusivity of the Solana C. difficile Assay was further evaluated by functional testing of toxigenic C. difficile in addition to those strains used in the LOD study. Twenty-three (23) additional strains of toxigenic C. difficile, representing at least five different toxinotypes, were tested in triplicate at the 2X LOD concentration (1.83E+04 CFU/mL as determined for ATCC BAA-1805) with Solana C. difficile Assay, in addition to the strains used for the LOD study. All twentythree additional strains were detected in all replicates by the Solana C. difficile Assay in this study.

StrainToxinotype
ATCC BAA-1805*III

{15}------------------------------------------------

StrainToxinotype
CCUG 20309*X
ATCC BAA-1870IIIb
CCUG 37770IV
ATCC BAA-1875V
ATCC 43598VIII
ATCC 37774XXIII
CCUG 9004Unknown
ATCC BAA-18740
ATCC 436000
ATCC BAA-18710
ATCC BAA-1803IIIc
ATCC BAA-18720
ATCC 7007920
ATCC 435990
CCUG 60276Unknown
CCUG 60275Unknown
CCUG 37778Unknown
CCUG 37777Unknown
CCUG 37776Unknown
CCUG 37773Unknown
ATCC 178570
ATCC 435940
ATCC 435960
ATCC 432550

3. Clinical studies:

Performance characteristics of the Solana C. difficile Assay were established during a prospective study conducted November 2016 to February 2017. Eight hundred fiftyfour (854) specimens used for this study were collected from patients suspected of having Clostridium difficile infection (CDI) at three distinct geographical sites across the United States. These specimens were tested with the Solana Assay at the sites. The Solana results were compared to an enhanced toxigenic bacterial culture (sensitivity/specificity) and a FDA-cleared molecular device (positive/negative percent agreement).

Performance in Comparison to Enhanced toxigenic bacterial culture

{16}------------------------------------------------

Eight hundred fifty-four (854) raw specimens were tested by both the Solana C. difficile Assay and enhanced toxigenic culture. For the toxigenic culture method, samples were inoculated into chopped-meat glucose (CMG) broth and sub-cultured after 48-hours onto CCFA-HB plates. Suspicious colonies were further characterized and C. difficile identified colonies were sub-cultured in CMG broth for subsequent cytotoxin testing. Two (2) specimens (0.2%) were invalid in the Solana C. difficile Assay when tested according to the Solana C. difficile Assay draft instructions for use. Both specimens remained invalid upon repeat testing. These specimens were removed from further analysis. The data below is for the remaining eight hundred fifty-two (852) specimens.

Combined Sites – Raw Specimen
Enhanced Toxigenic Culture95% CI
POSNEGTotalSensitivity93.0%86.9%96.4%
SolanaC. difficile AssayPOS1076*113Specificity99.2%98.2%99.6%
NEG8**731739
Total115737852
  • Three (3) of the six (6) specimens were positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device, three (3) were negative.

** Six (6) of eight (8) specimens were found positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device., and two (2) were negative

Performance in Comparison to FDA-cleared molecular device

Eight hundred fifty-four (854) specimens were tested by both the Solana C. difficile Assay and FDA-cleared molecular device. Two (2) specimens (0.2%) were invalid in the Solana C. difficile Assay when tested according to the Solana C. difficile Assay draft instructions for use. Both specimens remained invalid upon repeat testing. These specimens were removed from further analysis. The data below is for the remaining eight hundred fifty-two (852) specimens.

Combined Sites - Raw Specimen
FDA-cleared molecular device95% CI
POSNEGTotalPositive Percent97.0%91.59%99.0%

{17}------------------------------------------------

510(k) Summary

AgreementNegativePercentAgreement
SolanaC. difficile Assay
POS9716*11397.9%96.6%98.7%
NEG3**736739
Total100752852
  • Twelve (12) of the sixteen (16) specimens were positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device, four (4) were negative.

** Two (2) of three (3) specimens were found positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device, and one (1) was negative

5. Expected values:

The expected values of the Solana C. difficile Assay were established during a prospective study conducted between November 2016 to February 2017. Eight hundred fifty-four (854) specimens used for this study were collected from patients suspected of having Clostridium difficile-infection (CDI) at three distinct geographical sites across the United States. A single specimen was collected per patient. The specimens were processed and tested with Solana C. difficile Assay on the Solana instrument at the sites.

Combined Sites – Age and Gender Distribution
SexFMTotalTotal percent positive with theSolana C. difficile Assay in Rawspecimens
≤ 2 years33616.7% (1/6)
3 to 11 years461020.0% (2/10)
12 to 17 years410147.1% (1/14)
18 to 21 years11142524.0% (6/25)
22 to 59 years20613233814.2% (48/337*)
≥ 60 years26819346112.0% (55/460*)
Total49635885413.1% (112/852**)

Patient age and gender for the combined sites are presented below.

  • One (1) specimen was invalid

** Two (2) total specimens were invalid

§ 866.3130 Clostridium difficile toxin gene amplification assay.

(a)
Identification. AClostridium difficile toxin gene amplification assay is a device that consists of reagents for the amplification and detection of target sequences inClostridium difficile toxin genes in fecal specimens from patients suspected of havingClostridium difficile infection (CDI). The detection of clostridial toxin genes, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of CDI caused byClostridium difficile. (b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Toxin Gene Amplification Assays for the Detection ofClostridium difficile; Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.