K Number
K170464
Date Cleared
2017-11-07

(265 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The StatStrip Xpress Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood obtained for singertip. It is intended for single-patient home use and should not be shared. It is intended for self-testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or diabetes, and it is not intended for use with neonates.

The StatStrip Xpress Blood Glucose Monitoring System comprises the StatStrip Xpress Blood Glucose Monitor and StatStrip Xpress Glucose Test Strips.

· The StatStrip Xpress Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use).

Device Description

The Nova Biomedical StatStrip Xpress Blood Glucose Monitoring System consists of a hand held StatStrip Xpress Blood Glucose Monitor, StatStrip Xpress Blood Glucose Test Strips, StatStrip Xpress Blood Glucose Control Solutions (Levels 1, 2, 3, sold separately), and Instructions for Use Manual.

The modified StatStrip Xpress Blood Glucose Monitoring System is identical in fit, form, and functions to the StatStrip Xpress Blood Glucose Monitoring System, previously cleared in K160156. The Blood Glucose Monitor, Test Strip, and Control Solutions are identical. The only difference between the predicate and proposed device is the labeling.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: StatStrip Xpress Blood Glucose Monitoring System


1. Table of Acceptance Criteria and Reported Device Performance

The document describes a Human Factors Study related to the device's labeling and understanding of accuracy information. The specific acceptance criteria are presented as the "predetermined acceptance criteria" for the Human Factors Study. While the exact numerical targets for each topic aren't explicitly stated as "acceptance criteria," the study's conclusion indicates that the results met these criteria. We can infer the implicit acceptance criteria from the reported "better than average" performance and the conclusion.

Acceptance Criteria (Implied)Reported Device Performance (Average Score %)
Understanding Front Kit Carton (e.g., above a certain threshold)93.3%
Understanding Side Kit Carton (e.g., above a certain threshold)82.2%
Understanding of Accuracy Comparison Information System A vs. B (e.g., above threshold)86.0%
Understanding of Comparison Information System A vs. C (e.g., above threshold)92.1%
Understanding Comparison Information System A vs. D (e.g., above threshold)91.7%
Adequacy of Accuracy Summary Chart & Accuracy Grade depictions (overall effectiveness)Concluded as "effective way to present" and "meets the predetermined acceptance criteria."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 100 subjects
  • Data Provenance: The document does not explicitly state the country of origin. However, given it's an FDA submission, the study was likely conducted in the United States. It is a prospective study as it involves active participation of subjects to assess understanding of new labeling.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This study is a Human Factors Study evaluating user understanding, not a clinical accuracy study that would typically require expert-established ground truth. The "ground truth" for this study is essentially the correct interpretation of the accuracy information as intended by the device manufacturer and regulatory guidelines. The study itself assesses the subjects' ability to understand this information. Therefore, no external experts were used to establish "ground truth" in the traditional sense of diagnostic accuracy. The study staff provided the materials and questions designed to test understanding.


4. Adjudication Method for the Test Set

The document does not describe an adjudication method in the context of expert review for diagnostic discrepancies. This is because the study focuses on user comprehension of labeling information, not on diagnostic accuracy. Subjects answered questions, and their answers were presumably scored based on pre-defined correct responses from the study design.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for imaging devices or diagnostic tests where multiple readers interpret cases. The StatStrip Xpress Blood Glucose Monitoring System is a self-testing blood glucose meter, and this particular submission focuses on a Human Factors Study regarding labeling comprehension.


6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone performance study (algorithm only) was not explicitly detailed in this summary. The Human Factors Study is inherently "human-in-the-loop" as it assesses how human users understand the device's labeling. While the device itself (StatStrip Xpress) has an algorithm for glucose measurement, performance data for that algorithm would typically be covered in separate analytical and clinical accuracy studies, not this Human Factors summary. This submission focuses on a modification to labeling after a previous clearance (K160156), which likely covered the algorithm's standalone performance.


7. Type of Ground Truth Used

For this Human Factors Study, the ground truth is the intended and correct understanding of the accuracy information presented on the device's labeling. The study aimed to verify that laypersons could correctly interpret this information. This is distinct from ground truth based on pathology, expert consensus, or outcomes data, which are typically used for diagnostic performance validation.


8. Sample Size for the Training Set

The document does not report a sample size for a training set. This is because the Human Factors Study described is an evaluation of understanding, not the development or training of an algorithm. If this were a machine learning algorithm, there would be a distinct training set. The subjects in this study were an evaluation group.


9. How the Ground Truth for the Training Set Was Established

As there was no training set for an algorithm and this study is a Human Factors evaluation, the concept of establishing ground truth for a training set is not applicable here. The study assesses the "ground truth" of user comprehension against the pre-defined correct answers to comprehension questions.

{0}------------------------------------------------

November 7, 2017

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Nova Biomedical Corporation Eliza Wang Regulatory Affairs Project Manager 200 Prospect Street Waltham, Massachusetts 02454

Re: K170464

Trade/Device Name: StatStrip Xpress Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: October 4, 2017 Received: October 5, 2017

Dear Eliza Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170464

Device Name

StatStrip Xpress Blood Glucose Monitoring System

Indications for Use (Describe)

The StatStrip Xpress Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood obtained for singertip. It is intended for single-patient home use and should not be shared. It is intended for self-testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or diabetes, and it is not intended for use with neonates.

The StatStrip Xpress Blood Glucose Monitoring System comprises the StatStrip Xpress Blood Glucose Monitor and StatStrip Xpress Glucose Test Strips.

· The StatStrip Xpress Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use).

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for K170464

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Eliza Wang, RACRegulatory Affairs Project Manager
Date Prepared:November 3rd, 2017
Proprietary Name:StatStrip Xpress Blood Glucose Monitoring System
Common or Usual Name:System, test, blood glucose, over the counter
Classification Name:Glucose test system
Class Panel No.75
Regulation Number:21 CFR 862.1345
Product Codes:NBW
Class:II
Predicate Device:StatStrip Xpress Blood Glucose Monitoring SystemK160156

Device Description:

The Nova Biomedical StatStrip Xpress Blood Glucose Monitoring System consists of a hand held StatStrip Xpress Blood Glucose Monitor, StatStrip Xpress Blood Glucose Test Strips, StatStrip Xpress Blood Glucose Control Solutions (Levels 1, 2, 3, sold separately), and Instructions for Use Manual.

The modified StatStrip Xpress Blood Glucose Monitoring System is identical in fit, form, and functions to the StatStrip Xpress Blood Glucose Monitoring System, previously cleared in K160156. The Blood Glucose Monitor, Test Strip, and Control Solutions are identical. The only difference between the predicate and proposed device is the labeling:

    1. The accuracy information has been added to the modified StatStrip Xpress Blood Glucose Monitoring System kit carton label. This format is supported by the results of the newly conducted StatStrip Xpress Blood Glucose Monitoring System Human Factors Study.

{4}------------------------------------------------

    1. The Accuracy table on the original StatStrip Xpress Glucose Test Strips Insert Sheet has been replaced by the Accuracy Table suggested in the FDA guidance document "Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use" (October 11, 2016).
    1. The StatStrip Xpress Blood Glucose Monitoring System kit carton label has been updated with the following statement, "Proven hospital technology now available for personal use". This is supported in K160156 that the StatStrip Xpress Blood Glucose Monitoring System is equivalent to the Reference Device in K160156; StatStrip Xpress Glucose Hospital Meter System (K150461).
    1. The statement of Interferences on the StatStrip Xpress Blood Glucose Monitoring System kit carton label has been updated to "No Known interferences from common medicines like Ibuprofen, Acetaminophen, and Vitamin C".

Intended Use:

The StatStrip Xpress Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood obtained from the fingertip. It is intended for single-patient home use and should not be shared. It is intended for self-testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes, and it is not intended for use with neonates.

The StatStrip Xpress Blood Glucose Monitoring System comprises the StatStrip Xpress Blood Glucose Monitor, and StatStrip Xpress Glucose Test Strips. .

  • . The StatStrip Xpress Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use).

Summary of the Technological Characteristics:

The modified StatStrip Xpress Blood Glucose Monitor System that is the subject of this submission is identical in form, fit and function, to the device cleared in K160156. The StatStrip Xpress Glucose Monitoring System is intended for in vitro diagnostic use by lay people for the quantitative determination of glucose in whole blood.

The StatStrip Xpress Glucose Monitor is a hand-held testing device that works in conjunction with the StatStrip Glucose Test Strips. Monitor operation is self-prompting using a segmented liquid crystal display (LCD) and icons. Function and data selection is accomplished using 3 push buttons. The handheld monitor supports audible alerts and prompts with a built-in beeper. In addition to measuring glucose, the meter also stores up to 400 patient test records. The user can recall and review test results.

{5}------------------------------------------------

A single coin battery powers the device, and is expected to perform up to 600 tests before needing to be replaced. A low battery prompt will appear when it is time to replace the battery. All test data is stored in a non-volatile form to prevent data loss.

Test Strips:

The StatStrip Xpress Glucose Test Strips of this submission are identical in form, fit, function and packaging, to the glucose test strips originally cleared in K160156. The Test Strip is designed with an electrode that measures Glucose levels. Glucose in the blood sample mixes with reagent on the test strip that produces an electric current. The amount of current that is produced depends on the amount of Glucose in the blood.

The StatStrip Xpress Glucose Test Strip reacts with the glucose in the test sample. The reaction produces an electrical current which is proportional to the amount of glucose in the sample, and the electrical current is detected by the monitor and displayed to the user as a glucose value.

The Test Strip is designed such that when a drop of blood touched the end of the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be available in cartons of 100 strips (50 strips/vial).

  • Contents: Each glucose test strip contains a reaction layer that contains glucose ● enzyme (Aspergillus sp.) >1.0 IU, mediator >20 µg, and other nonreactive substances.
  • Function: The StatStrip Xpress Glucose Test Strips are intended for quantitative . determination of Glucose in fresh whole blood specimens. StatStrip Glucose Test Strips are for use only with the StatStrip Family of Monitors.
  • . Storage Conditions: Store the Test Strips in the vial between 34-86°F (1-30°C). Ensure that the vial is closed between uses. Once opened the test strips in the vial may be used for 180 days or until the expiration date printed on the label, whichever comes first.

{6}------------------------------------------------

CharacteristicPredicate Device (K160156)Proposed Device (K170464)
StatStrip Xpress Blood Glucose Monitoring SystemStatStrip Xpress Blood Glucose Monitoring System
Intended UseThe StatStrip Xpress Blood Glucose Monitoring System isintended to be used for the quantitative measurement ofglucose in fresh capillary whole blood obtained from thefingertip. It is intended for single-patient home use andshould not be shared. It is intended for self-testing outsidethe body by people with diabetes mellitus as an aid tomonitor the effectiveness of diabetes control. It is notintended for the diagnosis of or screening for diabetes, andit is not intended for use with neonates.The StatStrip Xpress Blood Glucose Monitoring Systemcomprises the StatStrip Xpress Blood Glucose Monitor,StatStrip Xpress Glucose Test Strips, and StatStrip XpressGlucose Control Solutions. The StatStrip Xpress GlucoseControl Solutions are intended for use with the StatStripXpress Blood Glucose Monitoring System as a qualitycontrol check to verify the accuracy of blood glucose testresults. There are 3 levels of controls (Level 1, 2, and 3).The StatStrip Xpress Blood Glucose Monitoring System is intended for use outside the body ( in vitro diagnostic use).The StatStrip Xpress Blood Glucose Monitoring System is intended to be used for the quantitativemeasurement of glucose in fresh capillary wholeblood obtained from the fingertip. It is intended forsingle-patient home use and should not be shared. It isintended for self-testing outside the body by peoplewith diabetes mellitus as an aid to monitor theeffectiveness of diabetes control. It is not intended forthe diagnosis of or screening for diabetes, and it is notintended for use with neonates.The StatStrip Xpress Blood Glucose Monitoring System comprises the StatStrip Xpress BloodGlucose Monitor and StatStrip Xpress Glucose TestStrips.The StatStrip Xpress Blood Glucose Monitoring System is intended for use outside the body ( in vitro diagnostic use).
CharacteristicPredicate Device (K160156)Proposed Device (K170464)
StatStrip Xpress Blood Glucose Monitoring SystemStatStrip Xpress Blood Glucose Monitoring System
EnzymeGlucose OxidaseSame
Operating PrincipleElectrochemical biosensor, amperometricSame
Sample typeCapillary whole blood (finger stick)Same
Sample size1.2 μLSame
Sample applicationTest Strip capillary drawSame
Measuring range20-600 mg/dLSame
Hematocrit range20-65%Same
Reported outputmg/dLSame
Time Resultto ~ 6 secondsSame
Quality control3 levelsSame
HandheldYesSame
CalibrationAutomatic, no calibration codeSame
Data storage400 test resultsSame
BarcodeNoSame
CharacteristicPredicate Device (K160156)Proposed Device (K170464)
StatStrip Xpress Blood Glucose Monitoring SystemStatStrip Xpress Blood Glucose Monitoring System
Power source3v DC Li coin cell batterySame
Dimensions3.6x2.3x0.9 in (91.4x58.4x22.9 mm)Same
Weight2.65 ounces (75grams)Same

Table 1: Comparison of Predicate and Proposed devices

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Summary of Testing:

Human Factors Study

A Human Factors study of 100 Intended Use Population (Type 1 and Type 2 Layperson Diabetics) was conducted to assess the overall understanding of the accuracy information presented on the proposed outer carton label of the StatStrip Xpress Blood Glucose Monitoring System.

Inclusion Criteria:

    1. Is 18 years of age or older
    1. Is able to read and write English
    1. Has been diagnosed with Type 1 Diabetes or,
    1. Has been diagnosed with Type 2 Diabetes or,

Exclusion Criteria:

    1. Does not meet the inclusion criteria or,
    1. Has a cognitive disorder or other condition, which, in the opinion of the investigator, would seriously compromise the integrity of the study.

Subject Demographics:

Table 1: Subject Age

AgeNumber of Subjects
25-302
31-4010
41-5010
51-6024
61-7032
71-8020
81-832
Total100

Table 2: Subject Gender

GenderNumber of Subjects
Male37
Female63
Total100

{10}------------------------------------------------

Table 3: Subject Race

RaceNumber of Subjects
African American54
Caucasian45
Native American1
Total100

Table 4: Subject Education

Education (Total number of years)Number of Subjects
82
92
102
112
1275
133
1411
ા ર0
163
Total100

Methods and Procedures:

    1. Potential Subjects were pre-screened by the study staff for inclusion into the study using the "Inclusion/Exclusion Form" (Form I/E).
    1. The Subject completed the Informed Consent (IC) process.
    1. Each Subject was assigned a Subject number by the study staff.
    1. A study staff member gave the Subject the "Mathematics Understanding Assessment Form" (Form MUA) and asked them to answer the questions.
    1. A study staff member provided the Subject with the Kit Carton label proposal for the StatStrip Xpress Blood Glucose Monitoring System.
    1. Each Subject was then asked to completely review the Kit Carton label contents of the StatStrip Xpress Blood Glucose Monitoring System.
    1. The Subject answered each of the questions on the forms to determine their level of understanding of the StatStrip Xpress Blood Glucose Monitoring System Kit Carton labels.
    1. A study staff member provided the Subject with the Kit Carton label for the StatStrip Xpress Blood Glucose Monitoring System alongside draft Kit Carton label for the Nova Max Plus Blood Glucose Monitoring System.
    1. Each Subject was asked to completely review and compare the labeling contents.

{11}------------------------------------------------

    1. The Subject answered each of the questions to determine their ability to correctly interpret the accuracy information when comparing one meter's accuracy information to another meter's accuracy information.
    1. The Subject answered each of the questions to determine how valuable the Kit Carton accuracy labeling information presented would be to them.

Study Findings:

Mathematics Assessment:

The average Subject score for the 10-question Mathematics Assessment was 71.3% across all races, educational levels, ages, and both genders.

Human Factors Assessment:

The data collected during this Human Factors Study shows that Subjects with minimal education and who demonstrated limited mathematics skills in the mathematics assessment are able to understand the accuracy information presented on the proposed Nova StatStrip Xpress kit packaging.

Although the Subjects exhibited an average mathematical assessment score of 71.3% when performing basic mathematical percentage calculations, the Subjects exhibited a better than average (82.2% - 93.3%) ability to understand the accuracy information proposed for the device's outer kit carton.

TopicForm NumberAverage Score (%)
Mathematics AssessmentMUA71.3
Understanding Front Kit CartonHFA 1A (Questions 1-3)93.3
Understanding Side Kit CartonHFA 1B82.2
Understanding of Accuracy Comparison Information System A vs. BHFA 286.0
Understanding of Comparison Information System A vs. CHFA 392.1
Understanding Comparison Information System A vs. DHFA 491.7

Table 5: Summary of Responses to Objective Questions

{12}------------------------------------------------

The Subjects answers to the subjective questions indicated that an overwhelming majority understood the importance of accuracy when choosing a glucose meter to monitor their diabetes.

Furthermore, the Subjects understood the information presented to them on the proposed kit packaging, and believed that there is value in accuracy ratings when selecting a Blood Glucose Monitoring System.

Conclusion:

The proposed Accuracy Summary Chart and Accuracy Grade depictions evaluated in this study are an effective way to present the performance of the Nova StatStrip Xpress Blood Glucose Monitoring System to layperson Subjects and meets the predetermined acceptance criteria.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.