K Number
K170360

Validate with FDA (Live)

Device Name
NAVIO
Date Cleared
2017-05-02

(85 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA),

The Navio system is indicated for use with cemented implants only.

Device Description

The NAVIO system is a computer-assisted orthopedic surgical navigation and surgical burring system. The system uses established technologies of navigation, via a passive infrared tracking camera, to aid the surgeon in establishing a bone surface model for the target surgery and in planning the surgical implant location, based on intraoperativelydefined bone landmarks and known geometry of the surgical implant. The NAVIO system then aids the surgeon in executing the surgical plan by using a standard off-theshelf surgical drill motor and bur (Anspach eMax2 Plus System, cleared via K080802), which has been adapted using a tracking system.

The surgical bur is inserted into a handpiece, which allows the bur to move within the handpiece. The NAVIO system software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved in two ways:

  • Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the NAVIO system retracts the bur inside the guard, disabling cutting.
  • . Speed control regulates the signal going to the drill motor controller itself and will limit the speed of the drill if the target surface is approached. This mode of operation is useful in shaping surfaces of the condyle as well as placing post holes and fixation features for femoral and tibial cut guides.

Additionally, the surgeon can disable both controls and operate the NAVIO system handpiece as a standard navigated surgical drill. The surgeon must press on a footpedal to activate the surgical bur and enable cutting in all modes.

AI/ML Overview

The provided document is a 510(k) premarket notification for the "Navio" surgical system, focusing on updates to its Total Knee Arthroplasty (TKA) software and support for additional implants. This document details non-clinical testing performed to establish substantial equivalence to a predicate device, rather than a standalone clinical study to prove novel device performance.

Therefore, the information regarding acceptance criteria and a study proving those criteria are met is derived from the non-clinical (bench) testing described in the submission, which aimed to demonstrate that the updated device is as safe and effective as its predicate.

Here's the breakdown of the information based on the provided text:


Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a dedicated table. However, the conclusion section implies the key acceptance criterion is that the total knee implant placement accuracy using the new "Bur All" method of bone removal is "acceptable and similar" to the implant placement accuracy required for unicondylar and patellofemoral implant placement. It also states that the placement accuracy using the TKA cut guides (including the new Tibia Twin Peg cut guides) is "acceptable and similar" to previous submissions.

The "reported device performance" is qualitative and states that these criteria were met.

Acceptance Criteria (Implied)Reported Device Performance
Total knee implant placement accuracy using the "Bur All" method is acceptable and similar to UKR and PFA implant placement accuracy."The accuracy verification testing performed indicates that the total knee implant placement accuracy using Bur All method of bone removal is acceptable and similar to the implant placement accuracy required for unicondylar and patellofemoral implant placement." (K170360, Page 10)
Placement accuracy using the NAVIO TKA cut guides (including new Tibia Twin Peg cut guides) is acceptable and similar to previous 510(k) submissions."Additionally, placement of the LEGION CR/PS and GENESIS II CR/PS Total Knee Implant systems using the NAVIO TKA cut guides (including the new Tibia Twin Peg cut guides) is acceptable and similar to the placement accuracy presented in the previous 510(k)." (K170360, Page 10)
Surgeon users are able to perform a TKA procedure in a simulated OR environment safely and effectively using the NAVIO system with TKA application."Usability validation testing demonstrated that surgeon users were able to perform a TKA procedure on cadaver in a simulated operating room environment safely and effectively using the NAVIO system with TKA application." (K170360, Page 10) and "Usability testing validated that surgeon users were able to perform a TKA procedure on cadaver in a simulated operating room environment safely and effectively using the NAVIO system with TKA application." (K170360, Page 9)

Analysis of the Study and Ground Truth Establishment

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: "Simulated knees (sawbones)" and "cadaver laboratory testing" were used. The exact number of simulated knees or cadavers is not specified in the provided document.
    • Data Provenance: The testing was non-clinical, conducted in a laboratory setting. No country of origin for the cadavers or sawbones is specified, nor is whether the data is retrospective or prospective, as it's a bench test.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • "Trained surgeons and technical support personnel" performed the verification accuracy testing and usability testing.
    • The number of surgeons/personnel is not specified.
    • Their specific qualifications (e.g., years of experience) are not detailed beyond "trained surgeons and technical support personnel."
  3. Adjudication method for the test set:

    • The document does not describe an adjudication method for establishing ground truth or evaluating the results of the non-clinical tests. This is typically more relevant for studies involving human interpretation or subjective measurements.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. The document explicitly states: "No human clinical testing was conducted to determine safety and effectiveness of the NAVIO system." (K170360, Page 10). The study described is a non-clinical bench test confirming the performance of the device itself and its updated features.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The primary evaluation described is a "human-in-the-loop" test, as "Trained surgeons and technical support personnel performed verification accuracy testing using simulated knees (sawbones) and cadaver laboratory testing." The device is intended to assist a surgeon. The accuracy testing evaluated the system's ability to achieve "acceptable" implant placement when operated by a surgeon.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Given the nature of the device (a surgical navigation and burring system) and the non-clinical testing, the ground truth for "accuracy" would likely be based on engineering measurements or predefined geometric targets. This means comparing the actual bone removal or implant placement (as measured by precision instruments on sawbones/cadavers) against the software-defined surgical plan/target. The "acceptable and similar" criteria suggest a comparison to known accurate outcomes for previous versions or similar surgical methods.
  7. The sample size for the training set:

    • The document does not mention a training set sample size. This submission is for updates to an existing system ("enhancements to the NAVIO system TKA software"), suggesting the core algorithms were likely developed and "trained" (if machine learning was involved, which is not indicated for this specific system update in the document) previously. The current submission focuses on verifying the impact of software enhancements and new implant support through non-clinical testing.
  8. How the ground truth for the training set was established:

    • As no training set is discussed, the method for establishing its ground truth is not provided. The system uses "established technologies of navigation... to aid the surgeon in establishing a bone surface model... and in planning the surgical implant location, based on intraoperatively-defined bone landmarks and known geometry of the surgical implant." This suggests a reliance on precise geometric modeling and established anatomical landmarks rather than a machine learning model requiring a traditional "training set" with ground truth labels in the typical sense.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DEPARTMENT OF HEALTH & HUMAN SERVICES - USA

May 2, 2017

Blue Belt Technologies, Inc. Richard Confer Director, Regulatory Affairs 2905 Northwest Blvd. Ste. 40 Plymouth, Minnesota 55441

Re: K170360

Trade/Device Name: Navio Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, HRY, KRR, NPJ, JWH, HSX Dated: February 2, 2017 Received: February 6, 2017

Dear Richard Confer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510{k) Number (if known)

K170360

Device Name Navio

Indications for Use (Describe)

The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA),

The Navio system is indicated for use with cemented implants only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K170360 - Page 1 of 8

Image /page/3/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.

2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com

enhancements to the NAVIO system TKA software.

510(k) Summary

510(k) OwnerBlue Belt Technologies, Inc.2905 Northwest Blvd Ste. 40Plymouth, MN 55441 USATel: (763) 452-4950Fax: (763) 452-4675
Contact PersonRichard G. ConferDirector, Regulatory AffairsTel: (412) 683-3844 x 4106Email: rick.confer@smith-nephew.com
Date of SubmissionFebruary 2, 2017
Classification Reference21 CFR 882.4560
Product CodeOLO
Supported CodesHSX, HRY, KRR, NPJ, JWH
Common/Usual NameOrthopedic Sterotaxic Instrument
Trade/Proprietary NameNAVIO° Surgical System (NAVIO system)
Predicate Device(s)NAVIO system (K160537)
Reason for SubmissionAddition of the Smith & Nephew LEGION® TotalKnee System CR/PS and GENESIS® II CR/PS TotalKnee System to the list of supported implants,updates to the NAVIO Total Knee Arthroplasty(TKA) instrument kit and components, and feature

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.

Intended Use

The NAVIO system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Indications for Use

The NAVIO system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA).

The NAVIO system is indicated for use with cemented implants only.

The Intended Use and Indications for Use statements are the same as the predicate device.

Device Description

The NAVIO system is a computer-assisted orthopedic surgical navigation and surgical burring system. The system uses established technologies of navigation, via a passive infrared tracking camera, to aid the surgeon in establishing a bone surface model for the target surgery and in planning the surgical implant location, based on intraoperativelydefined bone landmarks and known geometry of the surgical implant. The NAVIO system then aids the surgeon in executing the surgical plan by using a standard off-theshelf surgical drill motor and bur (Anspach eMax2 Plus System, cleared via K080802), which has been adapted using a tracking system.

The surgical bur is inserted into a handpiece, which allows the bur to move within the handpiece. The NAVIO system software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved in two ways:

  • Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the NAVIO system retracts the bur inside the guard, disabling cutting.
  • . Speed control regulates the signal going to the drill motor controller itself and will limit the speed of the drill if the target surface is approached. This mode of operation is useful in shaping surfaces of the condyle as well as placing post holes and fixation features for femoral and tibial cut guides.

Additionally, the surgeon can disable both controls and operate the NAVIO system handpiece as a standard navigated surgical drill. The surgeon must press on a footpedal to activate the surgical bur and enable cutting in all modes.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.

UKR and PFA Overview

In the UKR and PFA applications, the NAVIO system uses established technologies to prepare bone for attachment of implant components. The NAVIO system uses intraoperative data collection (image-free or non-CT data generation) to create a model of the patient's femur and/or tibia, depending on the procedure being performed, and allows the surgeon to prepare a surgical plan. The NAVIO system uses pre-defined boundaries generated during the planning process to control the motion of the surgical bur and limit the amount of bone removed in order to shape the condyles, tibial plateau, or patellofemoral joint in preparation for placement of the surgical implant.

TKA Overview

Similar to the UKR and PFA applications, the TKA application uses intraoperative data collection to create a model of the patient's femur and tibia and allows the surgeon to prepare a surgical plan. In the TKA application, the NAVIO system is used to plan the location of the femur and tibia implants, as well as cut guides and position the features used to secure the cut guides in place. The NAVIO system controls the cutting action of the bur during placement of the femur and tibia cut guides. The surgeon finishes the preparation of the bone surface using standard surgical saws, guided by the slots in the cut guides. Cutting action of the saw is not controlled by the NAVIO system. The surgeon follows the implant manufacturer's recommended procedure to remove bone and place the implant, just as if the cut guide had been placed manually. The manual instrumentation technique established by each implant's manufacturer is used to prepare the remainder of the bone surface to receive the total knee implant components.

One of the enhancements to the TKA application software is the addition of a Bur All feature that enables the surgeon to prepare bone for the attachment of total knee implant components. Using the same method of bone removal as the UKR and PFA applications, the NAVIO system controls the bur either by retracting the bur in a guard (Exposure Control) or by controlling the speed of the bur (Speed Control) as the target surface is approached. The Bur All feature also provides a way for a surgeon that performed bulk bone removal using a standard surgical saw to refine the bone model and make additional fine cuts to clean up the bone surface to receive the implant.

The following diagram shows the primary workflow steps in each application: UKR, PFA, and TKA. The three procedures are mutually independent and cannot be planned or completed in parallel. Note: The workflows for UKR and PFA procedures have not changed from the predicate device, cleared via K160537.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The logo is clean and modern, with a focus on the company name.

Figure 1: NAVIO System Workflows - UKR, PFA, and TKA

Image /page/6/Figure/4 description: This image shows three flowcharts for different applications: PFA, TKA, and UKR. Each flowchart outlines the steps involved in the respective application, starting with hardware and handpiece connections. The flowcharts progress through various stages, including bur and control selection, handpiece retraction and calibration, and homing validation. The TKA and UKR applications include additional steps specific to those procedures, such as surgical case options, bone tracker attachment, and component placement.

ReshapingMobility™

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.

Implant Model NameManufacturer510(k) Number
JOURNEY II CRSmith and NephewK121443
JOURNEY II BCSSmith and NephewK111711
GENESIS II CR/PSSmith and NephewK951987, K962557
LEGION CR/PSSmith and NephewK951987, K962557, K093746
NEO Total Knee SystemNew Era OrthopedicsK142388

Currently Supported Total Knee Implants

Discussion of Similarities and Differences

The NAVIO system presented in this 510(k) submission is substantially equivalent to the predicate NAVIO, K160537. The intended use, indications for use, and the general functionality of the NAVIO system are unchanged from the previously submitted device.

This submission supports the following updates to the NAVIO system:

  • the addition of the Smith & Nephew LEGION Total Knee System CR/PS and GENESIS II Total Knee System CR/PS to the list of supported implants,
  • . updates to the NAVIO TKA instrument kit and components, and
  • feature enhancements to the NAVIO system Total Knee Application software. ●

The implant product codes supported by the subject device are consistent with the predicate device. The LEGION Total Knee System CR/PS and GENESIS II Total Knee System are cemented implants that share the product code JWH, a currently supported NAVIO product code. These implants have been successfully imported and appropriately validated for use with the NAVIO system.

The NAVIO TKA instrument kit was updated to include two new components: a new femur drill guide to support the LEGION-GENESIS II CR/PS implants and a new Twin Peg Tibia Cut Guide.

Feature enhancements were made to the NAVIO TKA application software. The NAVIO system TKA workflow, described in K160537, has been updated to improve usability of the application, simplify the user interface to remove features that were not being utilized, and to introduce new features. One of the primary features that is being introduced is the option to use the NAVIO handpiece to bur away all bone necessary to accurately place a total knee implant component; this functionality is similar to the UKR and PFA applications.

The established technologies that are used by the NAVIO system to prepare bone for attachment of implant components or for the attachment of the TKA femur and tibia cutting guides are unchanged.

Figure 2: Summary of Technological Similarities with Predicate

ReshapingMobility™

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.

Subject DeviceNAVIO [Subject]Predicate DeviceNAVIO [K160537]
Intended useSame as Predicate.The NAVIO system is intended toassist the surgeon in providingsoftware-defined spatial boundariesfor orientation and referenceinformation to anatomical structuresduring orthopedic procedures.
Indications forUseSame as Predicate.The NAVIO system is indicated for usein surgical knee procedures in whichthe use of stereotactic surgery may beappropriate, and where reference torigid anatomical bony structures canbe determined. These proceduresinclude unicondylar kneereplacement, patellofemoralarthroplasty, and total kneearthroplasty.The NAVIO system is indicated for usewith cemented implants only.
SupportedProduct Code(s)Same as Predicate.HSX, HRY, KRR, NPJ, JWH
Environment ofUseSame as Predicate.Intended for use by trainedorthopedic surgeons in an orthopedicsurgical suite.
TechnologicalCharacteristicsThe NAVIO system uses establishedtechnologies to prepare bone forattachment of UKR, PFA, or TKAimplant components. In the case of atotal knee arthroplasty, the bonesurface may also be prepared toreceive the femoral and tibial cuttingguides.NAVIO uses intraoperative datacollection (image-free or non-CT datageneration) to create a model of thepatient's femur and/or tibia,dependent on the procedure beingThe NAVIO system uses establishedtechnologies to prepare bone forattachment of UKR and PFA implantcomponents, or in the case of a totalknee arthroplasty, the bone surface isprepared to receive the femoral andtibial cutting guides.NAVIO uses intraoperative datacollection (image-free or non-CT datageneration) to create a model of thepatient's femur and/or tibia,dependent on the procedure being
performed, and allows the surgeon toprepare a surgical plan.performed, and allows the surgeon toprepare a surgical plan.
The NAVIO system uses predefinedboundaries generated during theplanning process to control themotion of the surgical bur and limitthe amount of bone removed inorder to shape the condyles, tibialplateau, or patellofemoral joint inpreparation for placement of thesurgical implant.The NAVIO system uses predefinedboundaries generated during theplanning process to control themotion of the surgical bur and limitthe amount of bone removed in orderto shape the condyles, tibial plateau,or patellofemoral joint in preparationfor placement of the surgical implant,
During a TKA procedure, the surgeonmay choose to prepare the bonesurface for receiving the implant byutilizing the Bur All method or thebone surface may be prepared toreceive the femoral and tibial cuttingguides (if cut guides are utilized, thebone surface is prepared using astandard surgical saw).or in the case of a total kneereplacement procedure, the bonesurface is prepared to receive thefemoral and tibial cutting guides. Bonesurface for receiving the implant isprepared using a standard surgicalsaw.
Bur cutting is controlled either byretracting the bur in a guard, or bycontrolling the speed of the bur asthe target surface is approached.Bur cutting is controlled either byretracting the bur in a guard, or bycontrolling the speed of the bur as thetarget surface is approached.

ReshapingMobility™

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.

Non-Clinical Testing (Bench)

Design verification and validation tests were performed on the NAVIO system to support the updates presented in this submission. Testing included software database reviews, bench testing, labeling inspection, drawing inspections, and usability testing.

Trained surgeons and technical support personnel performed verification accuracy testing using simulated knees (sawbones) and cadaver laboratory testing. Usability testing validated that surgeon users were able to perform a TKA procedure on cadaver in a simulated operating room environment safely and effectively using the NAVIO system with TKA application.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.

Clinical Testing

No human clinical testing was conducted to determine safety and effectiveness of the NAVIO system.

Conclusions

The NAVIO system described in this submission has the same intended use and the same technological characteristics as the NAVIO system, most recently cleared per K160537. The differences in the methodology of bone removal between the previously cleared NAVIO system (K160537) and the additional supported total knee implant systems, as well as updated instrumentation, and software feature enhancements, do not raise any new questions of safety or effectiveness.

The key determining factor is whether NAVIO control can be applied accurately to accomplish the desired cutting in accordance with the plan without exceeding the predefined limits. The accuracy verification testing performed indicates that the total knee implant placement accuracy using Bur All method of bone removal is acceptable and similar to the implant placement accuracy required for unicondylar and patellofemoral implant placement. Additionally, placement of the LEGION CR/PS and GENESIS II CR/PS Total Knee Implant systems using the NAVIO TKA cut guides (including the new Tibia Twin Peg cut guides) is acceptable and similar to the placement accuracy presented in the previous 510(k). Usability validation testing demonstrated that surgeon users were able to perform a TKA procedure on cadaver in a simulated operating room environment safely and effectively using the NAVIO system with TKA application.

The NAVIO system presented in this 510(k) premarket notification demonstrates that the updated NAVIO system, with these additional supported total knee implants, instrument tray and component changes, and software feature enhancements, continues to be as safe and effective as the predicate NAVIO system (K121936, K140596, K143668, K152574, and K160537).

Image /page/10/Picture/9 description: The image shows the text "ReshapingMobility™" in a blue color. The text is the main focus of the image and appears to be a logo or a slogan. The background is a solid blue color, which complements the text.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).