K Number
K170284
Date Cleared
2017-03-31

(60 days)

Product Code
Regulation Number
866.3980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Great Basin Bordetella Direct Test is a qualitative in vitro diagnostic test for the detection of Bordetella pertussis DNA from nasopharyngeal swab specimens obtained from patients suspected of having a respiratory tract infection attributable to B. pertussis.

The Bordetella Direct Test is performed on the PA500 Portrait Analyzer and utilizes PCR amplification of the insertion sequence IS481. The IS481 sequence is also found in other organisms including Bordetella holmesii or Bordetella bronchiseptica. Respiratory infection with B. pertussis, B. holmesii ot B. bronchiseptica may yield positive test results with IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results for the Great Basin Bordetella Direct Test do not preclude B. pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the Great Basin Bordetella Direct Test should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of Bordetella pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.

Device Description

The Great Basin Bordetella Direct Test on the PA500 Portrait™ Analyzer System utilizes automated hot-start PCR technology to target and amplify the IS481 insertion sequence of B. pertussis. Genomic DNA is extracted from microbial cells and diluted to reduce potential inhibitors of PCR. During PCR, double-stranded DNA is separated and the target nucleic acid sequence is amplified by thermal cycling using biotin-labeled primers that target the IS481 sequence for identification of B. pertussis. Following PCR, biotin-labeled amplicon is hybridized to sequence-specific capture probes immobilized on the silicon chip surface, then incubated with anti-biotin antibody conjugated to the horseradish peroxidase enzyme (HRP). The unbound conjugate is washed away and tetramethylbenzidine (TMB) is added to produce a visible precipitate at the location of the probe/target sequence complex. The resulting signal is detected by the automated Portrait™ Optical Reader within the PA500 Portrait™ Analyzer System. The Specimen Processing Control (SPC), undergoes the extraction, and detection steps to monitor for inhibitory substances as well as process inefficiency due to instrument or reagent failure. No operator intervention is necessary once the clinical sample is loaded into the sample port and the Bordetella Direct Test cartridge is loaded into the Portrait Analyzer.

The PA500 Portrait™ Analyzer System is a fully automated system that includes: The Portrait™ Analyzer, single-use Bordetella Direct Test Cartridges, and the Portrait™ Data Analysis Software Program. The Portrait™ System is designed to perform automated sample preparation, PCR, and optical chip-based detection with integrated data analysis in less than two hours.

AI/ML Overview

This documentation describes acceptance criteria and the study that proves the Great Basin Bordetella Direct Test meets these criteria. The device is a qualitative in vitro diagnostic test for the detection of Bordetella pertussis DNA from nasopharyngeal swab specimens.

Here's the detailed breakdown of the acceptance criteria and study information:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Great Basin Bordetella Direct Test are implicitly defined by the performance observed in the analytical and clinical studies, which are designed to demonstrate the device's acceptable functioning for its intended use. While explicit numerical acceptance criteria (e.g., "PPA must be >X%") are not directly stated as pass/fail thresholds in the provided text, the reported performance metrics are presented as the results that demonstrate substantial equivalence.

Performance MetricAcceptance Criteria (Implied by Study Design & Comparison)Reported Device Performance
Analytical Studies
Limit of Detection (LoD)Lowest concentration of B. pertussis that can be reproducibly distinguished from negative samples with 95% confidence (19/20 replicates positive).Between 1.6 x 10^3 and 3.3 x 10^3 CFU/mL, with an average LoD of 2.4 x 10^3 CFU/mL. Ranges for individual strains: - ATCC 8467: 3.3 x 10^3 CFU/mL (20/20 detected)- ATCC 9797: 1.6 x 10^3 CFU/mL (20/20 detected)- ATCC BAA-589: 2.3 x 10^3 CFU/mL (19/20 detected)
Analytical Reactivity (Inclusivity)Correct detection of additional B. pertussis strains.All 8 additional B. pertussis strains tested were correctly detected (3/3 positive for each strain).
Analytical Specificity (Exclusivity)No cross-reactivity with non-target organisms commonly found in the human respiratory system, except for known IS481-positive Bordetella species, which should be noted as limitations.Most non-target organisms (48 bacteria, 20 viruses, 2 yeast, 1 human genomic DNA) showed no cross-reactivity (100% expected negative results for all tested).Observed cross-reactivity:- Bordetella bronchiseptica: 1 of 3 strains showed cross-reactivity (ATCC 4617).- Bordetella holmesii: All 3 strains showed cross-reactivity.- Bordetella hinzii: 1 of 2 strains showed cross-reactivity (ATCC 51784).These noted cross-reactivities for Bordetella species are consistent with the presence of the IS481 insertion sequence also detected by the test.
Microbial InterferenceNo interference (i.e., correct detection of B. pertussis) when B. pertussis is present with high concentrations of other organisms.83 of 84 tested organisms (48 bacteria, 19 viruses, 2 yeast, 14 Bordetella non-B. pertussis strains, 1 human genomic DNA) showed no interference (100% expected positive results). One exception: 1 replicate for M. tuberculosis initially reported 'NOT DETECTED', but subsequent retesting confirmed no interference.
Interfering SubstancesNo interference with common chemical substances found in patients with upper respiratory conditions.All 19 tested chemical substances showed no interference (100% expected positive and negative results).
Carry-over/Cross-ContaminationNo false positives in negative samples following high positive samples.No carry-over or cross-contamination observed (100% correct results for alternating high positive and negative samples).
Swab, Transport Media, Elution Buffer EquivalencyDevice compatibility with various swab types, transport media, and elution buffers.All tested swab types (Flocked Nylon, Polyester, Rayon), transport media (Remel M4/M4RT/M5/M6 VTM, BD Universal VTM, ESwab), and elution buffers (Molecular grade water, PBS, Saline, TE Buffer) showed no interference (3/3 expected results for each). Equivalence study confirmed 100% agreement when using different combinations of swab types and transport/elution media.
Sample Stability and StorageSpecimen stability for specified storage conditions (room temperature and refrigerated).100% agreement with expected results for samples stored up to 72 hours at room temperature and up to 168 hours (7 days) at 2-8°C, supporting specimen storage claims of 48 hours at room temperature or ≤120 hours at 2-8°C.
ReproducibilityConsistent results across different sites, operators, and reagent lots.100% agreement observed across replicates, runs, operators (6), and sites (3 clinical sites) for low positive, moderate positive, and negative samples.
Clinical Studies
Positive Percent Agreement (PPA)High agreement with the Reference NAAT for positive samples, demonstrating clinical sensitivity. (Implicitly, the confidence interval should be acceptable, likely meeting a lower bound, although not explicitly stated).Prospective Study (Fresh): 85.7% (95% CI: 65.4% - 95.0%) (18/21 true positives)Frozen Retrospective Study: 94.6% (95% CI: 86.1% - 98.3%) (56/59 true positives)
Negative Percent Agreement (NPA)High agreement with the Reference NAAT for negative samples, demonstrating clinical specificity. (Implicitly, the confidence interval should be acceptable, likely meeting a lower bound, although not explicitly stated).Prospective Study (Fresh): 99.6% (95% CI: 98.9% - 99.8%) (890/894 true negatives)Frozen Retrospective Study: 100.0% (95% CI: 94.3% - 100.0%) (63/63 true negatives)

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Study (Fresh Samples):
    • Sample Size: 915 samples (936 collected, 21 excluded due to improper storage or failed QC).
    • Data Provenance: U.S. clinical study sites (four external, geographically-diverse sites: Northwest, Southwest, Midwest, West). Samples collected prospectively from July 2016 to January 2017. These were excess, de-identified nasopharyngeal (NP) swab specimens submitted for standard of care B. pertussis testing.
  • Frozen Retrospective Study:
    • Sample Size: 122 samples (124 initially, 2 excluded due to failed QC).
    • Data Provenance: Archived frozen NP swab specimens from institutions across the U.S. (implied from the context of clinical studies and specimen collection methods), retrospective in nature. These were de-identified specimens previously characterized by a nucleic acid amplification test at their originating institution.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes a diagnostic test for Bordetella pertussis using molecular methods (NAAT). The "ground truth" for the clinical test sets (both prospective and retrospective) was established by comparison to an FDA-cleared Nucleic Acid Amplification Test (Reference NAAT).

  • No human experts (e.g., radiologists) were used to establish the ground truth for classification of positive/negative B. pertussis status in the clinical studies.
  • The ground truth relies on the performance of the established Reference NAAT.

4. Adjudication Method for the Test Set

  • For the clinical performance studies, discrepant results between the Great Basin Bordetella Direct Test and the initial Reference NAAT were investigated by testing in a second FDA-cleared NAAT (NAAT 2) which also detects Bordetella pertussis. This acts as an adjudication step.
  • The document states: "In total, there were six (6) false negative and four (4) false positive results. Two (2) of the six (6) false negatives were also negative by a second FDA cleared NAAT and two (2) of the four (4) false positives were also positive by the second FDA cleared NAAT." This indicates the NAAT 2 was used to help understand the nature of the discrepancies, but it does not specify if the NAAT 2 result was definitively taken as the final "truth" for the agreement calculations in all cases. Typically, in diagnostic studies, a third, independent method or a consensus of multiple reference methods is used for resolving discrepancies, but the exact arbitration rule (e.g., 2-out-of-3, or specific tie-breaking rules) is not detailed for the final truth determination for PPA/NPA beyond the investigation by NAAT 2.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test, specifically a nucleic acid amplification test (NAAT). The performance is measured by its ability to detect specific DNA directly from a patient sample, not by a human interpreting images or data to make a diagnosis. Therefore, comparative effectiveness with human readers (e.g., radiologists, pathologists) is not relevant for this type of device. The studies focused on its analytical performance and its agreement with a reference molecular diagnostic test.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the primary performance evaluation of the Great Basin Bordetella Direct Test is effectively "standalone" performance. This means the device itself, the automated PA500 Portrait Analyzer and its associated assay, is evaluated for its ability to produce positive or negative results. There is no human interpretation or intervention in the diagnostic output generation described beyond loading the sample into the automated system. The results are generated by the instrument's automated software.

7. The Type of Ground Truth Used

  • For the clinical studies, the ground truth for classifying samples as positive or negative for B. pertussis was established by a Reference FDA-cleared Nucleic Acid Amplification Test (NAAT). Discrepancies were investigated by a second FDA-cleared NAAT.
  • For the analytical studies (LoD, inclusivity, exclusivity, interference), the ground truth was established by known concentrations of cultured bacterial strains (CFU/mL) or genomic copies/TCID50 for viruses and other microbes, which were spiked into negative clinical matrix.

8. The Sample Size for the Training Set

  • The provided document is a 510(k) premarket notification summary for a test that is likely based on traditional molecular biology techniques (PCR) rather than a machine learning/AI algorithm that requires a separate training set.
  • Therefore, the concept of a "training set" in the context of machine learning, where an algorithm learns patterns from data, does not directly apply to the development and validation of this specific in vitro diagnostic device. Its "training" is more akin to traditional assay development and optimization to ensure primers, probes, and reaction conditions are specific and sensitive.
  • The document does not mention a machine learning component or a distinct "training set" data size.

9. How the Ground Truth for the Training Set was Established

  • As explained in point 8, the concept of a "training set" for a machine learning algorithm is not applicable here. The device's underlying technology is PCR amplification, which is based on established biological and chemical principles rather than machine learning from a specific training dataset. Development and optimization of such assays involve
    • Designing and testing primers/probes: Based on known genetic sequences of B. pertussis.
    • Optimizing reaction conditions: To achieve desired sensitivity and specificity using known positive and negative controls/samples.
    • Analytical validation: As described in the document (LoD, inclusivity, exclusivity) using characterized bacterial strains and other microbes.
  • The "ground truth" during the development and optimization phases would have been established by precisely characterized biological materials, such as pure cultures of B. pertussis at known concentrations, or panels of well-characterized positive and negative patient samples.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one in front of the other. The profiles are rendered in a simple, abstract style.

March 31, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Public Health Service

Great Basin Scientific, Inc. Suzette Chance 2441 S. 3850 West Suite 100 Salt Lake City, UT 84120

Re: K170284

Trade/Device Name: Great Basin Bordetella Direct Test Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OZZ Dated: December January 26, 2017 Received: January 30, 2017

Dear Ms. Chance:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

Page 2 - Ms. Chance

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Steven R. Gitterman -S

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170284

Device Name Great Basin Bordetella Direct Test

Indications for Use (Describe)

The Great Basin Bordetella Direct Test is a qualitative in vitro diagnostic test for the detection of Bordetella pertussis DNA from nasopharyngeal swab specimens obtained from patients suspected of having a respiratory tract infection attributable to B. pertussis.

The Bordetella Direct Test is performed on the PA500 Portrait Analyzer and utilizes PCR amplification of the insertion sequence IS481. The IS481 sequence is also found in other organisms including Bordetella holmesii or Bordetella bronchiseptica. Respiratory infection with B. pertussis, B. holmesii ot B. bronchiseptica may yield positive test results with IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results for the Great Basin Bordetella Direct Test do not preclude B. pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the Great Basin Bordetella Direct Test should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of Bordetella pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows a pattern of black ovals arranged in a diagonal, stair-step fashion. The ovals are oriented horizontally and decrease in size as they ascend. The arrangement creates a visual effect of depth or progression.

5.0 510(k) Summary - Bordetella Direct Test

A. Submitted by:

Great Basin Corporation 2441 South 3850 West Salt Lake City, Utah 84120

Contact Information

Suzette Chance, PhD Senior Director of Clinical Affairs Great Basin Scientific 2441 S. 3850 West Salt Lake City, Utah 84120 Phone: 385-215-3369 Email: schance@gbscience.com

B. Name of Device

Proprietary Name: Great Basin Bordetella Direct Test Common or Usual Names: Bordetella Test

ﻦ Regulatory Information:

a. Regulation Section:21 CFR 866.3980, Respiratory viral panel Multiplex Nucleic Acid Assay21 CFR 862.2570 – Instrumentation for clinical multiplex test systems
b. Classification:Class II (Bordetella Direct Test)Class II (PA500 Portrait Analyzer System)
c. Classification panel:Product Code:Microbiology Devices, OIVD (83) MicrobiologyOZZ Bordetella pertussis Nucleic Acid Amplification Assay SystemOOI Real-Time Nucleic Amplification System

D. Intended use(s)/Indications for Use:

The Great Basin Bordetella Direct Test is a qualitative in vitro diagnostic test for the detection of Bordetella pertussis DNA from nasopharyngeal swab specimens obtained from patients suspected of having a respiratory tract infection attributable to B. pertussis.

The Bordetella Direct Test is performed on the PA500 Portrait™ Analyzer and utilizes PCR amplification of the insertion sequence IS481. The IS481 sequence is also found in other organisms including Bordetella holmesii or Bordetella bronchiseptica. Respiratory infection with B.pertussis, B. holmesii or B. bronchiseptica may vield positive test results in IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results for the Great Basin Bordetella Direct Test do not preclude B. pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the Great Basin Bordetella Direct Test should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of Bordetella pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows a triangular arrangement of black oval shapes. The ovals are arranged in rows, with each row containing one more oval than the row above it. The bottom row has two ovals, the next row has two ovals, and the top row has two ovals. The ovals are all the same size and shape.

E. Device Description:

Test Principle:

The Great Basin Bordetella Direct Test on the PA500 Portrait™ Analyzer System utilizes automated hot-start PCR technology to target and amplify the IS481 insertion sequence of B. pertussis. Genomic DNA is extracted from microbial cells and diluted to reduce potential inhibitors of PCR. During PCR, double-stranded DNA is separated and the target nucleic acid sequence is amplified by thermal cycling using biotin-labeled primers that target the IS481 sequence for identification of B. pertussis. Following PCR, biotin-labeled amplicon is hybridized to sequencespecific capture probes immobilized on the silicon chip surface, then incubated with anti-biotin antibody conjugated to the horseradish peroxidase enzyme (HRP). The unbound conjugate is washed away and tetramethylbenzidine (TMB) is added to produce a visible precipitate at the location of the probe/target sequence complex. The resulting signal is detected by the automated Portrait™ Optical Reader within the PA500 Portrait™ Analyzer System. The Specimen Processing Control (SPC), undergoes the extraction, and detection steps to monitor for inhibitory substances as well as process inefficiency due to instrument or reagent failure. No operator intervention is necessary once the clinical sample is loaded into the sample port and the Bordetella Direct Test cartridge is loaded into the Portrait Analyzer.

Test Device:

The PA500 Portrait™ Analyzer System is a fully automated system that includes: The Portrait™ Analyzer, single-use Bordetella Direct Test Cartridges, and the Portrait™ Data Analysis Software Program. The Portrait™ System is designed to perform automated sample preparation, PCR, and optical chip-based detection with integrated data analysis in less than two hours. The Portrait System was granted 510(k) clearance for the Portrait Toxigenic C. difficile Assay (K113358), Portrait GBS Assay (K143312), Staph ID/R Blood Culture Panel (K152470) and the Shiga Toxin Direct Test (K152955).

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows a series of black ovals arranged in a triangular pattern. The ovals are oriented horizontally, and their size decreases as they ascend in the pattern. The base of the triangle consists of two large ovals, and the apex consists of two small ovals. There are a total of 10 ovals in the image.

F. Substantial Equivalence Information:

Predicate Device: illumigene® Pertussis DNA Amplification Assay (K133673) The following table provides a comparison of the Bordetella Direct Test and the predicate device:

Feature/CharacteristicBordetella Direct Test (K170284)Predicate (K133673)
ManufacturerGreat Basin Scientific, Inc.Illumigene® Pertussis AssayMeridian Biosciences, Inc.
Trade NameGreat Basin Bordetella Direct Testillumigene® Pertussis DNA Amplification Assay
510(k) NumberK133673
ClassificationClass IISame
Intended Use/ Indicationsfor UseThe Great Basin Bordetella Direct Test is aqualitative in vitro diagnostic test for thedetection of Bordetella pertussis DNA fromnasopharyngeal swab specimens obtainedfrom patients suspected of havingrespiratory tract infection attributable to B.pertussis .The Bordetella Direct Test is performed onthe PA500 Portrait™ Analyzer and utilizesPCR amplification of the insertionsequence IS481. The IS481 sequence isalso found in other organisms includingBordetella holmesii or Bordetellabronchiseptica . Respiratory infection withB.pertussis , B. holmesii or B.bronchiseptica may yield positive testresults in IS481 assays. B. holmesiiinfection may cause clinical illness similarto B. pertussis , and mixed outbreaksinvolving both B. pertussis and B. holmesiiinfection have been reported. Additionaltesting should be performed if necessaryto differentiate B. holmesii and B.pertussis . B. bronchiseptica is a rarecause of infection in humans. Whenclinical factors suggest that B. pertussismay not be the cause of respiratoryinfection, other clinically appropriateinvestigation(s) should be carried out inaccordance with published guidelines.Negative results for the Great BasinBordetella Direct Test do not preclude B.pertussis infection and positive results donot rule out co-infection with otherrespiratory pathogens. Results from theGreat Basin Bordetella Direct Test shouldbe used in conjunction with informationobtained during the patient's clinicalevaluation as an aid in diagnosis ofBordetella pertussis infection and shouldnot be used as the sole basis for treatmentor other patient management decisions.The illumigene Pertussis DNAAmplification Assay, performed on theillumipro-10™ , is a qualitative in vitrodiagnostic test for the direct detection ofBordetella pertussis in humannasopharyngeal swab samples taken frompatients suspected of having respiratorytract infection attributable to Bordetellapertussis .The illumigene Pertussis assay utilizesloop-mediated isothermal DNAamplification (LAMP) technology to detectB. pertussis by targeting the IS481insertional element of the B. pertussisgenome. The IS481 insertional elementcan also be found in B. holmesii and someB. bronchiseptica strains. Respiratoryinfections with B. pertussis , B. holmesii orB. bronchiseptica may yield positive testresults in IS481 assays. B. holmesiiinfection may cause clinical illness similarto B. pertussis , and mixed outbreaksinvolving both B. pertussis and B. holmesiiinfection have been reported. Additionaltesting should be performed if necessaryto differentiate B. holmesii and B.pertussis . B. bronchiseptica is a rarecause of infection in humans. Whenclinical factors suggest that B. pertussismay not be the cause of respiratoryinfection, other clinically appropriateinvestigation(s) should be carried out inaccordance with published guidelines.Negative results for the illumigenePertussis DNA Amplification Assay do notpreclude Bordetella pertussis infection andpositive results do not rule out co-infectionwith other respiratory pathogens. Resultsfrom the illumigenePertussis assay should be used inconjunction with information obtainedduring the patient's clinical evaluation asan aid in diagnosis of B. pertussis infectionand should not be used as the sole basisfor treatment or other patientmanagement decisions.illumigene Pertussis is intended for use inhospital, reference or state laboratorysettings. The device is not intended forpoint-of-care use.
Feature/CharacteristicBordetella Direct TestPredicate (K133673)Illumigene® Pertussis Assay
Qualitative/QuantitativeQualitativeSame
Test CartridgeDisposable, single-use, self-containedfluidic cartridgeillumigene® Test Device with TEST andCONTROL chambers
Specimen TypeNasopharyngeal swab (rayon, nylon orflocked)Same
Organism DetectionB. pertussisSame
CalibrationNot requiredSame
Target SequenceDetectedIS481Same
Sample Lysis and DNAExtractionAutomatedManual
DNA AmplificationTechnologyPCRLoop-mediated amplification (LAMP)
Detection TechnologyColorimetric target specific hybridization toprobe on a chip surface, optical reader,automated software with built-in resultinterpretationMeasures the change in light transmissiondue to turbidity formation which is a by-product of the amplification reaction.
ControlsOne internal assay control - SPC (wholeorganism). The SPC controls for allanalytical steps in the procedure, includingDNA extraction from organisms present inthe specimen, amplification of target DNAsequences, hybridization, and detection onthe chip surface.One internal control ( S. aureus DNA) tomonitor amplification inhibition, assayreagent performance and sampleprocessing effectiveness.
InstrumentPA500 PortraitTM Analyzer SystemIllumipro-10TM
Testing Time~100 minutes60 - 70 minutes

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are arranged in rows, with each row containing a different number of ovals. The bottom row contains two ovals, the next row contains two ovals, the next row contains two ovals, and the top row contains two ovals. The ovals are all the same size and shape.

The following summarizes the differences and similarities between the Bordetella Direct Test and the predicate device:

The Bordetella Direct Test is similar to the illumigene® Pertussis Assay in the following ways:

  • . The Bordetella Direct Test and the illumigene® Pertussis Assay have similar Intended Uses and both detect the B. pertussis analyte.
  • The Bordetella Direct Test and the illumigene® Pertussis Assay are both qualitative, ● automated nucleic acid based tests.
  • The Bordetella Direct Test and the illumigene® Pertussis Assay both target the IS481 DNA . sequence of the B. pertussis genome.
  • The Bordetella Direct Test and the illumigene® Pertussis Assay use the same sample type ● (nasopharyngeal swab – rayon, nylon or flocked).

The Bordetella Direct Test differs from the illumigene® Pertussis Assay in the following:

  • The Bordetella Direct Test uses automated sample lysis and DNA extraction whereas these . are manual steps in the illumigene® Pertussis Assay.
  • . The Bordetella Direct Test uses hot start PCR for nucleic acid amplification whereas the illumigene® Pertussis Assay used Isothermal Loop-mediated amplification (LAMP).

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are arranged in rows, with each row containing fewer ovals than the row below it. The bottom row contains two ovals, and the top row contains two ovals. The ovals are all the same size and shape.

G. Performance Summary - Analytical Studies

a. Analytical Sensitivity

The Analytical Sensitivity as defined by the limit of detection (LoD) of the Bordetella Direct Test was assessed by testing three (3) B. pertussis strains: ATCC 9797 and ATCC BAA-589. The limit of detection (LoD) is defined as the lowest number of colony forming units (CFU) that can be reproducibly distinguished from negative samples with 95% confidence, or the lowest concentration at which 19/20 replicates are positive.

The B. pertussis strains were prepared using fresh isolates from Bordet-Gengou (BG) agar plates incubated for 3-5 days at 37°C until colony isolates were visible. The colony isolates were suspended and vortexed in Mueller Hinton II Broth (MH II) to create a uniform cell suspension with an optical density (OD600 nm) of 0.05, then serially diluted in MH II to a targeted cell concentration based on the OD600 nm value. The targeted cell concentration was spiked into natural negative nasopharyngeal (NP) matrix collected in Viral Transport Media (VTM) for testing. The tested sample cell concentrations (CFU/mL) were determined by plating and enumerating on BG agar plates. The LoDs for each strain tested is shown in Table 1.

B. pertussis ATCCStrainBordetellapertussisDETECTEDLoD(CFU/mL)
846720/203.3 x 103
979720/201.6 x 103
BAA-58919/202.3 x 103

Table 1. Bordetella Direct Test Limit of Detection (LoD)

Conclusion: The established LoD for the Bordetella Direct Test for the detection of B. pertussis is between 1.6 x 103 and 3.3 x 103 CFU/mL with an average LoD of 2.4 x 103 CFU/mL.

b. Analytical Reactivity (Inclusivity)

The Bordetella Direct Test Analytical Reactivity (Inclusivity) was confirmed by testing an additional eight (8) B. pertussis strains beyond the three (3) tested in the Analytical Sensitivity Study.

B. pertussis cultures were prepared and enumerated as described in the LoD Study. All testing was conducted in a natural negative NP matrix collected in VTM. Each strain was tested in triplicate at a targeted range of approximately 2-3X LoD based on the 2.4 x 10 CFU/mL average LoD obtained in the LoD study. The concentrations tested along with the results from the Inclusivity Study are shown in Table 2.

B. pertussisStrainConcentration tested(CFU/mL)Bordetella pertussisDETECTED
ATCC 8478$5.5 x 10^3$3/3
ATCC 9340$2.6 x 10^3$3/3
ATCC 10380$5.8 x 10^3$3/3
ATCC 12742$4.9 x 10^3$3/3
ATCC 51445$5.3 x 10^3$3/3
ATCC 53894$3.2 x 10^3$3/3
ATCC BAA-1335$4.3 x 10^3$3/3
Zeptometrix A639$1.8 x 10^4$3/3
$3.6 x 10^3$3/3

Table 2. Analytical Reactivity (Inclusivity) Panel Results

{8}------------------------------------------------

Conclusion: The Bordetella Direct Test correctly identified the additional eight (8) organisms tested in the Inclusivity Study indicating that the Bordetella Direct Test can detect additional strains of B. pertussis.

c. Analytical Specificity (Exclusivity)

The potential for cross-reactivity was evaluated in an Exclusivity Study, by testing non-target organisms commonly found in the human respiratory system in the Bordetella Direct Test. The study included 90 organisms including 48 bacteria, 20 viruses, 2 yeast, 19 Bordetella species (non-B. pertussis) and human genomic DNA. For those organisms that were classified as Biosafety level III, unavailable as intact organisms or difficult to culture via standard clinical microbiology techniques, genomic DNA was tested in place of whole organism.

Whole organism stocks were prepared at high concentrations in liquid media appropriate for each organism. The liguid media stocks were enumerated and frozen at ≤-20°C until testing. On the day of testing, each organism or nucleic acid stock was thawed and spiked into a natural negative NP matrix collected in VTM.

All bacterial and yeast strains were tested at concentrations ≥ 1 x 10° CFU/mL. Genomic DNA and viral strains were tested at ≥ 1.4 x 10° genomic copies/mL and ≥ 1.6 x 10° TCID50 respectively. A minimum of three (3) replicates were tested for each organism evaluated for cross-reactivity. The results from the testing, including specific concentration at which each organism was tested are shown in Tables 3 and 4.

OrganismStrain IDConcentration TestedExpected NegativeResults /Total
Bacteria
Acinetobacter baumanniiATCC 196061.2 x 107 CFU/mL6/6
Acinetobacter calcoaceticusATCC 230551.5 x 107 CFU/mL3/3
Acinetobacter haemolyticusATCC 190022.2 x 107 CFU/mL3/3
Actinomyces odontolyticusATCC 179291.5 x 106 CFU/mL3/3
Arcanobacterium haemolyticumATCC BAA-17842.4 x 107 CFU/mL3/3
Bacteroides fragilisATCC 237454.6 x 107 CFU/mL3/3
Bordetella aviumATCC 350862.7 x 107 CFU/mL1/3
1.2 x 107 CFU/mL3/3
1.2 x 107 CFU/mL10/10
Bordetella parapertussisATCC 93051.2 x 107 CFU/mL3/3
Bordetella parapertussisATCC 152371.7 x 107 CFU/mL3/3
Bordetella parapertussisATCC 153112.4 x 107 CFU/mL3/3*
Bordetella parapertussisATCC 159893.2 x 107 CFU/mL2/3
3.2 x 107 CFU/mL10/10
Bordetella parapertussisATCC 538921.4 x 107 CFU/mL3/3
Bordetella parapertussisATCC 538933.2 x 107 CFU/mL3/3
Bordetella parapertussisATCC BAA-5873.8 x 107 CFU/mL3/3
Bordetella parapertussisZeptometrix A7472.8 x 107 CFU/mL3/3
Bordetella petriiATCC BAA-4612.7 x 107 CFU/mL1/3
1.2 x 107 CFU/mL3/3
1.2 x 107 CFU/mL10/10*
Bordetella trematumATCC 7003092.0 x 107 CFU/mL0/3
1.3 x 107 CFU/mL3/3
1.3 x 107 CFU/mL10/10
Burkholderia cepaciaATCC 254163.1 x 107 CFU/mL3/3
Chlamydia trachomatisATCC VR-879D1.5 x 108 copies/mL3/3
Citrobacter freundiiATCC 80901.7 x 107 CFU/mL3/3
Clostridium difficileATCC 432551.5 x 107 CFU/mL3/3
OrganismStrain IDConcentration TestedExpected NegativeResults /Total
Corynebacterium diptheriaeATCC 138129.3 x 106 CFU/mL3/3
Enterobacter aerogenesATCC 150381.5 x 107 CFU/mL3/3
Enterobacter cloacaeATCC 130471.6 x 107 CFU/mL3/3
Enterococcus faecalisATCC 292128.6 x 106 CFU/mL3/3
Escherichia coliATCC 438951.4 x 107 CFU/mL3/3
Haemophilus influenzaATCC 90063.0 x 107 CFU/mL3/3
Haemophilus parainfluenzaATCC 333923.6 x 107 CFU/mL3/3
Klebsiella pneumoniaeATCC BAA-17052.1 x 107 CFU/mL3/3
Lactobacillus acidophilusATCC 43561.1 x 106 CFU/mL3/3
Lactobacillus plantarumATCC 80144.6 x 107 CFU/mL3/3
Legionella pneumophilaATCC 33152D-51.3 x 108 copies/mL3/3
Moraxella catarrhalisATCC 81769.8 x 106 CFU/mL3/3
Morganella morganiiATCC 258297.3 x 107 CFU/mL3/3
Mycobacterium tuberculosisATCC 25177D-24.1 x 107 copies/mL3/3
Mycoplasma pneumoniaeATCC 15531D2.0 x 107 copies/mL3/3
Neisseria gonorrhoeaeATCC 194242.2 x 107 CFU/mL3/3
Neisseria meningitidisATCC 130774.2 x 106 CFU/mL3/3
Pandoraea oxalativoransDSM-235701.6 x 107 CFU/mL3/3
5.3 x 107 CFU/mL10/10
Peptostreptococcus anaerobiusATCC 273375.9 x 106 CFU/mL3/3
Proteus mirabilisATCC 259338.4 x 107 CFU/mL3/3
Proteus vulgarisATCC 68962.2 x 107 CFU/mL3/3
Pseudomonas aeruginosaATCC 101454.3 x 107 CFU/mL3/3
Pseudomonas flourescensATCC 135251.8 x 107 CFU/mL3/3
Serratia liquefaciensATCC 355517.3 x 106 CFU/mL3/3
Staphylococcus aureus (MRSA)ATCC 335911.6 x 107 CFU/mL3/3
Staphylococcus aureus (MSSA)ATCC 259233.6 x 107 CFU/mL3/3
Staphylococcus epidermidis (MS)ATCC 122282.4 x 106 CFU/mL3/3
Stenotrophomonas maltophiliaATCC 136372.0 x 107 CFU/mL3/3
Streptococcus anginosusATCC 333979.2 x 106 CFU/mL3/3
Streptococcus bovisATCC 333171.2 x 107 CFU/mL3/3
Streptococcus dysgalactiae sspdysgalactiaeATCC 430781.5 x 107 CFU/mL3/3
Streptococcus dysgalactiae sspequisimilisATCC 356665.5 x 107 CFU/mL3/3
Streptococcus intermediusATCC 273353.4 x 107 CFU/mL3/3
Streptococcus mitisATCC 137702.3 x 106 CFU/mL3/3
Streptococcus mutansATCC 251752.5 x 107 CFU/mL3/3
Streptococcus pneumoniaeATCC 63031.3 x 106 CFU/mL3/3
Streptococcus pyogenesATCC 493991.4 x 107 CFU/mL3/3
Streptococcus salivariusATCC BAA-25931.6 x 107 CFU/mL3/3
Virus
AdenovirusATCC VR-846D8.5 x 109 copies/mL3/3
CoronavirusATCC VR-740D7.2 x 108 copies/mL3/3
CoxsackievirusATCC VR-1691.0 x 109 copies/mL3/3
CytomegalovirusATCC VR-538D2.5 x 109 copies/mL3/3
EchovirusATCC VR-1734D4.2 x 109 copies/mL3/3
Epstein-Barr VirusATCC VR-3247SD5.3 x 109 copies/mL3/3
Herpes Simplex Virus 1ATCC VR-539D1.0 x 106 copies/mL3/3*
Herpes Simplex Virus 2ATCC VR-540D1.6 x 106 TCID503/3
Human BocavirusATCC VR-3251SD4.8 x 108 copies/mL3/3
Human MetapneumovirusATCC VR-3250SD6.2 x 108 copies/mL3/3*
Influenza AATCC VR-1738D3.9 x 109 copies/mL5/5**
Influenza BATCC VR-1813D1.2 x 1010 copies/mL3/3
Measles virusATCC VR-24D1.6 x 1010 copies/mL3/3
Mumps virusATCC VR-106D1.3 x 1010 copies/mL3/3
Parainfluenza virus 1ATCC VR-94D1.8 x 109 copies/mL3/3
Parainfluenza virus 2ATCC VR-92D3.4 x 109 copies/mL3/3
OrganismStrain IDConcentration TestedExpected NegativeResults /Total
Respiratory Syncytial VirusATCC VR-1540D$3.0 x 10^9$ copies/mL3/3
RhinovirusATCC MBC091$1.4 x 10^6$ copies/mL3/3
Varicella Zoster VirusVR-1367D$1.0 x 10^9$ copies/mL3/3
Yeast
Candida albicansATCC 18804$1.1 x 10^7$ CFU/mL3/3
Candida glabrataATCC 66032$1.3 x 10^7$ CFU/mL3/3†
Human Genomic DNA
Human Genomic DNAATCC HTB-20D$3.8 x 10^7$ copies/mL3/3
*This set of test runs also contained one 'Test incomplete' run.†This set of test runs also contained one 'Invalid' run.

Table 3. Analytical Specificity (Exclusivity) Study Results

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows a series of black ovals arranged in a pyramid-like formation. The ovals are stacked on top of each other, with the base of the pyramid consisting of two larger ovals. As the pyramid ascends, the ovals become smaller and fewer in number, culminating in two small ovals at the apex. The ovals are solid black and have a smooth, rounded shape.

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are arranged in rows, with each row containing one more oval than the row above it. The bottom row has two ovals, the next row has two ovals, and so on, until the top row has two ovals. The ovals are all the same size and shape.

The Bordetella Direct Test correctly reported 'Bordetella pertussis NOT DETECTED' for all replicates reported in Table 3 as expected, with the following exceptions:

Unexpected 'Bordetella pertussis DETECTED' results were initially observed for one (1) of two (2) replicates for B. parapertussis (ATCC 15989), two (2) of three (3) replicates for B. avium (ATCC 35086) and B. petrii (ATCC BAA-461), and three (3) replicates for B. trematum (ATCC 700309). For these unexpected positives, an additional three (3) replicates were tested with a new preparation to test the veracity of the positive results. When the retesting gave the expected negative result for all three (3) replicates instead of the positive result seen with the first test, a minimum of ten (10) more replicates were tested to confirm the negative result. All testing done with new preparations yielded 'Bordetella pertussis NOT DETECTED' results. The negative results from new preparations suggest original positive results were likely due to sample contamination. These data demonstrate that B. avium, B. petrii, B. trematum, and B. parapertussis do not cross-react in the Bordetella Direct Test.

Testing for analytical specificity was also performed on three Bordetella (non-B. perfussis) species identified as potentially cross-reacting: Bordetella bronchiseptica, Bordetella holmesii, and Bordetella hinzii. Species were defined as potentially cross-reacting based on the IS481 insertion sequence reported in a subset of strains for these species. The results are summarized in Table 4.

Potential Cross-ReactingOrganismStrain IDConcentration TestedPositiveResults/Total
Bordetella bronchisepticaATCC 193952.1 x 10' CFU/mL0/3
Bordetella bronchisepticaATCC 46171.2 x 10' CFU/mL3/3
Bordetella bronchisepticaATCC 46172.8 x 10' CFU/mL3/3
Bordetella bronchisepticaATCC BAA-5883.1 x 10' CFU/mL2/3
Bordetella bronchisepticaATCC BAA-5883.0 x 10' CFU/mL0/3
Bordetella bronchisepticaATCC BAA-5883.0 x 10' CFU/mL0/10
Bordetella holmesiiATCC 7000521.4 x 10' CFU/mL3/3*
Bordetella holmesiiATCC 7000522.1 x 10' CFU/mL3/3
Bordetella holmesiiATCC 7000532.0 x 10' CFU/mL3/3
Bordetella holmesiiATCC 7000531.1 x 10' CFU/mL3/3
Bordetella holmesiiATCC 515412.1 x 10' CFU/mL3/3
Bordetella hinziiATCC 517303.0 x 10' CFU/mL0/3
Bordetella hinziiATCC 517843.3 x 10' CFU/mL3/3
Bordetella hinziiATCC 517843.6 x 10' CFU/mL3/3
*This set of test runs also contained one 'Test incomplete' run.

Table 4. Analytical Specificity (Exclusivity) Results for Potential Cross-Reacting Strains

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are arranged in rows, with each row containing fewer ovals than the row below it. The ovals are all the same size and shape, and they are evenly spaced apart. The pattern is simple and geometric, and it is visually appealing.

For B. bronchiseptica, three strains were tested: ATCC 19395, ATCC 4617, and ATCC BAA-588 (Table 4). For ATCC 19395, all three (3) replicates gave a 'B. pertussis NOT DETECTED' result, demonstrating no cross-reactivity. For ATCC 4617, all three (3) replicates gave a result of 'B. pertussis DETECTED.' A second preparation was tested and cross-reactivity of ATCC 4617 was confirmed. For ATCC BAA-588, two (2) out of three (3) replicates gave a result of 'B. pertussis DETECTED.' A second preparation was tested, and all three (3) results were 'B. pertussis NOT DETECTED.' Due to the confounding results, another ten replicates were tested, and 'B. pertussis NOT DETECTED' results were obtained for all replicates confirming no cross-reactivity for ATCC BAA-588. In total, cross-reactivity was observed in one (1) of three (3) B. bronchiseptica strains.

For B. holmesii, three strains were tested: ATCC 700052, ATCC 700053, and ATCC 51541 (Table 4). For both ATCC 700052 and ATCC 700053, all three (3) replicates gave a result of 'B. pertussis DETECTED.' Cross-reactivity was confirmed upon retest for both ATCC 700052 and 700053 with three (3) replicates each. For ATCC 51541. the result for all three (3) replicates was also 'B. pertussis DETECTED," agreeing with the previous B. holmesii results. In total, cross-reactivity was observed for all three (3) B. holmesii strains.

For B. hinzii, two strains were tested: ATCC 51730 and ATCC 51784 (Table 4). For ATCC 51730, all three (3) replicates gave a 'B. pertussis NOT DETECTED' result, as expected from the in silico analysis. For ATCC 51784, the first test triplicate gave 'B. pertussis DETECTED' results. Since this was unexpected, repeat testing was performed in triplicate on ATCC 51784, and the retest results confirmed the 'B. pertussis DETECTED' results. Further investigation revealed that this particular B. hinzii strain (ATCC 51784), isolated from chickens and a known avian pathogen (Register, KB, Kunkle RA, Avian Diseases. 53:50-54, 2009), has been reported in the literature to have a weak IS481positive signal (Roorda et. al., BMC Res Notes. 4:11, 2011), Therefore, the observed cross-reactivity is consistent with the presence of the IS481 insertion sequence. Specific in silico analysis on this strain could not be performed because an accession number does not exist in the NCBI database.

Further investigation of the negative-resulting B. hinzii strain (ATCC 51730) revealed that this a clinical isolate from a blood sample of an AIDS patient (isolated from human blood, AIDS patient; Vandamme et al., Int J. Syst. Bacteriol. 45: 37- 45, 1995). The entire sequenced genome of this isolate is found in the NCBI database (strain F582, accession number CP012076). allowing for specific in silico analysis, In silico analysis demonstrated partial homology with the reverse primer, but no homology with the forward primer, consistent with the negative result.

In summary, cross-reactivity was observed for one (1) of three (3) B. bronchiseptica strains (ATCC 4617), three (3) of three (3) B. holmesii strains (ATCC 700052, 700053, and 51541), and one (1) of two (2) B. hinzii strains (ATCC 51784). These results are likely due to the presence of the IS481 insertion sequence reported in a subset of strains for these species. This is noted in Section 11 for B. bronchiseptica and B. holmesii (in silico Analysis for Bordetella Direct Test) and in the literature for B. hinzii ATCC 51784, a pathogen which mainly infects avian species.

Conclusion: All organisms shown in Table 3 gave the expected 'B. pertussis NOT DETECTED' result indicating that there was no cross-reactivity in the Bordetella Direct Test. Organisms known to potentially contain the IS481 sequence showed some cross-reactivity in the Bordetella Direct Test (Table 4), including B. bronchiseptica, B. holmesii, and B. hinzii. This cross-reactivity is listed as a limitation in the product labeling (Section 13), with the exception of B. hinzii which is mainly an avian pathogen.

{12}------------------------------------------------

Image /page/12/Picture/2 description: The image shows a pattern of black ovals arranged in a step-like formation. The ovals are oriented horizontally, and their size decreases as they ascend the steps. The arrangement creates a visual effect of depth and progression, with the larger ovals at the bottom providing a base for the smaller ones above.

d. Microbial Interference

The potential for cross-reactivity in a mixed infection was evaluated in a Microbial Interference Study by testing a subset of the organisms used in the Exclusivity Study in the Bordetella Direct Test in the presence of B. pertussis.

Similar to the Exclusivity Study, bacteria, virus, yeast and human genomic DNA were prepared at high concentrations (> 1 x 10° CFU/mL, genomic copies/mL or TCID50) in a natural negative NP matrix collected in VTM. Samples were prepared by spiking B. pertussis (ATCC 9797) to a final concentration of 4.0 x 10° CFU/mL (2.5X LoD) in the presence of each high concentration non-B. pertussis organism or genomic DNA. A total of 84 organisms (48 bacteria, 19 viruses, 2 yeast, 14 Bordetella strains (non-B. pertussis) and human genomic DNA were evaluated.

A minimum of three (3) replicates of each sample were tested. The specific concentrations at which each organism was evaluated along with the results are shown in Table 5.

Table 5. Microbial Interference Study Results: Organisms Tested in the Presence of 2.5X LoD B. pertussis

OrganismStrain IDNon-Target MicrobeInput TestedExpected PositiveResults/Total
Bacteria
Acinetobacter baumanniiATCC 196061.2 x 107 CFU/mL3/3
Acinetobacter calcoaceticusATCC 230551.5 x 107 CFU/mL3/3
Acinetobacter haemolyticusATCC 190022.2 x 107 CFU/mL3/3
Actinomyces odontolyticusATCC 179291.5 x 106 CFU/mL3/3
Arcanobacterium haemolyticumATCC BAA-17842.4 x 107 CFU/mL3/3
Bacteroides fragilisATCC 237454.6 x 107 CFU/mL3/3
Bordetella aviumATCC 350861.2 x 107 CFU/mL3/3
Bordetella bronchisepticaATCC BAA-5883.0 x 107 CFU/mL3/3
Bordetella bronchisepticaATCC 193952.1 x 107 CFU/mL3/3
Bordetella hinziiATCC 517303.0 x 107 CFU/mL3/3*
Bordetella parapertussisATCC 93051.2 x 107 CFU/mL3/3
Bordetella parapertussisATCC 152371.7 x 107 CFU/mL3/3
Bordetella parapertussisATCC 153112.4 x 107 CFU/mL3/3
Bordetella parapertussisATCC 159893.2 x 107 CFU/mL3/3
Bordetella parapertussisATCC 538921.4 x 107 CFU/mL3/3
Bordetella parapertussisATCC 538933.2 x 107 CFU/mL3/3
Bordetella parapertussisATCC BAA-5873.8 x 107 CFU/mL3/3*
Bordetella parapertussisZeptometrix A7472.8 x 107 CFU/mL3/3
Bordetella petriiATCC BAA-4611.2 x 107 CFU/mL3/3
Bordetella trematumATCC 7003091.3 x 107 CFU/mL3/3
Burkholderia cepaciaATCC 254163.1 x 107 CFU/mL3/3
Chlamydia trachomatisATCC VR-879D1.5 x 108 copies/mL3/3
Citrobacter freundiiATCC 80901.7 x 107 CFU/mL3/3
Clostridium difficileATCC 432551.5 x 107 CFU/mL3/3
Corynebacterium diptheriaeATCC 138129.3 x 106 CFU/mL3/3
Enterobacter aerogenesATCC 150381.5 x 107 CFU/mL3/3
Enterobacter cloacaeATCC 130471.6 x 107 CFU/mL3/3
Enterococcus faecalisATCC 292128.6 x 106 CFU/mL3/3*
Escherichia coliATCC 438951.4 x 107 CFU/mL3/3
Haemophilus influenzaATCC 90063.0 x 107 CFU/mL3/3
Haemophilus parainfluenzaATCC 333923.6 x 107 CFU/mL3/3
Klebsiella pneumoniaeATCC BAA-17052.1 x 107 CFU/mL3/3
Lactobacillus acidophilusATCC 43561.1 x 106 CFU/mL3/3
Lactobacillus plantarumATCC 80144.6 x 107 CFU/mL3/3
Legionella pneumophilaATCC 33152D-51.3 x 108 copies/mL3/3
Moraxella catarrhalisATCC 81769.8 x 106 CFU/mL3/3
Morganella morganiiATCC 258297.3 x 107 CFU/mL3/3
Mycobacterium tuberculosisATCC 25177D-24.1 x 107 copies/mL13/14
Mycoplasma pneumoniaeATCC 15531D2.0 x 107 copies/mL3/3
OrganismStrain IDNon-Target MicrobeInput TestedExpected PositiveResults/Total
Neisseria gonorrhoeaeATCC 194242.2 x 107 CFU/mL3/3
Neisseria meningitidisATCC 130774.2 x 106 CFU/mL3/3
Pandoraea oxalativoransDSM-235705.3 x 107 CFU/mL3/3
Peptostreptococcus anaerobiusATCC 273375.9 x 106 CFU/mL3/3
Proteus mirabilisATCC 259338.4 x 107 CFU/mL3/3
Proteus vulgarisATCC 68962.2 x 107 CFU/mL3/3
Pseudomonas aeruginosaATCC 101454.3 x 107 CFU/mL3/3
Pseudomonas flourescensATCC 135251.8 x 107 CFU/mL3/3
Serratia liquefaciensATCC 355517.3 x 106 CFU/mL3/3
Staphylococcus aureus (MRSA)ATCC 335911.6 x 107 CFU/mL3/3
Staphylococcus aureus (MSSA)ATCC 259233.6 x 107 CFU/mL3/3
Staphylococcus epidermidis (MS)ATCC 122282.4 x 106 CFU/mL3/3
Stenotrophomonas maltophiliaATCC 136372.0 x 107 CFU/mL3/3
Streptococcus anginosusATCC 333979.2 x 106 CFU/mL3/3*
Streptococcus bovisATCC 333171.2 x 107 CFU/mL3/3
Streptococcus dysgalactiae sspdysgalactiaeATCC 430781.5 x 107 CFU/mL3/3
Streptococcus dysgalactiae sspequismilisATCC 356665.5 x 107 CFU/mL3/3
Streptococcus intermediusATCC 273353.4 x 107 CFU/mL3/3
Streptococcus mitisATCC 137702.3 x 106 CFU/mL3/3
Streptococcus mutansATCC 251752.5 x 107 CFU/mL3/3
Streptococcus pneumoniaeATCC 63031.3 x 106 CFU/mL3/3*
Streptococcus pyogenesATCC 493991.4 x 107 CFU/mL3/3
Streptococcus salivariusATCC BAA-25931.6 x 107 CFU/mL3/3
Virus
AdenovirusATCC VR-846D8.5 x 109 copies/mL3/3
CoronavirusATCC VR-740D7.2 x 108 copies/mL3/3
CoxsackievirusATCC VR-1691.0 x 107 copies/mL3/3
CytomegalovirusATCC VR-538D2.5 x 109 copies/mL3/3
EchovirusATCC VR-1734D4.2 x 107 copies/mL3/3
Epstein-Barr VirusATCC VR-3247SD5.3 x 107 copies/mL3/3
Herpes Simplex Virus 1ATCC VR-539D1.0 x 109 copies/mL3/3
Herpes Simplex Virus 2ATCC VR-540D1.6 x 106 TCID503/3
Human BocavirusATCC VR-3251SD4.8 x 107 copies/mL3/3
Human MetapneumovirusATCC VR-3250SD6.2 x 107 copies/mL3/3
Influenza AATCC VR-1738D3.9 x 109 copies/mL3/3
Influenza BATCC VR-1813D1.2 x 1010 copies/mL3/3
Measles virusATCC VR-24D1.6 x 1010 copies/mL3/3
Mumps virusATCC VR-106D1.3 x 1010 copies/mL3/3
Parainfluenza virus 1ATCC VR-94D1.8 x 109 copies/mL3/3
Parainfluenza virus 2ATCC VR-92D3.4 x 109 copies/mL3/3
Respiratory Syncytial VirusATCC VR-1540D7.9 x 108 copies/mL3/3
RhinovirusMBC0913.0 x 109 copies/mL3/3
Varicella Zoster VirusVR-1367D1.4 x 106 copies/mL6/6
Yeast
Candida albicansATCC 188041.1 x 107 CFU/mL3/3
Candida glabrataATCC 660321.3 x 107 CFU/mL3/3
Human Genomic DNA
Human Genomic DNAATCC HTB-20D3.8 x 107 copies/mL3/3
* This set of test runs also contained one 'Test incomplete' run.

{13}------------------------------------------------

Image /page/13/Picture/2 description: The image shows a pattern of black ovals arranged in a step-like formation. The ovals are oriented horizontally, and their size decreases as they ascend. The bottom row consists of two large ovals, followed by a row of two medium-sized ovals, and finally, two small ovals at the top.

All organisms shown in Table 5 gave the expected 'Bordetella pertussis DETECTED' result demonstrating that there was no interference in the Bordetella Direct Test, with one exception. One (1) of three (3) replicates tested with M. tuberculosis reported 'Bordetella pertussis as 'NOT DETECTED'. Additional testing with 11 replicates yielded the expected 'Bordetella pertussis DETECTED' results, demonstrating that M. tuberculosis does not interfere with the Bordetella Direct Test.

{14}------------------------------------------------

Image /page/14/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are arranged in rows, with each row containing one less oval than the row below it. The bottom row contains two ovals, and the top row contains two ovals. The ovals are all the same size and shape.

Conclusion: No interference in the Bordetella Direct Test was observed with the organisms tested in Table 5 at the concentrations indicated.

e. Interfering Substances

A panel of 19 chemical substances commonly used or present in patients with upper respiratory conditions were evaluated for interference with the Bordetella Direct Test.

Samples were prepared by mixing B. pertussis (ATCC 9797) at 4.0 x103 CFU/mL (2.5X LoD) contrived in a natural negative NP matrix collected in VTM with each substance. A clinical negative sample was also tested as a control to evaluate potential interference of each substance with the internal Sample Processing Control (SPC) in the absence of analyte. Each sample was tested in triplicate. The chemical substances, test concentration (mg/mL or %), and results are listed in Table 6.

Expected Results/Total
Interfering SubstanceConcentrationTestedB. pertussisDETECTED(2.5X LoD)Negative
Afrin Nasal Spray15% v/v3/33/3
Cepacol Sore Throat Pain Relief Lozenges5% w/v3/33/3
Children's Dimetapp Cold & Allergy15% v/v3/33/3
Chloraseptic Max Sore Throat Lozenges10% w/v3/33/3
Diphenhydramine HCl, 25mg1 mg/mL3/33/3
Erythromycin20 mg/mL3/33/3
Flonase Nasal Spray15% v/v3/33/3
Mucin (Bovine Submaxillary Gland, Type I-S)10 mg/mL3/33/3
Mupirocin10 mg/mL3/33/3
NasalCrom Nasal Spray15% v/v3/33/3
Vaseline Petroleum Jelly1% w/v3/33/3
Releev Cold Sore Treatment15% v/v3/33/3
Robitussin Cough Syrup5% v/v3/33/3
Saline Nasal Moisturizing Spray15% v/v3/33/3
Sucrets Complete Lozenges5% w/v3/33/3
Tobramycin1 mg/mL3/33/3
Vicks Vapor Rub1% w/v3/33/3
Blood (human), EDTA anticoagulated5% v/v3/33/3
Zicam Nasal Gel5% v/v3/3*3/3
*This set of test runs contained one 'Test incomplete' run.

Table 6. Interfering Substances Study Results

Conclusion: No interference in the Bordetella Direct Test was observed for the 19 substances tested at the concentrations shown in Table 6.

f. Carry-over/Cross-Contamination

The Bordetella Direct Test was evaluated for carry-over/cross-contamination by testing a series of alternating high positive and negative samples for 20 consecutive runs on 2 separate Analyzers for a total of 40 runs. The high positive and negative samples were prepared as follows:

  • High Positive: B. pertussis 1.55 x107 CFU/mL in natural negative NP matrix collected in VTM ●
  • Neqative: natural negative NP matrix collected in VTM ●

{15}------------------------------------------------

Image /page/15/Picture/2 description: The image shows a series of black oval shapes arranged in a pattern. The ovals are organized in rows, with each row containing a different number of ovals. The number of ovals decreases as you move up the rows, creating a pyramid-like structure. The ovals are all the same size and shape, and they are evenly spaced within each row.

The results are shown in Table 7.

Bordetella pertussis Result
RunsSamplePortrait Analyzer 5.075Portrait Analyzer 5.435
1High PositiveDETECTEDDETECTED
2NegativeNOT DETECTEDNOT DETECTED
3High PositiveDETECTEDDETECTED
4NegativeNOT DETECTEDNOT DETECTED
5High PositiveDETECTEDDETECTED
6NegativeNOT DETECTEDNOT DETECTED
7High PositiveDETECTEDDETECTED
8NegativeNOT DETECTEDNOT DETECTED
9High PositiveDETECTEDDETECTED
10NegativeNOT DETECTEDNOT DETECTED
11High PositiveDETECTEDDETECTED
12NegativeNOT DETECTEDNOT DETECTED
13High PositiveDETECTEDDETECTED
14NegativeNOT DETECTEDNOT DETECTED
15High PositiveDETECTEDDETECTED
16NegativeNOT DETECTEDNOT DETECTED
17High PositiveDETECTEDDETECTED
18NegativeNOT DETECTEDNOT DETECTED
19High PositiveDETECTEDDETECTED
20NegativeNOT DETECTEDNOT DETECTED
Total Runs Per Device20
Total Runs for Study40
Table 7. Carry-over/Cross-contamination Study Results
-------------------------------------------------------------

Conclusion: The expected results were obtained in each run. No carry-over or cross- contamination was observed in the Bordetella Direct Test.

g. Swab, Transport Media and Elution Buffer Equivalency Studies

Two studies were conducted to demonstrate the equivalency between various swab types, transport media types and elution buffers. In the first study, various types of swabs, transport media and elution buffers were tested for potential interference in the Bordetella Direct Test. To test the transport media and elution buffers, samples were prepared by mixing B. pertussis (ATCC 9797) at 4.0x103 CFU/mL (2.5X LoD) with each transport media or elution buffer. To test for interference of the swabs, each swab type was inserted into natural negative NP matrix that was spiked with B. pertussis (ATCC 9797) at 4.0x103 CFU/mL (2.5X LoD). A clinical negative sample was also tested as a control to evaluate potential interference of each substance with the internal assay control (SPC) in the absence of analyte. Each sample was tested in triplicate. The swab types, media and elution buffers tested along with the study results are shown in Table 8. Note: Copan UTM is the same transport media as BD Universal VTM and therefore was not also tested.

{16}------------------------------------------------

Image /page/16/Picture/2 description: The image shows a pattern of black ovals arranged in a step-like formation. The ovals are oriented horizontally, with each row containing a different number of ovals. The bottom row has the most ovals, and the top row has the fewest. The ovals are evenly spaced within each row, creating a visually balanced composition.

Table 8. Swab, Transport Media and Elution Buffer Interference Study Results

Expected Results/Total
Swab TypeB. pertussisDETECTED(2.5X LoD)Negative
Flocked Nylon3/33/3
Polyester3/33/3
Rayon3/33/3
Transport Media TypeB. pertussisDETECTED(2.5X LoD)Negative
Remel M4 VTM3/33/3
Remel M4RT VTM3/33/3
Remel M5 VTM3/33/3
Remel M6 VTM3/33/3
BD Universal VTM3/33/3
Elution Buffer TypeB. pertussisDETECTED(2.5X LoD)Negative
Molecular grade water3/33/3
1x Phosphate Buffered Saline (PBS)3/33/3
0.85% Saline3/33/3*
Tris-EDTA (TE) Buffer3/33/3
*This set of test runs contained one 'Test incomplete' run

Results: None of the swabs, transport media or elution buffers demonstrated interference in the Bordetella Direct Test.

In a separate study the Bordetella Direct Test was evaluated using the following 3 distinct combinations of nasopharyngeal swab types and transport/elution media:

  • Polyester swab with Viral Transport Media (M5 VTM) ●
  • Flocked nylon NP swab with Liquid Amies transport media (ESwab)
  • Rayon swab in Liquid Stuart transport media, eluted in water

Each swab/media combination listed above was collected as clinical negative NP matrix. B. pertussis strains (ATCC 8467, ATCC 9797, and ATCC BAA-589) were used to prepare contrived samples in the above 3 swab/media combinations. Positive samples were prepared by the addition of each B. pertussis strain at varying LoDs in clinical negative NP matrix. The negative clinical matrix served as the neqative sample. Each unique sample was tested in the Bordetella Direct Test. The test conditions and results are shown in Table 9.

Table 9. Swab, Transport Media and Elution Buffer Equivalence Study Results

Sample Type, Concentration, Expected Results/Total
Transport/ElutionMediaSwabB. pertussisATCC 8467(2.1X LoD)$6.9 x 10^3$CFU/mLB. pertussisATCC 9797(4.5X LoD)$7.3 x 10^3$CFU/mLB. pertussisATCC BAA-589(1.9X LoD)$4.3 x 10^3$ CFU/mLNegative
M5 VTM (3mL)Polyester3/33/33/33/3
ESwab (1mL)FlockedNylon3/33/33/33/3
Liquid Stuart /Eluted in water (1mL)Rayon3/33/33/33/3

{17}------------------------------------------------

There was 100% agreement with the expected results demonstrating equivalent Results: performance of the Bordetella Direct Test with various swab and media combinations tested.

Conclusion: The Bordetella Direct Test is compatible with the following types of swabs: Polyester, Rayon, Flocked Nylon and with the following type of transport media: VTM (M4, M4RT, M6), BD Universal VTM, Copan UTM, ESwab (Liquid Amies) and Liquid Stuart. Samples stored in VTM or ESwab are eluted in the same transport media. Samples stored in Liquid Stuart can be eluted with 1.0 mL of any of the following: Molecular Grade water, PBS, 0.8% Saline or Tris-EDTA (TE) buffer.

h. Sample Stability and Storage

Sample stability studies were conducted to support the storage conditions stated in the product labeling.

The B. pertussis positive and negative samples prepared in the second Swab. Transport Media and Elution Buffer Equivalency Study were also used to evaluate sample stability at room temperature (20 - 25°C) and at 2-8°C. Aliquots of each sample type were tested at the time of preparation (Ta), then stored at room temperature (RT) for 24, 48 and 72 hrs or refrigerated (2-8°C) for 24, 48, 72, 96 and 168 hrs. Samples were tested in triplicate at each time point. Results from the room temperature and refrigerated stability studies are listed in Tables 10 and 11, respectively.

Time Point (hrs), Expected Results/Total
Strain Tested andConcentrationMediaT 0T24T48T72
ATCC 8467(6.9 x 103 CFU/mL)M5 VTM3/33/33/33/3
ESwab3/33/33/33/3
Liquid Stuart3/33/33/33/3
ATCC 9797(7.3 x 103 CFU/mL)M5 VTM3/33/33/33/3
ESwab3/33/33/33/3
Liquid Stuart3/33/33/33/3
ATCC BAA-589(4.3 x 103 CFU/mL)M5 VTM3/33/33/33/3
ESwab3/33/33/33/3
Liquid Stuart3/33/33/33/3
NegativeM5 VTM3/33/33/33/3*
ESwab3/33/33/33/3
Liquid Stuart3/33/3+3/33/3
*This set of test runs also contains one 'Test incomplete' run.+This set of test runs also contains one 'Invalid' run.

Table 10. Sample Stability Study: Room Temperature Storage

{18}------------------------------------------------

Image /page/18/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are oriented horizontally and vary in size, with the largest ovals at the bottom and the smallest at the top. The ovals are arranged in rows, with each row containing a different number of ovals. The pattern is simple and geometric, with a clear sense of order and balance.

Table 11. Sample Stability Study: 2-8°C Storage
-----------------------------------------------------------
Strain Tested andConcentrationMediaTime Point (hrs), Expected Results/Total
T 0T 24T 48T 72T 96T 168
ATCC 8467(6.9 x 103 CFU/mL)M5 VTM3/33/33/33/33/33/3
ATCC 8467(6.9 x 103 CFU/mL)ESwab3/33/33/33/33/33/3
ATCC 8467(6.9 x 103 CFU/mL)Liquid Stuart3/33/33/33/33/33/3
ATCC 9797(7.3 x 103 CFU/mL)M5 VTM3/33/33/33/33/33/3
ATCC 9797(7.3 x 103 CFU/mL)ESwab3/33/33/33/33/33/3
ATCC 9797(7.3 x 103 CFU/mL)Liquid Stuart3/33/33/33/33/33/3
ATCC BAA-589(4.3 x 103 CFU/mL)M5 VTM3/33/33/33/33/33/3
ATCC BAA-589(4.3 x 103 CFU/mL)ESwab3/33/33/33/33/33/3
ATCC BAA-589(4.3 x 103 CFU/mL)Liquid Stuart3/33/33/33/3†3/33/3
NegativeM5 VTM3/33/33/33/33/33/3*
NegativeESwab3/33/33/33/33/33/3
NegativeLiquid Stuart3/33/33/33/32/2†3/3
*This set of test runs also contained one 'Test incomplete' run.†This set of test runs also contained one 'Invalid' run.

The results demonstrated 100% agreement with the expected results for all samples Conclusion: tested at each time point. These results support the specimen storage claims in the Product Insert of 48 hours at room temperature or ≤120 hours at 2-8°C.

Reproducibility i.

A Reproducibility Study was conducted at three clinical study sites. Two of the were external sites (Site 4: Providence Health Services Oregon and Site 5: Indiana University School of Medicine) and the third site was internal (Site 6: Great Basin Scientific). The study incorporated several variables including 6 different operators at 3 sites (2 operators/site), 33 different Portrait Analyzers and 3 different Cartridge lots.

A panel consisting of three (3) different samples: a low positive (B. pertussis at 1.9X LoD) a moderate positive (B. pertussis at 3.8X LoD) prepared in negative clinical matrix collected in VTM and a neqative sample (neqative clinical matrix collected in the Bordetella Direct Test. Each operator tested each sample in triplicate over five (5) non-consecutive days. Six (6) different operators performed the testing.

The results were analyzed for percent agreement of positive results, by operator, by site and in total (overall) and are shown in Table 12.

{19}------------------------------------------------

Image /page/19/Picture/2 description: The image shows a pattern of black ovals arranged in a step-like formation. The ovals are oriented horizontally, and their size decreases as they ascend the steps. The arrangement creates a visual effect of progression or a staircase, with each step consisting of one or two ovals.

AnalyteConc.SiteOperatorCorrectResultsAgreementCorrectResultsAgreementTotalCorrectResults
Bordetella pertussis Positive Samples1.9X LOD(Low Positive)4115/15100%30/30100%90/90100%
4215/15†100%
5315/15100%
5415/15*100%
6515/15100%
6615/15100%
Bordetella pertussis Positive Samples3.8X LOD(Moderate Positive)4115/15100%30/30100%90/90100%
4215/15100%
5315/15100%
5415/15*100%
6515/15100%
6615/15100%
Negative SampleN/A4115/15100%30/30100%90/90100%
4215/15*100%
5315/15100%
5415/15100%
6515/15100%
6615/15100%
*This set of test runs also contained one 'Test Incomplete' run.†This set of test runs also contained one 'Invalid' run.

Table 12. Reproducibility Study Results

Conclusion: There was 100% agreement of results between replicates, runs, operators and sites.

H. Performance Summary - Clinical Studies

The clinical performance of the Bordetella Direct Test was evaluated in a multi-site clinical study. The evaluation included a Prospective and a Frozen Retrospective method comparison study. These studies compared the performance of the Bordetella Direct Test to an FDA cleared Nucleic Acid Amplification Test for the detection of Bordetella pertussis (Reference NAAT). Discrepant results were investigated by testing in a second FDA cleared NAAT (NAAT 2) which also detects Bordetella pertussis.

The Prospective study was conducted at four external, geographically-diverse U.S. clinical study sites (Northwest, Southwest, Midwest, West) from July, 2016 to January, 2017. The specimens enrolled in the Prospective study were excess, de-identified nasopharyngeal (NP) swab specimens collected in VTM that were submitted for standard of care B. pertussis testing and which would have otherwise been discarded. A total of 936 samples were collected for all four sites combined. Subsequent to enrollment, 21 samples were excluded from the data set due to improperly stored sample or failed daily QC leaving 915 samples included in the analysis.

Samples used in the Frozen Retrospective Study were frozen archived NP swab specimens collected in VTM. The specimens were de-identified specimens that were previously characterized as positive or negative by a nucleic acid amplification test used at the particular institution from which they were obtained (historical result).

{20}------------------------------------------------

Image /page/20/Picture/2 description: The image shows a pattern of black ovals arranged in a triangular shape. The ovals are oriented horizontally and are stacked on top of each other, with the number of ovals decreasing as the stack gets higher. The bottom row has two ovals, the next row has two ovals, and the top row has two ovals.

The historical result for each sample was first confirmed by the Reference NAAT prior to enrolling the sample in the study. A total of 124 frozen samples were tested. Subsequent to testing, 2 samples were excluded from the data set due to failed daily QC leaving 122 samples included in the final analvsis.

The results from the method comparison studies comparing the Bordetella Direct Test to the Reference NAAT are summarized in Table 13. Table 13 shows the calculated PPA and the associated 95% confidence intervals for the results obtained in the Prospective (Fresh) and Frozen Retrospective Study (Frozen). The number of true-positive) and true-negative (conegative) samples is also included.

Specimenn% Agreement (95% CI)
PositiveNegative
Bordetella pertussisFresh91585.7%(65.4% - 95.0%)18/2199.6%(98.9% - 99.8%)890/894
Frozen12294.6%(86.1% - 98.3%)56/59100.0%(94.3% - 100.0%)63/63

Table 13. Comparison of Bordetella Direct Test to the Reference NAAT

In total, there were six (6) false negative and four (4) false positive results. Two (2) of the six (6) false negatives were also negative by a second FDA cleared NAAT and two (2) of the four (4) false positives were also positive by the second FDA cleared NAAT.

Conclusion:

In the Prospective Study, the results for 915 samples were analyzed comparing the Bordetella Direct Test to the reference NAAT. For all sites combined, the point estimate achieved for PPA was 85.7% with a 95% confidence interval of 65.4% - 95.0%. The point estimate for NPA was 99.6% with a 95% confidence interval of 98.9% - 99.8%.

In the Frozen Retrospective Study, the results for 122 samples were analyzed comparing the Bordetella Direct Test to the reference NAAT. The point estimate achieved for PPA was 94.6% with a 95% confidence interval of 86.1% - 98.3%. The point estimate for NPA was 100.0% with a 95% confidence interval of 94.3% - 100.0%.

Overall Conclusion .

The submitted information in this 510(k) pre-market notification is complete and supports substantial equivalence for the Great Basin Bordetella Direct Test.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.