(60 days)
No
The description focuses on signal conditioning, transmission, and display, with no mention of AI or ML algorithms for analysis or interpretation.
No
The device is described as receiving, transmitting, and reconditioning physiological signals for display, rather than directly treating a condition or performing therapy.
No
Explanation: The device is described as conditioning, transmitting, and reconditioning physiological signals for display and/or recording. It does not perform analysis or interpretation of these signals to reach a medical diagnosis.
No
The device description explicitly states that the FFR Link receives, transmits, and converts signals, and mentions both wireless transmission and cable connections for output, indicating it includes hardware components beyond just software.
Based on the provided information, the FFR Link is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: IVD devices are used to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- FFR Link Function: The FFR Link's intended use and device description clearly state that it conditions, transmits, and reconditions physiological signals (specifically blood pressure) from measuring devices (pressure guidewire and external pressure transducer). It is processing real-time physiological data from within the body, not analyzing samples taken from the body.
- No Specimen Analysis: There is no mention of the FFR Link interacting with or analyzing any biological specimens.
Therefore, the FFR Link falls under the category of a medical device that processes and transmits physiological signals, but it does not meet the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The FFR Link ™ is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLabTM POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable.
Product codes
DRG
Device Description
During a fractional flow reserve procedure, the FFR Link receives, and transmits aortic (Pa) and distal (Pd) blood pressure signals. On a continuous basis, the FFR Link receives the Pd input from a Boston Scientific pressure guidewire, and Pa input from an invasive blood pressure (IBP) transducer cable.
The FFR Link converts the Pd and Pa pressures to both a digital and an analog signal. The digital signal can be wirelessly transmitted to iLab Polaris Multi-Modality Guidance System. The analog signal for the Pa and Pd blood pressure can be connected by a cable for output to a hemodynamic display monitor.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
As no design changes were made to the currently marketed FFR Link device, no additional performance testing in required to establish substantial equivalence. Bench, animal, electrical safety and EMC testing data submitted in K151613 remain valid and applicable to the proposed version of the FFR Link device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2910 Radiofrequency physiological signal transmitter and receiver.
(a)
Identification. A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.(b)
Classification. Class II (performance standards).
0
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other and appear to be connected.
March 24, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Boston Scientific Corporation Dan Krause Regulatory Affairs Manager 47215 Lakeview Blvd. Fremont, California 94538
Re: K170204
Trade/Device Name: FFR Link, FFR Signal Processing Module Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: January 20, 2017 Received: January 23, 2017
Dear Dan Krause:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
1
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mude Yellen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K170204
Device Name FFR Link, FFR Signal Processing Module
Indications for Use (Describe)
The FFR Link ™ is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLabTM POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary
510(k) Summary Complying with 21 CFR 807.92
510(k) Summary
I. SUBMITTER
Boston Scientific Corporation 47215 Lakeview Blvd Fremont, CA. 94538 USA
Phone: 510-240-0048
Contact Person: Dan Krause Date Prepared: January 20, 2017
II. DEVICE
Name of Device: FFR Link, FFR Signal Processing Module
Common or Usual Name:
Radiofrequency physiological signal transmitter and receiver. FFR Link, FFR Signal Processing Module
Classification Name:
21 CFR 870.2910 - Radiofrequency physiological signal transmitter and receiver.
Regulatory Class: Class II
Product Code: DRG - transmitters and receivers, physiological signal, radiofrequency
III. PREDICATE DEVICE
iLab Polaris Multi-Modality Guidance System - K151613
IV. REFERENCE DEVICE
AO USB Receiver, Wi-Box, Wi-Box PSU Kit, Wi-Box Xpress Cable, PW USB Receiver - K111854
V. DEVICE DESCRIPTION
During a fractional flow reserve procedure, the FFR Link receives, and transmits aortic (Pa) and distal (Pd) blood pressure signals. On a continuous basis, the FFR Link receives the Pd input from a Boston Scientific pressure guidewire, and Pa input from an invasive blood pressure (IBP) transducer cable.
The FFR Link converts the Pd and Pa pressures to both a digital and an analog signal. The
4
digital signal can be wirelessly transmitted to iLab Polaris Multi-Modality Guidance System. The analog signal for the Pa and Pd blood pressure can be connected by a cable for output to a hemodynamic display monitor.
The figure below depicts a typical cath lab set up including:
- Hospital owned invasive blood pressure monitor (provides Pa in to FFR Link)
- Boston Scientific Pressure Guidewire (i.e. Comet Pressure Guidewire K151610)
- FFR Link Signal Processing Module (subject of this Traditional 510(k)) .
- Hospital owned Hemodynamic Monitor ●
- BSC iLab Polaris Multi-Modality Guidance System
Image /page/4/Figure/8 description: The image shows a diagram of a system for measuring and displaying hemodynamic parameters. The system includes an IBP (Invasive Blood Pressure) sensor, a Boston Scientific Pressure Guidewire, an FFR (Fractional Flow Reserve) Link, a Hemodynamic Display, and an iLab Polaris System. The IBP sensor measures aortic pressure (Pa in), which is sent to the FFR Link and then to the Hemodynamic Display as aortic pressure (Pa out). The Boston Scientific Pressure Guidewire measures distal pressure (Pd in), which is sent to the FFR Link and then to the Hemodynamic Display as distal pressure (Pd out). The FFR Link also sends Pa and Pd out to the iLab Polaris System via Bluetooth.
FFR Link Functional Block Diagram
VI. INDICATIONS FOR USE
The FFR Link is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab™ POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable.
VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The FFR Link remains identical to the currently marketed FFR Link device. The primary purpose of this traditional 510(k) is to update the FFR Link indication for use statement and clarify Directions for Use (DFU) to more precisely align with the specific function of the FFR Link device. The design, fundamental technology and manufacturing of the currently cleared FFR Link device remain unchanged. The proposed indication for use statement is aligned with the FDA identification terminology for radiofrequency physiological signal transmitters and receivers under 870.2910 and with indication for use statements of comparable competitive systems (Reference Device, St. Jude Wi-Box - K111854).
This submission also introduces Pd Out cables to accommodate customers who wish to capture
5
510(k) Summary
Pd out data on their hospital owned hemodynamic monitors in conjunction with or independent of iLab Polaris Multi-Modality Guidance System. No changes to FFR Link are required to support this capability as the analog Pd Out design was incorporated into the original FFR Link design.
VIII. PERFORMANCE DATA
As no design changes were made to the currently marketed FFR Link device, no additional performance testing in required to establish substantial equivalence. Bench, animal, electrical safety and EMC testing data submitted in K151613 remain valid and applicable to the proposed version of the FFR Link device.
IV. CONCLUSIONS
The design and technological characteristics of the proposed FFR Link device remain identical to the predicate FFR Link device cleared as a component of the iLab Polaris Multi-Modality Guidance System in K151613. This submission aims primarily to update the FFR Link Indications for Use to align more precisely with its specific function. In no way is this DFU update intended to expand the use of the FFR Link to applications other than fractional flow reserve diagnostic procedures. The target population and clinical use of the FFR signal (i.e., diagnostic procedures) between the predicate and proposed device remains identical. Thus, the proposed indication for use still falls within the broader system level intended use that was provided in the originally cleared FFR Link labeling.
Based on the fact that the design of the proposed FFR Link device remains unchanged and that the proposed indication for use falls within the originally cleared intended use of the iLab Polaris Multi-Modality Guidance System it is considered to be substantially equivalent to the FFR Link device cleared under K151613.